Principles of Temperature Measurement
Principles of Temperature Measurement
Temperature measurement
Principles of temperature measurement
Temperature measurement is very important in all spheres of life and especially so in the
process industries. However, it poses particular problems, since temperature measurement
cannot be related to a fundamental standard of temperature in the same way that the
measurement of other quantities can be related to the primary standards of mass, length and
time. If two bodies of lengths l1 and l2 are connected together end to end, the result is a body
of length l1+l2. A similar relationship exists between separate masses and separate times.
However, if two bodies at the same temperature are connected together, the joined body has
the same temperature as each of the original bodies.
This is a root cause of the fundamental difficulties that exist in establishing an absolute
standard for temperature in the form of a relationship between it and other measurable
quantities for which a primary standard unit exists. In the absence of such a relationship, it is
necessary to establish fixed, reproducible reference points for temperature in the form of
freezing and boiling points of substances where the transition between solid, liquid and
gaseous states is sharply defined.
1|Page
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
the individual expansions of the metallic elements. To maximize the bending of the actuator,
metals or alloys with greatly differing coefficients of thermal expansion are normally selected.
The metal having the largest thermal expansitivities is known as the active element, while the
metal having the smaller coefficient of expansion is known as the passive element. For
maximum actuation, the passive element is often an iron–nickel alloy, Invar, having an almost
zero thermal expansivities (actually between 0.1 and 1×10 –6K–1, depending upon the
composition).The active element is then chosen to have maximum thermal expansivities given
the constraints of operating environment and costs.
The width of the strip is taken as equal to unity. Equation 5.49 applies for several strip
configurations, including the simply supported strip and a strip clamped at one end (i.e., a
2|Page
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
cantilever). For a given configuration, the deflection of a strip can be determined by its
relationship with curvature, 1/R.
FIGURE 6.2 Bending of bimetallic strip uniformly heated withα2≥α1. (a) Bimetallic strip. A1B1–
A2B2is an element cut out from the strip. (b) Bending of the element A1B1–A2B2when
uniformly heated. Assumingα2>α1, the deflection is convex up. The total force acting over the
section of (1) is an axial tensile force P1and bending moment M1,whereas over the section of
(2) it is an axial compressive force P2and bending moment M2. (c) Sketch of the internal
resulting stress distribution. (Left): normal stresses over the cross section of the strip. The
maximum stress during heating is produced at the interface between the two components of
the strip. This stress is due to both axial force and bending. (Right): shearing stresses at the
ends of the strip.
3|Page
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
long-term stability over time at high temperatures. Platinum is a better choice than copper or
nickel because it is chemically inert, it withstands oxidation well, and works in a higher
temperature range as well.
In operation, the measuring instrument applies a constant current through the RTD. As the
temperature changes, the resistance changes and the corresponding change in voltage is
measured. This measurement is then converted to thermal values by a computer. Curve-fitting
equations are used to define this resistance vs. temperature relationship. The RTD can then be
used to determine any temperature from its measured resistance.
These instruments have functions to compensate for errors such as thermoelectric voltages
and element self-heating.
In addition to temperature, strain on and impurities in the wire also affect the sensor’s
resistance vs. temperature characteristics. The Matthiessen rule states that the resistivity () of
a metal conductor depends on temperature, impurities, and deformation. is measured in
(Ωcm):
Proper design and careful material selection will minimize these effects so that resistivity will
only vary with a change in temperature.
Resistance of Metals
Whether an RTD’s element is constructed of platinum, copper, or nickel, each type of metal has
a different sensitivity, accuracy, and temperature range. Sensitivity is defined as the amount of
resistance change of the sensor per degree of temperature change. Figure below shows the
sensitivity for the most common metals used to build RTDs. Platinum, a noble metal, has the
most stable resistance-to-temperature relationship over the largest temperature range –
184.44C (–300F) to 648.88C (1200F). Nickel elements have a limited temperature range
because the amount of change in resistance per degree of change in temperature becomes very
nonlinear at temperatures above 300C (572F). Copper has a very linear resistance-to-
temperature relationship. However, copper oxidizes at moderate temperatures and cannot be
used above 150C (302F). Platinum is the best metal for RTD elements for three reasons. It
follows a very linear resistance-to-temperature relationship; it follows its resistance-to-
temperature relationship in a highly repeatable manner over its temperature range; and it has
the widest temperature range among the metals used to make RTDs. Platinum is not the most
sensitive metal; however, it is the metal that offers the best long-term stability.
