CMP 200 Drawings
CMP 200 Drawings
CMP 200 Drawings
THE PART NUMBER OF THE ORIGINAL AEC BOARD IN THE GENERATOR FOR
WHICH THIS MANUAL WAS PREPARED IS LISTED BELOW:
See Drawing List for the AEC Board Part Number for this particular unit.
MF-0863E
REFER TO PAGE 4 FOR LOGIC LEVELS, NOTES,
ETC, REFERENCED BY THIS SYMBOL:
+12V +35 V
+35 V
F1
E1 +
J15-1 J1-1 F2
F2 INVERTER TP11
AC MAINS BOARD -12V SWITCHING
INPUT REGULATOR CIRCUIT J15-3 J1-3
DS1 E2 - (U15 AND ASSOCIATED
COMPONENTS)
TP2
-12 V
J15-5 J1-5
+5V
J5-5 J5-7 J10-1 J10-3 J5-3 J5-1 J15-2 J1-2 F1
+12 V +12 V
ON / OFF TP13
+24V
D7 COMMAND 2 +12 V
FROM MD-0832 -35 V -35 V
1
MAIN CONTACTOR DRIVE J15-4 J1-4
TP12
FROM MD-0832 D5 J6-1 J11-1 +5V SWITCHING FILAMENT SUPPLY BOARD
REGULATOR CIRCUIT
F7 J6-2 J11-2 (U10 AND ASSOCIATED
E16 E15
** D4 COMPONENTS) +12 V
R16
120/240 VAC
E14 TO LOW SPEED
F6 STARTER J7-13
MD-0837 +12V
FOR UNITS WITH
TOUCHSCREEN +24 V
TRANSFORMER,
REFER TO K2 J7-19 REFER TO MD-0838 (AEC) FOR
R11
MD-0868 F5 C11 PINOUTS OF THE +/-12 VDC
TP14 AND +24 VDC SUPPLIES
R17 J7-17
F4
J7-11
R14
+12V
J6-12 J11-12 AEC BOARD
+24V -12 V
2 5
SOFT-START
CONTACTOR DRIVE SOFT-START
D8 FROM MD-0832 U5 SENSE TO
MD-0832 +12V SWITCHING FLUORESCENT LAMP
REGULATOR CIRCUIT
(U12 AND ASSOCIATED
BUCKY DRIVE 1 4 J6-6 J11-6
D6 COMPONENTS)
F8 F9 TO MD-0833
E10 E7 R13
LCD DISPLAY ASSEMBLY
+35 V
*
K6
J9-1 E6
E8 J5-1 J5-5
J9-3 E3 +35 V J1-4 J12-6 +24 V +5V TP9 3
J1-5 J12-5 K1
J9-7 E4 E5 F10 F1
* C2 300 VAC
E11 E9 BACKLIGHT
+12V POWER SUPPLY C36
+24 V
E12 (Q1, Q2, T1, ETC)
D1
D39 +5V
220V J8-5 TP7
0V D10 2
C4 TP2
110V J8-3 ON / OFF + 5 / -20VDC
J1-8 J12-2 J3-8 J8-8 F1
COMMAND 1 POWER SUPPLY
J1-9 J12-1 FROM MD-0832 & REGULATOR
0V J8-1
F11 (U12, L1, D3-D7, ETC)
-35 V
120V TP9
24 VAC TO TP3
+24 V +15V TP8
26V J3-1 F1 MD-0833 F12 24 VDC TO +24 V -20V
MD-0833 TP10 J3-3 J8-3
208V -35 V +24V
J1-1 J12-9
0V
240V J1-2 J12-8
18V J3-4 F2 CONSOLE BOARD
R27
C6
400V +15V SWITCHING
0V J3-3 TP9 REGULATOR CIRCUIT
D3 DS1 (U9 AND ASSOCIATED
480V J1-6 J12-4
18V J3-5 F3 D2 COMPONENTS)
J1-7 J12-3
D4
J3-2 DRAWN DATE
G. SANWALD 11 FEB 2003 DC BUS & POWER
CHECKED DISTRIBUTION
Steve Blake 18 FEB 2003
DES.\MFG.\AUTH.
THIS SHEET DEPICTS 1 PH. UNITS. H.V. AUXILIARY BOARD GENERATOR CONTROL BOARD ED WOOD FEB 13/03 MD-0831 REV M
REFER TO PAGE 2 FOR 3 PH.
400/480 V UNITS, AND PAGE 3 * E3 TO E12 CONFIGURE THE BUCKY OUTPUTS FOR +24 VDC, 120 VAC, OR 240 VAC. REFER TO CHAPTER 8 FOR DETAILS. SHEET 1 OF 4
FOR 3 PH. 208 V UNITS ** E14, E15, E16 CONFIGURE THE LOW SPEED STARTER BOOST VOLTAGE FOR 120 OR 240 VAC. REFER TO CHAPTER 2 FOR DETAILS.
F3
~ +
E1 +
+12V
+35 V
+35 V
AC MAINS
INPUT
F2
~ * INVERTER
BOARD
-12V SWITCHING
TP11
J15-1 J1-1 F2
~
F1 DS1 E2 - REGULATOR CIRCUIT J15-3 J1-3
(U15 AND ASSOCIATED
- COMPONENTS)
TP2
-12 V
* 50 KW UNITS USE J15-5 J1-5
EMC CAPACITOR TWO INVERTER BOARDS -12 V J15-6 J1-6
BOARD
+5V
J5-5 J5-7 J10-1 J10-3 J5-3 J5-1 J15-2 J1-2 F1
+12 V +12 V
ON / OFF TP13
+24V
D7 COMMAND 2 +12 V
TP11 FROM MD-0832 -35 V -35 V
1
MAIN CONTACTOR DRIVE J15-4 J1-4
TP12
FROM MD-0832 D5 J6-1 J11-1 +5V SWITCHING FILAMENT SUPPLY BOARD
REGULATOR CIRCUIT
J6-2 J11-2 (U10 AND ASSOCIATED
F6
120/240 VAC D4 COMPONENTS) +12 V
TO LOW SPEED
E16 E15 F7 STARTER J7-13
** MD-0837 +12V
FOR UNITS WITH
R27
E14
TOUCHSCREEN +24 V
TRANSFORMER, K2
F5 R11
REFER TO J7-19 REFER TO MD-0838 (AEC) FOR
MD-0868
PINOUTS OF THE +/-12 VDC
F4 R17 AND +24 VDC SUPPLIES
TP14
J7-17
+12V
J7-11
R14
J6-12 J11-12 AEC BOARD
SOFT-START +24V -12 V
2 5
CONTACTOR DRIVE
D8 FROM MD-0832
SOFT-START
U5 SENSE TO
MD-0832 +12V SWITCHING FLUORESCENT LAMP
REGULATOR CIRCUIT
F8 F9 (U12 AND ASSOCIATED
BUCKY DRIVE 1 4 J6-6 J11-6
D6 COMPONENTS)
TO MD-0833
E10 E7 R13
J9-5 LCD DISPLAY ASSEMBLY
+35 V +35 V
*
K6
J9-1 E6 TP2
E8 J5-1 J5-5
J9-3 E3 J1-4 J12-6 +24 V +5V TP9 3
J1-5 J12-5 K1
J9-7 E4 E5 F10 F1
* C2 300 VAC
R28
E11 E9 BACKLIGHT
POWER SUPPLY C36
+24 V +12V
E12 (Q1, Q2, T1, ETC)
D1
D39 +5V
220V J8-5 -35 V TP7
0V D10 C4 2 TP2
R29
TP3
110V J8-3 ON / OFF + 5 / -20VDC
J1-8 J12-2 J3-8 J8-8 F1
24 VAC TO COMMAND 1 POWER SUPPLY
MD-0833 J1-9 J12-1 FROM MD-0832 & REGULATOR
0V J8-1
F11 (U12, L1, D3-D7, ETC)
120V 24 VDC TO TP9
MD-0833 TP3
+24 V +15V TP8
26V J3-1 F1 F12 +24 V -20V
TP1 J3-3 J8-3
-35 V TP10
208V J1-1 J12-9 +24V
0V
240V J1-2 J12-8
18V J3-4 F2 CONSOLE BOARD
R30
C6
400V TP12 +15V SWITCHING
0V J3-3 TP9 REGULATOR CIRCUIT
D3 DS1 (U9 AND ASSOCIATED
480V J1-6 J12-4
18V J3-5 F3 D2 COMPONENTS)
J1-7 J12-3
D4
J3-2 DRAWN DATE
G. SANWALD 11 FEB 2003 DC BUS & POWER
CHECKED DISTRIBUTION
Steve Blake 18 FEB 2003
THIS SHEET DEPICTS 3 PH. DES.\MFG.\AUTH.
