Truestart Ii: Installation & Service Instrument
Truestart Ii: Installation & Service Instrument
Truestart Ii: Installation & Service Instrument
MENU F1 F2
Main Menu
[Earth Fault Test]
Press ENT to select
ENT
To Test Connect
IDNet/IDNAC Red +, Black -
X-Talk Red to CHNL1+
Green to CHNL2+
IDC Red+, Black -
NAC Red to Alarm-
Black to Alarm+
ON OFF
To Navigate Menus
Use up/down arrow and keys to move
between menu choices
Press ENT ket to select a menu choice
Press MENU key to back out of a menu
choice
579-1030 Rev. C
Quick Reference
TrueSTART II Introduction............................................................. 3
Power Sources.............................................................................. 3
Battery Operation .......................................................................... 3
AC Adapter.................................................................................... 4
Aux Power Harness ...................................................................... 4
Testing & Safety ............................................................................ 5
Selecting a Test............................................................................. 6
Earth Fault Testing ........................................................................ 6
IDNet ............................................................................................. 7
Other IDNet Tests.......................................................................... 9
IDNet Shorts.................................................................................. 9
IDNAC ........................................................................................... 10
Power Option ................................................................................ 11
X-Talk Test..................................................................................... 12
Test IDC/zone................................................................................ 13
Test NAC/Signal ............................................................................ 14
Saving IDNet and IDNAC Test Results ......................................... 15
Recalling Test Results................................................................... 15
Upload to PC................................................................................. 15
System Settings ............................................................................ 16
Backlight Timer.............................................................................. 16
Battery Selection ........................................................................... 17
Software Revision ......................................................................... 17
Battery Charger............................................................................. 17
IDNet Device Types ...................................................................... 18
IDNet Error Messages................................................................... 19
IDNAC Device Types .................................................................... 20
IDNAC Error Messages................................................................. 20
TrueSTART Data Analyzer ............................................................ 22
Replacement Parts........................................................................ 23
22
TrueSTART II Introduction
The Simplex TrueSTART II Instrument (TSIT) is a portable, handheld
instrument for verification of the correct installation of Simplex peripherals
and wiring to Simplex circuits and modules.
The TSIT is used to verify wiring, addressable devices, conventional zones
(IDC) and signal circuits (NAC) before the Fire Alarm Control Panel (FACP) is
commissioned.
This manual describes the steps to verify that all wiring and peripheral
devices are correctly installed before connecting the wires to the FACP.
Power Sources
The TSIT operates from multiple power sources:
Lithium Ion Battery
AC Adapter
Any 24V DC source (using the Auxiliary Power Harness).
Battery Operation
Remove the battery compartment by squeezing in the tabs on the sides.
Connect a charged battery to the harness inside the tester case.
Replace the battery compartment.
ON OFF
To Navigate Menus
Use up/down arrow and keys to move
between menu choices
Press ENT ket to select a menu choice
Press MENU key to back out of a menu
choice
This is the
battery
If you can
read this
much too you are sta
close to
the manua nding
l
3
3
AC Adapter
Connect the AC Adapter to the power port in the right side of the TSIT. Plug
the adapter into a wall outlet.
The AC Adapter is rated for 100-240 VAC, 50-60 Hz. It is supplied with a US
standard wall plug.
44
The Power Plug (1) connects to the TSIT, bringing power from either the AC
Adapter or the Alligator Clips (3).
To use the Aux Power Harness with the AC Adapter, connect the power plug
from the adapter to the Power Jack (2).
Red & Black Alligator Clips (3) may be used as INPUT to provide power to
the tester, or as an OUTPUT to distribute 24V power to peripheral devices
such as IDNet ZAMs.
2
1 3
Power-Up
Do not connect the test leads to any field wiring until after the TSIT is
powered. Press and hold the Green ON button until the Initialization Screen
is displayed. The tester requires approximately 5 seconds to initialize, after
which it will display the first menu choice.
5
5
Selecting a Test
Use the or keys to scroll through the menu choices. When the desired
selection is displayed, press ENT to select that option.
To return to the main menu, follow the instructions on the display. Press
MENU to back out of a menu choice.
Earth Test
NORMAL
Press MENU to exit
ENT
To detect every earth fault, you must test the positive lead and the negative
lead of a circuit independently. If there are earth faults on both positive and
negative leads, it might result in a balanced condition, where no earth fault is
detectable when tested as a circuit.
Connect the Red lead per the yellow label on the TSIT, and the Green lead to
earth. Press the ENT key. The display shows the earth test status in real time.