The accuracy of an RTD is significantly better than that of a thermocouple within an RTD’s
normal temperature range of –184.44C (–300F) to 648.88C (1200F). RTDs are also known
for high stability and repeatability. They can be removed from service and recalibrated for
verifiable accuracy and checked for any possible drift.
4|Page
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
FIGURE 6.3 of the common metals, nickel has the highest sensitivity
Overview of Platinum RTDs
There are three main classes of Platinum Resistance Thermometers (PRTs): Standard Platinum
Resistance Thermometers (SPRTs), Secondary Standard Platinum Resistance Thermometers
(Secondary SPRTs), and Industrial Platinum Resistance Thermometers (IPRTs). Table Below
presents information about each.
Table 6.1
5|Page
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
FIGURE 6.4 The Secondary Standard Platinum Resistance Thermometer is intended for
laboratory environments.
6|Page
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
FIGURE 6.6 The Standard Platinum Resistance Thermometer is fragile and used only in
laboratory environments.
Thermistor
A Thermistor is a thermally sensitive resistor whose primary function is to exhibit a change in
electric resistance with a change in body temperature. Unlike a wire wound or metal film
resistance temperature detector (RTD), a Thermistor is a ceramic semiconductor. An RTD
exhibits a comparatively low temperature coefficient of resistance on the order of 0.4 to 0.5%
C–1. Depending on the type of material system used, a Thermistor can have either a large
positive temperature coefficient of resistance (PTC device) or a large negative temperature
coefficient of resistance (NTC device).
Two types of PTC thermistors are available. Silicon PTC thermistors rely on the bulk properties
of doped silicon and exhibit resistance–temperature characteristics that are approximately
linear. They have temperature coefficients of resistance of about 0.7 to 0.8% C–1. The most
7|Page
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
8|Page
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
coefficient of resistance becomes positive and the resistance begins to rise. At a threshold or switching
temperature, the rate of rise becomes very rapid and the PTC characteristic becomes very steep. Within its
switching range, the device exhibits a high resistance value. At temperatures above the switching range, the
resistance reaches a maximum value beyond which the temperature coefficient becomes negative again.
Thermocouple
A thermocouple is a junction between two different metals that produces a voltage related to a
temperature difference. Thermocouples are a widely used type of temperature sensor for
measurement and control and can also be used to convert heat gradient into electricity. They
are inexpensive and interchangeable, are supplied fitted with standard connectors, and can
measure a wide range of temperatures. The main limitation is accuracy: system errors of less
than one degree Celsius (C) can be difficult to achieve.
Any junction of dissimilar metals will produce an electric potential related to temperature.
Thermocouples for practical measurement of temperature are junctions of specific alloys which
have a predictable and repeatable relationship between temperature and voltage. Different
alloys are used for different temperature ranges. Properties such as resistance to corrosion may
also be important when choosing a type of thermocouple. Where the measurement point is far
from the measuring instrument, the intermediate connection can be made by extension wires
which are less costly than the materials used to make the sensor. Thermocouples are usually
standardized against a reference temperature of 0 degrees Celsius; practical instruments use
electronic methods of cold-junction compensation to adjust for varying temperature at the
instrument terminals. Electronic instruments can also compensate for the varying
characteristics of the thermocouple, and so improve the precision and accuracy of
measurements.
Thermocouples are widely used in science and industry; applications include temperature
measurement for kilns, gas turbine exhaust, diesel engines, and other industrial processes.
