Technician Manual: Electronic Laboratory Autoclaves Models 3840, 3850, 3870 ELVG-D
Technician Manual: Electronic Laboratory Autoclaves Models 3840, 3850, 3870 ELVG-D
1. Introduction 4
2. Installation Instructions 6
2.1 Installation Site 7
2.2 Connection of the pressure regulator 7
2.3 Front View 9
2.4 Front View for the Fan Configuration 10
2.5 Rear View 11
3. Tests 12
3.1 Installation Tests 12
3.2 Periodical Tests 12
4. Technical Data 14
4.1 Directives and Standards 14
4.2 Water Quality 14
4.3 Reverse Osmosis 15
4.4 Electrical data 16
4.5 Specifications 17
4.6 Overall Dimensions 18
4.7 Construction 18
4.8 Utilities 19
4.9 Symbol Description 20
5. Description of the Control System. 21
5.1 Hardware Control System Components: Block Diagram ................... 21
5.2 Application System Architecture 21
5.3 Interfaces to Users 22
5.4 Inputs and Outputs 22
5.5 Hardware Control System components 24
6. Checking and Changing Parameters and Other Data 28
6.1 Browsing through the menus 28
6.2 Changing a Parameter 29
6.3 Quick options screen 31
6.4 Logging in and entering the Main menu 37
6.5 Directories and subdirectories 38
6.6 System Parameters 41
6.7 Inputs/Outputs 43
6.8 Maintenance 49
6.9 Advanced Options 53
6.10 Version handling 59
Page 1
98
9. Maintenance and Replacement Procedures: the Generator 103
9.1 Access to Generator 103
9.2 Draining the generator 103
9.3 Cleaning and Replacing the Water Level Electrodes 104
9.4 Replacing the Heating Elements 106
Page 2
Page 3
PV
Page 4
98
9. Maintenance and Replacement Procedures: the Generator 103
9.1 Access to Generator 103
9.2 Draining the generator 103
9.3 Cleaning and Replacing the Water Level Electrodes 104
9.4 Replacing the Heating Elements 106
Page 2
Attention:
Connection of water system to the autoclave must be performed
through "BACK FLOW PREVENTION SYSTEM" installation as per
EN 1717.
Page 6
Page 7
FLOW DIRECTION
Page 8
6 10
5
7
8
11
4
9
12
2
13
1 14
Page 9
There is a configuration equipped with a fan that allows shortening the cooling
operation. The fan is located on the door of the autoclave.
Fan
Page 10
16
15
1
2
14
13
4
12
5
11
6 7 8 9 10
Page 11
a. Integrity Check
Perform a visual check to verify that there are no dents, scratches,
broken gauges, etc.
b Leveling Check
Check that the autoclave is leveled.
c. Leakage current test
Check the precise operation of the earth leakage relay.
d. Continuity Check
Check the continuity of the grounding connection.
At this stage operate the autoclave and continue with the tests:
e. Safety Check
Check the safety elements; safety valve and the door locking
mechanisms.
f. Programs Check
Run basic programs of the autoclave and check the operation
sequences, the sterilization parameters etc.
g. Validation
Validate the sterilization cycles, taking in consideration the interface of
packaging/goods/autoclave.
After the above steps are performed, the autoclave is ready for
operation.
Period Test
2 months Test the safety valve by operating it.
Remove the cover of the autoclave, tighten the, valves and
6 months
connectors in the control box.
Check the continuity of the grounding connections.
Year
Check the temperature and pressure calibration.
Page 12
Only an authorized technician shall perform the 6-months and yearly tests!
Page 13
Tuttnauer. Ltd. company also works in conjunction with and refers to:
Page 14
Note: The use of heavy scaled water for the vacuum pump cooling, can
cause blocking of the rotor and put the pump out of operation. This
invalidates the guarantee for the vacuum pump.
Page 15
FLOW DIRECTION
Page 8
Model
Properties
3840 3850 3870
Chamber diameter 380 380 380
Chamber depth 400 500 690
Chamber volume 52 liters. 65 liters. 85 liters.
Height 1000 1000 1000
Overall dimensions Width 730 730 730
Length 700 700 700
Maximum Width 800 800 800
dimensions
(door open) Length 1500 1500 1500
Page 17
4.7 Construction
Page 18
Switch Recommended
Power
box Circuit Breaker
208-220V (3 Ph + Earthing) No 20A
380-400V(3 Ph + Neutral +
No 16A
Earthing)
230V 1ph yes 32A
Frequency 50/60Hz
Compressed Air
1/2" 3-4 Bar (44-58 psi)
(ELVCPVG-D only)
Tap water 1/2", 2-3 Bar (29-44 psi)
Mineral free water 1/2", 2-3 Bar (29-44 psi)
2" Minimum Withstanding
Drain
temp. of 80°C (176°F)
Attention:
A switch or circuit-breaker must be included in the building installation.
This switch or circuit-breaker shall be in close proximity to the
equipment, within easy reach of the operator; and marked as the
disconnecting device for the equipment.
The electrical net must be protected with a current leakage safety
relay.
The electrical network must comply with local rules or regulations.
Verify that there is an easy access to the main power switch, to the
water cut-off valve and to the current leakage safety relay.
Make sure while placing the autoclave, to leave space around the
machine, to give the technician access to service the machine.
All water connections to autoclave must be performed through "BACK
FLOW PREVENTION SYSTEM" only, as per IEC 61770.
Page 19
On-Off
Page 20
Main board
Page 21
USB socket:
The USB socket is intended to load cycles' history from flash a
memory (disk on key). The received file is in txt format that can be
loaded onto a PC.
Analog inputs JP
Drain temperature J5
Chamber temperature J2
Bio hazard filter temperature J6
Ref temperature J3
Jacket temperature J4
Chamber pressure J7/1
Page 22
Digital Inputs JP
Door closed J12/1
Door unlocked J12/6
Door locked J12/8
Digital outputs
Digital Outputs JP
Steam to jacket J13/8
Vacuum pump J13/12
Exhaust to bio J13/14
Main exhaust J13/16
Atmospheric air valve J14/2
Compressed air to chamber J14/4
Slow exhaust J14/6
Fast exhaust J14/8
Vacuum valve J14/10
Cool Drain J14/12
Steam to chamber J14/14
Cooling water J14/16
Compressed air to jacket J15/2
Close door J15/4
Open door J15/6
Air to door J15/14
Actuators
The control system operates electrical valves (solenoids), Electric
cylinder motors, pneumatic cylinders, electric pumps, heaters.
On / off switch
A Rocker Switch 250V AC, 16A
Page 23
Main card
+ -
J5
BT1
P4
J8
T1
J3
P2
J4
P3
Page 24
IO Card
J1 J4
J12
Digital Inputs
Page 25
The IO card is controlled by the MAIN card. Only when requests are
received from the MAIN, the requests will be checked and performed.
The IO card will perform an electrical restart of the systems (IO card) if
it does not receive requests within 5 seconds. In the next request
received from the MAIN, the IO card will indicate to the requester of IO
card systems restart.
Page 26
Tuttnauer. Ltd. company also works in conjunction with and refers to:
Page 14
Now you will learn how to browse through the folders. When you read
the Directories and subdirectories chapter with links to specific menus,
you will need to know how to browse through the folders using the
autoclave control panel. Below is the instruction.
Login as Technician (see 6.4). The Main menu screen appears. To
browse through the menus:
Below are the example screens for the following menu: Cycle
Parameters\ Drying\ Dry Time:
Page 28
In the next chapter you will see how to change the required parameter
as desired.
You have browsed through the menus and reached the parameter
changing screen as explained above. Now you can change the
required parameter as desired. To do so:
1. Enter the required value as follows:
Press the Up and Down keys to change the value of the digit.
