Cobas Liat User Manual SW 3 1
Cobas Liat User Manual SW 3 1
Operator’s Manual
Version 7.2
©2017 Roche
Roche Diagnostics
9115 Hague Rd
P.O. Box 50457
Indianapolis, IN 46250-0457
USA
www.roche.com
https://usdiagnostics.roche.com
i
Publication information
Edition notice This publication is intended for operators of the cobas® Liat®
System.
Every effort has been made to ensure that the information is correct
at the time of publishing. Not all functionality described in this
manual may be available to all users. Roche Diagnostics reserves
the right to change this publication as necessary and without notice
as part of ongoing product development. Such changes may not
immediately be reflected in this document.
Screenshots Any screenshots in this publication are added exclusively for the
purpose of illustration. Configurable and variable data such as
parameters, results, path names etc. visible therein must not be
used for laboratory purposes.
All other product names and trademarks are the property of their
respective owners.
Document availability This document is available on the Roche Diagnostics USA website
at https://usdiagnostics.roche.com.
Roche Diagnostics
Operator’s Manual US Supplement
08276340001
ii
Roche Diagnostics
Operator’s Manual US Supplement
08276340001
US Supplement US - 1
Roche warrants that this product will meet the specifications stated
in the labeling when used in accordance with such labeling and
will be free from defects in material and workmanship until the
expiration date printed on the label.
Roche Diagnostics
Operator’s Manual US Supplement
08276340001
US - 2 US Supplement
Service policy questions Questions relating to your service policy can be referred to your
local Roche representative or the regional service manager by
calling the appropriate Customer Support Center number.
Customer Training
Ordering Information
Roche Diagnostics
Operator’s Manual US Supplement
08276340001
US Supplement US - 3
Customer Communications
Urgent Medical Device Correction or Removal Urgent Medical Device Correction or Removal is printed on Roche
Diagnostics letterhead with the words Urgent Medical Device
Correction or Urgent Medical Device Removal. These require
immediate attention by the analyzer operator and are to be filed for
reference until further notice.
Customer Letters Customer Letters are printed on Roche letterhead and contain
general information or attachments that you may place in your
files. You may discard Customer Letters after reading and following
any applicable instructions contained in them.
CLIA/CLSI Compliance
Roche Diagnostics
Operator’s Manual US Supplement
08276340001
US - 4 US Supplement
Please provide your name and account number when contacting us.
Your account number is provided to you by a Roche representative
when your analyzer is installed.
Email [email protected]
Roche Diagnostics
Operator’s Manual US Supplement
08276340001
cobas® Liat® System
Operator's Manual
Version 7.2
Publication information
Where to find information The cobas® Liat® System Operator’s Manual contains
all information about the product, including the following:
• Safety
• Installation
• Routine operation
• Maintenance and calibration
• Troubleshooting information
• Configuration information
• Background information
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
3
! General attention
To avoid incorrect results, ensure that you are familiar
with the instructions and safety information.
r Pay particular attention to all safety notices.
r Always follow the instructions in this publication.
r Do not use the software in a way that is not described
in this publication.
r Store all publications in a safe and easily retrievable
place.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
4
Feedback Every effort has been made to ensure that this publication
fulfills the intended use. All feedback on any aspect of
this publication is welcome and is considered during
updates. Contact your local Roche representative, should
you have any such feedback.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
5
Contact addresses
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
6
Table of contents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Symbols and abbreviations . . . . . . . . . . . . . . . . . . . 9
What is new in publication version 7.2 . . . . . . . . . 12
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Installing the analyzer . . . . . . . . . . . . . . . . . . . . . . . 17
About the analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
About the analyzer functions . . . . . . . . . . . . . . . . . 22
Overview of the analyzer operating elements. . . . 24
About assays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Overview of the touch screen . . . . . . . . . . . . . . . . . 26
Touch screen elements . . . . . . . . . . . . . . . . . . . 26
Common user actions . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Starting the analyzer . . . . . . . . . . . . . . . . . . . . . . . . 28
Logging on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Changing your password. . . . . . . . . . . . . . . . . . . . . 31
Assigning or changing your badge barcode. . . . . 34
Selecting items on the screen. . . . . . . . . . . . . . . . . 36
Selecting predefined values . . . . . . . . . . . . . . . . . . 36
Entering text and numbers . . . . . . . . . . . . . . . . . . . 37
Scanning barcodes. . . . . . . . . . . . . . . . . . . . . . . . . . 37
About locking the analyzer . . . . . . . . . . . . . . . . . . . 39
Locking the analyzer manually . . . . . . . . . . . . . . . . 41
Unlocking the analyzer . . . . . . . . . . . . . . . . . . . . . . 42
Switching off the analyzer . . . . . . . . . . . . . . . . . . . . 44
Performing a run. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Running an assay . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Aborting an assay run . . . . . . . . . . . . . . . . . . . . . . . 49
Result handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
About results . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Reviewing results . . . . . . . . . . . . . . . . . . . . . . . . 53
Exporting and deleting results . . . . . . . . . . . . . 57
Setting up the analyzer . . . . . . . . . . . . . . . . . . . . . . . . . 60
Defining values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Installing and updating assays . . . . . . . . . . . . . . . . 62
Installing or updating assays from a USB key 62
Installing or updating an assay from the
Roche remote service platform. . . . . . . . . . . . . 64
Installing an assay from a local source . . . . . . 66
Registering assays . . . . . . . . . . . . . . . . . . . . . . . 68
Adding an assay tube lot . . . . . . . . . . . . . . . . . . 70
Deleting assay tube lots. . . . . . . . . . . . . . . . . . . 74
Uninstalling assays . . . . . . . . . . . . . . . . . . . . . . . 76
Viewing and printing assay tube lot information . 77
Defining system settings . . . . . . . . . . . . . . . . . . . . . 78
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
7
System settings . . . . . . . . . . . . . . . . . . . . . . . . . . 79
About authentication modes. . . . . . . . . . . . . . . 80
Network definitions . . . . . . . . . . . . . . . . . . . . . . . . . 81
Network resources . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Defining network resources . . . . . . . . . . . . . . . 82
Using the share lot function . . . . . . . . . . . . . . . 83
Connections to a host system. . . . . . . . . . . . . . . . . 85
Defining settings for host systems . . . . . . . . . . 85
About data exchange with a DMS . . . . . . . . . . 86
Connecting the analyzer to a DMS . . . . . . . . . 88
Managing users . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Overview of user account access roles . . . . . . 90
About user IDs and passwords. . . . . . . . . . . . . 91
Defining new users. . . . . . . . . . . . . . . . . . . . . . . 91
Changing user information . . . . . . . . . . . . . . . . 94
Deleting a user account. . . . . . . . . . . . . . . . . . . 95
Updating the software . . . . . . . . . . . . . . . . . . . . . . . 97
Registering the software . . . . . . . . . . . . . . . . . . . . . 100
Viewing software and hardware information . . . . 102
Viewing information about the analyzer and
software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Viewing system information . . . . . . . . . . . . . . . 103
Quality control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Maintenance and calibration . . . . . . . . . . . . . . . . . . . . 105
About self-check . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
About auto monitoring. . . . . . . . . . . . . . . . . . . . . . . 105
About auto adjustment . . . . . . . . . . . . . . . . . . . . . . 106
About service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Calibrating the touch screen. . . . . . . . . . . . . . . . . . 107
Cleaning the analyzer. . . . . . . . . . . . . . . . . . . . . . . . 107
Cleaning the outside of the analyzer . . . . . . . . 109
Cleaning spillages or leakages from an assay
tube. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
List of error messages . . . . . . . . . . . . . . . . . . . . . . . 111
Generating a backup for troubleshooting
purposes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Viewing and printing the event log . . . . . . . . . . . . 124
Resetting the lost default administrator
password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
Cleaning the window of the barcode reader . . . . 127
Resolving a tube entry door error before the start
of a run. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Resending results manually . . . . . . . . . . . . . . . . . . 128
Restarting an unresponsive analyzer . . . . . . . . . . . 129
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Technical characteristics . . . . . . . . . . . . . . . . . . . . . 130
Environmental conditions . . . . . . . . . . . . . . . . . . . . 130
Storage and transport conditions. . . . . . . . . . . . . . 131
Standard supplies. . . . . . . . . . . . . . . . . . . . . . . . . . . 131
Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
8
Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Preface 9
Intended use
Preface
Intended use
The cobas® Liat® Analyzer is an automated nucleic acid
test instrument, that when used together with
cobas® Liat® assay tubes to perform in vitro diagnostic
tests, is referred to as the cobas® Liat® System. In
conjunction with a cobas® Liat® assay tube, the
instrument performs reagent preparation, target
enrichment, inhibitor removal, nucleic acid extraction,
polymerase chain reaction (PCR) amplification, real-time
detection, and result interpretation to automate the
detection or quantification of nucleic acid targets in a
biological sample.
y Product names
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
10 Preface
Symbols and abbreviations
Symbol Explanation
p Task. Used in cross-references to tasks.
