QUICKCHECK Software Manual
QUICKCHECK Software Manual
QUICKCHECK
Version 1.5 or higher
D871.131.01/03 en 2011-08 Sa
QUICKCHECK
General Information
General Information
− The product bears the CE-mark "CE-0124" in − In case of any questions concerning service,
accordance with the Council Directive support or warranty please contact your sup-
93/42/EEC about Medical Devices and fulfills plier.
the essential requirements of Annex 1 of this
directive. − This manual is in conformity with the
The product is a class IIb device (MDD). QUICKCHECK software (version as indicated
on the front page) at printing date. All rights
− The user manual is an integral part of the are reserved for devices, circuits, techniques,
product. It should always be kept near the software and names referred to in the manual.
product. Observance of the manual is a pre-
requisite for proper product performance and − Windows is a trademark of Microsoft Corpora-
correct operation. tion, Inc.
− Operator safety, specified measuring accuracy − No part of the technical documentation may be
and interference-free operation can be guaran- reproduced without written permission from
teed only if original products and parts are PTW-Freiburg.
used. Furthermore only the accessories listed
in this manual are approved by PTW-Freiburg − PTW-Freiburg is registered manufacturer ac-
and may be used in conjunction with the prod- cording to the ElektroG (Elektro- und Elek-
uct, or else accessories whose use has been tronikgeräte-Gesetz).
expressly permitted by PTW-Freiburg. Safe Elektro-Altgeräte-Register (EAR) Registration
operation and proper product performance are number DE15599992
not guaranteed if accessories or consumables
− PTW-Freiburg works in strict accordance with
from other manufacturers are used.
a quality management system which is con-
− PTW-Freiburg cannot be held liable for dam- tinuously updated according to national and in-
ages resulting from the use of accessories, ternational standards.
consumables from other manufacturers or
when the user ignores the instructions and in-
formation given in this manual.
2 D871.131.01/03 en
QUICKCHECK
Contents
Contents
General Information 2
Contents 3
Illustrations 5
1 Intended Use 6
2 Safety Information 6
4 Licensing QUICKCHECK 14
4.1 General Information on Licensing 14
4.2 PTW License Control Center 14
4.3 Verifying the License 15
4.4 Registering QUICKCHECK, an Option or an Update 15
5 Definitions of Terms 16
7 Main Screen 17
7.1 Context Menu 18
D871.131.01/03 en 3
QUICKCHECK
Contents
11 Repeating a Measurement 31
12 Printing 31
14 QUICKCHECK Options 33
14.1 Options for the Data Transfer 33
14.2 Options for Program Start 34
14.3 Options for the Graphical Representations 34
14.4 Printout Options 35
16 Troubleshooting 44
Literature 45
Index 46
4 D871.131.01/03 en
QUICKCHECK
Illustrations
Illustrations
D871.131.01/03 en 5
QUICKCHECK
Intended Use
QUICKCHECK is a software for storage and visu- This is the safety alert symbol. It is used to
alization of the evaluation values and the measured alert the user to potential hazards. Obey all
values, which are determined with the QUICK- safety messages that follow this symbol to
CHECKwebline. QUICKCHECK does not perform avoid possible bodily injury or equipment
evaluations of measurements. damage.
QUICKCHECK receives the evaluation values and All safety messages consist of the following com-
the measured values from the QUICKCHECKwebline ponents:
via the selected interface (LAN, USB, RS232). – Safety alert symbol and signal word
– Type of danger
– Source of danger
– Consequence
– Measures to prevent hazard
Signal Words
DANGER
Indicates an imminent hazard. If not avoided, the
hazard will result in death or serious injury.
WARNING
Indicates a hazard. If not avoided, the hazard
can result in death or serious injury.
CAUTION
Indicates a potential hazard. If not avoided, the
hazard could result in minor injury or product /
property damage.
NOTE
Provides useful information to assure that you
get the most from your equipment.
