20200605-P117 Portable Ultrasonic Flowmeter
20200605-P117 Portable Ultrasonic Flowmeter
20200605-P117 Portable Ultrasonic Flowmeter
FUNCTION SPECIFICATIONS
Outputs Analog output: 4~20mA, Max 750 Ω.
Storage: 8GB;
SD card Max: 512 files;
Interval: 1 ~ 60 seconds.
rechargeable Lithium Battery Power
Power supply
( continuous operation of mainbattery 10 hours ).
Keypad Tactile Keys.
3.5 inch TFT screen( 320 × 240 ),
Display
backlit LCD.
Transmitter:14℉~122℉(-10℃~50℃)
Temperature
Transducer:40℉~176℉(-40℃~80℃)
Humidity 0 to 99% RH,non-condensing
PHYSICAL SPECIFICATIONS
Transmitter NEMA13 ( IP54 ).
Encapsulated design, IP68;
Transducer
Standard cable length: 5m.
Weight Transmitter:approximately1.0kg.
Suitcase Transmitter and Pipe strips Coupling compound Card reader and
Transducer SD card
ABOUT P117
INTERFACE AND SIZE
WIRING DIAGRAM
TRANSMITTER DIMENSIONS
111mm 39mm
P117
Ultrasonic Flowmeter
209mm
4 5 6
hold
7
DYN.
8
sound
speed
9 Enter
time CAL. zero <
. 0
<
TRANSDUCER
40mm
30mm
300mm
ABOUT P117
TRANSDUCER INSTALLATION
METHODS
V method measuring pipe size : 50mm-400mm
ENCLOSURE
FLOW CERTIFICATION
ANBIENT :IP68
TEMPERATURE
:Ex
ia
Part II BT6
(MAX):80
℃
Ultrason No. CP036
ic Transdu
cer
cer
ic Transdu
Ultrason No. CP036 ℃
(MAX):80
Part ia
II BT6
:Ex
TEMPERATURE
ANBIENT :IP68
CERTIFICATION
ENCLOSURE
FLOW
Top View
Flow
Transducer space
cer
ic Transdu
Ultrason No. CP036 ℃
(MAX):80
Part ia
II BT6
:Ex
TEMPERATURE
ANBIENT :IP68
CERTIFICATION
ENCLOSURE
FLOW
w
Flo Downstream Transducer
Pipe Strap
Top View
Flow
Transducer space
ABOUT P117
INSTALLATION SITE
SELECTION
STRAIGHT LENGTH OF STRAIGHT LENGTH OF
UPSTREAM PIPING DOWNSTREAM PIPING
When selecting a measurement 90° Bend
site, it is important to select an
area where the fluid flow profile L≥5D
is fully developed to guarantee a recudsnarT cin
osartlU
WOLF
L≥10 D
FLOW
Part No.CP036
Ultrasonic Transducer
Part No.CP036
ANBIENT TEMPERATURE(MAX):80
CERTIFICATION ℃
:Ex ia II BT6
ENCLOSURE
:IP68
FLOW
FLOW
:IP68
ENCLOSURE
:Ex ia II BT6 ℃
CERTIFICATION
ANBIENT TEMPERATURE(MAX):80
Part No.CP036
Ultrasonic Transducer
Tee
Choose a section of pipe that is
always full of liquid, such as a L≥1 0 D
vertical pipe with flow in the L≥50 D
upward direction or a full ENCLOSURE
:IP68
CERTIFICATION
FLOW
:Ex ia II BT6
ANBIENT TEMPERATURE(MAX):80
Part No.CP036
Ultrasonic Transducer
℃
recudsnarT c
630PC.oN tra
℃ 08:)XAM
WOLF
inosartlU
(ERUTAREPM
6TB II ai xE
P
ET TNEIBNA
: NOITACIFI
86PI : ERUSO
TREC
LCNE
horizontal pipe.
