Sterilizers Autoclave
Sterilizers Autoclave
Sterilizers Autoclave
March 2009
By Hani al Meer
Introduction
n today's world, the need for infection control has never been greater .The patients and the
health care professionals have become more concerned about the transmission of pathogenic
organism then ever before.
Controlling microbial contamination through sterilization has long been considered the most
essential component of an infection control program. The result of proper instrument sterilization
is the protection of the patient, and the health care professional from various infection diseases.
The focus on infection control has sharpened.
Since infections increase the severity of illness, complicate recovery, and prolong inpatient stays
they raise the cost of care.
The ever growing expense of health care has focused providers' attention on reducing costs. Often
this means re-sterilizing and re-using expensive plastic items that might otherwise be discarded.
Instruments must be sterilized between usage.[1]
Until recently, the most advanced sterilization technology was found only in large hospital
sterilizers. Today there is a growing demand for more sophisticated sterilization technologies in
veterinary, dental and private clinics as well as tattoo parlors and beauty salons. [3]
There are several types of sterilization equipment. Dry heat sterilizers, heated chemical vapor
sterilizers, Steam sterilizers (autoclaves), and gas sterilizers.[1]
Dry Heat
The dry heat sterilizers employ higher temperatures to destroy microorganisms. They are
considered effective and safe for metal instruments because the process does not dull instrument
edges or rust/corrode the instruments. There are two kinds of 'Dry heat sterilizers'. The traditional
'toaster oven' style that is slow and take an hour to sterilize at 320f (160c), or the COX Rapid Heat
sterilizer which is using a patented forced air and baffle system technology to sterilize in 6
minutes at 375f (190c). CDC guidelines calls for weekly monitoring the working of the dry heat
sterilizer by doing a weekly spore test.
Chemical Solution
Liquid disinfectants, such as glutaraldehyde, are widely used in health care facilities. They have
short, inexpensive disinfection cycles. However, glutaraldehyde can not be used for all medical
materials. Typical cycle times and concentrations only provide disinfecting, which can fail to kill
resistant microorganisms. The solutions are highly corrosive and toxic; since they come in liquid
form, they can not be used with barrier packaging. The moment an instrument is removed from
the liquid, its sterility is compromised. Immersion in a liquid sterilant is not recommended
because sterilization by liquid chemicals can not be monitored biologically. In addition,
instruments disinfected by liquids must be handled aseptically, rinsed in sterile water, and dried
with a sterile towel. Furthermore, instruments immersed in liquid sterilants are not wrapped and,
therefore, must be used immediately or stored in a sterile container, according to the American
Dental Association Councils on Scientific Affairs and on Dental Practice's "Infection Control
Recommendations for the Dental Office and the Dental Laboratory," presented at the 1996 OSAP
Scientific Forum.
Glass bead sterilizer have long been employed for instruments re-used on the same patient at a
single appointment; however, they are not suitable for terminal sterilization of instruments prior
to re-use on other patients. The reason they are not effective is that there isn't a way to test the
temperature throughout the cycle. In fact, glass bead sterilizers are not FDA approved, and any
new machines must receive pre market approval, according to Dr. Janie Fuller with the infection
control devices branch of the U.S. Food & drug Administration.[1]
An autoclave is a pressurized device designed to heat aqueous solutions above their boiling point
at normal atmospheric pressure to achieve sterilization . It was invented by Charles Chamberland
in 1879.[2]
And it is also used to describe an industrial machine in which elevated temperature and pressure
are used in processing materials.[2]
Under ordinary circumstances (at standard pressure), liquid water cannot be heated above
approximately 100 °C/212 °F (99.99 °C at 101.325 kPa, 99.62 °C at 100 kPa) in an open vessel
(see here for special situations ). Further heating results in boiling , which is the transition from
liquid to gas, but does not raise the temperature of the liquid water. However, when water is
heated in a pressurized vessel such as an autoclave, it is possible to heat liquid water to a much
higher temperature. As the container is heated the pressure rises due to the constant volume of the
container (see the ideal gas law ). The boiling point of the water is raised because the amount of
energy needed to form steam against the higher pressure is increased. [2]
Steam sterilizers will allow you to sterilize a wide range of materials. Temperature, pressure, and
time are the main factors of this process. The higher the temperature, the greater the pressure and
the shorter the sterilization cycle. Maximum temperature will rise to 273f (133c). Steam
sterilization , however, may cause unprotected instruments to corrode and unprotected cutting
