Exactive & Q Exactive MS: Operational Qualification 1.1 Procedure & Worksheets
Exactive & Q Exactive MS: Operational Qualification 1.1 Procedure & Worksheets
Q Exactive MS
Operational Qualification 1.1
Procedure & Worksheets
Revision E
© 2012 Thermo Fisher Scientific Inc. All rights reserved.
Thermo Fisher Scientific Inc. provides this document to its customers with a
product purchase to use in the product operation. This document is copyright
protected and any reproduction of the whole or any part of this document is strictly
prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical
information in this document is for reference purposes only. System configurations
and specifications in this document supersede all previous information received by
the purchaser.
This document is not part of any sales contract between Thermo Fisher Scientific
Inc. and a purchaser. This document shall in no way govern or modify any Terms
and Conditions of Sale, which Terms and Conditions of Sale shall govern all
conflicting information between the two documents.
John
15 May 2009 A New document.
Jameson
Brian Heft
16 January 2012
Senior Regulatory Product Support Engineer
Justification Author
Bryan Shade
16 January 2012
Senior Regulatory Product Support Engineer
Thermo Scientific
Revision E
Page 3 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
Dan Furlano
16 January 2012
Regulatory Support Manager
Thermo Scientific
Revision E
Page 4 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
MS Operational Qualification
Introduction
Documented evidence must be provided in order to demonstrate the
integrity of data collected and to validate the results obtained on
laboratory instrumentation.
• Exactive
• Q Exactive
Recommended Documents
Use this document along with the Exactive or Q Exactive Operations
Manual (P/N: 1249360 or 1288120) and individual component
operators/service manuals in order to complete the OQ.
Thermo Scientific
Revision E
Page 5 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
Recommended Materials
• 1 liter MS grade water
• 1 liter MS grade acetonitrile
• MS grade methanol (MeOH)
• Formic acid
• Qualification Test Loop (25001-60001)
• 5 µL Injection Loop (00109-99-00023)
• Disposable transfer pipettes
• Two clean 1 L solvent bottles
• 1 mg/mL Buspirone specification solution
• Dilution vials
• Positive Ion Mode Calmix Solution (88323)*
• Negative Ion Mode Calmix Solution (88324)*
Thermo Scientific
Revision E
Page 6 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
Overview of Tests
Instrument Parameter Description OQ Limits
Mass Calibration Passes:
>62,000
Resolution of MRFA is Exactive
Mass Resolution (+/-) observed at m/z 524 at
Ultra High resolution. >87,000
Q Exactive
Three injections of
≥10:1
5 µL of 100 fg/µL Buspirone
Exactive
solution are injected into the
Signal-to-Noise
mass spec. The data is ≥100:1
collected and evaluated for Q Exactive
the signal-to-noise ratio.
Thermo Scientific
Revision E
Page 7 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
Mass Spec Signal-to-Noise (Sensitivity) Specification
Mass Calibration
Mass calibration checks are performed in positive ion mode at m/z 74,
138, 195, 524, 1222, 1422 and 1622.
Mass Resolution
Dynamic Range
After a period of two hours, the RMS error of the external mass
accuracy is checked to verify stability.
Thermo Scientific
Revision E
Page 8 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
Thermo Scientific
Revision E
Page 9 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
Note: If using Office 2007 or 2010, ensure that the file is saved in the
.xls format and not the default .xlsx format. To save in .xls format,
choose, Save As, and choose the file type, Excel 97-2003
Workbook (*.xls).
Customer Information
Click on the tab labeled Customer Information. Complete this
information.
System Information
Click on the tab labeled System Information. Click on the cell
marked Choose Mass Spec. Using the drop down list for this cell,
choose the mass spec that is to be qualified.
The Thermo recommended specifications for each Operational
Qualification test are listed in the column labeled Suggested OQ
Limits. If different OQ limits are desired, the column labeled Tested
OQ Limits can be updated to reflect desired limits.
Note: Thermo will not be able to support limits more restrictive than
Suggested OQ Limits. If Tested OQ Limits do not match the
Suggested OQ Limits this will be denoted by a highlighted field in the
Tested OQ Limits section.
Thermo Scientific
Revision E
Page 10 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
Preparing Standards and Mobile Phase
Mobile Phase
1. Fill the bottle with HPLC grade water and label the bottle HPLC
Water.
2. Fill a bottle with a mix of 70:30 acetonitrile:water with 0.1% formic
acid. Label this bottle 70:30 acetonitrile:water with 0.1%
formic acid.
Standard Diluent
Thermo Scientific
Revision E
Page 11 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
3. Prepare 1 mL of 100 pg/µL Buspirone Stock Solution as follows:
a. Pipette 10 µL of the 10 ng/µL buspirone stock solution into a
clean dry vial.
b. Pipette 990 µL of the standard diluent into the vial.
c. Mix the solution thoroughly and label the vial 100 pg/µL
Buspirone.
