Scope-CC-2555 BAGSON
Scope-CC-2555 BAGSON
Scope-CC-2555 BAGSON
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 1 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Permanent Facility
ELECTRO-
TECHNICAL- Using 6 ½ DMM
Alternating Agilent 34401A by
1 AC Current At 50 Hz 0.5 A to 3 A 0.28%
Current (< 1 Comparison/Direct
GHz) Method
(Measure)
ELECTRO- Using 6 ½ DMM
TECHNICAL- Agilent 34401A with
Alternating Agilent Shunt
2 AC Current At 50 Hz 10 A to 20 A 0.71 % to 0.72 %
Current (< 1 30A/mV by
GHz) Comparison/Direct
(Measure) Method
ELECTRO-
TECHNICAL- Using 6 ½ DMM
Alternating Agilent 34401A by
3 AC Current At 50 Hz 100 mA to 0.5 A 0.65 % to 0.26 %
Current (< 1 Comparison/Direct
GHz) Method
(Measure)
ELECTRO- Using 6 ½ DMM
TECHNICAL- Agilent 34401A with
Alternating Agilent Shunt
4 AC Current At 50 Hz 3 A to 10 A 0.25 % to 0.71 %
Current (< 1 30A/30mV by
GHz) Comparison/Direct
(Measure) Method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 2 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Using DMM Escort
Alternating Ac High Voltage At 97 with HV Probe
5 1 kV to 25 kV 0.1 kV to 1.7 kV
Current (< 1 50 Hz Fluke 80 K-40 by
GHz) Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using DMM Escort
Alternating Ac High Voltage At 97 with HV Probe
6 1 kV to 6 kV 0.07 kV to 0.33 kV
Current (< 1 50 Hz Fluke 80 K-40 by
GHz) Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using 6 ½ DMM
Alternating Agilent 34401A by
7 AC Voltage At 50 Hz 10 mV to 30 mV 0.55 % to 0.25 %
Current (< 1 Comparison/Direct
GHz) Method
(Measure)
ELECTRO-
TECHNICAL- Using 6 ½ DMM
Alternating Agilent 34401A by
8 AC Voltage At 50 Hz 100 mV to 750 V 0.15%
Current (< 1 Comparison/Direct
GHz) Method
(Measure)
ELECTRO-
TECHNICAL- Using 6 ½ DMM
Alternating Agilent 34401A by
9 AC Voltage At 50 Hz 30 mV to 100 mV 0.55 % to 0.16 %
Current (< 1 Comparison/Direct
GHz) Method
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 3 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL-
Using Digital LCR
Alternating Capacitance At 1
10 Meter Aplab 4910 by 1 nF to 1 µF 0.64 % to 0.56 %
Current (< 1 kHz
Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using Digital LCR
Alternating
11 Inductance At 1 kHz Meter Aplab 4910 by 1 mH to 1 H 0.65 % to 0.5 %
Current (< 1
Direct Method
GHz)
(Measure)
ELECTRO-
Using Fluke 5522A
TECHNICAL-
AC Current (50 Hz - Multiproduct
12 Alternating 100 µA to 100 mA 0.27 % to 0.16 %
1kHz) Calibrator by direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Fluke 5522A
TECHNICAL-
AC Current (50 Hz - Multiproduct
13 Alternating 100 mA to 1 A 0.16 % to 0.08 %
1kHz) Calibrator by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
AC Current (50 Hz -
14 Alternating Product Calibrator 30 µA to 100 µA 0.52 % to 0.27 %
1kHz)
Current (< 1 by direct method
GHz) (Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 4 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
Using Fluke 5522A
TECHNICAL-
Multi Product
15 Alternating AC Current 50 Hz 1 A to 10 A 0.08 % to 0.15 %
Calibrator by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using 5522A
TECHNICAL-
Multiproduct
16 Alternating AC Current 50 Hz 10 to 20 0.15 to 0.21
Calibrator by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO- Using 5522A
TECHNICAL- Multiproduct
17 Alternating AC Current 50 Hz Calibrator with 50 50 A to 1000 A 0.8 % to 0.8 %
Current (< 1 Turns coil by Direct
GHz) (Source) Method
ELECTRO- AC Power @50
Using Fluke 5522A
TECHNICAL- Hz,7.5 W-12kW (15
Multiproduct
18 Alternating V to 600V , 0.5 A to 7.5 W to 12 kW 0.31%
Calibrator by Direct
Current (< 1 20 A)0.5 Lead & Lag,
Method
GHz) (Source) UPF
ELECTRO-
Using 5522A
TECHNICAL-
AC Voltage At 50 Hz Multiproduct
19 Alternating 1 V to 1000 V 0.04%
to 1 kHz Calibrator by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using 5522A
TECHNICAL-
AC Voltage At 50 Hz Multiproduct
20 Alternating 10 mV to 1 V 0.11 % to 0.04 %
to 1 kHz Calibrator by Direct
Current (< 1
Method
GHz) (Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 5 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
Using Decade
TECHNICAL-
Capacitance At 1 Capacitance Box
21 Alternating 100 pF to 10 µF 2.26%
kHz Sigma by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using
TECHNICAL-
Std.Inductance Box
22 Alternating Inductance At 1 kHz 100 µH to 10 H 2.26%
LDB-6 by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using 5522A
TECHNICAL-
Power Factor At 50 Multiproduct
23 Alternating 0.2 PF to 1 PF 0.006PF
Hz Calibrator by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using 6 ½ DMM
TECHNICAL-
Agilent 34401A by
24 DIRECT DC Current 0.1 mA to 1 mA 2.33 % to 0.39 %
Comparison/Direct
CURRENT
Method
(Measure)
ELECTRO-
Using 6 ½ DMM
TECHNICAL-
Agilent 34401A by
25 DIRECT DC Current 1 mA to 100 mA 0.29 % to 0.17 %
Comparison/Direct
CURRENT
Method
(Measure)
ELECTRO-
Using 6 ½ DMM
TECHNICAL-
Agilent 34401A by
26 DIRECT DC Current 100 mA to 3 A 0.07 % to 0.18 %
Comparison/Direct
CURRENT
Method
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 6 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 7 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
Using 6 ½ DMM
TECHNICAL-
Agilent 34401A by
33 DIRECT DC Voltage 1 mV to 100 mV 0.45 % to 0.04 %
Comparison/Direct
CURRENT
Method
(Measure)
ELECTRO-
Using 6 ½ DMM
TECHNICAL-
Agilent 34401A by
34 DIRECT DC Voltage 1 V to 1000 V 0.008%
Comparison/Direct
CURRENT
Method
(Measure)
ELECTRO-
Using 6 ½ DMM
TECHNICAL-
Agilent 34401A by
35 DIRECT DC Voltage 100 mV to 1 V 0.04 % to 0.008 %
Comparison/Direct
CURRENT
Method
(Measure)
ELECTRO-
Using Flluke 5522 A
TECHNICAL-
Multiproduct
36 DIRECT DC Current 10 A to 20 A 0.15 % to 0.21 %
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Flluke 5522 A
TECHNICAL-
Multiproduct
37 DIRECT DC Current 100 µA to 100 mA 0.27 % to 0.16 %
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Flluke 5522 A
TECHNICAL-
Multiproduct
38 DIRECT DC Current 100 mA to 3 A 0.16 % to 0.08 %
Calibrator by Direct
CURRENT
Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 8 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
Using Flluke 5522 A
TECHNICAL-
Multiproduct
39 DIRECT DC Current 3 A to 10 A 0.08 % to 0.15 %
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Fluke 5522A
TECHNICAL-
Multi Product
40 DIRECT DC Current 30 µA to 100 µA 0.52 % to 0.27 %
