Final - Schedule M III
Final - Schedule M III
FOR
AND
IN-VITRO DIAGNOSTICS
(12.06.2015)
1. 1
General
2. 1
Applicability
3. 1
Terms and definitions
4. 3
Quality management system
5. 5
Management responsibility
6. 7
Resource management
7. 8
Product realization
8. 14
Measurement, analysis and improvement
9. 17
Annexure A- Documents related to Site Master File
11. 20
Annexure C-Environmental requirement for Notified Devices
1. General
This schedule specifies requirements for a quality management system that shall be used by
the manufacturer for the design and development, manufacture, packaging, labeling, testing,
installation and servicing of medical devices and in-vitro diagnostics. If the manufacturer does
not carry out design and development activity, the same shall be recorded in the quality
management system and shall exclude the provisions of relevant clauses of this schedule.
The manufacturer shall ensure the claims of conformity with this schedule to reflect the
exclusions.
If any requirement(s) in Clause 7(product realization) of this schedule is(are) not applicable
due to the nature of the medical device(s) and in-vitro diagnostics(s) for which the quality
management system is applied, the manufacturer does not need to include such a
requirement(s) in its quality management system.
The processes required by this schedule, which are applicable to the medical device(s)and
in-vitro diagnostic devices, but which are not performed by the manufacturer are the
responsibility of the manufacturer and are accounted for in the manufacturer’s quality
management system.
If a manufacturer engages in only some operations subject to the requirements in this part,
and not in others, that manufacturer need only comply with those requirements applicable to
the operations in which it is engaged.
It is emphasized that the quality management system requirements specified in this schedule
are complementary to technical requirements for products.
Manufacturers of components or parts of finished devices and in-vitro diagnostics are
encouraged to use appropriate provisions of this regulation as guidance.
2. Applicability
The provisions of this Schedule shall be applicable to manufacturers of finished devices, In-
Vitro Diagnostics, mechanical contraceptives (Condoms, intrauterine Devices, Tubal Rings),
surgical dressings, surgical bandages, surgical staplers, surgical sutures and ligatures, Blood
and Blood Components Collection bags, intended for human or animal use that is
manufactured in India.
1|Page
3.3 Advisory notice
Notice issued by the manufacturer, subsequent to delivery of the medical device and in-vitro
diagnostic devices, to provide supplementary information and/or to advise what action should
be taken in
the use of a medical device and in-vitro diagnostic devices,
the modification of a medical device and in-vitro diagnostic devices,
the return of the medical device and in-vitro diagnostic devices to the organization that
supplied it, or
the destruction of a medical device and in-vitro diagnostic devices
3.6Component means any raw material, substance, piece, part, software, firmware, labeling,
or assembly which is intended to be included as part of the finished, packaged, and labeled
device.
3.7 Design input means the physical and performance requirements of a device that are used
as a basis for device design.
3.8Design output means the results of a design effort at each design phase and at the end
of the total design effort. The finished design output is the basis for the device master record.
The total finished design output consists of the device, its packaging and labeling, and the
device master record.
3.10 Finished device means any device or accessory to any device that is suitable for use or
capable of functioning, whether or not it is packaged, labeled, or sterilized.
3.11 In-vitro Diagnostic devices: In-vitro Diagnostics referred in this schedule means
diagnostics that fall under section 3(b)(i) of Drugs and Cosmetics Act
2|Page
procedures, that these procedures are implemented effectively, and that these procedures
are suitable to achieve quality system objectives.
3.15Quality policy means the overall intentions and direction of an organization with respect
to quality, as established by management with executive responsibility.
3.17Rework means action taken on a nonconforming product so that it will fulfill the specified
DMR requirements before it is released for distribution.
3.18Specification means any requirement with which a product, process, service, or other
activity must conform.
3| P a g e
4.2.1 General
The quality management system documentation shall include
For each type or model of medical device/ In-vitro Diagnostics, the manufacturer shall
establish and maintain a file either containing or identifying documents defining product
specifications and quality management system requirements. These documents shall define
the complete manufacturing process and, if applicable, installation.
