Notes Unit-4 Hve
Notes Unit-4 Hve
4 and Currents
4.1 INTRODUCTION
Transient measurements have much in common with measurements of steady state quantities but the
short-lived nature of the transients which we are trying to record introduces special problems. Frequently
the transient quantity to be measured is not recorded directly because of its large magnitudes e.g. when
a shunt is used to measure current, we really measure the voltage across the shunt and then we assume
that the voltage is proportional to the current, a fact which should not be taken for granted with transient
currents. Often the voltage appearing across the shunt may be insufficient to drive the measuring device;
it requires amplification. On the other hand, if the voltage to be measured is too large to be measured
with the usual meters, it must be attenuated. This suggests an idea of a measuring system rather than a
measuring device.
Measurements of high voltages and currents involves much more complex problems which a
specialist, in common electrical measurement, does not have to face. The high voltage equipments
have large stray capacitances with respect to the grounded structures and hence large voltage gradients
are set up. A person handling these equipments and the measuring devices must be protected against
these over voltages. For this, large structures are required to control the electrical fields and to avoid
flash over between the equipment and the grounded structures. Sometimes, these structures are re-
quired to control heat dissipation within the circuits. Therefore, the location and layout of the equipments
is very important to avoid these problems. Electromagnetic fields create problems in the measurements
of impulse voltages and currents and should be minimised.
The chapter is devoted to describing various devices and circuits for measurement of high voltages
and currents. The application of the device to the type of voltages and currents is also discussed.
110
MEASUREMENT OF HIGH VOLTAGES AND CURRENTS 111
of the spheres are met, the results obtained by the use of sphere gaps are reliable to within ±3%. It has
been suggested in standard specification that in places where the availability of ultraviolet radiation is
low, irradiation of the gap by radioactive or other ionizing media should be used when voltages of
magnitude less than 50 kV are being measured or where higher voltages with accurate results are to be
obtained.
In order to understand the importance of irradiation of sphere gap for measurement of impulse
voltages especially which are of short duration, it is necessary to understand the time-lag involved in
the development of spark process. This time lag consists of two components—(i) The statistical time-
lag caused by the need of an electron to appear in the gap during the application of the voltage. (ii) The
formative time lag which is the time required for the breakdown to develop once initiated.
The statistical time-lag depends on the irradiation level of the gap. If the gap is sufficiently
irradiated so that an electron exists in the gap to initiate the spark process and if the gap is subjected to
an impulse voltage, the breakdown will take place when the peak voltage exceeds the d.c. breakdown
value. However, if the irradiation level is low, the voltage must be maintained above the d.c. break-
down value for a longer period before an electron appears. Various methods have been used for irradia-
tion e.g. radioactive material, ultraviolet illumination as supplied by mercury arc lamp and corona
discharges.
It has been observed that large variation can occur in the statistical time-lag characteristic of a
gap when illuminated by a specified light source, unless the cathode conditions are also precisely
specified.
Irradiation by radioactive materials has the advantage in that they can form a stable source of
irradiation and that they produce an amount of ionisation in the gap which is largely independent of the
gap voltage and of the surface conditions of the electrode. The radioactive material may be placed
inside high voltage electrode close behind the sparking surface or the radioactive material may form
the sparking surface.
The influence of the light from the impulse generator spark gap on the operation of the sphere
gaps has been studied. Here the illumination is intense and occurs at the exact instant when it is re-
quired, namely, at the instant of application of the voltage wave to the sphere gap.
The formative time lag depends mainly upon the mechanism of spark growth. In case of second-
ary electron emission, it is the transit time taken by the positive ion to travel from anode to cathode that
decides that formative time lag. The formative time-lag decreases with the applied over voltage and
increase with gap length and field non-uniformity.
4
B
0·5D 0·2D
0·2D 0·5D
3
3
2
2D S 2D
2D 2D
A
Fig. 4.1
4
0·5D
5
0·20 2
P B
0·2D A
0·5D 3
1·5D
Fig. 4.2
MEASUREMENT OF HIGH VOLTAGES AND CURRENTS 113
According to BSS 358: 1939, when one sphere is grounded, the distance from the sparking
point of the high voltage sphere to the equivalent earth plane to which the earthed sphere is connected
should lie within the limits as given in Table 4.1.
Table 4.1
Height of sparking point of high voltage sphere above the equivalent earth plane.
S = Sparking point distance
Upto 25 cms. 7D 10 S 7D 5D
50 cms. 6D 8S 6D 4D
75 cms. 6D 8S 6D 4D
100 cms. 5D 7S 5D 3.5 D
150 cms. 4D 6S 4D 3D
200 cms. 4D 6S 4D 3D
In order to avoid corona discharge, the shanks supporting the spheres should be free from sharp
edges and corners. The distance of the sparking point from any conducting surface except the shanks
should be greater than
V
25 cms
3
where V is the peak voltage is kV to be measured. When large spheres are used for the measurement of
low voltages the limiting distance should not be less than a sphere diameter.
It has been observed that the metal of which the spheres are made does not affect the accuracy of
measurements MSS 358: 1939 states that the spheres may be made of brass, bronze, steel, copper,
aluminium or light alloys. The only requirement is that the surfaces of these spheres should be clean,
free from grease films, dust or deposited moisture. Also, the gap between the spheres should be kept
free from floating dust particles, fibres etc.
For power frequency tests, a protective resistance with a value of 1 /V should be connected in
between the spheres and the test equipment to limit the discharge current and to prevent high frequency
oscillations in the circuit which may otherwise result in excessive pitting of the spheres. For higher
frequencies, the voltage drop would increase and it is necessary to have a smaller value of the resistance.
For impulse voltage the protective resistors are not required. If the conditions of the spheres and its
associated accessories as given above are satisfied, the spheres will spark at a peak voltage which will
be close to the nominal value shown in Table 4.2. These calibration values relate to a temperature of
20°C and pressure of 760 mm Hg. For a.c. and impulse voltages, the tables are considered to be accurate
within 3% for gap lengths upto 0.5 D. The tables are not valid for gap lengths less than 0.05 D and
impulse voltages less than 10 kV. If the gap length is greater than 0.5 D, the results are less accurate and
are shown in brackets.
114 HIGH VOLTAGE ENGINEERING
Table 4.2
Sphere gap with one sphere earthed
Peak value of disruptive discharge voltages (50% for impulse tests) are valid for (i) alternating voltages
(ii) d.c. voltage of either polarity (iii) negative lightning and switching impulse voltages
Due to dust and fibre present in the air, the measurement of d.c. voltages is generally subject to
larger errors. Here the accuracy is within 5% provided the spacing is less than 0.4 D and excessive
dust is not present.
The procedure for high voltage measurement using sphere gaps depends upon the type of voltage
to be measured.