If the display indicates "Positive Earth Fault", you must troubleshoot and
correct the earth fault. If the display indicates "Normal", disconnect the Red
lead and connect the Black lead as described on the label.
66
When the TSIT display indicates that a circuit is NORMAL, the test leads may
be moved to another circuit. This permits sequential testing of multiple
circuits without any repeated menu selection.
IDNet
The user should have a list of IDNet devices and addresses when checking
out an IDNet installation. The list can be compared against the devices found
by the TSIT, making some problems easier to diagnose. The TSIT channel
information can be saved in one of the 60 internal memory slots, then
uploaded to a personal computer for easier review. (see the section on PC
Upload.)
To test an IDNet channel with the TSIT, connect as follows:
IDNet+ to Red Lead. Connect ALL IDNet+ wiring for the channel to
the Red Lead. It is important that the channel be complete when
scanned.
IDNet- to Black Lead. Connect ALL IDNet- wiring for this channel to
the Black Lead.
Earth to Green Lead. Connect to building steel or other earth point.
If ZAM or any 24V powered devices are present, be sure to connect the
Auxiliary Power Harness. Clip the red alligator lead to the +24V wiring, the
black to the 0V wiring.
7
7
The IDNet Test (Menu Selection Run IDNet Test) performs a series of tests
on the IDNet channel. Navigate through menu options until Run IDNet Test is
displayed, and press ENT.
This menu selection sequentially does the following:
Scans the channel to determine device types and addresses
Identifies any duplicated address setting
Polls every device to determine the state, for example:
Pull stations – Normal or Alarm?
ZAMs – Normal/Alarm/Open?
TrueAlarm – analog value/excessively dirty
If an earth fault exists, will attempt to isolate the fault if isolators are
present.
At the completion of the test, review the
Total Devices 249 summary of devices detected. Then
PHOTO 190 compare it to the devices that are
IPHOTO 58
-more- supposed to be installed.
88
If duplicate devices are detected, device type info may be incorrect due to the
simultaneous traffic from multiple devices causing an error.
To troubleshoot duplicates, go to the
Address: [192] TSIT Error or Duplicate menu and send a
PHOTO command to the device address to
DUPLICATE DEVICE activate the device’s LED. The devices
Press MENU to quit
set to the displayed address will have
ENT their LEDs lit as long as TSIT displays
their address. In the example shown, any
devices with LEDs at address 192 would
have the LEDs lit.
The technician can put the instrument down, but powered, and proceed to
find the lighted LEDs to help repair the fault.
IDNet Shorts
If the tester is connected to a loop that is shorted, or has a much greater
current demand than expected, it will not be able to communicate with IDNet
devices. It may receive faulty data if communications are attempted. The
TSIT has a warning screen when such a channel is tested. Follow the
instructions on the screen. A test can be forced, in cases where the high
current is due to excess IDNet devices (many more than 250). But the loop
should be checked for hard shorts and stray voltages before forcing a test.
9
9
IDNAC
As with IDNet, the user should have a list of IDNAC addressable notification
appliances and addresses when checking out an IDNAC installation. The
IDNAC information can be saved, same as the IDNet information. Refer to
the section on PC Upload.
The TSIT is connected to the IDNAC channel the same way as connection to
the IDNet channel. Red to IDNAC +, Black to IDNAC-, Green to earth. If the
channel has multiple T-Taps, be sure that all wires are connected for the
scan. When dealing with multiple T-Taps, scan each one separately then
merge the results using the PC tool.
The IDNAC Test (Menu Selection Detailed Scan) performs a series of tests
on the IDNAC channel, very similar to the tests for IDNet. The TSIT will
communicate with IDNAC devices, determine candela setting, detect
duplicates, and determine if a device is set for Alert or Alarm operation. Any
device LED can be activated from the TSIT. The TSIT will not power the
appliances into alarm mode.
At the completion of the scan, review the
Total Devices: 63 summary of devices detected and compare
HORN 31 to the list of devices that are supposed to be
A/V 32
**END** installed.
ENT
Follow the on-screen options to navigate
through the menu. Press ENT to review the
results of the scan. Using the up/down
arrows, navigate to the desired option. The choices are to review Totals,
review Totals with Mode, review by Address, review Errors or activate the
Power Option. When reviewing by address, the address of the device is
shown in brackets.