Calibrator by direct
CURRENT
method
(Source)
ELECTRO- Using Flluke 5522A
TECHNICAL- Multiproduct
41 DIRECT DC Current Calibrator with 50 50 A to 1000 A 0.72 % to 0.7 %
CURRENT Turns Coil by Direct
(Source) Method
ELECTRO-
Using Resistance
TECHNICAL-
Box (Discrete
42 DIRECT DC Resistance 1 mO 0.55%
Values)by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Resistance
TECHNICAL-
Box(Discrete
43 DIRECT DC Resistance 1O 0.23%
Values)by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Resistance
TECHNICAL-
Box(Discrete
44 DIRECT DC Resistance 10 mO 0.23%
Values)by Direct
CURRENT
Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 9 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
Using Resistance
TECHNICAL-
Box(Discrete
45 DIRECT DC Resistance 100 mO 0.23%
Values)by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Flluke 5522A
TECHNICAL-
Multiproduct
46 DIRECT DC Voltage 1 mV to 100 mV 0.17 % to 0.004 %
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Flluke 5522A
TECHNICAL-
Multiproduct
47 DIRECT DC Voltage 1 V to 1000 V 0.002%
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Flluke 5522A
TECHNICAL-
Multiproduct
48 DIRECT DC Voltage 100 mV to 1 V 0.004 % to 0.002 %
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Mega Ohm
TECHNICAL-
Box Sigma (Discrete
49 DIRECT High Resistance 1 GO 3.4%
Values)by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Mega Ohm
TECHNICAL-
Box Sigma (Discrete
50 DIRECT High Resistance 100 MO 3.4%
Values)by Direct
CURRENT
Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 10 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
Using Mega Ohm
TECHNICAL-
Box Sigma (Discrete
51 DIRECT High Resistance 2 GO 4.0%
Values)by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Mega Ohm
TECHNICAL-
Box Sigma (Discrete
52 DIRECT High Resistance 50 MO 3.4%
Values)by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Mega Ohm
TECHNICAL-
Box Sigma (Discrete
53 DIRECT High Resistance 500 MO 3.4%
Values)by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Flluke 5522 A
TECHNICAL-
Multiproduct
54 DIRECT Resistance 1 O to 100 O 0.12 % to 0.004 %
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Flluke 5522 A
TECHNICAL-
Multiproduct
55 DIRECT Resistance 1 MO to 100 MO 0.007 % to 0.07 %
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Flluke 5522 A
TECHNICAL-
Multiproduct
56 DIRECT Resistance 100 O to 100 kO 0.004%
Calibrator by Direct
CURRENT
Method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 11 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
Using Flluke 5522 A
TECHNICAL-
Multiproduct
57 DIRECT Resistance 100 kO to 1 MO 0.004 % to 0.007 %
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Flluke 5522 A
TECHNICAL-
Multiproduct
58 DIRECT Resistance 100 MO to 1000 MO 0.07 % to 1.26 %
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature ( B
59 TEMPERATURE Product Calibrator 600 °C to 1800 °C 0.64 °C to 0.64 °C
Type Thermocouple)
SIMULATION by direct method
(Measure)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature ( J
60 TEMPERATURE Product Calibrator -200 °C to 1000 °C 0.31 °C to 0.31 °C
Type Thermocouple)
SIMULATION by direct method
(Measure)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature ( K
61 TEMPERATURE Product Calibrator 0 °C to 1300 °C 0.19 °C to 0.46 °C
Type Thermocouple)
SIMULATION by direct method
(Measure)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature ( R
62 TEMPERATURE Product Calibrator 0 °C to 1750 °C 0.65 °C to 0.47 °C
Type Thermocouple)
SIMULATION by direct method
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 12 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature ( S
63 TEMPERATURE Product Calibrator 0 °C to 1750 °C 0.56°C
Type Thermocouple)
SIMULATION by direct method
(Measure)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature (E-
64 TEMPERATURE Product Calibrator -200 °C to 1000 °C 0.57°C
Type Thermocouple)
SIMULATION by direct method
(Measure)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature (N-
65 TEMPERATURE Product Calibrator 0 °C to 1300 °C 0.22 °C to 0.31 °C
Type Thermocouple)
SIMULATION by direct method
(Measure)
ELECTRO-
TECHNICAL- Using Masibus
Temperature (RTD -
66 TEMPERATURE Calibrator MC -12 by -200 ºC to 850 ºC 0.18 ºC to 0.40 °C
PT 100)
SIMULATION direct method
(Measure)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature (T-
67 TEMPERATURE Product Calibrator -200 °C to 390 °C 0.72 °C to 0.17 °C
Type Thermocouple)
SIMULATION by direct method
(Measure)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature ( B
68 TEMPERATURE Product Calibrator 600 °C to 1800 °C 0.64 °C to 0.64 °C
Type Thermocouple)
SIMULATION by direct method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 13 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature ( J
69 TEMPERATURE Product Calibrator -200 °C to 1000 °C 0.31 °C to 0.27 °C
Type Thermocouple)
SIMULATION by direct method
(Source)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature ( K
70 TEMPERATURE Product Calibrator 0 °C to 1300 °C 0.46 °C to 0.46 °C
Type Thermocouple)
SIMULATION by direct method
(Source)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature ( R
71 TEMPERATURE Product Calibrator 0 °C to 1750 °C 0.65 °C to 0.65 °C
Type Thermocouple)
SIMULATION by direct method
(Source)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature ( S
72 TEMPERATURE Product Calibrator 0 °C to 1750 °C 0.54 °C to 0.54
Type Thermocouple)
SIMULATION by direct method
(Source)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature (E
73 TEMPERATURE Product Calibrator 0 °C to 1000 °C 0.57 °C to 0.57 °C
Type Thermocouple)
SIMULATION by direct method
(Source)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature (N
74 TEMPERATURE Product Calibrator 0 °C to 1300 °C 0.21 °C to 0.31 °C
Type Thermocouple)
SIMULATION by direct method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 14 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature (RTD -
75 TEMPERATURE Product Calibrator -200 °C to 800 °C 0.09 °C to 0.27 °C
PT 100)
SIMULATION by direct method
(Source)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
Temperature (T
76 TEMPERATURE Product Calibrator -200 °C to 390 °C 0.72 °C to 0.17 °C
Type Thermocouple)
SIMULATION by direct method
(Source)
ELECTRO- 6.5 Digit Digital
