Nutech 3000T Portable THC Analyzer
The Latest Generation of
Portable VOCs Analyzer
for Stationary Emission Source & Fence Line
Technical Data
• Detection Method: FID
• Measuring Range: 0~10/100/1000/10000 ppm (Automatic)
• Detection Limit: ≤0.07 mg/m3 (Calculated by Carbon)
• Zero Drift: ≤±1% Full Scale/8H
• Span Drift: ≤±1% Full Scale/8H
• Repeatability: ≤±0.1% Full Scale
• Indication Error: ≤±1% Full Scale
• Warm-up Time ≤30min
• Data Resolution (T90): ≤2s
• Working Environment: temperature: -20 ℃ ~50 ℃ ; relative humidity: <95%; atmospheric
pressure: (80~106) kPa
• Power Supply: DC24V, AC (100-240±10%) V/(50±2%) Hz
• Dimensions: 370x260x397mm
• Weight: <12KG
Standard Configuration
Instrument main unit (including metal hydride cylinder), sampling line (including probe, filter
device), Microsoft Surface GO tablet, backpack.
Features
Lightweight & Portable
• Highly integrated design to ensure the total weight within 12KG (including metal hydride
cylinder, battery, zero air generators, etc)
• Equipped with a backpack and a handle, which can be easily transported and put into use by
one person.
Easy to Use
• Large-capacity battery (45Ah) power supply, which can run for a long time even without
utility power on site.
• Operated by a detachable touch screen tablet with Wi-Fi communication and a user-friendly
interface. Users can control the instrument and read test results within 50m (33ft) of the
site through the detachable tablet.
Most Accurate Test Result
• From the sampling line to the FID detector, the whole process is high temperature heating
to avoid condensation & sample loss.
• The sampling flow path is silanized (a special surface treatment technology to make the
surface utmost inert), resistant to high temperature and corrosion, and contains a filter
device to effectively filter particles to avoid sample contamination and adsorption.
• High precision EPC and FID are adopted to ensure accurate flow control and accurate
measurement.
• Support automatic & manual purge function to keep the system clean and thus deliver more
accurate test results.
Powerful & Easy Data Processing
• Data file automatic recording and storage, historical data query, reprocessing and printing
functions.
• Test data can be exported in Excel file
• Can equipped with a mini Bluetooth printer to print the data on-site.
Lowest Running Cost in Its Kind
• H2 is supplied by mini built-in metal hydride cylinder with very low pressure and thus is safe
and easy to handle and transport. The metal hydride cylinder is recyclable.
• Zero air is supplied by internal pump. No need disposable zero air bottle.
• Also support external gas cylinder
Wide Applications
• Applicable in VOCs measurement in stack (high concentration), inside the factory (medium
concentration), outside the factory/fence line (low concentration).