Autolumo Operation Manual
Autolumo Operation Manual
Autolumo Operation Manual
VERSION : V1.1
AUTOBIO tries to ensure that all contents of this manual in its print publication are accurate.
However, to meet the requirements of the development, AUTOBIO still reserves the rights to
modify the manual without any notice to customers.
E-mail : [email protected]
Tel : [86]-400-056-9995
Website : http://www.autobio.com.cn
EC Representative
OBELIS S.A
Bd.Général Wahis, 53
1030 Brussels,
Belgium
CHAPTER 1 INTRODUCTION
If you want any technical support, please call our service hotline, [86]-400-056-9995,
or contact our service support representatives. For Any technical support, the series
number of instrument is necessary to provide.
1.5 Manufacturer
The safety symbols on the instrument remind you of caution. Read contents of the
symbols and instructions of this manual carefully when running the instrument. If
any questions, please contact our technical service department of AUTOBIO.
AutoLumo A2000 Plus has protective device inside to avoid the user from harm.
When open the upper cover of the instrument, the protective device will stop all
moving parts. If open the upper cover when the instrument is working, the device
will stop all movements in process.
In case of operating AutoLumo A2000 Plus not according to the accurate safe
operation procedure, there are some areas may cause harms to body or instrument
damages. In consideration of safety, we paste the below security symbols on these
areas.
CAUTION:
Please check 1.9.3 part of operator’s manual. When you see the following symbols
on the instrument, you can understand the meaning of those symbols.
Symbol Description
Remind you of reading important information on
the label or in Operator’s manual, or of noticing
potential danger existed along with another
symbol nearby.
Biological Risks:
Crushing Of Hands:
Keep away from water: Keep away from water. Please reading
operator’s manual before operation.
The material derived from human, animal, tissue culture or in vitro culture must be
handled in accordance with the principles of potential infection risk. When dealing
with biological hazardous substances, operator should always wear suitable
protective outfit, such as approved disposable gloves, waterproof lab coats and
safety goggles. Handle biological harmful substances according to the process of
equipment biological hazards.
The following items must be carried out as potential biohazards substances:
The size of the instrument & weight of each parts are shown in the below sheet.
Instrument:when the cover and drawer of the instrument are Length=1374 mm
closed Height=1200 mm
Width=950 mm
Instrument:when the cover and drawer of the instrument are Length =1374 mm
opened Height =1700 mm
Width =1530 mm
Ventilation , safe and efficient operation, maintenance Back=340 mm
space required Top=800 mm
Front=800 mm
Space requirement for System control flat (including outer Width=600 mm
computer and go-cart) Height=1700 mm
Depth= 950 mm
CAUTION:
User must use the following accessories which are combined with instrument.
Accessories for safe operation of the equipment is described as follow:
CAUTION:
Change of analytical performance of the device:
When a message of error or a change of result comes out, please contact our
service support engineer.
AutoLumo A2000 Plus has to be operated indoor, and the environment should
meet the following standards.
Temperature Normal Working Temperature: 10℃ ~30℃
Humidity, Normal Working Humidity:≤85%
RH (No agglutination condition) Optimal Working Humidity:50%~70%
Maximum altitude 2000m
Atmospheric pressure 85kPa ~106kPa
Brightness Avoid direct Sunshine and Ultraviolet light
Operating ambient temperature 2℃ per 30 mins
change rate
Pollution degree Based on EN 61010-1-2010, AutoLumo
A2000 Plus applies to the pollution degree:
class II.
Electromagnetic interference Far from Electromagnetic interference
Other instruments interference Far from instruments and equipments with
high power, strong amplitude, strong light
Water quality requirements Sterility deionized water: < 10 cfu/mL
Conductivity:≤5μs/cm (at 25℃)
Hydraulic pressure:50~340 kpa (0.5~3.4
kg/cm²)
Water consumption: 10L/h
AutoLumo A2000 Plus has two separate power supply: one is for instrument and
the other is for outer computer, printer, etc. Each power should meet the specific
requirements.
1.11.1 Instrument
Power supply should meet the following requirements in order to protect the
instrument from damage:
Input power supply AC 100V—230V, 50/60Hz, Single phase
Power cord Attributes Special (only for power supply of
AutoLumo A2000 Plus)
Connector of Power cord AC 250V,10A
Plug of Power cord AC 250V,16A
Power cord 300/500V
Power cord length Less than 2m
Voltage Fluctuation No more than ±10%
Maximum resistance of Safety Grounding No more than 0.1Ω
between instrument and the laboratory
Overvoltage category II
CAUTION:
1) Please use Power cord which supplied by our company.
2) Do not replace Power cord of the instrument by inadequately rated cords.
1.11.2 Computer
Items Requirements
Processor Intel Pentium 4 1.8G
Internal storage 1G
Hard disk 80G
Safety certificate Compliance with CE certificate
CAUTION:
No Standby Power Supply (UPS) provided along with AutoLumo A2000 Plus. If
required, the UPS with Local Earthing Isolator and Low Battery Indicator is
recommended. For more information about UPS, please contact our technical
support department.
UPS Configuration Requirement:
Minimum Power Output 2400 VA
Output Voltage 100-230V AC
Output Frequency 50/60Hz
Output Wave Sine Wave
Working time without load 1000 VA, 15 mins at least
Certificate CE
CAUTION:
If you need to disconnect all parts of the instrument, the power supplies to the
equipment and supplies energy should be disconnected firstly.
1.12 Installation
The AutoLumo A2000 Plus can be unpacked and installed only by authorized staff
or technical person of AUTOBIO. Do not take out the instrument from its package
without engineers presenting to the site.
Interconnection of instrument and other equipment:
Interconnection instruction
CAUTION:
1) The ground of instrument installation position should be flat and hard, and its
bearing capacity is not less than 300kg/ ㎡ .The dimension for floor should be:
Length is not less than 1730 mm; Width is not less than 1380 mm.
1.13 Warranty
AUTOBIO warrants that AutoLumo A2000 Plus delivered to you will be repaired in
accordance with the warranty provisions in the purchase contract for A2000 Plus or
reagents. Clients shall be responsible for execution of the protective maintenance
and cleaning procedures. If misfunction caused by fail of execution of the above
procedures within specified time, fees of repairment and spare parts shall be born
by clients.
The following situations are not covered by warranty.
1. Loss of Warranty Label on the instrument.
2. Instrument serial number is not provided by our company or has been altered.
3. Installation, Power supply and operation of Instrument & Accessories fails to
comply with this manual and its appendix.
4. Damages caused during delivery.
CAUTION:
The Instrument is designed according to related standards, and substantially
conforms to EN 61010-1-2010 Safety Standards.