9|Page
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
Principle of operation
In 1821, the German–Estonian physicist Thomas Johann Seebeck discovered that when any
conductor is subjected to a thermal gradient, it will generate a voltage. This is now known as
the thermoelectric effect or Seebeck effect. Any attempt to measure this voltage necessarily
involves connecting another conductor to the "hot" end. This additional conductor will then
also experience the temperature gradient, and develop a voltage of its own which will oppose
the original. Fortunately, the magnitude of the effect depends on the metal in use. Using a
dissimilar metal to complete the circuit creates a circuit in which the two legs generate
different voltages, leaving a small difference in voltage available for measurement. That
difference increases with temperature, and is between 1 and 70 microvolt’s per degree Celsius
(µV/°C) for standard metal combinations. The voltage is not generated at the junction of the
two metals of the thermocouple but rather along that portion of the length of the two
dissimilar metals that is subjected to a temperature gradient. Because both lengths of dissimilar
metals experience the same temperature gradient, the end result is a measurement of the
temperature at the thermocouple junction.
Voltage–temperature relationship For typical metals used in thermocouples, the output voltage
increases almost linearly with the temperature difference (ΔT) over a bounded range of temperatures.
For precise measurements or measurements outside of the linear temperature range, non-linearity
must be corrected. The nonlinear relationship between the temperature difference (ΔT) and the output
voltage (mV) of a thermocouple can be approximated by a polynomial:
10 | P a g e
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
The coefficients an are given for n from 0 to between 5 and 13 depending upon the metals. In
some cases better accuracy is obtained with additional non-polynomial terms.
Thermocouples measure the temperature difference between two points, not absolute
temperature. To measure a single temperature one of the junctions—normally the cold
junction—is maintained at a known reference temperature, and the other junction is at the
temperature to be sensed. Having a junction of known temperature, while useful for laboratory
calibration, is not convenient for most measurement and control applications. Instead, they
incorporate an artificial cold junction using a thermally sensitive device such as a thermistors or
diode to measure the temperature of the input connections at the instrument, with special care
being taken to minimize any temperature gradient between terminals. Hence, the voltage from
a known cold junction can be simulated, and the appropriate correction applied. This is known
as cold junction compensation. Some integrated circuits such as the LT1025 are designed to
output a compensated voltage based on thermocouple type and cold junction temperature.
Alternatively cold junction compensation can be performed by computation using look-up
tables and polynomial interpolation.
After much investigation of thermoelectric circuits several basic concepts have been conceived. These
concepts can be summed up according to 3 fundamental "laws".
1.)The Law of Homogeneous Circuits
An electric current cannot be sustained in a circuit of a single homogeneous metal, however varying in
section, by the application of heat alone.
What this means for thermocouples, is that if there is a temperature distribution along the wires
between the hot and cold junctions, the total thermal EMF will be unaffected. Only the temperature at
the junction between the two dissimilar metals will have an effect on the EMF produced.
11 | P a g e
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
Seebeck Effect
In 1821 Thomas Johann Seebeck found that a circuit made from two dissimilar metals, with junctions at
different temperatures would deflect a compass magnet. Seebeck initially believed this was due to
magnetism induced by the temperature difference. However, it was quickly realized that it was an
electrical current that is induced, which by Ampere's law deflects the magnet. More specifically, the
temperature difference, produces an electric potential (voltage) which can drive an electric current in a
closed circuit. Today, this is known as the Seebeck effect.
The voltage produced is proportional to the temperature difference between the two junctions.
The Seebeck voltage does not depend on the distribution of temperature along the metals
12 | P a g e
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
between the junctions. This is the physical basis for a thermocouple, which is used often for
temperature measurement
The voltage difference, V, produced across the terminals of an open circuit made from a pair of
dissimilar metals, A and B, whose two junctions are held at different temperatures, is directly
proportional to the difference between the hot and cold junction temperatures, Th - Tc.
Peltier Effect
In 1834, a French watchmaker and part time physicist, Jean Charles Athanase Peltier found that an
electrical current would produce heating or cooling at the junction of two dissimilar metals. In 1838 Lenz
showed that depending on the direction of current flow, heat could be either removed from a junction
to freeze water into ice, or by reversing the current, heat can be generated to melt ice. The heat
absorbed or created at the junction is proportional to the electrical current. The proportionality
constant is known as the Peltier coefficient.
