Benchmark: Technical Ser Vices Manual
Benchmark: Technical Ser Vices Manual
L-K Industries, Inc. 6952 Lawndale ¤ Houston TX 77023 ¤ 713 926•2623, fax 713 926•7736 ¤ www.lk-ind.com
Technical Ser vices Manual
Lab Centrifuge
Model #’s LAB-A115C, LAB-A220C, LAB-B115C, LAB-
B220C, LAB-C115C, LAB-C220C, LAB-D115C, LAB-D220C
C
Cidunt 6
D
Dolore 6
E
ectetue 6
eret 6
F
faccum 6
folerit. See lorem
G
Galiwad 6
H
Hoodoo 6
Halisbo 6
I
illaoper. See Agniam
J
Jennicu 6
26 | Enicaper ficaed susta nondin is es nonim et dolore L-K INDUSTRIES, INC. | III
Table of Contents
2 Cum nullan 5
faci blan utem elestrud tinit . . . . . . . . . . . 6
Dolore magniamet . . . . . . . . . . . . . 6
Adit augiat, quat . . . . . . . . . . . . . . 6
iv | Table of Contents LOREM IPSUM | 25
24 | Enicaper ficaed susta nondin is es nonim et dolore | 1
Fuses
There are 5 fuses located behind
the front panel of your centrifuge.
These are used for over current
protection due to voltage spikes or
accident. If replacement becomes
necessary, disconnect power to the
centrifuge. Remove the 8 screws
on the upper front panel. Panel
swings down. Replace fuses with
250V “SLO-BLOW” or “FAST
ACTING” ceramic type that is
CSA or UL certified as indicated
at right:
2 Troubleshooting
Tachometer Timer
With the centrifuge turned off, With the centrifuge turned off,
there should be a “0” showing in there should be numbers show-
the LCD window. If this is not ing in the LCD window. If this is
the case, replace batteries and not the case, replace batteries and
follow the procedure for program- follow the procedure for program-
ming the tachometer. ming the timer.
If the tachometer is giving false If the timer is giving false read-
readings on rpm’s, follow the pro- ings or showing 8’s in the LCD
cedures for reprogramming. window, follow the procedures for
reprogramming.
3
RPM’s
Checking Accuracy
Battery Installation
Remove the battery cover by
pushing inward and down. Re-
move old batteries. Install new
Rate Display
Indicates the present rate value,
which is equal to the pulse fre-
quency multiplied by the input
calibration value.
batteries in both of the two slots.
Down Key Once the batteries are in place,
When the program input is active the unit will go into a self test
this key is used to scroll through mode, and all the segments on the
the menu items. After a menu LCD display will be illuminated.
item has been chosen for editing, The self test mode is exited by de-
the down key is used to set the pressing the Next key, which will
value for the currently selected then display the model number
(flashing) digit. (4). Depress the Next key again to
ready the unit for operation.
Next/Reset Each time the batteries are
removed and then replaced, the
When the program input is active
18 | Enicaper ficaed susta nondin is es nonim et dolore LOREM IPSUM | 7
RPM indicator will have to be this value and displays the results
reprogrammed. Run a jumper as the rate value. In combination
wire from terminal 5 to terminal 1 with the Rate Decimal Point
as shown below: parameter the calibrator value
can be set in a range from 0.001
Programming parameters can
to 9999. Set this value to 15 as
now be accessed by pressing the
shown below.
Down key on the front of the
RPM indicator. To edit a param-
eter, use the Down key scroll until Rate Display Decimal
the desired parameter appears
on the screen. Pressing the Next
key will cause the left most digit
of that value to begin to flash.
Point
Sets the decimal position to be
used for the rate display in a range
from Off to 0.000. Set this pa-
rameter to off as shown below.
Next/Reset Key
When in Operation Mode,
depressing the key will reset the Battery Installation
time value to the operator preset
value. When the preset input is Remove the battery cover by
active this key is used to select a pushing inward and down. Re-
menu item for editing (left most move old batteries. Install new
digit will begin to flash) and will batteries in both of the two slots.
move to the right each time the Once the batteries are in place,
key is depressed. the unit will go into a self test
mode, and all the segments on the
Preset Value. Indicates LCD display will be illuminated.
replaced, the timer will have the The self test mode is exited by de-
reset value. pressing the Next key, which will
Preset Value then display the model number
(8). Depress the Next key again to
ready the unit for operation.
Each time the batteries are
removed and then replaced, the
timer will have to be repro-
grammed. Run a jumper wire
from terminal 5 to terminal 1 as
shown below.
Battery Replacement
Make sure power to your centri-
fuge is disconnected at the main
breaker box. Open the front panel
of the centrifuge by removing the
8 screws at the top. Swing the
16 | Enicaper ficaed susta nondin is es nonim et dolore LOREM IPSUM | 9