OpenLAB and EZChrom PDA Analysis
OpenLAB and EZChrom PDA Analysis
OpenLAB and EZChrom PDA Analysis
PDA Analysis
Notices notice, in future editions. Further, to Restricted Rights Legend
the maximum extent permitted by
If software is for use in the
Copyright © Scientific Software, Inc applicable law, Agilent disclaims all
performance of a U.S. Government
1997-2003 © Agilent Technologies, Inc. warranties, either express or implied,
prime contract or subcontract,
2006-2007. with regard to this manual and any
Software is delivered and licensed as
information contained herein, including
No part of this manual may be "Commercial computer software" as
but not limited to the implied
reproduced in any form or by any defined in DFAR 252.227-7014 (June
warranties of merchantability and
means (including electronic storage 1995), or as a "commercial item" as
fitness for a particular purpose. Agilent
and retrieval or translation into a defined in FAR 2.101(a) or as
shall not be liable for errors or for
foreign language) without prior "Restricted computer software" as
incidental or consequential damages in
agreement and written consent from defined in FAR 52.227-19 (June 1987)
connection with the furnishing, use, or
Agilent Technologies, Inc. as governed or any equivalent agency regulation or
performance of this document or of any
by United States and international contract clause. Use, duplication or
information contained herein. Should
copyright laws. disclosure of Software is subject to
Agilent and the user have a separate
Agilent Technologies’ standard
written agreement with warranty terms
Edition commercial license terms, and non-
covering the material in this document
DOD Departments and Agencies of the
May, 2007 that conflict with these terms, the
U.S. Government will receive no greater
Document Revision 3.3B/3.2 warranty terms in the separate
than Restricted Rights as defined in
agreement shall control.
Printed in USA FAR 52.227-19(c)(1-2) (June 1987). U.S.
Technology Licenses Government users will receive no
Agilent Technologies, Inc.
greater than Limited Rights as defined
6612 Owens Dr. The hardware and/or software in FAR 52.227-14 (June 1987) or DFAR
Pleasanton, CA 94588-3334 described in this document are 252.227-7015 (b)(2) (November 1995),
furnished under a license and may be as applicable in any technical data.
used or copied only in accordance with
Warranty such license.
The material contained in this
document is provided “as is,” and is
subject to being changed, without
Contents
1 Using This Guide..............................................................................................................................1
Introduction ............................................................................................................................................................. 1
Export a Spectrum........................................................................................................................................ 17
Overlay Spectra............................................................................................................................................. 17
3D Data Graph............................................................................................................................................... 38
PDA Analysis iv
11 PDA Analysis and Calculations ...................................................................................................44
General ................................................................................................................................................................... 44
About Chromatograms Extracted from the Contour View ............................................................................ 45
Background Correction........................................................................................................................................ 47
Spectrum Interpolation........................................................................................................................................ 47
Spectrum Derivatives........................................................................................................................................... 48
General ................................................................................................................................................................... 49
Pre-Filters............................................................................................................................................................... 49
12 Index................................................................................................................................................53
PDA Analysis v
1 Using This Guide
Introduction
The PDA software provides functionality to extract and analyze spectral data from
Diode Array Detectors.
PDA Analysis 1
To select the view, choose PDA View from the View menu or by selecting Views from
the PDA Toolbar. A right mouse click within any of the view windows provides a
unique menu of options for that view. You must have a data file open in order to
access PDA Views.
Mixed View
3D Views
Contour View
Max Plot View
Chromatogram View
Multi-Chromatogram View
Spectrum View
Ratio View
PDA Analysis 2
Note: During a run, the user must manually refresh this display. It does not update automatically.
Right-click inside the 3D view to display the pop-up menu that gives you access to plot
rotation and axis setup features.
Hold down the left mouse button and move the cursor in the direction you wish
to move the plot. The plot will move as you move the cursor. The movement
option will remain in effect until you turn it off. When finished, click the right
mouse button, and de-select the movement option. If you wish to return to the
original view, select the Reset command.
Note: These values can be viewed or changed using the 3DPlot/Properties dialog.
PDA Analysis 3
Set the 3D Properties
To set or change the properties of the 3D data display,
7. From View menu, select PDA View.
8. Select Mixed View or one of the other views that includes 3D.
9. Do a right mouse click on the 3D view and select Properties.
10. Click the General tab to enter or change the properties for the 3D plot.
Click the Axis Setup tab to set or change the axis parameters.
Style
Select Color or Grayscale for the plot. Checking the Wire Mesh box will cause
the data to be rendered as a wire frame plot, rather than as a solid fill plot.
Colors
Range
Select how you want the coloration on the plot to appear. When Light & Dark
Range is selected, alternating light and dark bands are used. When Full
Spectrum is selected, then a continuous color spectrum is used.
PDA Analysis 4
Background
Select the color for the background of the plot.
Axes
Select the color for the plot axes.
Display
This specifies the relative quality of the displayed contour plot. Lower Display
Detail will result in faster drawing of the plot.
Print
This specifies the relative quality of the printed contour plot. Lower Print
Quality will result in faster printing.
Viewing
Use these fields to view current rotation settings, or to set them manually.
Light source
Selecting this option causes the plot to be shaded as if an external light source
were shining upon it.
Rotation
This reports the current level of rotation (front to back) in the aspect position
of the plot. When a new value is entered, the plot is redrawn to reflect the new
value upon exiting the dialog.
Elevation
This reports the current level of tilt (forward or backward) in the aspect
position of the plot. When a new value is entered, the plot is redrawn to reflect
the new value upon exiting the dialog.
Roll
This reports the current level of roll (side to side) in the aspect position of the
plot. When a new value is entered, the plot is redrawn to reflect the new value
upon exiting the dialog.
Zoom
This reports the current level of magnification in the plot. When a new value is
entered, the plot is redrawn to reflect the new value upon exiting the dialog.
PDA Analysis 5
Set the 3D Axis Properties
To set the axis properties for the 3D plot,
1. From View menu, select PDA View.
2. Select Mixed View or one of the other views that includes 3D.
3. Do a right mouse click on the 3D view and select Properties.
4. Click the Axis Setup tab to enter or change the axis properties for the 3D plot.
PDA Analysis 6
Limits
Enter the limits for the 3D plot.