Use the up/down arrows to scroll through
the list of detected addresses. Note that any [001]
missing devices will be represented by gaps A/V ALARM WALL
in the devices detected. If the scrolling goes 15Cd
F1 LED > NXT CFIG PG
from device address 3 to device address 5,
no device was detected at address 4. ENT
10
10
Mount A/V, set to 15Cd and “Alarm” mode. Note the LED can be activated by
pressing F1.
To review errors, press the menu key and
use the up/down arrows to scroll to the [003]
Errors choice. Press ENT. If multiple errors
DUPLICATE
are present, the up/down keys will scroll Press MENU to exit
through the list of all devices with abnormal
status. If duplicate devices are detected, ENT
IDNAC Tests
In the example above, two devices were set to address 003, and let’s assume
a device was missing from address 7. You’ve gone out and corrected those
two errors. If you wish to check them again to make sure there are no other
problems, you could run “IDNAC Test” again, or you could run the IDNAC
“Quick Device Srch” which is somewhat quicker, and will detect duplicate
devices. It will not detect device mode (alert or alarm) or any other
configuration information and will not isolate earth faults.
Power Option
The Power Option allows the user to power down the line after a test is
performed to save battery power.
Once this option is selected the user will have access to all the test results
but will no longer be able activate device LEDs.
To return to a normal state and regain the ability to activate device LEDs, the
user must repeat the scan.
11
11
X-Talk Test
X-TALK is shorthand for Cross-Talk. Cross-Talk is interference between two
signals, such as IDNet/MAPNET channels. The X-TALK test measures the
impedance between the POSITIVE lead of two IDNet channels. Coupling can
be caused by capacitive, resistive or inductive impedance between two
electrical signals. The greater the coupling between two electrical signals is,
the greater the likelihood of communications errors.
To Test Connect
Test Results are displayed and indicate the severity of the coupling between
the channels
HIGH Coupling - Indicates that coupling between the two IDNet channels is
too severe for reliable communications
12
12
Troubleshooting X-Talk Failure
Capacitive coupling is increased when the physical distance between two
conductors is very close over a long distance.
Things that increase coupling:
Insulation integrity is reduced by nicked or cut insulation. Exposed
conductors can have increased capacitance to other conduc-
tors.Especially earth. If shielded wire is used, it is very important to
maintain the integrity of the insulation and to prevent the shield from
contacting earth. Keep the drain wire insulated as close to the termi-
nation as possible.
Single conductor wire, such as THHN, coupling to other THHN
conductors and to conduit more than the coupling to the other
conductor in the pair. THHN or similar single conductor wires are not
recommended for IDNet or MAPNET, except for IDNet+ card used
with 4100U, 4100ES, 4010ES, 4007ES, and 4007ES Hybrid.
Inductive Coupling is caused by the electromagnetic field of a signal
interfering with the electrical signal on an adjacent conductor. Inductive
coupling interference is worse when a current changes rapidly in amplitude.
Signals that cause interference from inductive coupling include Strobes and
Voice NACs. Shielded wiring can eliminate inductive coupling by breaking the
path of the inductive field.
Resistive Coupling is usually caused by exposed conductors. Moisture and
contamination can create a conductive path to other conductors. Often, this
will be shown with the earth fault test. Nicked or cut insulation, combined with
moisture and contamination can result in electrical paths between wires.
Test IDC/zone
Test conventional Initiating Device Circuits with this menu selection. At the
top level menu, navigate to [Test IDC/zone] with the up/down arrow keys and
press ENT to select the IDC test.
If the leads are not connected to IDC wiring, the display should show the
results of the test as "Open", indicating that there is no End of Line Resistor
between the red and black leads.
13
13
The screen is updated constantly, so the
IDC Test Result tester will display the state of the circuit
Open
(Normal, Open, Short) without any further
Press MENU to exit menu commands. This makes it easy to test
multiple IDC by moving the wires from one
circuit to another.
ENT
Test NAC/Signal
Test conventional Notification Appliance Circuits with this menu selection. At
the top level menu, navigate to [Test NAC/signal] with the up/down arrow
keys and press ENT to select the NAC test.
As with the IDC test, the NAC test updates in real-time once the test has
been selected.
Because of the reverse-polarity activation of Notification Appliances, the
NACs must be tested in the Standby condition.
Connect the Red Lead to the -NAC Alarm wire. Connect the Black Lead to
the +NAC Alarm wire. Connect the Green Lead to earth.
The TSIT will detect two 10k EOLRs on the circuit as a Short condition.
For Short circuits, also check that devices are not wired backwards.
For Open circuits, check to make sure the EOLR is intact.
14
14
Saving IDNet and IDNAC Test Results
IDNet and IDNAC test results can be saved to one of the 60 memory slots.
