This document provides an overview of the menu structure and options for configuring the operation and settings of the EnergyTech 301 Display. The menu includes sections for diagnostics, setup, parameters, signals, outputs, real-time clock, and EEPROM functions. Navigation is through a keypad membrane and menu options allow calibrating sensors, configuring communication protocols and addresses, adjusting signal processing settings, and programming output relays and current loops.

Uploaded by

Cindy Henao
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
126 views46 pages

EnergyTech 301 LCD Interface User Manual

This document provides an overview of the menu structure and options for configuring the operation and settings of the EnergyTech 301 Display. The menu includes sections for diagnostics, setup, parameters, signals, outputs, real-time clock, and EEPROM functions. Navigation is through a keypad membrane and menu options allow calibrating sensors, configuring communication protocols and addresses, adjusting signal processing settings, and programming output relays and current loops.

Uploaded by

Cindy Henao
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 46

CODEL Page 1 Operation and Maintenance

EnergyTech 301 Display Menu


Display Interface Manual

CODEL International Ltd.


Station Building, Station Road, Bakewell, Derbyshire DE45 1GE United Kingdom
t : +44 (0) 1629 814 351 f : +44 (0) 8700 566 307 e : [email protected] web : www.codel.co.uk

Issue: A Revision: 3 Date: 01/05/2019 Ref:100255


CODEL Page 2 Operation and Maintenance

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 3 Operation and Maintenance

Contents

1 Introduction ..................................................................................................................................... 1
1.1 Keypad Membrane .................................................................................................................. 1
1.2 Welcome Screen ..................................................................................................................... 1
1.3 Menu structure ........................................................................................................................ 2
1.3.1 Main Menu ...................................................................................................................... 4
1.4 Measurements ......................................................................................................................... 4
1.4.1 Select Channel................................................................................................................. 4
2 Diagnostics...................................................................................................................................... 6
2.1 Running Mode ........................................................................................................................ 6
2.2 SIG Data.................................................................................................................................. 7
2.2.1 SIG ADC Smoothed Count ............................................................................................. 7
2.2.2 Exit .................................................................................................................................. 9
2.3 Status Mode ............................................................................................................................ 9
2.3.1 Exit ................................................................................................................................ 11
3 Setup Mode ................................................................................................................................... 12
3.1 Zero Cal ................................................................................................................................ 12
3.1.1 Full Zero Cal ................................................................................................................. 13
3.1.2 Bias Ct Zero Cal............................................................................................................ 13
3.1.3 FZC Ref Time ............................................................................................................... 13
3.1.4 FZC Frequency ............................................................................................................. 14
3.1.5 Exit ................................................................................................................................ 15
3.2 Span Cal ................................................................................................................................ 15
3.2.1 Span Cal Target............................................................................................................. 15
3.2.2 InSitu Span Cal ............................................................................................................. 16
3.2.3 Exit ................................................................................................................................ 17
3.3 Parameters ............................................................................................................................. 17
3.3.1 Comms (B, D, P, S)....................................................................................................... 18
3.3.2 Protocols ....................................................................................................................... 18
3.3.3 Comms Address ............................................................................................................ 19
3.3.4 Running Mode .............................................................................................................. 20
3.3.5 Exit ................................................................................................................................ 20
3.4 Signal .................................................................................................................................... 21
3.4.1 Damping Time .............................................................................................................. 21
3.4.2 Stage 1 Gain .................................................................................................................. 21

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 4 Operation and Maintenance

3.4.3 Stage 2 Gain .................................................................................................................. 23