It’s safe to operate according to this Operator’s Manual. Don not try to make any
changes to the instrument, otherwise, the below results will happen.
• No more quality guarantee supplied by the manufacturer.
• Not complying with the requirements of EN 61010-1-2010 Safety Standards.
• Potential safety hazards.
AUTOBIO won’t be responsible for any damages or defects caused by using the
instrument for any purpose other than the intended purpose, or repaired by any
unauthorized person.
Use period of instruments associated with the instrument operation environment,
frequency of use, etc., and regular maintenance can prolong life.
If any questions, please contact the service provider.
Parts List:
NO. NAME Quantity Instruction for use
1 PVC tube 15m Used for connection between
distilled water and waste dilution, cut
to suitable length according to
installation situation at the site
2 Φ1×Φ3 0.5m Cut to suitable length according to
Silicone maintenance
tube
3 Screw lid 10Ps Use when instrument is In short of
screw lips
CAUTION:
If operation can not comply with the reagent instructions or attempting to use other
company's reagents, any report results and responsibilities arise from above action
will not be borne by AUTOBIO.
1) Structure and components: This product contains three parts: main instrument,
computer(Optional) and software. The main instrument is composed by sample
transportation system, sample adding system, reaction system, reagent storehouse,
test system, bar code reader, communication interface and power device.
2) Automobile A2000 Plus System has two major subsystems: Functional
instrument executing all sample process, & system control platform providing a
human-machine interface. Interactive operation will be available at the several
points as below:
⒃ Input Device ⒄ PC
AutoLumo A2000 Plus has 4 Status Symbols at the left top of software interface.
They can be controlled by 3 control buttons at the right bottom of software interface.
Status Indication Description
Ready:System is ready to use
Pause : No new assay added, but current assays are still in the
process.
Total 6 System Status Buttons are colorless during the normal running of
instruments. View each Status interface by clicking each button. In the following
cases, the button will become red or yellow, the left amount of consumables reach
alert level, or unexpected problem with Sample processing, or event Logs error. The
buttons will return back to be colorless until these problems or alarms got solved.
Status Button Description Alarm Color
Access to “Abnormal Assay” lists RED indicates One or some
shown on sample management Abnormal Assays happened in
interface by clicking. View which process.
samples were detected abnormally. If
the list got cleared, the button will
NOTE
RESTART
At the system status of READY, Attend to the below situation:
Don’t restart the system while the amount of RV, the left space of
solid waste container, left volume of dilution and substrate reach
the alert level.
Make sure the left volume of dilution is sufficient, otherwise, dilution
adding operation can’t be executed during test.
Don’t restart until the temperature of Substrate, Reaction Tray and
Reagent Carousel reach to the normal range.
1. Time Display
2. Temperature Display
Normal Warning Serious Warning
Mode Description
LIS HOST QUERY MODE:AutoLumo A2000 Plus may be interfaced to a
LAB INFORMATION SYSTEM (LIS). Through the communication with
LIS, all the information of samples as well as assay requests can be
acquired by scanning the bar code of samples.
Enable method:[ System Setup] — [LIS Communication Setup] —
[Receiving LIS Requests]—[LIS Interface]—Select[Open]
STAND-ALONE MODE:All the information of sample and assay requests
needs to be input by hand on the operational interface.
Enable method:[ System Setup] — [LIS Communication Setup] —
[Receiving LIS Requests]—[LIS Interface]—Select[Close]
NOTE
Under LIS HOST QUERY MODE, system can read sample data and test request
information input by hand, and can acquire these information from LIS system.
WARM TIP
18 AutoLumo A2000 Plus © AUTOBIO LABTEC INSTRUMENTS CO.,LTD . Rev.V1.1
Operator’s Manual
AutoLumo
As for the samples which system failed to scan its barcode under LIS HOST
QUERY MODE, we may enter into STAND-ALONE MODE (LIS communication
must be set “OPEN” in system setting interface), and input these samples’ barcode
by hand into the systems. Then, Open LIS communication, and perform the
operation of acquiring LIS request. These samples information and assay requests
will be acquired from LIS system.
During the running process of AutoLumo A2000 Plus, you need to shutdown PC or
Instrument or both of them. Please follow the below steps of SHUTDOWN &
RESTART.
WARNING
The below steps have to be followed, otherwise, the potential damage may happen
to the instrument or system database.
CAUTION
Press switch on the left bottom of the instrument, the power (O position) will be off.
The main interface of software on PC shows that the status of system is “NOT
READY”. It means the instrument powers off.
NOTE
1. Use the user interface (UI) to execute standard shutdown
2. In the condition of long time do not use the instrument, select “Instrument of
dormancy” on the interface of maintenance, make the instrument to enter power
saving mode.
System Status: READY NOT READY
WARM TIP
WARNING
During the process of the above performance, it’s possible to touch the potential
infectious materials. Please dispose of these biological hazards strictly according to
the regulation of Lab. Attend to the protection of hand, eye and face especially.
NOTE
When the instrument shut down, the cooling of Reagent Carousel shut down too.
System Status: READY NOT READY
1. Run “ROUTINE CLEAN” program.
2. Take all the RV, sample tubes and racks out from Sample Supplies Unit and RV
supplies Station.
3. Unpack all reagents out from reagents Carousel, put them to 2-8℃ warehouse.
4. Clean large volume waste solution container.
5. Clean Solid waste container.
6. Shut the instrument and PC down, and turn the power off.
In “SLEEP” mode, the instrument stays at low power consumption status, and only
provides the power to thermal insulation systems for keep working, and other parts
stop working.
System Status: READY NOT READY
1. Select “MAINTENANCE” on the main interface.
2. Select “SLEEP MODE” on the “MAINTENANCE” interface.
3. Select “CONFIRM” on the POP-UP window
When the power of instrument is off, follow the below steps to restart AutoLumo
A2000 Plus.
Facing the instrument, and move the bottom switch to I position.
WARM TIP
The system interface of software will stay at “OFF-LINE” status when the PC just
starts.
1. Make sure upper cover of the instrument is at “Closed” status.
2. Turn on the power switch behind the instrument and confirm that the
emergency stop button is at back-out status.
3. Please perform “RESTART” if the computer is not running and start AutoLumo
A2000 Plus system.
4. If the user interface software is running, perform "reinitialization" in the
maintenance interface
5. If the instrument fails to intialize, please contact with our Technical Support
Department
6. Don’t load samples or open the cover of the instrument until the internal
temperature of instrument reach the required level. Normally, it will take 15-20
minutes for the system to recover the temperature to the level required.