Figure. Used in figure titles and cross-
w
references to figures.
Table. Used in table titles and cross-references
y
to tables.
Equation. Used in cross-references to
z
equations.
Code example. Used in code titles and cross-
k
references to codes.
y Symbols used in the publication
Humidity limits.
Keep upright.
Keep dry.
Recyclable.
Abbreviation Definition
AD Amplification and detection
ANSI American National Standards Institute
y Abbreviations
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Preface 11
Symbols and abbreviations
Abbreviation Definition
CFR Code of Federal Regulations
CISPR Comité International Spécial des
Perturbations Radioélectriques (International
Special Committee on Radio Interference)
CSA Canadian Standards Association
DML Device messaging layer
DMS Data management system
EC European Community
EN European standard
FTP File transfer protocol
HIS Hospital information system
HL7 Health Level Seven
IEC International Electrical Commission
IP Internet provider
IVD In vitro diagnostic
LAN Local area network
LIS Laboratory information system
n/a not applicable
NA Not available
NAT Nucleic acid test
PCR Polymerase chain reaction
POCT Point of care testing
RF Radio frequency
QC Quality Control
SD Standard deviation
SL Shared lot
SNTP Standard network time protocol
UPS Uninterruptible power supply
URI Uniform resource identifier
WEEE Waste Electrical and Electronic Equipment
WLAN Wireless LAN
WPA2 Wireless protected access 2
y Abbreviations
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
12 Preface
What is new in publication version 7.2
Audit trail Information about the audit trail feature was added.
u About the audit trail (23)
Logon procedures The procedure how to log on was reworked to cover the
different authentication modes. The procedure how to
change your password was reworked. A procedure how
to assign or change your badge barcode was added.
u Logging on (29)
u Changing your password (31)
u Assigning or changing your badge barcode (34)
User management Requirements for user IDs and passwords were added.
The procedures how to define a new user and how to
change user information were reworked to cover the new
features (e.g., locking users, assigning badge barcodes,
change password at first logon).
u Managing users (90)
Export and delete results The former functions for moving results and deleting
results were combined to one function for exporting and
deleting results. The corresponding procedures were
combined and reworked.
u Exporting and deleting results (57)
Sharing assay tube lots Information about ways to share assay tube lots between
different analyzers was added.
u Sharing assay tube lots (70)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Preface 13
What is new in publication version 7.2
DMS connectivity Information about the data exchange with a DMS was
added. A procedure how to connect the analyzer to a
DMS was added. Information how to disconnect the
analyzer from a host was added.
u Connections to a host system (85)
Changed user interface The user interface was changed with software version 3.1.
Screenshots and text were updated to reflect the
changes.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
14 Preface
Safety instructions
Safety instructions
Safety classification The safety precautions and important user notes are
classified according to the ANSI. Familiarize yourself with
the following meanings and icons:
! WARNING
Warning…
r …indicates a hazardous situation that, if not avoided,
could result in death or serious injury.
! CAUTION
Caution…
r …indicates a hazardous situation that, if not avoided,
could result in minor or moderate injury.
NOTICE
Notice…
r …indicates a hazardous situation that, if not avoided,
may result in damage to the analyzer.
Safety summary To avoid serious injury, read and comply with the
following safety precautions.
! WARNING
Insufficient knowledge and skills
As an operator, ensure that you know the relevant safety
precaution guidelines and standards and the information
and procedures contained in these instructions.
r Only trained personnel may operate the
cobas® Liat® Analyzer. Operators must have received
and read the comprehensive instructions for the use,
quality control, and care of the
cobas® Liat® Analyzer.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Preface 15
Safety instructions
! WARNING
Infectious materials
There is a potential risk of infection. Staff using the
cobas® Liat® Analyzer to perform tests on patient
samples must be aware that any object coming into
contact with biologic specimens is a potential source of
infection.
r Use lab gloves.
r Use a new disposable pipette (provided) for each
sample.
r Follow all health and safety regulations applicable to
your institution.
Laser transmission
Label Explanation
There is a danger of contact with
laser light or severe damage to the
eyes. Do not stare into the laser
transmitter.
! WARNING
Blindness due to intense barcode reader light
The intense light of a laser barcode reader may severely
damage your eyes or result in exposure to hazardous
radiation.
r Do not stare into the beam of a laser barcode reader.
r Do not remove covers from barcode readers.
r Do not perform any maintenance actions on barcode
readers. If you experience problems with the barcode
readers, contact your Roche representative.
r Perform only the procedures described in operating
instructions. Performing unauthorized procedures may
result in exposure to hazardous radiation.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
16 Preface
Safety instructions
! CAUTION
Electromagnetic interference
Malfunction of system and incorrect results due to
interfering electromagnetic fields
This system is designed and tested to CISPR 11 Class A
standard. In a domestic environment it may cause radio
interference, in which case, you may need to take
measures to mitigate the interference.
r The electromagnetic environment should be evaluated
prior to operation of the device.
r Do not operate this system in close proximity to
sources of strong electromagnetic fields (for example
unshielded intentional RF sources), as they may
interfere with proper operations.
Biohazard warning
This label indicates that there are potential biohazards
within the vicinity of this label. The user is responsible
for cleaning the area if biohazardous material was
spilled. Follow Good Laboratory Practices for working
with biohazardous materials
General warning
Potential hazards located near this label may lead to
death or serious injury
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Installation 17
Installing the analyzer
Installation
! CAUTION
Incorrect results or malfunctions due to incorrect
installation
Performing installation actions other than those
mentioned in this documentation may lead to malfunction
and incorrect results.
r Do not carry out any installation actions that are not
described in this documentation or for which you were
not appropriately trained.
! CAUTION
Injury by tripping over cables
Cables between the analyzer and wall outlets cause a
potential tripping hazard.
r Position the analyzer as close to the wall outlet as
possible, taking care to ensure sufficient room for
ventilation (at least 10 cm).
r Take care not to trip if you must walk behind the
analyzer.
! CAUTION
Power interruption
A power failure or momentary drop in voltage may
damage the system or lead to data loss.
r It is recommended to use an uninterruptible power
supply (UPS).
r Ensure periodic maintenance of the UPS.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
18 Installation
Installing the analyzer
NOTICE
Malfunction due to use of inappropriate power
supply
Using an inappropriate power supply may cause
malfunction.
r Only use the supplied power supply.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Installation 19
Installing the analyzer
2
2 Connect the power supply with the analyzer.
I Only use voltage of 100-240 V AC, 50-60 Hz.
4
4 Press and release the power button to power on the
analyzer.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
20 Installation
Installing the analyzer
1 When you start the analyzer for the first time, make
sure the USB printer is not connected.
3
3 Connect the printer to the analyzer.
5
5 Choose the printer type and connection:
• For a laser printer, choose the PCL Laser item and
choose the LPT1: option.
• For an inkjet printer, choose the PCL Inkjet item
and choose the LPT1: option.
6
6 Choose the Select button. Choose the Printing
Mode item:
• For a grayscale printer, keep the default
Grayscale option.
• For a color printer, choose the Color option.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Installation 21
Installing the analyzer
1
1 Connect the analyzer to the local data network.
I Use a standard Ethernet data cable and connect to
an appropriate wall connector or switch.
3
3 Choose the printer type and connection:
• For a laser printer, choose the PCL Laser item and
choose the NET0: option.
