H-100 User Manual
H-100 User Manual
Adopting the advanced “high luminosity cold light source reflection determination” technology, H-100
urine analyzer gets the specialties of resisting the interference of the ambient light and has longer
lifespan. It can finish the testes on 11 kinds of biochemical components in urine within 30 seconds,
and it also can revise the affects toward the test result which is caused by ambient temperature,
ambient light, acid-base scale and abnormally colored sample. It can also connect with the urine
sediments analyzer.
1.2Technical Specifications
Light wavelength 525nm 572nm 610nm 660nm
Test Items Urobilinogen(UBG)Bilirubin(BIL)Ketone(KET)Blood(BLD)
Protein(PRO)Nitrite(NIT)Leucocytes(LEU)Glucose(GLU)
Specific Gravity(SG)pH(pH)Vitamin C (VC)(or Microalbumin (MALB))
Test Speed Fast mode: 120 specimens/hour Slow mode :60 specimens/hour
Data Memory 1,000 patient results, 50 quality control records
Language Chinese, English optional
Environment requirements Required temperature range is 15℃-35℃, optimal temperature
20℃-25℃
relative humidity ≤80%
Communication interface RS-232 serial interface, parallel printing interface
Bar code connecting with bar code reader
Baud rate 9600 bps,1200 bps
Power supply AC 100V~240V, 50/60Hz
Fuse 250V 1A
Power 40VA
Net weight 3.60kg
Dimensions 376mm × 316mm ×170mm
Printer Inner thermal printer, outer stylus printer is available
Adopted reagent strip H8, H10, H11
1
H-100 Urine Analyzer Operation Manual
Tm × C r
R=
Tr × C m
In the principle:
R ----- The reflection rate,
Tr ----- the reflected intensity of the blocks on the strips under the reference light,
Cr ----- the reflected intensity of the blank block under the reference light,
Tm ----- the reflected intensity of the blocks on the strips under the predetermined light,
Cm -----the reflected intensity of the blank block under the predetermined light.
① ②
⑥ ⑦ ⑧ ⑨ ⑩
⑤
Figure 1-1
2
H-100 Urine Analyzer Operation Manual
The urine sample may have potential infectivity, protective measures should be taken when
testing, cleaning or doing maintenance to the instrument.
Please dispose the urine sample and the waste strips according to the local lab regulation.
Figure 1-2
1.6 Introduction of the Keyboard
.
These number keys are used to enter the corresponding number.
(Number keys)
After pressing the key, the analyzer will sound a warning tone to dip the test
strip into urine. 40 seconds later it begins testing. No response when press the
key during the testing time
(Start)
Under the main menu, touching this key will make the screen turn to the main
menu. Under the menu status, touching this key will make the screen return to
(Menu) the upper menu.
Press this key to confirm the option under the menu status. Press the key before
current test finished, the analyzer will stop testing after current test. Touching this
(Enter)
key could check the result.
Under the number input status, touching this key will turn the corresponding
(Clear) number to zero.
Touching this key will make the printer to go ahead, leaving two lines blank.
(Line)
Under the date recall status, touching this key will print out the test data on the
current screen.
(Print)
Under the menu status, touching the two keys will move the cursor upward or
downward. Under the data recall status, touching the two keys will turn to the
(Up/Down) previous or the next screen (the previous or the next record).
3
H-100 Urine Analyzer Operation Manual
1 2 3 4 5 6
7 8 9 10 11 12
Figure 2-1
4
H-100 Urine Analyzer Operation Manual
2.3Installation
2.3.1 Install printing paper (Figure 2-2)
a)Obtain a roll of thermal printer paper with width of 57mm and diameter less than 50mm
b)Remove the cover of the printer.
c)Put the new roll of paper into the paper box and release the paper roll, with the end of the paper
toward the near side of the instrument.
d)witch out the handle and feed the paper roll to the printer perpendicularly. Pull the paper up and
down and press down the handle.
e)Pull the paper through the hole on the covering tab of the printer and set the covering tab back into
position.