400/480V UNITS. REFER TO H.V. AUXILIARY BOARD GENERATOR CONTROL BOARD ED WOOD FEB 13/03 MD-0831 REV M
PAGE 1 FOR 1 PH. UNITS, AND * E3 TO E12 CONFIGURE THE BUCKY OUTPUTS FOR +24 VDC, 120 VAC, OR 240 VAC. REFER TO CHAPTER 8 FOR DETAILS. SHEET 2 OF 4
PAGE 3 FOR 3 PH. 208 V UNITS ** E14, E15, E16 CONFIGURE THE LOW SPEED STARTER BOOST VOLTAGE FOR 120 OR 240 VAC. REFER TO CHAPTER 2 FOR DETAILS.
F3
~ +
E1 + +12V +35 V
~
+35 V
AC MAINS F2 INVERTER
BOARD J15-1 J1-1 F2
INPUT
~
F1 TP11
DS1 E2 - -12V SWITCHING
J15-3 J1-3
- REGULATOR CIRCUIT
(U15 AND ASSOCIATED
COMPONENTS)
TP2
-12 V
EMC CAPACITOR J15-5 J1-5
BOARD -12 V J15-6 J1-6
+5V
J5-5 J5-7 J10-1 J10-3 J5-3 J5-1 J15-2 J1-2 F1
+12 V +12 V
ON / OFF TP13
+24V
D7 COMMAND 2 +12 V
FROM MD-0832 -35 V -35 V
1
MAIN CONTACTOR DRIVE J15-4 J1-4
TP12
FROM MD-0832 D5 J6-1 J11-1 +5V SWITCHING FILAMENT SUPPLY BOARD
REGULATOR CIRCUIT
F7 J6-2 J11-2 (U10 AND ASSOCIATED
E16 E15
** D4 COMPONENTS) +12 V
R16
120/240 VAC
E14 TO LOW SPEED
F6 STARTER J7-13
MD-0837 +12V
FOR UNITS WITH
TOUCHSCREEN +24 V
TRANSFORMER,
REFER TO K2 J7-19 REFER TO MD-0838 (AEC) FOR
R11
MD-0868 F5 C11 PINOUTS OF THE +/-12 VDC
TP14 AND +24 VDC SUPPLIES
R17 J7-17
F4
J7-11
R14
+12V
J6-12 J11-12 AEC BOARD
+24V -12 V
2 5
SOFT-START
CONTACTOR DRIVE SOFT-START
D8 FROM MD-0832 U5 SENSE TO
MD-0832 +12V SWITCHING FLUORESCENT LAMP
REGULATOR CIRCUIT
(U12 AND ASSOCIATED
BUCKY DRIVE 1 4 J6-6 J11-6
D6 COMPONENTS)
F8 F9 TO MD-0833
E10 E7 R13
LCD DISPLAY ASSEMBLY
+35 V
*
K6
J9-1 E6
E8 J5-1 J5-5
J9-3 E3 +35 V J1-4 J12-6 +24 V +5V TP9 3
J1-5 J12-5 K1
J9-7 E4 E5 F10 F1
* C2 300 VAC
E11 E9 BACKLIGHT
POWER SUPPLY C36
+24 V +12V
E12 (Q1, Q2, T1, ETC)
D1
D39 +5V
220V J8-5 TP7
0V D10 C4 2 TP2
110V J8-3 ON / OFF + 5 / -20VDC
J1-8 J12-2 J3-8 J8-8 F1
24 VAC TO COMMAND 1 POWER SUPPLY
MD-0833 J1-9 J12-1 FROM MD-0832 & REGULATOR
0V J8-1
F11 (U12, L1, D3-D7, ETC)
24 VDC TO -35 V
120V TP9
MD-0833 TP3
+24 V +15V TP8
26V J3-1 F1 F12 +24 V -20V
TP10 J3-3 J8-3
208V
-35 V +24V
J1-1 J12-9
0V
240V J1-2 J12-8
18V J3-4 F2 CONSOLE BOARD
R27
C6
400V +15V SWITCHING
0V J3-3 TP9 REGULATOR CIRCUIT
D3 DS1 (U9 AND ASSOCIATED
480V J1-6 J12-4
18V J3-5 F3 D2 COMPONENTS)
J1-7 J12-3
D4
J3-2 DRAWN DATE
G. SANWALD 11 FEB 2003 DC BUS & POWER
CHECKED DISTRIBUTION
Steve Blake 18 FEB 2003
THIS SHEET DEPICTS 3 PH. DES.\MFG.\AUTH.
208 V UNITS. REFER TO PAGE H.V. AUXILIARY BOARD GENERATOR CONTROL BOARD ED WOOD FEB 13/03 MD-0831 REV M
1 FOR 1 PH. UNITS, AND PAGE * E3 TO E12 CONFIGURE THE BUCKY OUTPUTS FOR +24 VDC, 120 VAC, OR 240 VAC. REFER TO CHAPTER 8 FOR DETAILS. SHEET 3 OF 4
2 FOR 3 PH. 400/480 V UNITS ** E14, E15, E16 CONFIGURE THE LOW SPEED STARTER BOOST VOLTAGE FOR 120 OR 240 VAC. REFER TO CHAPTER 2 FOR DETAILS.
NOTE
REMARKS
REFERENCE
“LOW” (APPROXIMATELY 0 VDC) DISABLES THE +5 V, +/-12 V, AND +15 V REGULATORS (GENERATOR SWITCHED OFF). “HIGH” (APPROXIMATELY 24 VDC) ENABLES THESE REGULATORS
1 (GENERATOR SWITCHED ON VIA THE CONSOLE ON/OFF SWITCHES, OR VIA THE ON/OFF SWITCHES ON THE GENERATOR CONTROL BOARD) .
2 +24 VDC IS PRESENT AT THIS POINT WHEN THE GENERATOR IS SWITCHED ON, ENERGIZING K1 ON THE GENERATOR CONTROL BOARD.
3 THE VOLTAGE WAVEFORM AT THIS TEST POINT IS DEPICTED IN FIGURE 1 BELOW.
18 s
4V
FIGURE 1
-8 V
DRAWN DATE
G. SANWALD 11 FEB 2003 DC BUS & POWER
CHECKED DISTRIBUTION
Steve Blake 18 FEB 2003
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0831 REV M
SHEET 4 OF 4
REFER TO PAGE 2 FOR
LOGIC LEVELS, NOTES, +24 V
ETC, REFERENCED BY
THIS SYMBOL:
ON / OFF
Q1
COMMAND 1
J7-16 J8-7 J3-7 TO MD-0831
ON / OFF
R235
OFF J7-15 J8-6 J3-6 COMMAND 2
Q2
TO MD-0831
ON TP3 TP9 Q21
S1
R236
J7-17 J8-3 J3-3 OFF
Q3
S2
CONSOLE ON
KEYBOARD ASSEMBLY BOARD
+5 V
U54
R232
TP4
J6-12 J11-12
BUFFER
MAIN CONTACTOR 5
DRIVE TO
MD-0831 SOFT START
SENSE FROM U5
MD-0831
4
J6-6 J11-6
D7
+12 V
TP11
J6-1 J11-1
J6-4 J11-4
3
TP5 DRIVER CPLD
DRAWN DATE
G. SANWALD 11 FEB 2003
CHECKED SYSTEM “ON”
Steve Blake 18 FEB 2003
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0832 REV A
SHEET 1 OF 2
NOTE
REMARKS
REFERENCE
1 THE VOLTAGE WAVEFORM AT THIS TEST POINT IS DEPICTED IN FIGURE 1 BELOW.
“LOW” (APPROXIMATELY 0 VDC) ENERGIZES THE MAIN POWER CONTACTOR IN THE GENERATOR, (”HIGH”, + 12 VDC = NOT ENERGIZED). THIS CONTACTOR IS ENERGIZED AFTER THE MAIN BUS CAPACITORS
2 ARE CHARGED, APPROXIMATELY 10 SECONDS AFTER POWER-ON.
“LOW” (APPROXIMATELY 0 VDC) ENERGIZES THE SOFT START CONTACTOR K2 ON THE H.V. AUXILIARY BOARD, (”HIGH”, + 12 VDC = NOT ENERGIZED). THIS CONTACTOR IS ENERGIZED FOR A MAXIMUM OF
3
APPROXIMATELY 10 SECONDS AFTER POWER-ON IN ORDER TO CHARGE THE DC BUS CAPACITORS.
MAXIMUM 1 SEC.