Press the Start/Stop key to move the cursor to the next digit to the left.
2. When finished, press the Start/Stop key repeatedly until you move the
cursor to Set.
3. Press the Up or Down key to confirm the new value and to exit the
parameter changing screen.
Page 29
Page 30
Page 31
his subdirectory allows you to export settings and cycles history to the
USB device.
To export settings:
Page 32
Page 33
Serial Software
number version
This subdirectory enables the operator to start the cycle at the time set
by this paramter.
1. Enter the Start cycle by clock screen. The following screen will appear:
Page 34
1. Exit the Enabling the Start Cycle by Clock. The start cycle by clock icon
appears on the display:
Page 35
On the Set date and time screen, the time is displayed in the upper row
in the form "HH:MM:SS". The hour range is 24 hour (i.e. from "0" to
"24"). The date is displayed in the lower row in the form "DD: MMM:
YYYY".
1. Set time and date
2. Exit the Set date and time screen. The following screen will appear:
Page 36
Below you can find instructions how to login and enter the Main menu.
Section 7.1 above explains how to browse through the menus, section
7.2 explains how to change a parameter.
Page 37
On-Off
Page 20
The following table lists the options that the technician may perform.
Directory Subdirectory
Temperature sensors
Displayed inputs
Purge (generator models only)
Drying
Create Pulse
Keep Heat
Exhaust
Drying
Ending
Global
Print Rate All
Print Rate Sterilization
Screen Saver
Pressure calibration high
System Parameters
Pressure calibration low
Temperature calibration high
Temperature calibration low
Cycle Print Gap
View digital inputs state
Inputs/Outputs View digital outputs state
Test digital outputs
1
See cycle parameters in more detail in Cycle Parameters.
Page 39
Page 40
This menu is listing the system parameters that are the same for all
cycles. Browse to the following folder:
In this menu you can define the time interval for printing out the cycle
status, for all the stages except sterilization: The default is 3 minutes.
See the Printer handling section of the Operation and maintenance
manual.
Page 41
In this menu you can define the time interval for printing out the cycle
status during the sterilization stage: The default is 1 minutes.
Browse to the following folder:
System parameters\Print Rate Sterilization
Change the parameter as desired
In this menu you can define the screensaver delay time, i. e. how long
the keyboard will be untouched before the screensaver activates.
Browse to the following folder:
System parameters\Screen Saver
Change the parameter as desired.
In this menu you can define the high calibration point for pressure.
Browse to the following folder:
System parameters\Pressure calibration high
Change the parameter as desired.
In this menu you can define the low calibration point for pressure.
Browse to the following folder:
System parameters\Pressure calibration low
Change the parameter as desired.
In this menu you can define the high calibration point for temperature.
Browse to the following folder:
System parameters\Temperature calibration high
Change the parameter as desired
Page 42
In this menu you can define the low calibration point for temperature..
Browse to the following folder:
System parameters\Temperature calibration low
Change the parameter as desired
This parameter defines the time interval between printing the current
values of the cycle (See the printer output section).
6.7 Inputs/Outputs
It is important for the technician to control the system down to the level
of specific inputs and outputs. Viewing the outputs state at specific
stages of the cycle helps you diagnose the malfunctions. You can also
test and calibrate inputs and outputs.
Below is the typical screen for viewing/testing the inputs/outputs.
In this menu you can view, at any stage, which digital inputs are being
controlled at the moment.
Page 43
In this menu you can view, at any stage, which digital outputs are being
controlled at the moment.
Browse) to the following folder:
Inputs/Outputs\View digital outputs state
In this menu you can view the normal function of all the autoclave
elements: valves, heating elements, pumps, etc.
Browse to the following folder:
Inputs/Outputs\Test digital outputs
Note: on the screen that appears, scroll down for more outputs.
Check the checkboxes of the outputs you want to test.
In this menu you can view, at any stage, the values of the analog
inputs at the moment.
Browse to the following folder:
Inputs/Outputs\View analog inputs state
In this menu you can calibrate water level, mineral water level,
pressure, and temperature sensors of the autoclave chamber.
1. Browse to the following folder:
Inputs/Outputs\Analog inputs calibration
The following screen will appear:
Page 44
Page 45
2. In the Actual field, change the high and low values of the input as
desired.
Page 46
There is an option to restore the gain and offset values set at previous
calibration.
1. On the calibration options screen, choose Restore last calibration.
2. Using the up and down keys, move your cursor to confirm and press
Start/Stop.
Page 47
There is an option to restore the default (factory) gain and offset values
1. On the calibration options screen, choose Restore last calibration.
2. Using the up and down keys, move your cursor to confirm and press
Start/Stop.
Page 48
The IO card is controlled by the MAIN card. Only when requests are
received from the MAIN, the requests will be checked and performed.
The IO card will perform an electrical restart of the systems (IO card) if
it does not receive requests within 5 seconds. In the next request
received from the MAIN, the IO card will indicate to the requester of IO
card systems restart.
Page 26
In this menu you can import from the USB the gain and offset
documents you have got as a result of calibration to the autoclave.
Browse to the following folder:
Maintenance\Import gain and offset from USB
Press Start/Stop. The following screen will appear:
In this menu you can reset the atmospheric pressure value. To do so:
1. Browse to the following folder:
Maintenance\Reset atmospheric pressure
The following screen will appear:
Page 50
In this menu you can check the two clocks. The two clocks work during
the interval set by the Test RTC parameter, and the results are
displayed. If the results are different for more than 2 seconds, one of
the clocks is faulty.
1. Browse to the following folder:
Maintenance\Test RTC
The following screen will appear:
2. Set the time interval as desired. Then move the cursor to start and
press Start/Stop.
Page 51
In this menu you can check the normal function of the printer. The
printer will print the list of errors.
Browse to the following folder:
Maintenance\Printer test
The following screen will appear to confirm that the test has been done.
See the printout shown in the Printer handling chapter of the user
manual.
In this menu you can print all the gain and offset for all the pressures
and temperatures in the autoclave (Chamber Temperature, Chamber
Pressure, Chamber Water Level, Mineral Free Water Level.
Page 52
Chamber Temperature
G:000.0385;O:-008.0976
Chamber Pressure
G:000.1238;O:-098.5985
Chamber Water Level
G:001.0000;O:000.0000
Mineral Free Water Level
G:001.0000;O:000.0000
Page 53
In this menu you can enable only the cycles you want to use. The
cycles you do not want to use will not appear on the screen.
Browse to the following folder:
Advanced Options\Enable cycles
In this menu you can choose the language for all the interface screens:
menus, cycle information, system messages.
1. Browse to the following folder:
Advanced Options\Set Language
Page 54
The machine will be restarted and the main screen (current cycle) will
appear in the chosen language.
In this menu you can set the temperature units (Celsius or Fahrenheit)
for screens and printouts.
Browse to the following folder:
Advanced Options\Set temperature units
Change the parameter as desired
Page 55
In this menu you can you can set the pressure units (kPa, Psia, Psig,
BarA, BarG) for screens and printouts.
Browse to the following folder:
Advanced Options\Set pressure units
Change the parameter as desired
In this menu you can create a copy of one of the cycles with all its
parameters (the parameters can be changed then). You can give any
name to the new custom cycle.
Browse to the following folder:
Advanced Options\Duplicate cycles
1. Using the Up and Down keys, move the cursor to the cycle you want to
duplicate, then press Start/Stop to select the cycle. A custom program
has been created with the same settings. You will be able to change
the settings later. The following screen will appear:
Page 56
Page 57
In this menu you can delete the custom cycles that were created.