edges to dull. Furthermore, excess water in the steam can provide a portal for microorganism to
penetrate wet instrument packages. Steam sterilizers (autoclaves) are the most popular method of
sterilization and are made and sold throughout the world. CDC guidelines calls for weekly
monitoring the working of the steam sterilizer by doing a weekly spore test.[1]
Uses
Autoclaves are widely used in microbiology, medicine, sterilizing instruments for body piercing ,
veterinary science, dentistry, podiatry and metallurgy. The large carbon-fiber composite parts for
the Boeing 787, such as wing and fuselage parts, are cured in large autoclaves.[2]
Autoclave is also used - as well as - in tattoo parlors and beauty salons .[3]
Non-sterilization applications
Many other systems use hot water or steam and Priorclave autoclaves have been adapted for the
following applications: food processing (simultaneously cooking and shrink wrapping
vegetables); plastics (re-hydration of molded nylon components); accelerated curing; construction
(testing conditions for curing roof tiles) .[7]
Types of autoclave
Front loading (cylindrical chamber or rectangular chamber from 40 liter to 700 liter)
Top loading (cylindrical chamber from 60 liter to 200 liter)
Pass through or double-ended (so that load enters in one room and is taken out when
autoclaved via other door in an adjoining
room). Cylindrical chambers of 150 or 200
liter and rectangular chambers of 350, 450
or 700 liter).[6]
Sterilizer chamber
The chamber size differs from one to other by type , uses , needs , and
stainless steel chamber
the location of the sterilizer .
The chamber should also be provided with one or tow doors and
interlocks that make opening the sterilizer door(s) during the
cycle is not possible .
Evacuation system
When the goal of autoclaving is to achieve sterility, it is very important to ensure that all of the
trapped air is removed. The reason for this is that hot air is very poor at achieving sterility. Steam
at 134 °C can achieve in 3 minutes the same sterility that hot air at 160 °C takes two hours to
achieve. Autoclaves may achieve air removal by various means including:
Downward displacement (or gravity type) - As steam enters the chamber, it fills the
upper areas as it is less dense than air. This compresses the air to the bottom, forcing it
out through a drain. Often a temperature sensing device is placed in the drain. Only when
air evacuation is complete should the discharge stop. Flow is usually controlled through
the use of a steam trap or a solenoid valve, but bleed holes are sometimes used, often in
conjunction with a solenoid valve. As the steam and air mix it is also possible to force out
the mixture from locations in the chamber other than the bottom.
Steam pulsing - Some autoclaves remove air by using a series of steam pulses, in which
the chamber is alternately pressurized and then depressurized to near atmospheric
pressure.
Vacuum pumps - Some autoclaves use vacuum pumps to suck air or air/steam mixtures
from the chamber.
Superatmospheric - This type of cycle uses a vacuum pump. It starts with a vacuum
followed by a steam pulse and then a vacuum followed by a steam pulse. The number of
pulses depends on the particular autoclave and cycle chosen.
Subatmospheric - Similar to superatmospheric cycles, but chamber pressure never
exceeds atmospheric until they pressurize up to the sterilizing temperature.[2]
Aeration system
When the sterilizer chamber is ventilated during the aeration for the purpose of desorption and to
release the vacuum at the end of the process, microbial recontamination of the sterilization load shall be
prevented by a filter (see Annex A).
Temperature
Measuring devices
o When used for process control, monitoring or registration purposes, temperature
measuring devices
o shall have maximum permissible errors of 1 K or less over the scale range 20 °C to 100
°C and be adjusted to± 0,5 K or less at the sterilization temperature.
o Temperature control systems shall be protected against unauthorized operation by the use
of an access device.
o At least two independent sensors shall be used for the measurement of the chamber
temperature,
being dedicated to indication, control and registration. The sensors shall be located in
the most representative position in the chamber specified as being the reference
measuring point.
o The registration and monitoring/controlling devices shall be mutually independent as
given by any of
the arrangements shown in Figure 3.
o The temperature sensor including the measuring system it is connected to shall have a
response time
τ0,9 < 5 s when tested in flowing water according to EN 60751.
o Temperature control systems shall have a function which causes a fault to be indicated if
a sensor fails.
Indicating devices
In addition to these indicators, autoclaves have timers, temperature and pressure gauges that can
be viewed from the outside.
There are certain plastics that can withstand repeated temperature cycling greater than the 121 °C
or 249 °F required for the autoclaving process. PFA, polypropylene, polysulfone and Noryl are
examples.
Some computer-controlled autoclaves use an F0 (F-naught) value to control the sterilization cycle.