+5 V
FIGURE 1
0V
GENERATOR
SWITCHED ON
DRAWN DATE
G. SANWALD 11 FEB 2003
CHECKED SYSTEM “ON”
Steve Blake 18 FEB 2003
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0832 REV A
SHEET 2 OF 2
BUCKY DRIVE +12V
FROM MD-0831
24 VDC, 1
RETURN 110 / 220 VAC K1
D2 TP13
U40 U2
J6-13 J11-13
J2-10 DRIVER CPLD
BUCKY RETURN
2 TP14
BUCKY 1 J2-9 +5V
GROUND J6-14 J11-14
1 5
J2-8 +24V
R199
BUCKY 1 OUT
U44
BUCKY 1 J2-7 +24V U2
R5
START K1 BUFFER
3
1 5 2 4
R3
TP15
J2-6 +5V
J6-15 J11-15
U1
BUCKY 1
READY J2-5
R200
2 4 +24V 4 U44
J2-4 TP16 BUFFER
INTERLOCK +5V
R7
J2-3 J6-16 J11-16
#1
1 5
R201
J2-2 +12V 5 U44
INTERLOCK #2 / U3
TOMO EXPOSURE TP19 BUFFER
J2-1
INPUT *
2 4 K3 D6
J4-4 U40 U2
BUCKY RETURN
J6-19 J11-19
TOMO / BUCKY 2 J4-3 DRIVER CPLD
GROUND
+5V
J4-2
TOMO / BUCKY 2 OUT
24 VAC
R203
TOMO / BUCKY 2 J4-1 FROM 6 TP18
K3 U44
START MD-0831
J6-18 J11-18
BUFFER
1 5
24 VDC
J11-6
24 VAC FROM
OUT J11-5 MD-0831 U6 +5V
+24V 7
R202
J11-2 2 4
24 VDC TP17
U44
R9
OUT J11-1
J6-17 J11-17
+24V BUFFER
1 5
J4-6 R10
+5V
BUCKY 2 U4
READY J4-5 8
R198
2 4 TP6
U44
J4-10
J6-10 J11-10
DOOR BUFFER
INTERLOCK J4-9 +24V 1 5
+12V U3
J7-1 R26 R25 U9
D45 CPU
THERMAL
SWITCH J7-2
2 4 R264
J6-9 J11-9
+12V Q23 DATA, ADDRESS, &
J11-3 CONTROL BUS
R263
ROOM K7 D11 J6-20 J11-20
LIGHT
J11-4 DRAWN DATE
H.V. AUXILIARY BOARD GENERATOR CONTROL BOARD G. SANWALD 11 FEB 2003 ROOM
CHECKED INTERFACE
Steve Blake 18 FEB/03
REFER TO CHAPTER 3 OF THE SERVICE MANUAL FOR ADDITIONAL DETAILS REGARDING INTERFACING OF BUCKYS, INTERLOCKS, ETC. DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0833 REV E
* THIS IS THE TOMO EXPOSURE INPUT FOR ANY IMAGE RECEPTOR THAT IS PROGRAMMED FOR TOMO OPERATION. THIS REFER TO PAGE 2 FOR LOGIC LEVELS, NOTES,
INPUT IS AN INTERLOCK FOR ALL REMAINING RECEPTORS THAT DO NOT HAVE TOMO ENABLED. THIS INPUT MUST NORMALLY
BE PROGRAMMED AS INACTIVE ON ALL NON-TOMO RECEPTORS IF THIS IS USED AS A TOMO INPUT ON THE TOMO RECEPTOR.
ETC, REFERENCED BY THIS SYMBOL: SHEET 1 OF 2
NOTE
REMARKS
REFERENCE
1 “LOW” (APPROXIMATELY 0 VDC) = BUCKY 1 START. “HIGH” (APPROXIMATELY +12 VDC) = BUCKY 1 NOT REQUESTED TO START.
2 “LOW” (APPROXIMATELY 0 VDC) = BUCKY 1 READY. “HIGH” (APPROXIMATELY +5 VDC) = BUCKY 1 NOT READY.
3 “LOW” (APPROXIMATELY 0 VDC) = 40” S.I.D. INTERLOCK CLOSED. “HIGH” (APPROXIMATELY +5 VDC) = 40” S.I.D. INTERLOCK OPEN.
4 “LOW” (APPROXIMATELY 0 VDC) = 72” S.I.D. INTERLOCK CLOSED. “HIGH” (APPROXIMATELY +5 VDC) = 72” S.I.D. INTERLOCK OPEN.
5 “LOW” (APPROXIMATELY 0 VDC) = BUCKY 2 START. “HIGH” (APPROXIMATELY +12 VDC) = BUCKY 2 NOT REQUESTED TO START.
6 “LOW” (APPROXIMATELY 0 VDC) = BUCKY 2 READY. “HIGH” (APPROXIMATELY +5 VDC) = BUCKY 2 NOT READY.
7 “LOW” (APPROXIMATELY 0 VDC) = DOOR INTERLOCK CLOSED. “HIGH” (APPROXIMATELY +5 VDC) = DOOR INTERLOCK OPEN.
8 “LOW” (APPROXIMATELY 0 VDC) = THERMAL SWITCH CLOSED. “HIGH” (APPROXIMATELY +5 VDC) = THERMAL SWITCH OPEN.
DRAWN DATE
G. SANWALD 11 FEB 2003 ROOM
CHECKED INTERFACE
Steve Blake 18 FEB/03
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0833 REV E
SHEET 2 OF 2
+5V +5V
+24V
R229
R228
R195
U3
1 5 CPU
U42
REFER TO PAGE 2 FOR 2 4
LOGIC LEVELS, NOTES,
ETC, REFERENCED BY X-RAY ENABLE
THIS SYMBOL: 1 1 5 TO MD-0835
TP13 U43
U40 U2
DS6
J8-5 J3-5 2 4 4 15
DRIVER CPLD
1 5
+5V
U10 +24V
R67
J3-1 2 4
R194
2 1 5
TP12 U41
DS5
J8-4 J3-4 2 4
+5V
DATA, ADDRESS, &
CONTROL BUS
U17
R66
X-RAY
CPU
1 5
U9
J3-3 2 4
PREP JW2
J3-5
JW1
HANDSWITCH REMOVE JW1 DATA, ADDRESS, &
ASSY TO DISABLE THE CONTROL BUS
CONSOLE PREP
BUTTON
REMOVE JW2
TO DISABLE THE
J7-2 CONSOLE X-RAY
BUTTON
X-RAY J7-3
PREP
J7-1
DRAWN DATE
G. SANWALD 11 FEB 2003 X-RAY EXPOSURE
CHECKED (RADIOGRAPHIC)
Steve Blake 13 FEB/03
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0834 REV B
SHEET 1 OF 2
NOTE
REMARKS
REFERENCE
1 “LOW” (APPROXIMATELY 0 VDC) = X-RAY REQUESTED. “HIGH” (APPROXIMATELY +24 VDC) = X-RAY NOT REQUESTED.
2 “LOW” (APPROXIMATELY 0 VDC) = PREP REQUESTED. “HIGH” (APPROXIMATELY +24 VDC) = PREP NOT REQUESTED.
DRAWN DATE
G. SANWALD 11 FEB 2003 X-RAY EXPOSURE
CHECKED (RADIOGRAPHIC)
Steve Blake 13 FEB/03
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0834 REV B
SHEET 2 OF 2
REFER TO PAGE 2 FOR
LOGIC LEVELS, NOTES, INVERTER “SHOOT THRU” INVERTER “SHOOT THRU”
ETC, REFERENCED BY SENSE, FROM PAGE 2 SENSE, FROM PAGE 2
+5V +5V +12V
THIS SYMBOL: 1V = 20 kV OF R126
1 H.V. FEEDBACK
TP26 J10-1 J10-3 J16-1 J16-3 TP8 R127 -
2
1 R120 J9-6
+
3
- R122
U2 U28A
R225
R124
+ T3 T4
R125
J9-5
TP27 U39
CPLD 2 - 6
7
R128
5 J9-7 KV FEEDBACK SIGNAL
- R193 R246 +
FROM PAGE 2
U28B
+ R130
R131
U57
R133 6
R135
7 -
5 R134 J9-8
INVERTER “SHOOT THRU” +
DETECTOR CIRCUITS U30B
+5V +5V
R121
R123
GATE DRIVE CIRCUIT
TP23
U2 FOR MOSFET INVERTER
3 (INCLUDES Q9, 10, 12, 13)
7 - J5-1
CPLD 2 R129
+ J5-2
U29
R132
J5-3
J5-4
J13-3
J13-4
+12V +12V
R161
R164
U3 Q16
R165
R166
CPU Q17
R169
1V = 15 kV OF TP5
R171
TP16
U46
R64
ERROR AMPLIFIERS
R76
R65
D/A 1 2- INCLUDES U36A, U36B
R177 R176 U17A U18A
CONV 1 2- R178 R179
3+ 1 2
3 - D35 3 2 7
+ - R183 2- VCO 1
U50A + 1
U35A D34 +
3+
R181
R187
6 4 5
4 4
TP23
U3
R78
CPU Q14
R81
TP15
5 +5V +5V
3 R96 D16-D19
7 - 7 9 J9-9 J3-9
2 CURRENT SENSE
+12V
R69
R68
+
TP19 TP17
R97
R99
U22
R98
R95
R94
T2 T1
6
U2
R71
R70
TP18 H.V. OIL TANK
-12V
CPLD R72
Q11
J6-1 J6-4
R73
DRAWN DATE
X-RAY G. SANWALD 11 FEB 2003 KV CONTROL
ENABLE CHECKED
DATA BUS H.T. PRIMARY CURRENT SENSE
FROM MD-0834 Steve Blake 18 FEB/03
& FEEDBACK
D0..D7 FROM PAGE 2
DES.\MFG.\AUTH.