6 D871.131.01/03 en
QUICKCHECK
Safety Information
Safety Information
WARNING CAUTION
Hazards originating from other system compo- QUICKCHECK is operated by non-qualified per-
nents. sons.
Maloperation! Maloperation!
Observe the safety information provided in the The use of QUICKCHECK is limited to persons
user manuals of the system components. who are familiar with the software and who are
adequately trained or experienced in the use of
such programs and who own the necessary de-
CAUTION cision-making competence to ensure proper use.
Influences by other software. This includes the following activities:
D871.131.01/03 en 7
QUICKCHECK
Safety Information
NOTE Disclaimer
It is recommended to back up the QUICKCHECK QUICKCHECK has been developed and tested with
data at regular intervals. all due care. Yet, PTW-Freiburg is not in a position
The data are located, for example, in the direc- to guarantee that the program is completely free of
tory '<PTW documents>\QUICKCHECK webline\ errors.
Data' (refer to section 3.5).
PTW-Freiburg is not responsible in any way for
consequences resulting from the use of
NOTE QUICKCHECK. Any conclusions drawn from these
Please observe the user manuals of all con- measurements are the sole responsibility of the
nected devices! user.
8 D871.131.01/03 en
QUICKCHECK
Installing, Starting and Exiting QUICKCHECK
Software: – MS Internet Explorer 6.0 or higher • Enter 'd:\setup.exe' (for drive D:\) and confirm
– Acrobat Reader 7.0 or higher with OK.
D871.131.01/03 en 9
QUICKCHECK
Installing, Starting and Exiting QUICKCHECK
Æ The window Customer Information is displayed. PTW-Freiburg recommends selecting the setup
type Complete especially if you use QUICKCHECK
for the first time.
With setup type Complete all available program
features will be installed in the directory
'C:\Program Files\'.
10 D871.131.01/03 en
QUICKCHECK
Installing, Starting and Exiting QUICKCHECK
If the data entered during installation are invalid, an 3.3 Repairing and Uninstalling
appropriate message appears. The window Pro- QUICKCHECK
gram Registration remains displayed to repeat en-
tering of registration data or to terminate registering You can uninstall QUICKCHECK or repair installa-
the program by clicking the Cancel button. In this tion errors.
case, a demo version of QUICKCHECK will be li-
censed which can be used for a period of 30 days. • Click Start → Settings → Control panel.
The licensing information can be entered at a later
date via the PTW License Control Center (see sec- • Double-click on Software.
tion 4, Licensing QUICKCHECK).
• Select QUICKCHECK and click the Change
If you had already installed a previous version with button.
valid licensing, abort registering by clicking the
Cancel button. In this case, the licensing remains Æ The window Program Maintenance is displayed.
also valid for the actual installed version. Please
keep in mind that this applies only if the major re-
lease numbers of the previous and of the actual
version are identical.
CAUTION
Working with an uncompleted version of
QUICKCHECK.
Malfunction!
If a restart of the system is required after the
installation of QUICKCHECK, it is essential to
restart the PC before the first use of the program
in order to ensure a correct installation of Figure 4: Window Program Maintenance
QUICKCHECK.
The window Program Maintenance offers the fol-
lowing options:
D871.131.01/03 en 11
QUICKCHECK
Installing, Starting and Exiting QUICKCHECK
2. <PTW documents>
e.g. demo data, measuring data or evaluation
data
− Windows XP:
'C:\Documents and Settings\All Users\
Shared Documents\PTW'
− Windows Vista / Windows 7:
'C:\Users\Public\Public Documents\PTW '
12 D871.131.01/03 en
QUICKCHECK
Installing, Starting and Exiting QUICKCHECK
D871.131.01/03 en 13
QUICKCHECK
Licensing QUICKCHECK
4 Licensing QUICKCHECK
2
3
1
1 List of all licensed PTW applications including 2 Click button to open the Registration dialog
version number and license status 3 Click button to view the software options
Registered for … = correct registration with user
4 Click button to close the PTW License Control
information
Center
Demo expired in … days = remaining days of
the demo version trial period
Demo expired = expired demo version
Unregistered = unregistered program version
14 D871.131.01/03 en
QUICKCHECK
Licensing QUICKCHECK
In the Modules info box you will see a list of all mo-
dules and their version numbers that are included in
the base version or in the option.
