Model DS1200-AMT: Single Channel Dew Point Hygrometer Ranges Available Between - 120°C To +20°C (-166°F To +68°F) Dewpoint
Model DS1200-AMT: Single Channel Dew Point Hygrometer Ranges Available Between - 120°C To +20°C (-166°F To +68°F) Dewpoint
Combination of advanced digital technology of the display electronics and the transmitter, makes this the
system of choice for a wide range of applications where precise indication, control and monitoring is critical.
Model DS1200-AMT system can be factory configured to read in °C dewpoint, °F dewpoint, ppm(v), ppb(v),
g/m³ or lbs/MMSCF. The measured value is displayed on a large 4 digit, 13.8mm high, fourteen segment
LED display. Model DS1200-AMT Dewpoint Hygrometer is fully configured to the customer's specifications.
The instrument offers two fully programmable alarm relays, which can be independently set across the
full operating range. An optically isolated analogue signal, linear with the measured unit is also available
(to interface with any external device).
Model DS1200-AMT can be supplied with any one of a range of dewpoint transmitters, which have all
been specially manufactured to provide long term stability, so ensuring a long and trouble free service life.
The long lasting sensing element offers excellent sensitivity, repeatability and response speed - especially
on dry down. Integrated electronics in each transmitter allow calibration data to be stored within its
memory, making transmitters completely field interchangeable.
Each transmitter is calibrated to provide overall system accuracy of better than ±2°C dewpoint and
supplied with a certificate traceable to international humidity standards.
Model DS1200-AMT also incorporates the invaluable “AUTOCAL” feature for easy field verification of the
system response, without recourse to any additional equipment. Accidental or unauthorised operation is
prevented by a simple initiation routine.
The system allows the display unit to be sited in excess of 1000m from the transmitter, connected by a
simple twisted pair cable.
© Alpha Moisture Systems Ltd. 1 OF 4
Alpha Moisture Systems Model DS1200-AMT
Specifications
MODEL DS1200-AMT Dewpoint Hygrometer SENSOR PROTECTION: 316 Sintered stainless steel filter - 50 micron
OPERATING PRESSURE: From 1kPa (0.01 barA) to Maximum
DS1200 DISPLAY INDICATOR SPECIFICATIONS 35,000kPa ( 350 barA)
ENCLOSURE: 1/8 DIN panel mounting ABS case SAMPLE FLOW RATE: Flow independent but ideally 2 to 5 standard
DISPLAY: 4 digit, 14 segment LED indicator. Red characters litres per minute. Max: 25 litres/min
13.8mm high. Factory set to display in 0C, 0F, ppm(v), ppb(v), g/m3 WEATHERPROOF CLASSIFICATION: IP66 / NEMA4X when Connector
or lbs/MMSCF mated to Transmitter.
SUPPLY VOLTAGE: Universal 22Vac - 253Vac 50/60Hz & 20Vdc - MECHANICAL CONNECTION: 3/4” UNF (16tpi) with integral Viton
300Vdc 150mA to 15mA “O” ring seal.
POWER CONSUMPTION: 3.5W Maximum WEIGHT: Transmitter and connector - 175grams
}
OUTPUT SIGNAL: Linear 4 to 20mA, fully isolated. Maximum load CABLE: Supplied with 2m length standard two core cable.
800Ω / 16Vdc Nominal diameter 3.4mm, 92ohms/km at 200C
ISOLATION: Voltage Test / Operation - 2.3kVac / 250Vac ELECTROMAGNETIC COMPATIBILITY (EMC):
ALARM RELAYS: 2 independent fully programmable switched Complies with BS EN 61326-1
contacts. Rated 2A @ 250Vac / 1A @ 24Vdc. TEMPERATURE COEFFICIENT: Transmitter temperature
ANNUNCIATORS: AL1 & AL2 LED’s on front panel indicating status compensated for operating range.