Thomson Effect
Twenty years later, William Thomson (later Lord Kelvin) issued a comprehensive explanation of the
Seebeck and Peltier Effects and described their interrelationship. The Seebeck and Peltier coefficients
are related through thermodynamics. The Peltier coefficient is simply the Seebeck coefficient time’s
absolute temperature. This thermodynamic derivation leads Thomson to predict a third thermoelectric
effect, now known as the Thomson effect. In the Thomson effect, heat is absorbed or produced when
current flows in a material with a temperature gradient. The heat is proportional to both the electric
current and the temperature gradient. The proportionality constant, known as the Thomson coefficient
is related by thermodynamics to the Seebeck coefficient.
The thermopile
The thermopile is the name given to a temperature-measuring device that consists of several
thermocouples connected together in series, such that all the reference junctions are at the
same cold temperature and all the hot junctions are exposed to the temperature being
measured, as shown in Figure 6.9. The effect of connecting n thermocouples together in series
is to increase the measurement sensitivity by a factor of n. A typical thermopile manufactured
by connecting together 25 chromel–constantan thermocouples gives a measurement resolution
of 0.001⁰C.
13 | P a g e
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
Flow Measurement
Flow measurement is an everyday event. Whether you are filling up a car with petrol (gasoline)
or wanting to know how much water the garden sprinkler is consuming, a flow meter is
required. Similarly, it is also difficult to think of a sector of industry in which a flow meter of one
type or another does not play a part. Given the size of the flow meter market, and the value of
product being measured, it is somewhat surprising that both the accuracy and capability of
many flow meters are poor in comparison to those instruments used for measurement of other
common process variables such as pressure and temperature. For example, the orifice plate
flow meter, which was first used commercially in the early 1900s and has a typical accuracy of ±
2% of reading, is still the only flow meter approved by most countries for the fiscal
measurement of natural gas. Although newer techniques such as Coriolis flow meters have
become increasingly popular in recent years, the flow measurement industry is by nature
conservative and still dominated by traditional measurement techniques.
The rate at which fluid flows through a closed pipe can be quantified by either measuring the
mass flow rate or measuring the volume flow rate. Of these alternatives, mass flow
measurement is more accurate, since mass, unlike volume, is invariant. In the case of the flow
of solids, the choice is simpler, since only mass flow measurement is appropriate.
Mass flow rate
The method used to measure mass flow rate is largely determined by whether the measured
quantity is in a solid, liquid or gaseous state. Some of these techniques are discussed below.
Conveyor-based methods
These methods are concerned with measurement of the flow of solids that are in the form of
small particles. Such particles are usually produced by crushing or grinding procedures in
process industries, and the particles are usually transported by some form of conveyor. This
mode of transport allows the mass flow rate to be calculated in terms of the mass of material
on a given length of conveyor multiplied by the speed of the conveyor. Figure 6.10 shows a
typical measurement system. A load cell measures the mass M of material distributed over a
length L of the conveyor. If the conveyor velocity is v, the mass flow rate, Q, is given by:
As an alternative to weighing the flowing material, a nuclear mass-flow sensor can be used, in
which a gamma-ray source is directed at the material being transported along the conveyor.
14 | P a g e
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
The material absorbs some radiation, and the amount of radiation received by a detector on
the other side of the material indicates the amount of material on the conveyor. This technique
has obvious safety concerns, and is therefore subject to licensing and strict regulation.
15 | P a g e
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
Figure 6.11 (a) Coriolis flow meter shapes; (b) detail of U-shaped Coriolis flow meter.
Coriolis meters give excellent accuracy, with measurement uncertainties of ±0.2% being typical. They also have low
maintenance requirements. However, apart from being expensive, they suffer from a number of operational
problems. Failure may occur after a period of use because of mechanical fatigue in the tubes. Tubes are also
subject to both corrosion caused by chemical interaction with the measured fluid and abrasion caused by particles
within the fluid. Diversion of the flowing fluid around the flow meter causes it to suffer a significant pressure drop,
though this is much less evident in straight tube designs.