Table 4.3
Sphere Gap with one sphere grounded
Peak values of disruptive discharge voltages (50% values).
Positive lightning and switching impulse voltages
Peak Voltage kV
Sphere Gap Sphere dia in cms
For the measurement of a.c. or d.c. voltage, a reduced voltage is applied to begin with so that the
switching transient does not flash over the sphere gap and then the voltage is increased gradually till the
gap breaks down. Alternatively the voltage is applied across a relatively large gap and the spacing is
116 HIGH VOLTAGE ENGINEERING
then gradually decreased till the gap breaks down. Corresponding to this gap the value of peak voltage
can be read out from the calibration tables. However, it is reminded that the calibration tables values
correspond to 760 mm Hg pressure and 20°C temperature. Any deviation from the value, a correction
factor will have to be used to get the correct value of the voltage being measured.
For the measurement of 50% impulse disruptive discharge voltages, the spacing of the sphere
gap or the charging voltage of the impulse generator is adjusted in steps of 3% of the expected disruptive
voltage. Six applications of the impulse should be made at each step and the interval between two
applications is 5 seconds. The value giving 50% probability to disruptive discharge is preferably obtained
by interpolation between at least two gap or voltage settings, one resulting in two disruptive discharges
or less out of six applications and the other in four disruptive discharges or more out of again six
applications.
Another method, simple though less accurate, is to adjust the settings such that four to six
disruptive discharges are obtained in a series of ten successive applications.
The breakdown voltage of a sphere gap increases with increase in pressure and decreases with
increase in temperature. The value of disruptive voltages as given in Tables 4.2 and 4.3 correspond to
760 mm Hg pressure and 20°C. For small variation in temperatures and pressures, the disruptive voltage
is closely proportional to the relative air density. The relative air density is given by
293b
=
760 ( 273 t )
where b and t are the atmospheric conditions (pressure in mm of Hg and temperature in °C respectively)
during measurement. The disruptive voltage V is given V = KdV0
Where V0 is the disruptive voltage as given in the Tables 4.2 and 4.3 and Kd is a correction factor
given in Table 4.4. Kd is a slightly non-linear function of a result explained by Paschen's law.
Table 4.4
Air density correction factor Kd
Some of the other factors which influence the breakdown value of air are discussed here.
V = m ln
MEASUREMENT OF HIGH VOLTAGES AND CURRENTS 117
650
S/D = 0.6
600
550
500
S/D = 0.4
450
400
0
0 2 4 6 8 10 12
A/D (diameter)
Where V = per cent reduction in voltage in the breakdown voltage from the value when the
clearance was 12.6 D, and m and C are the factors dependent on the ratio S/D.
Fiegel and Keen have studied the influence of nearby ground plane on impulse breakdown
voltage of a 50 cm diameter sphere gap using 1.5/40 micro sec. negative polarity impulse wave. Fig. 4.3
shows the breakdown voltage as a function of A/D for various values of S/D. The voltage values were
corrected for relative air density.
It is observed that the voltage increases with increase in the ratio A/D. The results have been
compared with those given in Table 4.2 and represented in Fig. 4.3 by dashed lines. The results also
agree with the recommendation regarding the minimum and maximum values of A/D as given in Table 4.1.
Influence of Humidity
Kuffel has studied the effect of the humidity on the breakdown voltage by using spheres of 2 cms to
25 cms diameters and uniform field electrodes. The effect was found to be maximum in the region 0.4
mm Hg. and thereafter the change was decreased. Between 4–17 mm Hg. the relation between breakdown
voltage and humidity was practically linear for spacing less than that which gave the maximum humidity
effect. Fig. 4.4 shows the effect of humidity on the breakdown voltage of a 25 cm diameter sphere with
spacing of 1 cm when a.c. and d.c voltages are applied. It can be seen that
(i) The a.c. breakdown voltage is slightly less than d.c. voltage.
(ii) The breakdown voltage increases with the partial pressure of water vapour.
It has also been observed that
(i) The humidity effect increases with the size of spheres and is largest for uniform field elec-
trodes.
(ii) The voltage change for a given humidity change increase with gap length.
118 HIGH VOLTAGE ENGINEERING
Fig. 4.4 Breakdown voltage humidity relation for a.c. and d.c.
for 1.0 cm gap between 25 cms diameter spheres
The increase in breakdown voltage with increase in partial pressure of water vapour and this
increase in voltage with increase in gap length is due to the relative values of ionisation and attachment
coefficients in air. The water particles readily attach free electrons, forming negative ions. These ions
therefore slow down and are unable to ionise neutral molecules under field conditions in which electrons
will readily ionise. It has been observed that within the humidity range of 4 to 17 g/m3 (relative
humidity of 25 to 95% for 20°C temperature) the relative increase of breakdown voltage is found to be
between 0.2 to 0.35% per gm/m3 for the largest sphere of diameter 100 cms and gap length upto
50 cms.
improvement as compared with the equivalent sphere gap arrangement. Fig. 4.5 shows a half-contour
of one electrode having plane sparking surfaces with edges of gradually increasing curvature.
The portion AB is flat, the total diameter of the flat portion being greater than the maximum
spacing between the electrodes. The portion BC consists of a sine curve based on the axes OB and OC
and given by XY = CO sin (BX/BO . /2). CD is an arc of a circle with centre at O.
Bruce showed that the breakdown voltage V of a gap of length S cms in air at 20°C and 760 mm
Hg. pressure is within 0.2 per cent of the value given by the empirical relation.
V = 24.22S + 6.08 S
This equation, therefore, replaces Tables 4.2 and 4.3 which are necessary for sphere gaps. This
is a great advantage, that is, if the spacing between the spheres for breakdown is known the breakdown
voltage can be calculated.
The other advantages of uniform field spark gaps are
(i) No influence of nearby earthed objects
(ii) No polarity effect.
However, the disadvantages are
(i) Very accurate mechanical finish of the electrode is required.
(ii) Careful parallel alignment of the two electrodes.
(iii) Influence of dust brings in erratic breakdown of the gap. This is much more serious in these
gaps as compared with sphere gaps as the highly stressed electrode areas become much
larger.
Therefore, a uniform field gap is normally not used for voltage measurements.
2 26 80 435
4 47 90 488
6 62 100 537
8 72 120 642
10 81 140 744
15 102 160 847
20 124 180 950
25 147 200 1054
30 172 220 1160
35 198
40 225
50 278
60 332
70 382
The breakdown voltage is a rod gap increases more or less linearly with increasing relative air
density over the normal variations in atmospheric pressure. Also, the breakdown voltage increases with
increasing relative humidity, the standard humidity being taken as 15.5 mm Hg.