of each alarm REPEATABILITY: Better than ±0.30C dewpoint
ELECTROMAGNETIC COMPATIBILITY (EMC): FACTORY CALIBRATION: Supplied with Certificate of Calibration
Complies with BS EN 61326-1 traceable to NPL / NIST
WEATHERPROOF CLASSIFICATION: Front panel only - IP65 (NEMA 4X). ACCURACY: ± 20C dewpoint (NPL / NIST traceable for range -900C to
WEIGHT: 230 grams +200C)
OPERATING TEMPERATURE: -200C to +600C
TRANSMITTER & GENERAL SPECIFICATIONS STORAGE TEMPERATURE: -200C to +700C
SENSOR TYPE: Model AMT Dewpoint Transmitter OPERATING HUMIDITY (External): Maximum - 95% RH
SENSING ELEMENT: Ultra High Capacitance - Aluminium Oxide Type Non-condensing
RANGES: See Step 1 below. WARM UP TIME: 10 seconds
AUTOCAL: Field calibration / Span check facility. Activated and MECHANICAL WARRANTY: 24 months in case of faulty
operated via two button switches on the Transmitter body. Locked workmanship and defective parts.
in normal operating mode to avoid tamper. CALIBRATION WARRANTY: 12 months (not valid if transmitter is
TRANSMITTER ENCLOSURE: 316 Stainless steel body with size C, subject to corrosives or mis-use)
DIN EN 175301 connector.
GYP:
SP:
1
1
to 6000
to 1000
ppm(v)
ppm(v)
resolution = 2ppm
resolution = 1ppm
How to Order - Step 3 of 3
MP: 0.1 to 100.0 ppm(v) resolution = 0.1ppm Select a transmitter holder below
LP: 0.01 to 10.00 ppm(v) resolution = 0.01ppm Comes compete with mounting plate.
2 grades available: 303 or 316 Stainless Steel
ADSH-AMT 303 +
PB: 1 to 1000 ppb(v) resolution = 1ppb
[X] [X]
Pipe Fittings
ADSH-AMT 316 +
HG: 0.01 to 10.00 g/m3 resolution = 0.01g/m3
LG: 0.001 to 1.000 g/m3 resolution = 0.001g/m3
LLB: 0.1 to 10.0 lbs/MMSCF resolution = 0.01 lbs/MMSCF X = Choose INLET Pipe Fitting
SLB: 0.1 to 25.0 lbs/MMSCF resolution = 0.01 lbs/MMSCF 0 = None 1/8” NPT Port
4 = 1/4” - Pipe OD
HLB: 0.1 to 50.0 lbs/MMSCF resolution = 0.01 lbs/MMSCF 8 = 1/8” - Pipe OD
VHLB: 0.1 to 300.0 lbs/MMSCF resolution = 0.1 lbs/MMSCF 6 = 6mm - Pipe OD
X = Choose OUTLET Pipe Fitting
0 = None 1/8” NPT Port
NOTE: ppm, ppb, g/m3 & lbs/MMSCF ranges are stated at atmospheric pressure.
4 = 1/4” - Pipe OD
Model AMT Transmitter can be configured to provide corresponding ranges for operation at other
8 = 1/8” - Pipe OD
constant pressure. This must be stated at time of order. 6 = 6mm - Pipe OD
Dimensions
Model AMT Dewpoint Transmitter
0.625 dia
15.9ø
28mm (1.1)
Process thread ¾ x 16 UNF
46mm 39mm
38mmø 3
4 Mounting holes 5.5mmø
55mm
90mm
65mm
25mm
15mm
45mm 45mm
Note: The Assembly is shown with 1/4@ OD tube fittings. The dimensions across the tube fittings will vary for all other size fittings.
Corrosive Gases: The Sensor should not be exposed to corrosive gases (or corrosive contaminants in the
gas sample) as these can chemically attack the sensor, impairing calibration accuracy and/or damaging
it beyond economic repair. Examples of such gases are mercury (Hg), ammonia (NH3), chlorine (Cl2) etc.
Strong oxidising agents such as ozone (O3) should also be prevented from coming into contact with the
sensor.