3.4.4 CF_m............................................................................................................................. 23
3.4.5 CF_c .............................................................................................................................. 24
3.4.6 Pressure ......................................................................................................................... 24
3.4.7 Temperature .................................................................................................................. 25
3.4.8 Exit ................................................................................................................................ 26
3.5 mA Output ............................................................................................................................ 27
3.5.1 mA Output Unit ............................................................................................................ 27
3.5.2 Span............................................................................................................................... 28
3.5.3 Zero ............................................................................................................................... 28
3.5.4 Simulation ..................................................................................................................... 29
3.5.5 Set Output ..................................................................................................................... 29
3.5.6 Exit ................................................................................................................................ 30
3.6 Relay Output ......................................................................................................................... 31
3.6.1 Rly Output Unit............................................................................................................. 31
3.6.2 Direction ....................................................................................................................... 32
3.6.3 Level ............................................................................................................................. 33
3.6.4 Exit ................................................................................................................................ 34
3.7 RTCC .................................................................................................................................... 34
3.7.1 Time .............................................................................................................................. 35
3.7.2 Date ............................................................................................................................... 36
3.7.3 Exit ................................................................................................................................ 36
3.8 EEPROM .............................................................................................................................. 37
3.8.1 Read 0xXXXX .............................................................................................................. 37
3.8.2 Write ............................................................................................................................. 38
3.8.3 Exit ................................................................................................................................ 39
3.9 PassCode ............................................................................................................................... 39
3.9.1 Set New PassCode ........................................................................................................ 39
3.9.2 Set Read Address .......................................................................................................... 39
3.9.3 Enter PassCode ............................................................................................................. 40
3.9.4 Exit ................................................................................................................................ 41

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 5 Operation and Maintenance

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 1 Operation and Maintenance

1 Introduction

This document is a user manual for LCD menu display for instrument Energy tech 301 which helps to understand and
navigate to menus and sub menus using the unit’s keypad membrane.

1.1 Keypad Membrane


Using the keypad buttons user can access the menus and sub menus. The following buttons are used to do tasks.

- To exit to main screen

- To Select/Apply/Exit

- To Scroll up/Increment

- To Scroll Down/Decrement

1.2 Welcome Screen


The welcome screen displays Codel-FM (manufacturer’s name) and EnergyTech

The system will now enter initialising, this process will last 30 seconds.

Few seconds later the measurement screen appears.

Issue: A Revision: 3 Date: 01/05/2019 Ref:100255


CODEL Page 2 Operation and Maintenance

1.3 Menu structure

Below is a visual representation of the menu and submenu path system of an EnergyTech 301 that this document is
based upon. Any further Submenus will be detailed later in the document.

Main Menu Sum menu 1 Sub menu 2

Measurements Select Channel %


Select Channel mg/m3
Select Channel mg/Nm3
Select Channel All Channels
Exit

Diagnostics Running Mode Normal/Maintenance

SIG Data ADC Smt Count


ADC Raw Count
SIG Amplitude
SIG Range Ratio
SIG DC Offset
SIG PreAmp Gain
SIG Amp Gain

Device Status Status


Performance

Exit

Setup Mode Zero Calibration Full Zero Cal


BIAS Current Cal
Zero Ref Time
Zero Frequency
Exit

Span Calibration Span Cal Target


InSitu Span Target
Exit

Parameters Comms (B,D,P,S)


Protocols
Comms Address
Running Mode
Exit

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 3 Operation and Maintenance

Main Menu Sum menu 1 Sub menu 2

Signal Damping Time


Stage 1 Gain
Stage 2 Gain
CF_m
CF_c
Pressure
Temperature
Exit

mA Outputs mA Output Unit


Span
Zero
mA Simulation
Exit

Relay Output Rly Output Unit


Direction
Level
Exit

RTCC Time
Date
Exit

EEPROM Read 0xXXXX


Write 0xXXXX to 0xXXXX
Exit

PassCode Set New PassCode


Enter PassCode
Exit

Exit

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 4 Operation and Maintenance

1.3.1 Main Menu


The Instrument has following main menus.

➔ Measurements
➔ Diagnostics
➔ Setup Mode

1.4 Measurements
Displays measurement reading. An overview of Measurements sub-menus is shown below.

Press to exit to main screen and press , the following screen appears.

1.4.1 Select Channel


Allows user to select measurement based on units. The following are the submenus.

Main Menu Sum menu 1 Sub menu 2

Measurements Select Channel %


Select Channel mg/m3
Select Channel mg/Nm3
Select Channel All Channels
Exit

Unit: %

Displays dust levels in percentage.

Use or to scroll through to % and press to select %. The display returns to measurement screen
displaying % measurement.

1.4.1.1 mg/m3
Displays dust measurement in milligrams per cubic meter.

