Syringe Pump Operating Manual: SP-8800 / UME-03-REV.8
Syringe Pump Operating Manual: SP-8800 / UME-03-REV.8
Syringe Pump Operating Manual: SP-8800 / UME-03-REV.8
0 SP-8800 / UME-03-REV.8
Thank you for purchasing our SP-8800 Syringe Pump.
In order to use this pump correctly and safely, read this manual carefully before operating SP-8800 Syringe Pump.
If you have any questions as you are reading through this manual, call the local authorized dealer in your country.
Retain this manual together with the unit for future reference.
This SP-8800 Syringe Pump is intended for the infusion of chemicals such as vasopressor drugs, depressor
drugs, anti-coagulants, anesthetic drugs, anti-cancer drugs, oxytocic, nutrition, drug for chemotherapy medication
and blood infusion at ICU, CCU, NICU, or operation room.
This device is designed for high flow-rate accuracy and ease of handle in the infusion of solutions with the
equipped audio-visual alarms and state-of-the-art sensors.
Features
Self testing – Every time the unit is turned on, self testing is proceeded.
Automatic syringe detection – Detect syringe size automatically. (10, 20, 30, 50ml )
Shortcut key – Syringe brand set, Occlusion set, Buzzer level set
K.V.O.(Keep Vein Open) – When the delivery volume has been reached, flow rate turns into K.V.O.
rate (adjustable from 0.1ml/h to 10ml/h) automatically.
Keypad lock function – Keypad is locked with or without password depends on its using
condition.
Infusion setup – Flow rate/ Delivery volume/ Infusion time
Retain memory – Last setting value is retained.
Infusion remaining time display
Alarm repeat function
Open system – Calibration up to 10 syringe brands in a single unit.
Buzzer level – 3steps
Occlusion level – 9steps (3.8psi~19.3psi)
Purge rate – Adjustable (0.1~1200ml/h)
Bolus rate – ON/OFF, Adjustable (0.1~1200ml/h, 0.1~1000ml)
Anti bolus – Reduce significantly bolus after occlusion release. (ON/OFF)
History call back – Call back last 10 infusion data.
Dosage mode(Body weight mode) – Automatically calculate and set a proper flow rate when dose
rate, body weight, drug mass and solution volume are entered.
Unit setup for Dosage mode – ml/h, mg/kg/h, mg/kg/min, ㎍/ kg/h, ㎍/ kg/min
Nurse call – Each unit can be connected with nurse call system (DC 24V, 0.5A) (Option)
Profile function – Program infusion condition (flow rate, delivery volume, delivery time)
differently for every hour, up to 24hours. (Option)
Central system – Connect with main PC for monitoring (Option)
0 SP-8800 / UME-03-REV.8
INTRODUCTION / FEATURES ············································· 1
1-5.Components ······························································································ 8
2-3.Precautions ································································································ 11
2-5.Storage ······································································································ 12
3. OPERATION ······································································ 14
3-1.Initial setup ································································································· 15
1 SP-8800 / UME-03-REV.8
CONTENT
INTRODUCTION / FEATURES
NTS
4. SYSTEM SETUP ······························································· 24
4-1.Buzzer level setup (Short-cut key) ···························································· 25
4-5.Display ······································································································· 30
4-6.Config ········································································································ 31
5-2.History ········································································································ 35
5-3.Profile (Option spec) ·················································································· 36
SPECIFICATIONS ································································· 42
SYMBOLS ············································································· 43
WARRANTY ·········································································· 44~45
2 SP-8800 / UME-03-REV.8
1. DESCRIPTION
OF PUMP
1-1.Front View
1-2.Rear View
1-3.Side View
1-4.Syringe View
1-5.Components
3 SP-8800 / UME-03-REV.8
1-1. FRONT VIEW
4 SP-8800 / UME-03-REV.8
1. DESCRIPTION OF PUMP
제품 설명
6 SP-8800 / UME-03-REV.8
1. DESCRIPTION OF PUMP
제품 설명
① Coupler : Joins two syringe pumps when one syringe pump is laid on another syringe pump
1-4. SYRINGE
Barrel
Finger-grips
Plunger
1-5. COMPONENT
AC power cord
7 SP-8800 / UME-03-REV.8
1. DESCRIPTION OF PUMP
제품 설명
2. PRIOR TO
PUMP USE
2-1.Explanation of Symbols
2-2.Warnings
2-3.Precautions
2-4.Cleaning & Sanitation
2-5.Storage
2-6.Maintenance & Repair
8 SP-8800 / UME-03-REV.8
2. PRIOR TO PUMP USE
안전 및 올바른 사용 정보
Caution is used to indicate the presence of a hazard which will cause minor personal
injury or property damage if the warning is ignored.