Calibration Solutions
Thermo Scientific
Revision E
Page 12 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
Thermo Scientific
Revision E
Page 13 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
Mass Resolution: Positive Ion Mode (+)
Resolution is inversely proportional to the square root of the m/z. The
resolution and peak shape are checked at m/z 524.
1. In the Instrument Control/Scan Parameters window set the
following:
• Exactive:
i. First Mass (m/z): 300
ii. Last Mass (m/z): 2000
iii. Resolution: Ultra High (100,000 @ 1Hz)
iv. AGC Target: Balanced (1e6)
• Q Exactive
i. Scan type: Full MS-SIM
ii. First Mass (m/z): 200
iii. Last Mass (m/z): 1000
iv. Resolution: Ultra High (140,000 @ 1.5Hz)
v. AGC Target: Balanced (1e6)
Note: the above settings may need to be adjusted to see the base
peak.
2. Zoom into the MRFA peak m/z = 524.
3. To display the resolution, right-click the Spectrum window and
choose Display options. Enable Display option, Show
Resolution.
4. Print the spectra or copy and paste into the Exactive Log.doc.
5. Record the results on the FTMS tab of the OQ Worksheets.
6. Expected results are as follows:
• 100,000 (Exactive):
i. Resolution at m/z 524 > 62,000
• 140,000 (Q Exactive):
i. Resolution at m/z 524 > 87,000
• Intensity of Shoulders (vs. main signal) < 4%
Thermo Scientific
Revision E
Page 14 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
HCD Performance: Positive Ion Mode (+), if installed
To check the HCD performance, compare the intensity of caffeine m/z
138 with and without HCD fragmentation.
1. In the Instrument Control/Scan Parameters set the following:
• Exactive:
i. First Mass (m/z): 70
ii. Last Mass (m/z): 300
iii. Max Inject Time (ms): 50
iv. Resolution: Ultra High (100,000 @ 1Hz)
v. AGC Target: Ultimate mass accuracy (5e5)
vi. Fragmentation: HCD: 20-30 eV
• Q Exactive
i. Scan type: MS/MS
ii. Precursor Mass: 195.1 Width = 2
iii. Scan range (m/z): 70-1000
iv. Max Inject Time (ms): 30
v. Resolution: Ultra High (140,000 @ 1.5Hz)
vi. AGC Target: 5e4
vii. Fragmentation: HCD: NCE = 10
2. For the Exactive:
• In the Instrument Status Tree, choose
Control/Settings/HCD Event. If needed, optimize HCD by
maximizing the peak height of the caffeine fragment at
m/z = 138.
• Fix the vertical scale in the Spectrum window and print or
copy both spectra (with and without fragmentation) into the
Exactive Log.doc.
3. For the Q Exactive:
• Fix the vertical scale in the Spectrum window and print or
copy the spectra with low fragmentation into the Exactive
Log.doc.
• Increase the NCE setting to increase the peak height of the
caffeine fragment at m/z= 138. Optimize the normalized
collision energy to increase the 138 fragment to a relative
peak height above 15%. The expected result is a NCE
between 30 and 40.
• Print or copy the spectra with high fragmentation into the
Exactive Log.doc.
Thermo Scientific
Revision E
Page 15 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
4. Record the results on the FTMS tab of the OQ Worksheets.
5. Expected results are as follows:
• Relative Peak Height at m/z 138 > 15%
• Sum of Relative Peak Heights at m/z 138 and 195 > 30%
Thermo Scientific
Revision E
Page 16 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
External Mass Calibration Stability (+)
Prior to running the external mass calibration stability check:
• Assure the Mass Calibration was performed at least 12 hours
after the system was baked out.
• Assure that the instrument electronics were on for at least 2
hours.
• Assure that the H/ESI ion source is stable.
1. In Instrument Control/Scan Parameters set the following:
• First Mass (m/z): 120
• Last Mass (m/z) : 2,000
• Max Inject Time (ms): 30 (20 for Q Exactive)
• Fragmentation: in-source CID: 20 eV
• Resolution: High
• AGC Target: Balanced (1e6)
2. The stability check runs for 2 hours. Be sure to fill the 500 μL
syringe pump and set the pump flow to 4 μL/min (240 μL/hour)
or less so that sufficient sample is available for a 2-hour run. It is
necessary to have stable H/ESI conditions.
3. Choose Control/Procedures/Evaluate/Extra Evaluation and
run Long-Term Mass Accuracy Test for 2 hours.
Note: these menu options may differ slightly in older versions of the
software.
4. When the test has completed, print the results of the test or copy
and paste into the Exactive Log.doc.
5. Record the results on the FTMS tab of the OQ Worksheets.
6. Expected results are as follows:
• RMS Error After 2 Hours < 5 ppm
Thermo Scientific
Revision E
Page 17 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
Mass Calibration: Negative Ion Mode (-)
1. Switch the instrument to negative polarity.
A complete calibration (negative ion mode) must be performed prior
to running the calibration checks to prevent the incorrect masses
from being utilized.