Note :
H-100 can connect to printers such as EPSON LQ1600K、EPSON LQ300K+。
5
H-100 Urine Analyzer Operation Manual
Communication
External Printer
Interface
Interface
Figure 2-4 Figure 2-5
6
H-100 Urine Analyzer Operation Manual
Locate
Data Recall
ID
Calibration
QC Record
Edit ID 8 PARAMETERS
11 PARAMETERS
ID
Abnormals
Set up menu
Plus Used
Status
SI
Int Printer
PC Port DR200
Clarity conventional
中文
Language Set
English Data Output Unit
FAST MODE
Parameter Select
Test Mode SLOW MODE
More Set
Baud Rate Select 9600
Set scaler
1200
Set abnormals
7
Figure 3-1
H-100 Urine Analyzer Operation Manual
3.2 Setup
3.2.1 Set Sequence Number
URINE ANALYZER 07-01-01
#0001 12:00:00
VERSION 7.81
MENU START
Figure 3-2
In the main screen (Figure 3-2) press the key “Menu”, enter into the sequence number setup, the
screen shows as follows:
MAIN MENU
Data Recall Set Seq No.
Calibration Edit ID
QC Liquid Set Up Menu
Figure 3-3
Press the key “Menu” return to the main screen (Figure 3-2).
Press the key“ ▲ ”and“ ▼”to move the cursor to the position of “Set Seq. No.”, press the key
“Enter” for confirmation, and enter sequence number setup Submenu.
Figure 3-4
The sequence number can be any numerical value between 001-999, the sequence number will add
1 automatically after each test.
Press the number key “0”to “9” on the keyboard to setup the sequence number.
Press the key “Enter” to confirm and return to the upper menu.
Press the key “Menu” to discard changes and return to the upper menu.
3.2.2 ID Edit
Choose “Edit ID” in the figure 3-3 and press “Enter” to confirm. In the ID interface, the display as
follows:
Input ID:
********************
ABCDEFGHIJKLMNOPQRSTUVWXYZ
abcdefghijklmnopqrstuvwxyz
Figure 3-5
enter the number on the keyboard to input the number, press” Line”, the cursor will stay under the
capital letter. Press “Line” again, the cursor will stay under the small letter. Press the key“▲ ”and
“▼”to move the cursor. Choose the letter and press “Start” to input the letter. After input the ID,
press “Enter” to confirm the ID, it displays main screen (Figure 3-3). Enter the next testing, the ID
input will display in the printout.
8
H-100 Urine Analyzer Operation Manual
In main menu (Figure 3-3), press the key“▲ ”and“ ▼ ”to select the “Data Recall”, press the key
“Enter” to confirm, the screen will show as follows:
Data Recall:
Figure 3-6
The user may check the test result by four ways: current record, Locate, QC Record and ID.
3.2.3.1 Current record recall
Select Current Record and press “Enter”, the screen will show current records. Press the key” Line”
could turn record page and press the key“▲ ”and“ ▼ ”to select the corresponding record. Press
the key “Print ”to print the showed record.
3.2.3.2 Sequence number Recall
Select the “Locate” and press the key “Enter” to confirm, the screen will show as follows:
Sequence NO.0001
Figure 3-7
The cursor stops at the first number, enter the recall sequence number through the number key. And
then, press the key “Enter” to check the result by the sequence number. If there is record in the
instrument, the screen will display the record. If there is no such record, the screen will display as
empty record (Shown in Figure 3-8).
Figure 3-8
press the key “Menu” to return to the figure3-6.
9
H-100 Urine Analyzer Operation Manual
3.2.3.3 ID record:
Choose ”ID” in figure3-6 and press ”Enter” to confirm. The display as follows:
Input ID:
********************
ABCDEFGHIJKLMNOPQRSTUVWXYZ
abcdefghijklmnopqrstuvwxyz
Figure 3-9
Enter the number on the keyboard to input the number, press“ Line”, the cursor will stay under the
capital letter. Press “Line” again, the cursor will stay under the small letter. Press the key“▲“and “▼”
to move the cursor. Choose the letter and press “Start” to input the letter. Press the Line for the third
time, figures could be input. After input the ID, press “Enter” to confirm the ID. If there is ID, the record
will be displayed. If not, “Record Not Found!” will be displayed.