Time
Click Autoscale if you want to have the software automatically scale the time
axis to the maximum values. To enter a manual range, de-select the autoscale
box, and add your own limits, or click the Get Limits button to enter the limits
displayed on the current 3D graph.
Wavelength
Click Autoscale if you want to have the software automatically scale the
wavelength axis to the maximum values. To enter a manual range, de-select the
autoscale box, and add your own limits, or click the Get Limits button to enter
the limits displayed on the current 3D graph.
Absorbance
Click Autoscale if you want to have the software automatically scale the
absorbance axis to the maximum values. To enter a manual range, de-select the
autoscale box, and add your own limits, or click the Get Limits button to enter
the limits displayed on the current 3D graph.
PDA Analysis 7
Labels
You can customize the labeling of the 3D plot using the parameters in this area.
You can select font, size, color and style using the selections provided.
While the option is active, you can move the plot in the designated way by
holding down the left mouse button and moving the mouse in the desired
direction on the plot. When you have the plot in the desired position, click the
right mouse button and turn off the option by selecting it again (checkmark
removed). If you want to return to the original plot view, click the right mouse
button and select Reset.
PDA Analysis 8
Right-click inside the window to display the pop-up menu. Select Properties to display
the Contour Properties dialog box.
To generate a chromatogram view from contour view of Mixed View or Mixed View w/
3D:
1. Select PDA Views from the View menu, and then select Mixed View to
display the Contour Map, Chromatogram, and Spectrum.
11. Move the cursor to the triangle-shaped handle located on the left- hand
wavelength axis of the Contour Map and press the left mouse button.
12. Drag the cursor up or down to the desired wavelength and release the
mouse button.
13. The chromatogram associated with the specified wavelength is displayed in
the Chromatogram View.
To generate a spectrum view from contour view of Mixed View or Mixed View w/ 3D:
14. Select PDA Views from the View menu, and then select Mixed View to
display the Contour Map, Chromatogram, and Spectrum.
15. Move the cursor to the triangle-shaped handle located on the time axis of
the Contour Map and press the left mouse button.
16. Drag the cursor to the desired peak and release the mouse button
17. The Spectrum associated with the specified retention time value is
displayed in the Spectrum View.
PDA Analysis 9
Style
Select how you want the plot to appear, Grayscale or Color.
Colors
Range: Use this to select how the coloration of the plot is to be displayed. When
Light & Dark Range is selected, alternating light and dark bands are used.
When Full Spectrum is selected, then a continuous color spectrum is used.
Background: Select the color to be used for the background of the plot.
Display
This specifies the relative quality of the displayed contour plot. Less Display
Detail will result in faster rendering of the plot.
Print Quality
This specifies the relative quality of the printed contour plot. Coarse Print
Quality will result in faster printing.
Cursors
The values in these boxes reflect the current contour cursor positions. When
you change the values, the cursors on the plot will be changed when you exit
the dialog or when you click the Apply button.
PDA Analysis 10
Time
This specifies the time position (X Value) of the cursor. When a value is
entered, the X cursor is updated to the new position upon exiting the dialog.
Wavelength
This specifies the wavelength position (Y Value) of the cursor. When a value is
entered, the Y cursor is updated to the new position upon exiting the dialog.
Bandwidth
This specifies the wavelength band that will be averaged when a chromatogram
is extracted from the contour plot. The extracted chromatogram is an average
of the absorbances at each wavelength in the wavelength band. The wavelength
band is equal to the selected wavelength (see above) +/- one half of the
bandwidth.
PDA Analysis 11
Time
Click Autoscale if you want to have the software automatically scale the time
axis to the maximum values. To enter a manual range, de-select the autoscale
box, and add your own limits, or click Get Limits to enter the limits displayed
on the current contour graph.
Wavelength
Click Autoscale if you want to have the software automatically scale the
wavelength axis to the maximum values. To enter a manual range, de-select the
autoscale box, and add your own limits, or click Get Limits to enter the limits
displayed on the current contour graph.
Absorbance
Click Autoscale if you want to have the software automatically scale the
absorbance axis to the maximum values. To enter a manual range, de-select the
autoscale box, and add your own limits, or click Get Limits to enter the limits
displayed on the current contour graph.
Labels
You can customize the labeling of the contour plot using the parameters in this
area. You can select font, size, color and style using the selections provided.
PDA Analysis 12
About the Mixed View
This view displays the contour, chromatogram and spectrum views along with the
fourth pane displaying similarity, purity and peak profile.
To zoom in on a portion of either the contour plot, chromatogram or spectrum, hold the
left mouse button down, move the mouse over the plot until the area of interest is
highlighted and release the mouse button.
To quickly move to the previous level of zoom, double-click on the plot.
To Unzoom to the full plot after multiple zooming operations, use Ctrl+Z, or use Shift +
double-click in the window, or click the right-mouse button anywhere in the window
and select Full Unzoom from the pop-up menu.
Note: You can synchronize the axis and zoom limits with the other two plots using the
Synchronize button on the PDA toolbar. When the contour plot is zoomed, the time
range of the chromatogram and the wavelength range of the spectrum plot are
automatically adjusted to the same range as the contour. When synchronized zooming
has been performed, the x axis plot limits for the chromatogram and spectrum plot are
changed. To unzoom these plots to the limits of the data, you must select Full Unzoom
from the contour plot context menu.
This button presents options that enable you to toggle the view for the lower right pane
of the mixed view:
• Show 3D Plot
• Show Peak Similarity and Threshold Plot
• Show Peak Purity Plot
• Show Spectrum Similarity Table
PDA Analysis 13
Mixed View Actions Button
PDA Analysis 14
About the Max Plot
To view a Max Plot
1. From the Channel Selection box on the toolbar, click the arrow and then
select Spectrum Max Plot.
A Max Plot is a chromatogram with each point plotted at its maximum absorbance. This
plot gives an indication of the appearance of the chromatogram when the wavelengths
are optimized for each peak. This plot is useful in selecting an appropriate Analysis
Channel for each peak. Note: The Max Plot is not an analysis channel and is not used
to analyze chromatographic data.
PDA Analysis 15
About the Spectrum View
This view displays the spectrum associated with a time on the chromatogram.
To change the displayed spectrum,
1. From the View menu, select PDA View and then choose Mixed View, or
select Views from the PDA Toolbar, and then select Mixed View.