• For an inkjet printer, choose the PCL Inkjet item
and choose the NET0: option.
5
5 Enter Server Name, Share Name, User Name, and
Password settings.
I For information about the settings for the print
server, contact your network administrator.
6
6 Choose the Printing Mode item:
• For a grayscale printer, keep the default
Grayscale option.
• For a color printer, choose the Color option.
u Related topics
• Logging on (29)
• Changing your password (31)
• System settings (79)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
22 About the analyzer
About the analyzer functions
Insert and test automation The cobas® Liat® System automates all testing
processes from sample preparation to amplification and
real-time detection. By reducing complex nucleic acid
testing to three simple steps, cobas® Liat® technology
enables non-specialized personnel to conduct
sophisticated testing.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
About the analyzer 23
About the analyzer functions
Rapid testing This technology uniquely uses liquid flow and mixing to
enhance reaction rates while allowing for relatively large
reaction volumes. For example, the cobas® Influenza A/B
Nucleic Acid Test for use on the cobas® Liat® System,
using a nasopharyngeal swab sample, takes
approximately 20 minutes from sample collection to result
generation.
NOTICE
Damage to the analyzer due to using unauthorized
assay tubes
The use of unauthorized or unapproved assay tubes or
accessories may damage the analyzer. Damage caused
by unauthorized tubes or accessories voids the analyzer
warranty.
r Only use assay tubes listed in the standard supplies
table in this publication.
Running assays The testing process has been condensed to a few simple
steps:
• Preparing a patient sample.
• Transferring the sample into an assay tube.
• Scanning the assay tube barcode label.
• Inserting the assay tube into the analyzer.
• The analyzer automatically runs all the required assay
steps and reports the test results.
• Disposing of the assay tube.
• Reviewing the result.
About the audit trail The system logs each user action that triggers a system
change in an audit trail entry together with user
information and time stamp.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
24 About the analyzer
Overview of the analyzer operating elements
You can export the audit trail using the advanced tools
key.
u For information on exporting the audit trail, refer to the
cobas® Liat® Advanced Tools Key Guide.
u Related topics
• Setting up the analyzer (60)
• Installing or updating assays from a USB key (62)
• Installing or updating an assay from the Roche remote
service platform (64)
• Installing an assay from a local source (66)
• Adding an assay tube lot (70)
A
F
B
C G
w Analyzer components.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
About the analyzer 25
About assays
Function buttons The function buttons are located beneath the touch
screen and are aligned with the touch screen buttons.
Navigation buttons You use the 4 buttons (left, right, up, and down) to
highlight an item for selection, scroll through menus,
position the cursors, and toggle selections.
Barcode reader The barcode reader aims the scanning beam down in
front of the analyzer. Choosing the Scan button activates
the reader for approximately 5 seconds to read barcodes.
Tube entry door Opens and closes automatically during an assay run so
that you can insert and remove the assay tube.
Touch screen You touch the items on the screen with your finger (also
when wearing lab gloves), or the stylus included with the
analyzer.
About assays
The analyzer runs a specific assay as determined by the
scanned barcode of an assay tube.
u Related topics
• Installing an assay from a local source (66)
• Adding an assay tube lot (70)
• Running an assay (46)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
26 About the analyzer
Overview of the touch screen
NOTICE
Damage to the touch screen due to using sharp
objects
Using sharp objects on the touch screen may damage the
screen and lead to malfunction.
r Do not use sharp objects for operating the touch
screen.
A
B
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
About the analyzer 27
Overview of the touch screen
User
Supervisor
Administrator
• Current main menu or assay name, including
indication whether the share lot function is enabled
(SL).
• User ID
Note that user IDs shown in screenshots in this
publication are for illustration purposes only.
• Possibly alert icon:
Warning
Service
• In case of a configured DMS or HL7 connection,
connectivity monitor icon:
Touch screen buttons The touch screen buttons vary depending on the current
content of the screen.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
28 Common user actions
Starting the analyzer
In this section
Starting the analyzer (28)
Logging on (29)
Changing your password (31)
Assigning or changing your badge barcode (34)
Selecting items on the screen (36)
Selecting predefined values (36)
Entering text and numbers (37)
Scanning barcodes (37)
About locking the analyzer (39)
Locking the analyzer manually (41)
Unlocking the analyzer (42)
Switching off the analyzer (44)
NOTICE
Damage to the analyzer due to inappropriate
handling
Inserting an assay tube into the analyzer prior to
completion of initialization results in damage to the tube,
possible damage to the analyzer and failure of the assay
tube to provide a result.
r Do not insert assay tubes or any other object into the
tube chamber while it is starting up.
n Up to several minutes
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Common user actions 29
Logging on
1
1 Press the power button.
I If the temperature inside the analyzer is low, the
analyzer warms up until the internal temperature
reaches the threshold temperature. This may take
some time.
f Power indicator lights up
f Touch screen lights up
f Tube entry door opens and closes
f Actuators inside the analyzer move
f Self-checks are performed
2
2 Wait until the login panel is displayed.
I Initialization may take several minutes.
Logging on
At logon, the analyzer prompts you for your user ID and
password, for your badge barcode and password, or for
your badge barcode only.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
30 Common user actions
Logging on
1
1 On the login panel, choose the Login button.
2
2 To enter your user ID, do one of the following:
• Enter your user ID manually and choose the OK
button.
• Place your user ID barcode parallel to the analyzer
directly beneath the barcode reader. Choose the
Scan button.
3
3 Enter your password and choose the OK button.
f In the title bar of the Main menu, your access role
and user ID are indicated.
1
1 On the login panel, choose the Login button.
I If a badge barcode authentication mode is set on
the analyzer, the barcode reader is enabled
automatically.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Common user actions 31
Changing your password
2
2 Scan your badge barcode.
I If you want to log on with your user ID instead, wait
until the time to scan your badge barcode runs out.
On the message dialog box, choose the Manual
button.
3
3 If required, enter your password and choose the OK
button.
f In the title bar of the Main menu, your user ID and
access role are indicated.
u Related topics
• Scanning barcodes (37)
• About authentication modes (80)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
32 Common user actions
Changing your password
Changing the password of another user Users with Supervisor or Administrator access role can do
the following:
• Change the password of another user up to their own
access role.
• Request a password change at next logon for another
user up to their own access role.
2
2 Choose the Info button.
3
3 Choose the Password item.
I If you are logged on with User access role, you
have access to the Password item and the Badge
Barcode item only.
4
4 Enter your current password and choose the OK
button.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Common user actions 33
Changing your password
5
5 Enter your new password and choose the OK button.
I For passwords, you can use alphanumeric
characters, symbols and/or spaces.
Passwords are restricted in length to 4–20
characters.
6
6 To confirm your new password, enter the password
again and choose the OK button.
f Your password is changed.
1
1 When prompted to change your password, choose the
OK button.
2
2 Enter your new password and choose the OK button.
I For passwords, you can use alphanumeric
characters, symbols and/or spaces.
Passwords are restricted in length to 4–20
characters.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
34 Common user actions
Assigning or changing your badge barcode
3
3 To confirm your new password, enter the password
again and choose the OK button.
f Your password is changed.
Changing the badge barcode of another user Users with Supervisor or Administrator access role can
assign or change the badge barcodes of other users up to
their own access role.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Common user actions 35
Assigning or changing your badge barcode
2
2 Choose the Info button.
3
3 Choose the Badge Barcode item and choose the
Change button.
I If you are logged on with User access role, you
have access to the Password item and the Badge
Barcode item only.
4
4 To assign a badge barcode, scan your badge barcode
twice.
5
5 To change your badge barcode, do one of the
following:
• To remove an assigned badge barcode, choose the
Remove button.
• To replace your badge barcode, choose the
Replace button. Scan your badge barcode twice.
u Related topics
• Scanning barcodes (37)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
36 Common user actions
Selecting items on the screen
1
1 Use the and navigation buttons to select
the desired item.
I Alternatively, you can touch the item on the
screen.
2
2 Choose the Select button.
f This may lead to detailed information on the
selected item or to a new panel. It may also
confirm an action.
2
2 Use the Left and Right navigation buttons to
select the desired value.