NOTE Note is used to notify the user of installation, operation or maintenance information
which is important but not hazard-related.
2-2. WARNINGS
• If used in the vicinity of the surgical operation equipment which generates a high frequency current such as
mobile (cellular) phone, radio, or defibrillator, the pump may malfunction since electrical interference.
Please carefully check for any sources of electrical interference in the vicinity before use.
• Before using the pump, make sure that the manufacturer of the syringe to be used is the same as the one
specified for this pump. If it is not matched, the flow rate accuracy and alarm system cannot be guaranteed.
In order to change the syringe manufacturer, please contact your local authorized dealer.
• When using the pump concurrently with the surgical operation equipment, please note the following:
- Do not use the pump together with any surgical operation equipment that generates high noise level.
- Be sure that the pump is kept a sufficient distance from the surgical operation equipment.
- The pump and such device should not be powered from the same outlet.
- Check and confirm the normal operation of the pump periodically.
※ In case of malfunction, immediately turn off the power and remove the infusion set from the patient’s
skin. After this action, please contact your local authorized dealer.
• Avoid using the pump in presence of flammable gases and flammable anaesthetic mixture with air, oxygen or
nitrousoxide.
• The use of any mobile (cellular) phone near the pump is not allowed since the high frequency noise during the
conversation could cause malfunction of the pump.
• The use of the pump in MRI rooms such as high-pressure rooms or places where high electromagnetic radiation
is generated is not allowed.
• In case of using new brand of syringe other than suggested by AMPall, new brand of syringe must be calibrated
by AMPall or local authorization. If not, new syringe may cuase problem on infusion accuracy and alarm state.
※ Please contact your local authorized dealer if you wish to use syringe that are not supported.
• This pump does not detect ‘over-infusion’ or ‘under-infusion’. Please check infusion state with patient regularly.
• Make sure about correct syringe installation, drug may flow backward if syringe is installed incorrect way.
• Eliminating a needle from the patient to prevent a release-bolus is required when the tube line got an occulsion.
9 SP-8800 / UME-03-REV.8
2. PRIOR TO PUMP USE
• The pump does not detect damage to the infusion tube such as a leak in the tube or a rupture in the filter due to
안전 및 올바른 사용 정보
pressure exertion. Therefore, regularly check for any damage to the infusion line during infusion.
• When the flow is obstructed due to kinking of the tube or clogging of the needle or filter, it can increase pressure
in the infusion tube and cause the tube to be inflated with the solution. Complete removal of the obstruction will
allow the solution to be delivered to the patient. If the flow is obstructed, take appropriate actions.
• The syringe is disposable. Therefore, the used syringe should not be reused.
• Never modify the pump by yourself.
• The pump is connected to the AC power outlet to be operated. If there is no available AC power outlet, pump
can be operated with only its internal battery. (Alternatively, the pump may be operated by DC power).
• The spill of the solution on the AC power inlet may cause a short circuit.
• Make sure that there is any damage with the pump and components. In case of the unit and components that
were shocked, do not use them even if visible damages are not observed. Please contact your local authorized
dealer.
• In case of malfunction, do not try to take the unit apart or attempt to repair by yourself. Please contact your local
authorized dealer immediately. If the user does not comply with these warnings, AMPall does not take liable and
the warranty does not apply.