GENERATOR CONTROL BOARD
ED WOOD FEB 13/03 MD-0835 REV A
SHEET 1 OF 3
TO J9-8 PART OF HT TANK
kV FEEDBACK TO J9-7
TO PAGE 2
TO J9-5
HV ANODE
T2 J2-1 TO J9-6
TO J16-1 INVERTER BOARD
PART OF INVERTER “SHOOT “SHOOT THRU”
THROUGH” DETECTOR CIRCUIT J2-3
TO J16-3 SENSE, TO PAGE 1
*
E1: 560 to 670 VDC (+)
MOSFET INVERTER J1
(INCLUDES Q1-Q16) ANODE
C S
L
J13-1 J1-1
J3-8
J13-2 J1-2 HV MULT
ASSY +
E3 J3-7 (ANODE)
J13-3 J1-3
J13-4 J1-4 J3-5
J3-6
E4
E9
INVERTER BOARD E2: 560 to 670 VDC (-)
FROM
PAGE 1
T2 J2-1
TO J10-1 INVERTER
PART OF INVERTER “SHOOT “SHOOT THRU” E10
THROUGH” DETECTOR CIRCUIT J2-3
TO J10-3 SENSE, TO PAGE 1
E1: 560 to 670 VDC (+) HV MULT
MOSFET INVERTER
ASSY-
(INCLUDES Q1-Q16)
(CATHODE)
J5-1 J1-1
*
J5-2 J1-2
E3
C L
J5-3 J1-3
S
J5-4 J1-4
HV CATHODE J2
TANK LID CATHODE
E4 BOARD
BOARD
DRAWN DATE
G. SANWALD 11 FEB 2003 KV CONTROL
CHECKED & FEEDBACK
Steve Blake 18 FEB/03
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0835 REV A
SHEET 2 OF 3
T2 J2-1
TO J10-1 INVERTER
PART OF INVERTER “SHOOT “SHOOT THRU”
THROUGH” DETECTOR CIRCUIT J2-3
TO J10-3 SENSE, TO PAGE 1
E1: 325 to 670 VDC (+)
MOSFET INVERTER
(INCLUDES Q1-Q16)
TO E9 AND E10
FROM J5-1 J1-1
OF H.V. TANK
PAGE 1 J5-2 J1-2 (PAGE 2)
E3
J5-3 J1-3
J5-4 J1-4
E17
GENERATOR
CONTROL BOARD INVERTER BOARD E2: 325 to 670 VDC (-)
NOTE
REMARKS
REFERENCE
A NARROW PULSE WILL BE PRESENT AT THIS TEST POINT IF AN INVERTER “SHOOT THROUGH” HAS BEEN DETECTED. THE VOLTAGE WAVEFORM AT THIS TEST POINT IS DEPICTED IN FIGURE 1 BELOW.
1 THIS PULSE MAY BE VERY DIFFICULT O DETECT, AS THE INVERTER DRIVE WILL BE SHUT DOWN WHEN A “SHOOT THROUGH” IS DETECTED, THUS REMOVING THE FAULT CONDITION.
2 AS PER # 1.
A NARROW PULSE WILL BE PRESENT AT THIS TEST POINT IF KV OVER VOLTAGE HAS BEEN DETECTED (130 KV FOR 125 KV UNITS, 163 KV FOR 150 KV UNITS). REFER TO FIGURE 1. THIS PULSE MAY BE VERY
3 DIFFICULT TO DETECT, AS THE HIGH VOLTAGE WILL BE SHUT DOWN WHEN THE OVER VOLTAGE CONDITION IS DETECTED, THUS REMOVING THE FAULT CONDITION.
4 LOW (APPROXIMATELY 0 VDC) = X-RAY REQUESTED BY THE CPU, HIGH (APPROXIMATELY +5 VDC) = NO X-RAY REQUEST BY THE CPU.
A NARROW PULSE WILL BE PRESENT AT THIS TEST POINT IF INVERTER OVER CURRENT HAS BEEN DETECTED. THE VOLTAGE WAVEFORM AT THIS TEST POINT IS DEPICTED IN FIGURE 1 BELOW.
5
THIS PULSE MAY BE VERY DIFFICULT O DETECT, AS THE INVERTER DRIVE WILL BE SHUT DOWN WHEN AN OVER CURRENT CONDITION S DETECTED, THUS REMOVING THE FAULT CONDITION.
6 LOW (APPROXIMATELY 0 VDC) = NO FAULT PRESENT, ALLOW AN X-RAY EXPOSURE. HIGH (APPROXIMATELY +5 VDC) = X-RAY EXPOSURE INHIBITED.
7 LOW (APPROXIMATELY 0 VDC) = X-RAY REQUESTED BY CONSOLE, HIGH (APPROXIMATELY +5 VDC) = X-RAY NOT REQUESTED.
8 LOW (APPROXIMATELY 0 VDC) = NAND GATES U17A, U17B DISABLED. HIGH (APPROXIMATELY +5 VDC) = NAND GATES ENABLED, THUS ALLOWING INVERTER GATE DRIVE.
9 LOW (0 VDC) = H.V. TANK CONNECTED, HIGH (APPROXIMATELY +5 VDC) = H.V. TANK NOT CONNECTED.
10 THE VOLTAGE AT TP2 AND TP3 SHOULD BE A 50% DUTY CYCLE SQUARE WAVE, RANGING IN FREQUENCY FROM APPROXIMATELY 80 kHz TO APPROXIMATELY 250 kHz, DEPENDING ON GENERATOR
OUTPUT POWER. SEE FIGURE 2.
+5 V
FIGURE 1 12 VDC
FIGURE 2
0V 0 VDC
DRAWN DATE
G. SANWALD 11 FEB 2003 KV CONTROL
CHECKED & FEEDBACK
Steve Blake 18 FEB/03
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0835 REV A
SHEET 3 OF 3
REFER TO PAGE 3 FOR
LOGIC LEVELS, NOTES, MAXIMUM FILAMENT CURRENT
ETC, REFERENCED BY LIMIT CIRCUIT
THIS SYMBOL: SET MAX
CURRENT ERROR AMPLIFIER, PWM REGULATOR,
JW1 AND FILAMENT CURRENT DRIVERS
3
+
2 6.5 A
-
R76 5.5 A
U1A 1 +35V
1 -12V
TP4
D6
R218
U48 R72 5
Q6 Q12
D/A 6 - R206 J14-6 J2-6 R65 TP3
CONV 7 6- R95 R77 U3
5 +
7 -
U52B 5+
J14-5 J2-5 R66 U1B +
U4B PWM
REGULATOR
R71
Q13
Q7
R85
C22
J2
L CATHODE
-35V
4
C L
TP21 TP4 J5-2 J4-2
S
+12V
J5-1 J4-1
R84
+12V 2 K1 K1
J5-4 J4-4
TP29
D38 J5-3 J4-3
J14-11 J2-11
U2
R227
CPLD Q19
S
FILAMENT CURRENT SENSE,
R226
R21
TANK LID
BOARD
DATA BUS
D0..D7
PART OF H.V. OIL TANK
DRAWN DATE
G. SANWALD 11 FEB 2003 FILAMENT DRIVE
CHECKED & MA CONTROL
Steve Blake 18 FEB/03
DES.\MFG.\AUTH.
GENERATOR CONTROL BOARD FILAMENT SUPPLY BOARD
ED WOOD FEB 13/03 MD-0836 REV C
SHEET 1 OF 3
U38A
3+ R223
1
2-
1V = 100 mA
R222
OF ANODE
CURRENT R214
R221
TP7
R163
U45
R224
2- A/D
R212
1 CONV
3+
U33A
R207
R162
R168
J1 HV ANODE
R160
ANODE BOARD
S C
R155
L
+5V +5V
R149
R148
J3-1 J9-1 R151 R152 R150 TP25
3-
6- R154 7
R153
7 2+
5+
U32
J3-2 J9-2 R157 R158 U33B
R156
R159
mA
TEST
E18 JACK
- +5V
R139
R138
U2
-12V TP24
+
3- CPLD
E17 7
2+
R136
U31
R137
J3-4 J9-4 R140 R141
2- R143
R142
1
3+
J3-3 J9-3 R145 R146 DATA BUS
R144
U30A
D0..D7
L C
R147
S
J2
CATHODE HV CATHODE TANK LID
BOARD BOARD
DRAWN DATE
G. SANWALD 11 FEB 2003 FILAMENT DRIVE
CHECKED & MA CONTROL
Steve Blake 18 FEB/03
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0836 REV C
SHEET 2 OF 3
NOTE
REMARKS
REFERENCE
1 1 VOLT AT THIS TEST POINT = 1 AMP OF FILAMENT DEMAND.