2. Drag the X-Axis arrow on the Contour plot to a time where the desired
spectrum is displayed in the Spectrum pane.
To change the working chromatogram,
1. From the Mixed View click and drag the arrow in the Spectrum pane to
the desired wavelength. The chromatogram associated with this
wavelength will be displayed in the Chromatogram pane.
PDA Analysis 16
Notes:
Background Correction may only be performed once on a spectrum. If a second
background correction is attempted on an already corrected spectrum, a message box
is displayed and the operation ignored.
If the current spectrum has not been extracted from an integrated peak of the working
chromatogram, selecting this menu item will have no effect on the spectrum.
Background subtraction must always be the first operation performed on a spectrum. If
a background correction is attempted on a spectrum after another operation has
already been performed (including operation specified in the Spectrum tab of PDA
Options), a message box is displayed and the background correction request ignored.
Interpolate Spectrum
A spectrum interpolation can be performed if the current spectrum has been
extracted from an integrated peak of the working chromatogram.
To perform a spectrum background correction,
1. Do a right mouse click in the Spectrum window and then select Operations
followed by Interpolate.
2. Click the spectrum on which you wish to perform the calculation. A 10:1
interpolation will be performed on the spectrum. The new interpolated
spectrum is then displayed along with the original spectrum.
Refer to Spectrum Interpolation in the PDA Analysis and Calculations
section for details on how this calculation is performed.
Export a Spectrum
To export a spectrum,
1. In the Spectrum pane, do a right mouse click. Select Utilities followed by
Export. A check-mark will appear next to the Export command.
2. Click on the spectrum to be exported as an ASCII file.
3. The Export Spectrum As dialog box will appear. Browse to the folder
where you want to save the exported spectrum and then type a name for
the file in File Name and then click Save.
Overlay Spectra
To overlay more than one spectra in the Spectrum pane,
1. Click Actions from the PDA toolbar, followed by Overlay Spectra...
2.In the dialog box presented, enter a time for which you want the spectrum
displayed and then click Overlay, or drag the cursor on the Contour or
Chromatogram View to the time for which you want the spectrum
displayed.
To remove the spectra you have added to the view,
1. Do a right mouse click in the Spectrum pane, and then select Clear
overlays.
You can also clear selected spectra using the Trace Setup dialog accessed by doing a
right mouse click in the view and then selecting Properties.
PDA Analysis 17
About the Peak Purity Plot
To show the Peak Purity Plot in the 4th pane of the Mixed View
1. From the View menu select PDA View followed by Mixed View.
2. Click Options followed by Show Peak Purity Plot.
3. Click the Action button, and choose Select Peak for Purity/Similarity
Display to select a peak for purity display.
The Total Peak Purity view displays the purity profile for a chromatogram extracted
from the 3D data. The view displays purity information for the Max Plot chromatogram
that is displayed in the Chromatogram pane of the Mixed View. The pane will be blank
until the data has been analyzed, and a peak has been selected for the purity
calculation.
PDA Analysis 18
By default, the Similarity View will use the current method settings from the Purity tab
of PDA Options. To override the method settings, click the Override method settings
box, then enter the values you wish to use.
Wavelength range
Specify the wavelength range over which the purity calculations will be
performed, for example, from 200nm to 400nm.
Wavelength Step
Specify the wavelength spacing (in nm) to be used when purity calculations are
performed.
Purity threshold
The comparison of two spectra gives a Similarity Index (SI), the closer to 1.00
number is the more similar the spectra are. The purity threshold is used to
eliminate spectra that do not match. If the SI is greater than the purity
threshold, the spectra will be considered pure. In general, a spectra with an SI
greater than 0.9900 would have a high probability of being the same as the apex
spectra. A SI greater than 0.9000 but less than 0.9900 shows some similarity
but would need to be evaluated with care and caution. SI less then 0.9000
should be considered non-similar.
Absorbance threshold
This value represents the percentage of peak height that spectra will included
in purity calculation. Spectra in sections of the peak that do not exceed this
threshold will not be included in the purity calculation. This provides a
method of eliminating spectra where the concentration of the compound is so
low that the solvent spectrum interferes.
Background compensation
Checking this box will cause spectra to be corrected for background using the
peak baseline prior to being used in the calculation of purity.
PDA Analysis 19
1. Make sure the data has been analyzed. (If you are not sure, click the
analyze button on the toolbar.)
2. Click Actions followed by Select Peak for Purity/Similarity Display.
3. Select a peak by holding down the CTRL button and then clicking on a peak
from the Chromatogram or Contour pane. The retention time of the
reference spectrum is displayed along with a value indicating peak purity.
4. Continue to select or change the peak. When you are finished, click Close.
By default, the Similarity View will use the current method settings. To override the
method settings, click the Override method settings box, then enter the values you wish
to use.
Wavelength range
Specify the wavelength range over which the purity calculations will be
performed, for example, from 200nm to 400nm.
PDA Analysis 20
Wavelength Step
Specify the wavelength spacing (in nm) to be used when purity calculations are
performed.
Purity threshold
The comparison of two spectra gives a Similarity Index (SI), the closer to 1.00
number is the more similar the spectra are. The purity threshold is used to
eliminate spectra that do not match. If the SI is greater than the purity
threshold, the spectra will be considered pure. In general, a spectra with an SI
greater than 0.9900 would have a high probability of being the same as the apex
spectra. A SI greater than 0.9000 but less than 0.9900 shows some similarity
but would need to be evaluated with care and caution. SI less then 0.9000
should be considered non-similar.
Absorbance threshold
This value represents the percentage of peak height that spectra will included
in purity calculation. Spectra in sections of the peak that do not exceed this
threshold will not be included in the purity calculation. This provides a
method of eliminating spectra where the concentration of the compound is so
low that the solvent spectrum interferes.
Background compensation
Checking this box will cause spectra to be corrected for background using the
peak baseline prior to being used in the calculation of purity.
PDA Analysis 21
The Ratio view displays two PDA wavelength channels and the ratio of those two
channels. These may be viewed during real-time acquisition as well as during post-run
analysis. The flat tops on the ratio peaks are a preliminary indication of peak purity.
The ratio wavelengths and parameters are set in the Ratio tab of PDA Options.