2-3. PRECAUTIONS
• The pump does not detect about correct infusion state with blood vessel on patient. Please check the puncture
site and monitor the patient’s condition carefully.
• Fix the pump securely to a pole stand and check its stability. If the pump is dropped or given a hard shock, do
not use it. Please contact your local authorized dealer.
• The pump must be used in accordance with this instruction manual by trained medical personnel.
• When alarm sounds, please take corrective actions.
• Do not apply a strong force or damage to the clamp or slider.
If the user does not comply with these warnings, the company is not
responsible for its result and the warranty is not applied.
10 SP-8800 / UME-03-REV.8
2. PRIOR TO PUMP USE
안전 및 올바른 사용 정보
< Casing >
• When the pump casing is stained, use a gauze cloth or similar, moistened with cold or tepid water and
let it dry, especially the AC power inlet before use. Clean AC power inlet, Adoptor connectorm Slider Slider hook,
Clamp, Slit and Clutch regularly.
• In case of disassemble Clutch or Clamp for cleaning, contact local authorization.
• Do not use a drier to dry casing parts.
2-5. STROAGE
• Avoid the following environment for storage and transport for SP-8800 Syringe Pump.
- Where the unit is exposed to dirt or heavy dust.
- Where the unit is exposed to salty atmosphere.
- Where the unit is exposed to severe vibration or corrosive GAS.
- Where the unit is exposed to rough handling.
- Where the unit is exposed to direct sunlight or UV light.
- Where the unit is exposed to water.
- Where the unit is exposed to extreme temperature and humidity.
• The optimum storage conditions for the unit
11 SP-8800 / UME-03-REV.8
2. PRIOR TO PUMP USE
안전 및 올바른 사용 정보
the product and its electronic accessories (e.g. Power cord) should not be disposed of
with other household waste at the end of their working life. To prevent possible harm
to the environment or human health from uncontrolled waste disposal, please
separate these items from other types of waste and recycle them responsibly to
promote the sustainable reuse of material resources.
12 SP-8800 / UME-03-REV.8
3. OPERATION
3-1.Initial setup
3-2.Syringe instollation
3-3.Setup flow rate (ml/h)
3-4.Setup delivery volume (ml)
3-5.Confirm total infused volume (Σ㎖/CLEAR)
3-6.Setup delivery time (hours)
3-7.Insert the needle into the patient
3-8.Start infusion
3-9.Infusion completion
3-10.Stop infusion
3-11.[POWER] button
3-12.Battery remain
3-13.Error alarm
3-14.Adoptor operation
3-15.Battery operation
3-16.Nurse call function (Option)
13 SP-8800 / UME-03-REV.8
3-17.Keypad lock
- LCD displays on the left side of LCD when it connected with AC power.
- LCD displays connected with adoptor, and connected with battery for power supply.
NOTE • Please contact a local authorized dealer or agency when you have problem with the
14 SP-8800 / UME-03-REV.8
unit.
3. OPERATION
NOTE • If the syringe is not properly placed in the slit, the flow rate accuracy and alarm
system are not guaranteed.
5. After infusion tube is placed in Line Guide, push the slider slowly until it hits push-button of the syringe.
When the slider hits the push-button, slider hook automatically gets down to fix the syringe.
When the syringe is set in the pump, [SYRIGNE SIZE] LED lights on according to the size of syringe.
6. Press and hold [PURGE] button until a drop of solution at the end of the infusion set appears to remove
air bubble in the infusion tube.
State of menu
2. Input desired value for FLOW RATE(ml/h) with numerical buttons and press [SEL] button to save it.
Display turns into previous display automatically.
3. Press [ESC/2s CLR] button for 2seconds to delete complete value, press button to delete previous
digit only.
4. When setup value exceeds the range, “RANGE OVER” will come up and maximum value is saved
automatically. Make sure setup value before start.