3.2.3.4 Quality Control Record
Analyzer could store 50 pieces QC records. Choose QC recall in figure 3-6 and then press “Enter”
key to confirm. Press the key Line could turn record page and press the key“▲“and “▼” to select the
corresponding record. Press the key “Print” to print the showed record. Press “Menu” to return.
3.2.3.5 Clear memory
Select “Clear Memory” in figure 3-6 and press the key “Enter”, the screen displays as follows:
Clear Memory?
No
Yes
Figure 3-10
Select “No”, the screen will return to the upper menu, select “Yes” and press the key “Enter”, all the
records will be cleared, the sequence number set as 001 automatically.
SET UP INSTRUMENT
Set Clock Language Set
Strip Select Set abnormals
Status More Set
Figure 3-11
Press the key”▲” or “▼“to move the cursor to the item which the operator want to change, press the
key “Enter” to confirm, the screen will show the corresponding display. Press the key “Menu” to return
the upper screen after setup.
3.2.5.1 Set clock
Select “Set Clock” in the figure 3-11, press the key “Enter” to confirm, the screen shows as follows:
10
H-100 Urine Analyzer Operation Manual
Set Clock:
Set Time
Set Date
Date Format
Figure 3-12
Set time
Select “Set Time” in figure 3-12, press the key “Enter” to confirm, the screen shows as follows:
12: 00
24 Hour
Figure 3-13
The cursor stop at the position of the first number, press the corresponding number key to enter the
current time, press Menu could give up setup and back to upper menu.
After each entering, the cursor will move to the number next to its right. After the last number has
been entered, the cursor will return to the fist number. After operation, press the key “Enter” to
confirm and return to the upper menu.(Figure 3-12)
If the number entered is less than 10, add “0” to its front.
Set date
Select “Set Date” in figure 3-12, press the key “Enter” to confirm, the screen shows as follows:
06/01/01
YY/MM/DD
Figure 3-14
The cursor stop at the position of the first number, press the corresponding number key to enter the
current time.After each entering, the cursor will move to the number next to its right. After the last
number has been entered, the cursor will return to the fist number. After operation, press the key
“Enter” to confirm and return to the upper menu.
If the number entered is less than 10, add “0” to its front.
Data Format
Select “Date Format” in figure 3-12, press the key “Enter” to confirm, the screen shows as follows:
11
H-100 Urine Analyzer Operation Manual
Strip select:
8 PARAMETERS
10 PARAMETERS
11 PARAMETERS
Figure 3-16
H-100 urine analyzer could test 8 PARAMETERS、10 PARAMETERS、11 PARAMETERS strips
respectively. The user could choose one according to their request. The cursor stop at the current
item, press the key “▲” or “▼” to select the strip.
Note:
H-100 urine analyzer is only fit for H8、H10、H11 reagent strips of DIRUI. For getting correct result,
please do not use other strips.
Strip selection function could be used after the supplier’s authorization.
3.2.5.3 Set Status
Select “Status” in figure 3-11, press the key “Enter” to confirm, the screen shows as follows:
Figure 3-17
In this menu, the operator can change the status of “ID”, “Abnormals”, “Plus Used”, “Int Printer” ,“Ext
Printer”, “PC Port”, “color ” and “Clarity”.
ID: when “ID” is “ON” and the connected bar code reader could scan the bar code on tube, the
output display ID as figure 3-18. When “ID” is “OFF”, the output not display ID as figure 3-19
Note:
Each scanning limit is less than 30 samples. The over parts are invalid code, the length of the bar
code should be less than 15 character. Operator could scan bar code when waiting for put in strip and
printing result. The scanned bar code during the strip table moving is invalid. Without tone response
from the scanner, the scanned result is invalid.