TECHNICAL- Multimeter Agilent
77 TIME & Frequency 34401A by 10 Hz to 300 kHz 0.08 % to 0.02 %
FREQUENCY Comparision/Direct
(Measure) Method
ELECTRO-
TECHNICAL-
Using digital timer
78 TIME & Time 1 S to 10 s 0.022S
by direct method
FREQUENCY
(Measure)
ELECTRO-
TECHNICAL-
Using digital timer 0.022 sec to 0.083
79 TIME & Time 10 s to 1000 s
by direct method sec
FREQUENCY
(Measure)
ELECTRO-
TECHNICAL-
Using digital timer 0.083 Sec to 1.42
80 TIME & Time 1000 s to 10000 s
by direct method Sec
FREQUENCY
(Measure)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 15 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL-
Using digital timer
81 TIME & Time 10000 s to 86400 s 1.42 s to 6.98 s
by direct method
FREQUENCY
(Measure)
ELECTRO-
TECHNICAL- Fluke-5522A Multi
82 TIME & Frequency Product Calibrator 10 Hz to 1 MHz 0.007 % to 0.007 %
FREQUENCY by direct method
(Source)
Non- Contact Type
MECHANICAL- Digital Tachometer
RPM Indicator/ 100 RPM to 25000
83 ACCELERATION & RPM generator 2.99 % to 0.13 %
Tachometer, RPM
AND SPEED source
Centrifuge
MECHANICAL- Sound Level meter Sound Level
84 94 dB to 114 dB 0.64dB
ACOUSTICS (1 kHz) Calibrator
Hydrometer of
resolution 0.0005
MECHANICAL- g/ml and
Density 0.600 g/ml to 1.600
85 DENSITY AND Appropriate Liquids 0.0018g/ml
Hydrometers g/ml
VISCOSITY BY Comparision
method as per
based on IS 3104
Hydrometer of
resolution 0.0005
MECHANICAL- g/ml and
Density 0.600 g/ml to 1.800
86 DENSITY AND Appropriate Liquids 0.001g/ml
Hydrometers g/ml
VISCOSITY BY Comparision
method as per
based on IS 3104
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 16 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION
Using Sine bar,
(BASIC
87 Angle Gauge Gauge Block & Dial Upto 60 degree 7.4Sec of arc
MEASURING
gauge
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Master
(BASIC Angle Plate / Box
88 Cylinder, Surface Upto 600 mm 3.0µm
MEASURING Angle Plate
Plate & Gauge Block
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using sine bar,
(BASIC Angle protector L.C gauge block, master
89 50 mm to 1000 mm 0.09°
MEASURING 0.1 ° cylinder, surface
INSTRUMENT, plate
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Dial Indicator,
(BASIC Bench Centre (
90 Electronic Level & Upto 1000 mm 8µm
MEASURING Parallelism)
Mandrel
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Dial Indicator,
(BASIC Bench Centre (Co -
91 Electronic Level & Upto 1000 mm 8µm
MEASURING Axiality)
Mandrel
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 17 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION
Using Dial Indicator,
(BASIC Bench Centre
92 Electronic Level & Upto 1000 mm 4µm
MEASURING (Flatness)
Mandrel
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Bevel Protractor &
(BASIC
93 Combination Set , LC Angle Gauges 0 ° to 360 ° 4min. of arc
MEASURING
- 5'
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Dial Calibration
Bore Gauge
(BASIC Tester, Dial Gauge,
94 Transmission Upto 2 mm 3.0µm
MEASURING Electronic Probe &
movement only
INSTRUMENT, gauge Block
GAUGE ETC.)
MECHANICAL-
DIMENSION
Caliper Checker,
(BASIC
95 Caliper Checker Length Bars & Lever Upto 600 mm 5.0µm
MEASURING
Dial gauge
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Coating Thickness
96 Using Thickness Foil 10 µm to 700 µm 3µm
MEASURING Gauge
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 18 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION
(BASIC Coating Thickness Using Thickness
97 Above 700 µm 5µm
MEASURING Gauge Foils
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Surface Plate
(BASIC Comparator Stand (
98 & Dial Gauge with 300X300 mm 2.0µm
MEASURING Flatness of Table)
stand
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Digital caliper of LC 25 x 25 x 25 mm to
99 Cube Mould 13.7µm
MEASURING 0.01 mm 150 x 150 x 150 mm
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Master
(BASIC Cylindrical Square
100 cylinder & Gauge Upto 600 mm 3.0µm
MEASURING (Squareness Only)
Block
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Depth Gauge ( LC - Using Caliper
(BASIC
101 0.01 mm) Vernier / Checker, Gauge 0 mm to 600 mm 7.2µm
MEASURING
Dial, Digital Block & Length Bars
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 19 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION Dial Gauge
Using Dial Gauge
(BASIC (Plunger/Lever
102 Calibrator and 0 mm to 25 mm 2.1µm
MEASURING Type), (LC -
Gauge Blocks
INSTRUMENT, 0.001mm)
GAUGE ETC.)
MECHANICAL-
DIMENSION Dial Gauge
Using Dial Gauge
(BASIC (Plunger/Lever
103 Calibrator and 0 mm to 50 mm 6.11µm
MEASURING Type), (LC -
Gauge Blocks
INSTRUMENT, 0.01mm)
GAUGE ETC.)
MECHANICAL-
DIMENSION
Dial Gauge
(BASIC Using Gauge Blocks
104 Calibrator (LC - 0 mm to 25 mm 2.0µm
MEASURING & Electronic Probe
0.001 mm)
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Dial Thickness
(BASIC
105 Gauge , (LC - Using Gauge blocks 0 mm to 25 mm 2.0µm
MEASURING
0.001mm)
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Caliper
(BASIC
106 Digital Caliper checker and Length 0 mm to 1000 mm 10.0µm
MEASURING
bars
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 20 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION
Using Master
(BASIC Engineer's Square/
107 Cylinder, Slip Gauge Upto 600 mm 3.0µm
MEASURING Try Square
& Surface Plate
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC External Micrometer Using Gauge Block &
108 100 mm to 600 mm 7.0µm
MEASURING ( LC - 0.01 mm) Length Bar
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC External Micrometer Using Gauge Block &
109 Upto 100 mm 3.0µm
MEASURING ( LC - 0.01 mm) Length Bar
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC External Micrometer Gauge Block &
110 0 mm to 150 mm 0.83µm
MEASURING ( LC - 0.001 mm) Length Bar
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Using Digital
111 Feeler Gauge 0.03 mm to 1 mm 3.4µm
MEASURING Micrometer
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 21 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION
Floating Carriage Using Gauge Block &
(BASIC
112 Micrometer (LC - Cylindrical 0 mm to 175 mm 1.20µm
MEASURING
0.2µm) Standards
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Gauge Block
(BASIC Calibrator and
113 Gauge Block 10 mm to 50 mm 0.20µm
MEASURING Gauge Block Set
INSTRUMENT, Grade - 00
GAUGE ETC.)