• Use or to scroll through to mg/m3 and press to select mg/m3. The display returns
to measurement screen displaying mg/m3 measurement.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 5 Operation and Maintenance

1.4.1.2 mg/Nm3
Displays normalised dust measurement in milligrams per cubic meter.

• Use or to scroll through to mg/Nm3 and press to select mg/Nm3. The display
returns to measurement screen displaying mg/Nm3 measurement.

1.4.1.3 All Channels


Displays normalised dust measurement in milligrams per cubic meter.

• Use or to scroll through to All Channels and press to select. The display returns to
measurement screen displaying which now cycles through all the display channels.

1.4.1.4 Exit
To exit from Select Unit submenu.

• Use or to scroll through to Exit and press to exit. The display returns to
measurement screen.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 6 Operation and Maintenance

2 Diagnostics

Allows user to view diagnostic data of instrument and are read only information. An overview of Parameters
submenus is shown below.

Main Menu Sum menu 1 Sub menu 2

Diagnostics Running Mode

SIG Data

Device Status

Exit

• Press to exit to main screen, use or to scroll to diagnostic menu.

• Press to Access Diagnostics submenus.

2.1 Running Mode


Displays instrument running mode. The following are the submenus.

Main Menu Sum menu 1 Sub menu 2

Diagnostics Running Mode Normal/Maintenance

• Use or to scroll through to Running Mode.

• Press to Access Running Mode information.

• Press to exit to Diagnostics submenu.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 7 Operation and Maintenance

2.2 SIG Data


Allows user to access the diagnostic Signal Data at different amplification stages in circuit and it is read only
information. The following are the submenus

Main Menu Sum menu 1 Sub menu 2

Diagnostics SIG Data ADC Smt Count


ADC Raw Count
SIG Amplitude
SIG Range Ratio
SIG DC Offset
SIG PreAmp Gain
SIG Amp Gain

• Use or to scroll through to diagnostic Sig Data Submenu.

• Press to access submenus and use or to scroll to different stages of signal


diagnostic data.

2.2.1 SIG ADC Smoothed Count


Displays analogue signal ADC Count smoothed.

• Use or to scroll through to diagnostic Sig Data Submenu.

2.2.1.1 SIG ADC Raw Count


Displays analogue signal Raw ADC Count.

• Use or to scroll through to diagnostic Sig Data Submenu.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 8 Operation and Maintenance

2.2.1.2 SIG Amplitude


Displays analogue signal amplitude in millivolts before the signal digitisation.

• Use or to scroll through to diagnostic Sig Data Submenu.

2.2.1.3 SIG Range Ratio


Displays SIG Range Ratio which is ratio of signal amplitude to the total voltage range.

• Use or to scroll through to Sig Range Ratio Submenu.

2.2.1.4 SIG DC Offset


Displays an offset added to the signal amplitude.

• Use or to scroll through to Sig DC Offset Submenu.

2.2.1.5 SIG PreAmp Gain


Displays the amount of pre-amp gain added to the Signal.

• Use or to scroll through to Sig pre-amp gain Sub-menu.

2.2.1.6 SIG Amp Gain


Displays the amount of post-amp gain added to the Signal.

• Use or to scroll through to Sig Amp Gain Submenu.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 9 Operation and Maintenance

2.2.2 Exit

1.1.1. Press to exit SIG diagnostic data submenu.

2.3 Status Mode


Display the current system status and performance information for the device.

Main Menu Sum menu 1 Sub menu 2

Diagnostics Device Status Status


Performance

• Use or to scroll through to Device Status.

• Press to Access Device Status information.

• Use or to toggle between “Status” and “Performance”.


• Status:
• Device is performing a Bias Current Calibration

• Device is performing a Electronic Zero Calibration

• Device is performing a Full Zero Calibration

• Device is performing an Insitu Span Calibration

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 10 Operation and Maintenance

• Device is in Maintenance Mode

• Device is in Power-Up condition

• Device has no status flags set

• Performance:
• The Detector Retrieve value is below the threshold

• The Detector Retrieve value is in saturation

• The Detector Transmission value is below the threshold

• The Detector Transmission value is in saturation

• A Hardware Gain Error has been detected

• The retrieved signal has been contaminated

• A temperature error has been detected

• The Tolerance level has been exceeded

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 11 Operation and Maintenance

• Device has no status flags set

• Press to exit to Diagnostics submenu.

