DRX-T7445HD-H: Toshiba X-Ray Tube (Water-Cooled Type)

The document provides product data and specifications for the Toshiba X-ray tube model DRX-T7445HD-H/S. It is a water-cooled rotating anode X-ray tube intended for high-speed inspection of circulatory organs. The document lists technical details such as electrical ratings, mechanical characteristics, cooling methods, and absolute maximum ratings.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
234 views4 pages

DRX-T7445HD-H: Toshiba X-Ray Tube (Water-Cooled Type)

The document provides product data and specifications for the Toshiba X-ray tube model DRX-T7445HD-H/S. It is a water-cooled rotating anode X-ray tube intended for high-speed inspection of circulatory organs. The document lists technical details such as electrical ratings, mechanical characteristics, cooling methods, and absolute maximum ratings.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

Product Data

No. MPDXR0119EA
TOSHIBA X-RAY TUBE
DRX–T7445HD–H/S
(WATER–COOLED TYPE)

APPLICATION
High-speed rotating anode X-ray tube
for circulatory organ inspection.

FEATURES
● The extra-heavy anode disc is con-
structed with specially processed rhe-
nium-tungsten-faced molybdenum,
graphite backed. (132 mm (5.2")
diameter)
● This tube is installed with the water-
cooled heat exchanger model HEX-
119. The maximum housing heat
dissipation rate is 3 kW with HEX-
119.
● This tube has foci of 0.8 mm, 0.5
mm and 0.3 mm, and is rated for a
maximum tube voltage of 150 kV
with DC or three-phase generator.

GENERAL DATA
IEC classification: Class I, Type B
● Motor ratings (1)
Electrical ratings Duty Starting Running
● Circuit: DC or three-phase
Power source (Hz) 180 60 180 60
full-wave rectified
(12-peak) Input power (W) 3550 2000 200 175
(center-grounded) Voltage (V) 350 190 80 62
● Operating tube voltage Current (A) 10.8 12.6 2.7 3.6
Radiography Min. speed-up time (s) 3.0(2) 2.0 − −
(less than 1.0 s): 50 kV to 150 kV
Capacitor (µF) 6 44 6 44
Radiography
(more than 1.1 s): 50 kV to 125 kV Min. braking (s) 4.5 (100 VDC)(3)
Fluoroscopy: 50 kV to 125 kV Notes (1) To be obtained with TOSHIBA high-speed starter RS-200P or
● Focal spots equivalent.
Large focus: 0.8 mm (2) The speed-up time from normal speed of 3000 rpm (3600 rpm) to high
Middle focus: 0.5 mm speed of 10800 rpm is 3.0 seconds.
Small focus: 0.3 mm (3) To be applied for high-speed rotation.
● Input power (at 0.1 s)
Large focus: 98 kW
Middle focus: 49 kW
Small focus: 20 kW
Product Data
No. MPDXR0119EA

● Anode rotation speed ● Maximum filament current


180 Hz: 9700 rpm min. (large focus 0.8 mm): 5.9 A
● Resistance between (middle focus 0.5 mm): 5.2 A
housing and low (small focus 0.3 mm): 5.4 A
voltage terminals: 2 MΩ min. ● Filament voltage:
● Input voltage of oil (at max. filament current)
pump in the tube (large focus 5.9 A): 13.1 V to 17.8 V
housing: 100 VAC, 2A (middle focus 5.2 A): 12.0 V to 16.2 V
(small focus 5.4 A): 3.7 V to 5.0 V
Mechanical characteristics ● Maximum input energy
● Dimensions Refer to outline drawing (continuous): 710 W (1000 HU/s)
• Overall length: 654 mm (25.7") ● Thermal characteristics
• Max. diameter: 242 mm (9.5") Anode heat storage
● Target capacity: 1300 kJ (1800 kHU)
• Angle (effective): 8° Maximum anode heat
• Material: Rhenium-tungsten dissipation rate: 4000 W (5700 HU/s)
● Inherent filtration: At least 1.1 mm Al Housing heat storage
equivalent at 75 kV capacity: 1550 kJ (2200 kHU)
● Radiation protection Maximum housing heat dissipation rate
(To meet the requirements of IEC 601-1-3/1994) With heat exchanger: 3000 W (253 kHU/min)
Leakage dose
measurement factor: 150 kV, 5 mA Ambient conditions
● X-ray irradiation field: 354 × 354 mm ● Temperature: 10°C to 40°C
(13.9" × 13.9") at ● Relative humidity: 30% to 85%
SID 1300 mm (51.2") (no condensation)
● Mass (approx.): 45 kg (99 lb) ● Atmospheric pressure: 70 kPa to 106 kPa
● High tension terminals: CLAYMOUNT MINI-75
● Cooling method: Water-cooled heat exchanger Transport and storage conditions (while packed)
● Inner cooling method: Oil circulation ● Temperature
● Tube housing • Before filling the
model No.: XH-164 cooling water: −20°C to 70°C
• After filling the
Absolute maximum and minimum ratings cooling water: 2°C to 60°C
(these values must not be exceeded at any time.) ● Relative humidity: 20% to 90%
● Maximum tube voltage (no condensation)
Radiography ● Atmospheric pressure: 50 kPa to 106 kPa
(less than 1.0 s): 150 kV
Radiography Atmosphere
(more than 1.1 s): 125 kV The equipment must not be used in an explosive or
Fluoroscopy: 125 kV corrosive gas environment.
● Maximum voltage to
ground: 78 kV
● Minimum tube voltage: 50 kV
● Maximum tube current
Refer to rating charts
(large focus 0.8 mm): 1100 mA
(middle focus 0.5 mm): 650 mA
(small focus 0.3 mm): 200 mA
TOSHIBA X-RAY TUBE
DRX–T7445HD–H/S (WATER–COOLED TYPE)
ABSOLUTE MAXIMUM RATING CHARTS
Conditions: Three-phase 50/60 Hz, 10800 rpm
Focal spot: 0.8 mm Focal spot: 0.5 mm
1400 700
70
80 k kV 60 kV
V
1200 90 kV 600 90 k
100 kV V
80 kV 70 kV 100 k
110 kV V 50 kV
1000 500 110 kV
125 k
Tube current (mA)