Because of the large variation in breakdown voltage for the same spacing and the uncertainties
associated with the influence of humidity, rod gaps are no longer used for measurement of a.c. or
impulse voltages. However, more recent investigations have shown that these rods can be used for d.c.
measurement provided certain regulations regarding the electrode configurations are observed. The
arrangement consists of two hemispherically capped rods of about 20 mm diameter as shown in Fig. 4.6.
The earthed electrode must be long enough to initiate positive breakdown streamers if the high
voltage rod is the cathode. With this arrangement, the breakdown voltage will always be initiated by
positive streamers for both the polarities thus giving a very small variation and being humidity dependent.
Except for low voltages (less than 120 kV), where the accuracy is low, the breakdown voltage can be
given by the empirical relation.
MEASUREMENT OF HIGH VOLTAGES AND CURRENTS 121
V= (A + BS) 4 . 10 –2 (h
51 8.65) kV
where h is the absolute humidity in gm/m3
and varies between 4 and 20 gm/m3 in the above relation.
The breakdown voltage is linearly related with the gap spacing and the slope of the relation
B = 5.1 kV/cm and is found to be independent of the polarity of voltage. However constant A is polarity
dependent and has the values
A = 20 kV for positive polarity
= 15 kV for negative polarity of the high voltage electrode.
The accuracy of the above relation is better than 20% and, therefore, provides better accuracy
even as compared to a sphere gap.
1 T 1 1 V 2 (t ) 1 A 1 1 V2
F= F(t )dt A dt . V 2 ( t )dt A rms
T 0 T 2 d2 2 d2 T 2 d2
122 HIGH VOLTAGE ENGINEERING
Electrostatic voltmeters measure the force based on the above equations and are arranged such
that one of the plates is rigidly fixed whereas the other is allowed to move. With this the electric field
gets disturbed. For this reason, the movable electrode is allowed to move by not more than a fraction of
a millimetre to a few millimetres even for high voltages so that the change in electric field is negligibly
small. As the force is proportional to square of Vrms, the meter can be used both for a.c. and d.c. voltage
measurement.
The force developed between the plates is sufficient to be used to measure the voltage. Various
designs of the voltmeter have been developed which differ in the construction of electrode arrangement
and in the use of different methods of restoring forces required to balance the electrostatic force of
attraction. Some of the methods are
(i) Suspension of moving electrode on one arm of a balance.
(ii) Suspension of the moving electrode on a spring.
(iii) Pendulous suspension of the moving electrode.
(iv) Torsional suspension of moving electrode.
The small movement is generally transmitted and amplified by electrical or optical methods. If
the electrode movement is minimised and the field distribution can exactly be calculated, the meter can
be used for absolute voltage measurement as the calibration can be made in terms of the fundamental
quantities of length and force.
From the expression for the force, it is clear that for a given voltage to be measured, the higher
the force, the greater is the precision that can be obtained with the meter. In order to achieve higher
force for a given voltage, the area of the plates should be large, the spacing between the plates (d)
should be small and some dielectric medium other than air should be used in between the plates. If
uniformity of electric field is to be maintained an increase in area A must be accompanied by an increase
in the area of the surrounding guard ring and of the opposing plate and the electrode may, therefore,
become unduly large specially for higher voltages. Similarly the gap length cannot be made very small
as this is limited by the breakdown strength of the dielectric medium between the plates. If air is used as
the medium, gradients upto 5 kV/cm have been found satisfactory. For higher gradients vacuum or SF6
gas has been used.
The greatest advantage of the electrostatic voltmeter is its extremely low loading effect as only
electric fields are required to be set up. Because of high resistance of the medium between the plates,
the active power loss is negligibly small. The voltage source loading is, therefore, limited only to the
reactive power required to charge the instrument capacitance which can be as low as a few picofarads
for low voltage voltmeters.
The measuring system as such does not put any upper limit on the frequency of supply to be
measured. However, as the load inductance and the measuring system capacitance form a series resonance
circuit, a limit is imposed on the frequency range. For low range voltmeters, the upper frequency is
generally limited to a few MHz.
Fig. 4.7 shows a schematic diagram of an absolute electrostatic voltmeter. The hemispherical
metal dome D encloses a sensitive balance B which measures the force of attraction between the movable
disc which hangs from one of its arms and the lower plate P. The movable electrode M hangs with a
clearance of above 0.01 cm, in a central opening in the upper plate which serves as a guard ring. The
MEASUREMENT OF HIGH VOLTAGES AND CURRENTS 123
diameter of each of the plates is 1 metre. Light reflected from a mirror carried by the balance beam
serves to magnify its motion and to indicate to the operator at a safe distance when a condition of
equilibrium is reached. As the spacing between the two electrodes is large (about 100 cms for a voltage
of about 300 kV), the uniformity of the electric field is maintained by the guard rings G which surround
the space between the discs M and P. The guard rings G are maintained at a constant potential in space
by a capacitance divider ensuring a uniform spatial potential distribution. When voltages in the range
10 to 100 kV are measured, the accuracy is of the order of 0.01 per cent.
Hueter has used a pair of sphares of 100 cms diameter for the measurement of high voltages
utilising the electrostatic attractive force between them. The spheres are arranged with a vertical axis
and at a spacing slightly greater than the sparking distance for the particular voltage to be measured.
The upper high voltage sphere is supported on a spring and the extension of spring caused by the
electrostatic force is magnified by a lamp-mirror scale arrangement. An accuracy of 0.5 per cent has
been achieved by the arrangement.
Electrostatic voltmeters using compressed gas as the insulating medium have been developed.
Here for a given voltage the shorter gap length enables the required uniformity of the field to be
maintained with electrodes of smaller size and a more compact system can be evolved.
One such voltmeter using SF6 gas has been used which can measure voltages upto 1000 kV and
accuracy is of the order of 0.1%. The high voltage electrode and earthed plane provide uniform electric
field within the region of a 5 cm diameter disc set in a 65 cm diameter guard plane. A weighing balance
124 HIGH VOLTAGE ENGINEERING
arrangement is used to allow a large damping mass. The gap length can be varied between 2.5, 5 and
10 cms and due to maximum working electric stress of 100 kV/cm, the voltage ranges can be selected
to 250 kV, 500 kV and 100 kV. With 100 kV/cm as gradient, the average force on the disc is found to be
0.8681 N equivalent to 88.52 gm wt. The disc movements are kept as small as 1 m by the weighing
balance arrangement.
The voltmeters are used for the measurement of high a.c. and d.c. voltages. The measurement of
voltages lower than about 50 volt is, however, not possible, as the forces become too small.
dq(t ) d
i (t) = ( a)da
dt dt
Where (a) is the electric field density or charge density along some path and is assumed constant over
the differential area da of the pick up electrode. In this case (a) is a function of time also and da the
area of the pick up electrode P exposed to the electric field.