2.3.1 Exit
Allows user to exit to main menu.

• Use or to scroll through to Exit and press to exit to main menu.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 12 Operation and Maintenance

3 Setup Mode

Setup Mode allows users to Setup instruments. An overview of Setup mode submenus are shown below.

Main Menu Sum menu 1 Sub menu 2

Setup Mode Zero Calibration

Span Calibration

Parameters

Signal

mA Outputs

Relay Output

RTCC

Exit

• Press to exit to main screen, use or to scroll through to Setup mode.

3.1 Zero Cal


Allows user to activate Zero calibration. The following are the submenus.

Main Menu Sum menu 1 Sub menu 2

Setup Mode Zero Calibration Full Zero Cal


BIAS Current Cal
Zero Ref Time
Zero Frequency
Exit

• Press to access setup mode sub menus.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 13 Operation and Maintenance

3.1.1 Full Zero Cal


Allows user to activate Full Zero Cal.

• Use or to scroll through to full Zero Cal.

• Press to activate Full Zero Cal.


The Data valid light shall extinguish, and the user will be returned to the Zero Calibration
Sub-Menu. If the user re-enters the menu “Wait for FZC” shall be displayed. The user
should leave the menu and wait for the data valid light to resume.

3.1.2 Bias Ct Zero Cal


Allows user to activate Bias current Zero Calibration.

• Use or to scroll through to Bias Current Zero calibration.

• Press to activate Bias Current Zero calibration.

3.1.3 FZC Ref Time


Allows users to setup Zero Calibration time. The following are the submenus.

 Set Full Zero Cal.

• Use or to scroll through to FZC Ref Time press .

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 14 Operation and Maintenance

3.1.3.1 Set Full Zero Cal


Allows users to setup time for Zero Calibration.

• Press to setup FZC Ref Time.

• Use or to increment or decrement most significant digit on hours.

• Press to move cursor to next most significant digit, use or to


increment or decrement.

• Repeat the above procedure to set minutes and seconds, finally press enter to set
FZC Ref Time.

3.1.4 FZC Frequency


Allows users to setup time interval for next Full Zero Calibration time. The following are the
submenus.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 15 Operation and Maintenance

3.1.5 Exit
Allows user to exit to main menu.

• Use or to scroll through to Exit and press to exit to Setup mode submenus.

3.2 Span Cal


Allows user to activate Span calibration. The following are the submenus.

Main Menu Sum menu 1 Sub menu 2

Setup Mode Span Calibration Span Cal Target


InSitu Span Target
Exit

• Use or to scroll through to Span Cal.

• Press to access Span Cal sub menu.

3.2.1 Span Cal Target


Allows users to setup the span calibration target. The following are the submenus.

 Span Cal Target


 Set Span Cal Target

• Use or to scroll to Span Cal Target.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 16 Operation and Maintenance

3.2.1.1 Set Span Cal Target


Allows users to set the span calibration target.

• Press to setup Span calibration target.

• The most significant digit is highlighted use or to Increment or


Decrement.

• Press to move cursor to next digit and finally press to exit Span Target
Setup.

3.2.2 InSitu Span Cal


Allows user to activate Insitu Span calibration.

• Use or to scroll through to Insitu Span calibration.

• Press to activate Insitu Span calibration.


After activation, the user shall be put back to the span menu.

Entering the menu again the user shall see a message explaining the InSitu calibration is in progress.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 17 Operation and Maintenance

3.2.3 Exit
Allows user to exit to setup mode submenus.

• Use or to scroll through to Exit and press to exit to Setup mode


submenus.

3.3 Parameters
Displays Communication parameters and Communication protocols. The following are the submenus.

Main Menu Sum menu 1 Sub menu 2

Setup Mode Parameters Comms (B,D,P,S)


Protocols
Comms Address
Running Mode
Exit

• Use or to scroll through to Parameters.