Tube current (mA)


V 125 kV
60 kV
800 150 kV 400
150 kV
50 kV
600 300

400 200

200 100

0 0
0.01 0.03 0.07 0.2 0.5 1 3 7 20 40 0.01 0.03 0.07 0.2 0.5 1 3 7 20 40
0.02 0.05 0.1 0.3 0.7 2 5 10 30 0.02 0.05 0.1 0.3 0.7 2 5 10 30
Exposure time (s) Exposure time (s)

Focal spot: 0.3 mm


200
110 kV 90 kV
180 100
125 kV kV 80 kV
160 70 kV
140 150 kV 60 kV
Tube current (mA)

120 50 kV
100
80
60
40
20
0
0.01 0.03 0.07 0.2 0.5 1 3 7 20 40
0.02 0.05 0.1 0.3 0.7 2 5 10 30
Exposure time (s)

THERMAL CHARACTERISTICS
Housing thermal characteristics Anode thermal characteristics
1600 1300
1200
1400

1000
1200
Cooling Cooling
Heat storage (kJ)

Heat storage (kJ)

1000 800
1000 W

800 710 W
600

600
400
400
710 W
Heating 500 W
200
200
Heating

0 0
0 10 20 30 40 50 60 70 80 0 2 4 6 8 10 12

Time (min.) Time (min.)


Product Data TOSHIBA X-RAY TUBE
No. MPDXR0119EA
DRX–T7445HD–H/S (WATER–COOLED TYPE)
OUTLINE DRAWING

126 4-M8
(5.0) 74
(2.9)

45°
φ160
φ126 (6.3)

φ112
φ128

(4.4)
(5.0)

(8.1)
φ9 6)

205
(3.
(5.0)

2
φ209
(8.2)

(5.3)
135
4-M6 Depth 8
(2.4±0.04)
60±1

Focal spot

(1.1)
(4.4)

28
111

111 (2.8)
72

(4.4)
(5.0)
(5.1)

127
130

(1.0)
26
80 140
(3.1)135 (5.5) 212 80
94 (3.1)
(3.7) (5.3) 180 (8.3) 175
317 (7.1) (6.9) 338
(12.5) (13.3)
655
(25.8)

Unit: mm (in)

★ The information contained herein is presented only as a guide for the applications
of our products. No responsibility is assumed by TOSHIBA for any infringe-
ments of patents or other rights of third parties which may result from its use. No
license is granted by implication or otherwise under any patent or patent right of
TOSHIBA or others.
★ The information contained herein may be changed without prior notice. It is
therefore advisable to contact TOSHIBA before proceeding with the design of
equipment incorporating this product.

TOSHIBA CORPORATION MEDICAL SYSTEMS COMPANY


1-1, SHIBAURA 1-CHOME, MINATO-KU, TOKYO 105-8001, JAPAN PHONE: 3457-3269 FAX: 5444-9345 TELEX: J22587 TOSHIBA
© Toshiba Corporation 1997 All Rights Reserved. Design and specifications subject to change without notice. 97-10 TME / SB Printed in Japan