12
H-100 Urine Analyzer Operation Manual
Abnormals: When Abnirmals is “ON”, if the test result of a certain item is more than or equal to
the set critical value, the symbol “*” will appear on the report (3-20) otherwise no report will be printed
out (3-21).
Plus Used: When the “plus Used” is “ON”, the test result will print the “+” together with the
relative test value under the current unit select (figure 3-22). When the plus system is “OFF”, the “+”
will not print out (figure 3-23).
13
H-100 Urine Analyzer Operation Manual
Internal Print: When “Int. printer” is “ON”, the internal printer will print the result, otherwise it will
not print.
External Print: When “Ext. printer” is “ON”, the external printer will print the result, otherwise it will
not print.
Computer Interface: when “PC Port” is “ON”, analyzer can transmit data to computer; otherwise
it can not.
Color and Clarity
When “Color” is “ON”, the printed test result will show “Color” Item; when “Clarity” is “ON”, the printed
test result will show “Clarity” item. Operator could write the relevant color and Clarity results. When
they are “OFF”, “Color” and “Clarity” will not show in the result.
3.2.5.4 Set Language
In figure 3-11, select the “Language Set”, press the key “Enter” to confirm, the screen will show as
follows:
Language Select:
中文
English
Figure 3-24
Press the key “▲” and “▼” to select language, press the key “Enter” to confirm and return the upper
menu.
Under the Chinese mode, the instrument screen has Chinese indication, and the print result is
Chinese; Under the English mode, the instrument screen has English indication, and the print result
is English.
3.2.5.5 Set abnormals
Select “Set abnormals” in figure 3-11, press the key “Enter” to confirm, the screen will show as
follows:
UBG PRO
BIL NIT
KET LEU
BLD GLU
Figure 3-25
By pressing the key “▲” or “▼” to move the cursor to the selected item, press the key “Enter”. For
example, to change the critical value of urobilinogen, move the cursor to the “UBG” item, press the
key “Enter”. The screen will show as follows:
Normal 17umol / L
1+ 34umol / L
2+ 68umol / L
3+ >=135umol / L
Figure 3-26
The cursor will stop at the current critical value. Press the key “▲” or “▼” to Move the cursor to the
needed critical value, and press the key “Enter” for confirmation and return to the upper menu.
14
H-100 Urine Analyzer Operation Manual
Set Options:
Parameter Select
Set Scaler
Reset All Parameter
Figure 3-27
a) Choose “Paramerer Select”,Press”Enter”to confirm,the screen displays as follows.
1 SI
2 CT200
3 DR200
4 Plus System
5 Conventional
Table 3-1
Note: CT200, DR200 is the format for data output (Appendix B).
The test mode: Select the “Test Mode” in figure3-28, press the key “Enter” to confirm, the screen
will show the option one by one between two modes: The test mode including “FAST MODE” and
“SLOW MODE”.
Under fast mode, put the next strip on the table without pressing other key after testing one strip. The
fast mode is suitable for the continuous test of large quantity of urine samples; the test rate for each
strip is 30 seconds. Under slow mode, the instrument will stop after complete one test. Press the key
“Start” if the next test is conducted. The slow mode is suitable for the non-continuous test; the test
rate for each strip is 60 seconds.
Baud rate select:it is used to set up the transmission speed of urine analyzer. In figure
3-28,select “Baud Rate Select” and press the key “ Enter”, the content after “Baud Rate Select” will
be switched between“9600” and“1200”.
Levels for MALB: it is used to set up the sensitivity of microalbumin. In figure 3-28,select “Levels
for MALB” and press the key “ Enter”, the content after “Levels for MALB” will be switched
between ”HIGH” and ”LOW”.
b)Reset All Parameter: Select the “Reset All Parameter” In figure3-27, press the key “Enter” to
confirm,After hearing a long ring,it retrurns to figure 3-27 the scaler recover to original setup.
15
H-100 Urine Analyzer Operation Manual
● If the calibration strip has besmirched or damaged, please contact with the supplier, do not
In order to make sure to get correct test result, it is recommended to test the urine analyzer by the
calibration strip every one or two weeks.