MECHANICAL-
DIMENSION Gauge Block
(BASIC Calibrator and
114 Gauge Block 50 mm to 100 mm 0.41µm
MEASURING Gauge Block Set
INSTRUMENT, Grade - 00
GAUGE ETC.)
MECHANICAL-
DIMENSION Gauge Block
(BASIC Calibrator and
115 Gauge Block Up to 10 mm 0.12µm
MEASURING Gauge Block Set
INSTRUMENT, Grade - 00
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Caliper
(BASIC Height Gauge (LC - Checker, Gauge
116 0 mm to 300 mm 8µm
MEASURING 0.01mm) Block ,Length Bars &
INSTRUMENT, Lever Dial gauge
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 22 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION Using Caliper
(BASIC Height Gauge (LC - Checker, Gauge 300 mm to 1000
117 10.0µm
MEASURING 0.01mm) Block , Length Bars mm
INSTRUMENT, & Lever Dial Gauge
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Caliper
(BASIC Height Master (LC - Checker, Gauge
118 0 mm to 600 mm 3.5µm
MEASURING 0.001mm) Block ,Length Bars &
INSTRUMENT, Lever Dial gauge
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Indside Dial Caliper
119 Using Gauge blocks 4 mm to 75 mm 8µm
MEASURING (LC - 0.010mm)
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Caliper Checker,
Inside Micrometer ,
(BASIC Gauge Block
120 LC - 0.01mm, (Two 5 mm to 25 mm 3.4µm
MEASURING Accessories &
Jaw & Stick type)
INSTRUMENT, Length Bars
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Caliper
Inside Micrometer,
(BASIC Checker, Gauge
121 (Two Jaw & Stick 25 mm to 50 mm 8.0µm
MEASURING Block Accessories &
type) LC - 0.01mm,
INSTRUMENT, Length Bars
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 23 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION Using Caliper
Inside Micrometer,
(BASIC Checker, Gauge 300 mm to 1000
122 (Two Jaw & Stick 10.0µm
MEASURING Block Accessories & mm
type) LC - 0.01mm,
INSTRUMENT, Length Bars
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Caliper
Inside Micrometer,
(BASIC Checker, Gauge
123 (Two Jaw & Stick 50 mm to 300 mm 8.0µm
MEASURING Block Accessories &
type) LC - 0.01mm,
INSTRUMENT, Length Bars
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Length Bars,
(BASIC
124 Length Bars LVDT Probe & 100 mm to 300 mm 1.10µm
MEASURING
Surface Plate
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Length Bars,
(BASIC
125 Length Bars LVDT Probe & 300 mm to 600 mm 4.3µm
MEASURING
Surface Plate
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Length Bars,
(BASIC 600 mm to 1000
126 Length Bars LVDT Probe & 5.0µm
MEASURING mm
Surface Plate
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 24 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION
Using Length Bars,
(BASIC
127 Length Bars LVDT Probe & Upto 100 mm 0.40µm
MEASURING
Surface Plate
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Using Caliper
DIMENSION
Checker, Length
(BASIC Linear Height Gauge
128 Bars , Lever Dial 0 mm to 1000 mm 4.2µm
MEASURING ( LC - 0.001 mm)
gauge and Gauge
INSTRUMENT,
Blocks
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Linear probe
129 Using Gauge Blocks 0 mm to 2 mm 0.60µm
MEASURING (Digital), LC - 0.1µm
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Master
(BASIC
130 Master Cylinder Cylinder, Surface Upto 600 mm 3.0µm
MEASURING
Plate & Gauge Block
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
using scale
(BASIC Pie Tape LC - 0.1
131 calibration unit with 1 mm to 25 meter 7.53µm
MEASURING mm
DRO of 0.001 mm
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 25 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION
(BASIC Using Length
132 Pin Gauge Upto 10 mm 1.0µm
MEASURING Measuring Machine
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Using Profile
133 Pitch gauge Upto 10 mm 5.0µm
MEASURING Projector
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Length Measuring
134 Plain Plug Gauge 0.5 mm to 100 mm 1µm
MEASURING Machine
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Length
(BASIC Measuring Machine
135 Plain Ring Gauge 3 mm to 300 mm 2.0µm
MEASURING by Comparision
INSTRUMENT, method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Profile Projector
(BASIC
136 linear scale LC = Glass Scale Upto 300 mm 2.3µm
MEASURING
0.001mm
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 26 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION
Profile Projector
(BASIC
137 Angular Scale LC = Angle gauge Upto 90 ° 33Sec of arc
MEASURING
1"
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Glass Scale
(BASIC Profile Projector
138 ,Standard wires & 5 x to 100 x 0.17%
MEASURING Magnification
Vernier Caliper
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC
139 Radius gauge Profile Projector Upto 50 mm 9.6µm
MEASURING
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Caliper
(BASIC
140 Riser Block checker, Lever Dial Upto 100 mm 1µm
MEASURING
gauge & Length bar
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Caliper
(BASIC
141 Riser Block checker, Lever Dial Upto 300 mm 2.0µm
MEASURING
Gauge & Length Bar
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 27 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION
Using Caliper
(BASIC
142 Riser Block checker, lever Dial Upto 600 mm 3.0µm
MEASURING
Gauge & Length Bar
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Profile projector &
(BASIC
143 Sieves Digital Vernier 10 mm to 150 mm 30µm
MEASURING
Caliper
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Profile Projector &
(BASIC
144 Sieves Digital Vernier Upto 10 mm 6.2µm
MEASURING
Caliper
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Angle Gauge,
(BASIC 4.5 µm to 6.5 sec of
145 Sine bar Dial gauge & Gauge Upto 300 mm
MEASURING arc
Block
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC
146 Snap Gauge Using Gauge Blocks 2 mm to 100 mm 3.0µm
MEASURING
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 28 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION
Spirit Level ,
(BASIC Electronic Level &