However, if the voltage V to be measured is constant (d.c voltage), a current i(t) will flow only
if it is moved i.e. now (a) will not be function of time but the charge q is changing because the area of
the pick up electrode exposed to the electric field is changing. The current i(t) is given by
MEASUREMENT OF HIGH VOLTAGES AND CURRENTS 125
d d
i(t) = ( a)da E (a)da
dt A(t ) dt A(t )
where (a) = E(a) and is the permittivity of the medium between the high voltage electrode and the
grounded electrode. The integral boundary denotes the time varying exposed area.
The high voltage electrode and the grounded electrode in fact constitute a capacitance system.
The capacitance is, however, a function of time as the area A varies with time and, therefore, the charge
q(t) is given as
q(t) = C(t)V(t)
dq dV (t ) dC(t )
and i(t) = C( t ) V (t )
dt dt dt
For d.c. voltages dV (t )
0
dt
dC(t )
Hence i(t) = V
dt
If the capacitance varies linearly with time and reaches its peak value Cm is time Tc /2 and again
reduces to zero linearly in time Tc /2, the capacitance is given as
Cm
C(t) = 2 t
Tc
For a constant speed of n rpm of synchronous motor which is varying the capacitance, time Tc is
given by Tc = 60/n.
n n
Therefore I = 2Cm V C V
60 30 m
If the capacitance C varies sinusoidally between the limits C0 and (C0 + Cm) then
C = C0 + Cm sin wt
and the current i is then given as
i(t) = im cos wt where im = VCm
Here is the angular frequency of variation
of the capacitance. If is constant, the current
measured is proportional to the voltage being
measured. Generally the current is rectified and
measured by a moving coil meter. Generating
voltmeters can be used for a.c. voltage
measurement also provided the angular frequency
is the same or equal to half that of the voltage
being measured. Fig 4.9 shows the variations of
C as a function of time together with a.c. voltage,
Fig. 4.9 Capacitance and voltage variation
the frequency of which is twice the frequency of
C (t).
It can be seen from Fig. 4.9 that whatever be the phase relation between voltage and the
capacitance, over one cycle variation of the voltage is same (e.g. V(t1) – V(t2)) and the rate of change of
126 HIGH VOLTAGE ENGINEERING
capacitance over the period Tv is equal to Cm/Tv. Therefore, the instantaneous value of current i(t) = Cm
fvV(t) where fv = 1/Tv the frequency of voltage.
Since fv = 2fc and fc = 60/n we obtain
I(t) = n/30 CmV(t)
Fig. 4.10 shows a schematic diagram of a generating voltmeter which employs rotating vanes
for variation of capacitance. The high voltage electrode is connected to a disc electrode D3 which is
kept at a fixed distance on the axis of the other low voltage electrodes D2, D1, and D0. The rotor D0 is
driven at a constant speed by a synchronous motor at a suitable speed. The rotor vanes of D0 cause
periodic change in capacitance between the insulated disc D2 and the high voltage electrode D3. The
number and shape of vanes are so designed that a suitable variation of capacitance (sinusodial or linear)
is achieved. The a.c. current is rectified and is measured using moving coil meters. If the current is
small an amplifier may be used before the current is measured.
Generating voltmeters are linear scale instruments and applicable over a wide range of voltages.
The sensitivity can be increased by increasing the area of the pick up electrode and by using amplifier
circuits.
The main advantages of generating voltmeters are (i) scale is linear and can be extrapolated
(ii) source loading is practically zero (iii) no direct connection to the high voltage electrode.
However, they require calibration and construction is quite cumbersome.
The breakdown of insulating materials depends upon the magnitude of voltage applied and the
time of application of voltage. However, if the peak value of voltage is large as compared to breakdown
strength of the insulating material, the disruptive discharge phenomenon is in general caused by the
instantaneous maximum field gradient stressing the material. Various methods discussed so far can
measure peak voltages but because of complex calibration procedures and limited accuracy call for
MEASUREMENT OF HIGH VOLTAGES AND CURRENTS 127
more convenient and more accurate methods. A more convenient though less accurate method would
be the use of a testing transformer wherein the output voltage is measured and recorded and the input
voltage is obtained by multiplying the output voltage by the transformation ratio. However, here the
output voltage depends upon the loading of the secondary winding and wave shape variation is caused
by the transformer impedances and hence this method is unacceptable for peak voltage measurements.
v (t) C
C
ic (t) Rd
D1 D2
D1 D2
A A
( ) ( )
The displacement current ic(t), Fig. 4.12 is given by the rate of change of the charge and hence
the voltage V(t) to be measured flows through the high voltage capacitor C and is subdivided into
positive and negative components by the back to back connected diodes. The voltage drop across these
diodes can be neglected (1 V for Si diodes) as compared with the voltage to be measured. The measuring
instrument (M.C. ammeter) is included in one of the branches. The ammeter reads the mean value of
the current.
1 t2 dv(t ) C I
I= C . dt . 2Vm 2Vm fC or Vm
T t1 dt T 2 fC
The relation is similar to the one obtained in case of generating voltmeters. An increased current
would be obtained if the current reaches zero more than once during one half cycle. This means the
wave shapes of the voltage would contain more than one maxima per half cycle. The standard a.c.
voltages for testing should not contain any harmonics and, therefore, there could be very short and
rapid voltages caused by the heavy predischarges, within the test circuit which could introduce errors in
measurements. To eliminate this problem filtering of a.c. voltage is carried out by introducing a damping
resistor in between the capacitor and the diode circuit, Fig. 4.11 (b).
128 HIGH VOLTAGE ENGINEERING
Fig. 4.12
Also, if full wave rectifier is used instead of the half wave as shown in Fig. 4.11, the factor 2 in
the denominator of the above equation should be replaced by 4. Since the frequency f, the capacitance
C and current I can be measured accurately, the measurement of symmetrical a.c. voltages using Chubb
and Fortescue method is quite accurate and it can be used for calibration of other peak voltage measuring
devices.
Fig. 4.13 shows a digital peak voltage measuring circuit. In contrast to the method discussed just
now, the rectified current is not measured directly, instead a proportional analog voltage signal is derived
which is then converted into a proportional medium frequency for using a voltage to frequency convertor
(Block A in Fig. 4.13). The frequency ratio fm/f is measured with a gate circuit controlled by the a.c.
power frequency (supply frequency f) and a counter that opens for an adjustable number of period
t = p/f. The number of cycles n counted during this interval is
p
n = tfm = fm
f
where p is a constant of the instrument.
fm fm fm 1
Let A= .
Ric R2 Vm fC f 2 RVm C
Therefore, n = p 2ARVmC
where A represents the voltage to frequency conversion factor.