7. Make sure the system enters into “READY” mode. If the system cannot get
ready after 60 minutes, please contact technical support team.
8. “Wash substrate” is suggested after “DAILY CLEAN” at the first time. Then, run
the system regularly.
After long term shutdown, please restart AutoLumo A2000 Plus according to the
following steps.
WARNING
During the process of the above performance, it’s possible to touch the potential
infectious materials. Please dispose of these biological hazards strictly according to
the regulation of Lab. Attend to the protection of hand, eye and face especially.
System Status: Not Applicable
1. Restart the instrument
2. Restart the computer and user interface software.
NOTE
“Test & Analysis” icon on the main screen of AutoLumo A2000 Plus software
provides access to “Test & Analysis” interface. Select “Test Requests” there to the
functional areas for the specimen programming.
Interface Description:
NAME DESCRIPTION
Sample Rack Code Click “?” icon below the picture of Sample Rack on the left side
to activate the input box.
Position Fixed automatic numbering from 1 to 5.
Sample ID Sample bar code or Number.
Patient ID/Control Patient ID or control name.
Sample Type Sample types include:Serum, Plasma & Urine etc., totally 10
sample types.
NOTE: Currently, the instrument only supports serum, plasma
and urine, if you need to do sample tests of other types, please
contact our technical support team.
Dilution Factor Default Factor is “1”. It’s optional depending on your
requirement.
CAUTION
1.Maximum 5 samples tube can be put into involved in each sample rack. Both
information of sample racks and samples are empty while entering into “Test
Requests” interface. Sample rack code need to be inputted by hand, once rack
code gets identified by system, samples information on the rack will be collected
automatically. If rack code cannot be identified, it will be regarded as a new one,
and the samples information have to be re-added.
2. Please note that the required sample volume must be greater than 150 μl.
WARNING
During the process of the above performance, it’s possible to touch the potential
infectious materials. Please dispose of these biological hazards strictly according to
the regulation of Lab. Attend to the protection of hand, eye and face especially.
During loading and unloading of samples and reagents, operator must wear medical
glove to protect themselves from biological hazards.
1. Click “?” icon below the picture of Sample Rack on the right side to activate the
inputting column of sample Rack Code
2. Input sample rack code into the column, for example, “HZ01”, then, the sample
ID column is activated too. Sample ID can be input by hand or automatically by
scanning the sample rack bar code.
3. After entering the sample number 1, you can enter the name of the medical
record number/quality control product, setup the sample type, the external
dilution factor, and whether there is any comment, etc. First of all, enter the
name of the medical record number/quality control product.
4. Sample Type Setting: Serum
5. External Dilution Factor: 5
6. Add the sample comment by clicking “Comment” button. Input the sample
information to the pop-up window, and then click “OK” button. Thus, all the
information of sample rack and samples had been input into the databases.
7. Put the sample and sample rack to their position on Sample Loading Platform.
Then, Loading sample work is over.
CAUTION
1. Don’t make any difference to input the information of samples and sample
racks firstly or put the sample rack to its position on Sample Loading Platform
firstly before test start. But, if test started, it’s necessary to input the information
of samples and sample racks firstly, and then, put the sample rack to Sample
Loading Platform. Otherwise, the system cannot identify the samples added, so
that test request is pending, and the result will be marked as “No Test” after test
is over.
2. The Sample Rack Code inputted should be the same as actual Sample Rack
Code loaded.
3. To ensure the reliability of test results, please be sure to the volume of
remaining sample before starting, be supposed to the volume of remaining
sample ≥ 150ul (inherent volume flow rate of a sample cup) + the sample
volume actual tests need.
NOTE
3. Click “OK”, then the new edited information will be saved automatically.
Click the “Sample Details” button again, the pop-up patient information is the saved
information.
Select “Test Request” for current samples. One or more tests can be carried out for
each sample.
1. Select the sample which needs test request.
2. Add test request in selection box of normative test or combination box and set
Dilution Ratio.
1. Deleting sample
a) Input Sample Rack code in the column.
b) Select the sample that you want to delete.
c)
Click “Delete Sample” icon on the right side of interface.
d)
Click “confirm” on the pop-up window, then, the sample and the test request
information it included got deleted, Remove the samples from the sample
rack.
2. Deleting Sample Rack
a) Input Sample Rack code in the column.
b) Click “Delete Rack” icon on the right side of interface.
CAUTION:
The samples under the sampling line, the software will not be able to edit, nor can it
be moved or removed.
4.1 Requested
Enter into “Test & Analysis” screen from the main menu screen by clicking its icon,
and then select “Requested” button within the column of “Queue”. The sample
Status is displayed on the “Requested” screen, which test had been ordered and
scheduled for processing, but without being pipetted.
Screen Description
Button Name Definition
Close Exit the current screen and return back to its previous
(Labeled F12) screen of “Test & Analysis”, “Maintenance” and
“Diagnostic” or directly back to the main menu if you didn’t
open “Test & Analysis” interface before.
Find(Labeled F2) Search for a sample of specified Sample ID.
Edit Request (Labeled F3) Switch to “Test Requests” screen for editing the samples
selected.
Test Result (Labeled F10) Switch to “Test Result” screen.
Test Request (Label F11) Switch to “Test Request” screen for editing.
CAUTION:
Sample Rack Code and Sample Position cannot be edited unless being deleted and
inputted again.
4.2 In Progress
Enter into “Test & Analysis” screen from the main menu screen by clicking its icon,
and then select “In Progress” icon within the column of “Queue”. The samples which
tests are in process are displayed on the “In Progress” screen.
Screen Description
Button Name Definition
4.3 Exception
Enter into “Test & Analysis” screen from the main menu screen by clicking its icon,
and then select “Exception” icon within the column of “Queue”. The samples which
tests are exceptional are displayed on the “Exception” screen.
Screen Description
Button Name Definition
Close Exit the current screen and display its previous “Test &
(Labeled F12) Analysis” screen or directly back to the main menu screen
if you didn’t open “Test & Analysis” screen before.
Clear All(Labeled F2) Clear all Exception Tests on the “Exception” Interface.
Clear Selection(Labeled F3) Clear Exception Tests selected.
Test Result (Labeled F10) Switch to “Test Result” screen.
Test Request (Labeled F11) Switch to “Test Request” screen for editing.
1. Clear all exception tests:
Click “Clear All” icon, and then all exception tests get deleted.
2. Clear Exception Tests selected
Select exception tests you want to delete, and click “Clear Selection” icon. Then, the
selected exceptional tests get deleted.
NOTE
1. Information shown “Exception” interface include “Rack”, “Position”, ”Sample ID”,
“Exception test”, “Occurred Time”, “Flag”.