D871.131.01/03 en 15
QUICKCHECK
Definitions of Terms
A data set contains the administrative data as well • QUICKCHECK will automatically search for a
as the measured values and the evaluation values connected QUICKCHECKwebline and will define a
of one measurement. connection for data transfer. The connection will
be shown in the Device tab via Tools → Op-
tions….
Displayed Data Sets
• Transfer data sets from QUICKCHECKwebline:
In QUICKCHECK only that part of the loaded data Download button
sets will be displayed which is within the adjusted and
period and complies with the adjusted filter parame- Save transferred data sets: Save button
ters (refer to Figure 9).
or
WorklistGenerator
Load saved data sets: Open button
The PTW software WorklistGenerator is used to
configure the measurement tasks.
• Display transferred or loaded data sets
QUICKCHECK is used to display the measurement − Adjust period
results (data sets) of the QUICKCHECKwebline.
− Adjust filter parameters
− Check status information on the evaluation
values
• Exit QUICKCHECK
16 D871.131.01/03 en
QUICKCHECK
Main Screen
7 Main Screen
9 10 11
1
2
3
12
13
4
14
5 15
6 16
7 17
D871.131.01/03 en 17
QUICKCHECK
Main Screen
Print data
18 D871.131.01/03 en
QUICKCHECK
Loading, Saving and Closing of Data Sets
8.1 Setting up a Connection • Select by means of the serial number the de-
between PC and sired QUICKCHECKwebline.
QUICKCHECKwebline • Confirm the selection with the OK button.
• Before you start QUICKCHECK, connect the Æ For the selected QUICKCHECKwebline, a connec-
QUICKCHECKwebline to the PC via LAN, USB or
tion will be created automatically in the Connec-
RS232 (refer also to QUICKCHECKwebline user
tion Control Center dialog (refer to section 15).
manual).
Æ This connection will be displayed in the Device
• Start QUICKCHECK. tab via Tools → Options… as current setting.
D871.131.01/03 en 19
QUICKCHECK
Loading, Saving and Closing of Data Sets
NOTE
After the transfer of data sets from
QUICKCHECKwebline a backup copy of these data
sets will be created automatically. The backup
copies will be designated with date and time and
will be located in the directory
'<PTW application files>\QUICKCHECK webline\
Download'. Thus it is ensured that data sets
remain available even in case of deleted data
sets in QUICKCHECKwebline.
20 D871.131.01/03 en
QUICKCHECK
Loading, Saving and Closing of Data Sets
NOTE
Files with identical data sets (identical ID) can
result from copying files or downloading data
sets several times.
D871.131.01/03 en 21
QUICKCHECK
Representation of the Data Sets
1 2 3 4 5 6 7
1 List box to select a period 4 Selection field to set the start date of the period
2 Two buttons to move the period by one month 5 Selection field to set the end date of the period
(backward resp. forward) 6 Date and time of the selected data set
3 Two buttons to scale up resp. down the period 7 Number of data sets in the adjusted period
by 20 days
22 D871.131.01/03 en
QUICKCHECK
Representation of the Data Sets
9.2 Selecting the Filter The list boxes Modality, Field Size and Gantry An-
Parameters gle contain always all possible parameters.
By adjusting filter parameters you can determine Furthermore you can select the <all> setting for the
the data sets to be displayed. parameters Field Size and Gantry Angle. The
values of these two parameters will be displayed for
the selected data set in the Parameters area.
After loading data sets the last (recent) data set will
be selected and its parameters will be displayed.
− Energy: The All filter options dialog contains a table with the
Energy in [MV] or [MeV] different combinations of filter parameters of all
− Field Size: loaded data sets.