The Y-axis of the ratio chromatogram is auto-scaled to 1.
Right-click within the window to display the pop-up context menu. The context menu is
the same as for a standard chromatogram graph window.
PDA Utilities
To access the Utilities menu,
1. Do a right mouse click from any PDA View and then select Utilities.
The Utilities menu allows displayed spectra to be printed, copied, saved or exported.
Some or all of the following options will be presented.
Print
Select Print to automatically print the currently displayed spectra.
Copy to Clipboard
Select Copy to Clipboard to copy the displayed spectra to the Clipboard. The
contents of the Clipboard may then be pasted into other software.
Save trace
Select Save trace to save the spectrum to a file with an .spc extension for later
inclusion in a library or report.
Export
Select Export to enable export of the item in the method.
PDA Analysis 22
About the Spectrum Similarity Table
The Spectrum Similarity table is displayed in the fourth pane of the PDA Mixed View
when the Options Mode is set to Show Spectrum Similarity Table.
To view the Spectrum Similarity Table
1. In the Mixed View, click Options.
2. Select Show Spectrum Similarity Table.
Spectra are automatically added to and deleted from this table whenever a spectrum is
added to and deleted from the Spectrum pane of the Mixed View.
Spectrum Name
This column displays identifications for each of the spectra added to the table.
For spectra extracted from the 3D data, the identification includes the time.
For spectra extracted from a chromatogram based on a peak name, the
identification includes the peak name. For spectra loaded from a file, the
identification includes the retention time.
A check mark next to the colored line for a spectrum indicates that the
spectrum will serve as the Reference Spectrum for the similarity calculation.
Spectra may be added to the table in the Mixed View by selecting the Add to
Similarity Table from the Actions button menu.
Similarity
This column displays the similarity of the spectrum on that row relative to the
Reference Spectrum. The Reference Spectrum is determined by double-clicking
on a row of the table or by highlighting a row and pressing the ‘Set Reference’
button.
Until a reference spectrum has been selected the similarity for all spectra in the
table will be zero.
Refer to the PDA Analysis and Calculations section for details on how this
calculation is performed.
Source
This column displays the source of the spectrum. (Current Data) refers to
working spectra extracted form the current 3D data. When spectra come form a
stored data file, this column will display the source filename.
PDA Analysis 23
Set Reference
Click this button will designate the spectrum in the currently highlighted row
of the table as the new reference spectrum. The similarities for all of the rows
will be recalculated based on the new reference spectrum.
Print
Click this button will output a simple text report representation of the table to
the default printer.
Properties
Click this button will display a dialog with parameters related to the
calculation of similarity.
Note: During analysis, it is possible that a portion of this wavelength range will be outside the
range of the acquired data. In that case, the wavelength range will be truncated to the limits of
the acquired data.
Extract a Chromatogram
To generate a chromatogram view (extract a working chromatogram) from contour
view of Mixed View or Mixed View w/ 3D:
1. From the View menu, select PDA View followed by Mixed View, or click
Views from the PDA Toolbar and then select Mixed View. This will
display the Contour Map, Chromatogram, and Spectrum.
2. Move the cursor to the triangle-shaped handle located on the left- hand
wavelength axis of the Contour Map and press the left mouse button.
3. Drag the cursor up or down to the desired wavelength and release the
mouse button.
4. The chromatogram associated with the specified wavelength will be
displayed in the Chromatogram View.
PDA Analysis 24
6. Click the desired tab to set the parameters for data analysis.
Tab Used for
Library Enabling spectral libraries and specifying search
parameters.
Purity Entering parameters for purity calculations.
Spectrum Defining filtering and per-peak spectrum
parameters
Multi-chromatogram Enabling multi-wavelength chromatograms
Ratio Setting up ratio plots
Library/Enabled
Enter the spectral library to be searched or select from available libraries by
clicking the Library field followed by clicking the file button. You can select
more than one library. To enable the library for searching, click the Enabled
box. If this list is empty, or no library is enabled, no hits will be returned when
a search is performed.
PDA Analysis 25
Search parameters
Enter the parameters to use for the search.
Wavelength range
Enter the wavelength range to search
Wavelength step
Enter a step number for the search. Larger numbers will make the search
faster, but if you use too large of a step, spectral details may not be picked up.
Max hits
Specify the number of hits that will be reported in the results of a library
search. Note that this works in conjunction with the Similarity Threshold
parameter to limit the number of hits reported.
Similarity threshold
Enter a number for threshold of similarity. The library search results will only
display matches whose similarity to the unknown exceeds this value.
Pre-filters
The options in this group allow you to specify search pre-filters that will be
performed on library spectra prior to the test for similarity. All pre-filters are
optional.
Retention time range
When values are entered in these fields, PDA limits its search to library spectra
obtained from peaks whose apex are within the specified retention time range.
This pre-filter is optional and may be left blank.
Lambda max
When one or more of these values is specified, library search will be restricted
to those library entries containing lambda max value(s) within 5nm of all of the
specified values. Entries without matching lambda max value(s) are
automatically excluded from the search (no similarity calculation is made).
Entering values for lambda max is optional.
Compound name filter
If you are searching only for spectra that contain a certain name, enter the
name here. Only spectra containing that name will be searched.
PDA Analysis 26
Purity Calculations
From/To – wavelength range
Specify the wavelength range over which the purity calculations will be
performed, for example, from 200nm to 400nm.
Step
Specify the wavelength spacing (in nm) to be used when purity calculations are
performed.
Purity threshold
The comparison of two spectra gives a Similarity Index (SI), the closer to 1.00
number is the more similar the spectra are. The purity threshold is used to
eliminate spectra that do not match. If the SI is greater than the purity
threshold, the spectra will be considered pure. In general, a spectra with an SI
greater than 0.9900 would have a high probability of being the same as the apex
spectra. A SI greater than 0.9000 but less than 0.9900 shows some similarity
but would need to be evaluated with care and caution. SI less than 0.9000
should be considered dissimilar.
Absorbance threshold
This value represents the percentage of peak height that spectra will included
in purity calculation. Spectra in sections of the peak that do not exceed this
threshold will not be included in the purity calculation. This provides a method
of eliminating spectra where the concentration of the compound is so low that
the solvent spectrum interferes.
Background compensation
Checking this box will cause spectra to be corrected for background using the
peak baseline prior to being used in the calculation of purity.