MEASUREMENT OF HIGH VOLTAGES AND CURRENTS 129
Thus the indicator can be calibrated to read Vm directly by selecting suitable values of A, p and
R.
The voltmeter is found to given an accuracy of 0.35%.
for 150 Hz. The third source of error is related to this discharge error. During the conduction time
(when the voltage across Cs is lower than that across C2 because of discharge of Cs through Rd) of the
diode the storage capacitor Cs is recharged to the peak value and thus Cs becomes parallel with C2. If
discharge error is ed, recharge error er is given by
Cs
er 2ed
C1 C2 Cs
Hence Cs should be small as compared with
C2 to keep down the recharge error.
It has also been observed that in order to
keep the overall error to a low value, it is desirable
to have a high value of R2. The same effect can be
obtained by providing an equalising arm to the low
voltage arm of the voltage divider as shown in
Fig. 4.15 Modified peak voltmeter circuit
Fig. 4.15. This is accomplished by the addition of
a second network comprising diode, Cs and Rd for negative polarity currents to the circuit shown in Fig.
4.14. With this, the d.c. currents in both branches are opposite in polarity and equalise each other. The
errors due to R2 are thus eliminated.
Rabus developed another circuit shown in Fig. 4.16. to reduce errors due to resistances. Two
storage capacitors are connected by a resistor Rs within every branch and both are discharged by only
one resistance Rd.
D2 D2 D1 D1
Rs
Here because of the presence of Rs, the discharge of the storage capacitor Cs2 is delayed and
hence the inherent discharge error ed is reduced. However, since these are two storage capacitors within
one branch, they would draw more charge from the capacitor C2 and hence the recharge error er would
increase. It is, therefore, a matter of designing various elements in the circuit so that the total sum of all
the errors is a minimum. It has been observed that with the commonly used circuit elements in the
voltage dividers, the error can be kept to well within about 1% even for frequencies below 20 Hz.
The capacitor C1 has to withstand high voltage to be measured and is always placed within the
test area whereas the low voltage arm C2 including the peak circuit and instrument form a measuring
unit located in the control area. Hence a coaxial cable is always required to connect the two areas. The
cable capacitance comes parallel with the capacitance C2 which is usually changed in steps if the
MEASUREMENT OF HIGH VOLTAGES AND CURRENTS 131
voltage to be measured is changed. A change of the length of the cable would, thus, also require
recalibration of the system. The sheath of the coaxial cable picks up the electrostatic fields and thus
prevents the penetration of this field to the core of the conductor. Also, even though transient magnetic
fields will penetrate into the core of the cable, no appreciable voltage (extraneous of noise) is induced
due to the symmetrical arrangement and hence a coaxial cable provides a good connection between the
two areas. Whenever, a discharge takes place at the high voltage end of capacitor C1 to the cable
connection where the current looks into a change in impedance a high voltage of short duration may be
built up at the low voltage end of the capacitor C1 which must be limited by using an over voltage
protection device (protection gap). These devices will also prevent complete damage of the measuring
circuit if the insulation of C1 fails.
These three elements form a voltage generating system. The lead L consists of a lead wire and
a resistance to damp oscillation or to limit short-circuit currents if of the test object fails. The measuring
system starts at the terminals of the test object and consists of a connecting lead CL to the voltage
divider D. The output of the divider is fed to the measuring instrument (CRO etc.) M. The appropriate
ground return should assure low voltage drops for even highly transient phenomena and keep the ground
potential of zero as far as possible.
It is to be noted that the test object is a predominantly capacitive element and thus this forms an
oscillatory circuit with the inductance of the load. These oscillations are likely to be excited by any
steep voltage rise from the generator output, but will only partly be detected by the voltage divider. A
resistor in series with the connecting leads damps out these oscillations. The voltage divider should
always be connected outside the generator circuit towards the load circuit (Test object) for accurate
measurement. In case it is connected within the generator circuit, and the test object discharges (chopped
wave) the whole generator including voltage divider will be discharged by this short circuit at the test
object and thus the voltage divider is loaded by the voltage drop across the lead L. As a result, the
voltage measurement will be wrong.
Yet for another reason, the voltage divider should be located away from the generator circuit.
The dividers cannot be shielded against external fields. All objects in the vicinity of the divider which
132 HIGH VOLTAGE ENGINEERING
may acquire transient potentials during a test will disturb the field distribution and thus the divider
performance. Therefore, the connecting lead CL is an integral part of the potential divider circuit.
In order to avoid electromagnetic interference between the measuring instrument M and C the
high voltage test area, the length of the delay cable should be adequately chosen. Very short length of
the cable can be used only if the measuring instrument has high level of electromagnetic compatibility
(EMC). For any type of voltage to be measured, the cable should be co-axial type. The outer conductor
provides a shield against the electrostatic field and thus prevents the penetration of this field to the
inner conductor. Even though, the transient magnetic fields will penetrate into the cable, no appreciable
voltage is induced due to the symmetrical arrangement. Ordinary coaxial cables with braided shields
may well be used for d.c. and a.c. voltages. However, for impulse voltage measurement double shielded
cables with predominently two insulated braided shields will be used for better accuracy.
During disruption of test object, very heavy transient current flow and hence the potential of the
ground may rise to dangerously high values if proper earthing is not provided. For this, large metal
sheets of highly conducting material such as copper or aluminium are used. Most of the modern high
voltage laboratories provide such ground return along with a Faraday Cage for a complete shielding of
the laboratory. Expanded metal sheets give similar performance. At least metal tapes of large width
should be used to reduce the impedance.
The resistance potential dividers are the first to appear because of their simplicity of construction, less
space requirements, less weight and easy portability. These can be placed near the test object which
might not always be confined to one location.
The length of the divider depends upon two or three factors. The maximum voltage to be measured
is the first and if height is a limitation, the length can be based on a surface flash over gradient in the
order of 3–4 kV/cm irrespective of whether the resistance R1 is of liquid or wirewound construction.
The length also depends upon the resistance value but this is implicitly bound up with the stray capacitance
of the resistance column, the product of the two (RC) giving a time constant the value of which must
not exceed the duration of the wave front it is required to record.
It is to be noted with caution that the resistance of the potential divider should be matched to the
equivalent resistance of a given generator to obtain a given wave shape.
Fig. 4.19 (a) shows a common form of resistance potential divider used for testing purposes
where the wave front time of the wave is less than 1 micro sec.