Press to view Serial Communication Parameters sub menu.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 18 Operation and Maintenance

3.3.1 Comms (B, D, P, S)


Displays serial communication parameters and are read only.

B is Baud Rate.
D is Data Bits
P is Parity
S is Stop Bits

• Use or to scroll through Serial Communication parameters.

• In the above screenshot baud rate is 9600, Data bits are 8, Parity is 0 and Stop bits are 1.

• Press to exit communication parameters.

3.3.2 Protocols
Allows user to setup communication protocol. The following are the submenus.

 Protocols
 Set Protocols

• Use or to scroll through Protocols

3.3.2.1 Set Protocols


Allows user to setup communication protocol.

• Press to access Set Protocols sub menu.

• Use or to change protocol and press to exit.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 19 Operation and Maintenance

3.3.3 Comms Address


Allows user to setup communication address of instrument. The following are the submenus.

 Comms Address.
 Set Comms Address.

• Use or to scroll through Comms Address.

3.3.3.1 Set Comms Address


Allows user to set communication address of instrument.

• Press to Set Comms Address.

• Use or to Increment or Decrement.

• Press to apply address and exit.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 20 Operation and Maintenance

3.3.4 Running Mode


Allows the user to toggle between maintenance mode and normal running mode.

• Use or to scroll through to Running Mode and press to enter the setup
menu.

• Use or to toggle between the maintenance option and the normal running

option and press to confirm the decision.

• When in maintenance mode ‘Mm’ will be displayed to the right of the display
when on the display is showing the first tier of menu (measurements, diagnostics
and Setup Mode)

3.3.5 Exit
Allows user to exit to setup mode submenus.

• Use or to scroll through to Exit and press to exit to Setup mode


submenus.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 21 Operation and Maintenance

3.4 Signal
Allows the user to alter how the device handles the incoming and outgoing signals.

Main Menu Sum menu 1 Sub menu 2

Setup Mode Signal Damping Time


Stage 1 Gain
Stage 2 Gain
CF_m
CF_c
Pressure
Temperature
Exit

3.4.1 Damping Time


Allows the user to configure the signal damping time for the device.

• Press key to set the device damping time.

• Use or to increment or decrement the time.

• Press to enumerate through the menu and finalise the new Damping Time.
3.4.2 Stage 1 Gain
Allows the user to change the preliminary gain.

• Press key to set the Stage 1 Gain.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 22 Operation and Maintenance

• Use or to increment or decrement the gain (0-255).

• Press to enumerate through the menu and finalise the new Stage 1 gain.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 23 Operation and Maintenance

3.4.3 Stage 2 Gain


Allow the user to change the secondary gain.

• Press key to set the Stage 2 Gain.

• Use or to increment or decrement the gain (0-255).

• Press to enumerate through the menu and finalise the new Gain 2.

3.4.4 CF_m
Alloooolorem ipsum. To Complete (0 - 9.999)

• Press key to set the Stage 2 Gain.

• Use or to increment or decrement the gain (0.000 - 9.999).

• Press to enumerate through the menu and finalise the new CF_c Gain.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 24 Operation and Maintenance

3.4.5 CF_c
Alloooolorem ipsum. To Complete (00000 - 65535)

• Press key to set the CF_c Gain.

• Use or to increment or decrement the gain (0.000 - 9.999).

• Press to enumerate through the menu and finalise the new CF_c Gain.

3.4.6 Pressure
Alloooolorem ipsum. To Complete

• Press key to set the CF_c Gain.

• Use or to increment or decrement the gain (0.000 - 9.999).

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 25 Operation and Maintenance

• Press to enumerate through the menu and finalise the new CF_c Gain.

3.4.7 Temperature
Alloooolorem ipsum. To Complete

• Press key to set the CF_c Gain.

• Use or to increment or decrement the gain (0.000 - 9.999).

• Press to enumerate through the menu and finalise the new CF_c Gain.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 26 Operation and Maintenance

3.4.8 Exit
Allows user to exit to setup mode submenus.

• Use or to scroll through to Exit and press to exit to Setup mode submenus.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 27 Operation and Maintenance

3.5 mA Output
Allows user to setup mA output configuration. The following are the submenus.