F0 values are set as the number of minutes of equivalent sterilization at 121 °C or 249 °F (e.g.: F0
= 15 min.). Since exact temperature control is difficult, the temperature is monitored, and the
sterilization time adjusted accordingly.[2]
Autoclave load thermocouple assemblies are specially designed to provide the best possible
accuracy and stability in the rigorous environment of sterilizer chambers.
The proven, Labfacility design has been subjected to successful, thorough evaluation by CMI
Healthcare including leak testing in a vacuum chamber and processing through normal, porous
load sterilization cycles.
The sensor complies with test standard HTM 2010, clause 6, 19/6.21.
Harsh conditions in the autoclave subject the thermocouple to high levels of pressure and
moisture.
A special multiple seal construction in this design of autoclave thermocouple ensures that
moisture ingress is prevented, even at elevated pressure.
The combination of a type T thermoelement and exceptionally rugged construction ensure high
accuracy, high stability and long term reliability.
Ordinarily, in order to achieve the sort of accuracy demanded for such an application, typically +/-
1C or better, the sensor would utilize a platinum resistance thermometer.
However, such a sensor would be prone to damage in harsh physical conditions and also be
particularly vulnerable to deterioration due to the ingress of moisture if any seals were damaged.
This design of type T thermocouple assembly uses close tolerance, selected thermocouple wire to
achieve the high accuracy and stability needed for sterilizer applications but combined with the
ability to withstand arduous conditions and handling.
The sensor is protected by a 316 grade stainless steel sheath and bulkhead tube: strong, flexible
stainless steel armour conduit protects the extension lead with an additional length of silicone
rubber exiting the bulkhead tube.
By locating the sensor in close proximity to the 'load' inside the autoclave, an accurate measure of
the temperature applied to the actual product is obtained as opposed to simply monitoring the
internal temperature of the autoclave in the chamber wall; control of the autoclave temperature is
typically based on such a location.[5]
Pressure
Measurement devices
Recording instruments
o The recorder shall be independent such
that the measuring chain as well as value
data processing and printed values are
separate from the automatic controller.
NOTE 1 This does not exclude the transfer of
informative data from the automatic controller to
the recorder and vice versa , via a combined
system for data transfer.
NOTE 2 Sterilizer identification, cycle number and
load identification may be recorded automatically.
o For operational inspection as well as for batch documentation, analogue or digital
recorders shall record pressure and temperature data versus time. The records shall allow
evaluation of the data for compliance with specified temperature and pressure profiles
throughout the operating cycle .
Services Required
Electrical Supply
380/415 volts, 50 Hz, three phase rated at 15 amps per phase with earth and neutral or
240 volts, 50Hz, single phase rated at 30 amps with earth and neutral terminated at a
suitably fused switched socket RS Components Stock Number 367 4086 within 2 meters
run of the rear of the Priorclave.
Steam supply (if required)
Not required, but you can ask for one if you want more service.
Water Supply
Not required, manual fill. If the Auto Water Filling option is supplied then a 15mm cold
water supply terminated at a stopcock within 2 meters run of the rear of the Priorclave
should be supplied.
Drain Service
A 35mm sealed drain with an untapped 22mm entry, capable of withstanding steam at
136 ° C, with a constant fall to waste, vented at a high level outside of the building to
satisfy the requirements of BS 2646 Part 2 1990. A separately trapped, 22mm drain
should be provided for the Auto Water Fill tank. All drains should be provided at a low
level within 2 meters run of the rear of the Priorclave.[6]
Other considerations
Vertical power doors are an option on front loading rectangular section autoclaves from
230 to 700 liter sizes
Load-lite lifting hoist - helps out with heavy loads from top loading autoclaves
Loading trolleys - to help roll loads in and out of front loading laboratory autoclaves
Automatic waterfill - fills the autoclave between cycles (usually R.O. water)
Air intake filter - filters air taken into the autoclave (standard with vacuum drying or
cooling cycles)
Serial interface - for monitoring by computer
Self-monitoring - TACTROL® continuously monitors itself for correct calibration
Recorders - for independent monitoring. Load and pressure sensing channels available
Waste containers - for operator safety and protection of your autoclave with waste loads.
These contain spillages whilst allowing good steam penetration
Load baskets - check the maximum number of load baskets for your size of autoclave
The choice of an autoclave, such as those in the Priorclave range, can represent a big investment
financially and you want to make sure it performs to all your expectations and sterilizes your loads
safely and correctly.[6]
How to sterilize medical instruments
This short tutorial will provide the basics for proper preparation of instruments before sterilization.
