TrollByte Plus shall only be used as a holder for dental x-ray.
The manufacturer and its dealers are not liable
for any other use of the product.
TrollByte Plus is available in several sizes to fit different sensors, image plate or film.The user is responsible
for the selection of size and model.
Always use a sensor cover to protect the sensor while taking x-rays.
Clean prior to first use.
Procedures to clean the product:
• Spirits
• Dishwashing machine/disinfector
• Autoclave
Autoclave procedure
All included parts are autoclavable and shall be autoclaved separately.
Pre-programmed sterilization cycles according to the manufacturer of the autoclave. Max temperature
134ºC (273ºF). Place in the middle tray of the autoclave, away from autoclave walls and heating element..
Caution: If pressure or weight from other objects or instruments is applied during autoclaving, the holder may
be deformed. Minor deformations can easily be corrected by gently reshaping the holder.
Please note: Any method of sterilization will shorten the life span of plastic parts.
Warranty
TrollByte Plus is manufactured from high a grade polymer. Plastic parts have a limited life span and should be
replaced periodically.
Warranty Period for all TrollByte Plus sensor holders, under normal use, is limited to 12 months from delivery,
or 75 autoclave cycles, whichever occurs first .
Trollhätteplast AB
Installatörvägen 9
461 37 Trollhättan • Sweden 2011-02-16
0520-48 90 50 • www.trolldental.se
1510003/001
Basic alignment Page 4
Bite block and aiming ring Page 5
Maxillary Central Incisor
Maxillary Lateral Incisor
Maxillary Cuspid
Page 6
Mandibular Central Incisor
Mandibular Lateral Incisor
Page 7
Maxillary Molar
Page 8
Mandibular Molar
Page 9
Horizontal Bitewing
Page 10
Horizontal Bitewing
Page 11
Vertical Bitewing
Page 12
Basic alignment features.
X-ray tube
The tip of the small pin (a) aligns perfectly with the center of the
holder. It is also parallel to the claw (b).
The aiming pin (c) indicates
a 90 degree angle to the sensor.
So, the combination of the aiming pin and the small pin makes
it easy to determine both the vertical and horizontal position of
the sensor. This applies when the sensor is centered, as in bitewing
images. When capturing anterior and posterior periapical images, it
is advised to use an aiming ring.
4
YELLOW
TrollByte Yellow
For upper and lower
molar and premolars.
Can also be used for
horizontal bitewing.
BLUE
TrollByte Blue
For anterior periapicals and
vertical bitewings.
You can finely adjust the
RED tension.
The adjustable range is
TrollByte Red quite wide as you can see in
For horizontal bitewings. these pictures
Aiming ring
The color markings on the ring Mount the ring on the holder as shown in the
are linked to the color of the picture,the dots facing the sensor. You can slide the
holder. That makes it easy to put ring along the holder.
the ring in the correct position.
Bite block
You will need a bite block for all
anterior periapical images. The bite
block works as an extension of the
biteplane and allows you to place Place the bite block on the holder by sliding it
sensor as posterior as possible. over the biteplane and small alignment pin of
the holder.
5
Maxillary Central Incisor
BLUE Maxillary Lateral Incisor
Maxillary Cuspid
You will need a bite block for all anterior
periapical images. See page 5 for as- Position for the ring,
sembly instruction. dots facing the sensor.
To use the most of the sensor´s ef- The sensor should be placed as high Ready for use. The bite block al-
fective image area, move the sen- as possible as shown in this picture. lows you to position the sensor
sor further in an apical direction. The bottom edge of the sensor as posterior as possible.
aligns with the midline of the claw.
Put the sensor in place. The ring shows the position of Align and center the x-ray tube
the sensor. It also has guidance on the aiming ring and capture
markings for centering a square the image.
6 or a rectangular x-ray tube.
Mandibular Central Incisor
Mandibular Lateral Incisor BLUE
You will need a bite block for all anterior
periapical images. See page 5 for as- Position for the ring,
sembly instruction. dots facing the sensor.
To use the most of the sensor´s ef- The sensor should be placed as low Ready for use. The bite block al-
fective image area, move the sen- as possible as shown in this picture. lows you to position the sensor
sor further in an apical direction. The top edge of the sensor aligns as posterior as possible.
with the midline of the claw.
Put the sensor in place. The ring shows the position of Align and center the x-ray tube
the sensor. It also has guidance on the aiming ring and capture
markings for centering a square the image.
or a rectangular x-ray tube. 7
Maxillary Molar
YELLOW
Position for the ring,
dots facing the sensor.
To use the most of the sensor´s ef- The sensor should be placed as high Ready for use.
fective image area, move the sen- as possible as shown in this picture.
sor further in an apical direction. The bottom edge of the sensor
aligns with the midline of the claw.
Put the sensor in place. The ring shows the position of Align and center the x-ray tube
the sensor. It also has guidance on the aiming ring and capture
markings for centering a square the image.
8 or a rectangular x-ray tube.
Mandibular Molar
YELLOW
Position for the ring,
dots facing the sensor.
To use the most of the sensor´s ef- The sensor should be placed as low Ready for use.
fective image area, move the sen- as possible as shown in this picture.
sor further in an apical direction. The top edge of the sensor aligns
with the midline of the claw.
Put the sensor in place. The ring shows the position of Align and center the x-ray tube
the sensor. It also has guidance on the aiming ring and capture
markings for centering a square the image.
or a rectangular x-ray tube. 9
Horizontal Bitewing
YELLOW
Position for the ring,
dots facing the sensor.
Center position. Position the sensor in the middle. Ready for use.
Put the sensor in place. The ring shows the position of Align and center the x-ray tube
the sensor. It also has guidance on the aiming ring and capture
markings for centering a square the image.
10 or a rectangular x-ray tube.
Horizontal Bitewing
RED
Position for the ring,
dots facing the sensor.
Center position. Position the sensor in the middle. Ready for use.
Put the sensor in place. The ring shows the position of Align and center the x-ray tube
the sensor. It also has guidance on the aiming ring and capture
markings for centering a square the image.
or a rectangular x-ray tube. 11
Vertical Bitewing
BLUE
Position for the ring,
dots facing the sensor.
Center position. Position the sensor in the Ready for use.
middle with the cord down.
Put the sensor in place. The ring shows the position of Align and center the x-ray tube
the sensor. It also has guidance on the aiming ring and capture
markings for centering a square the image.
12 or a rectangular x-ray tube.