NOTE • Please do not change the flow rate during the infusion.
To change the flow rate, the infusion should be stopped by pressing 'Stop' button first.
• Druing flow rate setup, if press [STOP] button for more than 2seconds, the value will be
reset as “0000”.
• The pump does not start when the flow rate is setup at 0 ml/h. The message “NO SET”
will come up on the LCD.
16 SP-8800 / UME-03-REV.8
3. OPERATION
State of menu
Previous setup value
Delivery volume(ml)
3. Press [ESC/2s CLR] button for 2seconds to delete complete value, press button to delete previous
digit only.
NOTE • If delivery volume is set with 0ml, it is considered as maximum value, 1000ml and
operates delivery volume with 1000ml.
(When remaining time is exceeded 99hours, infusion remaining time is not
displayed.)
• During infusion state, delivery setup is not available. Setup is available only “STAND
BY” mode.
State of menu
Previous value
17 SP-8800 / UME-03-REV.8
3. OPERATION
NOTE • Third value is calculated and displayed automatically when two setup values are
entered among FLOW RATE, DELIVERY VOLUME, DELIVERY TIME.
• Please make sure about total infused volume when you setup infusion by time,
previous total infused volume is included if it is not deleted.
• DELIVERY TIME range : 1min~99hour 59min
• Alarm does not go off when needle comes off the vein, resulting in an out-of-the-vein
infusion. Therefore, regularly check the puncture site and monitor the patient’s
condition carefully.
NOTE • Before operating the pump, make sure to check the flow rate, delivery volume and
the condition of infusion tube again.
[INFUSING DISPLAY]
18 SP-8800 / UME-03-REV.8
3. OPERATION
[K.V.O. RATE]
3. Press [STOP] button to stop K.V.O. rate. LCD display turns into initial display.
• K.O.V. rate is adjustable, please make sure K.V.O. rate before start infusion.
Otherwise over infusion may be cuased.
• Over infusion may be cuased by user’s setup condition, manufacturer does not take
any responsibility on the problem.
4. When drug in the syringe is left about 3~4cc, “NEAR EMPTY” alarm goes off with alarm sound
depends on setup flow rate. OCC alarm indicator is blinked in red.
[NEAR EMPTY]
5. When the syringe is empty, “ WARNING SYRINGE OCCLUSION” alarm goes off with alarm sound and
OCC alarm indicator is on red.
[SYRINGE OCCLUSION]
6. Press [STOP] button to stop. LCD display turns into initial display.
19 SP-8800 / UME-03-REV.8
3. OPERATION
NOTE • Alam goes off if do not restart in 2minutes after stop infusion temporally.
NOTE • Reset Total volume as “0” when restart infusion with new infusion condition.
Remainning battery
operating time
• Remainning battery operating time is different as its infusion setup condition and
battery condition.
• Check about battery condition and operating condition with you refer to “BATTERY
REMAIN”.
• Charging LED on the front should be green when you refer to “BATTERY REMAIN”.
20 SP-8800 / UME-03-REV.8
3. OPERATION
작동 순서
NOTE • When alarm goes off, error message shows up in LCD, alarm signals are transferred
to NURSE CALL SYSTEM every 5 second.
• Nurse call code number: 06LA548﹡
21 SP-8800 / UME-03-REV.8
3. OPERATION
작동 순서
22 SP-8800 / UME-03-REV.8
4. SYSTEM SETUP
23 SP-8800 / UME-03-REV.8
4. SYSTEM SETUP
작동 순서
4. SYSTEM SETUP
• Press MENU button to get into Menu mode and select “SYSTEM SETUP” with [SEL] button.
• SYSTEM SETUP MENU
1. BUZZER
2. OCCLUSION
3. BRAND SET
4. INFUSION tks
Menu move buttons Total menu number
5. DISPLAY
6. CONFIG
• It you do not press any button for 10seconds, LCD turns back to previous display with alarm sound.
• Press [ESC / 2sCLR] button to go back previous display. And press [SEL] button to save.