Steps
1. Make sure the instruments are free from all residual matter, such as blood or organic
tissue. Instruments must also be dry and free from mineral deposits. Such substances may
cause damage to the instruments or sterilizer.
2. Clean instruments immediately after use. It is recommended that instruments be cleaned
using the Clean and Simple Ultrasonic Cleaning System.
3. After ultrasonic cleaning, rinse for 30 seconds and pat dry.
4. Follow the instrument manufacturer’s instructions for cleaning and lubricating
instruments.
5. Sterilize in an open position.
6. Place a sterilization indicator in each tray or inside each wrapped pack.
7. Make sure that all instruments remain apart during the sterilization cycle.
8. Place empty canisters upside-down in order to prevent accumulation of water.
9. Do not overload sterilizer trays. Overloading will cause inadequate sterilization &drying.
10. Allow a distance of about 1" between trays or cassettes to permit steam circulation.
11. Place wrapped instruments in material which will allow steam penetration and promote
drying, such as an autoclave bag, autoclavable paper, or muslin towels.
12. Do not stack pouches.[3]
Tips
At least once per week use a biological spore test in any load to insure proper
sterilization. The spore test should be placed in an area that is the most difficult for the
steam to reach. (Be aware testing standards may vary) .
Warnings
Be sure that instruments of dissimilar metals (stainless steel, carbon steel, etc.) are
separated. Carbon steel instruments should be bagged or placed on autoclavable towels
and not directly on stainless steel trays. (Mixing will result in the oxidation of those
metals.
Check manufacturer's instructions as to the proper procedure for sterilizing each item.
The instrument manufacturer can give specific information on proper sterilization
temperature and sterilization time
for any item.[3]
Types of autoclave loads
Health Care
Priorclave are well acquainted with the U.K. Health Service HTM2010 requirements and offer a
compliance package suitable for all their laboratory autoclaves. Their technical department are
well experienced in performance testing to HTM2010 requirements. Priorclave's calibration and
performance testing procedures are fully accredited by UKAS.
Priorclave also recommend fitting a process printer for clear, automated record keeping and a
multi-programmed memory where more than one type of cycle is used.[6]
Automatic Autoclave Series
Tuttnauer Automatic Autoclaves produce the highest return on your investment by offering :
Rapid cycles
Low maintenance
Durability
Safety
Tuttnauer Semi-Automatic autoclaves are an effortless Plug n' Play solution. After installation,
the sterilizer is virtually maintenance free with reduced running costs.
Benefits:
Sterilizing process
Steam generator
Drying process
TSO3 has developed a unique, patented process to create and control ozone so that it can be used
to sterilize commonly used medical instruments. In a self-contained unit, ozone is generated,
circulated, and catalyzed back into oxygen – simply by pressing a single button.
Benefits
Fast
Safe for the environment
Safe for heat and moisture-sensitive
instruments
Large throughput
Inexpensive to operate
Reliable
Safe for operator
Easy to operate
Ozone technology
en.wikipedia.org/wiki/Sterilizers.html [2]
The famous internet encyclopedia
www.wikihow.com/Sterilize-Medical-Instruments [3]
Want to find all your answers of questions begin with the expression " how " ?
may you look in this website .
www.wisegeek.com [4]
This website give the science geek wisdom for all .
www.laboratorytalk.com [5]
Educational website for all instruments in laboratory.
www.progensci.co.uk/page918/Laboratory-Autoclaves [6]
Educational and business website for all instruments in laboratory.
www.arab-eng.org/vb/t80147.html [7]
The famous internet club for engineers.
Standard references
BS 2646 Part 2 1990 Ventilation standard for equipment
EN 13445 & EN 14222 design & constructing European standards
EN 866–1, Biological systems for testing sterilizers and sterilization processes — Part 1: General
requirements.1
EN 866–5, Biological systems for testing sterilizers and sterilization processes — Part 5: Particular
systems for use
in low temperature steam and formaldehyde sterilizers. 2
EN 867–5, Non-biological systems for use in sterilizers — Part 5: Specification for indicator systems
and process
challenge devices for use in performance testing for small sterilizers Type B and Type S.
EN 868–5, Packaging materials and systems for medical devices which are to be sterilized — Parts 5:
Heat and
self-sealable pouches and reels of paper and plastic film construction — Requirements and test
methods.
EN 60584–2, Thermocouples — Part 2: Tolerances (IEC 60584–2:1982 + A1:1989).
EN 60751, Industrial platinum resistance thermometer sensors (IEC 60751:1983 + A1:1986).