2 “LOW” (APPROXIMATELY 0 VDC) ENERGIZES K1 ON THE FILAMENT SUPPLY BOARD (SMALL FILAMENT), “HIGH” (APPROXIMATELY +12 VDC) DE-ENERGIZES K1 (LARGE FILAMENT).
3 0.6 VOLT AT THIS TEST POINT = 1 AMP OF ACTUAL FILAMENT CURRENT.
4 1 VOLT AT THIS TEST POINT = 1 AMP OF ACTUAL FILAMENT CURRENT.
5 PWM OUTPUT. THE WAVEFORM WILL BE AS PER FIGURE 1 FOR LOW AND HIGH FILAMENT CURRENT DEMAND.
A NARROW PULSE WILL BE PRESENT AT THESE TEST POINTS DURING SEVERE ANODE OR CATHODE OVER CURRENTS (I.E. TUBE OR TANK ARCS). REFER TO FIGURE 2. THESE PULSES MAY BE VERY
6, 7 DIFFICULT TO OBSERVE, AS THE HIGH VOLTAGE WILL SHUT DOWN WHEN A FAULT IS DETECTED, THUS REMOVING THE OVER CURRENT SITUATION.
0V
FIGURE 1
+12 V
HIGH FILAMENT
DEMAND
0V
+5 V TP25
0V
FIGURE 2
+5 V
TP24
0V
DRAWN DATE
G. SANWALD 11 FEB 2003 FILAMENT DRIVE
CHECKED & MA CONTROL
Steve Blake 18 FEB/03
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0836 REV C
SHEET 3 OF 3
F7 J7-4
COM
120/240 VAC
R19 C12 J7-6
TO LOW SPEED
STARTER
SHIFT
FROM MD-0831 F6 R20 J7-5
MAIN
J7-3
GROUND
R23
R21
R24
R22
2 1 2 1
4 5 4 5 +5 V +5 V
U44
2
U8
R197
R196
U7
1 TP7
J6-6 J11-6
3 U2
K5
TP9 U40
J6-5 J11-5
K4
D9 TP10 J6-3 J11-3 CPLD
4
J6-4 J11-4 DRIVER
D7
TP11 +12 V
J6-1 J11-1
J6-2 J11-2
DRAWN DATE
G. SANWALD 11 FEB 2003 LOW SPEED
CHECKED STARTER
Steve Blake 18 FEB/03
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0837 REV B
SHEET 1 OF 2
NOTE
REMARKS
REFERENCE
1 MAIN STATOR CURRENT SENSE. IF MAIN CURRENT IS LOW, THIS WILL BE APPROXIMATELY +5 VDC. PULSES AT 120 HZ WILL BE PRESENT AS SHOWN IN FIGURE 1 AT NORMAL STATOR CURRENT.
2 PHASE-SHIFT STATOR CURRENT SENSE. IF SHIFT CURRENT IS LOW, THIS WILL BE APPROXIMATELY +5 VDC. PULSES AT 120 HZ WILL BE PRESENT AS SHOWN IN FIGURE 1 AT NORMAL STATOR CURRENT.
3 “LOW” (APPROXIMATELY 0 VDC) FOR APPROXIMATELY 1.8 SEC DURING PREP, THEN PULSED LOW FOR 500 MSEC EVERY 5 SECONDS DURING PREP. REFER TO FIGURE 2.
“LOW” (APPROXIMATELY 0 VDC) ENERGIZES K4 ON THE H.V. AUXILIARY BOARD, (”HIGH”, + 12 VDC = NOT ENERGIZED). THIS RELAY IS ENERGIZED AFTER THE MAIN BUS CAPACITORS ARE CHARGED,
4
APPROXIMATELY 10 SECONDS AFTER POWER-ON.
+5 V
FIGURE 1
THIS APPROACHES 0 V AS CURRENT INCREASES
0V
5V K5 DE-ENERGIZED
FIGURE 2
0V K5 ENERGIZED
PREP STARTED
DRAWN DATE
G. SANWALD 11 FEB 2003 LOW SPEED
CHECKED STARTER
Steve Blake 18 FEB/03
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0837 REV B
SHEET 2 OF 2
REFER TO PAGE 7 FOR
LOGIC LEVELS, NOTES,
ETC, REFERENCED BY
THIS SYMBOL:
U2 U26
J7-2
CHAMBER 4 SELECT
J7-3
CHAMBER 3 SELECT
J7-4
CHAMBER 2 SELECT
J7-5
CHAMBER 1 SELECT
DRIVER J7-6
CPLD START
J7-7
LEFT FIELD SELECT
J7-8 TO AEC BOARD.
MIDDLE FIELD SELECT
J7-9
REFER TO SHEETS 2-6
RIGHT FIELD SELECT AS APPLICABLE FOR THE
AEC BOARD IN YOUR UNIT
+5V
VOLTAGE
AMPLIFIER
R100
TP21 1 COMPARATOR -
- R101 R102 J7-15
+ PT RAMP
+
U23B J7-10
U24 PT REF
R104
R107
J7-1
PT STOP
2
TP6
R108
R219
U47
D/A 6-
CONV 7
5+
U50B
U3 U56D TP22
3
CPU
R10
DRAWN DATE
G. SANWALD 13 FEB 2003
CHECKED AEC
Steve Blake 18 FEB/03
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0838 REV C
SHEET 1 OF 7
+12V +12V +12V +12V “L” SELECT J1-1 AEC CH 1
DS1 DS2 DS3 DS4 J1-2
“R” SELECT ANODE
J1-3
“M” SELECT J1-4
R37
R38
R39
R40
CATH (L)
J1-5
CATH (R)
R43 R75 R74 J1-6
J7-2 J5-2 D6 CATH (M)
CHAMBER 4 SELECT
2 J1 (shell)
U2D 1 -
R72 R71 GROUND
J7-3 J5-3 D12 R31 3
+
CHAMBER 3 SELECT
J11-1
U2C U11A GROUND
R73
R30
JW1
J7-4 J5-4 D10 J11-2
CHAMBER 2 SELECT ANODE
J11-3
R1
U2B CATH (L)
J7-5 J5-5 D11 R29 J11-4
CHAMBER 1 SELECT CATH (M)
J11-5
U2A CATH (R)
R28
R32
R42
R41
U6A
D13 D16 D15 D14
“L” SELECT J2-1 AEC CH 2
J7-6 J5-6
START J2-2
“R” SELECT ANODE
J7-7 J5-7 J2-3
LEFT FIELD SELECT
“M” SELECT J2-4
J7-8 J5-8 +12V CATH (L)
MIDDLE FIELD SELECT J2-5
J7-9 J5-9 +12V CATH (R)
R69 R76 R77 J2-6
RIGHT FIELD SELECT CATH (M)
+12V
R87 6 J2 (shell)
R61 7 -
R58 R57 GROUND
R63 Q1 5
R54 +
Q2 J12-1
R55 U11B GROUND
“L” SELECT
R59
JW2
J12-2
R86
Q3 ANODE
“M” SELECT J12-3
R62
R2
CATH (L)
“R” SELECT J12-4
R64
FROM CATH (M)
PAGE 1 J12-5
D30 CATH (R)
R82
D24
+12V U6B
D18
R22