Field sizes 10 cm x 10 cm and 20 cm x 20 cm
• Select by means of a double-click the desired
− SDD: combination of filter parameters.
Source-Detector-Distance in [cm]
− Gantry Angle: Æ The filter parameters will be selected in the list
gantry angles 0°, 90°, 180° and 270° boxes and the appropriate data sets will be dis-
played.
− Wedge Angle:
wedge angle in [°]
D871.131.01/03 en 23
QUICKCHECK
Representation of the Data Sets
9.3 Representation of the When passing the Summary area with the mouse
Analysis the data sets exceeding the tolerances will be dis-
played as tooltip.
The Summary area shows a comprehensive survey
on the evaluation values of the adjusted period • To select a data set exceeding tolerances click
(Figure 11). the right mouse button and select the desired
data set.
This comprehensive survey does not depend on
the selected filter parameters (Figure 12). Æ The selected data set will be marked according
to the information given in section 9.4.
NOTE
If you have selected a data set exceeding the
tolerances in such a way the filter parameters will
be adapted according to that data set.
Figure 14: Summary area
Green:
all evaluation values in the adjusted
period are within the adjusted toler-
ances
Red:
at least one evaluation values of the
selected data set is outside the ad-
justed tolerances
Red flashing:
at least one evaluation values outside
of the selected data set but in the ad-
justed period is outside the adjusted
tolerances
24 D871.131.01/03 en
QUICKCHECK
Representation of the Data Sets
9.4 Graphical Representation of By means of the scrollbar on the right side of the
the Data Sets graphics you can scroll through the graphics.
As default the following evaluation values as well as The X axes of the graphics represent the adjusted
the measured values of all measuring chambers will period. Start and end value comply with the se-
be represented graphically in the main screen: lected date.
− Target:
expected evaluation value
− Norm:
normalization factor
D871.131.01/03 en 25
QUICKCHECK
Representation of the Data Sets
Figure 15: Graphical representation of the data sets (example: evaluation values)
26 D871.131.01/03 en
QUICKCHECK
Representation of the Data Sets
D871.131.01/03 en 27
QUICKCHECK
Representation of the Data Sets
9.5.2 Display of the Parameters and the 9.6 Views of the Main Screen
Comment for the Selected Data Set
The default setting shows the main screen with
If a data set is selected all additional measurement vertical split. To select this split click the View but-
parameters of that data set will be displayed in the ton and select the Vertical entry or select View →
Parameters area. Vertical.
The additional measurement parameters are: The advantage of the vertical split of the main
screen is that a larger number of graphical repre-
T Temperature at the moment of the
sentations of evaluation values and measured val-
measurement
ues can be displayed.
p Air pressure at the moment of the
measurement To select the horizontal split click the View button
ExpTime Irradiation time and select the Horizontal entry or select View →
Horizontal.
DoseRate Mean dose rate during irradiation
The advantage of the horizontal split of the main
Field Actual field size - important informa- screen is that the period can be displayed with a
tion in case of filter setting <all> better resolution.
Gantry Actual gantry angle - important infor-
mation in case of filter setting <all> The number of displayed graphical representations
can be adjusted in the Graphics tab via Tools →
MU Number of monitor units
Options… (refer also to section 14.3).
Prot Protocol for calculation of the evalua-
tion values
Device Serial number of the measuring de-
vice
28 D871.131.01/03 en
QUICKCHECK
Editing Measured Values
In the Edit Data Values dialog you can edit the 10.1 Highlighting Data Sets
measured values. To display this dialog select Edit
→ Data Values…. Highlight a data set by clicking the appropriate line.
The selected data set is highlighted blue.
The Edit Data Values dialog contains a table with all
data sets which are within the adjusted period and Select the previous data set with the cursor key ↑
comply with the adjusted filter parameters. Evalua- and the next data set with the cursor key ↓.
tion values which are not within the adjusted toler-
ances are highlighted with a red background. Press the Ctrl key and click the corresponding line
to add single data sets to your selection or press
You can delete complete data sets or edit individual the Shift key and click in the corresponding line to
measured values (values with a white background). select a range of data sets.