EN 61010–1, Safety requirements for electrical equipment for measurement, control and laboratory
use — Part 1:
General requirements (IEC 61010-1:2001).
EN 61010–2–042:1997, Safety requirements for electrical equipment for measurement, control and
laboratory
use — Part 2–042: Particular requirements for autoclaves and sterilizers using toxic gas for the
treatment of
medical materials, and for laboratory processes (IEC 61010–2–042:1997).
EN 61326:1997, Electrical equipment for measurement, control and laboratory use — EMC
requirements
(IEC 61326:1997).
EN ISO 3746, Acoustics — Determination of sound power levels of noise sources using sound
pressure — Survey
method using an enveloping measurement surface over a reflecting plane (ISO 3746:1995).
HTM2010 U.K. Health Service requirements which offer a compliance package suitable for all
their laboratory autoclaves
ISO 228–1, Pipe threads where pressure-tight joints are not made on the threads — Part 1:
Dimensions, tolerances and designation.
Geobacillus stearothermophilus
holding time
period for which the temperature, the steam pressure and the formaldehyde concentration of the steam
are held
within pre-set values and their tolerances to achieve the required inactivation efficacy in the sterilizer
chamber
NOTE The holding time follows immediately after the equilibration time.
inoculated carrier
a carrier on which a defined number of test organisms has been deposited
[EN 866-1:1997, definition 3.8]
installation qualification
IQ
obtaining and documenting evidence that equipment has been provided and installed in accordance
with its
specification
[EN ISO 14937:2000, definition 3.9]
loading door
door in a double-ended sterilizer through which the sterilizer load is put into the sterilizer chamber
prior to
sterilization
[EN 285:1996, definition 3.21] (See also 3.43 unloading door)
medical device
any instrument, apparatus, appliance, material or other article, whether used alone or in combination.
microbicidal solution
aqueous solution containing formaldehyde to feed the vaporizer for generating sterilant in the sterilizer
operating cycle
the automatic sequence of operating stages performed in a sterilizer
[EN 1422:1997, definition 3.24]
operational qualification
OQ
process of obtaining and documenting evidence that installed equipment operates within predetermined
limits when
used in accordance with its operational procedures [EN ISO 14937:2000, definition 3.12]
operator
person operating equipment for its intended purpose
override
means intended only for maintenance or safety, by which the operating cycle can be interrupted or
modified
Porous
Full of pores , able to absorb fluids
post-cycle flushing
stage after "cycle complete" indication, during which the sterilization load is left in the closed chamber
and the
internal chamber atmosphere is exchanged
pressure vessel
a vessel consisting of the sterilizer chamber, door(s) and other components that form a permanent unit
with the
sterilizer chamber and that are pressurized by the same pressure
process challenge device
item designed to simulate product and used to assess the penetration performance of the sterilization
cycle
NOTE The device is so constituted that a biological or chemical indicator can be put in the place which is the most
difficult
to reach by sterilizing agent(s).The indicator should not interfere with the function of the process challenge device.
production test
series of tests performed to demonstrate compliance of each sterilizer with its type test performance
reference measuring point
the point where the temperature sensor for the sterilization cycle control is located
requalification
repetition of part of validation for the purpose of confirming the continued acceptability of a specified
process
[EN ISO 14937:2000, definition 3.20]
R-O Read Only
RS Recommended Standard
unloading door
door in a double-ended sterilizer through which the sterilizer load is removed from the sterilizer
chamber after a sterilization cycle [EN 285:1996, definition 3.42] (See also 3.18 loading door)
UKAS
United Kingdom Accreditation Service
usable space
space inside the sterilizer chamber, which is not restricted by fixed parts and which is, according to the
manufacturer’s specification, available to accept the sterilization load .
validation
documented procedure for obtaining recording and interpreting the results required to establish that a
process will
consistently yield product complying with predetermined specifications
[EN ISO 14937:2000, definition 3.32]
Annex (A)
1 Technical requirements
1.1 Sterilizer chamber
1.1.1 Materials
1.1.1.1 The surfaces of the materials for the pressure vessel (including, for example, welds) that can
come into
contact with process chemicals shall be of materials which, under the designed operating conditions,
are not
impaired by these chemicals. They shall not release any substances known to be toxic in such quantities
that can
create a health or environmental hazard.
NOTE In the selection of materials for pressure parts and their integral attachments, due attention should be paid to
the
effects of contact corrosion and differential expansion when dissimilar metals are used in contact.
