Please see drawings below before starting
with the instructions Installation Instructions
Ambient Conditions
Our balances are designed to provide reliable
weighing results under normal ambient conditions in
the laboratory and in industry. When choosing a location
to set up your balance, observe the following so that
you will be able to work with added speed and accuracy:
3
Connecting the BBC 22
Balance to the Electronics Operating the Balance
Box
Plug the connecting cable Warmup Time
into the socket
on the electronics box. Turning the Display On and Off (Standby Mode)
Press the ON/OFF key to turn the display on and off.
Preparing the Weighing Chamber
Place the components listed below Self-Test
on the balance in the order given:
After the balance has been
– Shield disk
turned on, an automatic self-test
– Weighing pan
of the balance’s electronic
circuitry is performed.
Connecting the Balance to AC Power
At the end of the self-test,
Model BBC 22:
a zero readout is displayed.
The balance is powered by This means that the balance
an AC adapter. Make sure that is ready to operate.
the voltage printed on this unit is
identical to your local line voltage. Important Note Concerning Verified Balances Approved
Plug the cord of the AC adapter for Use as Legal Measuring Instruments:
into DC jack on the balance.
For verified balances that have a verification scale
interval “e” which is greater than the scale interval “d,”
the last digit on the display is bordered.
4
Centre the calibration weight on the weighing pan.
Calibration/Adjustment The balance then calibrates automatically. At the end of
calibration, the calibration weight readout and the
During calibration1), the sensitivity of the balance is adjusted stability symbol “g” are displayed.
to changes in the ambient conditions.
You must adjust or calibrate your new balance at the place 3
) factory setting for standard balances with a readability
of installation after each warmup period and before the first of ≥ 1 mg
measurement. You must also re-adjust or recalibrate your
balance each time you set it up in a different area or when
the ambient conditions change (especially the temperature). Sensitivity Test for Balances with a Built-in
Verified balances approved for use as legal measuring Calibration Weight
instruments must be adjusted/calibrated at least once Menu code selection: 1 9 4
a day.
Rather substantial changes in barometric pressure and
The balance offers you various calibration and adjustment temperature may affect the display response of the balance.
functions. You can select these functions by setting the To ensure that you obtain the full accuracy of the balance,
appropriate menu codes even when you use the entire weighing range, we have
implemented a sensitivity test function. It will help you decide
Using Verified Balances as Legal Measuring Instruments: quickly whether or not you need to recalibrate your balance
Before using your balance as a legal measuring instrument, (for example, to maintain the same accuracy
you must carry out an “Internal Calibration” operation at the during long-term weighing series).
place of installation after the warmup period.
Unload the balance and tare.
Internal Calibration for Balances with a Built-in
Calibration Weight When a zero readout is
displayed, press the CAL key.
Menu code selection: 1 9 3 2) The built-in calibration weight is
now internally applied by
The built-in calibration weight servomotor. At the same time,
is internally applied by servo- CAL is displayed. After the
motor and removed at the end display has stabilized, the
of adjustment or calibration. deviation of the momentary
readout from the target weight
If any interference affects (in grams only) is indicated.
the calibration procedure, If external interference affects the
you may obtain a brief display sensitivity test, you may obtain a
of the error code “Err 02.” brief display of the error code
In this case, tare and press “Err 02.” In this case, tare; then
the CAL key again. press the CAL key again.
Important Note
Important Note
For standard balances: the balance should
During calibration, the weighing pan must be unloaded. be calibrated if the deviation of the readout from zero is more
than the reproducibility specified for the balance.
1
) “Calibration” technically means to determine the difference between For verified balances approved for use as legal measuring
the balance readout and the actual weight on the pan to determine the instruments: the balance should be calibrated
accuracy. Adjustment means to bring a balance into the state of if the deviation of the readout from zero is more than
accuracy required for its use. Therefore, “calibration,” as used in this
manual, actually means “adjustment.” the verification scale interval.
2) factory setting for standard balances with a readability of ≤ 0.1mg
and for verified balances approved for use as legal measuring instruments . To calibrate the balance with
the built-in weight:
External Calibration press the CAL key briefly
Menu code selection: 1 9 1 3)
or
Use only calibration weights
with nominal mass values and To quit the sensitivity test:
tolerances equal to or better press the CF key briefly
than the accuracy class specified
for your balance.
Blocking the Calibration/Adjustment Functions
When a zero readout is You can block these functions by setting code 1 9 7
displayed, press the CAL key. (when the menu access switch is locked).
This starts calibration. The
calibration weight in grams is then
displayed. Errors or interference
at the start of the calibration
routine are indicated by the error
code “Err 02.” If this is the case,
tare and press the CAL key
again when a zero readout
appears.
5
Troubleshooting Guide
Problem Causes Solution
No segments appear on – No AC power is available – Check the AC power supply
the weight display – The AC adapter is not plugged in – Plug in the AC adapter
No segments appear – The surface on which the balance – Make sure that the ambient conditions
on the weight display after rests is not stable are stable
calibration/adjustment – Internal stability has not – Prevent vibrations from affecting
been reached the surface on which the scale rests
– Close the draft shield
The weight display shows “H” – The load exceeds the capacity – Unload the balance
of the balance
The weight display shows “L” – The weighing pan and/or the pan – Position the pan and/or pan support disk
or “Err 54” support disk is not in place (depending on the balance model)
The weight display briefly shows – The display did not show – Press the TARE key;
“Err 02” a zero readout when the CAL key then press the CAL key again
was pressed to calibrate
– The balance is loaded – Unload the balance
The special code “b” – None of the keys has been pressed – Press a key
remains displayed since the balance was turned on
The weight readout changes constantly – Unstable ambient conditions – Set up the balance in another area
– Too much vibration, or the balance – Access the menu to select the correct code
is exposed to a draft for the weighing environment
– The draft shield is not completely closed – Close the draft shield
– A foreign object is caught between – Remove the foreign object
the pan and the balance housing – Close the port for
– The below-balance weighing below-balance weighing
port is open
– The sample does not have a stable
weight (absorbs moisture or evaporates)
– The sample is electrostatically charged
The weight readout is obviously wrong – The balance has not been calibrated – Calibrate the balance
– The balance was not tared – Tare before weighing
before weighing
– The air bubble of the level indicator – Level the balance
is not within the circle
6
– Confirming a Menu Code
Balance Operating Menu Setting
Press the TARE key to confirm
The factory-set menu codes are identified by an “*” the code you have just
in this instruction manual. You can select the functions not set (this is indicated by
identified by an “*” by setting the respective menu codes. the “o” after the code).