Each slot can be tagged with a 20-character name for easy recall at a later
time. To save test results, select Save Channel Data from the main menu.
The first available memory slot will be displayed.
The second line on the LCD contains a 20-
character, user editable name for the data Save Results
Channel 06
being saved. The cursor (underline Memory Slot 06
character) is under the first letter of the F1 EDIT, ENT SAVE
name. To enter a new name, use the or
keys to change the character and the left/ ENT
Upload to PC
Once IDNet and IDNAC test data is
saved to one of the memory slots, it
may be uploaded to the TrueSTART
Analyzer PC software for review. Attach
the download harness, 733-794, to the
connector on the back of the TSIT.
Orient the connector so that the striped
edge is on the left. The other end of the
cable is a DB9 connector, which
attaches to an RS232 port on your PC.
A USB to RS232 adapter is provided in case the PC does not have RS232
ports.
15
15
Start the analyzer software on the PC and press the “Upload” button on the
button bar.
Press the Start button to begin the data upload. The progress bar indicates
the status of the upload and the TSIT display will also indicate that an upload
is in progress. All 60 slots of information will be transferred to the PC. If the
“Erase Data” checkbox is checked, the TSIT will automatically erase the
contents of the memory slots once the upload is completed. If the box is
unchecked, the TSIT memory contents will not be erased.
System Settings
The System Settings menu option includes menu choices that allow you to
view information, clear memory slots and set TSIT options.
System Settings > Backlight Timer
> Battery Selection
> Clear Memory
> Software Revision
Backlight Timer
The TSIT LCD backlight automatically turns on whenever a key is pressed.
The Backlight Timer setting controls how long the TSIT will keep the backlight
on after a key-press. The possible settings are 0-60 seconds, where 0 is
always on. Leaving the backlight always on is not recommended while
running from the internal battery.
A TSIT may be equipped with either an LED or LCD. If LED, no backlight will
be present, and this selection has no effect on operation.
16
16
Battery Selection
This setting allows you to select the number of cells for the TSIT internal
lithium-ion or lithium-polymer battery. You can select either a 4-cell (14.4 -
14.8V) or 5-cell (18-18.5V). The standard battery that is supplied with the
TSIT is a 4 -cell, 1200 mAh. The default setting is 4-cell.
Be sure that the meter is set for the correct size battery. If your meter will not
stay ON, it is possibly set for a 5 cell battery, but has a 4 cell battery installed.
The cut-off voltage for the 5-cell battery is the operating voltage for the 4-cell.
It might be necessary to power the TSIT from the AC adapter, and change
the menu option if the unit will not power ON from a fully charged 4-cell
battery.
Clear Memory
The contents of any or all memory slots can be cleared with this menu
choice. Use the orkeys to select the slot to erase, or select “Clear
All” to erase all memory slots. The ENT key must be pressed twice to erase
the memory. The second key-press is a confirmation of the requested
operation.
Software Revision
This menu screen displays the TSIT software revision.
Battery Charger
The battery charger supplied with the TSIT is for a Lithium Ion, 4 cell battery.
The charger will recharge the standard battery in about 4 hours. The green
LED indicates that the battery is fully charged. The red LED indicates that the
battery is charging.
To charge the battery:
Press the power OFF switch
Remove the battery compartment (squeeze tabs on the sides)
Unplug and remove the battery from the TSIT
Connect the charger to an AC power source
Connect the battery to the charger
The battery is fully charged when the red LED is OFF and the green
LED is ON.
The charger is rated for 100-240 VAC, 50-60 Hz. It is supplied with a US plug.
For use in other countries, a local wall outlet adapter must be used (not
supplied).