•. Adjust buzzer level with UP , DOWN buttons. There are three levels, “HIGH” , “MIDDLE” and “LOW”.
Select and save with [SEL] button. After save buzzer level display turns into previous mode.
※ Short-cut key for buzzer level : Press + [1] button in initial display.
24 SP-8800 / UME-03-REV.8
4. SYSTEM SETUP
• OCCLUSION is adjusted with UP , DOWN buttons, it is from LEVEL 1 to LEVEL 9. Press [SEL] button
to select and save. After select and save it, it turns into previous menu.
※ Short-cut key for occlusion level : Press + [2] button in initial display.
• Syringe should be calibrated by supplier, select the syringe brand will use with UP , DOWN button.
Select and save the syringe brand with [SEL] button. After select and save it, it turns into previous menu.
25 SP-8800 / UME-03-REV.8
4. SYSTEM SETUP
SPECIAL FUNCTION
• Here is the list of basic calibration syringe brands.
(Calibrated syringe brand may be different depends on local authorizaed dealer.)
NO Brand Manufacturer
1 B-D Backton
2 MEDI Medihut
3 KOVA Korean Vaccine
4 SHIN Shinchang
5 HWAJ Hwajin
6 SUNG Sungshin
※ Short-cut key for syringe brand : Press + [3] button in initial display.
26 SP-8800 / UME-03-REV.8
4. SYSTEM SETUP
SPECIAL FUNCTION
4-4-1. PURGE SET UP
• PURGE is setup from 0.1~1200ml/h.
Previous setup
PURGE (㎖/h)
Range
• If PURGE rate is not setup, beep sould will come out and infusion will not start.
• When Purge rate is less then floow rate, Purge rate works as setup flow rate.
- BOLUS FLOWRATE SET : Setup flow rate of BOLUS with numerial buttons. Save desired value with [SEL]
button. Press [ESC / 2sCLR] buttons for 2seconds to delete the total value and press to delete previous
digit in order.
※ Press [ESC / 2sCLR] button to move previous menu.
Previous setup
27 SP-8800 / UME-03-REV.8
4. SYSTEM SETUP
SPECIAL FUNCTION
- BOLUS VOLUME : Setup total amount of BOLUS. Save desired value with [SEL] button. Press [ESC /
2sCLR] buttons for 2seconds to delete the total value and press to delete previous digit in order.
• Make sure K.V.O rate setup before use it to avoid excess infusion during K.V.O.
function.
• Manufacturer does not cover the result from K.V.O. setup failure by user. So please
make sure about K.V.O. setup condition before start infusion.
28 SP-8800 / UME-03-REV.8
4. SYSTEM SETUP
• Move with , , , buttons and input with numerical buttons for time setup. Press [SEL] button to
select and save time set.
• There is no extra delet button because you can move with , , , buttons at any point.
NOTE • Once time setup, the unit keeps time whenevery it turns on or off.
Move with , buttons and press [SEL] button to select and save the mode. After select and save it, it
turns into previous menu.
29 SP-8800 / UME-03-REV.8
4. SYSTEM SETUP
4-5-3. LCD BLACK LIGHT BRIGHT SPECIAL FUNCTION
• Get into “BLIGHT BRIGHT” menu in order to MENUSYSTEM SETUPDISPLAYB.LIGHT BRIGHT.
• There are brightness levels from LEVEL 1 to LEVEL 9. As level number is up, LCD goes brighter. Adjust it with
4-6. CONFIG
There are “KEY LOCK P/W” , “BOLUS ON/OFF” and “ANTI-BOLUS” for CONFIG menu.
Problems from incorrect setup by user, are charged on user. It does not cover
by manufacturer.
30 SP-8800 / UME-03-REV.8
4. SYSTEM SETUP
4-6-2. BOLUS ON/OFF SPECIAL FUNCTION
• Get into BOLUS INFUSE menu in order to MENUSYSTEM SETUPCONFIGBOLUS ON/OFF.