1.1.1.2 Materials for sterilizer furniture including load supporting systems shall be selected to avoid
corrosion
and galvanic attack.
1.1.2 Chamber size
For the usable space the following dimensions shall be specified in millimeters, as applicable:
a) For cylindrical horizontal or cylindrical vertical usable spaces: 000 x 000 in which:
the first three digits give the diameter of the usable space; and
the last three digits give the depth of the usable space.
b) For rectangular parallelepiped usable spaces: 000 x 000 x 000 in which:
the first three digits give the width of the usable space;
the next three digits give the height of the usable space; and
the final three digits give the depth of the usable space.
c) For other configurations the usable space shall be specified in analogy to a) or b).
d) If any dimension exceeds 1000 mm then four digits shall be used, without a decimal point.
1.1.3 Doors and interlocks of the sterilizer chamber
1.1.3.1 Sterilizer chambers shall be provided with one or two doors.
1.1.3.2 After closing the sterilizer door, it shall be possible to open it before a cycle has been started.
1.1.3.3 It shall not be possible to open a sterilizer door(s) during a cycle.
1.1.3.4 In case of an interrupted cycle (e. g. due to a fault), opening of the sterilizer door e. g. to gain
access to
the load shall require the use of an access device.
1.1.3.5 Means shall be provided to allow access to sealing surfaces for cleaning purposes and for
replacing
the door seal(s).
1.1.3.6 For double-ended sterilizers it shall not be possible to open the unloading door until a "cycle
complete"
indication is obtained.
1.1.3.7 For operating cycles dedicated for test or maintenance purposes only, the cycle complete
indication
shall be different from that of a normal sterilization cycle. For double-ended sterilizers such cycle
complete
indications shall not permit the unloading door to be opened.
1.1.3.8 The control used to start the automatic operating cycle shall be located at the loading side of the
sterilizer.
1.1.3.9 Except for maintenance purposes it shall not be possible to open both doors simultaneously on
double ended
sterilizers.
1.1.3.10 For double-ended sterilizers both ends of the sterilizer shall be fitted with a device to indicate
whether
the door at that end can be opened.
1.1.3.11 The indication "cycle complete" shall be cancelled when a door is opened. For double-ended
sterilizers
the loading door shall remain locked until the unloading door has been opened, closed and locked
again.
1.1.4 Heating the sterilizer chamber
Inner sterilizer chamber surfaces shall be heated to achieve pre-set temperatures. Initiation of the
sterilization cycle
shall not be possible until this condition has been fulfilled.
1.1.5 Test connections
1.1.5.1 The sterilizer chamber shall be provided with a test connection, which is used for the
connection of a
test pressure measuring instrument. This connection shall be at a point of easy access, but not in a pipe
for media
transport or evacuation, and shall terminate in a pipe thread ISO 228 –G1/2A.
The test connection shall be provided with a cap marked PT (Pressure Test) and sealed with a sterilant
proof and
mechanically resistant O-ring seal or flat seal.
NOTE If national regulations require the calibration of all pressure instruments connected to the pressure vessel,
test tees
and valve cocks with sealing plugs may be required to permit connection of reference instruments.
1.1.5.2 The sterilizer chamber shall be provided with a straight thermometry entry connection.This
connection shall be at a point of easy access.
The connection shall be provided with a cap marked TT (Temperature Test) and sealed with a sterilant
proof and
mechanically resistant O-ring seal or flat seal.
1.2 Design and construction
1.2.1 General
A Council Directive on the approximation of the laws of the member states concerning pressure
equipment was
released on 29 May 1997 (97/23/EC) and corresponding European Standards EN 13445 and EN 14222,
may
apply.
1.2.2 Pipework and fittings
1.2.2.1 Pipeworks and fittings (including e. g. seals) which may come into contact with process
chemicals shall
be of materials which, under the designed operating conditions, are not impaired by these chemicals.
They shall
not release any substances known to be toxic in such quantities that can create a health or
environmental hazard.
NOTE To avoid heat loss or condensation, all pipework carrying media to the sterilizer should be isolated except
where this
will interfere with the function of the sterilizer.
1.2.2.2 Connections for water and steam supply and drainage shall be provided with means to prevent
the
ingress of particles which could affect the performance of a sterilizer.
NOTE For connections to potable water supply and draining, national or local regulations may apply.
1.2.2.3 Pipework shall be designed to prevent accumulation of condensate.
1.2.2.4 All control valves in the pipework shall be marked with permanent identification in relation to
their functions
1.2.3 Evacuation system
1.2.3.1 Sterilizers shall be provided with a vacuum system to remove air, water and sterilant. The
lowest
absolute pressure needed for fulfilling the requirements of clause 6 when tested according to A.3 shall
be specified.