“L” SELECT
D19 J3-1 AEC CH 3
R45
DS5 J3-2
“R” SELECT ANODE
J3-3
R11
R19
JW3
J13-2
PAGE 3 SAMPLE ANODE
J7-11 J5-11 & HOLD J13-3
R3
-12V R51 CATH (L)
J7-12 J5-12 U12 J13-4
CATH (M)
J13-5
4 TP3 C37 CATH (R)
TP1 2 U6C
R23 - R60 “L” SELECT
J7-15 J5-15 R14
1 -
2 -
J4-1 AEC CH 4
PT RAMP 3 +
J4-2
+ +
J7-10 J5-10 R88 U10B “R” SELECT ANODE
R81
PT REF U10A U9B J4-3
R80
R33
J14-2
JW4
7 -
2 ANODE
+ J14-3
R4
CATH (L)
R50
U13 J14-4
D33 CATH (M)
J14-5
CATH (R)
U6D
GENERATOR CONTROL BOARD AEC BOARD
THIS SHEET APPLIES TO AEC BOARD ASSEMBLY 733347 DESIGNATES AN ANALOG SWITCH. THESE ARE I.C. “SWITCHES”
THAT ARE SWITCHED ON / OFF BY APPLYING THE APPROPRIATE DRAWN DATE
REFER TO CHAPTER 3D FOR INSTALLATION AND CALIBRATION DETAILS. LOGIC LEVEL (0V = OFF, 12V = ON). G. SANWALD 13 FEB 2003
CHECKED AEC
Steve Blake 18 FEB/03
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0838 REV C
SHEET 2 OF 7
+500V
RIGHT +/-300V
J1-2
+/-300 V AEC CH 1
+12V +12V +12V +12V +12V TP10 J11-1
6 JW8 H.V. J1-1
+12V 1 +/-300/+500 V
*
RN6C
RN6D
RN6B
RN6A
RN6E
TP22 2 J1-3
+45V +45 V
+/-300V 3 J11-4
TP24 H.V. START J1-8
DS5 DS1 DS2 DS3 DS4 LEFT START
J11-3
MIDDLE J1-10
RN4A JW7 M FIELD SEL
J7-2 J5-2 D6 J11-2
CHAMBER 4 SELECT CH 4 1 L/R J1-9
TP9 2 L FIELD SEL
U3C 3 J11-6
J7-3 J5-3 RN3D CONVERTER R/L J1-11
D5 R FIELD SEL
CHAMBER 3 SELECT CH 3 +12V +24V CIRCUIT J11-5
R66
12 VDC TO 45 VDC TP11 J1-12
U3B R89 SIGNAL
J7-4 J5-4 D12 RN3C AND 300 / 500 VDC 9 J11-9
CHAMBER 2 SELECT CH 2 J1-7
+45V CH 1 GROUND
U3F * INCLUDES U7 AND T1
J7-5 J5-5 D4 RN3A
CHAMBER 1 SELECT CH 1 S1D R90
D16 **
U3D
J7-6 J5-6 D11 RN3B Q1
SET
R79
START STRT
VALUE RIGHT +/-300V
J2-2
+/-300 V AEC CH 2
U3E U3A * START J12-1
JW6 H.V. J2-1
1 +/-300/+500 V
2 J2-3
+12V +24V +12V +24V +12V +24V +45V +45 V
3 J12-4
START J2-8
LEFT START
J12-3
MIDDLE J2-10
JW5 M FIELD SEL
J5-7
* * * 1 J12-2
L/R
J7-7 J2-9
LEFT FIELD SELECT 2 L FIELD SEL
D30 D40 D65 3 J12-6
J7-8 J5-8 R/L J2-11
MIDDLE FIELD SELECT Q2 Q3 Q4 R FIELD SEL
J12-5
J7-9 J5-9 TP12 J2-12
RIGHT FIELD SELECT * * * R67 SIGNAL
LEFT MIDDLE RIGHT 9 J12-9
J2-7
CH 2 GROUND
FROM
PAGE 1 S1C R68
RIGHT +/-300V
J3-2
+/-300 V AEC CH 3
TP18 J13-1
TP17 4 TP20 5 JW4 H.V. J3-1
J7-19 J5-19 CH 1 1 +/-300/+500 V
+24V J3-3
2 +45V +45 V
TP23 - STRT R1 S2A 3 J13-4
START J3-8
- CH 2 START
J7-13 J5-13 TP5 TP6 TP7 + S4 LEFT
+12V J13-3
FROM +
U4A MIDDLE J3-10
J7-17 J5-17 U4B R2 S2B JW3 M FIELD SEL
MD-0831 R32 J13-2
C11 CH 3 1 L/R J3-9
PAGE 3 * 2 L FIELD SEL
3 J13-6
TP4 C4 S2C R/L J3-11
R3 R FIELD SEL
CH 4 J13-5
J7-11 J5-11 CH 1 R53 TP13 J3-12
-12V - R57 SIGNAL
R4 S2D 9 J13-9
S3A + J3-7
CH 3 GROUND
U2B
R11 CH 2 S1B R58
TP3 2 TP19 7
S3B
TP2 8
J7-15 J5-15 R54 RIGHT +/-300V
J4-2
+/-300 V AEC CH 4
PT RAMP R12 J14-1
RN4C CH 3 JW2 H.V. J4-1
J7-10 J5-10 +/-300/+500 V
PT REF 1
2 J4-3
J7-1 J5-1 S3C +45V +45 V
PT STOP 3 J14-4
START J4-8
R13 CH 4 - LEFT START
- J14-3
+
MIDDLE J4-10
+
JW1 M FIELD SEL
S3D U1A 1 J14-2
D27 U2A L/R J4-9
R56 STRT 2 L FIELD SEL
R14 3 J14-6
R/L J4-11
RN4B R FIELD SEL
3 J14-5
7 - TP14 J4-12
2 R35 SIGNAL
+
SAMPLE 9 J14-9
U6 J4-7
D22 & HOLD CH 4 GROUND
S1A R36
GENERATOR CONTROL BOARD AEC BOARD
THIS SHEET APPLIES TO AEC BOARD ASSEMBLY 734614 DESIGNATES AN ANALOG SWITCH. THESE ARE I.C. “SWITCHES” THAT ARE DRAWN DATE
REFER TO CHAPTER 3D FOR INSTALLATION AND CALIBRATION DETAILS. SWITCHED ON / OFF BY APPLYING THE APPROPRIATE LOGIC LEVEL. G. SANWALD 13 FEB 2003
CHECKED AEC
THE +/- 12V OUTPUTS ON J1 TO J4 AND J11 TO J14 ARE NOT SHOWN ON DESIGNATES A FACTORY CONFIGURED LOGIC OR SIGNAL LEVEL. AEC Steve Blake 18 FEB/03
THIS DIAGRAM. THESE ARE DETAILED ON THE CONNECTOR PIN OUT BOARDS ARE CONFIGURED AT THE TIME OF ORDER TO BE COMPATIBLE DES.\MFG.\AUTH.
TABLES IN CHAPTER 3D. * WITH THE SPECIFIED AEC CHAMBER(S). FOR EXAMPLE, THE START
ED WOOD FEB 13/03 MD-0838 REV C
** SIGNAL TO THE CHAMBER MAY BE FACTORY CONFIGURED TO BE ACTIVE
R79 ADJUSTS THE +45V, +300V, AND +500V OUTPUTS FROM THE DC TO DC LOW (0 V), ACTIVE HIGH (+12 V), OR ACTIVE HIGH (+24 V). SHEET 3 OF 7
CONVERTER CIRCUIT. REFER TO CHAPTER 3D FOR DETAILS.