PDA Analysis 27
Per-peak spectrum calculations
Checking any of these boxes indicates that the indicated value will be
calculated on a per-peak basis during analysis. The result of this calculation
will then be available in reports and as chromatogram annotations.
Unchecking values that are not of interest will speed up analysis. If a box is unchecked
and the field appears in a run report, it will be reported as zero
Total Purity
Checking this box will cause total peak purity to be calculated on a per-peak
basis during analysis. The result of this calculation will then be available in
reports and as chromatogram annotations. If this box is unchecked and a
report contains the field, a zero will be reported.
3 Point Purity
Checking this box will cause 3-point peak purity to be calculated on a per-peak
basis during analysis. The result of this calculation will then be available in
reports and as chromatogram annotations. If this box is unchecked and a
report contains the field, a zero will be reported.
Note: The processing specified on this page is performed prior to any use of the spectra in
the software, including display, searching and reporting.
PDA Analysis 28
Spectral filtering
In this area you designate how spectral filtering (if any) will be performed.
Filtering type
Choose the type of filtering for the spectral plot. Choices are None, Smooth, 1st
Derivative, and 2nd Derivative. Selecting one of these smoothing algorithms can
remove noise from the spectrum.
Background correction
If this box is checked, then, prior to display, a correction for spectral
background is made, as follows:
1. The spectra from the baseline start and baseline stop times for the peak are
extracted from the data.
2. For each spectrum in the peak, a corresponding background spectrum is
generated by linear interpolation between the baseline start and baseline stop
spectra.
3. This background spectrum is subtracted from the original spectrum.
Note: The baseline start and stop times that are used in calculating background
compensation are based on the detected peaks for the channel currently selected.
Interpolate spectra
Select this check box to automatically perform 10:1 interpolation of spectra
displayed in the spectrum window using a cubic spline curve fit. This
interpolation is performed after the applying any spectral filtering option (1st
derivative, 2nd derivative or smooth) to the display spectrum.
PDA Analysis 29
Per-peak spectrum calculations
Checking any of these boxes indicates that the indicated value will be
calculated on a per-peak basis during analysis. The result of this calculation
will then be available in reports and as chromatogram annotations.
Unchecking values that are not of interest will speed up analysis. If a box is
unchecked and the field appears in a run report, it will be reported as zero
Similarity
Checking this box causes the Peak Apex Similarity to the reference spectrum to
be calculated on a per-peak basis during analysis. The result of this calculation
will then be available in reports and as chromatogram annotations. If this box
is unchecked and a report contains the field, a zero will be reported.
Peak Apex Similarity applies only to named peaks and its use requires that a
reference spectrum be included in the peak table.
Upslope Similarity
Checking this box causes upslope similarity to be calculated on a per-peak
basis during analysis. Upslope similarity compares the peak apex spectrum to
the spectrum at the peak inflection point on the upslope side (to the left of the
apex). The result of this calculation will then be available in reports and as
chromatogram annotations. If this box is unchecked and a report contains the
field, a zero will be reported.
Downslope Similarity
Checking this box causes downslope similarity to be calculated on a per-peak
basis during analysis. Downslope similarity compares the peak apex spectrum
to the spectrum at the peak inflection point on the downslope side (to the right
of the apex). The result of this calculation will then be available in reports and
as chromatogram annotations. If this box is unchecked and a report contains
the field, a zero will be reported.
Lambda Max
Checking this box indicates causes the lambda max (wavelength at which the
highest absorbance occurs) to be computed for each peak of each multi-
chromatogram extracted from the 3D data during analysis. The result of this
calculation will then be available in reports and as chromatogram annotations.
If this box is unchecked and a report contains the field, a zero will be reported.
PDA Analysis 30
Multi-Chromatogram view is not functional unless one or more valid wavelengths are
enabled in the Multi-Chromatogram Definition tab.
Enabled
To enable a wavelength, click the checkbox. To disable a wavelength, click the
checkbox again to remove the checkmark. Only enabled wavelengths will
appear in the multi-chromatogram view.
Wavelength
Enter the wavelength you want to view.
Bandwidth
Specify the nm range to be averaged in generating the analog signal for each
channel.
The multi-chromatogram data is calculate by averaging the data points of
chromatograms across the wavelength range of "wavelength - 1/2*bandwidth"
to "wavelength + 1/2*bandwidth".
For example, given a wavelength of 600 nm and a bandwidth of 4 nm, for each
sample time, the data points from the chromatograms at 598, 599, 600, 601 and
602 nm are added, and the sum divided by 5.
If, in the example above, the detector's range is 190 - 600 nm (inclusive), the
software only averages the points across chromatograms at 598, 599 and 600
nm (dividing by 3). The range would not be exceeded.
PDA Analysis 31
Wavelength
Enter the wavelength at which you wish to extract a chromatogram to be used
for baseline subtraction.
Bandwidth
Specify a bandwidth range to be averaged in generating the extracted
chromatogram.
Ratio plot
These controls specify the wavelength of the ratio multi-chromatogram
channels. The extracted chromatograms will be centered about the specified
wavelength.
Channel 1, Channel 2
Wavelength
Enter the wavelength of the chromatogram channel.
Bandwidth
Enter the nm range to be averaged in generating the chromatogram channel.
Threshold
Enter the ratio threshold. The threshold is the minimum absorbance value
required in the chromatograms of both channels for calculation of the ratio. It
is a method of eliminating ratio calculation at minor absorbance values, such as
those that occur with noise. If the threshold value is not met, the ratio plot will
be zero.
PDA Analysis 32
5 Spectral Library Search
Define a Spectral Library
In order to define a library, you must have the spectra to be added saved on disk as
spectra files (.spc) or you can create a library from the spectra obtained during the
current run. To define a spectral library
1. From the File menu, select Spectral Library followed by New to create a
new spectral library. Select Open to browse for and open an existing
library to modify.
2. Click Spectrum Data Source and select Current Data, Spectrum File or
Named Peak.
If you select Current Data, a dialog box will appear where you enter the retention
time for the spectra to be added to the library.
If you select Spectrum File, browse to the spectrum file to be added to the library
and then click OK.
If you select Named Peak, chose a peak whose spectrum is to be added to the
spectral library from the dialog box presented. (The current chromatogram must be
Analyzed first in order to use this option.)