2. Refer to Appendix 1, the Description of “Fatal Flag” & “Non-Fatal Flag”.
Enter into “Test & Analysis” screen from the main menu screen by clicking its icon,
and then select “Work Pending” icon within the column of “Queue”. The samples
which tests assay errors happened in process are displayed on the “Work Pending”
screen.
Screen Description
Button Name Definition
Close Exit the current screen and display its previous “Test &
(Labeled F12) Analysis” screen or directly back to the main menu screen if
you didn’t open “Test & Analysis” screen before.
Clear All(Labeled F2) Clear all the work pending.
Clear Selection(Labeled F3) Select the work pending listed on the screen, and click “Clear
Selection” button. Then, the selected work pending is deleted.
Test Result (Labeled F10) Switch to “Test Result” screen.
Test Request (Labeled F11) Switch to “Test Request” screen for editing.
NOTE
1. “Work pending” interface shows all tests which assay errors happened in
process, such as “CLT(Clot of sample pipettor)”, “QNS(Quantity of sample is
Enter into “Test & Analysis” screen from the main menu screen by clicking its icon.
Click “Daily” icon, the “Daily” window come out with all the samples information
according to specified date.
Screen Description
Button Name Definition
Close Exit the current screen and display its previous “Test &
(Labeled F12) Analysis” screen or directly back to the main menu screen
if you didn’t open “Test & Analysis” screen before.
Find(Labeled F2) Search for the specified sample with its Sample ID.
Details (Labeled F3) Show all the samples information on the day, including
Rack, Position, Sample ID, Patient ID/Control, Status, and
Completed Time.
Test Result (Labeled F10) Switch to “Test Result” screen.
Test Request (Labeled F11) Switch to “Test Request” screen for editing.
Screen Description
Button Name Definition
Close Exit the current screen and display its previous “Test & Analysis”
(Labeled F12) screen or directly back to the main menu screen if you didn’t
open “Test & Analysis” screen before.
Custom(Labeled F2) Determine which type of results’ information to show and display
sequence.
Filter(Labeled F3) Establish the filter condition, and filter test results.
Find(Labeled F4) Search the test results from filtered results according to specified
condition.
Details(Labeled F9) Show all the detailed information of selected test result.
Send to LIS(Labeled F6) Send the test result to LIS.
Print(Labeled F7) Print the test result out.
Refresh(Labeled F5) Refresh current test results.
Determine which type of results’ information to show. Click “Custom” button, and
select the desired options on pop-up “Custom setting” window as well as display
sequence. After click “confirm” button, the results display on the screen according to
Custom setting.
5.2 Filter
Click “Filter” icon to enter into “Filter” screen for setting of existed filter condition or
setup of new filter condition. After setting filter condition, click “confirmation”, the test
results will show on the screen according to the filter conditions selected or
established.
5.3 Find
Click “Find” icon to search the test results among filtered results according to
specified condition.
5.4 Details
Select the test result, and click “Details” icon or double-click current row for more
information of selected result, which will be displayed on the pop-up window.
Select the test result, and click “send to LIS” icon to send the result to LIS.
5.6 Print
Print the selected test result out in report format. Click “Print” icon, and the results
will be printed as report form. Refer to description of “report form” section.
5.7 Update
Update the test results to the newest state, click the refresh button to update.
6.1 Calibrator
Open “Calibrator Setup” screen by clicking its icon on the “Test & Analysis” screen.
Get access to “Test & Analysis” screen from The Main Menu.
Screen Description
Button Name Definition
Close(Labeled F12) Exit the current screen and display its previous “Test &
Analysis” screen or directly back to the main menu screen if
you didn’t open “Test & Analysis” screen before.
Add Calibrator(Labeled F2) Select it, “Calibrator” dialog window pop up for adding new
calibrator.
Edit Calibrator(Labeled F3) Select one calibrator, and click “Edit Calibrator” button. Then,
“Calibrator” Dialog window of selected calibrator pop up for
re-editing. “Edit Calibrator” button void without calibrator
selected.
Calibration(Labeled F5) Go to the calibration results interface
1. Press “Add Calibrator” Button, “Calibrator” Dialog window pop up for adding
new calibrator.
1. Select the calibrator you want to edit, and click “Edit Calibrator” button to
display “Calibrator” Dialog window.
2. Scan or input the bar code of calibrator.
3. Click “OK” button.
Open “Request Calibration” screen by clicking its icon on the “Test & Analysis”
screen. Get access to “Test & Analysis” screen from The Main Menu.
CAUTION:
1. Each sample rack involves 5 samples at most, but one set of calibration
requests have 5-7 standards. So, it’s possible to load two sample racks for
calibration test.
2. In the selection of calibrator, only allowed to select the one which has the same
batch with its corresponding reagent.
Screen Description
Button Name Definition
Sample rack Code Click “?” icon below the picture of Sample Rack on the
left side to activate the inputting column of sample rack
code
Position Automatically number from 1-5.
Calibrator Fixedly mark calibrators from S0 to S6.
Test Times “Test times” is changeable. Qualitative assay calibration
can be executed 1-3 times, and each calibrator’s assay
times can be defined separately. Quantitative assay
calibration times can be executed 1-2 times, but, each
calibrator’s assay times have to be same. Default test
times for qualitative and quantitative assay calibration
are 2.
Comment Add the description of sample.
Item Name Differentiate different test items.
Lot Manufacture date of reagent
Expiration Expire date of reagent
Left times Left times of a reagent
1. Scan or input the code of first sample rack with calibrators which calibration
request needs to be edited on the inputting column.
2. Select the relevant reagents on board from “On Board Reagent” column.
3. Select the calibrators scheduled for test from “Calibration” column.
4. Click “Select Calibrator” button.
5. Scan or input the code of sample rack with the calibrators selected for test.
Open “Calibration” screen by clicking its icon on the “Test & Analysis” screen for
reviewing the calibration results. Get access to “Test & Analysis” screen from The
Main Menu.
Screen Description
Button Name Definition
Close(Labeled F12) Exit “Calibration” interface
Filter(Labeled F2) Filter the calibration results according to the defined
condition.
Review Data(Labeled F4) Review the test data and calibration curve of selected
calibration item.
Calibrator Setup(Labeled F5) Enter into “Calibrator Setup” interface.
Open “QC Setup” screen for managing QC by clicking its icon on the “Test &
Analysis” screen. Get access to “Test & Analysis” screen from The Main Menu.
Screen Description
Button Name Definition
Close(Labeled F12) Exit the current screen
Add Control (Labeled F2) Add new quality control to the system.