147 Sensitivity : 0 ' to 1 ' 4µm/m
MEASURING Titling table
0.005mm/m
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Spirit Level ,
(BASIC Electronic Level &
148 Sensitivity : 0 ' to 1 ' 6µm/m
MEASURING Titling Table
0.01mm/m
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Spirit Level ,
(BASIC Electronic Level &
149 Sensitivity : 0 ' to 1 ' 8µm/m
MEASURING Titling Table
0.02mm/m
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Steel Scale Fixture
(BASIC Steel Scale LC - 0.5
150 with DRO, LC - 0 mm to 2000 mm 9.48µm
MEASURING mm
0.001mm
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Steel Scale/ Tape
151 Using Length Bar Set Upto 1000 mm 6.9µm
MEASURING Calibrator
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 29 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION
Scale calibration unit
(BASIC Steel Tape (LC - 1 0 m to 1 Every
152 with DRO , LC = 7.04 µm
MEASURING mm) Subsequent m
0.001mm
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Straight Edge
DIMENSION
Thickness upto Gauge Blocks
(BASIC
153 10mm I-section ,Surface Plate & Upto 4 m 17µm/ m
MEASURING
Thickness 20mm & Electronic Level
INSTRUMENT,
above (Straightness)
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Electronic
(BASIC Surface Plate Upto size
154 Level lest count 0.60µm
MEASURING (Flatness) (6000X4000) mm
0.001mm/m
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Gauge Block &
(BASIC
155 Taper Plug Gauge Digital Micrometer, Upto 100 mm 3.0µm
MEASURING
Standard Wires
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Floating
(BASIC Taper Thread Plug Carriage Micrometer
156 Upto 100 mm 3.5µm
MEASURING Gauge (Only PCD) & Thread Measuring
INSTRUMENT, wires
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 30 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION
Using Gauge Block
(BASIC
157 Thickness Foil Comparator & 10 µm to 1000 µm 4.0µm
MEASURING
Gauge Block
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Floating
Thread Plug Gauge,
(BASIC Carriage Micrometer
158 wear Check Plugs Upto 100 mm 2.70µm
MEASURING & Thread Measuring
(Only PCD)
INSTRUMENT, wires
GAUGE ETC.)
MECHANICAL-
DIMENSION
Thread Ring Gauge Using LMM, Master
(BASIC
159 & Thread Check Ring Ring & Thread 6 mm to 100 mm 2.0µm
MEASURING
Gauge Measuring Probe
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Ultra Sound
160 Using Slip Gauge set 1 mm to 100 mm 49.0µm
MEASURING Thickness Gauge
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC V - Block (Flatness/ Using Dial gauge & Upto 200X150X100
161 3.2µm
MEASURING Parallelism) Test Mandrel mm
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 31 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION
Using Caliper
(BASIC Vernier Caliper (LC - 300 mm to 1000
162 Checker, Gauge 10.0µm
MEASURING 0.01mm) mm
Block & Length Bars
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Caliper
(BASIC Vernier Caliper (LC -
163 Checker, Gauge 0 mm to 300 mm 7.2µm
MEASURING 0.01mm)
Block & Length Bars
INSTRUMENT,
GAUGE ETC.)
Using digital single
MECHANICAL- pan electronic 0 shore A to 100
164 Shore A tester 0.5Shore A
DUROMETER balance of 0.1 g and Shore A
fixture. ASTM 2240D
Using digital single
MECHANICAL- pan electronic 0 Shore D to 100
165 Shore D tester 0.5Shore D
DUROMETER balance of 0.1 g and Shore D
fixture. ASTM 2240D
Using fixture with
MECHANICAL- aluminium and SS
MOBILE FORCE Mobile force gauge Hangers for push
166 10 N to 1000 N 0.77%
MEASURING (Push Pull Gauge) and pull mode and
SYSTEM F1 class SS weights.
VDI
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 32 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 33 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Using primary
MECHANICAL-
method, calibrated 1
TORQUE Torque Transducer
171 m beam and F1 100 Nm to 200 Nm 0.08%
GENERATING calibration,
class SS weights.
DEVICES
BS7882
MECHANICAL- Torque Wrench Type
Digital Torque
TORQUE - 1, Class, A,B,C,D,E
172 Calibrator IS 2 Nm to 20 Nm 0.64%
MEASURING Type - II, Class,
6789:2003
DEVICES A,B,C,D,E,F, & G
MECHANICAL- Torque Wrench Type
Digital Torque
TORQUE - 1, Class, A,B,C,D,E
173 Calibrator IS 2 Nm to 200 Nm 0.64%
MEASURING Type - II, Class,
6789:2003
DEVICES A,B,C,D,E,F, & G
MECHANICAL- Torque Wrench Type
Digital Torque
TORQUE - 1, Class, A,B,C,D,E
174 Calibrator IS: 200 Nm to 2000 Nm 0.75%
MEASURING Type - II, Class,
6789:2003
DEVICES A,B,C,D,E,F, & G
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 34 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class , precision
balances, double
distilled water of
known density as
per Gravimetric
method based on
Measuring Cylinder/
ISO 4787, 10 µl to
MECHANICAL- Volumetric Flask/
175 200 ml, LC of 0.1 ml to 100 ml 2µl
VOLUME Graduated Jar/ Cane
Balance = 0.01 / 0.1
etc.
mg, Range 80 / 220
g, 200 ml to 5000
ml, LC of Balance =
0.1 g, Range - 5000
g, 5000 to 20000 ml,
LC of Balance = 5 g,
Range – 60 kg.
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 35 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class , precision
balances, double
distilled water of
known density as
per Gravimetric
method based on
Measuring Cylinder/
ISO 4787, 10 µl to
MECHANICAL- Volumetric Flask/
176 200 ml, LC of 0.1 ml to 100 ml 2µl
VOLUME Graduated Jar/ Cane
Balance = 0.01 / 0.1
etc.
mg, Range 80 / 220
g, 200 ml to 5000
ml, LC of Balance =
0.1 g, Range - 5000
g, 5000 to 20000 ml,
LC of Balance = 5 g,
Range – 60 kg.
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 36 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class , precision
balances, double
distilled water of
known density as
per Gravimetric
method based on
Measuring Cylinder/
ISO 4787, 10 µl to
MECHANICAL- Volumetric Flask/
177 200 ml, LC of 1 l to 5 l 0.8ml
VOLUME Graduated Jar/ Cane
Balance = 0.01 / 0.1
etc.
mg, Range 80 / 220
g, 200 ml to 5000
ml, LC of Balance =
0.1 g, Range - 5000
g, 5000 to 20000 ml,
LC of Balance = 5 g,
Range – 60 kg.
