FAPAS QC MATERIAL DATA SHEET T27208AQC


Matrix Wheat Flour
Weight / Volume of Contents 20 g

Analyte Qualitative Results


Soya POSITIVE (99% consensus)

Quantitative Results
Assigned No. of data points
Analyte Value, xa Range for |z| ≤2 Units producing xa

Neogen – Veratox for Soy 96.2 48.1 – 144.3 mg/kg 21


(8410)
R-Biopharm – Ridascreen 32.1 16.1 – 48.2 mg/kg 20
Fast Soya (R7102)

ELISA Systems – Soy 89.2 44.6 – 133.8 mg/kg 15


(ESSOYPRD-48)

This data sheet is applicable until 31 Dec 2019


Recommended Storage on receipt Ambient
Notes

• Mix the QC material thoroughly before taking a representative analytical sample.


• The assigned value has been derived from the consensus of laboratories taking part in
the accompanying proficiency test, using a specific kit type. This is not a certified
reference value.
• Results for each analyte were separated into individual kit types and assessed separately.
• The Range for |z| ≤2 is the concentration range within the limits of ±2 z-scores. The
assigned value and its range have been established from the proficiency test data and are
suitable for use by laboratories as a fit-for-purpose quality control measure.
• Stability of the QC material has been established as sufficient for the scope of the
proficiency test from previous experience, expert advice and published literature. FAPAS
advises that the QC material is analysed within the recommended date. FAPAS QC
materials are intended to be used as single-analysis samples.
• Full details on the proficiency test procedure used to characterise this QC material are
available in the Protocol, Part 1 - Common Principles, freely available to download from
the FAPAS website.