PDA Analysis 33
If you chose to enter a spectrum from a named peak, a dialog box will appear
where you must select the named peak from which the spectrum will be
extracted.
Component Name
Enter the chemical name or description of the spectrum. This field is used for
display and search pre-filter purposes.
Retention time
When the spectrum was extracted from a chromatographic run, the software
automatically identifies the retention time of the spectrum and enters the value
in this field. When desired, the value in this field may be edited. This field can
also be used as a search pre-filter.
Comment
Enter any descriptive information desired related to the spectrum. This field is
used only for display and documentation purposes.
Library Notes
This field is used to document relevant information about the library as a
whole, e.g., documentation about run conditions or general sample information.
Note: Fields in existing library entries may be edited by selecting the fields with the cursor.
Right-click on a row in the library table to display a pop-up menu used to cut, paste,
copy, and insert and delete lines from the library as necessary.
PDA Analysis 34
Hits Displayed
Select the number of hits you wish to display. Hits will be displayed in order of
similarity.
Search
Select Method if you want to use the parameters from the current method for
your search. Select Quick if you want to enter or change the parameters
yourself before the search.
Display Params
Click this button if you want to display the parameters. It toggles between
display and no display. If you always use the method parameters, you may not
wish to display them when you search.
Search Now
When you click this button, a library search will be performed on the selected
spectra using the parameters, and the number of hits requested will be
displayed.
Search Spectrum
Click the arrow to select a new spectrum to search on. You can choose from the
current data, or from a stored spectrum file. The current spectra source will be
displayed.
Params
This tab contains search parameters. If you have elected to use the Method
search parameters, the parameters displayed will be those in your current
method in the Library tab of the PDA Options. If you have elected to do a
Quick search, the parameters will be default parameters.
Wavelength Range
These specify the wavelength range over which the library search will be
performed.
PDA Analysis 35
Wavelength Step
This specifies the data point spacing to be used when a library search is
performed.
Max Hits
This specifies the number of hits that will be reported in the results of a library
search. Note that this works in conjunction with the Similarity Threshold
parameter to limit the number of hits reported.
Similarity Threshold
Specifying a value for this control will cause the library search results to only
display matches whose similarity to the unknown exceeds this value. Note that
this control works in conjunction with the Max Hits parameter to limit the
number of hits reported.
Library
Displays the library from the method, or for Quick enables you to select a
library to be used.
Pre-Filters
When Quick is selected as the Search Mode, the items on this tab allow the
user to specify search pre-filters that will be performed on library spectra prior
to the test for similarity. All pre-filters are optional. If Method is selected as the
Search Mode, the items on this tab are read-only and reflect the parameters
values on the Library tab of PDA Options.
Retention Time Range
When a Retention Time Range is specified, library search will be restricted to
those library entries whose retention time is within the specified range. Entries
outside this range are automatically excluded from the search (no similarity
calculation is made). Entering a value for this is optional.
Lambda Max
When one or more of these values is specified, library search will be restricted
to those library entries containing a lambda max within 5 nm of one of the
specified values. Entries without a matching lambda max are automatically
excluded from the search (no similarity calculation is made). Entering values
for this is optional.
Compound Name Filter
When a Compound Name Filter is specified, library search will be restricted to
those library entries whose name contains the specified string as a case-
insensitive sub-string. Entries without a matching sub-string are automatically
excluded from the search (no similarity calculation is made). Entering a value
for this is optional.
PDA Analysis 36
4. Enter the Retention Time, Spectrum File, or Named Peak to be used for
the search.
PDA Analysis 37
spectrum to be added to the library. Files are saved automatically with the
.spc extension.
Note: To add the 1st or 2nd Derivative of a spectrum to the library, select the appropriate filter in the
Spectrum tab of PDA Options and repeat steps 3 and 4.
3D Data Graph
The 3D View provides a three-dimensional view of absorbance versus time and
wavelength. Wavelengths of appreciable absorbance and interferences, which may be
invisible in a single wavelength plot, are easy to locate with the 3D View. Select this
function to automatically enter the 3D map into the report. Click the right-mouse
button within the 3D map and select Properties to display the 3D Properties dialog box
and enter appropriate changes
The 3D map can be elevated and rotated around its axis for display from any angle.
These functions work the same as in the 3D view window.
PDA Analysis 38
1. Click Edit Custom Report to open the custom report.
2. On the custom report, do a right mouse click and select Insert Report
followed by Library Search Report. The Library Search Report template
will be inserted at the location of the cursor in the custom report.
To modify the parameters of the report, do a right mouse click in the report table, then
select Properties.
PDA Analysis 39
A Specific Spectrum
Select this to specify the spectrum to be searched. When this is selected, click
the button to select the spectrum source. The menu contains the following
choices: Current Data, Named File, or Spectrum Peak.
Spectral Display
The in this group specify the layout of the search results graphs.
Graph Height
This specifies the relative height of a search results graph. A value of 100%
corresponds to a standard sized graph. A larger value may be selected to
provide more a detailed plot.
Graph Properties
Pressing this button displays the graph properties dialog as found on the
standard Spectrum graph.
Enter or select the library you wish to include in the report, then click OK.
PDA Analysis 40
Detected on
This allows the user to select the channel or channels from which peaks will be
detected.
Limit to named peaks only
This allows the user to restrict the peak selection (made above) to named
peaks.
Limit to time range
This allows the user to restrict the peak selection (made above) to a time range.
When checked, only peaks within the specified time range will be searched.
When this box is unchecked, the Time Range edit fields are disabled.
Graph
Select the type of curve you want to graph - Purity Curve or
Similarity/Threshold Curve.
Height %
This value will enable you to "zoom" the graph. A value of 100% will maximize
the curve to the page.
Graph Properties...
Pressing this button displays the graph properties dialog as found on the
standard graph.
PDA Analysis 41
will contain either the apex spectrum, an average of the upslope, apex and downslope
spectra, or several averaged spectra.
To set parameters for this report, do a right mouse click on the table, and then select
Properties.
Report information
Checking any of these indicates that the indicated value will be calculated and
printed to the right of the graph. Unchecking values that are not of interest will
speed up analysis.
Peak Area
Checking this indicates that the peak area should be printed to the right of the
spectrum graph.