Edit Control (Labeled F3) Edit the current quality control
QC maintenance Setup the following values: Means Value (Target Value),
(Labeled F5) SD, Test Items and other basic data.
Quality Control (Labeled F6) Switch to “Quality Control” interface.
1. Enter into “QC setup” screen, and click “Add control” button. The dialog window
of “Control” comes out for adding control information.
2. Input the following information of quality controls, Bar code, name of quality
control, Lot, Expiration, in case quality controls are not manufactured by
AUTOBIO.
3. Only input Bar code of quality controls that are manufactured by AUTOBIO.
4. Select the QC rule on dialog window of “Control”.
5. Click “Ok” button.
NOTE
QC Rule:
1. 1±2s (1-2s) – Flag when one point is more than 2 SD from the mean.
2. 1±3s (1-3s) – Flag when one point is more than 3 SD from the mean.
3. 2±2s (2-2s) – Flag when 2 consecutive points are more than 2 SD from the mean in
the same direction. Only the second point is flagged
4. 4±1s (4-1s) – Flag when 4 consecutive points are more than 1 SD from the mean in
the same direction. Only the fourth point is flagged.
5. 7x (7x) – Flag when 7 consecutive points are on the same side of the mean (for
example, 7 points are above the mean). Only the seventh point is
flagged.
7.2 QC Maintenance
Click “QC Maintenance” button on the screen of “QC Setup”, and the screen of “QC
Maintenance” appears. Here, administrator privileges can maintain third-party QC.
Screen Description
Button Name Definition
Close(Labeled F12) Exit the current screen and back to the screen of “QC Setup”.
New(labeled F2) Add the new QC.
Delete(Labeled F3) Delete the current QC selected as well as its test items
included.
Add Item(Labeled F4) Add the selected item in the column of “Item Selection” to the
current QC
Delete Item(Labeled F5) Delete the selected item included in the current QC.
1. Enter into the “QC Maintenance” Interface, and click “new” button for adding
new control.
NOTE
1. “Control Number” has to be 3 digits number from 100-999.
2. “Control Number” doesn’t allow to be repeated.
3. “Control Number” & “Control Name” cannot be empty.
4. Control has to include one test item at least.
7.2.2 QC Request
Open “Quality Control” screen for reviewing Quality Control results by clicking its
icon on the “Test & Analysis” screen. Get access to “Test & Analysis” screen from
The Main Menu.
NOTE
The color of normal Quality Control result is blue, but for abnormal or failed Quality
control result, the color will become red.
Screen Description
Button Name Definition
Close(Labeled F12) Exit the current screen and display its previous “Test &
Analysis” screen or directly back to the main menu screen if
you didn’t open “Test & Analysis” screen before.
Filter(Labeled F2) Filter the quality control results according to the defined
condition.
Review Chart and data Review result data and QC drawing of selected Quality
(Labeled F4) control. The button void without selecting Quality control.
QC Setting(Labeled F5) Click to switch directly to the QC setting interface.
1. Click “Filter” button on the screen of “Quality control”, the dialog window come
out for filtering QC results.
2. Select filter condition.
3. Select “Sort Order”.
4. Click “Ok”.
Click “Review Chart and Data” button for reviewing QC result and Diagram on the
screen of “Quality Control”.
Screen Description
Button Name Definition
Basic information of 1, Basic information:”Start Date”, ”End Date”, ”Control”, “Control Lot”,
Quality Control “Expiration”, ”Reagent Lots”, “Mean”, “SD”, “%CV”, “Data Points”.
2, “Mean”: Average value of all test datum involved in calculation.
“SD”: Standard deviation of all the tests datum involved in
calculation.
“%CV”: Coefficient variation of all the tests datum involved in
calculation.
“Data Point”: Records count that participate in calculation.
Test data list of 1, Verify the test data according to QC criteria selected. In case of the
Quality Control breach of QC criteria, the criteria breached will be marked in red or “*
*”. As for the breach of the criteria of “2-2S”,”4-1S”,”7X”, only the last
data is marked as invalid.
2, Test data list only shows the test results of quality control with data,
and only mark two QC results maximum which are out of range.
3, For Qualitative Tests, the data is “S/CO” Value (Actually RLU/cutoff
1. Filter QC Point: Filter the QC results through “Data” and “Reagent Lots”. The
condition of “Date” includes “all dates”, “Last week”, “Last month”, “Defined
Date (“Start day”, “End day”)”. The condition of “Reagent Lots” includes “All
Lots”, and “Each Reagent Lots”.
2. “Delete QC Point”: Only administrator User has the right to delete the QC point.
Password is required while common user wants to delete QC point.
3. “Ignore”: The QC point ticked as “ignore” won’t be used to calculate the SD
and %CV value, and the QC Diagram won’t show this QC point.
relevant QC point has comment, and “ ” means the relevant QC point doesn’t
have comment.
Screen Description
Button Name Description
Close (Labeled F12) Exit the current screen and return to “Test & Analysis” screen.
Change (Labeled F2) Execute the operation of replacing the reagent
Reagent Inventory Switch to the reagent manage menu
(Labeled F3)
3. When the appointment indicator lights up, it is showed on Nixie tube that after
XXmins and XXs change the reagent.
4. When the count down showed on Nixie tube is not up to 00:00 , press the
“Appointment” button for more than 3s to cancel the online reagent replacement.
6. User can change the reagent when it is indicated on the customer software that
“The reagent on instrument can be replaced” and the countdown is up to 00:00.
7. Open the cover on left side of reagent tray to find the position for replaced
reagent or new reagent and put them into the tray.
8. Press “Rotate” button to turn reagent tray if the position for new reagent can not
be found in view. Press for once, the tray turn one-third of circle (eight reagent
position).
9. Bar code reading for slot number of the reagent can be started when the operator
clicks the ‘Appointment’ for more than 3s after the reagent being put into the tray
and cover being closed.
10. The reagent information will be showed on screen after scanning is completed.
11. If the slot number was scanned incorrectly, the indication will be showed on the
screen.
12. User can use hand-held bar code reader to scan the reagent and add reagent,
please refer to 8.1.2 for steps.
CAUTION
1.‘Appointment’ function can not be initiated when the instrument is under the status
of ‘Reinitialization’, ‘Daily Clean’, ‘Daily Maintenance’, ‘Substrate Tube Washing’,
‘Implementing Diagnostic Function’and ‘Changing Reagent Kit’
2. Mixing the reagent adequately by manual before putting the reagent into tray
during On-Board Change Reagent.
Select ‘Reagent Inventory’ button to enter reagent list management window from
the reagent management screen.