R1 R1 R1
R3 Z Z R3 Z
V1
R2 R2 R4 R2 R4
V2
( ) ( ) ( )
Fig. 4.19 Various forms of resistance potential dividers recording circuits ( ) Matching at divider end
( ) Matching at Oscillograph end ( ) Matching at both ends of delay cable
Here R3, the resistance at the divider end of the delay cable is chosen such that R2 + R3 = Z which
puts an upper limit on R2 i.e., R2 < Z. In fact, sometimes the condition for matching is given as
R1 R2
Z = R3 +
R1 R2
134 HIGH VOLTAGE ENGINEERING
But, since usually R1 > > R2, the above relation reduces to Z = R3 + R2. From Fig. 4.19 (a), the
voltage appearing across R2 is
Z1
V2 = V1
Z1 R1
where Z1 is the equivalent impedance of R2 in parallel with (Z + R3), the surge impedance of the cable
being represented by an impedance Z to ground.
(Z R3 ) R2 (Z R3 ) R2
Now Z1 =
R2 Z R3 2Z
(Z R3 ) R2 V1
Therefore, V2 =
2Z Z1 R1
However, the voltage entering the delay cable is
V2 Z (Z R3 ) R2 V1 R2
V3 = Z . V1
Z R3 Z R3 2Z Z1 R1 2( Z1 R1 )
As this voltage wave reaches the CRO end of the delay cable, it suffers reflections as the
impedance offered by the CRO is infinite and as a result the voltage wave transmitted into the CRO is
doubled. The CRO, therefore, records a voltage
R2
V3 = V1
Z1 R1
The reflected wave, however, as it reaches the low voltage arm of the potential divider does not
suffer any reflection as Z = R2 + R3 and is totally absorbed by (R2 + R3).
Since R2 is smaller than Z and Z1 is a parallel combination of R2 and (R3 + Z), Z1 is going to be
smaller than R2 and since R1 > > R2, R1 will be much greater than Z1 and, therefore to a first approximation
Z1 + R1 R1.
R2 R2
Therefore, V3 = V1 V1 as R2 < < R1
R1 R1 R2
Fig. 4.19 (b) and (c) are the variants of the potential divider circuit of Fig. 4.19 (a). The cable
matching is done by a pure ohmic resistance R4 = Z at the end of the delay cable and, therefore, the
voltage reflection coefficient is zero i.e. the voltage at the end of the cable is transmitted completely
into R4 and hence appears across the CRO plates without being reflected. As the input impedance of the
delay cable is R4 = Z, this resistance is a parallel to R2 and forms an integral part of the divider’s low
voltage arm. The voltage of such a divider is, therefore, calculated as follows:
Equivalent impedance
R2 Z R1 ( R2 Z ) R2 Z
= R1 +
R2 Z ( R2 Z )
V1 ( R2 Z )
Therefore, Current I=
R1 ( R2 Z ) R2 Z
IR2 Z V1 ( R2 Z ) R2 Z
and voltage V2 =
R2 Z R1 ( R2 Z ) R2 Z R2 Z
MEASUREMENT OF HIGH VOLTAGES AND CURRENTS 135
R2 Z
= V1
R1 ( R2 Z ) R2 Z
V2 R2 Z
or voltage ratio V1 R1 ( R2 Z ) R2 Z
Due to the matching at the CRO end of the delay cable, the voltage does not suffer any reflection
at that end and the voltage recorded by the CRO is given as
R2 Z V1 R2 ZV1 R2 V1
V2 =
R1 ( R2 Z ) R2 Z ( R1 R2 ) Z R1 R2 R1 R2
( R1 R2 )
Z
Normally for undistorted wave shape through the cable
Z R2
Therefore,
R2
V2 = V1
2 R1 R2
For a given applied voltage V1 this arrangement will produce a smaller deflection on the CRO
plates as compared to the one in Fig. 4.19 (a).
The arrangement of Fig. 4.19 (c) provides for matching at both ends of the delay cable and is to
be recommended where it is felt necessary to reduce to the minimum irregularities produced in the
delay cable circuit. Since matching is provided at the CRO end of the delay cable, therefore, there is no
reflection of the voltage at that end and the voltage recorded will be half of that recorded in the
arrangement of Fig. 4.19 (a) viz.
R2
V2 = V1
2( R1 R2 )
It is desirable to enclose the low voltage resistance (s) of the potential dividers in a metal screening
box. Steel sheet is a suitable material for this box which could be provided with a detachable close
fitting lid for easy access. If there are two low voltage resistors at the divider position as in Fig. 4.19 (a)
and (c), they should be contained in the screening box, as close together as possible, with a removable
metallic partition between them. The partition serves two purposes (i) it acts as an electrostatic shield
between the two resistors (ii) it facilitates the changing of the resistors. The lengths of the leads should
be short so that practically no inductance is contributed by these leads. The screening box should be
fitted with a large earthing terminal. Fig. 4.20 shows a sketched cross-section of possible layout for the
low voltage arm of voltage divider.
Circuit elements
R2, C2 From high voltage arm
Matching Metal
impedance if reqd. casing
Delay cable
R1
C1 C1
R Z, Cd R3 Cd C1 ( Z – R2 ) Z
C4 R2
C2 C2
R4
C2
( ) ( ) ( )
For voltage dividers in Fig. (b) and (c), the delay cable cannot be matched at its end. A low
resistor in parallel to C2 would load the low voltage arm of the divider too heavily and decrease the
output voltage with time. Since R and Z form a potential divider and R = Z, the voltage input to the
cable will be half of the voltage across the capacitor C2. This halved voltages travels towards the open
end of the cable (CRO end) and gets doubled after reflection. That is, the voltage recorded by the CRO
is equal to the voltage across the capacitor C2. The reflected wave charges the cable to its final voltage
magnitude and is absorbed by R (i.e. reflection takes place at R and since R = Z, the wave is completely
absorbed as coefficient of voltage reflection is zero) as the capacitor C2 acts as a short circuit for high
frequency waves. The transformation ratio, therefore, changes from the value:
C1 C2
C1
for very high frequencies to the value
C1 C2 Cd
C1
for low frequencies.
MEASUREMENT OF HIGH VOLTAGES AND CURRENTS 137
However, the capacitance of the delay cable Cd is usually small as compared with C2.
For capacitive divider an additional damping resistance is usually connected in the lead on the
high voltage side as shown in Fig. 4.21 (c). The performance of the divider can be improved if damping
resistor which corresponds to the aperiodic limiting case is inserted in series with the individual element
of capacitor divider. This kind of damped capacitive divider acts for high frequencies as a resistive
divider and for low frequencies as a capacitive divider. It can, therefore, be used over a wide range of
frequencies i.e. for impulse voltages of very different duration and also for alternating voltages.
Fig. 4.22 shows a simplified diagram of a resistance potential divider after taking into
considerations the lead in connection as the inductance and the stray capacitance as lumped capacitance.