Test Method
Select “Calibration” in figure 3-3, put the calibration strip on the center of the strip table, press the key
“Enter” to conduct testing test toward the urine analyzer. The screen will show as follows:
Figure 4-1
If the urine analyzer has passed the test, the test result is “Calibration OK”, otherwise the
result is “Calibration Not OK”
In order to make sure the correctness of the test result, the positive and negative quality control liquor
manufactured by Dirui Industrial Co., Ltd. should be often used to test the urine analyzer and the
reagent strip.
The quality control monitoring might be conducted under the following conditions:
z At the time before daily test
z At the time when replace another tube of strips
z At the time when the operator is changed
z At the time when there is query in the test result
Test the urine analysis quality control liquor in the same process of testing the urine sample.
Then compare the result with the reference value in the control liquor manual.
Test Method
16
H-100 Urine Analyzer Operation Manual
In the main menu (figure 3-3) select “ QC liquid”. Put the strip that dipped by QC on the strip table and
press “Enter” as follows.
SYSTEM IS TESTING…
Figure 4-2
After the testing, the QC result will be displayed. Refer to the QC result in the manual.
17
H-100 Urine Analyzer Operation Manual
Warnings
The urine sample may have potential infectivity; please wear protective gloves when
testing, cleaning, or doing maintenances to the instrument.
Please dispose the urine sample and the abandoned strip according to the local lab
regulation.
This chapter introduces the method that the urine analyzer conducts the conventional urine analysis.
Before the test, please carefully read the “Chapter 3 Functions and Setup of the H-100 Urine
Analyzer”.
Note:
When the instrument did not pass the self-testing, the screen displays the error information code.
Do not place the instrument in the place where there is direct projection of the sunshine.
Please check the type of the strip before test, in order to avoid the mistaken test result caused by
the mistaken strip type.
Do not use the strip which has passed the expiration date or deterioration.
CALIBRATING ……
Figure 5-1
After the calibrating, the screen displays as followed
URINE ANALYZER 07-01-01
#0001 12:00:00
VERSION 7.81
MENU START
Figure 5-2
18
H-100 Urine Analyzer Operation Manual
1) Press the key “Start” around 2 seconds in main screen, the instrument make a tone for dipping the
strip into the urine sample. On hearing the warning tone, dip the reagent area of the strip into the
urine sample which is fresh, fully mixed and not centrifugal, then place the strip on the center of the
strip table, and push it forward until the strip touch the end of the strip table (this one is marked as the
first strip).
2) On hearing next warning tone, dip another strip into the urine sample, and then move out the strip
quickly and put it on the absorbent paper to wait to be tested (this one is marked as the second strip).
3) On hearing the warning tone for the third time, the fist strip has been tested and its test result is
printed out, the strip table moves out, dip another strip and put it on the absorbent paper, remove the
first strip from the strip table and put the second strip on the table to conduct test.
4) When hearing the warning tone, repeat the operation in 3).
Note: if want to stop the test, please take the following measures:
(1)Press the key “Enter” before the strip table come out.
(2)Remove the strip from the strip table.
Note:
The tested strip must be placed in the right place before the strip table moves.
If no strip is placed on the testing slot, the instrument will automatically stop testing. The display
will show information as in the following picture. The strip table moves out. Touch the key “Start” to
continue the test.
#0001
STRIP MISSING!
Figure 6-4
During the test process, the keyboard is ineffective, press any key will get no response.
During the test process, do not hit the strip table.
19
H-100 Urine Analyzer Operation Manual
Note:
Do not clean the instrument when the power is turned on.
Do not use any organic solvent such as petrol, thinner lacquer, benzene compounds and things
that may erode the instrument to wipe it.
Do not clean the LCD with water.
Do not use anything that may scrap the strip table and white benchmark to wipe the strip table.
Do not use any solvent to clean the white benchmark.
If there is apparent scrape on the white benchmark, contact with the supplier.
Note:
Do not touch the white benchmark with the NaOH solution. Weekly cleaning is recommended.