I The value that can be chosen is displayed between
green arrows, for example .
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Common user actions 37
Entering text and numbers
Key Function
ABC Enter capital letters.
abc Enter lower case letters.
123 Switch to number/symbol entry.
abc Switch to letter entry.
ABC
g Delete the character to the left of the cursor.
y Selected keys on the alphanumeric keyboard
Key Function
CLR Delete the character to the left of the cursor.
Caps Switch to letter entry.
Symb Switch to symbol entry.
Enter Confirm the entry and close the numeric keyboard.
y Selected keys on the numeric keyboard
Scanning barcodes
You use the barcode reader for scanning assay tube,
sample, package insert, control, user ID, and badge
barcodes.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
38 Common user actions
Scanning barcodes
1
1 Choose the Scan button.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Common user actions 39
About locking the analyzer
1
2
2
2 NOTICE Placing the barcode inappropriately may lead
to reading errors or malfunction.
2 Place the barcode parallel to the analyzer directly
beneath the barcode reader and make sure the
scanner light extends beyond the barcode on both
sides.
4
4 If the barcode still cannot be read, choose the Enter
button and enter the barcode text manually.
Locking the analyzer The analyzer locks automatically or you can lock it
manually. In each case, the walk-by panel is displayed.
w Walk-by panel.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
40 Common user actions
About locking the analyzer
Unlocking the analyzer When your account is locked, unlock it by entering your
password or by scanning your badge barcode. You are
returned to the panel in which you were working when
the analyzer locked.
Logging on to use the analyzer When another user’s account is locked, you must log on
to use the analyzer. If the previous user has unsaved data,
it is lost when you log on.
Auto-lock and assay runs During an assay run, the auto-lock function is suspended
until the sample preparation, amplification, and detection
have completed. It is not possible to lock the analyzer
manually until the assay run is finished.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Common user actions 41
Locking the analyzer manually
u Related topics
• Locking the analyzer manually (41)
• System settings (79)
• Unlocking the analyzer (42)
• Logging on (29)
• Running an assay (46)
You can lock the analyzer manually at any time with the
following exceptions:
• When a message box is displayed.
• When a process is running (e.g. an assay run).
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
42 Common user actions
Unlocking the analyzer
1
1 In the top left corner of the panel, choose the user
icon.
f The walk-by panel is displayed.
u Related topics
• Unlocking the analyzer (42)
1
1 On the walk-by panel, choose the Unlock button.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Common user actions 43
Unlocking the analyzer
2
2 Enter your password and then choose the OK button.
f You are returned to the panel in which you were
working when the analyzer locked.
1
1 On the walk-by panel, choose the Unlock button.
I If the Barcode & Password authentication mode
or the Barcode authentication mode is set on the
analyzer, the barcode reader is enabled
automatically.
2
2 Scan your badge barcode.
f You are returned to the panel in which you were
working when the analyzer locked.
1
1 On the walk-by panel, choose the Login button.
f A warning message is displayed.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
44 Common user actions
Switching off the analyzer
2
2 To continue and log on, choose the Yes button.
I Choosing the No button, returns you to the walk-
by panel.
u Related topics
• About locking the analyzer (39)
• Locking the analyzer manually (41)
! CAUTION
Loss of power
r Do not turn power off while the analyzer is running an
assay.
r In the event of power loss while an assay is running,
do not attempt to remove the assay tube.
When power is restored, power on the analyzer and
only remove the assay tube when the message
“Please remove the tube slowly and carefully” is
displayed.
1
1 Press and hold the power button until both the power
indicator light and the touch screen turn dark.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Performing a run 45
Switching off the analyzer
Performing a run
Safety u Make sure you are familiar with the safety instructions
in Safety instructions (14).
! CAUTION
Incorrect results due to using inappropriate assay
tubes
Using non certified assay tubes may lead to incorrect
results.
r Only use cobas® Liat® Assay Tubes.
r Never reuse assay tubes.
! CAUTION
Incorrect results or malfunction due to
inappropriate handling of assay tubes
Inappropriate handling of assay tubes may lead to
incorrect results or malfunction.
r Do not remove or insert an assay tube while the
analyzer is performing an assay.
r Always use the barcode of the assay tube sleeve that
belongs to the assay tube that you are using.
r Do not break any seals of the assay tube. Take
particular care when adding sample.
r Do not force or rotate the assay tube when inserting it.
r Do not reuse spent assay tubes.
NOTICE
Malfunction due to inappropriate handling of the
analyzer
Inappropriate handling of the analyzer may lead to
malfunction.
r Do not manually open or handle the tube entry door,
unless instructed to do so on the screen.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
46 Performing a run
Running an assay
Running an assay
The analyzer runs a specific assay as determined by the
scanned barcode of an assay tube.
d m Assay tube
m Disposable pipette supplied with the assay tube
m Sample ID barcode
r To run an assay
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Performing a run 47
Running an assay
3
3 On the Main menu, select the Run Assay option and
choose the Select button.
4
4 Choose the Scan button.
5
5 Scan the assay tube barcode.
I Do not remove the assay tube sleeve.
6
6 Choose the Scan button again and then scan the
sample ID barcode.
I If necessary, you can enter the sample ID
manually. No spaces should be used when you
enter the values.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
48 Performing a run
Running an assay
9
9 Choose the Scan button and scan the assay tube
barcode again.
I If by mistake you removed the assay tube sleeve
completely, make sure that the serial numbers and
the lot numbers on the assay tube and on the
sleeve match, before you scan the barcode.
10
10 Remove the assay tube sleeve and insert the assay
tube into the analyzer until the tube clicks into place.
I The assay tube only fits in one way.
f The tube entry door closes automatically and
processing begins.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Performing a run 49
Aborting an assay run
11
11 Check the progress, if required.
I In the subtitle bar, a run number (for example
“R00056”) and the sample ID are displayed.
Choose the ViewData button to view messages or
to display the growth curve.
To abort the assay run, choose the Abort button.
12
12 When the assay run is complete, the tube entry door
opens. A message asks you to remove the assay tube.
Remove the assay tube and dispose of it in
accordance with the disposal policy of your institution.
f If configured, results will be sent to a host
automatically.
13
13 Do one of the following:
• To review the results, choose the Report button.
• To return to the Main menu - and for example to
start another assay - choose the Main button.
• To view the growth curve or messages, choose the
ViewData button.
u Related topics
• Reviewing results (53)
• About assays (25)
• Scanning barcodes (37)
• Aborting an assay run (49)
1
1 On the assay run status panel, choose the Abort
button.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
50 Performing a run
Aborting an assay run
2
2 On the confirmation message, choose the Yes button.
3
3 Enter your password and choose the OK button.
• Depending on the setting of the analyzer, scan
your badge barcode to abort the run.
I You can choose the Back button to return to the
previous panel.
f The analyzer stops assay processing.
f The analyzer writes a result report.
4
4 When asked to do so, remove the assay tube.
6
6 Do one of the following:
• To view the result report, choose the Report
button.
• To return to the Main panel, choose the Main
button.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Performing a run 51
Result handling
Result handling
In this section
About results (52)
Reviewing results (53)
Exporting and deleting results (57)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
52 Performing a run
Result handling
About results
The way results are calculated and displayed depends on
the intended use of each individual assay.
Result reports
Result report
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Performing a run 53
Result handling
Reviewing results
Users with Supervisor or Administrator access role can
accept or reject results.
1
1 From the assay status panel, you can do the following:
• Display the growth curve (see step 2).
• Display and print run status information (see step
3).
• Review and print the result report (see steps 4 to
9).
I You can perform these actions when the assay run
is complete.
2
2 To display the growth curve, on the assay status panel,
choose the ViewData button. Choose the Real-time
PCR option.
I From the graph panel, you can print the graph.
You can zoom into an area of the graph by
dragging a finger, or the stylus, across the area
into which you want to zoom.
To zoom out, touch the screen and from the drop-
down list that displays do one of the following.
Either, choose the Zoom Out option to zoom out
incrementally, or the Zoom Normal option to the
original magnification.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
54 Performing a run
Result handling
3
3 To display run status information, on the assay status
screen, choose the ViewData button. Choose the
Message option.