Data may be recorded by electronic data processing systems or other reliable means, but
documents and record relating to the system in use shall also be available in a hard copy to
facilitate checking of the accuracy of the records. Wherever documentation is handled by
electronic data processing methods, authorized persons shall enter or modify data in the
computer. There shall be record of changes and deletions. Access shall be restricted by
‘passwords’ or other means and the result of entry of critical data shall be independently
checked. Batch records electronically stored shall be protected by a suitable back-up. During
the period of retention, all relevant data shall be readily available.
The manufacturer shall prepare documentation in a form of a Site Master File containing
specific information about the facilities, personnel and other details as prescribed in
Annexure A.
Changes to document shall be reviewed, approved and maintained. Change records shall
include a description of the change, identification of the affected documents, the signature of
the approving individual(s), the approval date, and when the change becomes effective.
The manufacturer shall ensure that changes to documents are reviewed and approved either
by the original approving function or another designated function which has access to
pertinent background information upon which to base its decisions.
The manufacturer shall define the period for which at least one copy of obsolete controlled
documents shall be retained. This period shall ensure that documents to which medical
devices/In-vitro Diagnostics have been manufactured and tested are retained for at least one
year after the date of expiry of the medical device/In-vitro Diagnostic as defined by the
manufacturer.
4.2.4 Control of records
Records shall be established and maintained to provide evidence of conformity to
requirements and of the effective operation of the quality management system. Records shall
remain legible, readily identifiable and retrievable. A documented procedure shall be
established to define the controls needed for the identification, storage, protection, retrieval,
retention time and disposition of records.
The manufacturer shall retain the records for a period of time at least one year after the date
of expiry of the medical device/In-vitro Diagnostics as defined by the manufacturer, but not
less than two years from the date of product release by the manufacturer.
5 Management responsibility
5.1 Management commitment
Top management of the manufacturer shall provide evidence of its commitment to the
development and implementation of the quality management system and maintaining its
effectiveness by
a) communicating to the employees the importance of meeting customer as well as statutory
and regulatory requirements,
b) establishing the quality policy,
c) ensuring that quality objectives are established,
d) conducting management reviews, and
e) ensuring the availability of resources.
5.2 Customer focus
Top management of the manufacturer shall ensure that customer requirements are
determined and are met.
5.3 Quality policy
Top management of the manufacturer shall ensure that the quality policy
a) is appropriate to the purpose of the manufacturing facility,
b) includes a commitment to comply with requirements and to maintain the effectiveness of
the quality management system,
c) provides a framework for establishing and reviewing quality objectives,
d) is communicated and understood within the manufacturers organization, and
e) is reviewed for continuing suitability.
5.4 Planning
5| P a g e
5.4.1 Quality objectives
Top management of the manufacturer shall ensure that quality objectives, including those
needed to meet requirements for product, are established at relevant functions and levels
within the manufacturing organization. The quality objectives shall be measurable and
consistent with the quality policy.
5.4.2 Quality management system planning
Top management of the manufacturer shall ensure that
a) the planning of the quality management system is carried out in order to meet the specified
requirements, as well as the quality objectives, and
b) the integrity of the quality management system is maintained when changes to the quality
management system are planned and implemented.
6| P a g e
5.6.3 Review output
The output from the management review shall include any decisions and actions related to
a) improvements needed to maintain the effectiveness of the quality management system
and its processes,
b) improvement of product related to customer requirements, and
c) resource needs.
6 Resource Management
6.1 Provision of resources
The manufacturing organization shall determine and provide the resources needed
a) to implement the quality management system and to maintain its effectiveness, and
b) to meet regulatory and customer requirements.
6.2 Human resources
6.2.1 General
Personnel performing work affecting product quality shall be competent on the basis of
appropriate education, training, skills and experience. Number of personnel employed shall
be adequate and in direct proportion to the workload. Prior to employment, all personnel,
shall undergo medical examination including eye examination, and shall be free from
communicable or contagious diseases. Thereafter, they should be medically examined
periodically, at least once a year. Records shall be maintained thereof. All personnel shall
bear clean body covering appropriate to their duties. Smoking, Eating, Drinking, Chewing or
keeping food and drink shall not be permitted in production, laboratory and storage areas.