Browse to the following folder:
Advanced Options\Delete custom cycles
1. Move the cursor to the cycle you wish to delete. Press Start/Stop key to
select/deselect the cycles.
2. Move the cursor to Delete selected cycles and press Start/Stop key to
delete.
The following screen will appear:
Page 58
In this menu you can replace the autoclave software with the
application software from the USB device to the autoclave.
1. Browse to the following folder:
Version handling\Import application from USB
The system will prompt you to confirm import
2. Move the cursor to Confirm and press Start/Stop.
The application will be imported from the USB device.
Page 59
his subdirectory allows you to export settings and cycles history to the
USB device.
To export settings:
Page 32
In this menu you can replace both the application software and the
parameters of the autoclave with the software and parameters from the
USB device.
4. Browse to the following folder:
Version handling\Import application and setting from USB
5. Move the cursor to Confirm and press Start/Stop.
In this menu you can restore the default values of all the changeable
parameters of the cycle and the system. Default settings are those your
autoclave had when it left the factory.
1. Browse to the following folder:
Version handling\Return to factory default settings
2. Move the cursor to Confirm and press Start/Stop. The default factory
settings will be restored.
Page 61
Subdirectory Property
Temperature sensors Chamber Temperature
First
Displayed inputs Second
Third
Purge time
Purge (generator models only)
Purge temperature
Pulse A Count
Pulse A Stay Time
Pulse A Low Pressure
For custom programs only
Page 62
Sterilization Temperature
Sterilization
Sterilization Time
Cooling (“C” Cool Mode
mod
Cool End Temperature
els
only) Cool Exhaust Rate
Exhaust Exhaust Mode
Dry Time
Dry Heat On 1
For custom programs only
Check RTC
Jacket Temperature
Page 63
This menu defines 2 (or 3, according to the cycle type) input values to
be shown on the display.
You can choose, for each of them: chamber temperature, chamber
pressure, chamber water level, and mineral free water level. Below is
the example of setting the first input.
1. Browse to the following folder:
Cycle parameters\Displayed inputs\First
The following screen will appear:
Page 64
This menu allows to set parameters for each pulse of the program (see
below).
Page 65
This parameter defines how many times the pulse of each type (A, B)
is repeated. For example, if the Pulse count for pulse A (low) is 4, and
the Pulse count for pulse B (high) is also 4, the cycle will have the
following pulses: low-high-low-high-low-high-low-high.
1. Browse (to the following folder:
Cycle parameters\Create Pulse\Pulse A Count
Change the parameter as desired.
Note: Pulse A and B counts for standard programs equal one, with the
exception of Hollow load, Waste, and Bowie and Dick test (pre-vacuum
models only), whose pulse count equals four.
6.15.2 Pulse A Stay Time
This parameter defines the value of delay after the required pressure
for the pulse is reached.
1. Browse to the following folder:
Cycle parameters\Create Pulse\Pulse A Stay Time
Change the parameter as desired.
Note: Same parameters are available for other pulses. The following cycles
have four pulses pre-vacuum models only): Hollow load, Waste, and
Bowie and Dick test. Other cycles have only one pulse.
6.15.3 Pulse A Low Pressure
This parameter defines the lower pressure value for the A pulse.
1. Browse to the following folder:
Cycle parameters\Create Pulse\Pulse A Low Pressure
2. Change the parameter as desired.
Note: Same parameters are available for other pulses. The following cycles
have four pulses pre-vacuum models only): Hollow load, Waste, and
Bowie and Dick test. Other cycles have only one pulse.
Page 66
This parameter defines the higher pressure value for the A pulse
Browse to the following folder:
Cycle parameters\Create Pulse\Pulse A High Pressure
Change the parameter as desired.
Note: Same parameters are available for other pulses. The following cycles
have four pulses pre-vacuum models only): Hollow load, Waste, and
Bowie and Dick test. Other cycles have only one pulse.
6.16 Keep Heat
There are two optional stages, within the sterilization cycle, when constant
temperature (Temperature 1 or 2 stay) is maintained for certain time
period (Temperature 1 or 2 stay time). Keep Heat starts before heating
and is used to melt solid loads of substances such as agar. Keep Heat
is needed for such loads to become completely liquid before
sterilization starts.
Below is the example of changing the Temperature 1 stay/stay time.
This parameter allows you to set two temperatures at which the process will
be paused (Keep Heat). It is possible to set the 2 values (Temp 1 stay
and Temp 2 stay).
Page 67
6.17 Heating
6.18 Sterilization
Page 68
This parameter defines duration of the sterilization stage for the cycle.
3. Browse to the following folder:
Cycle parameters\ Sterilization\Sterilization Time
4. Change the parameter as desired.
If the setup of your model has the cooling option, and the cycle has the
cooling stage, the Cool mode parameter equals 1, otherwise it equals
0. You can set this parameter as 0 or 1 therefore enabling or disabling
the cooling option.
1. Browse to the following folder:
Cycle parameters\Cooling\Cool Mode
2. Change the parameter as desired.
Page 69
Below you can find instructions how to login and enter the Main menu.
Section 7.1 above explains how to browse through the menus, section
7.2 explains how to change a parameter.
Page 37
This parameter defines duration of the drying stage for the cycle.
1. Browse to the following folder:
Cycle parameters\Drying\Dry Time
2. Change the parameter as desired.
Drying stage is divided into two stages. For each stage you can set the
total time, on time and off time. So during the Dry First Stage, the
heating elements are on for the time set in Dry Heat On 1, then they go
off for the time set in Dry Heat Off 1, and then this sequence is
repeated during the entire Dry First Stage Time. The same is true for
the Dry Second Stage.
This parameter defines how long will the heating elements be on for
the Dry First Stage.
Browse to the following folder:
Cycle parameters\Drying\Dry Heat On 1
Change the parameter as desired.
Note: There are Dry heat On 2 and Dry heat Off 2 parameters for the Dry
Second Stage of the drying.
This parameter defines how long will the heating elements be off for
the Dry First Stage.
1. Browse to the following folder:
Page 71
Note: There are Dry heat On 2 and Dry heat Off 2 parameters for the Dry
Second Stage of the drying.
This parameter defines duration of the Dry First Stage. The Dry
Second Stage time is calculated as Dry Time minus the Dry First Stage
Time, that is why there is no Dry Second Stage Time menu.
This subdirectory allows you to change the Dry Time parameter for the
current cycle.
Browse to the following folder:
Cycle parameters\Drying\Add Dry Time
Change the parameter as desired.
6.22 Ending
This menu defines the End temperature parameter for the Ending
stage.
This parameter defines the temperature when the cycle ends. The
cycle will not end and the door will not open unless the autoclave
chamber is cooled down to this temperature.
1. Browse to the following folder:
Cycle parameters\Ending\End Temperature
2. Change the parameter as desired
Page 72
This parameter checks how long has the load been in sterilization
conditions (pressure, temperature) before the sterilization stage
actually begins. This data helps preventing damage to the load by
excessive exposure to heat and high pressure.
1. Browse to the following folder:
Cycle parameters\Global\F0 mode
2. Change the parameter as desired
In this menu you can set the time interval for checking the two clocks.
The two clocks work during this interval, and the results are displayed.
If the results are different for more than 2 seconds, one of the clocks is
faulty. For the procedure of the check itself, see Maintenance\Test
RTC.
1. Browse to the following folder:
Cycle parameters\Global\Check RTC
2. Change the parameter as desired).
This parameter allows you to repeat, for testing purposes, the same
cycle several times without reloading the autoclave. The number or
repetitions is defined by this parameter.
1. Browse to the following folder:
Cycle parameters\Global\Multiple Cycles
2. Set the parameter as desired.
Page 73
This parameter allows you set the time interval between the multiple
cycles (see above).