The date and the evaluation values (values with
blue background) cannot be changed. Use context menu Deselect All to undo the selec-
tion.
D871.131.01/03 en 29
QUICKCHECK
Editing Measured Values
• Select context menu Delete or press the Del • Enter the desired value and confirm the entry
key. with the Enter key.
Æ After confirming the acknowledgement the data Æ After confirming the acknowledgement the
set will be deleted. measured value will be changed and highlighted
red. Furthermore the evaluation values will be
• Close the Edit Data Values dialog with OK. recalculated and the data set will be marked with
Edited.
Æ The deleted data set will be removed from the
graphical representations. NOTE
• Save the changes (refer to section 8.3). For the calculation formulas for the evaluation
values please refer to the QUICKCHECKwebline
user manual.
30 D871.131.01/03 en
QUICKCHECK
Repeating a Measurement
QUICKCHECK offers the option to delete the most Display of the Print Preview
recently performed measurement in order that it can
be repeated. NOTE
In order to generate a print preview an applica-
NOTE tion which can process PDF files (e.g. Acrobat
A measurement can only be deleted on the day Reader) must be installed on the PC.
of the measurement. You can only delete the
most recently performed measurement.
• To display a print preview of the currently dis-
played data sets select File → Print Preview….
• Select Edit → Advanced….
Æ The print preview will be displayed in the appro-
Æ The Edit Advanced dialog will appear: priate application.
Æ The most recently performed measurement will The print preview resp. the printout contains the
be deleted and can be repeated. data depending on the selected template.
D871.131.01/03 en 31
QUICKCHECK
Exporting Data Sets
32 D871.131.01/03 en
QUICKCHECK
QUICKCHECK Options
14 QUICKCHECK Options
Via Tools → Options… you display the Options In the Device tab the current setting for the data
dialog. transfer from QUICKCHECKwebline is displayed.
In the Options dialog you will find the following 4 ... Button for displaying the Connection Control
tabs: Center dialog to change the settings for the
data transfer (refer to section 15)
− Device: options for the data transfer from and to
the QUICKCHECKwebline
If you have enabled the Search device after pro-
− Data: options for program start gram start option, QUICKCHECK will search auto-
matically for a connected QUICKCHECKwebline (refer
− Graphics: options for the graphical representa-
to section 8.1), when the software is started.
tion of the data sets
− Print: printout options If you have disabled the Search device after pro-
gram start option, you have to set the required con-
nection manually in the Connection Control Center
14.1 Options for the Data dialog. For further information on this refer to sec-
Transfer tion 15.
NOTE
You have to adjust the settings for the data trans-
fer for QUICKCHECK and WorklistGenerator
only in one of the programs. These settings af-
fect always both programs.
D871.131.01/03 en 33
QUICKCHECK
QUICKCHECK Options
14.2 Options for Program Start 14.3 Options for the Graphical
Representations
34 D871.131.01/03 en
QUICKCHECK
QUICKCHECK Options
D871.131.01/03 en 35
QUICKCHECK
Settings for Data Transfer
Data transfer is possible by the following communi- From the Connection Control Center dialog you
cation channels: can:
− LAN Network Connection − create a new connection
36 D871.131.01/03 en
QUICKCHECK
Settings for Data Transfer
15.2.1 Connecting via LAN Network • In the Connection Control Center dialog click the
Connection New… button.
• For this you have to connect the QUICK- Æ The New Connection dialog will appear:
webline
CHECK and the PC to the corporate net-
work via a network cable each.
D871.131.01/03 en 37
QUICKCHECK
Settings for Data Transfer
• In the Connection Control Center dialog select You can search existing LAN connections by
the desired connection. means of the Find function.
• Check the connection by clicking the Check • Connect QUICKCHECKwebline to the PC (refer
button. also to QUICKCHECKwebline user manual).