6.2.2 Competence, awareness and training
The manufacturer shall
a) determine the necessary competence for personnel performing work affecting product
quality,
b) provide training or take other actions to satisfy these needs,
c) evaluate the effectiveness of the actions taken,
d) ensure that its personnel are aware of the relevance and importance of their activities and
how they contribute to the achievement of the quality objectives, and
e) maintain appropriate records of education, training, skills and experience.
f) establish documented procedures for identifying training needs and ensure that all
personnel are trained to adequately perform their assigned responsibilities.
6.3 Infrastructure
The organization shall determine, provide and maintain the infrastructure needed to achieve
conformity to product requirements. Infrastructure includes, as applicable
a) buildings, workspace and associated utilities.
b) process equipment (both hardware and software), and
c) supporting services (such as transport or communication).
The manufacturer shall establish documented requirements for maintenance activities,
including their frequency, when such activities or lack thereof can affect product quality.
Records of such maintenance shall be maintained.
6.4 Work environment
The organization shall determine and manage the work environment needed to achieve
conformity to product requirements. The following requirements shall apply.
a) The manufacturer shall establish documented requirements for health, cleanliness and
clothing of personnel if contact between such personnel and the product or work environment
could adversely affect the quality of the product.
b) If work environment conditions can have an adverse effect on product quality, the
manufacturer shall establish documented requirements as per annexure-C of this schedule
for the work environment conditions and documented procedures or work instructions to
7|Page
monitor and control these work environment conditions.
c) The manufacturer shall ensure that all personnel who are required to work temporarily
under special environmental conditions within the work environment are appropriately trained
and supervised by a trained person.
d) If appropriate, special arrangements shall be established and documented for the control
of contaminated or potentially contaminated product in order to prevent contamination of
other product, the work environment or personnel.
7 Product realization
7.1 Planning of product realization
The manufacturer shall plan and develop the processes needed for product realization.
Planning of product realization shall be consistent with the requirements of the other
processes of the quality management system. In planning product realization, the
manufacturer shall determine the following, as appropriate:
a) quality objectives and requirements for the product;
b) the need to establish processes, documents, and provide resources specific to the product;
c) required verification, validation, monitoring, inspection and test activities specific to the
product and the criteria for product acceptance;
d) records needed to provide evidence that the realization processes and resulting product
meet requirements.
The output of this planning shall be in a form suitable for the manufacturer’s method of
operations.
The manufacturer organization shall establish documented requirements for risk
management (as per the IS/ISO 14971) throughout product realization. Records arising from
risk management shall be maintained.
7.2 Customer-related processes
7.2.1 Determination of requirements related to the product
The manufacturer shall determine
a) requirements specified by the customer, including the requirements for delivery and post-
delivery activities,
b) requirements not stated by the customer but necessary for specified or intended use,
where known,
c) statutory requirements related to the product, and
d) any additional requirements determined by the manufacturer.
8|Page
7.2.3 Customer communication
The manufacturer shall determine and implement effective arrangements for communicating
with customers in relation to
a) product information,
b) enquiries, contracts or order handling, including amendments,
c) customer feedback, including customer complaints, and
d) advisory notices.
7.3 Design and development
7.3.1 Design and development planning
The manufacturer shall establish documented procedures for design and development. The
manufacturer shall plan and control the design and development of product. During the
design and development planning, the manufacturer shall determine
a) the design and development stages,
b) the review, verification, validation and design transfer activities (see Note) that are
appropriate at each design and development stage, and
c) the responsibilities and authorities for design and development.
The manufacturer shall manage the interfaces between different groups involved in design
and development to ensure effective communication and clear assignment of responsibility.
Planning output shall be documented, and updated as appropriate, as the design and
development progresses.
NOTE Design transfer activities during the design and development process ensure that design and
development outputs are verified as suitable for manufacturing before becoming final production
specifications.