1. Browse to the following folder:
Cycle parameters\Global\Multiple Cycles Gap
2. Change the parameter as desired.
This parameter defines the temperature of the jacket (for the relevant
models only). The system is not ready unless the jacket reaches this
temperature.
1. Browse to the following folder:
Cycle parameters\Global\Jacket Temperature
2. Change the parameter as desired
Page 74
Caution!
It is strictly forbidden to use soft water; this may results in
corrosion of the generator.
Page 75
The fault light on the generator box will light up, and the main heater
circuit breakers will be switched OFF.
If this happens, do not try to correct the fault yourself, but call for
assistance.
Page 76
The water level control system consists of three electrodes. (HIGH and
LOW water level), and a reference electrode, all connected to the
GICAR 2DRLHG controller.
Page 77
Page 78
Warning!
Do not touch door microswitches in case of malfunction! Call
Tuttnauer service.
6. A Vacuum Test
7. A B&D Test.
By a qualified technician
Page 79
Once in 5 years
Check the door closing system for excessive wear.
Performing safety tests: pressure vessel, efficiency, electrical,
according to local rules or regulations.
safety tests shall be perform only, by an authorized inspector.
Caution!
Before starting, be sure that the electric cord is disconnected from the
power source and that there is no pressure in the autoclave generator.
Allow the autoclave to cool for about an hour before removing outer
covers.
The circuit breaker is located on the rear of the autoclave (See Rear
View).
8. Disconnect the wires (1, 2) from the circuit breaker.
Page 80
In this menu you can define the time interval for printing out the cycle
status during the sterilization stage: The default is 1 minutes.
Browse to the following folder:
System parameters\Print Rate Sterilization
Change the parameter as desired
In this menu you can define the screensaver delay time, i. e. how long
the keyboard will be untouched before the screensaver activates.
Browse to the following folder:
System parameters\Screen Saver
Change the parameter as desired.
In this menu you can define the high calibration point for pressure.
Browse to the following folder:
System parameters\Pressure calibration high
Change the parameter as desired.
In this menu you can define the low calibration point for pressure.
Browse to the following folder:
System parameters\Pressure calibration low
Change the parameter as desired.
In this menu you can define the high calibration point for temperature.
Browse to the following folder:
System parameters\Temperature calibration high
Change the parameter as desired
Page 42
3
4
2
5
5 Rear wall
Rear wall
6 7 7 8 6 7 7 8
ASME PED
Page 82
Caution!
Before starting, disconnect the instrument from the power
source and ensure that there is no pressure in the chamber or
the generator.
Allow the autoclave to cool before opening the service doors.
1. Pull off the gasket from the door groove and install the new gasket
referring to the directions below.
2. Remove the door gasket and clean the groove (plain water is okay).
3. Line the inside of both sides/walls of the gasket groove with a small
amount of silicon lubricant. This can be sprayed in or brushed in
depending on the type lubricant you use.
Page 83
Caution!
If insufficient lubricant is applied, replacing the gasket will be
difficult. If excess lubricant is applied, the gasket will 'spring' out
of the groove.
Caution!
Make sure the gasket sits evenly
without waves, humps, or cavities.
This drawing shows the correct
direction of the gasket.
Gasket
Door
Caution!
Page 84
2
1
1 2
2. With a screwdriver, inserted through the holes in the rear of the door
cover (3), release the 4 screws fastening the door stoppers. Be careful
not to disassemble the stopper.
Page 85
16
5. Unscrew the screws assembling the axis bridge (11) to the door
bracket (5).
11
1. Assemble door bracket (5) to the door cover (4) with 10 screws (6).
Page 86
2. Assemble the axis bridge (10) to the axis seat (7) with a pin (9) and
cotter pin (8).
10
3. Place the door cover (4) on the door (12) so that the axis bridge (11)
will cover the door bracket (5) and assemble the door to the axis bridge
with screws.
Page 87
11
12
4. Lift the spring holder (13) and place the spring holder support P/N
CMT387-0148 (14, supplied with the autoclave), as shown in the
picture below. Perform this on the right and on the left spring holder.
13
14
5. Lift the door and insert the edge of the spring holder (16) into the slot
(15).
Page 88
16
6. After the edge of the spring holder is in the slot the door cover is
attached to the door. At this stage remove the spring holder supports
(14).
7. Close the door.
8. With a screwdriver, inserted through the holes in the rear of the door
cover (3), fasten the 4 screws of the door stoppers.
Caution
Before starting, disconnect the instrument from the power source
and ensure that there is no pressure in the autoclave.
Allow the autoclave to cool before removing the outer cover.
Page 89
2. Turn the panel to reveal the internal part and the electrical connectors.
3. Unplug the board connectors (3) and the printer connectors (2). Verify
that you know where plug each is connected.
3 2
4. The printer is fastened (6) to its seat with two fasteners (5). To release
the fasteners, press slightly on the part marked (4) and pull the fattener
upward. Repeat this with the second fastener.
5. Pull the printer out of its seat (the hole in the panel).
6. Replace the printer with a new printer.
7. Inset the fasteners. Verify that they are inserted firmly.
8. Connect the printer's connectors
9. Connect all the connectors of the board.
Page 90
Warning!
Do not touch door micro-switches in case of malfunction! Call
Tuttnauer service.
Page 91
There is an option to restore the gain and offset values set at previous
calibration.
1. On the calibration options screen, choose Restore last calibration.
2. Using the up and down keys, move your cursor to confirm and press
Start/Stop.
Page 47
Caution!
Before starting, be sure that the electric cord is disconnected
from the power source and that there is no pressure in the
autoclave generator.
Allow the autoclave to cool for about an hour before removing
outer covers.
Note: The locking motor can be replaced with an open door or with a closed
door.
Page 93
4
8
5
7 6
Page 94
Caution!
Before starting, disconnect the instrument from the power source
and ensure that there is no pressure in the chamber.
Allow the autoclave to cool before opening the service doors.
1. Unlock and remove the left service door.
2. Remove nut (1).
3. Remove the coil (2).
4. Unscrew the plunger + spring (3) with the plunger housing (4) and
replace it with a new one. The plunger and the spring are replaced as a
kit.
5. Reassemble the coil (2) and the nut (1).
6. Close the autoclave service door.
3 4
Page 95
The solenoid valves may be out of order due to faulty plunger or coil.
To repair the solenoid valve – replace the faulty plunger or solenoid
Caution!
Before starting, disconnect the instrument from the power source
and ensure that there is no pressure in the chamber, coil or
generator.
Allow the autoclave to cool before removing the outer cover.
1. Open the relevant service door to reveal the faulty BACCARA solenoid
valve.
2. Unscrew and remove nut (1).
3. Remove the coil (2).
4. Unscrew the plunger + spring (4) with the plunger housing (3) and
replace it with a new one. The plunger the spring and the housing are
replaced as a kit.
5. Reassemble the coil (2) and the nut (1).
6. Re-install the autoclave cover.
5 4 3 2 1
Caution!
Before starting, ensure that there is no pressure in the chamber,
coil or generator.
Allow the autoclave to cool before removing the outer cover.
Page 96
Caution!
Before starting, be sure that the electric cord is disconnected from the
power source and that there is no pressure in the autoclave generator.
Allow the autoclave to cool for about an hour before removing outer
covers.
1. Open left service door.
2. Disconnect the electrical connections (1).
3. Unscrew and remove the pressure switch (2) from its seat (3).
4. Replace the pressure switch with a new one, use only an original
pressure switch supplied by Tuttnauer.
5. Assemble the pressure switch (2) to its seat (3).
6. Connect the electrical connections (1).
7. Close the left service door.
Page 97
Caution!