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 37 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class , precision
balances, double
distilled water of
known density as
per Gravimetric
method based on
Measuring Cylinder/
ISO 4787, 10 µl to
MECHANICAL- Volumetric Flask/
178 200 ml, LC of 1 l to 5 l 0.8ml
VOLUME Graduated Jar/ Cane
Balance = 0.01 / 0.1
etc.
mg, Range 80 / 220
g, 200 ml to 5000
ml, LC of Balance =
0.1 g, Range - 5000
g, 5000 to 20000 ml,
LC of Balance = 5 g,
Range – 60 kg.:
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 38 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class , precision
balances, double
distilled water of
known density as
per Gravimetric
method based on
Measuring Cylinder/
ISO 4787, 10 µl to
MECHANICAL- Volumetric Flask/
179 200 ml, LC of 100 ml to 1 l 0.5ml
VOLUME Graduated Jar/ Cane
Balance = 0.01 / 0.1
etc.
mg, Range 80 / 220
g, 200 ml to 5000
ml, LC of Balance =
0.1 g, Range - 5000
g, 5000 to 20000 ml,
LC of Balance = 5 g,
Range – 60 kg.
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 39 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class , precision
balances, double
distilled water of
known density as
per Gravimetric
method based on
Measuring Cylinder/
ISO 4787, 10 µl to
MECHANICAL- Volumetric Flask/
180 200 ml, LC of 5 l to 20 l 7.1ml
VOLUME Graduated Jar/ Cane
Balance = 0.01 / 0.1
etc.
mg, Range 80 / 220
g, 200 ml to 5000
ml, LC of Balance =
0.1 g, Range - 5000
g, 5000 to 20000 ml,
LC of Balance = 5 g,
Range – 60 kg.
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 40 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class , precision
balances, double
distilled water of
known density as
per Gravimetric
method based on
Measuring Cylinder/
ISO 4787, 10 µl to
MECHANICAL- Volumetric Flask/
181 200 ml, LC of 5 l to 20 l 7ml
VOLUME Graduated Jar/ Cane
Balance = 0.01 / 0.1
etc.
mg, Range 80 / 220
g, 200 ml to 5000
ml, LC of Balance =
0.1 g, Range - 5000
g, 5000 to 20000 ml,
LC of Balance = 5 g,
Range – 60 kg.
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 41 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 42 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 43 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 44 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
Weighing Machine Standard Weights of
WEIGHING
189 Readability =1 mg & F1 Class II Weighing 0 to 1 kg 2.5mg
SCALE AND
coarser Balance & Coarser
BALANCE
Standard Weights of
MECHANICAL-
Weighing Machine F1Calibration &
WEIGHING
190 Readability =100 Weighing Balance of upto 5 kg 0.11g
SCALE AND
mg & coarser accuracy Class II &
BALANCE
Coarser
MECHANICAL-
Weighing Machine
WEIGHING Weighing Balance of
191 Readability =1g & upto 100 kg 2g
SCALE AND Class IV
coarser
BALANCE
MECHANICAL-
Weighing Machine
WEIGHING Weighing Balance of
192 Readability =20g & upto 200 kg 26g
SCALE AND Class IV
coarser
BALANCE
MECHANICAL-
Weighing Machine
WEIGHING Weighing Balance of
193 Readability =20g & upto 200 kg 26g
SCALE AND Class IV
coarser
BALANCE
MECHANICAL-
Weighing Machine
WEIGHING Weighing Balance of
194 Readability =5 g & upto 50 kg 5.1g
SCALE AND Class III & Coarser
coarser
BALANCE
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 45 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.01mg
MECHANICAL-
195 Conventional Mass) Procedure based on 1 g 0.10mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.01mg
MECHANICAL-
196 Conventional Mass) Procedure based on 1 mg 0.02mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.01mg
MECHANICAL-
197 Conventional Mass) Procedure based on 10 g 0.20mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 46 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.01mg
MECHANICAL-
198 Conventional Mass) Procedure based on 10 mg 0.02mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.1mg
MECHANICAL-
199 Conventional Mass) Procedure based on 100 g 0.5mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.01mg
MECHANICAL-
200 Conventional Mass) Procedure based on 100 mg 0.05mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 47 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.01mg
MECHANICAL-
201 Conventional Mass) Procedure based on 2 g 0.12mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.01mg
MECHANICAL-
202 Conventional Mass) Procedure based on 2 mg 0.02mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.01mg
MECHANICAL-
203 Conventional Mass) Procedure based on 20 g 0.25mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 48 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.01mg
MECHANICAL-
204 Conventional Mass) Procedure based on 20 mg 0.03mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.1mg
MECHANICAL-
205 Conventional Mass) Procedure based on 200 g 1.0mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.01mg
MECHANICAL-
206 Conventional Mass) Procedure based on 200 mg 0.06mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 49 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.01mg
MECHANICAL-
207 Conventional Mass) Procedure based on 5 g 0.16mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.01mg
MECHANICAL-
208 Conventional Mass) Procedure based on 5 mg 0.02
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.01mg
MECHANICAL-
209 Conventional Mass) Procedure based on 50 g 0.30mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 50 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.01mg
MECHANICAL-
210 Conventional Mass) Procedure based on 50 mg 0.04mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.01mg
MECHANICAL-
211 Conventional Mass) Procedure based on 500 mg 0.08mg
WEIGHTS
of F2 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.2 g
MECHANICAL-
212 Conventional Mass) Procedure based on 1 kg 0.1g
WEIGHTS
of M1 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 51 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 5 g
MECHANICAL-
213 Conventional Mass) Procedure based on 10 kg 5g
WEIGHTS
of M1 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 5 g
MECHANICAL-
214 Conventional Mass) Procedure based on 10 kg 6g
WEIGHTS
of M1 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.2mg
MECHANICAL-
215 Conventional Mass) Procedure based on 2 kg 0.1g
WEIGHTS
of M1 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 52 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 5 g
MECHANICAL-
