Quality
Quality
Immunoassay Analyzer
REAGENTS
REF C89003G 2×50 Tests Reagent kit, 100 tests, 2 packs, 50 tests/pack
Tg coated microparticles, 3.5 mL/pack,
R1
0.05% ProClin 300.
INTENDED USE
The iFlash-Anti-Tg assay is a paramagnetic particle Anti-Tg acridinium-ester-labeled conjugate,
R2
chemiluminescent immunoassay (CLIA) for the 6.5mL/pack; 0.05% ProClin 300.
quantitative determination of antibody to thyroglobulin (Tg)
Sample diluent, phosphate buffer, 10
in human serum and plasma using the iFlash R3
mL/pack; 0.05% ProClin 300.
Immunoassay Analyzer.
Calibrator 1, 1 bottle, 1.0 mL, phosphate
SUMMARY AND EXPLANATION CAL1 buffer with protein stabilizers, 0.05%
Thyroglobulin is synthesised in the thyrocytes, which ProClin 300.
release it into the follicular volume, where it plays an
important part in the production and storage of thyroid Calibrator 2, 1 bottle, 1.0 mL, Anti-Tg in
CAL2 phosphate buffer with protein stabilizers,
hormones. A concomitant elevation of autoantibodies to
0.05% ProClin 300.
both thyroid peroxidase (TPO) and Tg is indicative of a
hypertrophic form of autoimmune thyroiditis, namely Calibrator 3, 1 bottle, 1.0 mL, Anti-Tg in
lymphadenoid goitre (Hashimoto's disease), as well as of CAL3 phosphate buffer with protein stabilizers,
the atrophic form of autoimmune thyroiditis. 0.05% ProClin 300.
(NaOH) and should be avoided contact with eyes. Ensure that residual fibrin and cellular matter have
been removed prior to analysis.
REAGENT HANDLING
Use with caution in handling patient specimens to
The reagents may not be used after the stated prevent cross-contamination.
expiration date.
Do not use heat-inactivated samples.
Avoid the formation of foam with all reagents.
Ensure that the patient samples, calibrators and
The reagents in the pack and calibrators are ready for controls are at ambient temperature (20–25°C) before
use measurement.
Close the bottles of calibrator right after calibration and Due to the possible evaporation, specimens and
store at 2–8°C. calibrators on the analyzers should be measured
Do not pool reagents within a reagent kit or between within 2 hours.
reagent kits.
ASSAY PROCEDURE
Prior to loading the iFlash-Anti-Tg reagent pack on the
system for the first time, suspend the microparticles by Refer to the system operating instruction or the online
inverting the reagent pack slightly. help system for detailed information on preparing the
system.
For further information on reagent handling
precautions during system operation, refer to the The test-specific parameters stored in barcode on the
iFlash system operating instruction. reagent pack are read in. In case the barcode cannot
be read, enter the sequence numbers.
STORAGE AND STABILITY Carry out calibration, if necessary.
Storage: Place the calibrators CAL1, CAL2 and CAL3 in the
Store at 2–8°C in an upright position. calibrator rack in the sample zone. Only keep
calibrators open during calibration.
The kit may be used immediately after removal from
2-8°C storage. Test application.
Load samples (Use 20 μL of sample for each
Stability: determination in addition to the sample container and
Unopened at 2–8°C: up to the stated expiration date. system dead volumes).
Opened at 2–8°C: 28 days. Click RUN, the iFlash System performs all the
Store on-board: 28 days. functions automatically and calculates the results.
determinations at least once every 24 hours when the test The assay is unaffected by icterus (bilirubin < 10
is in use, once per reagent kit and after every calibration. mg/dL), hemolysis (Hb < 500 mg/dL), lipemia
Include commercially available quality control materials (Intralipid < 1,800 mg/dL) and total serum protein (< 10
that cover at least two levels of analyte. Follow g/dL).
manufacturer’s instructions for reconstitution and storage. No interference was observed from rheumatoid factors
Each laboratory should establish mean values and up to a concentration of 2,000 IU/mL.
acceptable ranges to assure proper performance. Quality
No interference was observed from anti-nuclear
control results that do not fall within acceptable ranges
antibodies up to a concentration of 500 AU/mL.
may indicate invalid test results.
No interference was observed from HAMA up to a
RESULT concentration of 600 ng/mL.
1st-4th Floor, No.5 Building, Lishan Industrial Pictograms for Hazardous to the
Area, Xinghai Road, Nanshan District, aquatic environment
Shenzhen 518054, P.R. China