Chapter 1 Specifications
■ Model:XK3190-A19
■ Accuracy: Class III ,N=3000
■ A/D Conversion theory & sample rate: adopt ∑-∆, 20~40times/sed.
■ Simulating Voltage:DC+5V signal range:-2mV~13mV
■ Load cell:1~4 350 ohm
■ Display(EL backlight) : 6 bits LCD with 0.8 inch character height display
6 status indicating light
■ Keyboard: Slight keyboard with 6 keys
■ Power supply: AC 187~242V; 49~50HZ
Built in rechargeable battery,6V/4AH (If you connect one load cell, indicator can
work 100 hours while backlight is off)
■ fuse:500mA
■ Pre-heat time:15minutes
■ Operation Temp/RH:0~40℃ ;≤ 85%RH
■ Storage/Transportation Tem:-25℃ ~55℃
■ Dimension:260×190×170mm; Weight:2.2KG
■ Optional: RS232 , connected to PC or scoreboard, transmission distance ≤ 15m
A19-5628 V1.0 3
Chapter 2 Installation and Keypad
2-1 FRONT AND BACK VIEW OF THE INDICATOR
Front View
Back View
A19-5628 V1.0 4
2-2 CONNECTING LOAD CELL TO INDICATOR
The 9-pin socket is used for the connection of load cell, as shown below:
-V SHIELD
The below graph clearly shows the connection of load cell and indicator
1 2 3 4 5
6 +V +
BRIDGE OF POWER SUPPLY
+V -IN -IN _
1 -V
9 +IN +
SIGNAL OUTPUT
_
8 -I N
5 SHIELD
■ Warning: Connection of indicator and load cell must be reliable and the operation should be in
power off statement.
■ Warning: After connection, the 9-pin socket must be fixed to ensure the security. No plugging
or pulling the connection freely. And be sure the connection must be operated under power off
statement.
■ Warning: As both the load cell and indicator are sensitive to the static electricity, anti-static
measures must be taken. No soldering operation or other strong electricity operation is
permitted. And anti-thunder measurements also should be taken to avoid any injury to the
devices or the operators.
2-3 KEY
ZERO KEY
Indicator will come to zero if it is within the manual zero setting range, the ”Zero” light will be on; if
it is without this range,it will be invalid.
TARE KEY
The current weight value will be regarded as tare weight, the “Tare” light will be on ; Repress
[Tare] ,it will come into the gross weighing status and the “Tare” light will be off.The maximum
A19-5628 V1.0 5
value for tare is 99999.Also used as confirm key while setting parameter and calibration.
M+ KEY
The current weight value will be added to the accumulated weight value when pressing this key.
The “Accum” light will be on.After about 2 sencods showing the accumulated weight value,it will
come to show the accumulated number for about 2 seconds.Then the “Accum” light will be off,the
indicator return to the weighing status.
CH KEY
Press this key,the display value will show the value prior to the selected division(exception of
division 1).That is to say,if the selected division is 2,then it will show the value with division 1;if the
selected division is 5,then it will show with division 2;If the selected division is 10,it will show with
division 5….But if the selected division is 1,press this key won’t make any change.The “CH” light
will be on,repress this key,the “CH” light will be off and show the valued with the selected
division.Also used as selection key or enter key while setting parameter.
CLEAR KEY
Press this key,it will show[C-Add] for about 2 seconds,the accumulated value and number
will be cleared out,then it returns to the weighing status.
COUNT KEY
Press this key,the “Count” light will be on,it will come into the counting status.Repress this
key,the “Count” light will be off,it returns to the weighing status. er
CLEAR + M+ KEY
Press the combined key of this,it will show the inner code of indicator.The full value of the
inner code is 20 times of the selected counts.For example,the counts is 3000,then the inner
code will show 0.6.0.0.0.0 marking with 5 decimal points flashing.Repress the combined
key,it will returns to the weighing status.