Here L represents the loop inductance of the lead-in connection for the high voltage arm. The damping
resistance Rd limits the transient overshoot in the circuit formed by test object, L, Rd and C. Its value has
a decided effect on the performance of the divider. In order to evaluate the voltage transformation of the
divider, the low voltage arm voltage V2 resulting from a square wave impulse V1 on the hv side must be
investigaged. The voltage V2 follows curve 2 in Fig. 4.23 (a) in case of aperiodic damping and curve 2
in Fig. 4.23 (b) in case of sub-critical damping. The total area between curves 1 and 2 taking into
consideration the polarity, is described as the response time.
2
1 –
1 –
V2 (t) + +
2 V2 (t) +
( ) t ( ) t
With subcritical damping, even though the response time is smaller, the damping should not be
very small. This is because an undesirable resonance may occur for a certain frequency within the
passing frequency band of the divider. A compromise must therefore be realised between the short rise
time and the rapid stabilization of the measuring system. According to IEC publication No. 60 a maximum
overshoot of 3% is allowed for the full impulse wave, 5% for an impulse wave chopped on the front at
times shorter than 1 micro sec. In order to fulfill these requirements, the response time of the divider
must not exceed 0.2 micro sec. for full impulse waves 1.2/50 or 1.2/5 or impulse waves chopped on the
138 HIGH VOLTAGE ENGINEERING
tail. If the impulse wave is chopped on the front at time shorter than 1 micro sec the response time must
be not greater than 5% of the time to chopping.
Positioning device
There are characteristic differences between the figures for positive and negative voltages.
However, for either polarity the radius of the figure (if it is symmetrical) or the maximum distance from
the centre of the figure to its outside edge (if it is unsymmetrical) is a function only of the applied
voltage. The oscillatory voltages produce superimposed effects for each part of the wave. Thus it is
possible to know whether the wave is unidirectional or oscillatory. Since the size of the figure for
positive polarity is larger, it is preferable to use positive polarity figures. This is particularly desirable
in case of measurement of surges on transmission lines or other such equipment which are ordinarily
operating on a.c. voltage and the alternating voltage gives a black band along the centre of the film
caused by superposition of positive and negative figures produced on each half cycle. For each surge
voltage it is possible to obtain both positive and negative polarity figures by connecting pairs of electrodes
in parallel, one pair with a high voltage point and an earthed plate and the other pair with a high voltage
plate and an earthed point.
MEASUREMENT OF HIGH VOLTAGES AND CURRENTS 139
Klydonograph being a simple and inexpensive device, a large number of elements can be used
for measurement. It has been used in the past quite extensively for providing statistical data on magnitude,
polarity and frequency of voltage surges on transmission lines even though its accuracy of measurement
is only of the order of 25 per cent.
Hall Generators
Hall effect (Fig. 4.25) is used to measure very high direct current. Whenever electric current flows
through a metal plate placed in a magnetic field perpendicular to it, Lorenz force will deflect the electrons
in the metal structure in a direction perpendicular to the direction of both the magnetic field and the
flow of current. The charge displacement results in an e.m.f. in the perpendicular direction called the
Hall voltage. The Hall voltage is proportional to the current I, the magnetic flux density B and inversely
proportional to the plate thickness d i.e.,
BI
VH = R
d
where R is the Hall coefficient which depends upon the material of the plate and temperature of the
plate. For metals the Hall coefficient is very small and hence semiconductor materials are used for
which the Hall coefficient is high.
B
I B I
I VH
VH d
(Constant)
B
R
E
( )
( )
When large d.c. currents are to be measured the current carrying conductor is passed through an
iron cored magnetic circuit (Fig. 4.25 (b)). The magnetic field intensity produced by the conductor in
the air gap at a depth d is given by
1
H=
2 d
The Hall element is placed in the air gap and a small constant d.c. current is passed through the element.
The voltage developed across the Hall element is measured and by using the expression for Hall voltage
the flux density B is calculated and hence the value of current I is obtained.
Fig. 4.26 Current transformers and electro-optical system for high a.c. current measurements
A voltage signal proportional to the current to be measured is produced and is transmitted to the
ground through the electro-optical device. Light pulses proportional to the voltage signal are transmit-
ted by a glass optical fibre bundle to a photodetector and converted back into an analog voltage signal.
The required power for the signal convertor and optical device are obtained from suitable current and
voltage transformers.
MEASUREMENT OF HIGH VOLTAGES AND CURRENTS 141
and the outer case. The frequency response of this element is almost a flat characteristic upto about
1000 MHz and the response time is a few nanoseconds. The upper frequency limit is governed by the
skin effect in the sensitive element.
() ( )
Squirrel cage shunts are high ohmic shunts which can dissipate larger energies as compared to
coaxial shunts which are unsuitable due to their limitation of heat dissipation, larger wall thickness and
the skin effect. Squirrel cage shunt consists of thick metallic rods or strips placed around the periphery
of a cylinder and the structure resembles the rotor construction of a double squirrel cage induction
motor. The step response of the element is peaky and, therefore, a compensating network is used in
conjunction with the element to improve its frequency response. Rise times less than 8 n sec and band
width of 400 MHz have been obtained with these shunts.
Integration of v(t) can be carried out more elegantly by using an appropriately wired operational
amplifier. The frequency response of the Rogowski coil is flat upto 100 MHz but beyond that it is
affected by the stray electric and magnetic fields and also by the skin effect.
i(t) v(t)
v0(t)
Magnetic Links
These are used for the measurement of peak magnitude of the current flowing in a conductor. These
links consist of a small number of short steel strips on high retentivity. The link is mounted at a known
distance from the current carrying conductor. It has been found through experiments that the remanant
magnetism of the link after impulse current of 0.5/5 micro sec shape passes through the conductor is
same as that caused by a direct current of the same peak value. Measurement of the remanance possessed
by the link after the impulse current has passed through the conductor enables to calculate the peak
value of the current. For accurate measurements, it is usual to mount two or more links at different
distances from the same conductor. Because of its relative simplicity, the method has been used for
measurement of lightning current especially on transmission towers.
It is to be noted that the magnetic links help in recording the peak value of the impulse current
but gives no information regarding the wave shape of the current. For this purpose, an instrument
called Fulcronograph has been developed which consists of an aluminium wheel round the rim of
which are slots containing magnetic links of sufficient length to project on both sides of the wheel. As
the wheel is rotated, the links pass successively through a pair of narrow coils through which flows the
current to be measured. The current at the instant during which a particular link traverses the coil, can
be determined by a subsequent measurement of the residual flux in the link and, therefore, a curve
relating the variation of current with time can be obtained. The time scale is governed by the speed of
rotation of the wheel.