NOTE Vacuum systems mostly operate by means of water. Attention should be paid to optimize the use of water
in such
systems, as there could be a balance between the use of resources and diluting of formaldehyde into concentrations
harmless
to environment .
1.2.3.2 The sterilizer shall be provided with a means for leak testing which shall include the sterilizer
chamber
and all relevant connected pipework and fittings.
At leak testing the chamber and relevant pipework shall be evacuated to or below the lowest process
pressure. The
pressure rise shall not exceed 0,1 kPa/min over a period of not less than 5 min and not more than 15
min after
obtaining the lowest pressure.
1.2.4 Electrical and mechanical safety
1.2.4.1 For general design, EN 61010–1 and EN 61010–2–042 shall apply.
NOTE For guidance regarding specific design aspects, EN 60204–1 may apply. The guidance in EN 60204–1
may reduce
testing.
1.2.4.2 Sterilizers shall comply with EN 61326 regarding electromagnetic compatibility (EMC).
Sterilizers operating either in areas in which medical electrical equipment is intended to be used or in
the vicinity of
other sensitive equipment shall be regarded as class B equipment as specified in EN 61326.
For immunity, the testing requirements in EN 61326:1997, Table A.1 shall apply.
For the selection of immunity testing performance criteria, the general guidance of EN 61326 applies.
The selected performance criteria shall ensure that the sterilizer performance as specified in 5.2 will be
met during
normal operation, when exposed to disturbance phenomena given in EN 61326:1997, Table A1.
1.2.5 Aeration system
1.2.5.1 When the sterilizer chamber is ventilated during the aeration for the purpose of desorption and
to
release the vacuum at the end of the process, microbial recontamination of the sterilization load shall be
prevented.
1.2.5.2 When a filter is fitted to the sterilizer to prevent microbial recontamination during aeration or
pressure
equalization, it shall be readily accessible for replacement. The filter shall be capable of retaining at
least 99,5 % of
particles with a diameter of 0,3 μm at a pressure difference of 1 bar and at maximum airflow.
Means shall be provided between the filter and the sterilization chamber to prevent fluid flow from the
sterilizer
chamber into the filter.
1.2.6 Framework and paneling
1.2.6.1 If the sides of the sterilizer need not to be accessible for normal operation, they shall be
enclosed with
paneling.
NOTE 1 Sterilizers designed to be recessed into existing walls providing continuous joint with the sterilizer front
paneling, do
not need to be provided with side paneling.
The paneling shall be long-term corrosion-resistant. Instructions for cleaning of the paneling shall be
provided.
NOTE 2 Ventilation openings may be provided in the paneling.
1.2.6.2 The paneling of sterilizers shall allow access for maintenance work.
NOTE The access for maintenance should be positioned so that it will not compromise the safety of either the
product or
persons. Requirements for access are specified in EN 61010–2–042.
1.3 Indicating, measuring, operating and recording devices
1.3.1 General
1.3.1.1 Indicating and operating devices shall be identified as to their function. They shall be readily
accessible, clearly and durably marked with their function and designed to be easy to operate and read.
1.3.1.2 The devices shall be positioned and/or protected such that their performance is within the
specified
tolerances during the operation of the sterilizer.
1.3.1.3 If an indicating instrument is connected in turn to more than one sensing point, there shall be a
continuous indication of the active sensor that is being monitored.
1.3.1.4 Indicating and operating devices shall be readable when viewed at a distance of (1±0,15) m
with
normal or corrected vision in an illumination of (215±15) lx.
1.3.1.5 Indicating, measuring and recording devices shall have means to adjust in situ by the use of an
access device without dismantling the instrument. For analogue indicating instruments that are not
easily
demountable, means for adjustment in situ shall be provided.
NOTE Where digital pressure indicators are used, an additional mechanically attached indicator may be required
to comply
with national pressure vessel regulations. Where an analogue instrument is provided only for this purpose, the
requirement for
adjustment in situ is waived.
1.3.1.6 Additional functions fitted at recording or indicating devices shall not jeopardize the accuracy
of
registration or indication.
1.3.2 Temperature measuring devices
1.3.2.1 When used for process control, monitoring or registration purposes, temperature measuring
devices
shall have maximum permissible errors of 1 K or less over the scale range 20 °C to 100 °C and be
adjusted to
± 0,5 K or less at the sterilization temperature.