7.3.2 Design and development inputs
Inputs relating to product requirements shall be determined and records maintained. The
design requirements relating to a device are appropriate and address the intended use of the
device, including the needs of the user and patients.
These inputs shall include
a) functional, performance and safety requirements, according to the intended use,
b) applicable statutory and regulatory requirements,
c) where applicable, information derived from previous similar designs,
d) other requirements essential for design and development, and
e) output(s) of risk management.
These inputs shall be reviewed for adequacy and approved by designated individual(s)
Requirements shall be complete, unambiguous and not in conflict with each other.
7.3.3 Design and development outputs
The outputs of design and development shall be provided in a form that enables verification
against the design and development input and shall be documented, reviewed, and approved
prior to release.
Design and development outputs shall
a) meet the input requirements for design and development,
b) provide appropriate information for purchasing, production and for service provision,
c) contain or reference product acceptance criteria, and
d) specify the characteristics of the product that are essential for its safe and proper use.
Records of the design and development outputs shall be maintained.
Records of design and development outputs can include specifications, manufacturing
procedures, engineering drawings, and engineering or research logbooks.
7.3.4 Design and development review
At suitable stages, systematic reviews of design and development shall be performed in
accordance with planned arrangements
9|Page
a) to evaluate the ability of the results of design and development to meet requirements, and
b) to identify any problems and propose necessary actions.
Participants in such reviews shall include representatives of functions concerned with the
design and development stage(s) being reviewed, as well as other specialist personnel
Records of the results of the reviews and any necessary actions shall be maintained (see
4.2.4).
7.3.5 Design and development verification
Verification shall be performed in accordance with planned arrangements to ensure that the
design and development outputs have met the design and development input requirements.
Records of the results of the verification and any necessary actions shall be maintained.
7.3.6 Design and development validation
Design and development validation shall be performed in accordance with planned
arrangements to ensure that the resulting product is capable of meeting the requirements for
the specified application or intended use. Design validation shall be performed under defined
operating conditions on initial production units, lots, or batches or their equivalence. Design
validation shall include software validation and risk analysis, where appropriate Validation
shall be completed prior to the delivery or implementation of the product.
Records of the results of validation and any necessary actions shall be maintained.
As part of design and development validation, the manufacturer shall perform clinical
evaluations and/or evaluation of performance of the medical device/In-vitro Diagnostics.
NOTE 1 If a medical device /In-vitro Diagnostics can only be validated following assembly
and installation at point of use, delivery is not considered to be complete until the product has
been formally transferred to the customer.
NOTE 2 Provision of the medical device for purposes of clinical evaluations and/or evaluation
of performance is not considered to be delivery.
7.3.7 Control of design and development changes
Design and development changes shall be identified and records maintained. The changes
shall be reviewed, verified and validated, as appropriate, and approved before
implementation. The review of design and development changes shall include evaluation of
the effect of the changes on constituent parts and product already delivered. Records of the
results of the review of changes and any necessary actions shall be maintained.
Note: -Each manufacturer shall establish and maintain a Design History File for each type of
device. The Design History File shall contain or reference the records necessary to
demonstrate that the design was developed in accordance with the approved design plan
and the requirements of design and development.
7.4 Purchasing
7.4.1 Purchasing process
The manufacturer organization shall establish documented procedures to ensure that
purchased product conforms to specified purchase requirements. The type and extent of
control applied to the supplier and the purchased product shall be dependent upon the effect
of the purchased product on subsequent product realization or the final product.
The manufacturer shall evaluate and select suppliers based on their ability to supply product
in accordance with the manufacturer’s requirements. Criteria for selection, evaluation and re-
evaluation shall be established.
Records of the results of evaluations and any necessary actions arising from the evaluation
shall be maintained.
10 | P a g e
7.4.2 Purchasing information
Purchasing information shall describe the product to be purchased, including where
appropriate
a) requirements for approval of product, procedures, processes and equipment,
b) requirements for qualification of personnel, and
c) quality management system requirements.
The manufacturer shall ensure the adequacy of specified purchase requirements prior to their
communication to the supplier.
To the extent required for traceability, the manufacturer shall maintain relevant purchasing
information, i.e. documents and records.