Hall Generators
The high amplitude a.c. and impulse currents can be measured by Hall Generator described earlier. For
the Hall Generator, though a constant control current flows which is permeated by the magnetic field of
the current to be measured, the Hall voltage is directly proportional to the measuring current. This
method became popular with the development of semi-conductor with sufficient high value of Hall
constant. The band width of such devices is found to be about 50 MHz with suitable compensating
devices and feedback.
beam passes through a transparent crystal in the presence of a magnetic field, the plane of polarisation
of the light beam undergoes relation. The angle of rotation is given by
Bl
where = A constant of the cyrstal which is a function of the wave length of the light.
B = Magnetic flux density due to the current to be measured in this case.
l = Length of the crystal.
C
i(t)
P1 P2
L
CRO
PM
Fig. 4.30 Magneto-optical method
Fig. 4.30 shows a schematic diagram of Magneto-optic method. Crystal C is placed parallel to
the magnetic field produced by the current to be measured. A beam of light from a stabilised light
source is made incident on the crystal C after it is passed through the polariser P1. The light beam
undergoes rotation of its plane of polarisation. After the beam passes through the analyser P2, the beam
is focussed on a photomultiplier, the output of which is fed to a CRO. The filter F allows only the
monochromatic light to pass through it. Photoluminescent diodes too, the momentary light emission of
which is proportional to the current flowing through them, can be used for current measurement. The
following are the advantages of the method (i) It provides isolation of the measuring set up from the
main current circuit. (ii) It is insensitive to overloading. (iii) As the signal transmission is through an
optical system no insulation problem is faced. However, this device does not operate for d.c. current.
Example 4.1. Determine the breakdown voltage for air gaps of 2 mm and 15 mm lengths under uni-
form field and standard atmospheric conditions. Also, determine the voltage if the atmospheric pres-
sure is 750 mm Hg and temperature 35°C.
Solution: According to empirical formula which holds good at standard atmospheric conditions
Vb = 24.22 S + 6.08 S
where S is the gap length in cms.
(i) When S = 0.2 cm
V = 24.22 × 0.2 + 6.08 0.2 = 7.56 kV Ans.
MEASUREMENT OF HIGH VOLTAGES AND CURRENTS 145
1 V2
Force of attraction = A
2 d2
1 1 V2
0.2 × 10–3 × 9.8 = 10 9
6
16 10 4
2 36 16 10
1 7
1.96 × 10–3 = 10 V2
72
of V = 1188 volts. Ans.
Example 4.3. An electrostatic voltmeter has two parallel plates. The movable plate is 10 cm in diam-
eter. With 10 kV between the plates the pull is 5 × 10–3 N. Determine the change in capacitance for a
movement of 1 mm of movable plate.
1 1 9 188 4
Solution: 5 × 10–3 = . 10 25 10
2 36 d2
or d = 26.35 mm.
Therefore, change in capacitance
10 3 9 4 1 1
10 25 10 = 0.0959 pF Ans.
36 26.35 27.35
Example 4.4. A generating voltmeter is required to measure voltage between 15 kV to 250 kV. If the
indicating meter reads a minimum current of 2 A and maximum of 35 A, determine the capacitance
of the generating voltmeter. Assume that the speed of driving synchronous motor is 1500 rpm.
Solution:
VCm w
I rms
2
146 HIGH VOLTAGE ENGINEERING
15 1000 Cm 2 1500
2 × 10–6 =
2 60
Cm = 12 × 10–7 × 10–6 = 1.5 pF Ans.
At 250 kV, the current indicated will be
250
2× 33.3 A Ans.
15
Example 4.5. A peak reading voltmeter is required to measure voltage upto 150 kV. The peak voltmeter
uses an RC circuit, a microammeter and a capacitance potential divider. The potential divider has a
ratio of 1200 : 1 and the micrometer can read upto 10 A. Determine the value of R and C if the time
constant of RC circuit is 8 secs.
Solution: The voltage across the low voltage arm of the potential divider,
150 1000
= 125 volts.
1200
The same voltage appears across the resistance.
V 125
Therefore R 6
12.5 M
I 10 10
Since the time constant of the RC circuit is 8 sec.
8
C= 0.64 F Ans.
12.5 10 6
Example 4.6. A Rogowski coil is required to measure impulse current of 8 kA having rate of change of
current of 1010 A/sec. The voltmeter is connected across the integrating circuit which reads 8 volts for
full scale deflection. The input to the integrating circuit is from the Rogowski Coil. Determine the
mutual inductance of coil R and C of the integrating circuit.
RC
Solution: i(t) = V0 (t )
M
RC
8 × 103 = ×8
M
RC
or 10 3
M
Time taken for the current wave to reach to maximum value
8 1000
= = 8 × 10–7 sec.
1010
This is for 1/4th of a cycle. Therefore for one cycle, the time is
T = 32 × 10–7
1 1 10 7
f 7
Hz
T 32 10 32
MEASUREMENT OF HIGH VOLTAGES AND CURRENTS 147
4.1. What are the requirements of a sphere gap for measurement of high voltages? Discuss the disadvantages
of sphere gap for measurements.
4.2. Explain clearly the procedure for measurement of (i) impulse; (ii) a.c. high voltages using sphere gap.
4.3. Discuss the effect of (i) nearby earthed objects (ii) humidity and (iii) dust particles on the measurements
using sphere gaps.
4.4. Describe the construction of a uniform field spark gap and discuss its advantages and disadvantages for
high voltage measurements.
4.5. Explain with neat diagram how rod gaps can be used for measurement of high voltages. Compare its
performance with a sphere gap.
4.6. Explain with neat diagram the principle of operation of an Electrostatic Voltmeter. Discuss its advantages
and limitations for high voltage measurements.
4.7. Draw a neat schematic diagram of a generating voltmeter and explain its principle of operation. Discuss
its application and limitations.
4.8. Draw Chubb-Fortescue Circuit for measurement of peak value of a.c. voltages discuss its advantages over
other methods.
4.9. Discuss the problems associated with peak voltmeter circuits using passive elements. Draw circuit devel-
oped by Rabus and explain how this circuit overcomes these problems.
4.10. What are the problems associated with measurement of very high impulse voltages? Explain how these
can be taken care of during measurements.
4.11. Discuss and compare the performance of (i) resistance (ii) capacitance potential dividers for measurement
of impulse voltages.
4.12. Discuss various resistance potential dividers and compare their performance of measurement of impulse
voltages.
4.13. Discuss various capacitance, potential dividers and compare their performance for measurement of im-
pulse voltages.
4.14. Draw a simplified equivalent circuit of a resistance potential divider and discuss its step response.
4.15. Discuss various methods of measuring high d.c. and a.c. currents.
4.16. Discuss various methods of measuring high impulse currents.
4.17. What is Rogowski Coil? Explain with a neat diagram its principle of operation for measurement of high
impulse currents.