You will find these codes in the section entitled
“Balance Operating Parameters”. – To store the new menu code
setting, press the TARE key
Important Note Concerning Verified Balances Approved for more than 2 seconds!
for Use as Legal Measuring Instruments:
The balance operating menu can also be changed when
the balance is being used as a legal measuring instrument. The current code setting in the balance operating menu
Codes that are not permitted for operation of the balance as is identified by a small, superscript “o” after the last number.
a legal measuring instrument are blocked and cannot be When you access the operating menu, the previously
selected. The balance operating menu on verified balances set code will be displayed after you have selected
cannot be locked with the menu access switch the right-hand number.
(“–” not displayed). If you would like to change several menu code settings,
you do not have to press the TARE key after each change
Changing Menu Code Settings to exit the balance operating menu. You can also confirm
individual settings.
To select specific functions, you will need to set the
respective menu codes.
Important Note Concerning Standard Balances:
There are three steps to changing a code: Please do not forget to relock
– Accessing the menu the balance operating menu to
– Setting a code avoid inadvertent changes to
– Confirming and storing the code the settings. The “–” symbol
indicates that the menu is
The keys have special functions for setting menu codes: locked.
CAL = Increases a number by one with each press Exiting the Menu without Storing Code Changes
(the numbers change in cycles) Changes to the code settings are not stored if you turn off
the balance by pressing ON/OFF while selecting the code
TARE = Confirms and stores a code setting, numbers or before pressing TARE to save a setting.
and exits the menu
Undoing All Menu Code Changes – Reset Function
PRINT = Moves to the next of the three numbers The reset function lets you undo all menu code changes,
of a code (1st-2nd-3rd-1st, etc.) which means that you will obtain the original factory-set
menu codes identified by an “*.” To use this function, select
Accessing the Menu (for example, code 8 5 4) code 9 -- 1°.
– Press ON/OFF to turn off
the balance. Reset function Code
– Turn the balance back on. On 9——1
While all segments are Off 9——2
displayed, briefly hold down
TARE.
– Release TARE when “1”
is displayed. Balance Operating Parameters
– For standard balances,
if “–” is displayed next to the Adapting the Balance to
left-hand number, proceed Ambient Conditions Code
as follows to unlock the menu: Very stable conditions 1 1 1*
– Remove the protective cap Stable conditions 1 1 2*
located on the front right of the Unstable conditions 1 1 3
balance and move the menu Very unstable conditions 1 1 4
access switch in the direction
of the arrow. Standard Weighing Mode –
– Press CAL several times Manual Filling Mode Code
until “8” appears. Standard weighing mode 1 2 1*
– Press PRINT until the 2nd Manual filling mode 1 2 2
number of the code appears.
– Press CAL until “5” appears. Stability Range Code
– Press PRINT until the 3rd
number appears 0.25 digit 1 3 1
(when you move to the third 0.5 digit 1 3 2*
number, the previously 1 digit 1 3 3
set menu code will appear). 2 digits 1 3 4*
– Press CAL to select “4.” 4 digits 1 3 5
8 digits** 1 3 6
7
Tare Parameter Code** Symbol/ Code
At any time 1 5 1 Conversion factor
Not until the readout is stable 1 5 2*
1g= 1st level 2nd level R1
Auto Zero Function Code Milligrams mg 1000. 1 7 13 3 1 13 *
Auto Zero on 1 6 1* Parts
Auto Zero off 1 6 2 per pound** o (/lb)2) 1.1287667712 1 7 14 3 1 14
Chinese
taels ** tl (thlc)2) 0.02645547175 1 7 15 3 1 15
Calibration/Adjustment and Mommes** m
Linearization Functions Using CAL Code (mom)2) 0.2667 1 7 16 3 1 16
Carats** k (K)2) 5. 1 7 17 3 1 17
External calibration 1 9 1*
Tola** t (tol)2) 0.0857333381 1 7 18 3 1 18
Internal calibration
Baht** b
for balances with a built-in automatic
(bat)2) 0.06578947436 1 7 19 3 1 19
calibration function 1 9 3*
Mesghal** m
Sensitivity test
(MS)2) 0.217 1 7 20 3 1 20
for balances with a built-in automatic
calibration function 1 9 4
Internal linearization (only for BBC22) 1 9 6 * = factory setting; depends on the balance model in some cases
Calibration/adjustment ** = not applicable to verified balances approved for use as
functions blocked 1 9 7 legal measuring instruments
1) = not applicable to verified balances of accuracy class k
2) = unit symbols output via the data interface
* = factory setting
8
Utilities for Printouts or Application Programs
Data Transfer In addition to the functions implemented for weighing,
our balances offer you a variety of application programs.
Data Output Parameter Code
Print on request regardless of stability 6 1 1 To select an application program or a specific function,
Print on request after stability, set the appropriate code in the menu.
with storage of the function 6 1 2*
Print on request at stability, Additional settings for running a particular program and
without storage of the function 6 1 3 displaying or printing the data on hard copy are listed in
Auto print regardless of stability 6 1 4 a table of codes. For your convenience,
Auto print at stability 6 1 5 we have indicated all factory-set codes with an “*.”
Auto Print Code Using Verified Balances as
Start/stop auto print using Legal Measuring Instruments:
the PRINT key 6 2 1 All application programs can be selected on balances
Auto print not stoppable 6 2 2* used as legal measuring instruments. Non-metric values
are indicated as follows:
Data Output at Defined Intervals Code
1 display update 6 3 1* Percent =%
2 display updates 6 3 2 Piece count (counting) = pcs
Computed value =o
Automatic Taring after Data Output
(Print on Request) Code
Data output without automatic taring 6 4 1* Tare Memory NET
Data output with automatic taring 6 4 2
Code 2 1 6
Automatic Output of
the Application Parameters Code Press the FUNC. key to store the tare weight. The balance
Off 7 1 1* is now automatically tared so you can weigh again starting
On 7 1 2 with a zero readout.
If you have stored a value in the tare memory, a “1” will be
Data ID Codes Code printed or output after the ID code of this value to identify it
Without 7 2 1* as such on the printout or on the computer screen.