7.4.3 Verification of purchased product
The manufacturer shall establish and implement the inspection or other activities necessary
for ensuring that purchased product meets specified purchase requirements. Where the
manufacturer intends to perform verification at the supplier’s premises, the manufacturer
shall state the intended verification arrangements and method of product release in the
purchasing information. Records of the verification shall be maintained.
If the product is cleaned in accordance with a) or b) above, the requirements content in 6.4
(a) & (b) do not apply prior to the cleaning process
11 | P a g e
If the agreed customer requirements allow installation to be performed other than by
manufacturer or its authorized agent, the manufacturer shall provide documented
requirements for installation and verification. Records of installation and verification
performed by the manufacturer or its authorized agent shall be maintained.
7.5.1.3 Particular requirements for sterile medical devices
The manufacturer shall maintain records of the process parameters for the sterilization
process which was used for each sterilization batch. Sterilization records shall be traceable
to each production batch of medical devices.
7.5.2 Validation of processes for production and service provision
7.5.2.1 General requirements
The manufacturer shall validate any processes for production and service provision where
the resulting output cannot be verified by subsequent monitoring or measurement. This
includes any processes where deficiencies become apparent only after the product is in use.
Validation shall demonstrate the ability of these processes to achieve planned results.
The manufacturer shall establish arrangements for these processes including, as applicable
a) defined criteria for review and approval of the processes,
b) approval of equipment and qualification of personnel,
c) use of specific methods and procedures,
d) requirements for records, and
e) revalidation.
The manufacturer shall establish documented procedures for the validation of the
application of computer software (and changes to such software and/or its application)
for production and service provision that affect the ability of the product to confirm to
specified requirements. Such software applications shall be validated prior to initial use.
Records of validation shall be maintained.
7.5.2.2 Particular requirements for sterile medical devices
The manufacturer shall establish documented procedures for the validation of sterilization
processes. Sterilization processes shall be validated prior to initial use. The records of
validation of each sterilization process shall be maintained.
12 | P a g e
7.5.3.2.2 Particular requirements for active implantable medical devices and
implantable medical devices
In defining the records required for traceability, the manufacturer shall include records of all
components, materials and work environment conditions, if these could cause the medical
device not to satisfy its specified requirements.
The manufacturer shall require that its agents or distributors maintain records of the
distribution of medical devices/In-vitro Diagnostics to allow traceability and that such records
are available for inspection. Records of the name and address of the shipping package
consignee shall be maintained.
13| P a g e
take appropriate action on the equipment and any product affected. Records of the results of
calibration and verification shall be maintained.
When used in the monitoring and measurement of specified requirements, the ability of
computer software to satisfy the intended application shall be confirmed. This shall be
undertaken prior to initial use and reconfirmed as necessary.
8 Measurement, analysis and improvement
8.1 General
The manufacturer shall plan and implement the monitoring, measurement, analysis and
improvement processes needed
a) to demonstrate conformity of the product,
b) to ensure conformity of the quality management system, and
c) to maintain the effectiveness of the quality management system.
This shall include determination of applicable methods, including statistical techniques, and
the extent of their use.
Note: If relevant Indian standards are not available, International standards are applicable.
In case no Indian/International standards are available, validated testing process of the
manufacturer is applicable
The responsibilities and requirements for planning and conducting audits, and for reporting
results and maintaining records shall be defined in a documented procedure.
The management responsible for the area being audited shall ensure that actions are taken
without undue delay to eliminate detected nonconformities and their causes. Follow-up
activities shall include the verification of the actions taken and the reporting of verification
results.
8.2.3 Monitoring and measurement of processes
The manufacturer shall apply suitable methods for monitoring and, where applicable,
measurement of the quality management system processes. These methods shall
demonstrate the ability of the processes to achieve planned results. When planned results
14 | P a g e
are not achieved, correction and corrective action shall be taken, as appropriate, to ensure
conformity of the product.