216 Conventional Mass) Procedure based on 20 kg 5g
WEIGHTS
of M1 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 5 g
MECHANICAL-
217 Conventional Mass) Procedure based on 20 kg 6g
WEIGHTS
of M1 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 5 g
MECHANICAL-
218 Conventional Mass) Procedure based on 20 kg 6g
WEIGHTS
of M1 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 53 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.2 g
MECHANICAL-
219 Conventional Mass) Procedure based on 5 kg 0.3g
WEIGHTS
of M1 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 5 g
MECHANICAL-
220 Conventional Mass) Procedure based on 50 kg 6g
WEIGHTS
of M1 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Standard Weights F1
Class & Precision
Mass Calibration of Balance of
weights ( readability 0.2 g
MECHANICAL-
221 Conventional Mass) Procedure based on 500 g 0.2g
WEIGHTS
of M1 class and OIML R 111
coarser Substitution method
of ABA Weighing
cycle
Using light chamber
OPTICAL- and calibrated LUX
222 Digital Lux meter 100 lx to 10000 lx 5.9 % to 6.5 %
EQUIPMENTS meter by
comparison method
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 54 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Humidity Sensor
with Indicator
Relative Humidity
(Dial/Digital)/Thermo
THERMAL- Sensor with
Hygrometer, 10 %RH to 95
223 SPECIFIC HEAT Indicator, Humidity 1.16%RH
Humidity Chamber, %RH @ 25 ºC
& HUMIDITY Generator by
Environmental/Clima
Comparision Method
tic Humidity
Chamber
Temperature R Type
THERMAL- Indicator of Dry Thermocouple with
224 50 °C to 600 °C 0.65°C
TEMPERATURE Block furnace , Indicator by
Muffle Furnace Comparision Method
Temperature R Type
THERMAL- Indicator of Dry Thermocouple with
225 600 °C to 1200 °C 1.88°C
TEMPERATURE Block furnace , Indicator by
Muffle Furnace Comparision method
Temperature
Transmitter, RTD,
Thermocouples with
& without RTD (PT-100) 4 wire
THERMAL- controller/Data with Indicator, Oil
226 50 °C to 300 °C 0.62°C
TEMPERATURE Logger/recorder, bath by Comparision
Temperature Gauge, method
Glass Thermometer,
Digital Thermometer
with inbuilt sensor
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 55 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Temperature
Transmitter, RTD,
Thermocouples with
R Type
& without
Thermocouple with
THERMAL- controller/Data
227 Indicator, Dry Block 600 °C to 1200 °C 1.88°C
TEMPERATURE Logger/recorder,
Furnace by
Temperature Gauge,
Comparision method
Glass Thermometer,
Digital Thermometer
with inbuilt sensor
Temperature
Transmitter, RTD,
Thermocouples with
& without
R Type
controller/Data
Thermocouple with
THERMAL- Logger/recorder,
228 Indicator, Dry Block 300 °C to 600 °C 1.5°C
TEMPERATURE Temperature Gauge,
Furnace by
Glass Thermometer,
Comparision method
Digital Thermometer
with inbuilt sensor,
Thermohygrometers
with Sensor
Temperature
Transmitter, RTD,
Thermocouples with
& without RTD (PT-100) 4 wire
THERMAL- controller/Data with Indicator, Liquid
229 -30 °C to 50 °C 0.21°C
TEMPERATURE Logger/recorder, bath by Comparision
Temperature Gauge, method
Glass Thermometer,
Digital Thermometer
with inbuilt sensor.
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 56 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Thermohygrometer,
Using 4 Wire RTD
Temperature
(PT-100) sensor with
Indicator/Thermome
THERMAL- Indicator using 20 ºC to 50 ºC
230 ter/Data 0.82 ºC
TEMPERATURE Temperature @50%RH
logger/Temperature
chamber by
Gauge with inbuilt
comparison method
Temperature Sensor
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 57 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Site Facility
ELECTRO-
TECHNICAL- Using DMM Escort
Alternating Ac High Voltage At 97 with HV Probe
1 1 kV to 25 kV 0.1 kV to 1.7 kV
Current (< 1 50 Hz Fluke 80 K-40 by
GHz) Direct Method
(Measure)
ELECTRO-
Using DMM Escort
TECHNICAL-
97 with HV Probe
2 DIRECT DC High Voltage 1 kV to 30 kV 0.06 kV to 1.2 kV
Fluke 80 K-40 by
CURRENT
Direct Method
(Measure)
ELECTRO-
TECHNICAL- Masibus Calibrator
Temperature ( J
3 TEMPERATURE MC -12 by direct -200 °C to 1050 °C 0.35 °C to 0.35 °C
Type Thermocouple)
SIMULATION method
(Source)
ELECTRO-
TECHNICAL- Masibus Calibrator
Temperature ( K
4 TEMPERATURE MC -12 by direct 0 °C to 1300 °C 0.46 °C to 0.46 °C
Type Thermocouple)
SIMULATION method
(Source)
ELECTRO-
TECHNICAL- Masibus Calibrator
Temperature ( R
5 TEMPERATURE MC -12 by direct 0 °C to 1750 °C 0.57 °C to 0.57 °C
Type Thermocouple)
SIMULATION method
(Source)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 58 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
ELECTRO-
TECHNICAL- Masibus Calibrator
Temperature ( S
6 TEMPERATURE MC -12 by direct 0 °C to 1750 °C 0.57 °C to 0.57 °C
Type Thermocouple)
SIMULATION method
(Source)
ELECTRO-
TECHNICAL- Using Masibus
Temperature (RTD -
7 TEMPERATURE Calibrator MC -12 by -200 °C to 850 °C 0.18 °C to 0.40 °C
PT 100)
SIMULATION direct method
(Source)
Non- Contact Type
MECHANICAL- Digital Tachometer
RPM Indicator/ 100 RPM to 25000
8 ACCELERATION & RPM generator 2.99 % to 0.13 %
Tachometer, RPM
AND SPEED source
Centrifuge
MECHANICAL- Sound Level meter Sound Level
9 94 dB to 114 dB 0.64dB
ACOUSTICS (1 kHz) Calibrator
MECHANICAL-
DIMENSION
Using Master
(BASIC Angle Plate / Box
10 Cylinder, Surface Upto 600 mm 3.0µm
MEASURING Angle Plate
Plate & Gauge Block
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Dial Indicator,
(BASIC Bench Centre (
11 Electronic Level & Upto 1000 mm 8µm
MEASURING Parallelism)
Mandrel
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 59 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION
Using Dial Indicator,
(BASIC Bench Centre (Co -
12 Electronic Level & Upto 1000 mm 8µm
MEASURING Axiality)
Mandrel
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Dial Indicator,
(BASIC Bench Centre
13 Electronic Level & Upto 1000 mm 4µm
MEASURING (Flatness)
Mandrel
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Floating Carriage Using Gauge Block &
(BASIC
14 Micrometer (LC - Cylindrical 0 mm to 175 mm 1.20µm
MEASURING
0.2µm) Standards
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Caliper
(BASIC Linear Height Gauge Checker, Length
15 0 mm to 1000 mm 4µm
MEASURING LC - 0.0001mm Bars & Lever Dial
INSTRUMENT, Gauge
GAUGE ETC.)
MECHANICAL-
DIMENSION
Profile Projector
(BASIC
16 linear scale LC = Glass Scale Upto 300 mm 2.3µm
MEASURING
0.001mm
INSTRUMENT,
GAUGE ETC.)