PDA Analysis 42
Lambda Max
Checking this indicates that the three largest lambda max values should be
printed to the right of the spectrum graph.
Lambda Min
Checking this indicates that the three largest lambda min values should be
printed to the right of the spectrum graph.
Total Purity
Checking this indicates that total peak purity should be printed to the right of
the spectrum graph.
3 Point Purity
Checking this indicates that 3 point peak purity should be printed to the right
of the spectrum graph.
Similarity
Checking this indicates that peak apex similarity to the reference spectrum
should be printed to the right of the spectrum graph. The use of this value
requires that a reference spectrum be included in the peak table. See the
section on Analysis and Calculations for details.
Spectral Display
The selections in this group specify the content and labeling of the spectrum
graphs.
Show Lambda Maxima
Checking this box will annotate each spectrum graph with the 3 largest lambda
max absorbance.
Show Lambda Minima
Checking this box will annotate each spectrum graph with the 3 largest lambda
minima absorbance.
Background Correction
Checking this will cause each apex spectrum to be corrected for background
using the chromatographic baseline prior to being used elsewhere in analysis.
Refer to the PDA Analysis and Calculations section for details on the formula
used.
Filtering Type
This allows the user to specify a mathematical filtering function to be
performed on all spectra extracted from the 3D data during analysis. Refer to
the Analysis and Calculations section for details on the formula used for each
filter.
Graph Height
This specifies the relative height of the spectrum graph. A value of 100%
corresponds to a standard sized graph. A larger value may be selected to
provide more a detailed plot.
Graph Properties
Click this button to display the graph properties dialog as found on the
standard Spectrum graph.
PDA Analysis 43
10 Changes to the Peak Table for PDA Option
When using the PDA option, certain columns will be added to the peak table that
enable you to analyze peaks from the PDA detector.
Detection
Select the basis for the identification of the peak. If you choose Ret Time, only
the retention time will be used for identification of the peak. If you choose Ret
Time with Spectral Confirm, the Similarity of the peak’s spectrum to that of a
designated reference spectrum will be used in addition to the retention time as
the basis of peak identification.
Spectrum
If you want Similarity to be used as a basis for peak identification, then click
on the arrow to the right of this field to specify the stored reference spectrum
to be used for comparison. During identification, this reference spectrum is
compared to the peak apex spectrum and a similarity index is computed. A
peak is considered identified if this calculated similarity index is at least the
value specified in the Similarity column of the peak table.
If Similarity is not specified as a basis for peak identification, then this field is
ignored.
Similarity
If Similarity is specified as a basis for peak identification, then this field
specifies required minimum similarity index for a peak to be considered
identified. During identification, the reference spectrum (see previous section)
is compared to the peak apex spectrum and a similarity index is computed. A
peak is considered identified if the calculated similarity index is at least the
value in this field.
If Similarity is not specified as a basis for peak identification, then this field is
ignored.
Analysis Channel
Specify which Diode Array wavelength channel is to be used for analysis of the
peak. The choices will be those specified in Instrument Setup/DAD.
This section describes the various types of calculations that are related to PDA data
and how the data is analyzed.
General
In order to maintain accuracy during the application of multiple operations, all
calculations are performed using double-precision floating-point numbers.
Analysis spectra
Background correction
PDA Analysis 44
Calculating total purity
Chromatograms extracted from the 3D data
Lambda max/min calculations
Library search calculations
Multi-chromatogram channels
Peak purity calculations
Pre-filters
Ratio chromatogram calculation
Similarity calculations
Spectra extracted from the 3D data
Spectrum derivatives
Spectrum interpolation
Spectrum operations
Spectrum smoothing
Three-point purity calculations
Upslope and downslope spectra
Working chromatogram
Working spectrum
PDA Analysis 45
• In the Channel Selection drop down list, each multi-chromatogram channel
have a unique identifier that includes the wavelength and bandwidth of the
channel.
• Each Multi-Chromatogram Channel have a separate Integration Events Table,
Manual Integration Table, Export Table and Performance Table associated with
it.
• All Multi-Chromatogram Channels share a common peak and group table.
Within the peak and group tables, any channels may be selected as the analysis
channel for quantitative information.
• When an analysis is performed, all multi-chromatogram channels are
automatically analyzed.
• Each multi-chromatogram channel is an average of the absorbances monitored
at each wavelength in the wavelength range. The wavelength range is equal to
the selected wavelength +/- one half of the bandwidth.
PDA Analysis 46
About Analysis Spectra
The following applies to the analysis spectra:
• Prior to being used elsewhere, all analysis spectra extracted from the data are
filtered according to the settings on the Spectrum tab of PDA Options.
Background Correction
Background Correction may only be performed on a spectrum as a result of the settings
on the Spectrum tab of PDA Options, the settings on the Purity tab of PDA Options,
or by setting the properties of the Peak Purity Pane.
Background subtraction is always the first operation performed on a spectrum. If a
background correction is attempted on a spectrum after another operation has already
been performed (including an operation specified in the Spectrum tab of PDA
Options), a message box is displayed and the background correction request ignored.
Background subtraction may only be performed once on a spectrum. If a second
background correction is attempted on a spectrum after the operation has already been
performed (including the operation specified in the Spectrum tab of PDA Options), a
message box is displayed and the background correction request ignored.
A background correction is performed as follows:
1. The spectra from the baseline start and baseline stop times for the peak are
extracted from the 3D data. The Max Plot is used to determine the peak that is used.
2. For each spectrum in the peak, a corresponding background spectrum is generated
by linear interpolation between the baseline start and baseline stop spectra.
3. These background spectra are subtracted from the original spectra.
Note: While compensation for background will provide more precise purity calculations, it can slow
down re-analysis of large data files.
Spectrum Interpolation
Spectrum interpolation may be performed on a spectrum as a result of the settings on
the Spectrum tab of PDA Options or by doing a right mouse click in the Spectrum
window and then selecting Operations followed by Interpolate.
Interpolated spectra may not be stored in spectrum libraries.
The calculation is performed by doing a 10:1 interpolation of the spectrum data points
using a cubic spline curve fit. This interpolation is performed after the applying any
spectral filtering option (1st derivative, 2nd derivative or smooth) to the spectrum.