Æ The connection will be selected and displayed in • In the New Connection dialog or the Edit Con-
the Options dialog on the Device tab in the Con- nection dialog click the Find… button.
nection pane.
Æ The QUICKCHECK Devices dialog will appear:
• Adjust the parameters required for the connec- • If required repeat the search by clicking the Find
tion according to section 15.2.3. button in the QUICKCHECK Devices dialog.
38 D871.131.01/03 en
QUICKCHECK
Settings for Data Transfer
Æ The name in network or the IP address (if a • Connect the QUICKCHECKwebline directly to the
name in network is not available) of the selected PC via a USB cable.
QUICKCHECKwebline will be displayed in the New
Connection dialog or the Edit Connection dialog. The USB cable L178085 is included in the QUICK-
CHECKwebline delivery.
2. In case of a LAN direct connection the IP ad- The installation of the USB driver will be per-
dresses of QUICKCHECKwebline and PC are not formed automatically when you connect a
in the same subnet (e.g. in the range QUICKCHECKwebline to the PC for the first time.
169.254.x.x). The devices will be distinguished by means of
You have to configure the network interface card the serial number. Each QUICKCHECKwebline will
of the PC manually. thus be assigned to its own virtual COM port.
For further information please contact the PTW • Connect QUICKCHECKwebline to the PC (refer
support. also to QUICKCHECKwebline user manual).
• Switch on QUICKCHECKwebline.
15.2.8 Deleting a LAN Connection
Æ The installation of the USB driver will be started.
• In the Connection Control Center dialog click the
connection to delete. • To perform the installation follow the prompts on
the screen.
• Click the Delete button.
• In QUICKCHECK in the Connection Control
Æ After confirming the acknowledgement the con- Center dialog click the New… button.
nection will be deleted.
D871.131.01/03 en 39
QUICKCHECK
Settings for Data Transfer
Æ The New Connection dialog will appear: 15.3.3 Selecting an USB Connection
Figure 29: New Connection dialog for USB connec- Æ The connection will be selected and displayed in
tion the Options dialog on the Device tab in the Con-
nection pane.
• Adjust the parameters required for the connec-
tion:
15.3.4 Editing an USB Connection
Type: USB
Name: user-defined name of connection • In the Connection Control Center dialog click the
connection to edit.
COM: used virtual COM port
• Click the Edit… button.
You can also adjust the connection by means of
the Find… button (refer to section 15.3.5). Æ The Edit Connection dialog will appear:
NOTE
If no QUICKCHECKwebline is connected via an
USB connection, the COM list box is empty.
40 D871.131.01/03 en
QUICKCHECK
Settings for Data Transfer
You can search existing USB connections by If communication problems occur, please contact
means of the Find function. the PTW support.
Æ The QUICKCHECK Devices dialog will appear: Æ After confirming the acknowledgement the con-
nection will be deleted.
D871.131.01/03 en 41
QUICKCHECK
Settings for Data Transfer
15.4.1 Connecting via RS232 Æ In the Connection Control Center dialog the new
connection will be displayed.
• Connect the QUICKCHECKwebline directly to the
PC via a RS232 cable.
15.4.3 Selecting a RS232 Connection
The RS232 cable L178012 is included in the
QUICKCHECKwebline delivery. • In the Connection Control Center dialog select
the desired connection.
• Switch on QUICKCHECK webline
.
• Check the connection by clicking the Check
button.
15.4.2 Creating a New RS232 Connection
• Confirm with the OK button.
• In the Connection Control Center dialog click the
New… button. Æ The connection will be selected and displayed in
the Options dialog on the Device tab in the Con-
Æ The New Connection dialog will appear: nection pane.
Figure 32: New Connection dialog for RS232 con- Æ The Edit Connection dialog will appear:
nection
Type: RS232
Name: user-defined name of connection
COM: used COM port
Baudrate: desired baud rate Figure 33: Edit Connection dialog for RS232 con-
nection
The baud rate in QUICKCHECK has to match
the baud rate set in QUICKCHECKwebline. • Adjust the parameters required for the connec-
Default: 57600 tion according to section 15.4.2.
