Dropsens Stat400
Dropsens Stat400
STAT400
μStat 400 is the portable BiPotentiostat/Galvanostat from DropSens that can be applied for Voltammetric,
Amperometric or Potentiometric measurements, including 18 electroanalytical techniques, and can be used
with one- or two- working electrodes configuration.
The new portable bipotentiostat/galvanostat is Li-ion Battery powered (USB charger adapter compatible). It can
be easily connected to a PC via USB, RS232 and Bluetooth®.
μStat 400 has eight current ranges: 1 nA to 10 mA, and Auto (the instrument automatically selects the optimal
current range), with a maximum measurable current of 40 mA.
The supplied DropView 8400 software for Windows is used to control the instrument and to plot the
measurements and perform the analysis of results. DropView 8400 software provides powerful functions such
as:
Available techniques:
POTENTIOSTAT GALVANOSTAT
Voltammetry LSP Linear Sweep Potentiometry
LSV Linear Sweep Voltammetry CP Cyclic Potentiometry
CV Cyclic Voltammetry PD Potentiometric Detection (galvanostatic)
SWV Square Wave Voltammetry FP Fast Potentiometry (tint < 0.1s)
DPV Differential Pulse Voltammetry ZCP Zero Current Potentiometry
NPV Normal Pulse Voltammetry PSAG Potentiometric Stripping Analysis (galvanostatic)
NDPV Differential Normal Pulse Voltammetry PSAF Potentiometric Stripping Analysis (faradaic)
ACV AC Voltammetry
Amperometry
AD Amperometric Detection
FA Fast Amperometry (tint < 0.1 s)
PAD Pulsed Amperometric Detection
ZRA Zero Resistance Amperometry
Instrument Specifications
Control Specifications
General Conditioning stage duration: 0 – 1300 s
Pretreatment Deposition stage duration: 0 – 1300 s
Equilibration stage duration: 0 – 1300 s
General Begin, End, Base, Vertex potentials: -4.096 V to +4.096 V
Parameters Step potential: 1 mV to 500 mV
Pulse potential: 1 mV to 250 mV
Scan rate: 1 ms up to 1.3 s per step
Specific SWV Frequency: 1 Hz to 400 Hz
Parameters Amplitude: 1 mV to 250 mV
DPV, NPV, NDP Modulation time: 1 ms to 1300 ms
Pulse time: 1 ms to 1300 ms
ACV Frequency: 2 Hz to 250 Hz
Amplitude: 5 mV to 250 mV (RMS)
Chrono. Methods (AD, PD, ZCP, ZRA) Interval time: 0.1 s to 1300 s
Run time: Hours (65000 points)
Fast Chrono. Methods (FA, FP) Interval time: 1 ms to 1300 ms
Run time: Hours (65000 points)
PAD Pulse time: 1 ms to 1300 ms
Interval time: 10 ms to 1300 ms
Run time: Hours (65000 points)
PSA Potential limit: ±2.048 V