Before starting, be sure that the electric cord is disconnected
from the power source and that there is no pressure in the
autoclave generator.
Allow the autoclave to cool for about an hour before removing
outer covers.
For description of items – see table below.
Page 98
1 2 3 4 5
5 4 6 7
7
Description of Items
Page 99
Terminal, Feed-Through, 2-
5 THE039-0091 Conductor, Fuse Disconnect 1
1/4X1/4, 4mm²
Page 100
Description of Items
Page 101
Page 102
In this menu you can check the normal function of the printer. The
printer will print the list of errors.
Browse to the following folder:
Maintenance\Printer test
The following screen will appear to confirm that the test has been done.
See the printout shown in the Printer handling chapter of the user
manual.
In this menu you can print all the gain and offset for all the pressures
and temperatures in the autoclave (Chamber Temperature, Chamber
Pressure, Chamber Water Level, Mineral Free Water Level.
Page 52
Caution!
Before starting:
Make sure that the electric cord is disconnected and that there is
no pressure in the autoclave.
Make sure there is no pressure in the generator. If there is, open
the right service door and reduce the generator pressure by
opening the drain valve.
The three electrodes of the water level control system are located on
top of the steam generator.
1. Unlock and remove the right service door.
2. Unscrew the big nut (1).
3. Mark which wire belongs to which electrode and remove the three
electrodes (2) covered with a protecting sleeve (3).
4. Clean (only the tip) with soft cloth, or replace the electrodes.
5. Reassemble the three electrodes into their places.
6. Tighten the nuts to avoid leakage.
7. Verify that the electrodes height (20 mm) did not change (see detail A
on the figure below). If there was any change in the electrodes height,
loosen the small nut (4), adjust the height and tighten the nut.
8. Cover the electrode with the protecting sleeve (3).
9. Test the unit.
Page 104
Reference
Heater
Page 105
Caution!
Before starting:
Make sure that the electric cord is disconnected and that there is
no pressure in the autoclave.
Make sure there is no pressure in the generator. If there is, open
the left service door and reduce the generator pressure by
opening the drain valve.
Caution!
Before starting:
Make sure that the electric cord is disconnected and that there is
no pressure in the autoclave.
Page 106
Warning!
Do not exceed the recommended pressure of the gauge glass.
Caution!
Do not subject gauge glass to bending or torsional stresses.
Note: On each side of the glass tube, there are stainless steel balls that act
as a safety device by preventing the escape of steam and boiling water
if the glass is broken. Before replacing the glass water tube, switch
OFF the electrical power and release steam pressure from the vessel.
Page 107
Caution!
Before starting:
Make sure that the electric cord is disconnected and that there is
no pressure in the autoclave.
Make sure there is no pressure in the generator. If there is, open
the right service door and reduce the generator pressure by
opening the drain valve.
This thermostat cuts out the power to the generator in the event that all
other safety systems malfunction and the temperature continues to
rise. In order to restart the operation, turn trip coil of the generator (to
the right of the electric board) again (see below).
Page 108
Page 109
Caution!
Before starting:
Make sure that the electric cord is disconnected and that there is
no pressure in the autoclave.
Release pressure on the outlet side of the pump before doing any
maintenance work.
Check the pump for any sign of damage. If the pump has any indication
of failure, contact Tuttnauer for a complete motor assembly. Proceed
as follows to replace the pump.
1 2
Page 110
1. When pump is operated for the first time, check if it works normally. If
not, stop the electric current immediately.
2. Do not open the pump and by no means, try to free the rotor shaft, by
forcibly rotating the impeller, which can provoke an irreversible damage
to the pump.
3. Further described the way to proceed, for restoring the normal
operation of the pump.
4. Carefully remove the fan cover from the vacuum pump.
Page 111
Page 112
Page 113
Page 57
12. The door 12.1 Low pressure at the 12.1 Set the air pressure;
does not air supply line. 1.5 bar
close.
12.2 CLOSE DOOR 12.2 Fix or replace
(Applicable to solenoid valve solenoid valve.
sliding doors) transmitting air
pressure to the door
actuator is stuck or
damaged.
12.3 Door drive actuator 12.3 Fix or replace
damaged. actuator.
12.4 OPEN DOOR button 12.4 Replace the button.
is defective.
12.5 Control system 12.5 Replace the control
trouble. boards.
13. The door 13.1 Low pressure at the 13.1 Set the air pressure;
does not seal. air supply. 1.5 bar
(Applicable to 13.2 Low steam pressure 13.2 Set the steam
