UM GDS-3000-GDP Probe E
UM GDS-3000-GDP Probe E
UM GDS-3000-GDP Probe E
GDP-025/050/100 Series
USER MANUAL
GW INSTEK PART NO. 82DP-10000MA1
Table of Contents
SAFETY INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Safety Symbols ........................................................................ 4
Safety Guidelines..................................................................... 5
Power cord for the United Kingdom ...................................... 8
GETTING STARTED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Main Features ................................ ........................... 9
Instrument Overview ................................ ................. 11
GDP-025 ................................................................................ 11
GDP-050/ GDP-100............................................................... 11
MEASUREMENT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Operation ................................ ................................13
APENDIX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
GDP Series Specifications...........................................16
EC Declaration of Conformity ................................ ..... 20
3
GDP Series User Manual
SAFETY INSTRUCTIONS
This chapter contains important safety instructions
that should be followed when operating and
storing a differential probe. Read the following
before any operation to ensure your safety and to
keep the instrument in the best condition.
Safety Symbols
These safety symbols may appear in this manual or on the
instrument.
4
SAFETY INSTRUCTIONS
Safety Guidelines
5
GDP Series User Manual
6
SAFETY INSTRUCTIONS
7
GDP Series User Manual
8
GETTING STARTED
GETTING STARTED
The Getting Started chapter introduces the probe’s
features, appearance and functions.
Main Features
Model name Bandwidth Max differential voltage
GDP-025 25MHz 450Vrms
GDP-050 50MHz 6500Vrms
GDP-100 100MHz 6500Vrms
Features The GDP Series differential probes provide a
safe means to measure differential voltages for
all models of oscilloscopes.
The differential probes convert high voltages
(7000 peak for GDP-050, GDP-100) into low
voltages (≤7V with ref to ground) for display on
oscilloscopes.
The GDP Series is designed to operate with the
1MΩ impedance oscilloscopes. When combined
with a 50Ω load, the attenuation will be 2 times.
Accessories AC Adapter (110V or 220V depending on
GDP-025 region)
BNC to BNC cable: 50Ω resistance, RG58C UL,
100cm
Banana plug to banana plug silicon wire, UL
6kV, 18AWG, Length 60cm. (red x1, black x1)
IC clip, UL 1000V CAT III (red x1, black x1)
9
GDP Series User Manual
black x1)
Accessories Adapter (110V or 220V depending on region)
10
GETTING STARTED
Instrument Overview
GDP-025
GDP-050/ GDP-100
Attenuation Power supply
Intput terminals
function switch input
Over range
Power indicator Output
indicator
11
GDP Series User Manual
12
MEASUREMENT
MEASUREMENT
Risk of electric shock. The GDP Series differential
voltage probes are designed for use with high
High voltage voltages. When operating with high voltages
ensure proper safety precautions are taken at all
times. Please see the safety section for more details.
Operation
2 1 3
+
-
13
GDP Series User Manual
Deviation Table
Model GDP-050 GDP-050 All GDP-050 GDP-025 GDP-025
GDP-100 GDP-100 GDP-100
0.1 100 50 20 10 5 2
50m 50 25 10 5 2.5 1
20m 20 10 4 2 1 0.4
10m 10 5 2 1 .5 0.2
14
MEASUREMENT
15
GDP Series User Manual
APENDIX
GDP Series Specifications
GDP-025 Bandwidth DC-25MHz (-3dB)for x50, x200
DC-50MHz (for attenuation x20)
Attenuation x20, x50, x200
Accuracy ± 2%
Voltage Input ≤ 140Vp-p for x20 (≅ 45Vrms)
Range (DC+AC ≤ 350Vp-p for x50 (≅ 110Vrms)
peak to peak) ≤ 1400Vp-p for x200 (≅ 450Vrms)
Permitted Max Max differential voltage: 1400 V (DC+AC
Input Voltage peak to peak) or 450 Vrms
Max voltage between each input terminal
and ground: 600 Vrms
Input Impedance Differential: 4MΩ/1.2pF
Between terminals and ground:
2MΩ/2.3pF
Output ≤ ±7.0V
Output 50Ω
Impedance
Rise Time 14ns for x50, and x200; 23.4ns for x20
Rejection Rate on 60Hz: >80dB; 100Hz: >60dB; 1MHz:
Common Mode >50dB
Power Supply External 9V DC power supply
Consumption Max. 35mA (0.4 Watt)
Environmental Indoor Use only.
Operating Temperature: Reference: +20˚C to
Conditions
+30˚C; Use: 0˚C to +50˚C; Storage: -30˚C
to +70˚C
Relative Humidity: Reference: ≤70%;
Use: 10% to 85%: 10% to 90%
Storage: 10% to 90%
Dimensions and 195mm x 55mm x 30mm: 250g
Weight
16
APENDIX
17
GDP Series User Manual
18
APENDIX
19
GDP Series User Manual
EC Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
No.7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan
GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.
No. 69, Lushan Road, Suzhou New District Jiangsu, China
declares that the below mentioned product
GDP-025, GDP-050, GDP-100
Are herewith confirmed to comply with the requirements set out in the
Council Directive on the Approximation of the Law of Member States
relating to Electromagnetic Compatibility (89/336/EEC, 92/31/EEC,
93/68/EEC) and Low Voltage Equipment Directive (2006/95/EC). For
the evaluation regarding the Electromagnetic Compatibility and Low
Voltage Equipment Directive, the following standards were applied:
◎ EMC
Electrical equipment for measurement, control and laboratory
EN 61326-1 : use –– EMC requirements (1997+A1: 1998+A2: 2003)
Conducted and Radiated Emissions Electrostatic Discharge
EN 61326: 1997+A1: 1998+A2:2 003 Class B IEC 61000-4-2: 1995+A1: 1998+A2: 2000
Current Harmonic Radiated Immunity
EN 61000-3-2: 2006 IEC 61000-4-3: 2002+A1: 2002
Voltage Fluctuation Electrical Fast Transients
EN 61000-3-3: 1995+A1: 2001+A2 : 2005 IEC 61000-4-4: 2004
Surge Immunity
------------------------- IEC 61000-4-5: 1995+A1: 2000
Conducted Susceptibility
-------------------------
IEC 61000-4-6: 2003+A1: 2004
Voltage Dips/ Interrupts
-------------------------
IEC 61000-4-11: 2004
◎ Safety
Low Voltage Equipment Directive 2006/95/EC
Safety Requirements
IEC/EN 61010-031: 2002+A1: 2008
20