6.3 Disinfection
Because the test slot, select slot and the select plate contact with the urine sample, it should be
disinfected.
The following solutions can be used to conduct disinfection.
20
H-100 Urine Analyzer Operation Manual
Note:
The white benchmark does not contact the liquid.
21
H-100 Urine Analyzer Operation Manual
22
H-100 Urine Analyzer Operation Manual
Appendix A
Manufacturer’s warranty
Dear consumer:
Thank you for purchasing the H-100 urine analyzer. Our company provides the following services for
you:
1. Technical consultations are provided at any time.
2. Maintenance free of charge within a year from the day you purchase the instrument.
3. Maintenance will be charged in the following conditions:
1) Product which has pass the date for free maintenance.
2) Damage caused by accidental factor or improper use.
3) Damage caused by the operation that not according to the instruction manual.
4) Damage caused by your own repair that without our company’s permission.
4. With the development of technology, DIRUI will supply the service of update of analyzers
If you need any technological service, please reach us according to the following address:
Address: 95 Yunhe Street New & High Tech. Development Zone Changchun, Jilin 130012 P.R.China
E-mail: [email protected]
http://www.dirui.com.cn
23
H-100 Urine Analyzer Operation Manual
Appendix B
Verification: none
24
H-100 Urine Analyzer Operation Manual
SP D a t e : × × × × - × × - × × SP × × : × × CR LF
SP N o . SP : × × × CR LF
SP U B G SP × × × × × × × × × × × × × × × × CR LF
SP B I L SP × × × × × × × × × × × × × × × × CR LF
SP K E T SP × × × × × × × × × × × × × × × × CR LF
SP B L D SP × × × × × × × × × × × × × × × × CR LF
SP P R O SP × × × × × × × × × × × × × × × × CR LF
SP N I T SP × × × × × × × × × × × × × × × × CR LF
SP L E U SP × × × × × × × × × × × × × × × × CR LF
SP G L U SP × × × × × × × × × × × × × × × × CR LF
SP S G SP SP × × × × × × × × × × × × × × × × CR LF
SP P H SP SP × × × × × × × × × × × × × × × × CR LF
SP V C SP SP × × × × × × × × × × × × × × × × CR LF
SP A L B SP × × × × × × × × × × × × × × × × CR LF
ETX
SP D a t e : × × × × - × × - × × SP × × : × × CR LF
SP N o . SP × × × × CR LF
SP I D : × × × × × × × × × × × × × × × × × × CR LF
SP U B G SP × × × × × × × × × × × × × × × × CR LF
SP B I L SP × × × × × × × × × × × × × × × × CR LF
SP K E T SP × × × × × × × × × × × × × × × × CR LF
SP B L D SP × × × × × × × × × × × × × × × × CR LF
SP P R O SP × × × × × × × × × × × × × × × × CR LF
SP N I T SP × × × × × × × × × × × × × × × × CR LF
SP L E U SP × × × × × × × × × × × × × × × × CR LF
SP G L U SP × × × × × × × × × × × × × × × × CR LF
SP S G SP SP × × × × × × × × × × × × × × × × CR LF
SP P H SP SP × × × × × × × × × × × × × × × × CR LF
SP V C SP SP × × × × × × × × × × × × × × × × CR LF
SP A L B SP × × × × × × × × × × × × × × × × CR LF
ETX
Note:
The above tables are the transmission format of 12 items under international unit. If the test items are
less than 12, the untested item will not output. For example: we test 11 items, the ALB in the
transmission format will not output.
