E.M.F. K Thermocouples: Drift and Ofqpe
E.M.F. K Thermocouples: Drift and Ofqpe
Abstract: Type K thennocouples were exposed at some temperatures in order to estimate their E.M.F.
drift and inhomogeneity. About 3'C E.M.F. drift was observed at 7OO0Cexposure. And over 10°C E.M.F.
change caused by inhomogeneity was also observed at XOO°C exposure. It wiU be dangerous for type K
thermocouples to calibrate at around 800OC
Keywords: Type K, Thermocouple, Drift, Inhomogeneity.
1. Introduction
PIatindaUadium thermocouple has already been exam-
ined and reported by Ogra et 01'. 2). It is applied for the
transfer standard at Silver and Cupper fixed points on JCSS '.
Qpe R themocoupIe, which is apply for working stan-
dard, has also been reported of its E.M.F. drift and
inhomogeneity3-4, ').
However little reports of simiIar E.M.F. drift and inhomo-
geneity are seen of type K thermocouple@ in spite of the
grate use in not only laboratory but also industry.
So, we are going to report the E.M.F. drift and inhomo-
geneity of type K thermocouple after exposed at high e m -
peratures.
JIS C 1602-*9g5says that 65OoCis an upper limit for con-
tinuous measurement and 850T is another limit for short
t:
time measurement €or type K thermocouple with 4 0.65mm O O 20 40
wires. So, calibration often requires up to 85OoC. Distance from the hot juction, Ucm
Following the above, type K thermocouples were exposed
at 70OoC and 800T in order to evaluate their E.M.F. drift
and inhomogeneity in the experiment.
Figure 1: Temperature distributions of the heating fumace
and measuring furnace.
2. E.M.F.drift
There will be two method to measure the E.M.F. drift of ther- In this case, temperature distribution is changed be-
mocouples. Details are as follows. tween two furnaces, so that the measured E.M.F. con-
tains the effect of inhomogeneity. However it is possi-
1. in situ observation" ble to measure at fixed points, small change of E.M.F.
The test is performed in a furnace. The test thermocou- is detected,
ples keep being in the fumace and a new thermocou-
ple is installed into the furnace when the temperature As each method has merit and demerit, present experiment
and/or E.M.F. is measured. has performed with Two furuace method. From the view
point if the calibrated results is believed by users, how much
This method has a merit that there is no influence difference will be observed caused by inhomogeneity which
caused by inhomogeneity. is born the difference between the temperatwe distribution
of the furnaces among the calibration laboratory and users
2. N o fumace methods
one.
' h o furnaces are used for the test. The first one is for
only expose at high temperatures. When measure the
E.M.F. drift andlor inhomogeneity, thermocouples are 2.1 Experimental procedure
put into another furnace. 1. Q p e K thermocouple wires are installed into two bom
'Japan Calibration Service System insulators with 6 3" out diameter, 4 0.8" holes and
I 1
800 -
0 .
2
-
0 200 400 600
Temperature, T/"C
-990-
50 100 150 200
time, Vh
2. Temperature distribution of the furnace is measured Figure 4 shows the E.M.F. change, calculated to tempera-
with a type R thermocouple. (see Fig. 3) ture, from the reference standard type R thermocouple .
At the beginning of the test, the type K thermocouples
3. Reference standard Type R thermocouple with only a h - show little difference between center position and puiled po-
mina insulator and type K thermocouples with not only sition. However, the longer exposed time bring larger EMF.
alumina insulator but also in silica tube a e wounded change especially at 50cm and 57.5cm positions.
with platinum wire caring hot junction being same At last after 200 hours passed, the E.M.F. change caused
depth. by inhomogeneity over 10°C at 57.5cm position !!
4. Put the thermocouples at 35cm position in fig.3. Generally speaking, temperature stability and distribution
of calibration furnace should be good as well as possible. As
5 . E.M.F. o f each thermocouple is measured at this posi- the result, its immersion depth becomes larger.
tion. On the other hand, furnaces in laboratories andor facto-
ries do not always have long immersion depth. So that the
6. Pull the thermocouples to the 42.5cm position in fig.3 similar situation will occur, i.e. the thermocouple is cali-
and E.M.F. of each thermocouple is measured. brated with long immersion depth in a calibration laboratory
7. Pull them 50cm and measure. Then Pull them 57.5cm but it is set with short immersion depth in a factory.
and measure. In this case, the calibration makes no sense.
-991-
We would like to thank them and Professor Gotoh, Pro-
fessor of Tamagawa University, allowing the presentationof
their measurement.
References
[I] H. OGURA, 11. NIMAJIRI, K.YAMAZAWA, J . TAMBA,
M.IZUCH and M. ARAI : l.hmrtainty Evaluation on
the Freezing Point of Copper for Calibration of Themo-
couples, Transactions of SICE, Vol. 39, No. ll(2003)
[2] H. OGURA, H. NUMAJIRI,H. NARUSHIMA, M.
IZUCHI and M. ARAI: Emf Changes of Pt/Pd Ther-
mocouples in the Range from 420'C to 1080"C, SICE
in Fukui 2003,2692f2695
[3] Kazuaki KODAIRA and Tokio HAMADA : An Estima-
tion of Uncertainty at At Fixed Point, Journal of SICE,
Vol. 42 No. 1 1 9121917 (2003)
[4] TOKIO HAMADA and YUKIHARU SUYAMA : A re-
liability test of type R thermocouples as a reference stan-
dard, SICE 2003 in Fukui, L50/155
[5] T. HAMADA, S. HITOMI and H. YMASAKI : Long-term
Stability Test of Type R Thermocauples, Transactions of
SICE, Vol. 31., No. 4,532/534(1995)
[6] M. TOYODA,Y. TAMURA and M . OGAWA : Practi-
c a l thermocouples @Instrumentation
I, and Automation
'95.5 (1995)
-992-