Note:
1. If the reagent is unload already, the slot will display ‘--’
2. If the reagent is expired or its calibration is expired, it will be displayed in red
color.
Screen Description
Button Name Description
Close(Labeled F12) Exit the current window and return to the secondary menu (Test
& Analysis, maintenance and Diagnosis) or directly back to the
main menu if you didn’t open the secondary menu before.
Filter(Labeled F2) Set the filter factor to filtrate reagent records
Print(Labeled F3) Print filtrated reagents and their detailed information
Refresh(Labeled F4) Refresh on-board reagents
On-board Display all reagents on the machine
Reagent(Labeled F11)
8.2.1 Filter
In the main menu, click the "reagent" on the interface to enter the reagent alarm
form, and check the reagent alarm information.
CAUTION
1. It indicates serious alarm if the button background is red and general alarm if that
is yellow
Interface Description:
Button Name Description
Close Drop out the interface and return to secondary navigation interface
(Detect Analysis).
Reagent Inventory Return to reagent kit list management interface.
Reagents Tray Display all existing reagent kit.
Reagent Change Carry out the operation of reagent change on the instrument.
Enable/Disable Enable or disable the operation of reagent kit on the instrument.
Disenable Mixing Disenable reagent mixing function.
Enable Mixing Enable reagent mixing function.
1. It will display details information while clicking the button which represents
reagent kit (If there is an exception, the abnormal information column is marked in
red or yellow). If there is alarm information for the reagent kit, the alarm information
will be automatically oriented in the list.
Screen Description
Button Name Description
Close(Labeled F12) Exit the current window and return to the secondary
menu (Test & Analysis, maintenance and Diagnosis)
or directly back to the main menu if you didn’t open
the secondary menu before.
Change Substrate(Labeled F2) Change the CLIA substrate.
Switch Substrate(Labeled F3) In the “ready” state, the active substrate and the
standby substrate can be switched back and
forth;Other states are not available
Add RV(Labeled F4) Place the reaction vessel to the cup supplier deck
Clean Solid Waste(Labeled F5) Change the bag contains wasted reaction vessel
Change Diluent(Labeled F6) Change the diluent liquid
CAUTION
1. The system will prompt user to change the substrate if the substrate is not
enough when running. At this point the instrument is no longer planning a new
test, but the current test will still continue.
2. Only in “Ready” state the ‘Change Substrate’ button is active, please flush
substrate pipelines carefully after the change operation.
Warning
In the process you will be exposed to potentially infectious materials, please
execute the standard laboratory procedures for handling and disposal the biological
hazardous material. Take appropriate methods to protect hands, eyes and face.
8.4.2 Add RV
1. Click ‘Clean Solid Waste’ button, switching to "Clean Solid Waste" dialog.
2. Open the bottom cover of the instrument.
3. Take down the container of solid waste and put the waste into biological waste
bags
4. Install the container of solid waste.
5. Close the bottom cover of the instrument.
6. Click ‘OK’.
Warning
In the process you will be exposed to potentially infectious materials, please
execute the standard laboratory procedures for handling and disposal the biological
hazardous material. Take appropriate methods to protect hands, eyes and face.
1. Click ‘Change Special Wash’ button, switching to “Change Special Wash” dialog.
2. Open the top cover of the instrument.
3. Take down the old wash bottle and change into a new one. Put the old one into
biological waste bags.
4. Install the new bottle.
5. Close the cover of the instrument.
6. Click ‘OK’.
CAUTION
When doing dilution experiments, please make sure that dilution position has
placed sufficient diluent, if diluent shortage during the test, instrument will alarm
automatically, right now you need to add diluent manually, and click "change
diluent" in the interface of bulk, the system will detect automatically, and cancel the
alarm.
The default interface is the general settings when enter the system setting.
Screen Description
Button Name Description
System ID Set the System identification (1-12 characters)
Language The Language of the software includes:
(Chinese(Simplified, PRC)-Chinese
(Simplified, PRC)、English-English)
Screen Description
Button Name Description
Organization Set the name of organization
Laboratory Set the name of laboratory
Telephone Set the telephone number
Director Set the name of director
Report comment Set the comment of the test report
Printer Select the printer
Report Select report type, click “Format Adjustments” if it need to
change
1. Click ‘Save’ after achieving the settings.
2. Click ‘Close’ if you want to cancel the changes.
Screen Description
Button Name Description
LIS Interface【ON】 Open the LIS communication interface
LIS Interface【OFF】 Close the LIS communication interface
Auto send to LIS---Result Type Select send all results or close the function
Send mode Select the mode of send to LIS(By test or by sample)
Port Set the LIS communication interface(COM0、COM1、
COM2、COM3、…)
Baud Rate Set the Baud Rate of LIS communication interface
Date Bits Set the data bits of LIS communication
Parity Set the parity parameters of LIS communication
Flow Control Set the flow control of LIS communication
Stop Bits Set the stop bits of LIS communication
LIS communication path Set LIS export data location
1. Click ‘Save’ after achieving the settings.
2. Click ‘Close’ if you want to cancel the changes
Screen Description
Button Name Description
Old Password Enter the old password of administrator
New Password Enter the new password of administrator
Verify Reenter the new password of administrator
1. Click ‘Save’ after achieving the settings.
2. Click ‘Close’ if you want to cancel the changes
Screen Description
Button Name Description
Frequency Set the frequency of data backup(Daily or Weekly)
Start date Set the begin date of data backup(Monday, Tuesday,
Wednesday, Thursday, Friday, Saturday, Sunday)
Start time Set the start time of data backup(0-24 hour)
Next Backup Time Display the next time of data backup
1. Click ‘Save’ after achieving the settings.
2. Click ‘Close’ if you want to cancel the changes.
9.1.6 About
Within the permission of maintenance staff, the user can obtain the information of
version number and CanID number in client software.
Screen Description
Button Name Description
Filter test list Display test information by filter
Filter(Labeled F2) Set the filter factor to filtrate the test
Edit (Labeled F3) Edit the parameters of test items.
Print(Labeled F4) Print selected test
Update (Labeled F11) Update test items.
Click ‘Close’ if you want to cancel the operation.
9.2.1 Filter
Screen Description
Button Name Description
Name Display test item name and can not be edited
Full Name Display full name of test item and can not be edited
ID Edit test item ID number and the number range is from
100 to 999.The repeat test item number is not allowed
to enter.
Places Set the places number of the test results (0-5),
Basic Settings
exceeded the range will be handled by rounding up or
rounding down.
Units Set data units of the test results(IU/mL、mIU/mL、
NCU/mL、ng/dL、ng/mL、pg/mL、U/mL、μIU/mL、
pmol/L)
Enabled Set enable the test or not.