8.2.4 Monitoring and measurement of product
8.2.4.1 General requirements
The manufacturer shall monitor and measure the characteristics of the product to verify that
product requirements have been met. This shall be carried out at appropriate stages of the
product realization process in accordance with the planned arrangements and documented
procedures.
Evidence of conformity with the acceptance criteria shall be maintained. Records shall
indicate the person(s) authorizing release of product. Product release shall not proceed until
the planned arrangements have been satisfactorily completed.
8.2.4.2 Particular requirement for active implantable medical devices and implantable
medical Devices wherever applicable
The manufacturer shall record the identity of personnel performing any inspection or testing.
8.3 Control of nonconforming product
The manufacturer shall ensure that product which does not conform to product requirements
is identified and controlled to prevent its unintended use or delivery. The controls and related
responsibilities and authorities for dealing with nonconforming product shall be defined in a
documented procedure.
The manufacturer shall deal with nonconforming product by one or more of the following
ways:
a) by taking action to eliminate the detected nonconformity;
b) by authorizing its use, release or acceptance under concession;
c) by taking action to preclude its original intended use or application.
The manufacturer shall ensure that nonconforming product is accepted by concession only
if regulatory requirements are met. Records of the identity of the person(s) authorizing the
concession shall be maintained. Records of the nature of nonconformities and any
subsequent actions taken, including concessions obtained, shall be maintained.
When nonconforming product is corrected it shall be subject to re-verification to demonstrate
conformity to the requirements. When nonconforming product is detected after delivery or
use has started, the manufacturer shall take action appropriate to the effects, or potential
effects, of the nonconformity.
If product needs to be reworked (one or more times), the manufacturer shall document the
rework process in a work instruction that has undergone the same authorization and approval
procedure as the original work instruction. Prior to authorization and approval of the work
instruction, a determination of any adverse effect of the rework upon product shall be made
and documented.
8.4 Analysis of data
The manufacturer shall establish documented procedures to determine collect and analyze
appropriate data to demonstrate the suitability and effectiveness of the quality management
system and to evaluate if improvement of the effectiveness of the quality management
system can be made.
This shall include data generated as a result of monitoring and measurement and from other
relevant sources. The analysis of data shall provide information relating to
a) feedback,
b) conformity to product requirements,
c) characteristics and trends of processes and products including opportunities for preventive
action, and
d) suppliers.
15 | P a g e
Records of the results of the analysis of data shall be maintained.
8.5 Improvement
8.5.1 General
The manufacturer shall identify and implement any changes necessary to ensure and
maintain the continued suitability and effectiveness of the quality management system
through the use of the quality policy, quality objectives, audit results, analysis of data,
corrective and preventive actions and management review.
The manufacturer shall establish documented procedures for the issue and implementation
of advisory notices. These procedures shall be capable of being implemented at any time.
Records of all customer complaint investigations shall be maintained. If investigation
determines that the activities outside the manufacturer’s organization contributed to the
customer complaint, relevant information shall be exchanged between the organizations
involved.
If any customer complaint is not followed by corrective and/or preventive action, the reason
shall be authorized and recorded. Manufacturer require to notify adverse events and the
manufacturer shall establish documented procedures to such notification to regulatory
authorities.
8.5.2 Corrective action
The manufacturer shall take action to eliminate the cause of nonconformities in order to
prevent recurrence. Corrective actions shall be appropriate to the effects of the
nonconformities encountered. A documented procedure shall be established to define
requirements for
a) reviewing nonconformities (including customer complaints),
b) determining the causes of nonconformities,
c) evaluating the need for action to ensure that nonconformities do not recur,
d) determining and implementing action needed, including, if appropriate, updating
documentation,
e) recording of the results of any investigation and of action taken, and
f) reviewing the corrective action taken and its effectiveness.
