XRF Lecture Notes
XRF Lecture Notes
Becoming the most widely used method for elemental analysis of solids
* Simple spectra
* Spectral positions are almost independent of the chemical state of the analyte
* It is non-destructive
* X-ray penetration of the sample is limited to the top 0.01 - 0.1 mm layer
* Light elements (below Ti) have very limited sensitivity although C is possible on new
22
instruments
* Inter element (MATRIX) effects may be substantial and require computer correction
NOMENCLATURE
Simplified spectral lines observed in x-ray spectra (each energy shell actually comprises
several energy levels, thus transitions are more numerous than shown).
PRINCIPLES OF X-RAY FLUORESCENCE
X-Ray Excitation
Electron Excitation
AUGER ELECTRON EMISSION
(internal photoionisation)
The Auger effect is more common in elements of low Z because their atomic electrons are
more loosely bound and their characteristic X-rays more readily absorbed.
X-RAY FLUORESCENCE YIELD
The yield of X-ray photons is reduced by the Auger effect. The fluorescence yield ( ) is the
ratio of X-ray photons emitted from a given shell to the number of vacancies created in that
shell.
Since production of Auger electrons is the only other competing reaction the ratio of Auger
The probability of X-ray absorption as a function of path length through the sample is given by
Beer's law for X-rays:
I/I0 = exp(-µm d)
where I/I0 is the fraction of X-rays transmitted through a thickness d of a material of density
. The parameter µm is the mass absorption coefficient which is a function of the atomic
number Z and the energy (wavelength) of the X-ray.
The mass absorption coefficient for a complex sample is the weighted average of the
coefficients for the constituent atoms.
X-RAY MASS ABSORPTION COEFFICIENTS
A plot of mass absorption coefficient vs energy of the X-ray photon for Pb.
82
Abrupt changes are observed corresponding to absorption edges for K, L and M electrons. At
the energy (wavelength) of the edge, the photons first become sufficiently energetic to eject
K, L and M photoelectrons.
X-RAY SOURCES
The X-ray tube is energised by a high-voltage power supply with an output of 0.5 to 50 kV.
The head of the vacuum tube consists of a target (anode), which is often made of tungsten
and chromium. As accelerated electrons strike the target, X-rays are emitted. The tube also
forms the X-rays into a beam through a Beryllium window.
In the target the 74W is used to excite L and K lines of higher Z atoms and the Cr is
24
used to excite atoms of 22Ti and below.
Spectral Output
X-ray spectrum produced by electron bombardment of a tungsten target:
Continuum Spectrum: The continuum results from deceleration of electrons by the atoms in
the target. The German term "bremsstrahlung" means "braking radiation."
X-ray photons (as well as other energetic particles) can be measured using the following types
of detectors:
* gas-filled detectors register a current pulse from the collection of electron-ion pairs formed;
* a semiconductor detector register a current pulse from the formation of electron-hole pairs;
* a scintillation detector counts light pulses created when an X-rays passes through a
phosphor;
* a photographic plate.
Gas-Proportional Counters
A gas-proportional counter is filled with P90 gas (90% argon, 10% methane). X-rays ionise
the gas, leaving electrons that migrate to the anode and positive ions that move to the case.
Proportional counters use gas amplification: the detector voltage is raised to 500 to 700 V so
that the primary electrons and ions, first formed, are accelerated to produce secondary
electrons and ions when they collide with gas atoms. This yields a greatly increased signal
which is, nevertheless, proportional to the energy of the original x-ray.
Solid-State Detectors
When an electron enters the crystal it ejects a high-energy photoelectron which ultimately
dissipates its energy in multiple interactions which promote valence band electrons to the
conduction band, leaving holes in the valence band. The electron-hole pairs are then collected
by biasing the detector at -1000 V, giving rise to a current pulse for each x-ray entering the
detector. Charge collection is much more efficient than in a gas.
Instrumentation
In wavelength dispersive spectrometers, the several x-ray lines emitted from the sample are
dispersed spatially by crystal diffraction on the basis of wavelength. The detector then receives
only one wavelength at a time.
The crystal and detector are made to synchronously rotate through angles of θ and 2θ
respectively.
Bragg Diffraction
Wavelength dispersive X-ray spectrometers function by separating the X-rays of interest using
diffraction from a crystal. This follows from the Bragg equation:
n = 2 d sin( )
where n is the diffraction order, d is the interplanar spacing of the atomic layer and is the
angle of incidence.
Crystals
Si 0.055 - 0.598 nm
lead stearates 6 - 15 nm
* a range of alternative excitation sources can be used in place of high-power x-ray tubes with
their large, heavy, expensive and power-consuming supplies
* alternative sources include, low power x-ray tubes, secondary monochromatic radiators,
radioisotopes and ion beams.
Advantages of Wavelength Dispersion:
* higher individual intensities can be measured because only a small portion of the spectrum
is admitted to the detector
* with multichannel analysers sensitivity for weak lines in the presence of strong lines is
limited because the strongest line determines the counting time
SAMPLE PREPARATION
Reproducible sample preparation methods are essential. Samples must be in a form that are
similar to available standards in terms of matrix, density and particle size.
* Solids, generally solids must be polished as surface roughness may give erratic results.
* Powders and pellets, powdered samples are often pressed into pellets, suspensions may also
be analysed
* Liquids and solutions, a x-ray transparent cover and sample cup must be provided to
prevent volatility under vacuum conditions
MATRIX EFFECTS
2. The matrix absorbs the secondary analyte x-rays (secondary-absorption effect); it may
have a larger or smaller absorption coefficient for the analyte-line radiation
3. The matrix elements emit their own characteristic lines, which may lie on the short
wavelength side of the analyte absorption edge, thereby exciting the analyte to emit additional
radiation to that excited by the primary source of X-rays alone (enhancement)
Absorption-Enhancement Effects
Absorption-enhancement effects can be positive or negative on the basis of their effect upon
analyte intensity.
In the positive absorption effect, the matrix has a smaller absorption coefficient than the
analyte for the primary and analyte-line radiation, and the analyte-line radiation is higher than
would be predicted.
In the negative absorption effect, the matrix has a larger absorption coefficient than the
analyte, and the analyte-line intensity is lower than expected.
In the enhancement effect, one or more spectral lines of the matrix elements excite analyte-
line emission). This enhancement may take two forms: direct enhancement ( B and C
both excite A ) and the third element enhancement ( C excites B which in turn
excites A ).
QUANTITATIVE ANALYSIS
3. Matrix-Dilution Methods. The matrix of all samples is diluted to a composition such that the
effect of the matrix is determined by the diluent rather than the matrix.
4. Thin-Film Methods. The samples are made so thin that absorption-enhancement effects
substantially disappear.
5. Standard Addition and Dilution Methods. The analyte concentration is altered quantitatively
in the sample itself. The sample is subjected to one or more quantitative incremental
concentrations or dilutions of the analyte. The intensity of the analyte lines is measured for
effectively the same matrix in each case.