Automatic and at the door gasket. pressure; 1.5 - bar.
Page 115
Page 116
Page 117
Page 118
Page 119
Page 120
Page 121
Page 122
Page 123
Page 124
Subdirectory Property
Temperature sensors Chamber Temperature
First
Displayed inputs Second
Third
Purge time
Purge (generator models only)
Purge temperature
Pulse A Count
Pulse A Stay Time
Pulse A Low Pressure
For custom programs only
Page 62
Page 126
Page 127
Page 128
104.3 117.9 1.18 219.7 2.4 17.1 45.8 10 0.10 114.5 2.95 1.5
104.4 118.6 1.18 219.9 2.5 17.2 54.1 15 0.15 129.3 4.44 2.2
104.5 118.6 1.19 220.1 2.5 17.2 60.1 20 0.20 140.2 5.90 2.9
104.6 119.3 1.19 220.3 2.6 17.3 65.0 25 0.25 149.1 7.39 3.6
104.7 120.0 1.20 220.5 2.7 17.4 68.9 30 0.30 156.4 8.86 4.4
104.8 120.0 1.20 220.6 2.7 17.4 72.7 35 0.35 162.9 10.34 5.1
104.9 120.4 1.20 220.8 2.8 17.5 75.9 40 0.40 168.6 11.81 5.8
105.0 120.7 1.21 221.0 2.8 17.5 78.8 45 0.45 173.8 13.30 6.5
105.1 121.3 1.21 221.2 2.9 17.6 81.3 50 0.50 178.4 14.76 7.3
105.3 122.0 1.22 221.5 3.0 17.7 °C kPa Bar °F psig psia
105.4 122.7 1.23 221.7 3.1 17.8 100.0 101.3 1.01 212.0 0.0 14.7
105.5 122.7 1.23 221.9 3.1 17.8 100.1 101.7 1.02 212.2 0.1 14.8
105.6 123.4 1.23 222.1 3.2 17.9 100.2 102.1 1.02 212.4 0.1 14.8
105.7 124.1 1.24 222.3 3.3 18.0 100.3 102.4 1.02 212.5 0.2 14.9
105.8 124.1 1.24 222.4 3.3 18.0 100.4 102.8 1.03 212.7 0.2 14.9
105.9 124.7 1.24 222.6 3.4 18.1 100.5 103.2 1.03 212.9 0.3 15.0
106.0 125.1 1.25 222.8 3.5 18.2 100.6 103.6 1.04 213.1 0.3 15.0
106.1 125.5 1.26 223.0 3.5 18.2 100.7 104.0 1.04 213.3 0.4 15.1
106.2 126.0 1.26 223.2 3.6 18.3 100.8 104.3 1.04 213.4 0.4 15.1
106.3 126.2 1.26 223.3 3.6 18.3 100.9 104.7 1.05 213.6 0.5 15.2
106.4 126.8 1.27 223.5 3.7 18.4 101.0 105.1 1.05 213.8 0.5 15.2
106.5 127.2 1.27 223.7 3.8 18.5 101.1 105.4 1.05 214.0 0.6 15.3
106.6 127.7 1.28 223.9 3.8 18.5 101.2 105.8 1.06 214.2 0.7 15.4
106.7 128.1 1.28 224.1 3.9 18.6 101.3 106.2 1.06 214.3 0.7 15.4
106.8 128.5 1.29 224.2 3.9 18.6 101.4 106.6 1.07 214.5 0.8 15.5
Page 129
106.9 129.0 1.29 224.4 4.0 18.7 101.5 106.9 1.07 214.7 0.8 15.5
107.0 129.6 1.29 224.6 4.1 18.8 101.6 107.3 1.07 214.9 0.9 15.6
107.1 129.9 1.30 224.8 4.2 18.9 101.7 107.7 1.08 215.1 0.9 15.6
107.2 130.4 1.30 225.0 4.2 18.9 101.8 108.1 1.08 215.2 1.0 15.7
107.3 130.8 1.31 225.1 4.3 19.0 101.9 108.4 1.08 215.4 1.0 15.7
107.4 131.3 1.31 225.3 4.3 19.0 102.0 108.8 1.09 215.6 1.1 15.8
107.5 131.7 1.32 225.5 4.4 19.1 102.1 109.2 1.09 215.8 1.1 15.8
107.6 132.2 1.32 225.7 4.5 19.2 102.2 109.6 1.10 216.0 1.2 15.9
107.7 132.6 1.33 225.9 4.6 19.3 102.4 110.0 1.10 216.3 1.3 16.0
107.8 133.1 1.33 226.0 4.6 19.3 102.5 110.7 1.11 216.5 1.4 16.1
107.9 133.5 1.34 226.2 4.7 19.4 102.6 111.1 1.11 216.7 1.4 16.1
108.0 134.0 1.34 226.4 4.7 19.4 102.7 111.5 1.12 216.9 1.5 16.2
108.1 134.4 1.34 226.6 4.8 19.5 102.8 111.9 1.12 217.0 1.5 16.2
108.2 134.9 1.35 226.8 4.9 19.6 102.9 112.3 1.12 217.2 1.6 16.3
108.3 135.3 1.35 226.9 4.9 19.6 103.0 112.7 1.13 217.4 1.7 16.4
108.4 135.8 1.36 227.1 5.0 19.7 103.1 113.1 1.13 217.6 1.7 16.4
108.5 136.2 1.36 227.3 5.1 19.8 103.2 113.5 1.14 217.8 1.8 16.5
108.6 136.7 1.37 227.5 5.1 19.8 103.3 114.0 1.14 217.9 1.8 16.5
108.7 137.1 1.37 227.7 5.2 19.9 103.4 114.3 1.14 218.1 1.9 16.6
108.8 137.6 1.38 227.8 5.2 19.9 103.5 114.7 1.15 218.3 1.9 16.6
108.9 138.1 1.38 228.0 5.3 20.0 103.6 115.1 1.15 218.5 2.0 16.7
109.0 138.5 1.39 228.2 5.4 20.1 103.7 115.6 1.16 218.7 2.1 16.8
109.1 139.0 1.39 228.4 5.5 20.2 103.8 116.0 1.16 218.8 2.1 16.8
109.2 139.5 1.39 228.6 5.6 20.3 103.9 116.3 1.16 219.0 2.2 16.9
109.3 140.0 1.40 228.7 5.6 20.3 104.0 116.7 1.17 219.2 2.2 16.9
109.4 140.5 1.40 228.9 5.7 20.4 104.1 117.1 1.17 219.4 2.3 17.0
109.5 140.9 1.41 229.1 5.7 20.4 104.2 117.5 1.18 219.6 2.4 17.1
115.1 169.7 1.70 239.2 9.9 24.6 109.6 141.4 1.41 229.3 5.8 20.5
115.2 170.2 1.70 239.4 10.0 24.7 109.7 142.0 1.42 229.5 5.9 20.6
Page 130
115.3 170.8 1.71 239.5 10.0 24.7 109.8 142.4 1.42 229.6 5.9 20.6
115.4 171.3 1.71 239.7 10.1 24.8 109.9 142.9 1.43 229.8 6.0 20.7
115.5 171.8 1.72 239.9 10.2 24.9 110.0 143.3 1.43 230.0 6.1 20.8
115.6 172.4 1.72 240.1 10.3 25.0 110.1 143.9 1.44 230.2 6.2 20.9
115.7 173.1 1.73 240.3 10.4 25.1 110.2 144.3 1.44 230.4 6.3 21.0
115.8 173.6 1.74 240.4 10.5 25.2 110.3 144.8 1.45 230.5 6.3 21.0
115.9 174.1 1.74 240.6 10.6 25.3 110.4 145.3 1.45 230.7 6.4 21.1
116.0 174.7 1.75 240.8 10.6 25.3 110.5 145.8 1.46 230.9 6.4 21.1
116.1 175.3 1.75 241.0 10.7 25.4 110.6 146.2 1.46 231.1 6.5 21.2
116.2 175.9 1.76 241.2 10.8 25.5 110.7 146.7 1.47 231.3 6.6 21.3
116.3 176.4 1.76 241.3 10.9 25.6 110.8 147.2 1.47 231.4 6.6 21.3
116.4 177.0 1.77 241.5 11.0 25.7 110.9 147.7 1.48 231.6 6.7 21.4
116.5 177.6 1.78 241.7 11.1 25.8 111.0 148.2 1.48 231.8 6.8 21.5
116.6 178.2 1.78 241.9 11.2 25.9 111.1 148.6 1.49 232.0 6.9 21.6
116.7 178.7 1.79 242.1 11.2 25.9 111.2 149.6 1.49 232.2 7.0 21.7
116.8 179.3 1.79 242.2 11.3 26.0 111.3 149.6 1.50 232.3 7.0 21.7