NOTE Initial setup is OFF. [KEY LOCK P/W] , [BOLUS INFUSE] , [ANTI BOLUS]
NOTE • if “KEY SOUND” is OFF, there is no sound when the buttons are pressed
• Preset is ON for KEY SOUND
31 SP-8800 / UME-03-REV.8
5. SPECIAL
FUNCTIONS
5-1.Dosage
5-2.History
5-3.Profile (Option spec)
5-4.Drug label
5-5.Unit setup
5-6.Pro. mode
32 SP-8800 / UME-03-REV.8
5. SPECIAL FUNCTION
4. SYSTEM SETUP
5. SEPCIAL FUNCTION SPECIAL FUNCTION
• Press MENU button to getinto Menu mode and select SPEICAL FUNCTION with [SEL] button.
• SPEICAL FUNCTION MENU
1. DOSAGE
2. HISTORY
3. PROFILE (OPTION SPEC)
4. DRUG.LABEL (OPTION SPEC)
5. UNIT SETUP
6. PRO.MODE
5-1.DOSAGE
• DOSAGE SETUP
• Press buttons in order to MENUSPECIAL FUNCTIONDOSAGE and there come up “DOSE RATE” on the
LCD.
• DOSEGE : Automatically calculate and set the proper flow rate when the dose rate, body weight, drug mass and
solution volume are entered.
• After input each parameter, press [SEL] button. Each mode is comfirmed and turns into next parameter
automatically. Please refer to picture No. ① to ④.
• After setup complete parameter, it turns into picture No. ⑤ automatically and display complete information for
each parameter value.
• Each paramet, you can delet or change the value with [ESC / 2sCLR] or , , , buttons.
• Press [ESC / 2sCLR] button to back previous mode.
① ② ③ ④
• As soon as finish input each parameter value, complete data is displayed as picture No. ⑤. Select YES to
confirm the data for DOSAGE mode, it turns into STANDBY mode with confirm sound. Flow rate and Delivery
volume are displayed on FND automatically. Press [START] button to start. If you select NO on the picture No.
⑤, it goes back DOSE RATE mode automatically to setup each parameter again.
33 SP-8800 / UME-03-REV.8
5. SPECIAL FUNCTION
⑤ 4. SYSTEM SETUP
SPECIAL FUNCTION
* Change unit for DOSE RATE : Unit can be changed page 36 “UNIT SET”
1. ㎍ / ㎏ / min
2. ㎍ / ㎏ / hour
DOSE RATE
3. ㎎ / ㎏ / hour
4. ㎎ / ㎏ / min
5-2. HISTORY
5-2-1. Check HISTORY
• Get into HISTORY VIEW menu in order to MENUSPECIAL FUNCTIONHISTORY.
• HISTORY : Display last 10 infusion datas, infusion starting time, infusion completing time, FLOW RATE ,
INFUSED VOLUME.
34 SP-8800 / UME-03-REV.8
5. SPECIAL FUNCTION
• Move with , , , 4. SYSTEM
buttons and press [SEL] button to see detail information SETUP
for each data.
• HISTORY DATA : Last 10 infusion datas only.
SPECIAL FUNCTION
• PROFILE LOAD : Call back the previous setup data. When you select this mode, there is YES or NO selection
mode to start same profile.
• Select “YES” to start same profile with previous setup. The unit turns into STAND BY mode. Press [START]
button to start PROFILE, the unit start PROFILE mode with displaying “PROFILE” on LCD.
• Select “NO” to setup new PROFILE.
NOTE • Duringe PROFILE function, you can not setup Flow rate, Delivery volume. To stop
PROFILE function, make sure PROFILE LOAD setup as “ NO”.
• PROFILE : Program infusion condition(flow rate, delivery volume, delivery time) differently for every hour, up to
24hours. (Option).
• Select “PROFILE SET UP” and press [SEL] button to get into below picture mode. Press [SEL] button to setup
time for each profile, input each time with numerical buttons. Press [SEL] button to save each time and infusion
setup(Flow rate, Delivery volume).