With 7 2 2
Practical Example: Tare – Net – Gross Weights
Automatic Output of Menu code settings used in the example: Code
the Tare Memory Data Code Tare memory 2 1 6
Last net value (individual value N1) 7 3 1* Automatic output of all parameters 7 1 2
Tare memory data (total T1) 7 3 2 With data ID code 7 2 2
Print tare memory 7 3 2
* = factory setting
As an alternative, you can select this automatic
output parameter:
Print net value 7 3 1*
Additional Functions
Menu Access Switch Function on
Standard Balances Code
Accessible 8 1 1
Read only 8 1 2*
* = factory setting
9
Application: Obtaining printouts of tare, net and gross Changing the Reference Percentage
weights or outputting these values to a computer You can change the reference percentage in cycles.
Choose from the following settings: 5, 10, 20, 50 and 100.
Step/Key Readout Data Output/Printout
To activate the change function: Hold down FUNCTION
for more than 2 seconds
CF, TARE 0.00 g until “rEF 100 %”
Place container on pan + 22.65 g appears on the display
FUNCTION 0.00 g NET T1 + 22.65 g
Fill container with sample;
To change the setting: Briefly press FUNCTION
press PRINT + 250.24 g NET N1 + 250.24 g To store the percentage: Hold down FUNCTION
CF, PRINT + 272.89 g N + 272.89 g for more than 2 seconds
to store this value
permanently (in the non-
The data ID codes mean:
volatile memory) after
T1 + 22.65 g tare weight stored in the memory you turn off the power
(weight value)
N1 + 250.24 g net weight when tare weight is stored This setting is not canceled by the reset code 9 – – 1°!
N + 272.89 g gross weight = tare + net
As an alternative, you can select this output parameter: If the weight stored is too light to be displayed, the number
Printout of individual of decimal places is automatically decreased.
weight/total tare weight 7 3 2
* = factory setting; depends on the balance model in some cases
Application: Simple compounding and formulating of
several components with additive storage, automatic taring Practical Example:
and output of the component weights Determination of the Residual Weight in Percent
10
Counting Z Application:
Counting bulk quantities of items with the same weight
Code 2 1 4
Step/Key Readout Data Output/Printout
The counting program allows automatic conversion
of weights into piece counts based on a reference Place container on pan + 22.65 g
sample weight. A weight readout is stored as a reference CF, TARE 0.00 g
sample quantity (factory setting: 10 pcs = pieces). For the factory setting,
When you turn on the scale, the reference sample quantity add 10 counted parts
will be displayed as “rEF 10 pcs” before you enter Change the
the piece count. reference sample qty.,
if desired:
To use 20 as
Displaying the Reference Sample Quantity the reference qty.,
You can have the reference sample quantity (pcs) add 10 more parts
setting displayed by pressing the FUNCTION key when the to container.
balance is unloaded. To change the ref.
sample qty. to 20,
hold down FUNCTION
Changing the Reference Sample Quantity for more than
You can change the reference sample quantity in cycles. 2 seconds rEF 10 pcs
Choose from the following settings: 5, 10, 20, 50 and 100. Press FUNCTION rEF 20 pcs
To store: hold down
To activate the change function: Hold down FUNCTION FUNCTION for
more than 2 seconds
for more than 2 seconds
Press FUNCTION to confirm
until “rEF ... pcs” appears reference sample quantity rEF 10 pcs
on the display + 5.6546 g
To change the setting: Briefly press FUNCTION + 10 pcs nRef+ 10 pcs
To store the quantity: Hold down FUNCTION wRef+ 5.6546 g
for more than 2 seconds Fill container with desired
to store this quantity quantity of parts
(without counting them) + 500 pcs
permanently (in the non-
PRINT + 500 pcs Qnt + 500 pcs
volatile memory) after you
Press CF to clear the weight
turn off the power readout and the reference value + 2827.35 g
PRINT + 2827.35 g N +2827.35 g
This setting is not canceled by the reset code 9 – – 1°!
* = factory setting
Animal Weighing/Averaging V
Practical Example: Counting Small Parts
Code 2 1 7
Menu code settings used in the example: Code
Use this program to determine the weights of live animals
Counting 2 1 4
or weights under unstable ambient conditions.
Ref. qty. and ref. weight 7 1 2
During averaging, the number of individual subweighing
With data ID code 7 2 2
operations left is shown on the application display in
a “countdown” mode.
Once all subweighing operations have been performed,
the calculated mean value is indicated as a stable readout
on the weight display.
11
Animal weighing/averaging Practical Example:
will start by: Code Animal Weighing in the Automatic Start Mode
Manual mode 3 8 1
Automatic mode 3 8 2* Menu codes used in the example: Code
Animal weighing 2 1 7
Once averaging has been completed, the program Delay start until dif. is average 3 7 2*
will stop until the balance is unloaded to half the value Automatic start mode 3 8 2*
(50 display digits) of the storage threshold. Automatic output of all parameters 7 1 2
With data ID code 7 2 2
* = factory setting
As an alternative you can select this code:
Delayed Start Mode Other delay start 3 7 x
A rule of thumb to go by for selecting the right setting Manual start mode 3 8 1
to weigh animals is: the more active an animal is,
the greater the difference must be between two successive
subweights measured. You can change the number of subweighing operations.
Depending on individual requirements, starting the If you do not enter a number, averaging will be done based
averaging operation can be delayed either in the automatic on the number shown on the display
or manual mode until the animal you are weighing has (factory setting: 10 subweighing operations).
calmed down to a certain degree. Once you press FUNCTION the first time, all further
weighing operations will be performed automatically after
In this case, the start criterion is defined by the difference they have met the start criteria.
between two successive subweights measured.
If the animal moves, the start criterion is not met; therefore, * = factory setting
averaging will not start.
Once the animal has calmed down, the program checks Application:
whether two measured subweights are within the previously Automatic weighing of animals based on 20 subweighing
selected range. If so, the actual averaging operation will be operations
started.