Ultrasonic Transducer
Part No.CP036
ANBIENT TEMPERATURE(MAX):80
CERTIFICATION :Ex ℃
ia II BT6
ENCLOSURE :IP68
FLOW FLOW
ENCLOSURE :IP68
ia II BT6
CERTIFICATION :Ex ℃
ANBIENT TEMPERATURE(MAX):80
Part No.CP036
Ultrasonic Transducer
630PC.oN traP
TAREPMET TNEIBN
A
℃ 08:)XAM(ERU ITACIFITREC
6TB II ai xE : NO
86PI : ERUSOLCN
E
Ultrasonic Transducer
FLOW WOLF
Part No.CP036
ANBIENT TEMPERATURE(MAX):80
℃ ENCLOSURE :IP68
CERTIFICATION :Ex ia ia II BT6
II BT6 CERTIFICATION :Ex ℃
ENCLOSURE :IP68
ANBIENT TEMPERATURE(MAX):80
FLOW Part No.CP036
FLOW
ENCLOSURE :IP68
CERTIFICATION :Ex ia
II BT6 Ultrasonic Transducer
℃
ANBIENT TEMPERATURE(MAX):80
Part No.CP036
Ultrasonic Transducer
FLOW FLOW
FLOW
ENCLOSURE :IP68
ia II BT6
CERTIFICATION :Ex ℃
ANBIENT TEMPERATURE(MAX):80
:IP68
Part No.CP036 ENCLOSURE
:Ex ia II BT6 ℃
CERTIFICATION
ANBIENT TEMPERATURE(MAX):80
Part No.CP036
Ultrasonic Transducer
Ultrasonic Transducer
TEMPERATURE(MAX):80
Part No.CP036
II BT6
℃
L≥1 0 D
Ultrasonic
Transducer
recud
snarT
cinosa
rtlU
630PC
.oN tr
℃ 0
8:)XAM aP
(E
6TB I RUTAREPM
I ai x E
E : NO T TNEIBNA
86PI : TACIFITREC
I
ERUSO
LCNE
WOLF
Ultrasonic
Transducer
Part No.CP036
ANBIENT
TEMPERATURE(MAX):80
CERTIFICATION
ENCLOSURE :Ex ia
II BT6 ℃
:IP68
FLOW
FLOW
:IP68 II BT6
ENCLOSURE :Ex ia ℃
CERTIFICATION
TEMPERATURE(MAX):80
ANBIENT
Part No.CP036
Transducer
Ultrasonic
Vertical
L≥5D
ENCLOSURE :IP68
CERTIFICATION :Ex ia II BT6
ANBIENT TEMPERATURE(MAX):80 ℃
Ultrasonic Transducer
Part No.CP036
FLOW
Ultrasonic Transducer
ANBIENT TEMPERATURE(MAX):80 ℃
FLOW
Part No.CP036
CERTIFICATION :Ex ia II BT6
ENCLOSURE :IP68
L≥5 0 D
L≥10D
Ultrasonic
Transducer
Part No.CP036
ANBIENT
TEMPERATURE(MAX):80
CERTIFICATION
ENCLOSURE :Ex ia II
BT6
℃
:IP68
FLOW
FLOW
:IP68 BT6
ENCLOSURE :Ex ia II ℃
CERTIFICATION
TEMPERATURE(MAX):80
ANBIENT
Part No.CP036
Transducer
Ultrasonic
ABOUT P117
ORDERING INFORMATION
MODEL DESCRIPTION
Portable Ultrasonic Flowmeter
Installation method: Handheld
8G SD card high memory data logging,
maximum memorize 512 days data.
Flow Range: ±0.03 ft/s ~ ±20 ft/s (±0.01 m/s~ ±6 m/s)
P117 Accuracy: ±1%
Repeatability: 0.3%
Output: 4-20mA
Internal lithium power supply: 10hours
Pipe size range: 1ʺ~48ʺ(25mm~1200mm)
Transducer: IP54, CP magnet portable transducer, 5m cable
Version:P117-ZOZO-OA-B