LCMS-2020 Preparative System: Installation Guide
LCMS-2020 Preparative System: Installation Guide
LCMS-2020 Preparative System: Installation Guide
Jun. 2014
Installation Guide
225-18427A
Jun. 2014
CONTENTS
1. Introduction
1.1 Fractionation by LCMS
1.2 Splitter and Delay-time
1.3 Outline of Analog Output
2. System Configuration
2.1 Example Layout
2.2 Connecting Signal Cables
2.3 Flow Line Configuration
2.4 How to Connect the Main Flow Lines
3. How to Check and Adjust the Delay Time
4. Precautions for Using the Preparative System
5. Consumable Parts
6. Service Parts
7. Appendix
8.1 Equipment size and electric requirements
8.2 Footprint
2
225-18427A
Jun. 2014
1. Introduction
This manual is guide for installing the LCMS-2020 Preparative system. Use this guide in
conjunction with the LCMS-2020 instruction manual. For information regarding LC units,
refer to the instruction manuals for respective units. For information regarding how to
operate the software, refer to the LCMS Preparative System Operation Manual, which is
provided with this preparative system.
From Column
Splitter T1
LCMS
Fraction Collector
T2 T2>T1
Adjustment Line of Delay time
3
225-18427A
Jun. 2014
When flow lines are split to collect fractions, sample substances diverted by the
splitter must reach the fraction collector after they reach the MS unit. This difference
in arrival times (T2 minus T1) is referred to as “Delay time”, where T2 is the time
required for the sample to flow to the fraction collector and T1 is the time required for
the sample to flow to the MS unit. To become T2>T1, the LCMS-2020 preparative
system provides the adjustment line (Tube for the Delay time) that allows adjusting
the Delay time. (See Figure 1.1.)
We use an active splitter (APV) that has high speed switching valve to split the flow.
The APV can be set to several different split ratio configurations, depending on the
flow rate. (For more details, refer to the APV instruction manual.) Using an APV
(valve) requires a make-up pump (such as the LC-20AD). By using the APV, the
LCMS preparative system has two independent flow lines – the MS line and the
fraction collector line, where the APV diverts part of the sample from the fraction
collector line to the MS line. (See Figure 1.2.)
MS Line LCMS
Make-up Pump
FC Line
APV
The MS detector receives the digital input signal, performs a D/A conversion and
outputs and analog signal. With default settings, one MS count is output as 1 micro V,
4
225-18427A
Jun. 2014
with a maximum count of 1,000,000 (= 1 v). This maximum value can be changed by
editing the method parameters. For example, if the maximum value is set to
10,000,000, then 1 μV will be output for a count of 10. (The maximum analog value
remains 1 volt.) Consequently, even if chromatogram values for an MS detector are
exceptionally large, this maximum value setting can be changed in order to output
appropriate analog values.
The baseline for the analog output is automatically determined from the first three
signal points generated at the beginning of the analysis and for each change in scan
group. Therefore, even if the baseline increases after repeating an analysis or the
baseline varies due to changes in analytical parameters for different scan groups, the
auto-zero function allows ignoring these baseline variations.
Analog output settings are entered via the LCMSsolution software. Refer to the
LCMS Preparative System Operation Manual provided with the system regarding
how to enter these settings.
5
225-18427A
Jun. 2014
2. System Configuration
2.1 Example Layout
Figure 2.1 shows one example of units included in a system and their layout.
Different models can be selected for the pump, autoinjector and fraction collector
units depending on system objectives.
SPD-20A(prep cell)
LC-8A APV
LC-20AD PC Printer
SPD-M20A
(semimicro cell)
6
225-18427A
Jun. 2014
LC-8A PC
LAN
LC-8A CBM-20A
Optical link 1
APV
OUT 2 IN A (B) (A)
Gilson819
LC-8A PC
LAN
CBM-20A 2Ch-RS232C board
LC-8A 1
Optical link
M.inj 2 COM3 COM4
Event
LC-20AD
Event USB
USB
LCMS-2020
Cable of IFC-PAL
ANALOG IN ANALOG OUT
RS232C
SPD-20A (Trigger)
SPD-M20A
A1 A0 APV
INTERFACE 1
NI-6009
IFC PAL USB
P00&G
RS232C
B) When IFC-PAL is used
Figure 2.2 Signal cables Connection Diagram 7
225-18427A
Jun. 2014
8
225-18427A
Jun. 2014
CBM-20A
0.2ml/min
Semimicro cell
LC-8A
SPD-M20A LCMS-2020
Prep-mixer ESI/DUIS
Bypass ASSY
20ml/min
Injector
Prep-column APV bypass kit
APV
LC-8A
Delay time
Gilson215 or
SPD-20A
IFC-PAL
Drain Prep-cell Tube for the Delay time LC-20A
9
225-18427A
Jun. 2014
A Tube
LC-8A Column holder
CBM-20A
LCMS-2020
LC-8A Gilson215 or IFC-PAL
SPD-20A APV
LC-8A
LC-20AD SPD-M20A
CBM-20A
LCMS-2020
LC-8A Gilson215 or IFC-PAL
SPD-20A APV
LC-8A
LC-20AD SPD-M20A
10
225-18427A
Jun. 2014
Injection port
Sampling loop
Drain port
Tygon tubing
To drain
Drain tubing
From
ミキサよりmixer
To column
カラムへ
Bypass ASSY(228-20355-92)
Injection port
Sampling loop
Drain tubing
To column
カラムへ
From
ミキサよりmixer
Bypass ASSY(228-20355-92)
in the injector.
CBM-20A
LCMS-2020
LC-8A Gilson215 or IFC-PAL
SPD-20A
LC-8A
LC-20AD SPD-M20A
APV
B Tube
CBM-20A
LCMS-2020
LC-8A Gilson215 or IFC-PAL
SPD-20A APV
LC-8A
LC-20AD SPD-M20A
Prep-cell
12
225-18427A
Jun. 2014
About 20mm
About 55mm
Clamp、CKN20
Cover
13
225-18427A
Jun. 2014
Delay Time (min)= RT2 – RT1 + (Delay Time between SPD Detector and Fraction Collector)
= RT2 – RT1 + (Volume of piping between SPD Detector and Fraction Collector / flow Rate)
14
225-18427A
Jun. 2014
changes.
5. Consumable Parts
Part Name Part Number Remarks
APV Maintenance Kit 225-10799-02 Includes 3 rotors and 3
stators
Note: For information regarding other consumables, see the instruction manuals for
respective units.
6. Service Parts
For information regarding Service parts, see the instruction manuals for respective units.
7. Appendix
7.1 Equipment size and electric requirements (For IFC-PAL)
400mm
Rotary pump ゚
900mm
PC
IFC-PAL Printer
LC
(LC-8A) LC LC
LCMS-2020
APV
4000mm 15