Normal Range Min The minimum value of the normal range.
Max The maximal value of the normal range.
Reference Range Set the reference range for printing the test result
(0-60 characters) in range1 to range 6
Note:Reference range will not be printed in the
current report.
Qualitative Gray upper Set the value equal to or higher than the critical value
Judgment Gray lower Set the value equal to or lower than the critical value
Less than Set the result when the result is less than the gray
gray lower
Less than Set the result when it is less than critical value 1.
critical
value 1’s
results
results Set the result when it is between critical value 1 and
between critical value 2.
critical
value 1
and critical
value 2
More than Set the result when it is bigger than critical value 2
critical
value 2’s
results
1. Click ‘Save’ after achieving the settings.
2. Click ‘Close’ if you want to cancel the settings.
privileges
Screen Description
Button Name Description
New Create a new test panel
Delete Selected Delete the selected test panel
Save Save after modifying test panel content
Test Panel ID The ID cannot be changed if it is existed already.
Enabled Enabled the selected test panel or not.
Sample Type Select the type of sample.
Panel Name Enter a name for a new test panel or edit the name of an
existed test panel.
Optional Test Display all tests of current sample type
Selected Test Display the tests which are added into the test panel.
Add Add the test into a test panel(multi-test and repetitive are
allowable)
Remove Remove the selected test.
1. Click ‘Save’ after achieving the settings.
2. Click ‘Cancel’ if you want to cancel the changes.
3. Click ‘Close’ if you want to close the window.
privileges
This function is used to derive a new result through four arithmetic calculate
(addition, subtraction, multiplication, and division) according to one or more existing
test results. As a new test item, the test result will creat a new data in accordance
with derivation rules after patient sample is completed with all tests.
Screen Description
Button Name Description
Derived Results Display the existing derivation result information.
Derived Name Set the name for derived formula.
Formula Units Set the units for the test results.
Enabled Enabled the selected Derived Formula or not.
Decimal Set the decimal place of test results (0-5), exceeded the
Place range will be handled by rounding up or rounding down.
Formula Enter or edit the derived formula.
Symbol The button for derived formula.
Note:
Some command buttons may be not displayed due to different management
privileges. If you want to adjust the items, please contact with our engineers.
10.1 Maintenance
Data Clear:
The historical data can be deleted in the Data Clear menu by selecting time and
data type. This operation should be careful as the data cannot be recovered.
Restoration:
Choose restoration and it will give a dialog box asking whether to recover the entire
system, select Yes or No. Recovery system will initialize the system, and therefore
the database backup should be carried out first to prevent database corruption or
data loss cause by incorrect content.
10.2 Clean
Daily Clean
Cleaning procedures should be performed by operator before starting the test every
day. Due to the substrate tip has preheating function, it is recommended to clean at
least 20 times before the start of the experiment.
Enter the times for washing and then click’ Perform Routine Clean’, the progress of
cleaning will be displayed at the below of the window. Do not try to finish the
process of the clean, to avoid database damage or failure of the operation.
Notice
The substrate may affect test results if it is in high temperature for a long time, as
the substrate tips have preheating function. The operation of substrate spit out
should be executed strictly.
Routine Clean
It is suggested to run the cleaning procedures after finishing the test every day. The
instrument should be cleaned in the longest idle date if your laboratory is continuing
to run.
Wash Substrate
Replacement of substrate in two ways:
1. When changing a different batch of substrate, first pull the substrate tube out from
the substrate bottle, then emptied at least 100 times, and wipe outside of the tube
with distilled water, followed insert the substrate tube into a new bottle, last flush at
least 100 times.
2. To change the same batch of substrate or if the instrument is idle for a long time,
please flush at least 50 times. See the figure blow.
WARNING
CAUTION
To ensure the accuracy of experimental results, suggest that change the sample
pipettor once every half a year, change the liquid pipeline once every half a year.
10.5 Debug
Sensor List:
Display port status of each sensor, convenient for you to troubleshoot and feedback
problems.
10.6 System
Shutdown:
In case of relocation the instrument or shutdown it for more than 5 days, please
close the software and computer.
Low Power:
In the condition of do not use the instrument for a long time, select instrument
hibernated to save energy. At dormant state, instrument cannot run testing.
Reconnect:
Restart software and computer.
Close software:
Close the software.
Reset system:
Perform initialization of the control system and the operating system in the state of
“Not Ready”. During the process, the instrument will not be able to operate.
Change role:
Different user permissions switch.
10.7 Reinitialization
When the instrument cannot work normal caused by fault or other problems, you
can reinitialize the software and the operating system. Select the items you need
Notice
Before the reinitialization you should perform each needle returns to the initial
position, in order to reduce the risk of striker.
11.1 Preparation
2. Check whether the reagent kit on RDT is sufficient for test, if not please add the
reagent.(Refer to chapter 8)
1. Click the “Bulk” button and enter into “Bulk Supplies” interface
2. Check whether left count of Substrate I and Substrate II can meet the
requirements of test. Please add new substrate if they are not sufficient (Refer to
the section of 8.3.1).
Notice:
Do not place the substrate A and substrate B in contrary. Please refer to the
color of bottle cap.
Please check whether the pipette inside of the substrate bottle is bent and use
medical gloves to straighten it before place the substrate.
11.1.4.1 RV Check
1. Click ‘Bulk’ button and enter into ‘Bulk Supplies’ interface.
2. Check whether the left count of RV can meet the requirements of test, if it is not
sufficient, please execute the operation of ‘Adding RV’
3. Check if there are enough RV in the RV container. Please add RV without delay
and execute the operation of ‘Adding RV’
Notice:
The number of RV in the container can not exceed 1000 and the height can not
reach up to two third of the container.
After RV is added to the container, operation of ‘Adding RV’ will be executed.
3. Click ‘Reinitialization’
Second: Click on main menu ‘Analyse Test’--‘Patient Test Request’ and enter into
the interface of ‘Test Request’
After completion of sample edit, put the sample rack into SRK and close the SRK
cover, click ‘Start’ button and start to test.
Power off the instrument and computer after test is completed (the power switch is
on the rear of instrument).
Buttons Description:
Button Name Description
Close Close the window and return to the main window
Push Rack Push out the uninitialized sample rack in the fault status
Clean Incubation Units Clear away all reaction vessels in the reaction stage and
preachment tray.
Sampler Check Check the quantity, position and bar code of sample by
scanner inside.
Liquid Check The sample pipettor and reagent pipettor will detect the
liquid level of sample, reagent and diluent to check the
liquid detection function.
Quantitative Check The difference between the test value and set value is
measured by the balance to check dispense accuracy.