116.9 180.0 1.80 242.4 11.4 26.1 111.4 150.3 1.50 232.5 7.1 21.8
117.0 180.5 1.80 242.6 11.5 26.2 111.5 151.0 1.51 232.7 7.2 21.9
117.1 181.1 1.81 242.8 11.6 26.3 111.6 151.0 1.51 232.9 7.2 21.9
117.2 181.6 1.82 243.0 11.7 26.4 111.7 151.7 1.52 233.1 7.3 22.0
117.3 182.2 1.82 243.1 11.7 26.4 111.8 152.2 1.52 233.2 7.4 22.1
117.4 182.8 1.83 243.3 11.8 26.5 111.9 152.7 1.53 233.4 7.4 22.1
117.5 183.4 1.83 243.5 11.9 26.6 112.0 153.2 1.53 233.6 7.5 22.2
117.6 184.0 1.84 243.7 12.0 26.7 112.1 153.8 1.54 233.8 7.6 22.3
Page 131
117.7 184.5 1.85 243.9 12.1 26.8 112.2 154.3 1.54 234.0 7.7 22.4
117.8 185.1 1.85 244.0 12.1 26.8 112.3 154.8 1.55 234.1 7.7 22.4
117.9 185.7 1.86 244.2 12.2 26.9 112.4 155.3 1.55 234.3 7.8 22.5
118.0 186.3 1.86 244.4 12.3 27.0 112.5 155.8 1.56 234.5 7.9 22.6
118.1 186.9 1.87 244.6 12.4 27.1 112.6 156.3 1.56 234.7 8.0 22.7
118.2 187.5 1.88 244.8 12.5 27.2 112.7 156.8 1.57 234.9 8.1 22.8
118.3 188.2 1.88 244.9 12.6 27.3 112.8 157.3 1.57 235.0 8.1 22.8
118.4 188.8 1.89 245.1 12.7 27.4 112.9 157.9 1.58 235.2 8.2 22.9
118.5 189.4 1.89 245.3 12.8 27.5 113.0 158.4 1.58 235.4 8.3 23.0
118.6 190.0 1.90 245.5 12.9 27.6 113.1 158.9 1.59 235.6 8.4 23.1
118.7 190.6 1.91 245.7 13.0 27.7 113.2 159.4 1.59 235.8 8.4 23.1
118.8 191.2 1.91 245.8 13.0 27.7 113.3 159.9 1.60 235.9 8.5 23.2
118.9 191.8 1.92 246.0 13.1 27.8 113.4 160.4 1.60 236.1 8.6 23.3
119.0 192.4 1.92 246.2 13.2 27.9 113.5 160.0 1.61 236.3 8.7 23.4
119.1 193.0 1.93 246.4 13.3 28.0 113.6 161.5 1.62 236.5 8.7 23.4
119.2 193.7 1.94 246.6 13.4 28.1 113.7 162.1 1.62 236.7 8.8 23.5
119.3 194.3 1.94 246.7 13.5 28.2 113.8 162.6 1.63 236.8 8.9 23.6
119.4 194.9 1.95 246.9 13.6 28.3 113.9 163.1 1.63 237.0 9.0 23.7
119.5 195.5 1.95 247.1 13.7 28.4 114.0 163.7 1.64 237.2 9.0 23.7
119.6 196.1 1.96 247.3 13.8 28.5 114.1 164.2 1.64 237.4 9.1 23.8
119.7 196.7 1.97 247.5 13.9 28.6 114.2 164.8 1.65 237.6 9.2 23.9
119.8 197.3 1.97 247.6 13.9 28.6 114.3 165.3 1.65 237.7 9.3 24.0
119.9 197.9 1.98 247.8 14.0 28.7 114.4 165.9 1.66 237.9 9.4 24.1
120.0 198.5 1.99 248.0 14.1 28.8 114.5 166.4 1.66 238.1 9.4 24.1
Page 132
120.1 199.2 1.99 248.2 14.2 28.9 114.6 167.0 1.67 238.3 9.5 24.2
120.2 199.8 2.00 248.4 14.3 29.0 114.7 167.5 1.67 238.5 9.6 24.3
120.3 200.5 2.00 248.5 14.4 29.1 114.8 168.0 1.68 238.6 9.7 24.4
120.4 201.1 2.01 248.7 14.5 29.2 114.9 168.6 1.69 238.8 9.7 24.4
120.5 201.8 2.02 248.9 14.6 29.3 115.0 169.1 1.69 239.0 9.8 24.5
125.9 238.7 2.39 258.6 19.9 34.6 120.6 202.4 2.02 249.1 14.7 29.4
126.0 239.4 2.39 258.8 20.0 34.7 120.7 203.1 2.03 249.3 14.8 29.5
126.1 240.2 2.40 259.0 20.1 34.8 120.8 203.7 2.04 249.4 14.8 29.5
126.2 240.9 2.41 259.2 20.2 34.9 120.9 204.4 2.04 249.6 14.9 29.6
126.3 241.6 2.42 259.3 20.3 35.0 121.0 205.0 2.05 249.8 15.0 29.7
126.4 242.3 2.42 259.5 20.4 35.1 121.1 205.7 2.06 250.0 15.3 29.8
126.5 243.1 2.43 259.7 20.6 35.3 121.2 206.3 2.06 250.2 15.4 29.9
126.6 243.8 2.44 259.9 20.7 35.4 121.3 207.0 2.07 250.3 15.5 30.0
126.7 244.5 2.45 260.1 20.8 35.5 121.4 207.6 2.08 250.5 15.6 30.1
126.8 245.3 2.45 260.2 20.9 35.6 121.5 208.3 2.08 250.7 15.6 30.3
126.9 246.0 2.46 260.4 21.0 35.7 121.6 208.9 2.09 250.9 15.8 30.5
127.0 246.8 2.47 260.6 21.1 35.8 121.7 209.6 2.10 251.1 15.8 30.5
127.1 247.6 2.48 260.8 21.2 35.9 121.8 210.2 2.10 251.2 15.9 30.6
127.2 248.3 2.48 261.0 21.3 36.0 121.9 210.8 2.11 251.4 16.0 30.7
127.3 249.1 2.49 261.1 21.4 36.1 122.0 211.5 2.11 251.6 16.1 30.8
127.4 249.9 2.50 261.3 21.5 36.2 122.1 212.1 2.12 251.8 16.3 31.0
127.5 250.6 2.51 261.5 21.8 36.5 122.2 212.8 2.13 252.0 16.3 31.0
127.6 251.4 2.51 261.7 21.8 36.5 122.3 213.5 2.13 252.1 16.4 31.1
127.7 252.2 2.52 261.9 21.9 36.6 122.4 214.2 2.14 252.3 16.5 31.2
Page 133
127.8 252.9 2.53 262.0 22.0 36.7 122.5 214.8 2.15 252.5 16.6 31.3
127.9 253.7 2.54 262.2 22.1 36.8 122.6 215.2 2.16 252.7 16.7 31.4
128.0 254.5 2.54 262.4 22.2 36.9 122.7 216.2 2.16 252.9 16.8 31.5
128.1 255.2 2.55 262.6 22.3 37.0 122.8 216.9 2.17 253.0 16.9 31.6
128.2 256.0 2.56 262.8 22.4 37.1 122.9 217.6 2.18 253.2 17.0 31.7
128.3 256.8 2.57 262.9 22.5 37.2 123.0 218.3 2.18 253.4 17.1 31.8
128.4 257.5 2.58 263.1 22.7 37.4 123.1 218.9 2.19 253.6 17.1 31.8
128.5 258.3 2.58 263.3 22.8 37.5 123.2 219.6 2.20 253.8 17.2 31.9
128.6 259.1 2.59 263.5 22.9 37.6 123.3 220.3 2.20 253.9 17.3 32.0
128.7 259.8 2.60 263.7 23.0 37.7 123.4 221.0 2.21 254.1 17.4 32.1
128.8 260.6 2.61 263.8 23.1 37.8 123.5 221.7 2.22 254.3 17.5 32.2
128.9 261.4 2.61 264.0 23.2 37.9 123.6 222.4 2.22 254.5 17.6 32.3
129.0 262.2 2.62 264.2 23.3 38.0 123.7 223.1 2.23 254.7 17.7 32.4
129.1 263.0 2.63 264.4 23.4 38.1 123.8 223.7 2.24 254.8 17.8 32.5
129.2 263.8 2.64 264.6 23.6 38.3 123.9 224.4 2.24 255.0 17.9 32.6
129.3 264.6 2.65 264.7 23.7 38.4 124.0 225.1 2.25 255.2 17.9 32.6
129.4 265.4 2.65 264.9 23.8 38.5 124.1 225.8 2.26 255.4 18.0 32.7
129.5 266.2 2.66 265.1 23.9 38.6 124.2 226.5 2.26 255.6 18.1 32.8
