Quick Start Guide: T21P Indicator
Quick Start Guide: T21P Indicator
Quick Start Guide: T21P Indicator
ASSEMBLY
MOUNTING THE COLUMN BRACKET
The Indicator can be mounted to a 38mm scale
column (outside diameter) with the supplied
column bracket.
1. Route the cable through the column and
the column bracket.
2. Slide the column bracket down onto the
column and tighten the nut.
3. Solder the load cell cable to the supplied
load cell cable connector. (See Section
1.8.)
4. Slide the indicator onto the bracket as
shown below.
5. Plug the load cell cable into the Indicator’s load cell connector.
6. Remove the upper knob if necessary to change the viewing angle. Then re-install the
knob to fix the position.
LOAD CELL CONNECTION
Connect the load cell cable to the supplied load 5 4 3 2 1
cell cable connector.
9 8 7 6
NOTE: To use a 4-wire load cell cable, short
pins 2 to 8, and pins 6 to 7.
CAUTION:
Never connect a communication cable to the load cell connector of the Indicator.
Switch the indicator off when you insert or remove the load cell cable connector
from the Indicator.
NOTE: Connect the load cell cable connector to the load cell cable after routing the cable
through the column and column bracket. (The load cell cable connector cannot pass through
the supplied column bracket.)
• Verify that the input voltage printed on the power Adapter and the plug type matches
the local power supply.
• Make sure that the power cord does not pose an obstacle or tripping hazard.
• Disconnect the product from the power supply when cleaning.
• Do not operate this product in hazardous or unstable environments.
• Do not drop loads on the pan.
• Use only approved accessories and peripherals, as available.
• Operate the product only under ambient conditions specified in the Instruction Manual.
• Service should be performed by authorized personnel only.
CONTROLS
AC power
LED
Low Battery
LED Over, Accept &
Under LEDs
7-segment, 6-digit
LED display
Button
Menu Saves settings and N/A ENTER Decrement Increment value of Shift to next digit
Function exits menu Accepts current value of settings settings when setting
(Short Press) setting on display parameters
• Manually by pressing ZERO. (Scale must be stable and within the zero range.)
Manual Tare:
Place a container on the weighing pan and press TARE.
The display shows a net value of “0”.
To clear the Tare value, press TARE with the pan empty.
Weighing: Use this mode to measure the weight of a sample in the selected unit of measure.
1. Press ZERO to set the display to zero.
2. Put an empty container on the pan, and then press TARE to tare the container weight.
3. Add material to the container.
The display shows the weight of the material in the selected unit of measure.
Verify the indicator: place a known mass on the scale and check if the reading is accurate.
If not, calibrate, using the Calibration Guide.
ADDITIONAL OPERATIONS
Changing Units of Measure: Only one unit of measure can be active
at a time (kg or lb).
To select the active unit press and hold SET for three seconds.
Press repeatedly until F1-3 appears. Choose “0” for kg, “1” for lb.
F1-
F1-3
(If selecting “lb”, place the supplied “lb” label over “kg” on the overlay.)