Step/Key Readout Data Output/Printout
Delay start until Code
difference is slight 3 7 1 Place animal weighing
bowl on balance
difference is average 3 7 2*
CF, TARE 0.00 g
difference is considerable 3 7 3
Change number of
subweighing operations;
During averaging, the number of subweighing e.g., set number to 20:
operations left to perform is shown on the weight display Hold down FUNCTION
(countdown mode). for more than 2 seconds rEF 10
Press FUNCTION rEF 20
Changing the Number of Subweighing Operations To store: hold down FUNCTION
You can change the number of subweighing operations for more than 2 seconds rEF 20
used to average a weight. Change this number in cycles. Place 1st animal in bowl
You can choose from the following settings: FUNCTION 20
465.20 g mdef + 20
5,10, 20, 50 and 100. x-Net +465.20 g
Remove 1st animal
To activate the change function: Hold down FUNCTION Place 2nd animal in bowl 20
for more than 2 seconds 388.53 g mdef + 20
until “rEF 10” appears x-Net +388.53 g
on the display Remove 2nd animal
To change the setting: Briefly press FUNCTION Place 3rd animal in bowl 20
401.18 g mdef + 20
If you enter a wrong x-Net +401.18 g
number, press CF Repeat above steps for all animals
to clear it.
To store the number: Hold down FUNCTION
for more than 2 seconds The data ID codes mean:
to store this number
mdef 20 defined number of subweighing operations
permanently (in the non- x-Net 401.18 g calculated average
volatile memory)
after you turn off the
power
12
The printout can have the following lines:
GLP/GMP-compliant Printout
or Record ------------------
Model BBC22 : Balance model
Application: S/N 055600048 : Serial number
Use of the balance as a test and measuring instrument of the balance
in quality assurance systems in compliance with the Id : Space for entering a
requirements of ISO, GLP, GMP and EN workstation or operator ID
(European Standards) in which proof of the balance’s ------------------
accurate performance is required. Date : 21-Sep-95 : Current date
Start: 11:05:30 : Time at which the
The balance can record all completed calibration or application started
adjustment operations and print out data in compliance Cal. : Extern : Calibration mode (in this
with the requirements of Good Laboratory Practice (GLP). case, “external calibration”)
The balance, interfaced with a data printer or a computer, Set. : 200.00000 g : Calibration weight value
creates a document that records the date, time, serial (only for “external
number and model number, making it possible to clearly calibration”)
trace data to the balance that generated it and the time End : 11:05:45 : End of application
at which it was generated. Name : : Space for signature of
the operator responsible
Select the GLP/GMP-compliant printout or ------------------
record mode by setting the respective code in the
balance operating menu: Additional information for other calibration modes:
Cal. : Intern : Internal calibration
GLP/GMP-compliant Stat.: Complete : Status comment for
printout/record mode Code calibration or linearization
functions
Off 8 10 1 * Cal. : Test : Sensitivity test
Only for calibration/adjustment Diff.:- 0.00004 g : Data measured during
and linearization functions 8 10 2 the sensitivity test
Always on
(e.g., for calibration/adjustment Data Printout/Record (GLP/GMP-compliant)
and linearization functions,
application programs, weight readouts) 8 10 3 To have a data record printed out, perform the following:
– Press PRINT to output the printout heading and
The following menu code setting must be selected in order the first value (after you have turned on the balance or
to obtain an GLP/GMP-compliant printout/record: cleared a function by pressing CF)
– Press PRINT to output additional data
With data ID code 7 2 2 – To end printout generation and recording of data,
press CF
Important Note
Generation of an GLP/GMP-compliant printout or record
GLP/GMP-compliant printouts/records will not is also ended when an adjustment or calibration operation
be generated if the factory setting, code 7 2 1, is selected. is started.
In addition, do not select the “Auto print” data output
parameter (code 6 1 4 or 6 1 5). The printout can have the following lines:
------------------
Model BBC22 : Balance model
Operating the Balance with a Device that has an S/N 055600048 : Serial number
GLP/GMP Printing or Recording Capability of the balance
You can connect a special printer to balances for Id : Space for entering a
generating GLP/GMP-compliant printouts. workstation or operator ID
The GLP/GMP-compliant printer provides you with ------------------
the following functions: Date : 12-Sep-95 : Current date
– Turns the GLP/GMP function on and off Start: 11:00:30 : Time at which the
– Date/Time application started
– ID no. for identifying a workstation/operator Ser. : : Space for entering
– Printouts with balance-specific data a project number
N + 5.45390 g : Measured weights
* = factory setting N + 24.34586 g
N + 63.23450 g
Printout/Record for Adjustment or Calibration and End : 11:15:55 : End of application
Linearization Functions Name : : Space for signature of the
A printout or record is generated at the end of operator responsible
the following functions: ------------------
– all adjustment or calibration and linearization procedures
– sensitivity test
13
Data Printout/Record for Application Programs
For application programs, reference data (parameters)
can be included in the printout/record.
* = factory setting
14
Specifications Specifications
Model BBC22 Model BBC31 BBC32
Tare range (subtractive) g – 220 Tare range (by subtraction) g – 220 – 120
Power requirements V Via AC adapter: Power requirements V Via AC adapter: STNG 6, 230 VAC
STNG 6, 230 VAC or or 115 VAC, – 20%...+15%
115 VAC, – 20% ... +15%
Frequency Hz 48 – 63
Frequency Hz 48 – 63
Power consumption
Power consumption (average) VA 18.7: maximum; (average) VA 16: maximum; 8: average
10: average
Built-in interface RS-232C-S/V24-V28; 7-bit; parity:
Built-in interface RS-232C-S/V24-V28; 7-bit; even, odd; transmission rates:
parity: even, odd; 150 ... 19,200 baud; 1 or 2 stop bits;
transmission rates: software/hardware handshake
150 ... 19,200 baud;
1 or 2 stop bits; software/
hardware handshake
* = EC Directive No. 90/384/EEC on non-automatic weighing instruments * = EC Directive No. 90/384/EEC on non-automatic weighing instruments
applicable in the E.U. and within the European Economic Area applicable in the E.U. and within the European Economic Area
15
Specifications Specifications
Model BBC41 Model BBC51
Adaptation to ambient conditions By selection of one of Adaptation to ambient conditions By selection of one of
and application requirements 4 optimized filters and application requirements 4 optimized filters
Operating temperature range °C +10 ... +30 Operating temperature range °C +10 ... +30
Selectable application programs Mass unit conversion Selectable application programs Mass unit conversion
by toggling; tare memory; by toggling; tare memory;
net total; weighing in net total; weighing in
percent; animal weighing percent; animal weighing
Power requirements V~ Via AC adapter 230 VAC or Power requirements V~ Via AC adapter 230 VAC or
115 VAC, – 20% ... +15% 115 VAC, – 20% ... +15%
Frequency Hz 48 – 63 Frequency Hz 48 – 63
Power consumption (average) VA 16: maximum; Power consumption (average) VA 16: maximum;
8: average 8: average
* = EC Directive No. 90/384/EEC on non-automatic weighing instruments * = EC Directive No. 90/384/EEC on non-automatic weighing instruments
applicable in the E.U. and within the European Economic Area applicable in the E.U. and within the European Economic Area
16
73/23/EU “Electrical equipment designed for use
Declarations of Conformity within certain voltage limits”
Important Note:
The operator shall be responsible for any modifications to
the equipment and for any connections of cables or
equipment not authorized by the manufacturer and must
check and, if necessary, correct these modifications
and connections. On request, the manufacturer will provide
information on the minimum operating specifications
(in accordance with the Standards listed above for defined
immunity to interference).