• Setup profile data up to 24hours in order to same process.
• Press [START] button to save.
35 SP-8800 / UME-03-REV.8
5. SPECIAL FUNCTION
4. SYSTEM SETUP
SPECIAL FUNCTION
• Move with , , , buttons to select the unit and press [SEL] button to save it.
5-6.PRO.MODE
Contact local agency or local authorized dealer for this mode.
36 SP-8800 / UME-03-REV.8
6. TROUBLE SHOOTING
37 SP-8800 / UME-03-REV.8
6.TROUBLE SHOOTING - ALARM
Take the following actions if any trouble occurs. When the troubles could not be solved with the following
actions, Please contact your local authorized dealer immediately.
NOTE • Whenever alarm sounds, the pump stops infusion. That is, alarm sounds only in error
situation during infusion.
Pump cannot be • AC or DC power cord is not ▶Check the AC or DC power cord connection.
switched on. inserted properly • Never connect both AC and DC power to the pump
• Internal battery has deteriorated ▶Stop the operation of the pump and replace with a new battery through
• The voltage of the internal battery ▶Recharge the battery fully for more than 24 hours by connecting the
AMPall or distributor.
[OCCLUSION] LED • The infusion line is kinked or ▶1. Turn the alarm off by pressing [SILENCE] button.
flashes and alarm twisted ▶2. Check the infusion line and take a corrective action like untwisting or
alarm sounds again, and the ▶Turn the alarm off by pressing [SILENCE] button.
[FINISH] LCD display • The syringe is completely empty ▶1. Turn the alarm off by pressing [SILENCE] button.
and alarm sounds ▶2. To continue the infusion, replace the syringe with a new one.
[NEAR EMPTY] LED • The syringe is nearly empty ▶1. Press [STOP] button to stop infusion.
flashes and alarm (Finishing injection in 5 minutes ▶2. To continue the infusion, replace with the new syringe filled with
38 SP-8800 / UME-03-REV.8
6.TROUBLE SHOOTING - ALARM
[SLIDER HOOK] LED • Slider Hook does not fix Push- ▶ 1. Turn the alarm off by pressing [SILENCE] button.
flashes and alarm Button of syringe ▶ 2. Make Slider Hook fix Push-Button of syringe.
button.
[CLAMP] LED flashes • Clamp does not fix Barrel of syringe ▶ 1. Turn the alarm off by pressing [SILENCE] button.
and alarm sounds ▶ 2. Pull the clamp, turn it to the left up to 90° and put the finger
button.
[CLUTCH] LED flashes • Clutch is not properly engaged ▶ 1. Turn the alarm off by pressing [SILENCE] button.
button.
“BATT-LOW” on the • The voltage of internal battery is low ▶ 1. Press [STOP] button to stop infusion.
LCD and alarm sounds ▶ 2. Recharge the battery fully for 24 hours by connecting the pump
to an AC power outlet.
“KEEP-VEIN” on the • In the operation, the total volume ▶ Stop K.V.O. rate by pressing [STOP] button
• The total volume delivered exceeds ▶ Stop K.V.O. rate by pressing [STOP] button
the delivery limit ▶ When infusion is newly needed, set the delivery volume again or
When [PURGE] button • The syringe is not set properly ▶ Set the syringe correctly.
is pressed, the alarm • Even when the syringe is set again, ▶ The system is in trouble.
sounds. purging does not occur Please contact your local authorized dealer
“ERROR” is on the LCD • The system is in trouble ▶ Please contact your local authorized dealer
39 SP-8800 / UME-03-REV.8
6.TROUBLE SHOOTING - ALARM
Even when [START] • The value of “flow rate, delivery ▶ Confirm the values that were set.