25
H-100 Urine Analyzer Operation Manual
STX CR LF
# × - × × × SP SP SP SP SP SP × × - × × - × × CR LF
U B G SP × × × × × × × × × × × × × × × × CR LF
B I L SP × × × × × × × × × × × × × × × × CR LF
K E T SP × × × × × × × × × × × × × × × × CR LF
B L D SP × × × × × × × × × × × × × × × × CR LF
P R O SP × × × × × × × × × × × × × × × × CR LF
N I T SP × × × × × × × × × × × × × × × × CR LF
L E U SP × × × × × × × × × × × × × × × × CR LF
G L U SP × × × × × × × × × × × × × × × × CR LF
S G SP SP × × × × × × × × × × × × × × × × CR LF
P H SP SP × × × × × × × × × × × × × × × × CR LF
V C SP SP × × × × × × × × × × × × × × × × CR LF
A L B SP × × × × × × × × × × × × × × × × CR LF
ETX
STX CR LF
SP N O . × × × SP SP × × - × × - × × × × CR LF
SP SP SP SP SP SP SP SP SP SP SP × × : × × : × × CR LF
CR LF
SP p H SP SP × × × × × × × × × × × × × × CR LF
SP N I T SP × × × × × × × × × × × × × × CR LF
SP G L U SP × × × × × × × × × × × × × × CR LF
SP V C SP SP × × × × × × × × × × × × × × CR LF
SP S G SP SP × × × × × × × × × × × × × × CR LF
SP B L D SP × × × × × × × × × × × × × × CR LF
SP P R O SP × × × × × × × × × × × × × × CR LF
SP B I L SP × × × × × × × × × × × × × × CR LF
SP U R O SP × × × × × × × × × × × × × × CR LF
SP K E T SP × × × × × × × × × × × × × × CR LF
SP W B C SP × × × × × × × × × × × × × × CR LF
CR LF
ETX
26
H-100 Urine Analyzer Operation Manual
STX CR LF
SP N O . × × × SP SP × × - × × - × × × × CR LF
SP SP SP SP SP SP SP SP SP SP SP × × : × × : × × CR LF
CR LF
SP p H SP SP × × × × × × × × × × × × × × CR LF
SP N I T SP × × × × × × × × × × × × × × CR LF
SP V C SP SP SP SP SP SP SP SP SP SP SP SP SP SP SP SP CR LF
SP G L U SP × × × × × × × × × × × × × × CR LF
SP S G SP SP × × × × × × × × × × × × × × CR LF
SP B L D SP × × × × × × × × × × × × × × CR LF
SP P R O SP × × × × × × × × × × × × × × CR LF
SP B I L SP × × × × × × × × × × × × × × CR LF
SP U R O SP × × × × × × × × × × × × × × CR LF
SP K E T SP × × × × × × × × × × × × × × CR LF
SP W B C SP × × × × × × × × × × × × × × CR LF
CR LF
ETX
STX CR LF
SP N O . × × × SP SP × × - × × - × × × × CR LF
SP SP SP SP SP SP SP SP SP SP SP × × : × × : × × CR LF
CR LF
SP p H SP SP × × × × × × × × × × × × × × CR LF
SP N I T SP × × × × × × × × × × × × × × CR LF
SP V C SP SP SP SP SP SP SP SP SP SP SP SP SP SP SP SP CR LF
SP G L U SP × × × × × × × × × × × × × × CR LF
SP S G SP SP SP SP SP SP SP SP SP SP SP SP SP SP SP SP CR LF
SP B L D SP × × × × × × × × × × × × × × CR LF
SP P R O SP × × × × × × × × × × × × × × CR LF
SP B I L SP × × × × × × × × × × × × × × CR LF
SP U R O SP × × × × × × × × × × × × × × CR LF
SP K E T SP × × × × × × × × × × × × × × CR LF
SP W B C SP SP SP SP SP SP SP SP SP SP SP SP SP SP SP CR LF
CR LF
ETX
Note:
STX=0X02 CR=0X0D LF=0X0A SP=0X20
SP1=0XAB ETX=0X03 -=0X2D X= Any ASCII character
27
H-100 Urine Analyzer Operation Manual
Appendix C
3+ >= Ca200 Ery/μL >= Ca200 Ery/μL 3+ >= Ca200 Ery/μL >= Ca200 Ery/μL 3+
NIT
Pos Pos Pos Pos Pos
28
H-100 Urine Analyzer Operation Manual
29
H-100 Urine Analyzer Operation Manual
Appendix D
Rev. 2007/12
30