Acuson X300: System Voltage Test Points

Download as pdf or txt
Download as pdf or txt
You are on page 1of 14

ACUSON X300

US

Troubleshooting Guide
System
Voltage Test Points

© Siemens AG 2006
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
Siemens Medical Solutions
lne
Ultrasound
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: US11-107.840.01.01.02 English


Replaces: n.a. Doc. Gen. Date: 09.06
n.a.

Part No.:
sd
2 Revision / Disclaimer
1Revision / Disclaimer

Copyright
Copyright © 2006 by Siemens Corporation. All rights reserved. No part of this publication
may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into
any language or computer language, in any form or by any means, electronic, mechanical,
magnetic, optical, chemical, manual, or otherwise, without the prior written permission of
Siemens Corporation. Siemens Corporation reserves the right to change its products and
services at any time. In addition, this manual is subject to change without notice. Siemens
Corporation welcomes customer input on corrections and suggestions for improvements
to this manual.Although Siemens Corporation has attempted to ensure accuracy through-
out this manual, Siemens Corporation assumes no liability for any errors or omissions, nor
for any damages resulting from the application or use of this information.
Trademarks
ACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE,
SONOVISTA, Adara, Antares, G20, G40, G50, G60S, X300, X500, Omnia, Premier and
KinetDx are trademarks of Siemens Corporation registered in the U.S. Patent and Trade-
mark Office. CWS3000, DS3000, WS3000, and WebPro are trademarks of Siemens Cor-
poration.
Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet
Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are
registered trademarks of Netscape Communications Corporation. Java and the JavaScript
name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered
trademark of Adobe Systems, Inc.
Other products and brand names are trademarks of their respective owners.
Disclaimer
The service of equipment described herein is to be performed by qualified personnel who
are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie-
mens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
License Agreement
All computer programs copyright 1990-2006 by Siemens Corporation or its suppliers. Pro-
grams are licensed under the following agreement:
Siemens or its suppliers retain(s) ownership of and title to any computer program supplied
with the equipment and to the trade secrets embodied in such computer programs. Sub-
ject to the Buyer’s acceptance and fulfillment of the obligations in this paragraph, Siemens
grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any
computer program supplied with the Equipment that is necessary to operate the Equip-
ment solely on the medium in which such program is delivered for the purpose of operat-
ing the equipment in accordance with the instructions set forth in the operator’s manuals
supplied with the Equipment and for no other purpose whatsoever. Buyer may not
reverse-assemble, reverse-compile or otherwise reverse-engineer such computer pro-
grams nor may Buyer make a copy of such program or apply any techniques to derive the

ACUSON X300 US11-107.840.01.01.02 Page 2 of 14 Siemens


09.06 sd Medical Solutions USA, Inc.
Revision / Disclaimer 3

trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms
of this license, the license granted by this paragraph shall terminate. Further, because
unauthorized use of such computer programs will leave Siemens without an adequate
remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to
restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s
suppliers of software is a direct and intended beneficiary of this end-user sublicense and
may enforce it directly against Buyer with respect to software supplied by such supplier,
and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indi-
rect, consequential, incidental or other damages arising out of the sublicense of the com-
puter programs supplied with the equipment.

Siemens US11-107.840.01.01.02 Page 3 of 14 ACUSON X300


Medical Solutions USA, Inc. 09.06 sd
4 Table of Contents
1- 0Table of Contents

1 _______ Voltage Test Points ______________________________________________ 5

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
BE Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
I/O Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
RC Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
TR Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
AC Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
DC Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

2 _______ Changes to Previous Version _____________________________________ 13

Changes to Previous Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

ACUSON X300 US11-107.840.01.01.02 Page 4 of 14 Siemens


09.06 sd Medical Solutions USA, Inc.
Voltage Test Points 5
2-
Overview
1Voltage Test Points

The Voltage Test Point section provides the test point locations of each board, power sup-
ply and their corresponding values.

BE Board 0

Fig. 1: BE Board Test Points

Siemens US11-107.840.01.01.02 Page 5 of 14 ACUSON X300


Medical Solutions USA, Inc. 09.06 sd
6 Voltage Test Points
Tab. 1 BE Board Test Points

Reference Title Test Points Function


Number
TP101 PD3_3V +3.3V Supplying power for Digital ICs
(FPGA, ASIC)
TP100 PD2_5V +2.5V Supplying internal power for
FPGA, ASIC
TP108 PD5_0V_PC +5.0V FOR ETX Supplying power for ETX
POWER
TP17 PD1_8V +1.8V Supplying internal power for
FPGA, DSP
TP106 PD5_0V +5.0V Supplying power for Audio DAC
TP109 PA12_0V +12.0V Supplying power for Audio ICs
TP110 MA12_0V -12.0V Supplying power for Audio ICs
DS1 DS1 Emitting green lights Checking BE Board function