I From the message panel, you can print the
messages.
4
4 To view the result report, on the assay status panel,
choose the Report button.
5
5 On the Result Report panel, if the result information
does not fit into one screen, use the and
navigation buttons to display the rest of the
information.
8
8 On the Approval message, choose the Release
button to accept the result or the Reject button to
reject it.
f On the result report, your user ID is added to the
Approved By line.
9
9 To print the result report, on the Result Report panel,
choose the Print button.
I If the analyzer is set up to print the results
automatically or to send the results automatically
to an information system or a POCT1-A interface,
this step is not required.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Performing a run 55
Result handling
2
2 Select the filter.
Use the and navigation buttons to select a
specific filter.
• All files
• Sample ID
• Assay
• User ID
• Calib
• Unreleased
• Needs Approval
• Export and delete results
I With filter options ending with a colon, you need to
enter a value, for example an ID or name.
Choose the Properties button to display the total
number of result reports, the space used so far
(Flat File), and the remaining space available.
4
4 On the next filter panel, define the time period for
which the results should be included.
I Use the and navigation buttons to select
a specific period.
This week
All dates
Today
Last 2 days
Last month
MMDDYY-MMDDYY (you need to enter the two
dates)
6
6 On the Results panel, select the desired result entry.
I + Positive
- Negative
? Indeterminate
! Invalid
@ Aborted
Approved
Rejected
Sent, but not yet confirmed by host
Sent to host successfully
Double-touch a column header to sort the table
according to this criterion.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
56 Performing a run
Result handling
9
9 On the Result Report panel, if the result information
does not fit into one screen, use the and
navigation buttons to display the rest of the
information.
12
12 On the Approval message, choose the Release
button to accept the result or the Reject button to
reject it.
14
14 To print the result report, on the Result Report panel,
choose the Print button.
I If the analyzer is set up to print the results
automatically or to send the results automatically
to an information system or a POCT1-A interface,
this step is not required.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Performing a run 57
Result handling
15
15 To save the result report or to send it to a host, on the
Results panel, choose the File button. From the File
panel, you can do the following:
• Save the report in a specific location (see step 16).
• Send the report to a host (see step 17).
! CAUTION
Data security
Results backed up by the advanced tools key are stored
as plain text files and so their content is accessible to
unauthorized persons.
r Ensure that you store backed up results safely and
securely so that they are not accessible by
unauthorized persons.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
58 Performing a run
Result handling
2
2 Use the and navigation buttons to select
the Export and Delete Results option.
4
4 On the next filter panel, define the kind of results you
want to delete.
Use the and navigation buttons to select a
specific criterion.
• Export and Delete all results
• Export and Delete results over 7 days old
• Export and Delete results over 30 days old
• Export and Delete results over 60 days old
• Export and Delete results over 90 days old
• Export and Delete results over 180 days old
• Export and Delete results over 365 days old
• Export and Delete results before MMDDYY
You must enter the required date, for example
“103115” to delete all results older than October
31, 2015.
6
6 Choose the export location:
• USB Key
• Share Folder
• Service FTP
• LAN FTP
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Performing a run 59
Result handling
8
8 If you use a USB key, insert the USB key at the rear of
the analyzer. Choose the OK button.
9
9 On the message dialog box, choose the Yes button to
confirm that this is the correct criterion and that you
want to export and delete all results that meet this
criterion.
u Related topics
• Network definitions (81)
• Network resources (81)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
60 Setting up the analyzer
Defining values
In this section
Defining values (60)
Installing and updating assays (62)
Viewing and printing assay tube lot information (77)
Defining system settings (78)
Network definitions (81)
Network resources (81)
Connections to a host system (85)
Managing users (90)
Updating the software (97)
Registering the software (100)
Viewing software and hardware information (102)
Defining values
You define values by entering text or by selecting a
predefined value.
r To enter text
1
1 Use the and navigation buttons to select
the configuration item.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 61
Defining values
3
3 Use the and navigation buttons to select
the specific item that you want to define, and choose
the appropriate button, typically Enter.
f The alphanumeric keyboard or the numeric
keyboard is displayed.
2
2 Use the Left and Right navigation buttons
to select the desired value.
4
4 To save the setting, choose the Enter or OK button,
whichever is available.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
62 Setting up the analyzer
Installing and updating assays
In this section
Installing or updating assays from a USB key (62)
Installing or updating an assay from the Roche remote
service platform (64)
Installing an assay from a local source (66)
Registering assays (68)
Adding an assay tube lot (70)
Deleting assay tube lots (74)
Uninstalling assays (76)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 63
Installing and updating assays
1
1 While on the Main menu, insert the USB key at the
rear of the analyzer.
3
3 Enter your password and choose the OK button.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
64 Setting up the analyzer
Installing and updating assays
4
4 Choose the Yes button to confirm that you want to
install or update the listed assays.
f A progress panel is displayed. All available assays
or updates are installed.
6
6 Do one of the following:
• When a message indicating a successful
installation or update is displayed, choose the OK
button to confirm.
You can check the new assay in the Assay Menu
panel.
You can check the updated assay script version by
choosing Settings > Versions.
• When a message indicating an unsuccessful
update is displayed, choose the OK button to
return to the Main menu.
You should repeat steps 1 to 5.
7
7 If assay revalidation is required, a message is
displayed.
• Choose the Display button to display a list of
assays that need to be revalidated.
• Choose the OK button to return to the Main menu.
u Related topics
• Installing or updating an assay from the Roche remote
service platform (64)
• Adding an assay tube lot (70)
• Registering assays (68)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 65
Installing and updating assays
1
1 When a new assay or an update to an assay is
available, a message is displayed on the analyzer
screen during logon.
3
3 Enter your password and choose the OK button.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
66 Setting up the analyzer
Installing and updating assays
4
4 On the message dialog box, choose the Yes button.
f A progress panel is displayed.
5
5 When the installation is complete, a message is
displayed. Choose the OK button.
u Related topics
• Adding an assay tube lot (70)
• Registering assays (68)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 67
Installing and updating assays
1
1 On the Main menu, select the Assay Menu option
and choose the Select button.
f All currently installed assays are listed.
2
2 Choose the [New Assay] option.
I You may have to scroll down to display this option.
3
3 Choose Scan.
5
5 Define where to download the software from.
• Select USB Key if you install from a USB key.
• If you have defined a different install source
setting, choose the appropriate option.
6 If you use a USB key, ensure the signed ZIP file of the
assay to be installed is in the root folder of USB key.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
68 Setting up the analyzer
Installing and updating assays
7
7 If you use a USB key, insert the USB key at the rear of
the analyzer.
10
10 Decide whether you want to validate (add an assay
tube lot) this assay now or later.
u Related topics
• Adding an assay tube lot (70)
• Registering assays (68)
• System settings (79)
• Defining network resources (82)
Registering assays
New and updated assays must be registered within 30
days.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 69
Installing and updating assays
2
2 Select the assay that needs registering and choose
the Select button.
I Unregistered assays are marked with a number in
the Days left column.
A red icon at the end of an entry indicates that
the registration period has expired.
3
3 On the Report panel, choose the Activate button.
4
4 From the message dialog box, note the registration
code.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
70 Setting up the analyzer
Installing and updating assays
7
7 On the message dialog box, choose the OK button.
8
8 Enter the activation code and choose the OK button.
9
9 On the message dialog box, choose the OK button.
Sharing assay tube lots Instead of adding and validating the same assay tube lot
on several analyzers, you can share assay tube lots
between analyzers in the following ways:
• By synchronizing the assay tube lots using the
advanced tools key.
• By using the share lot function over the share lot
folder.
• By connecting the analyzers to a DMS and enabling
the exchange of lots.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 71
Installing and updating assays
j m Assay installed
1
1 On the Main menu, select the Assay Menu option
and choose the Select button.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
72 Setting up the analyzer
Installing and updating assays
3
3 Obtain the package insert or the instructions for use
from the new lot. Choose the Scan button and scan
the package insert barcode on the first page of the
package insert or the instructions for use of the lot
you want to add.
5
5 Choose the Scan button and place the barcode of the
negative control under the barcode reader.