Volume flow rate
Volume flow rate is an appropriate way of quantifying the flow of all materials that are in a
gaseous, liquid or semi-liquid slurry form (where solid particles are suspended in a liquid host),
although measurement accuracy is inferior to mass flow measurement as noted earlier.
Materials in these forms are carried in pipes, and various instruments can be used to measure
the volume flow rate as described below.
Differential pressure (obstruction-type) meters Differential pressure meters involve the
insertion of some device into a fluid-carrying pipe that causes an obstruction and creates a
pressure difference on either side of the device. Such meters are sometimes known as
obstruction-type meters or flow-restriction meters. Devices used to obstruct the flow include
the orifice plate, the Venturi tube, the flow nozzle and the Dall flow tube, as illustrated in Figure
6.12.When such a restriction is placed in a pipe, the velocity of the fluid through the restriction
increases and the pressure decreases. The volume flow rate is then proportional to the square
root of the pressure difference across the obstruction. The manner in which this pressure
difference is measured is important. Measuring the two pressures with different instruments
and calculating the difference between the two measurements is not satisfactory because of
the large measurement error which can arise when the pressure difference is small. Therefore,
the normal procedure is to use a differential pressure transducer, which is commonly a
diaphragm type.
The Pitot static tube is a further device that measures flow by creating a pressure
difference within a fluid-carrying pipe. However, in this case, there is negligible obstruction of
flow in the pipe. The Pitot tube is a very thin tube that obstructs only a small part of the flowing
fluid and thus measures flow at a single point across the cross-section of the pipe. This
measurement only equates to average flow velocity in the pipe for the case of uniform flow.
The Annubar is a type of multi-port Pitot tube that does measure the average flow across the
16 | P a g e
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
cross-section of the pipe by forming the mean value of several local flow measurements across
the cross-section of the pipe.
Figure 6.12 Obstruction devices: (a) orifice plate; (b) venturi; (c) flow nozzle; (d) Dall flow
tube.
All applications of this method of flow measurement assume that flow conditions upstream of
the obstruction device are in steady state, and a certain minimum length of straight run of pipe
ahead of the flow measurement point is specified to ensure this. The minimum lengths required
for various pipe diameters are specified in British
Standards tables (and also in alternative but equivalent national standards used in other
countries), but a useful rule of thumb widely used in the process industries is to specify a length
of ten times the pipe diameter. If physical restrictions make this impossible to achieve, special
flow smoothing vanes can be inserted immediately ahead of the measurement point.
Flow-restriction type instruments are popular because they have no moving parts and are
therefore robust, reliable and easy to maintain. One disadvantage of this method is that the
obstruction causes a permanent loss of pressure in the flowing fluid. The magnitude and hence
importance of this loss depends on the type of obstruction element used, but where the
pressure loss is large, it is sometimes necessary to recover the lost pressure by an auxiliary
pump further down the flow line. This class of device is not normally suitable for measuring the
flow of slurries as the tapping’s into the pipe to measure the differential pressure are prone to
blockage, although the Venturi tube can be used to measure the flow of dilute slurries.
Figure 6.13 illustrates approximately the way in which the flow pattern is interrupted when
an orifice plate is inserted into a pipe. The other obstruction devices also have a similar effect to
this. Of particular interest is the fact that the minimum cross-sectional area of flow occurs not
within the obstruction but at a point downstream of there. Knowledge of the pattern of
pressure variation along the pipe, as shown in Figure 6.13, is also of importance in using this
technique of volume flow rate measurement. This shows that the point of minimum pressure
coincides with the point of minimum cross section flow, a little way downstream of the
obstruction. Figure 16.5 also shows that there is a small rise in pressure immediately before the
obstruction. It is therefore important not only to position the instrument measuring P2 exactly
17 | P a g e
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
at the point of minimum pressure, but also to measure the pressure P1 at a point upstream of
the point where the pressure starts to rise before the obstruction.