16 | P a g e
Annexure A
The manufacturer shall prepare a succinct document in the form of Site Master File containing
specific information about the production and/or control of device manufacturing carried out
at the premises. It shall contain the following information:
A GENERAL INFORMATION
IBrief information on the site (including name and address), relation to other sites
IIManufacturing activities
IIIAny other operations carried out on the site
IV Name and exact address of the site, including telephone, fax numbers, web site URL
and e-mail address
V Type of medical devices handled on the site and information about specifically toxic or
hazardous substances handled, mentioning the way they are handled and precautions
taken
VI Short description of the site (size, location and immediate environment and other
activities on the site)
VII Number of employees engaged in Production, Quality Control, warehousing, and
distribution
VIII Use of outside scientific, analytical or other technical assistance in relation to the
design, manufacture and testing
IX Short description of the quality management system of the company
X Devices details registered with foreign countries
B PERSONNEL
I Organisation chart showing the arrangements for key personnel
II Qualifications, experience and responsibilities of key personnel
III Outline of arrangements for basic and in-service training and how records are
maintained
IV Health requirements for personnel engaged in production
V Personnel hygiene requirements, including clothing
C PREMISES AND FACILITIES
I Layout of premises with indication of scale
II Nature of construction, finishes/fixtures and fittings
III Brief description of ventilation systems. More details should be given for critical areas
with potential risks of airborne contamination (including schematic drawings of the
systems). Classification of the rooms used for the manufacture of sterile products
should be mentioned
IV Special areas for the handling of highly toxic, hazardous and sensitizing materials
V Brief description of water systems (schematic drawings of the systems are desirable)
including sanitation
VI Maintenance (description of planned preventive maintenance programmes for premises
and recording system)
D EQUIPMENT
I Brief description of major production and quality control laboratories equipment (a list of
the equipment is required)
II Maintenance (description of planned preventive maintenance programmes and
recording system).
III Qualification and calibration, including the recording system. Arrangements for
computerized systems validation.
E SANITATION
I Availability of written specifications and procedures for cleaning the manufacturing
areas and equipments
17 | P a g e
F PRODUCTION
I Brief description of production operations using, wherever possible, flow sheets and
charts specifying important parameters
II Arrangements for the handling of starting materials, packaging materials, bulk and
finished products, including sampling, quarantine, release and storage.
III Arrangements for reprocessing or rework
IV Arrangements for the handling of rejected materials and products
V Brief description of general policy for process validation
G QUALITY CONTROLS
I Description of the Quality Control system and of the activities of the Quality Control
Department. Procedures for the release of finished products
H STORAGE
I Policy on the storage of medical device
E DOCUMENTATION
I Arrangements for the preparation, revision and distribution of necessary documentation,
including storage of master documents
F MEDICAL DEVICE COMPLAINTS AND FIELD SAFETY CORRECTIVE ACTION
I Arrangements for the handling of complaints
II Arrangements for the handling of field safety corrective action
G SELF INSPECTION
I Short Description of the internal audit system
H CONTRACT ACTIVITIES
I Description of the way in which the compliance of the contract acceptor is assessed
18 | P a g e
Annexure B
The manufacturer shall prepare a succinct document in the form of Device Master File
containing specific information about the device(s) manufacturing in the premises.
3.0 LABELLING
4.0 DESIGN AND MANUFACTURING INFORMATION
4.1 Device Design
4.2 Manufacturing Processes
5.0 ESSENTIAL PRINCIPLES (EP) CHECKLIST
6.0 RISK ANALYSIS AND CONTROL SUMMARY
7.0 PRODUCT VERIFICATION AND VALIDATION
7.2 Biocompatibility
7.3 Medicinal Substances
7.4 Biological Safety
7.5 Sterilisation
7.6 Software Verification and Validation
7.7 Animal Studies
7.8 Shelf Life/Stability Data
7.9 Clinical Evidence
7.10 Post Marketing Surveillance Data (Vigilance Reporting)
19| P a g e
Annexure C
“Environmental requirement for “Notified Devices”
Final Inspection 7
Component Preparation 8
Cutting, lathing 9
Moulding 9
20 | P a g e
Annexure –D
The Members of Sub Committee constituted by DTAB in its 68th meeting to examine the issue
relating to requirements of factory premises for manufacture of medical devices in totality
including the amendment of Schedule M-III under the Drugs and Cosmetic Rules
New
Dr. B Suresh
President, Pharmacy Council of India,
New Delhi
21| P a g e