6
6 Follow the instructions in the assay tube package
insert or the instructions for use on how to add the
negative control sample to an assay tube from this lot.
9
9 Place the prepared assay tube under the barcode
reader.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 73
Installing and updating assays
10
10 Within the time limit for inserting assay tubes, insert
the prepared assay tube in the analyzer.
f Processing starts automatically.
13
13 Wait until the information in the title bar stops
flashing, then choose the Back button.
14
14 Choose the Scan button and place the barcode of the
positive control under the barcode reader.
15
15 Follow the instructions in the assay tube package
insert or the instructions for use on how to add the
positive control sample to an assay tube from this lot.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
74 Setting up the analyzer
Installing and updating assays
17
17 Choose the Scan button.
19
19 Within the time limit for inserting assay tubes, insert
the prepared assay tube in the analyzer.
f Processing starts automatically.
21
21 When asked to do so, remove the assay tube from the
analyzer and dispose of it in accordance with the
disposal policy of your institution.
22
22 Wait until the information in the title bar stops
flashing, then choose the Back button.
23
23 Check the entry for the new assay tube lot.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 75
Installing and updating assays
2
2 Select the assay tube lot entry.
4
4 On the confirmation message dialog box, choose the
OK button.
f The lot entry NA on the Assay Menu panel
indicates that there are no validated lots for this
assay.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
76 Setting up the analyzer
Installing and updating assays
Uninstalling assays
Uninstalling an assay removes the assay script from the
analyzer.
r To uninstall an assay
2
2 Select the assay entry.
I If there are validated assay tube lots for the assay,
there may be several entries for this assay.
3 If there are assay tube lots for the assay you want to
uninstall, delete them all first.
• Select the assay tube lot entry.
• Choose the Delete button.
I When all assay tube lots are deleted, there is only
one entry left for the assay and the value in the
Lot# column is NA.
5
5 On the confirmation message dialog box, choose the
Yes button.
u Related topics
• Deleting assay tube lots (74)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 77
Viewing and printing assay tube lot information
1
1 On the Main menu, select the Assay Menu option
and choose the Select button.
2
2 To print the list of installed assays and assay lots,
choose the Print button.
f The list of installed assays and assay lots is sent to
the printer.
1
1 On the Main menu, select the Assay Menu option
and choose the Select button.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
78 Setting up the analyzer
Defining system settings
2
2 Select the assay tube lot you are interested in and
choose the Select button.
3
3 On the report panel for the control, do either of the
following:
• To print the results, choose the Print button.
• To display the report panel of other controls,
choose the Next button.
(Choose the Previous button to display the
previous report panel again.)
In this section
System settings (79)
About authentication modes (80)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 79
Defining system settings
System settings
To define the general operating environment, enter the
required information.
f Settings > System
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
80 Setting up the analyzer
Defining system settings
u Related topics
• Logging on (29)
• Managing users (90)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 81
Network definitions
Network definitions
Define the network environment.
f Settings > Network
! CAUTION
Breach of confidentiality and privacy
Unprotected network connections may allow access by
unauthorized persons and confidential personal
information may be obtained.
r When connecting the analyzer to a network make sure
the connection is secure and monitored for security
breaches.
r If parts of your network, which the analyzer uses to
exchange data, are connected by WLAN, make sure to
use WPA2 encryption.
Network resources
You can define network resources for installing and
updating assays, and for sharing assay tube lots. These
resources can also be used for backing up results.
In this section
Defining network resources (82)
Using the share lot function (83)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
82 Setting up the analyzer
Network resources
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 83
Network resources
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
84 Setting up the analyzer
Network resources
1
1 Choose Settings > Install Source > Share Lot
Folder.
• Choose the Server Name, Share Name, User
Name, and Password items and enter the
required information.
• Choose the Use Share Lot item and choose the
Yes option.
1
1 Choose Settings > Install Source > Share Lot
Folder. Choose the Use Share Lot item and choose
the No option.
3
3 Choose the Assay Menu option. Review the installed
assays and assay tube lots.
I You must revalidate assay tube lots that were
validated previously while the share lot function
was enabled.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 85
Connections to a host system
In this section
Defining settings for host systems (85)
About data exchange with a DMS (86)
Connecting the analyzer to a DMS (88)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
86 Setting up the analyzer
Connections to a host system
Data Synchronization o Auto Send Choose which data is exchanged with a connected DMS.
o Users
o Assay Lots
o Device Configuration q Enabled data synchronization affects analyzer
o Log Events functions, e.g. the local deletion of assay tube lots or
local user management.
u About data exchange with a DMS (86)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 87
Connections to a host system
About automatic locking Some data received from the DMS (e.g. user data and
assay tube lots) can only be applied while the analyzer is
locked or no user is logged on. While such data is
received and applied, you cannot unlock the analyzer.
u About locking the analyzer (39)
About lot management You can share assay tube lots between analyzers, if the
exchange of lots is enabled.
About user management If the analyzer exchanges user data with a DMS, user
management is disabled on the analyzer.
About changing your own password If the analyzer exchanges user data with a DMS, users
can only change their passwords based on a request from
the DMS.
About changing your own badge barcode If the analyzer exchanges user data with a DMS, users
cannot assign or change their badge barcodes directly on
the analyzer.
u Assigning or changing your badge barcode (34)
About analyzer settings If the analyzer accepts system settings from the DMS, the
received settings override local settings. Local changes of
the settings are only effective as long they are not
overwritten from the DMS.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
88 Setting up the analyzer
Connections to a host system
About event logs If configured, the analyzer sends events to the connected
DMS.
u Viewing and printing the event log (124)
1
1 Choose Settings > Connectivity. Choose the Type
item and choose the POCT1-A (DML) option.
2
2 Choose the Server Details item:
• Enter Server, Port, and Timeout settings.
• Set the TLS encryption item to the Yes option.
• Configure secure communication.
I For details, refer to the cobas® Liat® Analyzer
Host Interface Manual POCT1-A (DML) or contact
your Roche representative.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 89
Connections to a host system
4
4 Choose the Data Synchronization item. If you want
the analyzer to exchange data with the DMS, choose
the corresponding item and set it to the Yes option.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
90 Setting up the analyzer
Managing users
Managing users
You can add new users, change user information, and
delete complete user accounts.
In this section
Overview of user account access roles (90)
About user IDs and passwords (91)
Defining new users (91)
Changing user information (94)
Deleting a user account (95)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 91
Managing users
About user management on a DMS If the analyzer exchanges user data with a DMS, the
following rules apply additionally:
• Make sure the rules for user IDs and passwords on
the analyzer and the DMS match.
• User IDs and passwords defined on the DMS must
only consist of ASCII Printable Characters (32-126)
without #96 (‘ character).
u About data exchange with a DMS (86)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
92 Setting up the analyzer
Managing users
1
1 To display the User panel, do one of the following:
• On the Main menu, choose the User button.
• Choose Settings > User.
f The User panel lists all defined users. The icon
preceding the user ID indicates the user access
role.
2
2 Choose the [New User] option.
4
4 To define a user detail, select the item and choose the
Enter button. Then do either of the following:
• Enter the information and choose the OK button.
• Select the option and choose the Select button.
I To avoid cut-off user IDs in the title bar, it is
recommended to define short user IDs.
5
5 If the analyzer is configured for a badge barcode
authentication mode, assign and scan the badge
barcode:
• In the User Info panel, choose the Badge
Barcode item. Choose the Enter button.
• Select the Assigned option.
• Scan the badge barcode twice.
I New users cannot assign their badge barcodes at
logon. This is only possible when a user is logged
on.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 93
Managing users
6
6 To define which assays the user can perform, do the
following:
• Select the Allowed Assay: option and choose the
Enter button.
• On the Allowed Assay panel, select the assay you
want the user to be able to perform. Choose the
All button if you want to select all listed assays.
• Choose the Change button. The values of the
selected assays change from blank (not allowed)
to Allowed or vice versa.
• To keep the changes, choose the Done button.
7
7 On the User Info panel, choose the Add button to
save the definitions.
I By default, new users must change their
passwords on first logon.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
94 Setting up the analyzer
Managing users
About locking user accounts To prevent a user from logging on, lock the user account.
To allow access to the analyzer again, unlock the user
account.