17
18
Vea los dibujos que aparecen más abajo,
antes de comenzar con las instrucciones Instrucciones de instalación
Condiciones ambientales
La balanza está construida para que, al emplearse
en el laboratorio e industria bajo las condiciones normales
de funcionamiento, se obtengan resultados fiables de
pesada.
Por tal razón, en el lugar de instalación observe lo
siguiente:
19
Realizar la conexión de
balanza y caja electrónica Manejo de la balanza
para BBC 22
Insertar el conector del cable Tiempo previo de calentamiento: 30 minutos
en la hembrilla de conexión de
la caja electrónica. Encender y apagar (modo standby de funcionamiento)
Para encender y apagar la lectura active la tecla ON/OFF.
Montar cámara de pesada
Colocar las partes Autochequeo
en el siguiente orden: Al encender se realiza un
– anillo apantallador test automático de funciones de
– platillo de pesada la electrónica de la balanza.
El autochequeo finaliza con la
Conexión a la red lectura cero.
BBC 22: La balanza ya se encuentra lista
para el funcionamiento.
La alimentación de corriente
se realiza a través del adaptador Nota importante para modelos verificados
de CA. El voltaje impreso debe Para las balanzas verificadas con un escalón de
corresponder con la tensión verificación “e” mayor que el escalón real “d”, el último
de red local. dígito de la indicación se encuentra dentro de un marco.
Conecte el adaptador de CA
a la balanza.
Para su información, aparecen los siguientes símbolos
en el campo de indicación de la balanza:
20
Test de sensibilidad
Calibrar/ajustar
Selección de menú: 1 9 4
Al calibrar/ajustar se adapta la sensibilidad de la balanza
a las condiciones ambientales reinantes. Calibre su balanza Variaciones de consideración de la presión del aire y
en el lugar de instalación después de transcurrir el tiempo de la temperatura pueden influenciar el comportamiento de
previo de calentamiento, y antes de realizar la primera la balanza. Para garantizar también la exactitud total de
medición. En metrología legal, las balanzas han de ser lectura, al utilizar el rango de pesada total, tales variaciones
calibradas por lo menos una vez al dia. pueden controlarse mediante el test de sensibilidad.
La balanza le ofrece diferentes funciones de calibración, Esta función le facilita la decisión de realizar un nuevo
dependientes del código ajustado en el menú. ajuste, cuando es necesario (p. ej. series de medición
durante un largo intervalo de tiempo).
Utilización de las balanzas verificadas en ámbitos de
verificación obligatoria Quite la carga del platillo de pesada y tare.
Antes de la utilización de la balanza en ámbitos de
Con lectura cero, pulse la
verificación obligatoria, hay que llevar a cabo en el lugar
tecla CAL.
de instalación la función “calibración interna”, después
La pesa interna será aplicada
del tiempo de calentamiento
por motor.
Aparece CAL en la lectura.
Calibración/ajuste internos
Después de estabilizar, aparece
Selección de menú: 1 9 3 1) en la lectura la desviación del
valor de peso actual respecto al
La pesa de calibración es
valor debido (sólo en gramos).
aplicada por motor y retirada
Al producirse alguna distorsión
nuevamente al finalizar el
externa puede aparecer
proceso de calibración.
brevemente “Err 02”.
En este caso, tare nuevamente,
Con lectura cero, acceda a la
pulsando la tecla CAL.
función de calibración pulsando
la tecla CAL.
Advertencia:
Con interferencias durante La balanza ha de ser ajustada, si la desviación de cero
el proceso de calibración puede importa más que la repetibilidad.
aparecer brevemente el mensaje
de error “Err 02” en la lectura. La balanza ajusta automática-
Si éste fuera el caso, vuelva a mente con la pesa interna:
tarar pulsando nuevamente la Pulse brevemente la tecla CAL
tecla CAL.
o bien
Advertencia:
abandone la función test
Durante un proceso de calibración/ajuste,
de sensibilidad:
el platillo de pesada no debe tener carga puesta.
mantener pulsada la tecla CF
por más de 2 segundos.
1) = ajuste de fábrica en balanzas estándares con legibilidad de ≤ 0,1 mg
y uso en metrología legal Bloquear la función calibración/ajuste
La función calibración/ajuste puede ser bloqueada
Calibración/ajuste externos ajustando el código 1 9 7 (interruptor de bloqueo cerrado).