PDA Analysis 47
Upslope and Downslope Spectra
The upslope and downslope spectra of a peak are identified by calculating the second
derivative of the portion of the chromatogram containing the peak. The two times at
which the second derivative plot crosses zero are known as the inflection points. For
normal peaks (i.e. non-Negative peaks), the Upslope Spectrum is the spectrum at the
time represented by the first inflection point, while the Downslope Spectrum is the
spectrum at the time represented by the second inflection point.
Spectrum Operations
The following table describes what operations can be done on what type of spectra
Raw Spectra (no filtering Smoothed, 1st Derivative and Interpolated Spectra
or operations) 2nd Derivative Spectra
Search against Yes
Libraries
Similarity Yes
Peak Identification Yes, but generate an error if the operations that were performed on the reference
using spectral spectrum do not match the current method.
confirmation
Appear in Reports Yes Yes, but label must include what operations were
performed
Purity Yes No.
Add to Spectrum Yes No
Library
Spectrum Smoothing
Spectrum smoothing may be performed on a spectrum as a result of the settings on the
Spectrum tab of PDA Options or by doing a right mouse click in the Spectrum
window, and then selecting Operations followed by Smooth.
Smoothing may be performed repeatedly on a spectrum.
The calculation is performed by doing a 9 point Savitsky-Golay smooth on the spectrum
data points.
Spectrum Derivatives
Calculation of the 1st and 2nd derivatives of a spectrum may be performed on a
spectrum as a result of the settings on the Spectrum tab of PDA Options or by doing a
right mouse click in the Spectrum window, and then selecting Operations followed by
1st or 2nd Derivative .
Derivatives may be computed repeatedly on a spectrum.
The absorbance values of the 1st derivative of a spectrum are computed by calculating
the differences between adjacent absorbance values to create a new spectrum. The 2nd
PDA Analysis 48
derivative of a spectrum is defined as the 1st derivative of the 1st derivative of the
spectrum.
General
During analysis, if one or more search libraries are defined on the Library tab of PDA
Options, then an automated library search is performed on every integrated peak of
every PDA analysis channel.
Note: Unless a Library Search Results object is part of the method custom report, no
automated library searching will done when analysis is performed.
In this section, a Query Spectrum is defined the unknown spectrum that is being
searched. A Reference Spectrum is defined as a spectrum from a spectrum library file.
During a search, the apex spectrum of the peak (the query spectrum) is compared to
each spectrum contained in the libraries (reference spectra) to determine the similarity
of the query spectrum to the reference spectrum. The similarity is quantified through
the calculation of a Similarity Index for each query/reference pair. The Similarity
Indices are used to generate a hit list of the best matching entries. A perfect match will
have a Similarity Index of 1.0000. Similarity indices less than 1 indicate differences in
the spectral patterns.
If the query spectrum and the reference spectrum have different wavelength ranges,
then the intersection of the two ranges is used in the similarity calculation.
If the query spectrum and the reference spectrum have different wavelength steps, then
the higher resolution spectrum is de-resolved to match the other spectrum before being
used in the similarity calculation.
Pre-Filters
A pre-filter is a criterion on a reference spectrum that must be met before that
spectrum is used in similarity calculations. One or more pre-filters may be specified in
the search parameters. When multiple pre-filters are specified, a reference spectrum
must meet all of the individual pre-filters in order to be considered for similarity
calculations.
Reference spectra that do not meet all of the pre-filter criteria are automatically
excluded from calculations and from being a candidate for a hit list. No similarity
calculation is performed on these spectra.
The following pre-filters are supported:
Retention Time Range: When a Retention Time Range is specified, a reference
spectrum is excluded unless the retention time for the
spectrum in the library is within the specified range.
Lambda Max: Up to three Lambda Max values may be specified. When one or
more Lambda Max values are specified, a reference spectrum is
excluded unless all of the specified maxima are within 5 nm of
a maxima for the spectrum in the library.
Compound Name: When a Compound Name is specified, a reference spectrum is
excluded unless the specified string is a case-insensitive sub-
string of the compound name for the spectrum in the library.
PDA Analysis 49
Ratio Chromatogram Calculation
The ratio chromatogram plot consists of data points calculated as follows:
Ratio Pt. = abs1 / sqrt(abs1 * abs1 + abs2 * abs2)
where: abs1 = the absorbance in chromatogram 1 at this wavelength
abs2 = the absorbance in chromatogram 2 at this wavelength
If the absorbance of any point in chromatogram 1or chromatogram 2 is less than the
threshold, the Ratio Pt. for that wavelength is set to 0.
Similarity Calculations
The Similarity Index (SI) compares two spectra across the wavelength range defined in
Method > PDA Setup > Library. A Library Search is performed using the Similarity
Index, determined as follows:
A spectrum is considered as an array of absorbances at each wavelength.
(a (λ1), a (λ2), ...., (a (λn))
Where a(λi) is the absorbance at λi
The spectrum can also be represented as an n-dimensional vector, with each
absorbance corresponding to one dimension of the vector, as follows:
S = (a (λ1), a (λ2), ...., (a (λn))
When two spectra are obtained from the same compound, the ratio between
corresponding elements in S1 and S2 is constant; consequently, the vectors have the
same direction. In such a case, the angle between the two vectors would be 0. As a
general rule, the greater the angle between the two vectors, the less similarity they
have. This relationship is calculated as follows:
The closer the SI value is to 1, the more similar two spectra are considered to be. Two
spectra are considered to be identical when the SI value is close to 1.
PDA Analysis 50
Peak Purity Calculations
The purity of a peak is calculated by comparing each spectrum in the peak to the apex
spectrum using the calculation given in the Similarity Calculation section. To derive a
purity number, the peak area under the spectra whose similarity meets the purity
threshold will be divided by the total area of the peak.
PDA Analysis 51
12 Index
3 Changes to the Peak Table for PDA Option. 44
About the Ratio View....................................... 21 How to add spectra to an existing library .... 33
About the Spectrum Similarity Table............ 23 How to Collect Spectra for a Library............. 37
About the Working Chromatogram ............... 46 How to Perform a Manual Library Search ... 37
B I
C L
PDA Analysis 53
Library Search.................................................... 37 Purity View...................................................19, 21
PDA Analysis 54
T U
XY Ratio .............................................................. 32
PDA Analysis 55