AEC CH 1 AEC CH 3
RIGHT 1 RIGHT 1
SELECT SELECT
2 2
JW1 JW5
LEFT 3 R5 RN1 RN1 LEFT 3 R6 RN2 RN2
3 6 7 2 J1-2 3 6 7 2 J3-2
SELECT ANODE (R) SELECT ANODE (R)
2 J1-5 2 J3-5
R69 - CATH (R) R70 - CATH (R)
1 RN1 RN1 1 RN2 RN2
3 4 5 8 1 J1-1 3 4 5 8 1 J3-1
U4A + ANODE (M) U5A + ANODE (M)
U2A J1-6 U3A J3-6
CATH (M) CATH (M)
R10
R50
J1-3 J3-3
ANODE (L) ANODE (L)
J1-4 J3-4
CATH (L) CATH (L)
MIDDLE MIDDLE
SELECT J1 (shell) SELECT J3 (shell)
R13 RN7 RN7 GROUND R14 RN8 RN8 GROUND
3 6 7 2 3 6 7 2
R9 -
6 R11 -
6
7 RN7 RN7 7 RN8 RN8
+
5 4 5 8 1 +
5 4 5 8 1
U4B U5B
U2B U3B
R7
R8
LEFT 1 J11-5 LEFT 1 J13-5
ANODE (R) ANODE (R)
SELECT J11-6 SELECT J13-6
2 CATH (R) 2 CATH (R)
JW2 J11-3 JW6 J13-3
3 ANODE (M) 3 ANODE (M)
RIGHT R27 RN9 RN9 J11-4 RIGHT R28 RN10 RN10 J13-4
3 6 7 2 CATH (M) 3 6 7 2 CATH (M)
SELECT SELECT
2 J11-1 2 J13-1
R22 - ANODE (L) R23 - ANODE (L)
1 RN9 RN9 1 RN10 RN10
3 4 5 8 1 J11-2 3 4 5 8 1 J13-2
U4C + CATH (L) U5C + CATH (L)
U8A J11-7 U9A J13-7
R16
R17
CH 1 CH 3
SELECT U4D SELECT U5D
CH 1 CH 3
OUT OUT
(SHT 6) (SHT 6)
AEC CH 2 AEC CH 4
RIGHT 1 RIGHT 1
SELECT SELECT
2 2
JW3 JW7
LEFT 3 R43 RN11 RN11 LEFT 3 R44 RN12 RN12
3 6 7 2 J2-2 3 6 7 2 J4-2
SELECT ANODE (R) SELECT ANODE (R)
6 J2-5 6 J4-5
R40 - CATH (R) R41 - CATH (R)
7 RN11 RN11 7 RN12 RN12
5 4 5 8 1 J2-1 5 4 5 8 1 J4-1
U14A + ANODE (M) U15A + ANODE (M)
U8B J2-6 U9B J4-6
CATH (M) CATH (M)
R33
R34
J2-3 J4-3
ANODE (L) ANODE (L)
J2-4 J4-4
CATH (L) CATH (L)
MIDDLE MIDDLE
SELECT J2 (shell) SELECT J4 (shell)
R51 RN13 RN13 GROUND R52 RN14 RN14 GROUND
3 6 7 2 3 6 7 2
R47 -
2 R49 -
2
1 RN13 RN13 1 RN14 RN14
+
3 4 5 8 1 +
3 4 5 8 1
U14B U15B
U16A U17A
R45
R46
LEFT 1 J12-5 LEFT 1 J14-5
ANODE (R) ANODE (R)
SELECT J12-6 SELECT J14-6
2 CATH (R) 2 CATH (R)
JW4 J12-3 JW8 J14-3
3 ANODE (M) 3 ANODE (M)
RIGHT R57 RN15 RN15 J12-4 RIGHT R58 RN16 RN16 J14-4
3 6 7 2 CATH (M) 3 6 7 2 CATH (M)
SELECT SELECT
6 J12-1 6 J14-1
R55 - ANODE (L) R56 - ANODE (L)
7 RN15 RN15 7 RN16 RN16
5 4 5 8 1 J12-2 5 4 5 8 1 J14-2
U14C + CATH (L) U15C + CATH (L)
U16B J12-7 U17B J14-7
R54
R53
CH 2 CH 4
SELECT U14D SELECT U15D
CH 2 CH 4
OUT OUT
(SHT 6) (SHT 6)
THIS SHEET APPLIES TO AEC BOARD ASSEMBLY 737992. THIS DESIGNATES AN ANALOG SWITCH. THESE ARE I.C. “SWITCHES”
PAGE SHOWS THE INPUT CIRCUITS; THE SIGNAL PROCESSING THAT ARE SWITCHED ON / OFF BY APPLYING THE APPROPRIATE DRAWN DATE
CIRCUITS ARE CONTINUED ON THE NEXT PAGE. LOGIC LEVEL (0V = OFF, 5V = ON). G. SANWALD 13 FEB 2003
CHECKED AEC
Steve Blake 18 FEB/03
REFER TO CHAPTER 3D FOR INSTALLATION AND CALIBRATION DETAILS. DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0838 REV C
SHEET 4 OF 7
+5V +5V +5V +5V
8 7 6 5
RN3
RN3
RN3
RN3
1 2 3 4 U6
DS4 DS3 DS2 DS1
J7-2 J5-2 R67 2 18 CH 1 OUT
CHAMBER 4 SELECT CH 4 SELECT (FROM SHT 5)
+5V R15
R36
TP1 10 R18
R31
J7-11 J5-11 U12B
-12V
R37
R42
CH 1
SELECT
R59 R1
TP4 4 TP3 CH 2
2 C31 SELECT
SAMPLE
J7-15 J5-15 R26 & HOLD
PT RAMP R60 R2
-
2
J7-10 J5-10 1 CH 3
PT REF +
3 SELECT CH 4 OUT
J7-1 J5-1 (FROM SHT 5)
PT STOP U11A
+12V R3
CH 4
R19
+5V SELECT
R20
R4
3 D37
R12
R24
TP5
D38 R30
-
3
7
+
2
U10
THIS SHEET APPLIES TO AEC BOARD ASSEMBLY 737992. THIS DESIGNATES AN ANALOG SWITCH. THESE ARE I.C. “SWITCHES”
PAGE SHOWS THE SIGNAL PROCESSING CIRCUITS; THE INPUT THAT ARE SWITCHED ON / OFF BY APPLYING THE APPROPRIATE DRAWN DATE
CIRCUITS ARE SHOWN ON THE PREVIOUS PAGE. LOGIC LEVEL (0V = OFF, 5V = ON). G. SANWALD 13 FEB 2003
CHECKED AEC
Steve Blake 18 FEB/03
REFER TO CHAPTER 3D FOR INSTALLATION AND CALIBRATION DETAILS.
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0838 REV C
SHEET 5 OF 7
RIGHT AEC CH 1
+5V +5V +5V +5V SELECT
JW8 NO CONNECTION N/C
+5V 1
J1-2
R51
R49
R50
R52
2 LEFT/RIGHT
3 J1-6
U6 LEFT LEFT/RIGHT
R39
DS1 DS2 DS3 DS4 SELECT MIDDLE J1-3
R61 +12V JW7 MIDDLE
J7-2 J5-2 2 18 SELECT
CHAMBER 4 SELECT +12V
CH 4 SELECT START/ 1 J1-4
START START
2
R62 R40 D12 3 TP1 J1-8
J7-3 J5-3 3 17 Q5 +12V
CHAMBER 3 SELECT CH 3 SELECT 9
R1 J1-5
START SIGNAL
J7-4 J5-4 R63 4 16
CHAMBER 2 SELECT CH 2 SELECT R38 CH 1 J1-7
Q4 SELECT -12V
J7-5 J5-5 R64 5 15 J1-9
CHAMBER 1 SELECT GROUND
CH 1 SELECT U1D R2
INVERTING -12V
BUFFER
RIGHT AEC CH 2
SELECT
J7-6 J5-6 R65 D20 6 14
START JW6 NO CONNECTION N/C
R66 1
J7-7 J5-7 7 13 J2-2
LEFT FIELD SELECT 2 LEFT/RIGHT
R67 3 J2-6
J7-8 J5-8 8 12 LEFT/RIGHT
MIDDLE FIELD SELECT LEFT
J7-9 J5-9 R68 9 11 SELECT MIDDLE J2-3
RIGHT FIELD SELECT JW5 MIDDLE
SELECT
+12V
1 J2-4
START START
2
+12V +12V +12V 3 TP2 J2-8
9 +12V
R3 J2-5
SIGNAL
+12V +5V D11 D10 D9
CH 2 J2-7
R37
R36
R35
U3 RIGHT MIDDLE LEFT SELECT -12V
SELECT SELECT SELECT J2-9
+5 V GROUND
FROM U1C R4 -12V
REGULATOR Q3 Q2 Q1
PAGE 1
J7-13 J5-13
RIGHT AEC CH 3
SELECT
TP11 TP12 JW4 NO CONNECTION N/C
FROM 1
J7-17 J5-17 J3-2
2 LEFT/RIGHT
MD-0831 TP7 13 3
PAGE 3 CH 1 SELECT J3-6
LEFT LEFT/RIGHT
J7-11 J5-11 SELECT MIDDLE J3-3
-12V - R11 JW3 MIDDLE
CH 2 SELECT SELECT
START/ +12V
+ 1 J3-4
START START
U4A 2
R12 3 TP3 J3-8
CH 3 SELECT 9 +12V
* R5 J3-5
SIGNAL
R13 CH 3
CH 4 SELECT J3-7
SELECT -12V
- J3-9
R14 GROUND
TP9 4 TP8 + U1B R6 -12V
2 U8B
MIDDLE J4-3
+
JW1 MIDDLE
R34
+ SELECT
U7A +12V
U8A 1 J4-4
START/ START START
3 2
D25 3 TP4 J4-8
R33 +12V
TP10 9
R7 J4-5
R31 SIGNAL
3 SAMPLE
7 - CH 4 J4-7
2 & HOLD SELECT -12V
+
U9 J4-9
GROUND
U1A R8 -12V
THIS SHEET APPLIES TO AEC BOARD ASSEMBLY 737998 DESIGNATES AN ANALOG SWITCH. THESE ARE I.C. “SWITCHES” THAT ARE DRAWN DATE
REFER TO CHAPTER 3D FOR INSTALLATION AND CALIBRATION DETAILS. SWITCHED ON / OFF BY APPLYING THE APPROPRIATE LOGIC LEVEL. G. SANWALD 13 FEB 2003
CHECKED AEC
Steve Blake 18 FEB/03
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0838 REV C
SHEET 6 OF 7
NOTE
REMARKS
REFERENCE
1 GENERATES A PULSE PER FIGURE 1 WHEN THE AEC RAMP IS APPROXIMATELY 5 % OF THE AEC REFERENCE IF THE AEC RAMP IS ON THE CORRECT TRAJECTORY.