Selección de menú: 1 9 1 2)
21
Soluciones a imprevistos…
Problema … Causa … Solución
no aparecen segmentos – no hay suministro de corr. eléctrica – controle suministro eléctrico
en la indicación – adaptador de CA no está conectado – conecte el adaptador de CA
después del ajuste no aparecen – lugar de instalación de la – procure un lugar de instalación estable,
segmentos de indicación balanza es intranquilo, no se evite vibración de suelo, cierre protector
alcanza estabilidad interna contra corr. de aire.
en la lectura aparece “H” – capac. de pesada superada – descargue balanza
en la lectura aparece “L” – platillo de pesada o soporte – coloque platillo o soporte de platillo
o bien “Err 54” de platillo no están colocados (seg. modelo)
en la lectura aparece – tecla CAL al ajustar ha sido – active la tecla TARE y pulse
“Err 02” activada sin lectura cero nuevamente CAL
– la balanza está cargada – quite la carga de la balanza
en la lectura no desaparece – después de encender no se ha – pulse alguna tecla
la información especial “b” pulsado ninguna tecla
el resultado de pesada – lugar de instalación inestable – cambie de lugar de instalación
varía constantemente – mucha vibración, o bien, – realice adaptación en el
– hay corrientes de aire “menú funcionamiento”
– un objeto extraño roza entre platillo
y carcasa
– la tapa del dispositivo para pesar
por debajo de la balanza está abierta
– la muestra no tiene peso estable
(absorve humedad o se evapora)
– la muestra posee carga electrostática
resultado de pesada es – balanza sin ajustar – ajuste
obviamente falso – antes de pesar no se taró – tare antes de pesar
– burbuja de aire no está en – nivele
el centro del círculo del nivel
Advertencia:
Al utilizar tal dispositivo especial debería disponerse de un
apantallamiento contra las corrientes de aire.
22
– Confirmar ajuste
Menú de funcionamiento Para la confirmación del
de la balanza ajuste hay que pulsar la tecla
2 seg. TARE (identificación “o”
Los ajustes de fábrica del menú de funcionamiento tras el código).
de la balanza están identificados mediante asterisco “*”.
Las funciones no identificadas con “*” pueden ser – Para la memorización
ajustadas mediante la entrada del código correspondiente. del juste, la tecla TARE
Los códigos los encuentra Ud. a continuación bajo mantenerla pulsada
“Modo de pesada”. por más de 2 segundos
PRINT = saltar hacia la derecha paso a paso Función reset-anulación de las modificaciones
y hacia la izquierda, desde la tercera a la de ajustes
primera cifra La función reset posibilita la reintegración de todos los
ajustes de fábrica modificados que se identifican mediante
Acceso al menú (ejemplo: ajustar código 8 5 4) “*”. A esta función especial se accede ajustando
– Apagar balanza con ON/OFF. el código 9 – – 1°.
Encender balanza.
Durante la lectura de todos los Función reset Código
segmentos pulsar brevemente activada 9––1
tecla TARE. desactivada 9––2
– Soltar tecla TARE, cuando
en la lectura aparezca “|”
23
Condiciones tara Código** Símbolo/ Código
en todo momento 1 5 1 Factor cálculo
recién desp. de estabilidad 1 5 2*
1g= 1. nivel 2. nivel R1
Función auto-cero Código Quilates
activada 1 6 1* austr.** k (K)2) 5. 1 7 17 3 1 17
desactivada 1 6 2 Tolas t (tol)2) 0.0857333381 1 7 18 3 1 18
Bahts b (bat)2) 0.06578947436 1 7 19 3 1 19
Funciones de calibr./ajuste y Mesghales m (MS)2) 0.217 1 7 20 3 1 20
linealización con CAL Código
calibr./ajuste externos 1 9 1* * = ajuste de fábrica, a veces dependiente del modelo de balanza
** = excepto en metrología legal
calibr./ajuste internos para modelos 1) = excepto en balanzas verificadas, clase precisión k
con aplicación de pesa interna 1 9 3* 2) = caracteres impresos vía interfaz
test de sensibilidad para modelos
con aplicación de pesa interna 1 9 4
linealización interna sólo BBC22 1 9 6 Código
función calibr./ajuste bloqueada 1 9 7
Exactitud de lectura 1. rango 2. rango R1
Código
tecla FUNCTION bloqueada 2 1 1* Ajustes de los parámetros
conmutación de unidad 2 1 2
de la interfaz
Unidades de peso
Velocidad de transmisión en baudios Código
La unidad de peso básica es la unidad con la que
la balanza pesa al ser encendida. Esta unidad se ajusta 150 baudios 5 1 1
en el 1. rango. Las unidades del segundo rango pueden 300 baudios 5 1 2
ser seleccionadas ajustando el código individual 600 baudios 5 1 3
1200 baudios 5 1 4*
correspondiente.
2400 baudios 5 1 5
4800 baudios 5 1 6
Símbolo/ Código
9600 baudios 5 1 7
Factor cálculo
19200 baudios 5 1 8
1g= 1. nivel 2. nivel R1
Paridad Código
Gramos (o) o 1. 1 7 1 3 1 1 Paridad marca 5 2 1
Gramos g 1. 1 7 2* 3 1 2* Paridad espacio 5 2 2
Kilogramos1) kg 0.001 1 7 3* 3 1 3* Paridad impar 5 2 3*
Quilates** ct 5. 1 7 4 3 1 4 Paridad par 5 2 4
Libras** lb 0.0022046226 1 7 5 3 1 5
Onzas** oz 0.035273962 1 7 6 3 1 6 Cantidad de bitios de parada Código
Onzas Troy** ozt 0.032150747 1 7 7 3 1 7 1 bitio de parada 5 3 1*
Taeles 2 bitios de parada 5 3 2
Hongkong** tl (thl)2) 0.02671725 1 7 8 3 1 8
Taeles Tipo de acuse de recibo Código
Singapur** tl (ths)2) 0.02646063 1 7 9 3 1 9
Taeles Software-handshake 5 4 1
Taiwan** tl (tht)2) 0.02666666 1 7 10 3 1 10 Hardware-handshake
Granos** GN 15.43235835 1 7 11 3 1 11 con 2 caract. desp. de CTS 5 4 2*
Penny- Hardware-handshake
weight** dwt 0.643014931 1 7 12 3 1 12 con 1 caract. desp. de CTS 5 4 3
Miligramos mg 1000. 1 7 13 3 1 13 *
Piezas por * = ajuste de fábrica
libra** o (/lb)2) 1.1287667712 1 7 14 3 1 14
Taeles
China** tl (thlc)2) 0.02645547175 1 7 15 3 1 15
Momes** m
(mom)2) 0.2667 1 7 16 3 1 16
24
Servicios para la impresión Programas de aplicaciones
o la salida de datos Adicionalmente a las funciones de pesada, nuestra balanza
le ofrece otros programas útiles de aplicación.
Condiciones de salida de datos Código
Individual sin estabilidad 6 1 1 La selección de determinadas funciones o de un
Individual después de estabilidad programa de aplicación se realiza ajustando el código
– con función de memoria 6 1 2* correspondiente en el menú.