2. Through the Tune application open the Calibration dialog window.
3. With the CalMix infusing, run the Negative Mass Calibration
Check (this is called Evaluation in newer Tune versions).
4. When the test has completed, print the results of the test or copy
and paste into the Exactive Log.doc.
5. Record the results on the FTMS tab of the OQ Worksheets.
Thermo Scientific
Revision E
Page 19 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
• Change the scan parameters to the following:
i. Scan type: MS/MS
ii. Precursor Mass: 1480 Width = 8
iii. Scan range (m/z): 600-1484
iv. Fragmentation: HCD: NCE ~ 25
• Print or copy the spectra with high fragmentation into the
Exactive Log.doc.
3. Record the results on the FTMS tab of the OQ Worksheets.
4. Expected results are as follows:
• Relative Peak Height at m/z 1305 > 50%
Thermo Scientific
Revision E
Page 20 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
Thermo Scientific
Revision E
Page 21 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
7. In the Tune software, put the system in the On mode.
Note: Utilize positive mode.
8. Set up the source parameters as follows for the Exactive
(or the Q Exactive):
• H/ESI – Source (MaxIon, metal needle kit installed)
• Spray Voltage: 3.5 – 4.0 kV
• Sheath Gas Pressure: 20-30 constant spray conditions
• Aux Gas Pressure: 6-12 constant spray conditions
• Capillary Temperature: 275 (320 for the Q Exactive)
• Define Scan: Full MS-SIM
• Scan range: 300 to 1000
• RES: Ultra High 100,000 (70,000 for the Q Exactive)
• AGC target: balanced 1e6 (3e6 for the Q Exactive)
• Max Inj Time: 250 ms or higher, if MaxInjTime is reached
permanently
• microscans: 1
• Polarity: pos
9. Additionally, for the Q Exactive:
• S-Lens: 55%
• Heater Temperature: 310
• Lock Masses: On; 445.12003
10.Set the Divert/Inject valve to the Inject position.
11.Set up the LC Pump to run 70:30 Acetonitrile:water at a flow rate of
200 µL/min.
12.Put the LC Pump in the Run mode.
MS Detector
• Acquisition time: 5.50 min
• Segment 1: 5.50 min duration at 0.00
• Tune file: C:\Xcalibur\Operational Qualification\Hardware\
Exactive_ Buspirone_Test_ESI.mstune or
QE_Buspirone_Test_ESI.mstune
• Lock masses: Off
• Event 1:
o Positive Mode
o Scans: 1
o R=100,000 (140,000 for the Q Exactive)
o AGC=1,000,000
o Inject=250.00 ms [300.0 - 1,000.0] (120 – 1800 for the Q
Exactive)
o Source CID: disabled
o HCD: disabled
• Syringe Pump
o In use
o 1: 5.50 min duration at 0.00 in On
o End of run: Pump will be stopped
o Syringe type: Unimetrics 250 µL
o Syringe diameter: 2.300 mm
o Flow rate: 10.000 µL/min
• Divert valves
Divert valve 1: in use
o 1: 0.99 min duration at 0.00 in Pos2
o 2: 0.51 min duration at 0.99 in Pos1
o 3: 0.99 min duration at 1.50 in Pos2
o 4: 0.51 min duration at 2.49 in Pos1
o 5: 1.01 min duration at 3.00 in Pos2
o 6: 0.50 min duration at 4.01 in Pos1
o 7: 0.98 min duration at 4.51 in Pos2
Divert valve 2: not driven
Thermo Scientific
Revision E
Page 24 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
Thermo Scientific
Revision E
Page 25 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
Exceptions and Comments
1. If an exception is made during the qualification, the exception must
be recorded on Exceptions/Comments tab.
2. Open the OQ Worksheet and click on the tab labeled
Exceptions/Comments.
3. Record any exceptions, changes, or comments.
4. Click the Save button to save the changes to the OQ Worksheets.
Thermo Scientific
Revision E
Page 26 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
Thermo Scientific
Revision E
Page 27 of 28
Exactive & Q Exactive MS Operational Qualification 1.1
Procedure
OQ Completion
Required Documents
The following documents must be part of the OQ packet submitted for
customer review:
Customer Review
• The OQ results should be reviewed by the Customer. If the
qualification is accepted, both the Customer and the Qualification
Engineer should sign the OQ Results, below.
OQ Results
Based upon the actual results obtained, this OQ Passed Failed the acceptance criteria described
in Operational Qualification Scope, above.
Qualification Engineer
I have completed all aspects of the operational qualification and have concluded that the system has been
successfully verified to be operating as required.
Customer
I agree that all aspects of the operational qualification have been completed and that the system has been
successfully verified to be operating as required.
Thermo Scientific
Revision E
Page 28 of 28