You can also adjust the COM port and the baud • Confirm the settings with OK.
rate by means of the Find… button (refer to sec-
tion 15.4.5). Æ In the Connection Control Center dialog the
changed connection will be displayed.
42 D871.131.01/03 en
QUICKCHECK
Settings for Data Transfer
D871.131.01/03 en 43
QUICKCHECK
Troubleshooting
16 Troubleshooting
… the display of Check in the control panel the settings for the display:
QUICKCHECK shows
artifacts (e.g. text color and 1. Windows XP:
background color are
• Select Start → Settings → Control Panel → Display.
similar or identical)?
• Select the Appearance tab.
• Select one of the following color schemes. These schemes are recom-
mended by PTW:
– Windows Classic
– Windows Standard
3. Windows 7:
• Select Start → Control Panel → Appearance and Personalization → Per-
sonalization.
• Select one of the following color schemes. These schemes are recom-
mended by PTW:
– Windows 7 Basic
– Windows Classic
44 D871.131.01/03 en
QUICKCHECK
Literature
Literature
D871.131.01/03 en 45
QUICKCHECK
Index
Index
A L
Analysis, representation.................................................24 LAN connection
creating...................................................................... 37
C deleting...................................................................... 39
editing........................................................................ 38
Connection Control Center dialog..................................36 searching................................................................... 38
Context menu ................................................................18 selecting .................................................................... 38
License Control Center.................................................. 14
Licensing ....................................................................... 14
D Literature ....................................................................... 45
Data backup...................................................................12
Data set .........................................................................16 M
comment ....................................................................28
deleting ......................................................................30 Main screen................................................................... 17
highlighting.................................................................29 Measured value, changing ............................................ 30
measured values........................................................27 Measured values, editing .............................................. 29
parameters.................................................................28 Measurement, repeating ............................................... 31
selecting.....................................................................27
Data sets O
displayed....................................................................16
exporting ....................................................................32 Options of QUICKCHECK ............................................. 33
graphical representation ............................................24
saving ........................................................................20
Data storage ..................................................................12
P
Data transfer Period, setting ............................................................... 22
from QUICKCHECKwebline.......................................20 Printing .......................................................................... 32
settings ......................................................................36 Printing, options............................................................. 35
Data, closing ..................................................................21 Program start, options ................................................... 34
Data, loading..................................................................20
saved data sets..........................................................21
Data, merging ................................................................21
Q
Download.......................................................................20 Quick start ..................................................................... 16
E R
Exiting QUICKCHECK ...................................................13 Registering .................................................................... 15
Repairing QUICKCHECK .............................................. 11
F Restoring a backup ....................................................... 12
RS232 connection
Filter parameters, selecting............................................23 creating...................................................................... 42
deleting...................................................................... 43
G editing........................................................................ 42
searching................................................................... 43
General information .........................................................2 selecting .................................................................... 42
Graphical Representations, options ...............................34
S
I
Safety information ........................................................... 6
Icons, explanation..........................................................18 Starting QUICKCHECK ................................................. 13
Illustrations.......................................................................5
Installation of QUICKCHECK...........................................9
Intended use ....................................................................6
46 D871.131.01/03 en
QUICKCHECK
Index
T
Target directory ............................................................. 10
Terms, definitions.......................................................... 16
Troubleshooting ............................................................ 44
LAN connection......................................................... 39
RS232 connection ..................................................... 43
USB connection ........................................................ 41
U
Uninstalling QUICKCHECK........................................... 11
USB connection
creating ..................................................................... 39
deleting...................................................................... 41
editing........................................................................ 40
searching................................................................... 41
selecting .................................................................... 40
W
Workflow ....................................................................... 16
WorklistGenerator ......................................................... 16
D871.131.01/03 en 47