Individual en estabilidad
– sin función de memoria 6 1 3 En un ejemplo de trabajo se muestran ajustes adicionales
Automática sin estabilidad 6 1 4 para el desarrollo de programa con la documentación en
Automática en estabilidad 6 1 5 una tabla de códigos.
Autoprint Código Empleo de las balanzas verificadas
Automática mediante tecla PRINT en metrología legal
desconectable 6 2 1 Todos los programas de aplicación pueden ser
Automática no desconectable 6 2 2* seleccionados en metrología legal. Valores de pesada no
métricos están identificados con los siguientes símbolos:
Salida de datos en intervalos Código
1 cambio de indicación 6 3 1* Porcentaje =%
2 cambios de indicación 6 3 2 Cant. piezas = pcs
Valores cálculo =o
Tara automática después
de una salida de datos Código
Sin tara automática 6 4 1* Memoria de TARA NET
Con tara automática 6 4 2
Código 2 1 6
Salida automática de los
parámetros de aplicación Código El peso tara se memoriza pulsando la tecla FUNCTION.
Desactivada 7 1 1* La balanza es tarada automáticamente.
Activada 7 1 2 La pesada se realiza a continuación partiendo nuevamente
desde “cero”. Al utilizar la función memoria de tara,
Identificación de la salida de datos Código la asignación de memoria tara se indica, en la salida
Sin identificación 7 2 1* de datos o impresión con identificación “1”, a la derecha.
Con identificación 7 2 2 Ejemplo de trabajo “Tara – Neto – Bruto”
Salida automática datos de tara Código Ajustes en el ejemplo Código
Ultimo valor neto (val. indiv. N1) 7 3 1* Memoria tara 2 1 6
Contenido memoria (valor total T1) 7 3 2 Salida datos automática
de todos los parámetros 7 1 2
* = ajuste de fábrica – con identificación 7 2 2
Print memoria tara 7 3 2
* = ajuste de fábrica
25
Aplicación: Modificar valor porcentual de referencia
Documentación de valores Tara, Neto, Bruto El valor porcentual de referencia puede ser modificado
cuando se requiera.
Paso/tecla Lectura Salida de datos
Valores posibles para el porcentaje de referencia son
5, 10, 20, 50 y 100.
CF, TARE + 110.00 g
Colocar recipiente + 122.65 g Acceso a la función: Mantener pulsada tecla FUNCTION
FUNCTION + 110.00 g NET T1 + 22.65 g
Llenar con muestra, PRINT + 250.24 g NET N1 + 250.24 g
por más de 2 segundos,
CF, PRINT + 272.89 g N + 272.89 g hasta que en la lectura aparezca
“rEF 100 %”
La identificación designa
Ajustar: Pulsión breve de tecla FUNCTION
Memoria del valor: Mantener pulsada tecla FUNCTION
T1 + 22.65 g Peso memoria de tara (valor peso) por más de 2 segundos y el valor
N1 + 250.24 g Peso neto con memoria de tara asignada
N + 272.89 g Peso bruto = Tara + Neto se memoriza con seguro contra
los fallos de red
* = ajuste de fábrica
Este ajuste no se borra mediante reset: 9 – – 1°!
La identificación designa
26
Conteo Z Aplicación:
Conteo de piezas de igual peso masa
Código 2 1 4
Paso/tecla Lectura Salida de datos
El programa de conteo posibilita el cálculo del valor de peso Colocar recipiente + 22.65 g
en cantidad de piezas, en relación a un peso de referencia. CF, TARE + 20.00 g
El valor de peso indicado es memorizado como una
Poner 10 piezas
cantidad de piezas predefinida y fija (ajuste de fábrica:
10 pcs). eventualm, modificar
La cantidad de referencia, al iniciar, se indica cada cant. ref. p. ej. 20 piezas
vez como “rEF 10 pcs”, antes de definir la cantidad de piezas. ajustar cant. ref. 20
mantener pulsada
Modificar la cantidad piezas de referencia tecla FUNCTION por más
de 2 segundos rEF 10 pcs
La cantidad piezas de referencia puede modificarse
pulsar tecla FUNCTION rEF 20 pcs
cuando se requiera. Cantidades piezas de referencia
memorizar:
posibles son 5, 10, 20, 50 y 100. mantener pulsada
tecla FUNCTION por
Acceso a la función: Mantener pulsada tecla FUNCTION más de 2 segundos
por más de 2 segundos,
hasta que en la lectura aparezca Confirmar cant. piezas ref
“rEF … pcs” con tecla FUNCTION + 10 pcs nRef + 10 pcs
Ajustar: Pulsar brevemente tecla FUNCTION wRef + 5.6546 g
Poner la cantidad
Memoria del valor fijo de piezas deseada + 500 pcs
después de apagar PRINT + 500 pcs Qnt + 500 pcs
(seguro contra fallos
Borrar lectura de peso y
de red): Mantener pulsada tecla FUNCTION valor de referencia con CF + 2827.35 g
por más de 2 segundos PRINT + 2827.35 g N + 2827.35 g
Pesada de animales V
* = ajuste de fábrica
Código 2 1 7
Ejemplo de trabajo “Conteo de piezas”
Con este programa se calcula el peso de animales
Ajustes en el ejemplo Código
vivos o valores de peso en ambientes intranquilos.
Conteo 2 1 4 Aquí el peso se emite como valor promedio de una
Valor y peso de referencia 7 1 2 cantidad, a determinar, de pesadas individuales.
– con identificación 7 2 2
Durante la formación del valor promedio, en la lectura de
peso se indican permanentemente las pesadas individuales
que restan por realizar. Finalizadas las mediciones,
aparece el valor promedio calculado como valor estable
en la lectura de peso.