1
1 To display the User panel, do one of the following:
• On the Main menu, choose the User button.
• Choose Settings > User.
2
2 Select a user entry.
I If you are logged on with Administrator or
Supervisor access role, all currently defined users
of the same access role or lower are listed.
If you are logged on with User access role, your
own user account is listed only.
The icon preceding the user ID indicates the user
access role.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 95
Managing users
4
4 To lock a user, select the Locked item. Select the Yes
option and choose the OK button.
I Locked users cannot log on to the analyzer.
To unlock a user, change the Locked item to the
No option.
5
5 To change a user detail, select the item and choose
the Change button. Then do either of the following:
• Enter the information and choose the OK button.
• Select the option and choose the Select button.
I If you set the Change PW item to the On next
login option, the user must change the password
when logging on the next time.
u Related topics
• Defining new users (91)
• Changing your password (31)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
96 Setting up the analyzer
Managing users
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 97
Updating the software
About migration of user information During an upgrade of the analyzer software from version
3.0 to version 3.1, data is migrated. For user information,
the following migration rules apply:
• User IDs are no longer case-sensitive. Therefore, if 2
user IDs differ only in capitalization (e.g., user1 and
USER1), the newer user account will be deleted. The
older user account will be locked.
• If a user ID is longer than 20 characters, the user ID
will be truncated and the user account locked.
• If a user ID contains spaces, the spaced are removed
and the user account is locked.
• The default Admin account is migrated properly and is
not locked.
1
1 CAUTION! Do not disconnect the mains power supply
to the analyzer while updating software.
1 When a software update is available, a message is
displayed on the analyzer screen.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
98 Setting up the analyzer
Updating the software
3
3 Enter your password and then choose the OK button.
f The analyzer checks the available storage space
on the analyzer. If there is not enough storage
space, a message is displayed and you must make
storage space available by deleting either lots or
results.
4
4 On the Upgrade panel, when asked to do so, choose
the Reboot button.
I After successful restart, the analyzer attempts to
register the software automatically.
1
1 CAUTION! Do not disconnect the mains power supply
to the analyzer while updating software.
1 While on the Main menu, insert the USB key at the
rear of the analyzer.
2
2 On the message dialog box, choose the OK button.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 99
Updating the software
3
3 Enter your password and choose the OK button.
f The analyzer checks the available storage space
on the analyzer. If there is not enough storage
space, a message is displayed and you must make
storage space available by deleting either lots or
results.
4
4 On the message dialog box, choose the Yes button.
6
6 Do one of the following:
• When a message indicating a successful upgrade
is displayed, choose the Reboot button and wait
until the analyzer has restarted.
• When a message indicating an unsuccessful
upgrade is displayed, choose the OK button to
return to the Main menu.
Reboot the analyzer and repeat steps 1 to 6.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
100 Setting up the analyzer
Registering the software
u Related topics
• Registering the software (100)
• Exporting and deleting results (57)
• Deleting assay tube lots (74)
1
1 Start the registration process.
• If a message is displayed, choose the Activate
button.
• If no message is displayed, on the Main menu,
choose Settings > System Info.
2
2 From the System Info panel, note the registration
code.
6
6 Enter the activation code and choose the OK button.
7
7 On the message dialog box, choose the OK button.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
102 Setting up the analyzer
Viewing software and hardware information
In this section
Viewing information about the analyzer and
software (102)
Viewing system information (103)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Setting up the analyzer 103
Viewing software and hardware information
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
104 Quality control
Quality control
Internal control Internal Process Control (IPC) is included in every assay
tube. It verifies the adequate processing of the biological
sample. The IPC passes, if sample purification and target
amplification meet validated acceptance criteria.
Quality control kits Quality control kits for each cobas® Liat® Assay are
available from Roche. The quality control kits contain
positive and negative control sample materials. The
process of running positive and negative controls is
similar to that for adding lots.
u See Adding an assay tube lot (70) and refer to the
cobas® Liat® Assay package insert or the
instructions for use for detailed instructions.
NOTICE
Additional QC testing
Additional QC testing is not required by the manufacturer.
Quality control should be run as a patient sample if
additional quality control testing is performed at a site.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Maintenance and calibration 105
About self-check
In this section
About self-check (105)
About auto monitoring (105)
About auto adjustment (106)
About service (106)
Calibrating the touch screen (107)
Cleaning the analyzer (107)
About self-check
The analyzer performs self-diagnostics during startup
(initialization) and utilizes an advanced error diagnostics
system to monitor the analyzer’s performance during an
assay. Under normal operation, the analyzer alerts the
operator in the event that a malfunction or error is
detected.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
106 Maintenance and calibration
About auto adjustment
About service
Please contact your local Roche representative if you
have questions regarding the analyzer, its service needs,
or if you have other questions.
! CAUTION
Electrical hazards
Only a qualified Roche representative can service the
analyzer.
Disassembling the analyzer could result in electrical
hazards.
r Never attempt to repair or adjust the analyzer yourself.
r Contact your local Roche representative if the analyzer
fails to operate properly.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Maintenance and calibration 107
Calibrating the touch screen
2
2 Choose the Select button.
Safety u Make sure you are familiar with the safety instructions
in Safety instructions (14).
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
108 Maintenance and calibration
Cleaning the analyzer
! CAUTION
Electric shock
Spraying or applying liquid to the analyzer or the power
supply may damage the analyzer or pose an electrical
hazard.
r Do not spray or apply liquid directly on the analyzer or
the power supply.
r Do not attempt to clean the interior of analyzer
through the vents.
r To clean the tube chamber, only use the cobas®
Liat® Cleaning Tool when prompted by the analyzer.
NOTICE
Damage to the analyzer due to use of unsuitable
cleaning materials
Using unsuitable cleaning materials can damage the
touch screen and other surfaces.
r Only use the supplied cleaning tool when prompted
by the message. Do not attempt to clean the tube
chamber of the analyzer with anything other than the
cleaning tool kit.
r Do not use harsh, abrasive cleaners or wipes.
In this section
Cleaning the outside of the analyzer (109)
Cleaning spillages or leakages from an assay tube (110)
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Maintenance and calibration 109
Cleaning the analyzer
1
1 Wipe the touch screen gently with a soft lint-free
cloth.
1
1 Moisten a lint-free cloth with either 70% isopropanol
or 5-10% bleach solution.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
110 Maintenance and calibration
Cleaning the analyzer
! CAUTION
Contamination of cleaning tool
Do not use the cleaning tool for biohazardous
contamination. The cleaning tool does not decontaminate
the analyzer, but becomes contaminated itself.
r In case of spillages or leakages from an assay tube
follow the procedure below.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Troubleshooting 111
List of error messages
Troubleshooting
The analyzer monitors its operation and logs abnormal
events. Based on the severity, the analyzer tries to recover
or to fix the error while running. If the situation cannot be
resolved, the analyzer stops.
In this section
List of error messages (111)
Generating a backup for troubleshooting purposes (124)
Viewing and printing the event log (124)
Resetting the lost default administrator password (125)
Cleaning the window of the barcode reader (127)
Resolving a tube entry door error before the start of a
run (128)
Resending results manually (128)
Restarting an unresponsive analyzer (129)
Software error codes and error messages Software error codes for the cobas® Liat® Analyzer
are displayed in hexadecimal system, in the range from
“0xb01” to “0xbff”, or numbered as “Error 1” to “Error 10”.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
112 Troubleshooting
List of error messages
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Troubleshooting 113
List of error messages
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
114 Troubleshooting
List of error messages
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Troubleshooting 115
List of error messages
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
116 Troubleshooting
List of error messages
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Troubleshooting 117
List of error messages
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
118 Troubleshooting
List of error messages
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Troubleshooting 119
List of error messages
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
120 Troubleshooting
List of error messages
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Troubleshooting 121
List of error messages
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
122 Troubleshooting
List of error messages
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Troubleshooting 123
List of error messages
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
124 Troubleshooting
Generating a backup for troubleshooting purposes
Assay failure codes and assay pattern codes Assay failure codes are generated for invalid test
results.