Press the combined key of this,it will come into the parameter setting status
Press the combined key of this,it will come into the calibration status
A19-5628 V1.0 6
M+ + ZERO KEY
Press the combined key of this,it will come into the percentage display status
CH + ZERO KEY
Press the combined key of this,it will come into the high/low limit alarming setting
2-4-2 The communication method is continuous sending, the data format is:7 bits for weight
value(low data in front of high data ,1 bit for sign.)
For example:
A19-5628 V1.0 7
Chapter 3 Calibration
3-1 Connect load cell properly, then turn on the indicator, it will perform self-checking, then enter
into the weighing statement.(15~30 minutes pre-heat is necessary to assure excellent
performance)
3-2 Put the jumper inside the indicator at ”CN2” location(the default location is just “CN2”),then
calibrate as shown table:3-2-1
Table 3-2-1
STEP OPERATION DISPLAY NOTES
[CLEAR] + [ZERO] Enter the calibration, notes for inputting the
1 [C 000] password
A19-5628 V1.0 8
Press[TARE] to confirm
Example:012
Press [CH] FOR COUNTS Select counts optional(press[CH] for selection:
SELECTION ,Press[TARE] to [n 3000] 1000/2000/2500/3000/4000/5000/6000/7500/100
9 confirm 00/12000/15000/20000/25000/30000
Exmaple:3000
Table 3-2-2
W 0 1 2 3
Zero track speed 0.4second 0.3second 0.2second 0.1second
Table 3-2-3
X 0 1 2 3 4 5 6 7 8 9
Zero track No 0.5e 1.0e 1.5e 2.0e 2.5e 3.0e 3.5e 4.0e 4.5e
range track
Table 3-2-4
Y 1 2 3 4 5
manual zero setting range 2%F.S 4%F.S 10%F.S 20%F.S 100%F.S
Table 3-2-5
Z 1 2 3 4 5
auto zero setting range 2%F.S 4%F.S 10%F.S 20%F.S 100%F.S
A19-5628 V1.0 9
Z filter intensity: the high value it is, the more filter intensity and the lower response it is.
4-2 ZERO-SETTING
When the gross weight is within the zero setting range, press[zero],indicator will perform zero.
When tare is applied, pressing [zero]won’t effect.
4-3 TARE
When the value is stable, not negative, and without the zero range, press[TARE],indicator will
perform the corresponding operation, the “Tare” light will be on. Repress[Tare],it will quit the net
display mode.Then the “Tare” light will be off.The maximum tare value is 99999
A19-5628 V1.0 10
seconds)
[******] Return to weighing status,”Accum” light
will be off
4 Press[CLEAR] [CA----]
Note: 4-5-1
X:0 stands for the buzzer speak;1 stands for the buzzer doesn’t speak
Y:0 stands for check the fixed value whether stable or not;1 stands for check the fixed value while
it is stable
A19-5628 V1.0 11
Z:0 stands for alarm while it is less than the low limit;1 stands for alarm while it is between the
high limit and low limit;2 stands for alarm while it is more than the high limit;3 stands for alarm
while it is less than the low limit or more than the high limit
Table:4-6-1
Step Operation Display Notes
In weighing Low limit setting: indicating for low limit
status,press[CH]+ inputting
[L ***** ]
1 [ZERO],it will enter into
the high/low limit
setting
Input the low limit by If keep the original setting, just
[M+] and [CH],then [L 01000 ] press[TARE] to step 3,or press[ZERO]
2
press[TARE] to confirm directly to step 5,Note 4-6-1
Example:1000
[H ***** ] High limit setting: indicating for high limit
3
inputting
Input the high limit by If keep the original setting, just
[M+] and [CH],then [H 03000 ] press[TARE] to step 5,or press[ZERO]
4
press[TARE] to confirm directly to step 5,Note 4-6-1
Example:3000
Return to the weighing
5 [******]
status
Note 4-6-1:
If the low limit is 0,any value the high limit it is,the alarm function will be forbidden
Note 4-6-2:
XK3190-A19 has two methods of alarming:Single fixed value alarming and double fixed value
alarming.In single fixed value alarming,the high limit should be set 0,the alarming method should
be 0(under low limit) or 2(over high limit);While in double fixed value alarming,the high limit
should be over than the low limit.