Main Menu Sum menu 1 Sub menu 2

Setup Mode mA Outputs mA Output Unit


Span
Zero
mA Simulation
Exit

• Use or to scroll through to mA output submenu.

• Press to access mA output sub menus.

3.5.1 mA Output Unit

Allows user to set up the Units of Measurement for the mA Output. Press to access and sub menu.

• The most significant digit is highlighted use or to Increment or


Decrement in hexadecimal.

• Press to move cursor to next most significant digit. Repeat above


procedure to increment or decrement and finally press enter to set.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 28 Operation and Maintenance

3.5.2 Span
Allows user to set up the span range. The following are the submenus.

• Use or to scroll through to Span submenu.

3.5.2.1 Set Span


Allows user to set span range.

• Press to access and set up span range sub menu.

• Use or to change the span range.

• Press to move cursor to next most significant digit. Repeat above

procedure to increment and decrement and finally press to set Span


range and exit.

3.5.3 Zero
Allows user to set up range for Zero mA. The following are the submenus.

 Zero
 Set Zero

• Use or to scroll through to Zero submenu.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 29 Operation and Maintenance

3.5.3.1 Set Zero


Allows user to set range for Zero mA.

• Press to access and set up range for Zero mA.

• Use or to change the zero range.

• Press to move cursor to next most significant digit. Repeat above

procedure to increment and decrement and finally press to set Zero


range and exit.

3.5.4 Simulation
Allows the user to simulate any mA output between 4-20 mA.

 mA Simulation
 Set Output

• Use or to scroll through

3.5.5 Set Output


Allows user to set range for .

• Press to access and set up range for .

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 30 Operation and Maintenance

• Use or to change the .

• Press to move cursor to next most significant digit. Repeat above

procedure to increment and decrement and finally press to set and exit.

3.5.6 Exit
Allows user to exit to setup mode submenus.

• Use or to scroll through to Exit and press to exit to Setup mode


submenus.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 31 Operation and Maintenance

3.6 Relay Output


Allows user to setup Relay output configuration. The following are the submenus.

Main Menu Sum menu 1 Sub menu 2

Setup Mode Relay Output Rly Output Unit


Direction
Level
Exit

• Use or to scroll through to relay output submenu.

• Press to access Relay output sub menus.

3.6.1 Rly Output Unit


Allows user to set up the absolute data location for relay output configuration. It has the following
submenu.

 Abs Location
 Set Abs Location

• Use or to scroll through to Abs Location submenu.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 32 Operation and Maintenance

3.6.1.1 Set Abs Location


Allows users to set the absolute data location.

• Press to access and Set Abs Location.

• The most significant digit is highlighted use or to Increment or


Decrement.

• Press to set the output unit.

3.6.2 Direction
Allows user to set up direction of the relay. The following are the submenus.

 Direction.
 Set Direction

• Use or to scroll through to Direction submenu.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 33 Operation and Maintenance

3.6.2.1 Set Direction


Allows users to set direction status of relay.

• Press to access Set direction sub menu.

• The Relay State Options.


Normal
Reverse

• Use or to scroll through different relay state options and press


to exit.

3.6.3 Level
Allows user to set up the alarm level. The following are the submenus.

 Level
 Set Level.

• Use or to scroll through to Level submenu.

3.6.3.1 Set Level


Allows user to set alarm level.

• Press to access alarm level submenu.

• Use or to change alarm level in hexadecimal.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 34 Operation and Maintenance

• Press to move cursor to next most significant digit. Repeat above

procedure to increment or decrement and finally press to set Span range


and exit.

3.6.4 Exit
Allows user to exit to setup mode submenus.

• Use or to scroll through to Exit and press to exit to Setup mode


submenus.

3.7 RTCC
Allows user to setup real time calendar clock. The following are the sub menus.

Main Menu Sum menu 1 Sub menu 2

Setup Mode RTCC Time


Date
Exit

• Use or to scroll to RTCC.

• Press to view RTCC sub menu.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 35 Operation and Maintenance

3.7.1 Time
Allows user to setup clock. It has the following sub menu.

 Set Time.
3.7.1.1 Set Time
Allows user to set time.