17
17
IDNet Device Types
Type Description
4009A4 4009A - 4 NACs, no repeater
4009A8 4009A - 8 NACs, no repeater
R4009A4 4009A - 4 NACs, with IDNet repeater
R4009A8 4009A - 8 NACs, with IDNet repeater
ADRPUL Addressable Pull Station
MBZAM Class B Monitor ZAM
MAZAM Class A Monitor ZAM
IAM IAM
KACPUL KAC Pull Station
SECIAM Security Monitor IAM
CANPUL 2-Stage Pull Station
RIAM Relay IAM
SIGIAM Signal IAM
IDNETISO IDNet Isolator
POWERISO IDNet Power Isolator
RELEASE Suppression Release Peripheral
TRIAM Relay IAM with T-sense input
FRIAM Relay IAM with unsupervised feedback input
MLPTIO Multi-Point I/O module
PHOTO TrueAlarm photoelectric sensor (standard base)
SPHOTO TrueAlarm photoelectric sensor (sounder base)
RPHOTO TrueAlarm photoelectric sensor (relay base)
IPHOTO TrueAlarm photoelectric sensor (isolator base)
QPHOTO TrueAlarm quick-connect photoelectric sensor
ION TrueAlarm ionization sensor (standard base)
SION TrueAlarm ionization sensor (sounder base)
RION TrueAlarm ionization sensor (relay base)
IION TrueAlarm ionization sensor (isolator base)
HEAT TrueAlarm heat sensor (standard base)
SHEAT TrueAlarm heat sensor (sounder base)
RHEAT TrueAlarm heat sensor (relay base)
IHEAT TrueAlarm heat sensor (isolator base)
COMBO TrueAlarm photo/heat sensor (standard base)
SCOMBO TrueAlarm photo/heat sensor (sounder base)
RCOMBO TrueAlarm photo/heat sensor (relay base)
ICOMBO TrueAlarm photo/heat sensor (isolator base)
MCOP TrueAlarm photo/CO sensor (standard base)
MCOPS TrueAlarm photo/CO sensor (sounder base)
18
18
IDNet Device Types (continued)
MCOI TrueAlarm ionization/CO sensor (standard base)
MCOIS TrueAlarm ionization/CO sensor (sounder base)
MCOH TrueAlarm heat/CO sensor (standard base)
MCOHS TrueAlarm heat/CO sensor (sounder base)
MCOPH TrueAlarm photo/heat/CO sensor (standard base)
MCOPHS TrueAlarm photo/heat/CO sensor (sounder base)
ANAMON 4-20ma Analog Monitor ZAM
STANDARD BASE Standard TrueAlarm base (no head inserted)
SOUNDER BASE TrueAlarm sounder base (no head inserted)
RELAY BASE TrueAlarm relay base (no head inserted)
ISOLATOR BASE TrueAlarm isolator base (no head inserted
CO BASE TrueAlarm CO sensor base (no head inserted)
CO SOUNDER TrueAlarm CO sensor base with sounder (no head inserted)
BASE
3PII Third Party IDNet Interface
UNKNOWN Unrecognized device type
DEVICE
19
19
UNKNOWN DEVICE The TSIT has encountered a device that it is unable
to recognize. This occurs for several reasons:
More than one device is set to the same
address (run a duplicate address test).
Bad device.
A newer device type was found that is not
supported by the version of your TSIT
software.
20
20
IDNAC Configuration Information
Configuration Information Displayed on Definition
Information the TSIT
Audio Output Hi High
Level (AOL)
Lo Low
FACP FACP controlled AOL
Candela Rating 15Cd 15 Candela
20Cd 20 Candela
30Cd 30 Candela
75Cd 75 Candela
75WP Cd 75 Weatherproof Candela
110Cd 110 Candela
135Cd 135 Candela
185Cd 185 Candela
FACP Cd Candela configured by FACP
Lens Color Clear Clear lens cover
Color Colored lens cover
Mode ALERT Alert mode
ALARM Alarm mode
Mount CEIL Ceiling mount
WALL Wall mount
Style Indoor Indoor device
Weatherproof UL UL Weatherproof devices
Weatherproof ULC ULC Weatherproof devices
Tone Broadband Broadband Tone
21
21
TrueSTART Data Analyzer
The TrueSTART Data Analyzer is a Windows PC application that allows you
to upload test data from the TSIT for easier review, archiving, and report
generation.
Installing the Software
Simply insert the TrueSTART Data Analyzer CD into your computer’s CD or
DVD drive and the installer will automatically start. If it does not automatically
start, use Windows Explorer to open the CD folder and double-click on the file
“setup.exe”. You must accept the license agreement to install the Analyzer
software. A Simplex/TrueSTART program group will be created allowing you
to easily run the program from the Windows start menu.
Once you start the software, refer to the built-in help in the TrueSTART data
analyzer tool for details on uploading and reviewing the data, and generating
reports.
22
22
Replacement Parts
These replacement parts for the TrueStart and the TrueSTARTII instruments
may be ordered from stock:
SKU Description
23
23
R
© 2013-2014 Tyco Fire Protection Products. All rights reserved. All specifications and
other information shown were current as of document revision date and are subject to
change without notice.
TYCO, SIMPLEX, and the product names listed in this material are marks and/or
registered marks. Unauthorized use is strictly prohibited.