ACUSON X300 US11-107.840.01.01.02 Page 6 of 14 Siemens


09.06 sd Medical Solutions USA, Inc.
Voltage Test Points 7

I/O Board 0

Fig. 2: I/O Board Test Points


Tab. 2 I/O Board Test Points

Reference Title Test Points Function


Number
TP 2 PD3_3V +3.3V Supplying power for Digital ICs
TP 5 MA12_0V -12V Supplying power for Audio ICs
TP 6 PD5_0V_PC +5V FOR HDD / Supplying power for HDD, CD-R/W
CDRW through IO
TP 7 PD12_0V_PC +12V FOR HDD / Supplying power for HDD, CD-R/W
CDRW through IO
TP 3 PD5_0V +5.0V Supplying power for Digital ICs
TP 4 PD12_0V +12V Supplying power for Audio ICs

Siemens US11-107.840.01.01.02 Page 7 of 14 ACUSON X300


Medical Solutions USA, Inc. 09.06 sd
8 Voltage Test Points

RC Board 0

Fig. 3: RC Test Point


Tab. 3 RC Board Test Points

Reference Title Test Function


Number Points
H14 P3-3A_RX_AD +3.3V Supplying power for regulator generating
2.5V source to RXBF
H24 P3-3A_RX +3.3V Supplying power for ADC reference in
RXBF

ACUSON X300 US11-107.840.01.01.02 Page 8 of 14 Siemens


09.06 sd Medical Solutions USA, Inc.
Voltage Test Points 9

TR Board 0

Fig. 4: TR Board Test Points


Tab. 4 TR Board Test Points

Reference Title Test Function


Number Points
P3_3_AR0 P3_3_AR0 +3.3V Supplying power for AR-ASIC ampli-
(P3_AQ0 (P3_AQ0) fier (Channel 0~31)
P5A_0 (P5AQ0) P5A_0 +5.0V Supplying power for TX pulser (Chan-
(P5AQ0) nel 0~31)
M2_AR0 M2_AR0 -2.0V Supplying power for TX pulser (Chan-
(M2_0AQ0) (M2_0AQ0) nel 0~31)
P3_3D P3_3D +3.3V Supplying power for FPGA, TXPG and
(P3_3AQ2) (P3_3AQ2) other circuits
P5D (P5AQ2) P5D (P5AQ2) +5.0V Supplying power for TXPG
P5A_1 (P5AQ1) P5A_1 +5.0V Supplying power for TX pulser (Chan-
(P5AQ1) nel 32~63)
P3_3_AR1 P3_3_AR1 +3.3V Supplying power for AR-ASIC amplifier
(P3_3AQ1) (P3_3AQ1) (Channel 32~63)
M2_AR1 M2_AR1 -2.0V Supplying power for TX pulser
(M2-0AQ1) (M2-0AQ1)

Siemens US11-107.840.01.01.02 Page 9 of 14 ACUSON X300


Medical Solutions USA, Inc. 09.06 sd
10 Voltage Test Points

AC Power 0

Fig. 5: AC Test Points


Tab. 5 AC Power Test Points

Reference Title Test Points Tolerance


Number
TP1 48V Output +48.0V ± 1V
TP2 GND +48.0V ±1V
TP3 12V Output +12.0V ± 0.3V
TP4 12 Return +12.0V ± 0.3V
TP5 OEM Output +100~120V or ± 10V
+200~240V
TP6 OEM GND +100~120V or ± 10V
+200~240V
TP7 5.2V Battery +5.2V ± 0.3V
TP8 12V Battery +12.0V ± 0.3V

ACUSON X300 US11-107.840.01.01.02 Page 10 of 14 Siemens


09.06 sd Medical Solutions USA, Inc.
Voltage Test Points 11

DC Power 0

Fig. 6: DC Test Points

Reference Title Test Point Value Tolerance


Number
TP1 P2.5D +2.5V ± 5%
TP2 P5D_PC +5.1V ± 5%
TP3 P5.0D +5.0V ± 5%
TP4 P3.3D +3.3V ± 5%
TP5 M12D -12.0V ± 5%
TP6 P12D +12.0V ± 5%
TP7 P12D_PC +12.0V ± 5%
TP8 PSRTRXVO 0 ~ 18.0V ± 5%
TP9 M100HV -100.0V ± 5%
TP10 P100HV +100.0V ± 5%

Siemens US11-107.840.01.01.02 Page 11 of 14 ACUSON X300


Medical Solutions USA, Inc. 09.06 sd
12 Voltage Test Points

Reference Title Test Point Value Tolerance


Number
TP11 M14AQ -14.2V ± 5%
TP12 P14AQ +14.2V ± 5%
TP13 M2AG -2.0V ± 5%
TP14 P5AQ +5.0V ± 5%
TP15 P3.3AQ +3.3V ± 5%
TP16 GND NA
TP17 P3.3A +3.3V ± 5%
TP20 P5.0V-STBY +5.2V ± 0.5V

ACUSON X300 US11-107.840.01.01.02 Page 12 of 14 Siemens


09.06 sd Medical Solutions USA, Inc.
Changes to Previous Version 13
3-
Changes to Previous Version
2Changes to Previous Version

Change(s) Description
New document Initial release for ACUSON X300.

Siemens US11-107.840.01.01.02 Page 13 of 14 ACUSON X300


Medical Solutions USA, Inc. 09.06 sd
14 Changes to Previous Version

ACUSON X300 US11-107.840.01.01.02 Page 14 of 14 Siemens


09.06 sd Medical Solutions USA, Inc.