Reagent Check Scan the bar code of the reagent to check the information
and liquid level
Dark Value Check Check the dark value of the detection device with nothing
Note:
Please contact with engineers of AUTOBIO before diagnosis test, make sure
whether you can diagnose or adjust those projects or not.
Error 1:
Description:
Reaction vessels is not enough or slide is blocked, please check immediately.
Overtime for supplying of reaction vessels, lacking of reaction vessels.
Error analysis:
1. Software—consumables—warning of lacking of RV.
2. There is not enough RV in the container.
3. Reaction vessels slide get stuck.
4. Reaction vessels slide inlet get stuck.
Solutions:
1. Execute steps from the software: “consumables”—”adding reaction vessels”
and click “OK”.
2. Supply reaction vessels into the container in time and click ”OK”.
3. Open the cover of the container, clear off the stuck vessels in the slide, and close
the cover, then click ”OK”.
4. Open the cover of the container, clear off the stuck vessels at the inlet of slide,
and close the cover, then click ”OK”.
Error 2:
Description:
The lip of the reaction vessels alignment device is open.
Error analysis:
The cover of reaction vessels feeder is not closed.
Solution:
Closet the cover and click “OK”
Error 3:
Description:
Click “OK”, it is displayed on the screen “Disable” for test.
Error analysis:
The cover of the sample loading device is not covered.
Solution:
Reclose the cover of sample loading device and click “OK”.
Error 4:
Description:
Sample line jam
Error 5:
Description:
The user presses the emergency stop button on the instrument.
Error analysis:
The emergency stop button is pressed.
Solution:
1. Open the front cover of the instrument, clean off the waste reaction vessels.
2. Reclose the cover of reaction tray, remove each pipette tip and the transfer arm
to the home position.
3. Close the front cover, click emergency stop button.
4. Reinitialize the instrument, and reedit the sample, and click “OK” to start the test
again.
Error 6:
Description:
Front cover of the instrument is open, emergency stop.
Error analysis:
Front cover of the instrument is open.
Solution:
1. Open the front cover of the instrument, clean off the waste reaction vessels.
2. Reclose the cover of reaction tray, remove each pipette tip and the transfer arm
to the home position.
3. Close the front cover.
4. Reinitialize the instrument, and reedit the sample, and click “OK” to start the
test again.
Error 7:
Description:
Rear cover of the instrument is open, emergency stop.
Error analysis:
Rear cover of the instrument is open.
Solution:
1. Open the front cover of the instrument, clean off the waste reaction vessels.
2. Reclose the cover of reaction tray, remove each pipette tip and the transfer arm
to the home position.
3. Close the front and rear cover.
Error 8:
Description:
Transfer arm 1 can not put up the RV in right place constantly for 3 times, please
check if there is any problems for RV alignment device or transfer arm 1 .
Error analysis:
RV alignment device or transfer arm 1 is abnormal.
Solution:
1. Please refer to Error 1 if RV supply is abnormal.
2. If RV supply is normal, click “PAUSE” button, after the testing items is completed,
reinitialize the instrument and start test again.
Error 9:
Description:
Buffer area is full of unloaded sample racks, please clear off the sample racks
unloaded.
Error analysis:
Buffer area is full of unloaded sample racks.
Solution:
Open the cover of sample loading device, take out of the sample racks from buffer
area and close the cover, click “OK” and continue the test.
Error 10:
Description:
The instrument is not initialized and can not be started for test.
Error analysis:
Start the test in the status of “NOT READY”
Solution:
Reinitialize the instrument and click “OK” to start the test.
Error 11:
Description:
The sensor for the distilled water can not work properly, please check.
Error analysis:
Distilled water is not full in the distilled water bucket in the limit time, the outer join
for distilled water is not enough.
Solution:
Supply the distilled water without delay.
Error 12:
Error 13:
Description:
When RPP1V driving toward downside, the RPP1V stays above RTT/ORTT while
the PTT/ORTT are turning, or RPP1V stays above RDTT while the RDTT is turning.
Error analysis:
The reagent carousel cover is not closed.
Solution:
Circumstance 1: “PAUSE” is displayed on the screen: After the ongoing test is
completed, close the reagent carousel cover and front and rear cover of instrument,
then reinitialize the instrument and continue the test.
Circumstance 2: “NOT READY” is displayed on the screen, test is stopped:
1. Open the front cover of the instrument, clear off the waste reaction vessels in the
reaction tray.
2. Close the cover of reaction tray again, move each pipettor pin and transfer arm
to HOME position.
3. Close the front and rear cover of the instrument and reaction tray cover.
4. Reinitialize the instrument and reedit the sample, click “OK” to start test.
14.1 Transport
The transport of the instrument should be avoided of moisture, water, excessive
vibration and extrusion. Handling, loading and unloading should be executed under
the guidance of professional engineers. Temperature range between -40℃ ~ +55℃,
and relative humidity is 5% -70%RH .
Essential information of symbols on instrument’s package as below:
Symbol Description
This symbol shall be accompanied by the
manufacturer's serial number.
Serial Number:
Keep away from sunlight: Please keep away from sunlight when transport
Keep Dry:
Contents of the transport package are fragile
therefore it shall be handled with care
Fragile:
Indicates correct upright position of the transport
package
This way up:
Transport package shall not be rolled
Do not roll:
14.2 Storage
The instrument should be stored in a room with non-chemical, non-corrosive gas,
good ventilation, sanitation, temperature range between -40 ℃ ~ +55 ℃ , and
relative humidity is 5% -70%RH.
Fatal Flag
(1) Waste liquid and discard Other 1. When repairing instrument, service
sample cup may cause personnel must wear protective mask
biological hazard to service and glove.
personnel
2.Service personnel should get a
professional pre-service training
(4) The risks of crushing the Frame 1. Service personnel should get a
hand by the upper, front and professional pre-service training.
back covers.
2.Add warning symbol on instrument
(5) The power indicator may not ELEC 1. Before operating the electrical
lightened when the voltage cabinet, assure that the power switch is
is 12V, which may cause the off, the Power cord is cut off and the
risk of electric shock. instrument should stay out of operation
for 10 minutes.
2. Service personnel should get a
professional pre-service training.
Barcode Coat Barcode surface is coated with protective plastic after printing.
Requirements Mate coat is suggested. The barcode reader is same with
eyes. If the reflected light is too strong, the barcode can not be
read successfully.
Barcode installation Adjust by using of tooling.
distance At the distance of 35mm(from scanning window to barcode).
(Service staff)
No.199, 15th Ave, National Eco & Tech Zone, Zhengzhou 450016, China