27
Inicio de la pesada de animales Código Ejemplo de trabajo:
Manual 3 8 1 “Pesada de animales con inicio automático”
Automático 3 8 2*
Ajustes en el ejemplo Código
Realizada la formación del valor promedio, el programa Pesada de animales 2 1 7
queda bloqueado, hasta que la balanza sea descargada Retardo mediano 3 7 2*
hasta la mitad de umbral de memorización (50 dígitos Inicio automático 3 8 2*
de indicación). Salida de datos automática
de todos los parámetros 7 1 2
* = ajuste de fábrica – con identificación 7 2 2
28
El protocolo puede constar de las siguientes líneas:
Protocolos GLP/GMP
------------------
Aplicación: Modelo BBC22 : Modelo de balanza
Empleo de la balanza en sistemas de aseguramiento N/S 020300189 : N° de serie de balanza
de la calidad, de acuerdo con ISO/GLP/GMP/EN, etc. Id : Espacio para la entrada Id
------------------
La balanza puede documentar todos los procesos Fecha: 15-Dez-97 : Fecha actual
de calibración ejecutados y también todos los valores Inicio: 10:05:30 : Hora inicio de la aplicación
de medición, de acuerdo con los requerimientos de las Cal. : Externa : Modo función de calibración
Prácticas Buenas del Laboratorio (GLP). En combinación (aquí “Calibración externa”)
con una impresora especial u ordenador se edita un Pesa : 200.0000 g : Valor pesa calibración (sólo
documento, el que permite desde la entrada de fecha, con “Calibr. externa”)
hora, N° de serie y designación de modelo, hasta un Fin : 10:05:45 : Fin de la aplicación
ordenamiento claro del valor de medición de la balanza Nombre : : Espacio para la firma
utilizada y hora de la emisión del valor. del operador responsable
------------------
La selección del protocolo GLP/GMP se realiza
ajustando el código correspondiente en el menú de Informaciones adicionales con otras funciones
funcionamiento de la balanza: de calibración:
29
Salida automática
de datos de referencia (parámetros) Código
Desactivada 7 1 1*
Valor referencia y peso referencia 7 1 2
* = ajuste de fábrica
30
Datos técnicos Datos técnicos
Modelo BBC22 Modelo BBC31 BBC32
Adaptación a las condiciones 4 escalas optimizadas Adaptación a las condiciones 4 escalas optimizadas
de empleo e instalación de filtro de empleo e instalación de filtro
Rango temperatura de empleo °C +15 ... +25 Rango temperatura de empleo °C +15 ... +25
Altura de la cámara de pesada (área útil) mm 253 Altura de la cámara de pesada mm 225
Consumo eléctrico (típico) VA máximo 18,7; típico 10 Consumo eléctrico (típico) VA máximo 16; típico 8
* = Directiva N° 90/384/CEE para balanzas de funcionamiento * = Directiva N° 90/384/CEE para balanzas de funcionamiento
no automático aplicable en los Estados Miembros de la no automático aplicable en los Estados Miembros de la
Comunidad Europea Comunidad Europea
31
Datos técnicos Datos técnicos
Modelo BBC41 Modelo BBC51
Adaptación a las condiciones 4 escalas optimizadas Adaptación a las condiciones 4 escalas optimizadas
ambientales y de aplicación de filtro ambientales y de aplicación de filtro
Rango temperatura de funcionamiento °C +10 ... +30 Rango temperatura de funcionamiento °C +10 ... +30
Dimensión del platillo de pesada mm Ø 115 Dimensión del platillo de pesada mm 180 x 180
Programas de aplicación seleccionables conmutación de unidades, Programas de aplicación seleccionables conmutación de unidades,
memoria de tara, total neto, memoria de tara, total neto,
pesada en porcentaje, pesada en porcentaje,
conteo, pesada de animales conteo, pesada de animales
Consumo eléctrico (típico) VA 16 máx., 8 típico Consumo eléctrico (típico) VA 16 máx., 8 típico
Interfaz de datos integrada RS 232 C-S/V24-V28; Interfaz de datos integrada RS 232 C-S/V24-V28;
7 bitios; paridad: par, 7 bitios; paridad: par,
marca, impar, espacio; marca, impar, espacio;
velocidad de transmisión velocidad de transmisión
150…19200 baudios, 150…19200 baudios,
1 ó 2 bitios de parada, 1 ó 2 bitios de parada,
acuse de recibo del acuse de recibo del
software/hardware software/hardware
* = Directiva N° 90/384/CEE para instrumentos de pesaje * = Directiva N° 90/384/CEE para instrumentos de pesaje
de funcionamiento no automático aplicable en los Estados Miembros de funcionamiento no automático aplicable en los Estados Miembros
de la Comunidad Europea de la Comunidad Europea
32
Directiva 73/23/CEE
Declaraciones de conformidad “Material eléctrico a utilizarse con determinados
límites de tensión“
La marca C puesta en los aparatos indica que éstos
Normas europeas aplicables:
cumplen con los requisitos de las siguientes Directivas del
Consejo de la Unión Europea: NE 60950 Seguridad de los equipos de tratamiento
de la información, incluyendo los equipos
eléctricos de oficina
89/336/CEE “Compatibilidad electromagnética (CEM)”
NE 61010 Requisitos de seguridad de equipos
eléctricos de medida, control y uso en
1. Compatibilidad electromagnética:
laboratorio
1.1 Fuente de 89/336/CEE: Diario Oficial de las Parte 1: Requisitos generales
Comunidades Europeas, N° 2001/C105/03
Al utilizarse equipo eléctrico en instalaciones, bajo
condiciones ambientales que requieren altas medidas
EN 61326-1 Material eléctrico para medida, control y uso
de seguridad, han de observarse las descripciones
en laboratorio
correspondientes para la instalación.
Requisitos CEM
Parte 1: Requisitos generales Inmunidad definida:
área industrial, funcionamiento continuo no Verificaciones ulteriores en los Estados de la
vigilado Comunidad Europea
Limitación de las fuentes de distorsión:
residencial, clase B La validez de verificación será reglamentada por los
mismos Estados en donde el instrumento de medición sea
utilizado. Es responsabilidad del usuario informar a tiempo
al organismo correspondiente.
Advertencia:
Modificaciones en los aparatos o la conexión
de cables o aparatos no suministrados por BOECO
es de responsabilidad del usuario y deben ser
controladas por él mismo y, en caso necesario, realizar
las correcciones pertinentes.
BOECO pone a disposición, según consultas,
informaciones sobre los datos mínimos del funcionamiento
de los aparatos (según las normas sobre inmunidad, más
arriba mencionadas).
33
34
35
Boeckel+Co
(GmbH+Co.) scientific equipment
Brandstwiete 4, 20457 Hamburg
t (040) 32 56 27 0
f (040) 32 56 27 41
E-Mail: [email protected]
98648-008-33
Publication No.: WXX6043-al03092
W4A000 Boeco BBC e/s p65 · KT