Stratified Sampling of Capsule
Stratified Sampling of Capsule
Stratified Sampling of Capsule
Alex Viehmann
PQRI
06 October 2015
Disclaimer
This presentation is not official Agency policy. My
statements and advice do not bind or otherwise
obligate or commit the agency to the views expressed.
2
Agenda
• Sampling Methods
• Distribution considerations
• Quality levels – producer / consumer risk
• USP <905> - what does it provide?
• Acceptance criteria design
3
Sampling Methods
• Simple Random Sampling
– Randomly pick units from lot
• Stratified Sampling
– Divide Lot into strata where the combined strata
cover the entire lot
– Use random sampling within each strata.
• Systematic Sampling
– Sample at specified points usually equally spaced
throughout the lot.
• Random vs. Stratified/Systematic
– Random = no estimation of between/within location
variability
Comparison of Sampling Methods
•CU(%LC)
Distribution considerations
6
Parametric vs Distribution-free
• Parametric: Assume the data follow a specific distribution
– Example: Not enough evidence to conclude the distribution is non-
normal
• Distribution-free: Do not assume the data follow a specific
distribution
• Parametric tests can perform well with skewed and non-
normal distributions and when the spread of each group is
different
– Follow sample size guidelines
• Limited knowledge/data to confirm distribution or severely
skewed distributions (best represented by median) =
distribution-free techniques
7
Quality Levels
Acceptable: The quality level that would be
accepted with a high probability (e.g. 95%)
Producers risk
M is defined as follows:
If T is less than or equal to 101.5%LC, and
(i) If X is less than 98.5%LC, then M = 98.5%LC.
(ii) If X is between 98.5 and 101.5%LC, then M = X .
(iii) If X is greater than 101.5%LC, then M = 101.5%LC.
If T is greater than 101.5%LC, and
(i) If X is less than 98.5%LC, then M = 98.5%LC.
(ii) If X is between 98.5 and T, then M = X .
(iii) If X is greater than T, then M = T.
T is the Target content per dosage unit at the time of manufacture, expressed as percentage label
claim. Unless otherwise specified in the individual monograph, T is 100.0%LC.
9
USP <905> background
• The procedure is based on a two-sided tolerance interval
approach.
– An interval that contains p percent of the population
measurements
• N=10 / k2=2.4
• N=30 / k2=2.0
– k2 is a tolerance interval factor that is affected by
sample size, desired confidence, and coverage (k2 is
specific to a two sided tolerance interval)
– K2 is determined so that the interval covers at least a
proportion p of the population with a confidence c
• Based upon the criteria of the test (k2 value), the metrics
provided are – 84% confident that 91% of the population lies
between +/- 15%
• The second aspect of the procedure is no tablet will be outside ~
73.9-126.9%
10
USP <905> - Acceptance Region / OC curve
Acceptance Limits
8
7
6
5
Max SD
n=30
4
3 n=10
2
1
0
117
115
113
111
109
107
105
103
102
99.5
98.5
97
95
93
91
89
87
85
83.5
Mean
11
Blend Uniformity
• Draft guidance acceptance criteria was inconsistent
with current Agency thinking
– Old recommendation: sample multiple units (replicates) from
multiple locations (e.g. 3 samples from 10 locations) and only
assay 1 per location = total of 10 measurements
• Mean (90-110%) / SD < 5%
– Level II Q/A recommendation: assay all replicates
• Allow for estimation of between/within location variance
components
http://www.fda.gov/Drugs/GuidanceComplianceRegulatoryInforma
tion/Guidances/ucm124782.htm
12
Variance Components
• Partitions total variability
into its individual sources
of variation
• Examples:
– Between/within locations
– Between/within batches
13
Designing appropriate acceptance criteria
• What is the sampling method?
– Random, stratified or systematic
• What are the quality levels that are required for this
product/process?
14
Example I
• Sampling Method = Systematic
• Data = Continuous / Normal
– Sample size = 60 (15x4)
• Assurance requirements = 90% confident that at least
95% of samples will meet USP standard
– Coverage = true proportion of samples from the batch that
will meet USP <905>
• Method = ASTM E2709 / 2810; sampling plan 2
15
Example: Sampling Plan 2 (15 X 4) - Data
Result
Location 1 2 3 4
1 97.08 99.72 98.37 97.50
2 99.72 100.32 101.01 100.29
3 99.90 98.27 98.88 97.96
4 98.78 98.17 98.94 97.78
5 96.32 96.61 99.66 97.20
6 100.97 102.17 99.06 98.80
7 97.02 97.35 98.65 99.98
8 99.39 98.81 98.63 98.06
9 99.59 97.80 97.67 98.95
10 97.97 98.54 100.26 98.74
11 96.09 98.61 97.49 97.50
12 98.87 97.81 97.28 98.80
13 101.10 102.60 100.48 98.62
14 100.80 100.34 98.49 100.93
15 99.70 100.09 100.14 99.20
Example: Sampling Plan 2
Example: Sampling Plan 2
Descriptive Statistics
Overall Mean 98.93
SE (within-location Std Dev) 1.07
Standard deviation of Location Means 1.06
SE LL UL LL UL LL UL LL UL
AQL = 0.25% (95% probability the lot will be accepted if % defective ≤ 0.25%)
UQL = 2% (90% probability the lot will be rejected if % defective ≥ 2%)
20
Example III
• Sampling Method = Random
• Data = Continuous / Normal
– Spec: 90-110%
• Sample size = 60
• Assurance requirements = 95% confident that at
least 95% batch will be in spec.
– Coverage = true proportion of batch within desired spec
• NMT 2.5% below 90%; NMT 2.5% above 110
• Method = Two one-sided Tolerance Interval test
21
Example III acceptance criteria
• Xbar + (2.38*s) < 110
• Xbar – (2.38*s) > 90
22
Summary
• Sampling method needs to be defined
– Dictates the ability to quantitate between/within variance
components
• Distribution needs to be evaluated
– Dictates the method to apply
• Attribute (spec range) needs to be relevant for the
product
– Is 85-115% appropriate for all drug products?
• Quality levels = Risk Based
– Should not be static across all products
• Firm needs to clearly identify the assurance the plan
provides
– OC curve
23