A19-5628 V1.0 12
Attention:When the weigh value is zero,any alarming method won’t work.
Note:
(1) The sample quantity shouldn’t be too less.It is recommended over 50~100.
(2) When the weight value is negative,counting operation is invalid.(The counting number is
always 0).If the unit weight of the sample is less than 0.25 divison,it will show[Err 15] to show
sample is too less.But the counting operation can still continue.If the unit weight of the sample
is less than 0.05 divison,it will show[Err 10] to show sample is seriously less.The setting for
the sample quantity is invalid.Set the sample quantity again.
(3) In counting mode.operation of tare,zero,M+ is disabled.
(4) In the counting mode,if the sample quantity isn’t set,then counting operation will work
according to the last sample quantity setting.
Note:When the sample weight is less than 50 outer code,it will show[Err-12] to show the sample
isn’t enough.Press [ZERO] to quit from the percentage mode to weighing status.
A19-5628 V1.0 13
Turn off the power switch button to power off the indicator.To completely stop the
indicator from working,just pull off the AC wire.
For details of the error information, please refer to the following table:
[Err 12] Sample is too less for percentage Assure the sample weight is more
than 50 outer code
[Err 23] Error from memory First check the memory on the main
board and its concerned circuit to see
wheter it’s shorted or opened.If it’s
ok,then replace the memory
A19-5628 V1.0 14
Attention:
A. Red wire stands for positive while black wire stands for negative.
B. The first time charge for the battery, please take more than 24 hours to assure the security and
stability of the battery
2. When the low battery light is on, the battery should be charged soon. Under the statement of
low battery, the indicator can continue to work more than 10 minutes until it shows “bAttLo”.
Chapter 7 Maintenance
7- 1 To guarantee indicator clarity and using life, the indicator shouldn’t be placed directly under
sunshine and should be set in the plain space.
7- 2 The indicator can’t be placed into the place where the dust pollution and vibration are
serious.
7- 3 Load cell should connect with indicator reliably, and the system should be well connected
into ground. The indicator must be protected from high electrical field and high magnetic field.
■ In order to protect the operator, indicator and relevant device, you should mount lightning rod in
7- 4 It is strictly forbidden to clean the case of indicator with intensive solvents (for example:
benzene and nitro oils)
7- 5 Liquid and conducting particle should not be poured into the indicator, otherwise the
electronic components will be damaged and electric shock is likely to happen.
7- 6 You should cut off power supply of indicator and relevant device before you pull-in and out
the connecting line of indicator and external device.
■ You must cut off power supply of the indicator , before you pull out connecting line of the load
cell.
7- 7 During operation, if trouble occurs, operator must pull off the power supply plug immediately,
and user should return this indicator to our company for repair. Non-weighing manufacturer
should not repair it, or by yourself, otherwise further destruction may happen.
A19-5628 V1.0 15
7- 8 The storage is not granted the free repair guarantee, because it is easily exhausted products.
■ In order to prolong using life, please charge the cell fully before using it.
If you don’t use the indicator for a long time, you must charge the cell every two month and for
eight hours/each charging time.
■ Moving or installation must be carefully taken and must avoid strong vibration, impact and bump
7- 9 From invoice date, the indicator has a one-year free repair period. If any non-artificially
obstacle about the indicator happens under correct using conditions within this period, the user is
allowed to send the product with its guarantee card (of the correct number) back to our
corporation for free repair. The indicator shouldn’t be taken apart, otherwise free guarantee will
be cancelled.
A19-5628 V1.0 16