129.6 267.0 2.67 265.3 24.0 38.7 124.3 227.2 2.27 255.7 18.2 32.9
129.7 267.8 2.68 265.5 24.1 38.8 124.4 227.9 2.28 255.9 18.3 33.0
129.8 268.6 2.69 265.6 24.3 39.0 124.5 228.6 2.29 256.1 18.4 33.1
129.9 269.4 2.69 265.8 24.4 39.1 124.6 229.3 2.29 256.3 18.6 33.3
130.0 270.3 2.70 266.0 24.5 39.2 124.7 230.0 2.30 256.5 18.7 33.4
130.1 271.1 2.71 266.2 24.6 39.3 124.8 230.7 2.31 256.6 18.8 33.5
Page 134
130.2 271.9 2.72 266.4 24.7 39.4 124.9 231.5 2.31 256.8 18.9 33.6
130.3 272.7 2.73 266.5 24.8 39.5 125.0 232.2 2.32 257.0 19.0 33.7
130.4 273.5 2.73 266.7 25.0 39.7 125.1 232.9 2.33 257.2 19.1 33.8
130.5 274.3 2.74 266.9 25.1 39.8 125.2 233.6 2.34 257.4 19.2 33.9
130.6 275.1 2.75 267.1 25.2 39.9 125.3 234.4 2.34 257.5 19.3 34.0
130.7 275.9 2.76 267.3 25.3 40.0 125.4 235.1 2.35 257.7 19.4 34.1
130.8 276.7 2.77 267.4 25.4 40.1 125.5 235.8 2.36 257.9 19.5 34.2
130.9 277.5 2.78 267.6 25.6 40.3 125.6 236.5 2.37 258.1 19.6 34.3
131.0 278.3 2.78 267.8 25.7 40.4 125.7 237.3 2.37 258.3 19.7 34.4
131.1 279.1 2.79 268.0 25.8 40.5 125.8 238.0 2.38 258.4 19.8 34.5
135.2 315.0 3.15 275.4 31.2 45.7 131.2 280.0 2.80 268.2 25.9 40.6
135.3 315.9 3.16 275.5 31.3 45.8 131.3 280.9 2.81 268.3 26.0 40.7
135.4 316.8 3.17 275.7 31.5 45.9 131.4 281.7 2.82 268.5 26.2 40.9
135.5 317.7 3.18 275.9 31.6 46.1 131.5 282.6 2.83 268.7 26.3 41.0
135.6 318.6 3.19 276.1 31.7 46.2 131.6 283.4 2.83 268.9 26.4 41.1
135.7 319.5 3.20 276.2 31.9 46.3 131.7 284.3 2.84 269.1 26.5 41.2
135.8 320.5 3.20 276.4 32.0 46.5 131.8 285.1 2.85 269.2 26.7 41.4
135.9 321.4 3.21 276.6 32.1 46.6 131.9 286.0 2.86 269.4 26.8 41.5
136.0 322.4 3.22 276.8 32.3 46.8 132.0 286.8 2.87 269.6 26.9 41.6
136.1 323.3 3.23 277.0 32.4 46.9 132.1 287.7 2.88 269.8 27.0 41.7
136.2 324.3 3.24 277.2 32.6 47.0 132.2 288.5 2.89 270.0 27.1 41.8
136.3 325.2 3.25 277.3 32.7 47.2 132.3 289.4 2.89 270.1 27.3 42.0
Page 135
136.5 327.1 3.27 277.7 33.0 47.4 132.5 291.1 2.91 270.5 27.5 42.2
136.6 328.1 3.28 277.9 33.1 47.6 132.6 291.9 2.92 270.7 27.6 42.3
136.7 329.0 3.29 278.1 33.2 47.7 132.7 292.8 2.93 270.9 27.8 42.5
136.8 330.0 3.30 278.2 33.3 47.9 132.8 293.6 2.94 271.0 27.9 42.6
136.9 330.9 3.31 278.4 33.3 48.0 132.9 294.5 2.94 271.2 28.0 42.7
137.0 331.9 3.32 278.6 33.4 48.1 133.0 295.4 2.95 271.4 28.1 42.8
137.1 332.8 3.33 278.8 33.6 48.3 133.1 296.2 2.96 271.6 28.3 43.0
137.2 333.8 3.34 279.0 33.7 48.4 133.2 297.1 2.97 271.8 28.4 43.1
137.3 334.7 3.35 279.1 33.8 48.5 133.3 297.9 2.98 271.9 28.5 43.2
137.4 335.6 3.36 279.3 34.0 48.7 133.4 298.8 2.99 272.1 28.6 43.3
137.5 336.6 3.37 279.5 34.1 48.8 133.5 299.7 3.00 272.3 28.8 43.5
137.6 337.5 3.38 279.7 34.3 49.0 133.6 300.6 3.01 272.5 28.9 43.6
137.7 338.5 3.38 279.9 34.4 49.1 133.7 301.5 3.01 272.7 29.0 43.7
137.8 339.4 3.39 280.0 34.5 49.2 133.8 302.4 3.02 272.8 29.2 43.9
137.9 340.4 3.40 280.2 34.7 49.4 133.9 303.3 3.03 273.0 29.3 44.0
138.0 341.4 3.41 280.4 34.8 49.5 134.0 304.2 3.04 273.2 29.4 44.1
138.1 342.4 3.42 280.6 35.0 49.7 134.1 305.1 3.05 273.4 29.5 44.2
138.2 343.4 3.43 280.8 35.1 49.8 134.2 306.0 3.06 273.6 29.7 44.4
138.3 344.4 3.44 280.9 35.2 49.9 134.3 306.9 3.07 273.7 29.8 44.5
138.4 345.4 3.45 281.1 35.4 50.1 134.4 307.8 3.08 273.9 29.9 44.6
138.5 346.4 3.46 281.3 35.5 50.2 134.5 308.7 3.09 274.1 30.1 44.8
138.6 347.4 3.47 281.5 35.7 50.4 134.6 309.6 310 274.3 30.2 44.9
138.7 348.4 3.48 281.7 35.9 50.6 134.7 310.5 3.10 274.5 30.3 45.0
138.8 349.4 3.49 281.8 36.0 50.7 134.8 311.4 3.11 274.6 30.5 45.2
Page 136
139.0 351.4 3.51 282.2 36.3 51.0 135.0 313.2 3.13 275.0 30.7 45.4
139.1 352.4 3.52 282.4 36.4 51.1 135.1 314.1 3.14 275.2 31.1 45.6
Legend:
psia ─ absolute pressure in psi
Psig ─ gauge pressure in psi
kPa ─ absolute pressure in kilo-Pascal
InHg ─ pressure (vacuum) in inch-Mercury
Page 137
Page 138
81. Inlet
82. Main inlet
8. GAS
83. Inlet through humidifier
Page 139
Page 140
Page 141
Page 142
Page 143
Legend:
DILM12
CB1 16A KM1
L1 T1 L1 T1
R/L1
L3 T3 L3 T3
T/L3
1. HEATERS = 400V /9000W
TC
C
N
PE/GND
TB1
Heaters
TB1
To
2. KM1 = CONTATOR: DILM12-104. QS1 = PANEL SWITCH: N/A
Control
PE/GND
Panel
Q1
14.5 Electrical Wiring Diagram: The Generator
TB2
3. CB1 = CIRCUIT BREAKER: FAZ-C 16/3
Page 151
Preparation By: Description Date Rev. No ECO
Yan S. Use Water Valve 24VDC 14.02.2012 01
Yan S. Change connection to TRIPCOIL( TC ) 8.08.2012. 02
~ VAC
Q1/L1 Q1/N1
+ -
TB2 TB2
24VDC
Page 152
Pressure Temperature Steam
Level Elec. Low Heater Air Supply Water Supply
Cut off Cut off Pressure Level Elec. High 40
CB4 RL1 41 Pressure Water Relay
Pressure/Float RL2
L 2 1 2 A1 A2
T 1
38
C
32 33 36 37
4A 3 4 42 43 39
R1
NC
Option
34 34 34
For Air Valve Water Valve
1
D1 D2
TC
1
2 3
44 45 46
POWER SUPPLAY
L N
24VDC 1.0A
S S S S S S
R M m R M m
1 1 1 2 2 2
S E
RL1
GICAR 2DRLH
~ VAC
S E
Q1/L1
RL2
~ VAC
TB2
Q1/N1 Q1/L1
L N
Water Pump
CB3 R1 TB2
L T
M TB2
NO 36 33 34
4A Water
Heater Pump
14.6 Generator Control Electrical Cascade Diagram