• Use or to scroll through to Time sub menu.

• Press to access Set Time sub menu.

• Use or to increment or decrement most significant digit on hours.

• Press to move cursor to next most significant digit. Repeat above


procedure to increment and decrement.

• Repeat the above procedure to set minutes and seconds, finally press enter
to set time.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 36 Operation and Maintenance

3.7.2 Date
Allows user to setup date. It has the following sub menu.

 Set Date
3.7.2.1 Set Date
Allows user to set date.

• Use or to scroll through to Date.

• Press to access set date sub menu.

• Use or to increment or decrement most significant digit on date.

• Press to move cursor to next most significant digit. Repeat above procedure to
increment and decrement for month and year and finally press enter to set Date.

3.7.3 Exit
This is to exit from RTCC submenu.

• Use or to scroll to exit and press to exit.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 37 Operation and Maintenance

3.8 EEPROM
Allows user to read and write data into memory. The following are the sub menus.

Main Menu Sum menu 1 Sub menu 2

Setup Mode EEPROM Read 0xXXXX


Write 0xXXXX to 0xXXXX
Exit

• Use or to scroll through to EEPROM.

• Press to access Read sub menu. The first line is memory location 0x2180 and the
second line is data 0x1234.

3.8.1 Read 0xXXXX


Allows user to read data from memory location. The following are the sub menus.

 Read 0xXXXX
 Set Read Address.

• Press to Set Read Address.

3.8.1.1 Set Read Address


Allows user to enter memory location to read data.

• Use or to increment or decrement most significant digit in


hexadecimal.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 38 Operation and Maintenance

• Press to move cursor to next most significant digit. Repeat above

procedure to increment or decrement and finally press to set read


memory location.

3.8.2 Write
Allows user to write data into memory location. The following are the sub menus.

 Write 0xXXXX to 0xXXXX (Write DATA to Memory)


 Write 0xXXXX to 0xXXXX (Write DATA to Memory).

• Use or to scroll through to Write 0xXXXX to 0xXXXX.

3.8.2.1 Write 0xXXXX to 0xXXXX


Allows user to enter data to memory location.

• Press to access Write sub menu.

• Use or to increment or decrement most significant digit in


hexadecimal.

• Press to move cursor to next most significant digit. Repeat above

procedure to increment or decrement and finally press to set data


0x01F8 in memory location 0x7F80.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 39 Operation and Maintenance

3.8.3 Exit
Allows user to exit to setup mode submenus.

• Use or to scroll through to Exit and press to exit to Setup mode


submenus.

3.9 PassCode
Allows user to assign a new PassCode for setup menu entry.

Main Menu Sum menu 1 Sub menu 2

Setup Mode PassCode Set New PassCode


Enter PassCode
Exit

• Use or to scroll through to PassCode.

• Press to access Read sub menu. The first line is memory location 0x2180 and the
second line is data 0x1234.

3.9.1 Set New PassCode


Allows user to set a new passcode to enter the setup menu

 Set New PassCode


 Set New PassCode.

• Press to Set New PassCode.

3.9.2 Set Read Address


Allows user to enter memory location to read data.
Issue: A Revision:3 Date: 01/05/2019 Ref:100255
CODEL Page 40 Operation and Maintenance

• Use or to increment or decrement most significant digit in


hexadecimal.

• Press to move cursor to next most significant digit. Repeat above

procedure to increment or decrement and finally press to set read


memory location.

• Pressing shall move the user out of the setup menu and
back to the main screen.

3.9.3 Enter PassCode


Main password screen for Setup menu access.

 Enter PassCode
 Enter PassCode

• Use or to scroll through to Enter PassCode.

• Press to access Write sub menu.

• Use or to increment or decrement most significant digit in


hexadecimal.

• Press to move cursor to next most significant digit. Repeat above

procedure to increment or decrement and finally press to set the


PassCode and return to the initial menu screen.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255


CODEL Page 41 Operation and Maintenance

3.9.4 Exit
Allows user to exit to setup mode submenus.

• Use or to scroll through to Exit and press to exit to Setup mode


submenus.

Issue: A Revision:3 Date: 01/05/2019 Ref:100255