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 60 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
DIMENSION
Profile Projector
(BASIC
17 Angular Scale LC = Angle gauge Upto 90 ° 33Sec of arc
MEASURING
1"
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Glass Scale
(BASIC Profile Projector
18 ,Standard wires & 5 x to 100 x 0.17%
MEASURING Magnification
Vernier Caliper
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Straight Edge
DIMENSION
Thickness upto Gauge Blocks
(BASIC
19 10mm I-section ,Surface Plate & Upto 4 m 17µm/ m
MEASURING
Thickness 20mm & Electronic Level
INSTRUMENT,
above (Straightness)
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Electronic
(BASIC Surface Plate Upto size
20 Level lest count 0.60µm
MEASURING (Flatness) (6000X4000) mm
0.001mm/m
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Hardness Blocks
HARDNESS Hardness Rockwell
21 Indirect Method IS : 20 HRC to 70 HRC 0.45 HRC
TESTING Testing machines
1586
MACHINES
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 61 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 62 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
UTM, TENSION
CREEP AND Compression Testing Dynamometer
26 50 kN to 1000 kN 0.5%
TORSION machine (Class- 0/1) IS: 1828
TESTING
MACHINE
MECHANICAL-
UTM, TENSION
CREEP AND Tensile/ compression Using Digital load
27 0.5 N to 50 N 0.5%
TORSION testing machines cell class 0/1 IS 1828
TESTING
MACHINE
MECHANICAL-
UTM, TENSION
Tensile/ Digital Load Cell
CREEP AND
28 Compression Testing (class - 0/1) IS : 50 N to 50 kN 0.7%
TORSION
mchine 1828
TESTING
MACHINE
Standard Weights of
F1 Class as per
MECHANICAL-
Weighing Machine Procedure based on
WEIGHING
29 Readability = 2 mg OIML R76 to upto 200 g 3mg
SCALE AND
& coarser (2006)calibration of
BALANCE
Class I Weighing
Balance & Coarser
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 63 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Standard Weights of
F1 Class as per
MECHANICAL-
Weighing Machine Procedure based on
WEIGHING
30 Readability = 2 mg OIML R76 up to 200 g 3mg
SCALE AND
& coarser (2006)calibration of
BALANCE
Class I Weighing
Balance & Coarser
MECHANICAL-
Weighing Machine Standard Weights of
WEIGHING
31 Readability =1 mg & F1 Class II Weighing 0 to 1 kg 2.5mg
SCALE AND
coarser Balance & Coarser
BALANCE
Standard Weights of
MECHANICAL-
Weighing Machine F1Calibration &
WEIGHING
32 Readability =100 Weighing Balance of upto 5 kg 0.11g
SCALE AND
mg & coarser accuracy Class II &
BALANCE
Coarser
MECHANICAL-
Weighing Machine
WEIGHING Weighing Balance of
33 Readability =1g & upto 100 kg 2g
SCALE AND Class IV
coarser
BALANCE
MECHANICAL-
Weighing Machine
WEIGHING Weighing Balance of
34 Readability =20g & upto 200 kg 26g
SCALE AND Class IV
coarser
BALANCE
MECHANICAL-
Weighing Machine
WEIGHING Weighing Balance of
35 Readability =20g & upto 200 kg 26g
SCALE AND Class IV
coarser
BALANCE
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 64 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
MECHANICAL-
Weighing Machine
WEIGHING Weighing Balance of
36 Readability =5 g & upto 50 kg 5.1g
SCALE AND Class III & Coarser
coarser
BALANCE
Humidity Sensor
with Indicator of Relative Humidity
THERMAL-
Environmental/condi Sensor with 10 %RH to 95 %RH
37 SPECIFIC HEAT 1.16%RH
tioning Indicator(Single @25 ºC
& HUMIDITY
Chamber/Humidity point Calibration)
chamber
Calibration of Deep
Freezer, Freezer,
Incubator,BOD
Using Data Logger
Incubator, Oven,
THERMAL- with RTD Sensors
38 Water bath, -80 ºC to 300 ºC 1.0ºC
TEMPERATURE (PT-100)by Multi
Temperature
Position Calibration
Furnace, Chamber,
Autoclave (Industrial
purpose only)
Calibration of Using Data Logger
THERMAL- Oven,Thermostat,Ch with K-Type Sensors
39 300 ºC to 1200 ºC 3.0ºC
TEMPERATURE amber,Temperature by Multi Position
Furnace Calibration
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 65 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Temperature
Indicator of Deep
Freezer, Freezer,
Incubator,BOD
RTD (PT-100) 4 wire
THERMAL- Incubator, Oven,
40 with Indicator by -80 °C to 300 °C 0.26°C
TEMPERATURE Water bath,
Comparision method
Temperature
Furnace, Chamber,
Autoclave (Industrial
purpose only)
Temperature R Type
THERMAL- Indicator of Dry Thermocouple with
41 50 °C to 600 °C 0.65°C
TEMPERATURE Block furnace , Indicator by
Muffle Furnace Comparision Method
Temperature R Type
THERMAL- Indicator of Dry Thermocouple with
42 600 °C to 1200 °C 1.88°C
TEMPERATURE Block furnace , Indicator by
Muffle Furnace Comparision method
Temperature
Transmitter, RTD,
Thermocouples with
& without RTD (PT-100) 4 wire
THERMAL- controller/Data with Indicator, Oil
43 50 °C to 300 °C 0.62°C
TEMPERATURE Logger/recorder, bath by Comparision
Temperature Gauge, method
Glass Thermometer,
Digital Thermometer
with inbuilt sensor
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 66 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Temperature
Transmitter, RTD,
Thermocouples with
R Type
& without
Thermocouple with
THERMAL- controller/Data
44 Indicator, Dry Block 600 °C to 1200 °C 1.88°C
TEMPERATURE Logger/recorder,
Furnace by
Temperature Gauge,
Comparision method
Glass Thermometer,
Digital Thermometer
with inbuilt sensor
Temperature
Transmitter, RTD,
Thermocouples with
& without
R Type
controller/Data
Thermocouple with
THERMAL- Logger/recorder,
45 Indicator, Dry Block 300 °C to 600 °C 1.5°C
TEMPERATURE Temperature Gauge,
Furnace by
Glass Thermometer,
Comparision method
Digital Thermometer
with inbuilt sensor,
Thermohygrometers
with Sensor
This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, NEW DELHI, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 67 of 67
Validity 14/02/2020 to 13/02/2022 Last Amended on 08/04/2020
Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument
Temperature
Transmitter, RTD,
Thermocouples with
& without RTD (PT-100) 4 wire
THERMAL- controller/Data with Indicator, Liquid
46 -30 °C to 50 °C 0.21°C
TEMPERATURE Logger/recorder, bath by Comparision
Temperature Gauge, method
Glass Thermometer,
Digital Thermometer
with inbuilt sensor.
* CMCs represent expanded uncertainties expressed at approximately the 95% level of confidence, using a coverage factor of k = 2.