! CAUTION
Data security
Data backed up by the advanced tools key is stored as
plain text files and so their content is accessible to
unauthorized persons.
r Ensure that you store backed up data safely and
securely so that it is not accessible by unauthorized
persons.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Troubleshooting 125
Resetting the lost default administrator password
2
2 To print the event log, choose the Print button.
1
1 On the login panel, choose the Password? button.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
126 Troubleshooting
Resetting the lost default administrator password
3
3 Make a note of the key displayed in the message.
I Do not touch the screen until you have received
the reset code.
7
7 On the Enter Reset Code panel, enter the code
provided by your Roche representative.
9
9 Enter a new administrator password.
10
10 Choose the OK button.
f A success message is displayed.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Troubleshooting 127
Cleaning the window of the barcode reader
! CAUTION
Electric shock
r Do not spray or apply liquid directly on the analyzer.
NOTICE
Damage to the analyzer due to use of unsuitable
cleaning materials
Using unsuitable cleaning materials can damage the
window of the barcode scanner.
r Do not use harsh, abrasive cleaners or wipes.
3
3 Gently wipe the outside of the barcode reader
window.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
128 Troubleshooting
Resolving a tube entry door error before the start of a run
2
2 On the message dialog box, choose the Manual
button. Choose the OK button.
3
3 To run the assay, continue with step 9 of (u 46).
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Troubleshooting 129
Restarting an unresponsive analyzer
2
2 On the Results panel, review and send the results as
described in (u 53).
j m Analyzer unresponsive
1
1 If the analyzer is completely unresponsive, press the
power button and hold it down for 5–10 seconds until
the analyzer powers off.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
130 Technical data
Technical characteristics
Technical data
To ensure trouble-free operation, ensure that you use the
components listed below and that the conditions
mentioned in this section are met.
In this section
Technical characteristics (130)
Environmental conditions (130)
Storage and transport conditions (131)
Standard supplies (131)
Printer (132)
Technical characteristics
The analyzer has the following technical characteristics:
Analyzer
User interface Built-in touch screen & keypad
Internal storage capacity Approximately 20,000 test results with date and time can be
stored on the analyzer (depending on result file size)
Barcode reader Class 2 Laser Product
For barcodes, the following standards are supported: Code
39, 93, 128, Codabar, GS1 Databar-14.
Connectivity port Ethernet, RJ-45, TCP/IP
2 Universal Serial Bus (USB)
(maximum load of 250 mA)
Main connection Power supply adapter: Input 100-240 V AC / 50-60 Hz
Output: 15 V DC / 8.6 A
Power consumption 130 W on AC mains power
Safety class II
Dimensions (L x W x H) 24.1 cm x 11.4 cm x 19.0 cm (9.5 in x 4.5 in x 7.5 in)
Weight 3.76 kg (8.3 lbs)
y Technical data
Environmental conditions
For operation, the following environmental conditions
must be met.
Analyzer
Temperature range +15 °C to +32 °C (59 °F to 90 °F) recommended
Relative humidity 15% to 80% (non-condensing)
Maximum altitude 2000 m (6500 ft) above sea level
y Environmental conditions
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Technical data 131
Storage and transport conditions
Analyzer
Temperature range -20 °C to +60 °C (-4 °F to +140 °F)
Storage relative humidity 10% to 95% (non-condensing)
y Storage and transport conditions
Standard supplies
For trouble-free operation, use the following standard
supplies:
Name
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
132 Technical data
Printer
Printer
A standard PCL compatible printer (laser or inkjet) that
can be connected either to the USB port or via data
network.
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Index 133
Index
A
Abbreviations, 10
Aborting assays, 49
Access roles, 90
Adding
– assay tube lots, 70
Addresses, 5
Adjustment
– automatic, 106
Analyzer
– cleaning outside, 109
– functions, 22
– operating elements, 24
– setting up, 60
– starting, 28
– switching off, 44
– unresponsive, 129
Approvals, 4
Assay tube lots
– adding, 70
– deleting, 75
– sharing, 70, 83
Assay tubes, 22
Assays
– aborting, 49
– adding tube lots, 70
– deleting tube lots, 75
– installing from local source, 66
– installing from remote service platform, 64
– installing from USB key, 62
– overview, 25
– performing, 46
– registering, 68
– uninstalling, 76
– updating from remote service platform, 64
– updating from USB key, 62
Audit trail, 23
Auto adjustment, 106
Auto monitoring, 105
Auto-lock function, 39
Automatic adjustment, 106
Automatic restart, 105
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
134 Index
Buttons
– function, 25
– navigation, 25
Calibration
– touch screen, 107
Changing user information, 94
Characteristics, technical, 130
Checks, automatic, 23
Cleaning
– ouside of analyzer, 109
– spillages, 110
– touch screen, 109
Code
– assay failure, 124
– assay pattern, 124
– firmware error, 121
– hardware error, 121
– software error, 111
Conditions
– environmental, 130
– for operation, 130
– storage, 131
– transport, 131
Contact, 5
Controls, 77
Conventions used in this publication
– abbreviations, 10
– product names, 9
– symbols, 9
Copyright, 3
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Index 135
Documents, supporting, 2
Edition notice, 2
Entering text and numbers, 37
Environmental conditions, 130
Error
– assay, 124
– firmware, 121
– software, 111
– tube entry door, 128
Event log, 124
Feedback, 4
Flap, for assay tubes, 25
Function buttons, 25
Functions, analyzer, 22
Graphs, 52
HIS connections, 85
Hospital information system connections, 85
Labels, on analyzer, 16
Laboratory information system connections, 85
LIS connections, 85
List of messages, 111
Locking the analyzer manually, 41
Locking users, 94
Log files, 124
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
136 Index
Logging on, 29
– with badge barcode, 30
– with user ID, 30
Lots
– control results, 77
– information, 77
Maintenance, 105
Messages, 111
Monitoring, 105
Navigation buttons, 25
Network definitions, 81
Network resources, 81
New in version 7.2, 12
Numbers, entering, 37
Operating elements, 24
Overview
– assays, 25
– barcode reader, 25
– results, 52
– touch screen, 26
Password
– changing, 31
– changing anytime, 32
– changing at logon, 33
– resetting for Administrator, 125
Passwords
– about, 91
Patents, 4
Performing
– assays, 46
– runs, 45
Power switch, 25
Powering on, 28
Printer
– compatibility, 132
– installing network, 21
– installing USB, 20
Printing
– lot control results, 77
– lot information, 77
Product names, 9
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Index 137
Reading barcodes, 37
Registering
– assays, 68
– software, 100
Remote service platform, 64
Reports, 52
Resetting Administrator Password, 125
Results
– deleting, 57
– lot controls, 77
– overview, 52
– reviewing, 53
Reviewing results, 53
Revision history, 2
Running assays, 46
Runs, performing, 45
Safety
– classification, 14
– disposal, 16
– labels, 16
– summary, 14
Scanning barcodes, 37
Screen
– See Touch screen
Screenshot disclaimer, 3
Selecting items, on screen, 36
Self-checks, 23, 105
Service, 106
Settings
– See Definitions
Setup
– analyzer, 60
– HIS, LIS, DMS connections, 85
– network, 81
– network resources, 82
– system, 79
Share lot function
– disabling, 84
– using, 83
Shutting down analyzer, 44
Software
– installed versions, 102
– registering, 100
– updating, 97
Specifications, 130
Spillages, 110
Starting
– analyzer, 28
Storage conditions, 131
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
138 Index
Supplies, 131
Support, 4
Supporting documents, 2
Switching off analyzer, 44
Symbols used in this publication, 9
System elements
– versions and IDs, 103
System settings, 79
Uninstalling assays, 76
Unlocking the analyzer, 42
Unlocking users, 94
Updating assays
– from remote service platform, 64
– from USB key, 62
Updating software, 97
USB key
– for assay installation, 62
– for assay update, 62
User IDs
– about, 91
User rights, 90
Users
– changing information, 94
– defining, 91
– deleting, 95
– locking, 94
– unlocking, 94
Versions
– software, 102
– system elements, 103
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
Index 139
Warranty, 3
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001
140 Index
Roche Diagnostics
cobas® Liat® System · Software version 3.1 · Operator's Manual · Version 7.2 P/N: 08276340001