Q is given by:
In the absence of any heat transfer mechanisms, and assuming frictionless flow of an
incompressible fluid through the pipe, the theoretical volume flow rate of the fluid, where A1
and P1 are the cross-sectional area and pressure of the fluid flow before the obstruction, A2
and P2 are the cross-sectional area and pressure of the fluid flow at the narrowest point of the
flow beyond the obstruction, and ρ is the fluid density.
Turbine flow meters
A turbine flow meter consists of a multi-bladed wheel mounted in a pipe along an axis parallel
to the direction of fluid flow in the pipe, as shown in Figure 6.14. The flow of fluid past the
wheel causes it to rotate at a rate that is proportional to the volume flow rate of the fluid. This
rate of rotation has traditionally been measured by constructing the flow meter such that it
behaves as a variable reluctance tacho generator. This is achieved by fabricating the turbine
blades from a ferromagnetic material and placing a permanent magnet and coil inside the
meter housing. A voltage pulse is induced in the coil as each blade on the turbine wheel moves
past it, and if these pulses are measured by a pulse counter, the pulse frequency and hence
flow rate can be deduced.
In recent instruments, fiber optics are also now sometimes used to count the rotations by
detecting reflections off the tip of the turbine blades. Provided that the turbine wheel is
mounted in low friction bearings, measurement inaccuracy can be as low as ±0.2%. However,
turbine flow meters are less rugged and reliable than flow-restriction type instruments, and are
badly affected by any particulate matter in the flowing fluid. Bearing wear is a particular
problem and they also impose a permanent pressure loss on the measured system. Turbine
meters are particularly prone to large errors when there is any significant second phase in the
18 | P a g e
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
fluid measured. For instance, using a turbine meter calibrated on pure liquid to measure a
liquid containing 5% air produces a 50% measurement error. As an important application of the
turbine meter is in the petrochemical industries, where gas/oil mixtures are common, special
procedures are being developed to avoid such large measurement errors.
Turbine meters have a similar cost and market share to positive displacement meters, and
compete for many applications, particularly in the oil industry. Turbine meters are smaller and
lighter than the latter and are preferred for low-viscosity, high-flow measurements. However,
positive-displacement meters are superior in conditions of high viscosity and low flow rate.
19 | P a g e
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
L is the distance between the electrodes, which is the diameter of the tube, and B is a known
constant. Hence, measurement of the voltage E induced across the electrodes allows the flow
velocity v to be calculated from equation above. Having thus calculated v, it is a simple matter
to multiply v by the cross-sectional area of the tube to obtain a value for the volume flow rate.
The typical voltage signal measured across the electrodes is 1mV when the fluid flow rate is
1m/s.
The internal diameter of magnetic flow meters is normally the same as that of the rest of the
flow-carrying pipe work in the system. Therefore, there is no obstruction to the fluid flow and
consequently no pressure loss associated with measurement. Like other forms of flow meter,
the magnetic type requires a minimum length of straight pipe work immediately prior to the
point of flow measurement in order to guarantee the accuracy of measurement, although a
length equal to five pipe diameters is usually sufficient.
Whilst the flowing fluid must be electrically conductive, the method is of use in many
applications and is particularly useful for measuring the flow of slurries in which the liquid
phase is electrically conductive. Corrosive fluids can be handled providing a suitable lining
material is used. At the present time, magnetic flow meters account for about 15% of the new
flow meters sold and this total is slowly growing. One operational problem is that the insulating
lining is subject to damage when abrasive fluids are being handled, and this can give the
instrument a limited life.
Current new developments in electromagnetic flow meters are producing physically smaller
instruments and employing better coil designs which reduce electricity consumption and make
battery-powered versions feasible (these are now commercially available). Also, whereas
conventional electromagnetic flow meters require a minimum fluid conductivity of
10µmho/cm3, new versions can cope with fluid conductivities as low as 1µmho/cm 3.
20 | P a g e
A Great Teacher Deserves A Great Student
Instrumentation and measurement (TELT 2206) 2012
21 | P a g e
A Great Teacher Deserves A Great Student