1.3.2.2 Temperature control systems shall be protected against unauthorized operation by the use of an
access device.
1.3.2.3 At least two independent sensors shall be used for the measurement of the chamber
temperature,
being dedicated to indication, control and registration. The sensors shall be located in the most
representative
position in the chamber specified as being the reference measuring point.
1.3.2.4 The registration and monitoring/controlling devices shall be mutually independent as given by
any of
the arrangements shown in Figure 1.
1.3.2.5 The temperature sensor including the measuring system it is connected to shall have a response
time
τ0,9 < 5 s when tested in flowing water according to EN 60751.
1.3.2.6 Temperature control systems shall have a function which causes a fault to be indicated if a
sensor
fails.
1.3.3 Temperature indicating devices
In addition to the requirements in 4.3.2.1, the temperature indicating devices shall:
be graduated in degrees Celsius;
have a scale, which includes 0 °C to 100 °C;
for analogue instruments be graduated in divisions not greater than 2 °C;
for digital instruments have a resolution of at least 0,1 °C.
2 Process control
2.1 General
2.1.1 The sterilizer shall be operated by an automatic controller, which has one or more pre-
set sterilization
cycles.
2.1.2 The sterilizer shall operate with pre-set programs permanently stored in the automatic
controller. Any
change of the pre-set program or its parameters shall require the use of an access device.
NOTE Automatic loading and unloading may be initiated without the use of an access device.
2.1.3 The variables and parameters programmed into the automatic controller and the
tolerable limits that will still
enable the performance requirements of clause 6 to be met shall be specified. (See also 9.2
a)).
2.1.4 The automatic controller shall be protected against short circuit in components or
equipment, which are
directly or indirectly connected to the controller.
NOTE 1 Guidance is given in EN 60204–1.
NOTE 2 Verification may be achieved by assessing compliance with EN 60204–1.
2.1.5 The automatic controller shall be located such that the maximum values of temperature
and humidity
specified for the automatic controller are not exceeded.
NOTE Normally the temperature and humidity in the vicinity of the control system should not exceed 50
°C and 85 %
relative humidity respectively.
2.1.6 The automatic controller shall have status indicators for each digital inputs and outputs.
NOTE These can be located within the enclosure of the controller.
2.1.7 The reference measuring point shall be selected in such a way that throughout the
holding time the
temperature at this point correlates with the temperature in the usable space.
2.2 Sterilization cycle and automatic control
2.2.1 Leak testing shall be performed either during the first vacuum phase or during the
complete holding time.
For process control, if the test is carried out during the first vacuum phase, a pressure
stabilizing time not
exceeding 15 min may precede the test period.
The pressure change measured shall be 50 Pa (0,5mbar) per minute when tested over a
period of 10 min.
A pressure rise due to media injection during the holding time shall not be considered as a
leak for the evaluation
of the leak rate.
There shall be a visual and/or audible indication if the test fails. The process shall be
interrupted automatically by
aeration and, in case the sterilant has already been injected, the desorption process shall be
started immediately.
2.2.2 The parameters identified during process development as critical to the sterilization
cycle shall be
reproducible within the limits identified in 2.1.3. This shall be demonstrated by the tests
specified in Table B.1.
2.2.3 Means shall be provided to ensure that the specified amount of microbicidal solution has
been supplied to
the vaporizer and has been vaporized.
2.2.4 The automatic controller shall ensure that all chemical/physical conditions necessary for
the efficacy and
reproducibility of the sterilization cycle are achieved throughout the specific periods.
NOTE Ab-/adsorption of formaldehyde to the load and formation of paraformaldehyde may reduce the
concentration of
vaporised formaldehyde in the chamber during the process.
2.2.5 The sterilization cycle shall include a desorption phase (see 6.2 and EN 61010–2–042).
2.2.6 The stage advance system of the automatic controller shall ensure that during a
sterilization cycle,
switching from one process stage to another will only be possible when the values required
for advance have been
attained.
2.2.7 An automatic post cycle flushing shall run or be repeated until the unloading door is
opened for removal of
the sterilizer load. This flushing-sequence shall be continuous or otherwise ensure that when
opening the
unloading door, the formaldehyde concentration in the chamber and load does not constitute
a hazard to human
beings or environment.
NOTE 1 Desorption of condensate residuals from the sterilized load may continue when the sterilized
load is left in the
sterilizer chamber.
NOTE 2 In order to maintain this ability a separate maintenance process may be recommended to
remove accumulated
formaldehyde residuals from the sterilizer chamber and piping (see 6.2 a)).