2 AEC REFERENCE VOLTAGE, 0 TO +10 VDC, DEPENDING ON AEC TECHNIQUE. THE LENGTH OF THE AEC EXPOSURE IS PROPORTIONAL TO THE AEC REFERENCE VOLTAGE.
3 AEC STOP (PT STOP) SIGNAL. THIS IS NORMALLY HIGH (APPROXIMATELY +5 VDC), SWITCHING LOW WHEN THE AEC RAMP = THE AEC REFERENCE VOLTAGE. REFER TO FIGURE 1.
4 AEC RAMP. THIS IS A SIGNAL RAMPING FROM 0 TOWARD +10 VDC, THE ACTUAL MAGNITUDE WILL DEPEND ON THE AEC TECHNIQUE. REFER TO FIGURE 2.
5 AS PER # 4.
6 PWM OUTPUT. THIS WILL BE VARIABLE WIDTH PULSES (PULSE WIDTH INCREASING AT INCREASING LOAD), UP TO A MAXIMUM OF 50% DUTY CYCLE. REFER TO FIGURE 4.
7 AEC RAMP OR DC VOLTAGE. THIS IS A RAMP OR DC VOLTAGE, DEPENDING ON AEC CHAMBER TYPE (INTEGRATING OR NON-INTEGRATING). REFER TO FIGURE 3.
8 AS PER # 7, EXCEPT THAT THE RAMP OR DC VOLTAGE WILL BE POSITIVE GOING AND NOT OF THE SAME MAGNITUDE.
9 THE VOLTAGE AT THIS TEST POINT IS THE OUTPUT OF THE AEC CHAMBER. REFER TO THE AEC CHAMBER MANUFACTURERS DOCUMENTATION FOR DETAILS.
10 THIS IS THE START SIGNAL. “HIGH” (5 VDC) = START = ANALOG SWITCHES CLOSED, “LOW” (0 VDC) = START = ANALOG SWITCHES OPEN.
11 THIS WILL BE A NEGATIVE DC VOLTAGE. THE MAGNITUDE OF THE DC VOLTAGE IS DEPENDENT ON THE AEC TECHNIQUE IN USE.
12 AS PER # 7, EXCEPT THAT THE POLARITY WILL BE POSITIVE.
13 THE POLARITY AND MAGNITUDE OF THE RAMP AT THIS POINT SHOULD BE APPROXIMATELY THE SAME AS THE PT RAMP OUTPUT. NOTE REFERENCE 4.
+5 V
FIGURE 1
0V
AEC STOP
<10 VDC
FIGURE 2
0V
0V
FIGURE 3 OR
< -10 VDC
INTEGRATING NON-INTEGRATING
AEC CHAMBERS AEC CHAMBERS
200 kHz
12 VDC
FIGURE 4
0 VDC DRAWN DATE
G. SANWALD 13 FEB 2003
CHECKED AEC
Steve Blake 18 FEB/03
DES.\MFG.\AUTH.
ED WOOD FEB 13/03 MD-0838 REV C
SHEET 7 OF 7
+5V +5V
DS1 DS1
2 TP1 1 Hz +5V +5V +5V +5V 2 TP1 1 Hz
DS2 DS3 DS3 DS2
R10
R11
U3
U17 1
R19
R20
R17
R16
U8 U8
TP10 TP11
R23 J8-1 J3-1 R19
CPU
R24 J8-2 J3-2 R18
CPU
+5V
DS5
TXD
+5V RS-232
(LAPTOP)
R22
DS4 U11 J4
RXD R25
3 TXD
R21
R26
7 RTS DATA BUS
D0..D7
R27
2 RXD
DATA BUS R28
D0..D7 8 CTS
DRAWN DATE
G. SANWALD 11 FEB 2003 SERIAL
CHECKED COMMUNICATIONS
Steve Blake 18 FEB/03
DES.\MFG.\AUTH.
ED WOOD FEB 13/03
MD-0839 REV A
SHEET 1 OF 1
+15V
Q6
+15V
Q7
+5V
J4-1
SWITCHED +15V
Q4 J4-6
TEST +15V
R43
Q5 J4-7
TEST +15V
DS4 J4-2
+ DOSE
U11 J4-3
- DOSE
DAP
+5V +5V +5V J4-4
CHAMBER
OPTO
J4-5
RELAY
R42
R47
R48
J4-8
U13 GROUND
MICRO-
J4-9
CONTROLLER GROUND
RS-422
DRIVER
+15V
R50
R54
Q8
R53
D7
DATA BUS
D0..D7
DRAWN DATE
G. SANWALD 11 FEB 2003
CHECKED DAP
Steve Blake 18 FEB/03
DES.\MFG.\AUTH.
ED WOOD FEB 13/03
MD-0840 REV A
SHEET 1 OF 1
TOUCHSCREEN TRANSFORMER.
THE PRIMARY TAP WILL
BE SET TO MATCH THE
LINE VOLTAGE IN USE
480V
400V
230V
0V
F1 J2-1
E1 J2-4
J2-5
E2
E3
J3-4 F2
J3-1 F3
DRAWN DATE
G. SANWALD 07 JUNE 2005 TOUCHSCREEN
CHECKED
ISOLATED
JAC 07 JUNE 2005 115 VAC SUPPLY
DES.\MFG.\AUTH.
J. CUNNINGHAM 07 JUNE 2005
MD-0868 REV A
SHEET 1 OF 1
SEE MD-0837
J4 J2 J7
PRIMARY OF
H.V. AUXILIARY BOARD J8
AUXILIARY TRANSFORMER J9
SEE MD-0831 SHEET 1, 2, 3 208 / 230 VAC
24 VAC & 24 ASSEMBLY 736173 SCH 736171
VDC OUT, ROOM J5
J11
LIGHT OUT 400 / 480 VAC
SEE MD-0833 ASSEMBLY 737182 SCH 737180 12 VDC DRIVE FOR MAIN
J10 POWER CONTACTOR
SECONDARY OF
J6 SEE MD-0831 SHEET 1, 2, 3
AUXILIARY TRANSFORMER J3
J1
SEE MD-0831 SHEET 1, 2, 3
SEE
SHEET
2
FROM DAP
CHAMBER J4 J12 J15 J4
SEE MD-0840 J3 INTERNAL SYSTEM
J2 J1
GROUNDS
J3
X-RAY TUBE
J5 HOUSING GROUND
CONTROL BOARD FILAMENT SUPPLY HT TANK
ASSEMBLY 736763 BOARD
FOR FUTURE SCHEMATIC 736761 J13 ASSEMBLY 731407
J8
USE SEE SCHEMATIC 731405
J6 SHEET 2
J10
J14
J16
J9 J5
J1 J7 J2 J11
INVERTER BOARD
ASSEMBLY 736139 SCH 736137
ASSEMBLY 736885 SCH 736883
J5 E10
CONTROL BOARD E9
ASSEMBLY 736763
SCHEMATIC 736761 J13 J1
J8
HT TANK
J6
E17 E18
J10
THIS APPLIES TO UNITS WITH ONE INVERTER BOARD (30 - 40 kW). SEE PAGE 3 FOR UNITS WITH TWO INVERTER BOARDS (50 kW UNITS)
SEE
SHEET
3
SERIAL PORT TO
NOTE BOOK COMPUTER
SEE MD-0839
TO
HAND SWITCH
SEE MD-0834 J3 J4 AC OUT TO PRIMARY
J8
OF TOUCHSCREEN
TRANSFORMER
INVERTER BOARD
ASSEMBLY 732813 SCH 732811
J4 ASSEMBLY 736885 SCH 736883
J12 J15 J4
J3 SHARING
INDUCTOR
J1 J3
J5
CONTROL BOARD E10
ASSEMBLY 736763 E9
SCHEMATIC 736761 J13
J8
J6 HT TANK
J10
E17 E18
J14
J16 J2
J9 mA/mAs PORT
J1 J7 J2 J11
INVERTER BOARD
ASSEMBLY 732813 SCH 732811
ASSEMBLY 736885 SCH 736883
SHARING
INDUCTOR
J1
AEC BOARD (OPTIONAL)
REFER TO MD-0838
AND CHAPTER 3D DEPENDING ON MODEL, ONE OR TWO
OF SERVICE MANUAL INVERTER BOARDS WILL BE USED
J5
THIS APPLIES TO UNITS WITH TWO INVERTER BOARDS (50 kW). SEE PAGE 2 FOR UNITS WITH ONE INVERTER BOARD (30-40 kW UNITS)
DRAWN DATE
G. SANWALD 18 MAY 2004
INTERCONNECT
CHECKED
DIAGRAM
JAC 18 MAY 2004 (CMP 200)
DES.\MFG.\AUTH.
E. WOOD 18 MAY 2004 MD-0845 REV C
SHEET 3 OF 3