• The low voltage of the internal ▶ Recharge the battery fully for more than 24 hours by connecting
[SYRINGE SIZE] LED • The syringe, which is used, may be ▶ Make sure that the syringe, which is used, is the same as the
and [CLAMP] LED flash different from the one specified in one specified in the pump in terms of manufacturer.
the pump in terms of manufacturer To use the syringe of different manufacturer, please contact your
stops ▶1. Turn the alarm off by pressing [SIL ENCE] button.
The accuracy of flow • The syringe, which is used, may be ▶ Make sure that the syringe, which is used, is the same as the
rate is questionable. different from the one specified in one specified in the pump in terms of manufacturer.
the pump in terms of manufacturer To use the syringe of different manufacturer, please contact your
• Otherwise, the system is suspected ▶ Please contact your local authorized dealer for repair.
to be in trouble
The operation hour by • Internal battery has deteriorated ▶ Stop the operation of the pump and replace with a new battery
recharging the battery • Otherwise, the system is suspected ▶ Please contact your local authorized dealer for repair.
to be in trouble
Battery is not recharged • AC power cord is not inserted ▶ Check the AC power cord connection.
connected to AC power • AC power cord has deteriorated ▶ Replace with a new AC power cord.
outlet. • Otherwise, the system is suspected ▶ Please contact your local authorized dealer for repair.
to be in trouble
Battery is not recharged • DC power cord is not inserted ▶ Check the DC power cord connection.
connected to DC power • DC power cord has deteriorated ▶ Replace with a new DC power cord.
outlet. • Otherwise, the system is suspected ▶ Please contact your local authorized dealer for repair.
to be in trouble
40 SP-8800 / UME-03-REV.8
SPECIFICATION
DELIVERY PERFORMANCE
ALARM
OCCLUSION, NEAR EMPTY, LOW BATTERY, K.V.O. RATE, CLUTCH OPEN, SLIDER HOOK OPEN,
FEATURES
CLASSIFICATIONS ClassⅠ, CF, IPX3
KEYPAD LOCK, SHORT-CUT KEY, HISTORY CALL-BACK, RETAIN MEMORY, REMAINNING TIME,
OPEN SYSTEM, PURGE, BOLUS, K.V.O., BUZZER VOLUME / OCCLUSION PRESSURE CONTROL,
CENTRALL SYSTEM(OPTION)
OTHER PARAMETERS
POWER REQUIREMENTS / CONSUMPTION AC110 / 230V, 50 / 60Hz or DC9.6V, DC24V 1A / 40VA
※Specifications and design are subject to change for improvement without prior notice
41 SP-8800 / UME-03-REV.8
SYMBOL
Meaning Symbol
Manufacture Date
Lot number:
WEEE Mark
Waterproof level
IPX 2
DC
Type CF Equipment
(Protection against electrical shock)
AC
42 SP-8800 / UME-03-REV.8
AMPall warrants that the product shall be free from defects in workmanship
and materials for the warranty period. The warranty is void if failure of the
product has resulted from accident, alteration, abuse or misapplication to
include unauthorized opening of the product.
Under the warranty, AMPall shall repair or replace at its option and expense
any products found defective by the customer during the warranty period
and returned to an authorized AMPall representative.
Return of the repaired or replaced product to the customer’s original location
shall be at the expense of AMPall unless AMPall determines that the
products are not defective within the terms of the warranty.
In such case the customer shall pay AMPall all costs of handling,
transportation and labor.
• EC Representative : CE medical
Amerlandseweg 7
3621 ZC Breukelen, The Netherlands
TEL: +31 (0)85 007 3220
E-mail: [email protected]
43 SP-8800 / UME-03-REV.8
Warranty Card
Thank you for your purchasing for AMPall product. The warranty period
Of SP-8800 is valid from the purchasing date. Please notice the purchasing
date on the warranty card and keep this Warranty Card with the device
all the time to get full customer service.
Product Name Infusion Pump
Manufacturing Date
Name: Tel.:
Customer
Add.:
Name: Tel.:
Retailer
Add.:
Repairing Record
☆ Please show this Warranty Card when you request repairing service.
44 SP-8800 / UME-03-REV.8