0% found this document useful (0 votes)
59 views41 pages

Model 483C41 Eight-Channel ICP®/Voltage/Charge Sensor Signal Conditioner Installation and Operating Manual

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
59 views41 pages

Model 483C41 Eight-Channel ICP®/Voltage/Charge Sensor Signal Conditioner Installation and Operating Manual

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 41

Model 483C41

Eight-Channel ICP®/Voltage/Charge Sensor Signal Conditioner

Installation and Operating Manual

For assistance with the operation of this product,


contact PCB Piezotronics, Inc.

Toll-free: 800-828-8840
24-hour SensorLine: 716-684-0001
Fax: 716-684-0987
E-mail: [email protected]
Web: www.pcb.com
Service, Repair, and Return
Policies and Instructions

The information contained in this document supersedes all similar information that
may be found elsewhere in this manual.

also good practice to recalibrate after


Service – Due to the sophisticated exposure to any severe temperature
nature of the sensors and associated extreme, shock, load, or other
instrumentation provided by PCB environmental influence, or prior to any
Piezotronics, user servicing or repair is critical test.
not recommended and, if attempted,
may void the factory warranty. Routine PCB Piezotronics maintains an ISO-
maintenance, such as the cleaning of 9001 certified metrology laboratory and
electrical connectors, housings, and offers calibration services, which are
mounting surfaces with solutions and accredited by A2LA to ISO/IEC 17025,
techniques that will not harm the with full traceability to SI through
physical material of construction, is N.I.S.T. In addition to the normally
acceptable. Caution should be observed supplied calibration, special testing is
to ensure that liquids are not permitted also available, such as: sensitivity at
to migrate into devices that are not elevated or cryogenic temperatures,
hermetically sealed. Such devices phase response, extended high or low
should only be wiped with a dampened frequency response, extended range,
cloth and never submerged or have leak testing, hydrostatic pressure
liquids poured upon them. testing, and others. For information on
standard recalibration services or
Repair – In the event that equipment special testing, contact your local PCB
becomes damaged or ceases to Piezotronics distributor, sales
operate, arrangements should be made representative, or factory customer
to return the equipment to PCB service representative.
Piezotronics for repair. User servicing or
repair is not recommended and, if Returning Equipment – Following
attempted, may void the factory these procedures will ensure that your
warranty. returned materials are handled in the
most expedient manner. Before
Calibration – Routine calibration of returning any equipment to PCB
sensors and associated instrumentation Piezotronics, contact your local
is recommended as this helps build distributor, sales representative, or
confidence in measurement accuracy factory customer service representative
and acquired data. Equipment to obtain a Return Warranty, Service,
calibration cycles are typically Repair, and Return Policies and
established by the users own quality Instructions Materials Authorization
regimen. When in doubt about a (RMA) Number. This RMA number
calibration cycle, a good “rule of thumb” should be clearly marked on the outside
is to recalibrate on an annual basis. It is of all package(s) and on the packing
list(s) accompanying the shipment. A complete list of distributors and offices
detailed account of the nature of the can be found at www.pcb.com.
problem(s) being experienced with the Customers within the United States may
equipment should also be included contact their local sales representative
inside the package(s) containing any or a factory customer service
returned materials. representative. A complete list of sales
representatives can be found at
A Purchase Order, included with the www.pcb.com. Toll-free telephone
returned materials, will expedite the numbers for a factory customer service
turn-around of serviced equipment. It is representative, in the division
recommended to include authorization responsible for this product, can be
on the Purchase Order for PCB to found on the title page at the front of this
proceed with any repairs, as long as manual. Our ship to address and
they do not exceed 50% of the general contact numbers are:
replacement cost of the returned
item(s). PCB will provide a price PCB Piezotronics, Inc.
quotation or replacement 3425 Walden Ave.
recommendation for any item whose Depew, NY14043 USA
repair costs would exceed 50% of Toll-free: (800) 828-8840
replacement cost, or any item that is not 24-hour SensorLineSM: (716) 684-0001
economically feasible to repair. For Website: www.pcb.com
routine calibration services, the E-mail: [email protected]
Purchase Order should include
authorization to proceed and return at
current pricing, which can be obtained
from a factory customer service
representative.

Contact Information – International


customers should direct all inquiries to
their local distributor or sales office. A
PCB工业监视和测量设备 - 中国RoHS2公布表
PCB Industrial Monitoring and Measuring Equipment - China RoHS 2 Disclosure Table
有害物质
汞 镉
部件名称 铅 (Pb) (Hg) (Cd) 六价铬 (Cr(VI)) 多溴联苯 (PBB) 多溴二苯醚 (PBDE)
住房 O O O O O O
PCB板 X O O O O O
电气连接器 O O O O O O
压电晶体 X O O O O O
环氧 O O O O O O
铁氟龙 O O O O O O
电子 O O O O O O
厚膜基板 O O X O O O
电线 O O O O O O
电缆 X O O O O O
塑料 O O O O O O
焊接 X O O O O O
铜合金/黄铜 X O O O O O
本表格依据 SJ/T 11364 的规定编制。
O: 表示该有害物质在该部件所有均质材料中的含量均在 GB/T 26572 规定的限量要求以下。
X: 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB/T 26572 规定的限量要求。
铅是欧洲RoHS指令2011/65/ EU附件三和附件四目前由于允许的豁免。

CHINA RoHS COMPLIANCE


Component Name Hazardous Substances
Lead Mercury Cadmium Chromium VI Polybrominated Polybrominated
(Pb) (Hg) (Cd) Compounds Biphenyls Diphenyl
(Cr(VI)) (PBB) Ethers (PBDE)
Housing O O O O O O
PCB Board X O O O O O
Electrical O O O O O O
Connectors
Piezoelectric X O O O O O
Crystals
Epoxy O O O O O O
Teflon O O O O O O
Electronics O O O O O O
Thick Film O O X O O O
Substrate
Wires O O O O O O
Cables X O O O O O
Plastic O O O O O O
Solder X O O O O O
Copper Alloy/Brass X O O O O O
This table is prepared in accordance with the provisions of SJ/T 11364.
O: Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit
requirement of GB/T 26572.
X: Indicates that said hazardous substance contained in at least one of the homogeneous materials for this part is above
the limit requirement of GB/T 26572.
Lead is present due to allowed exemption in Annex III or Annex IV of the European RoHS Directive 2011/65/EU.

DOCUMENT NUMBER: 21354


DOCUMENT REVISION: D
ECN: 46162
SIGNAL CONDITIONER
Model 483C41

GENERAL OPERATION
MANUAL

For powering ICP®, and charge sensors, this signal conditioner provides an effective method for
managing large numbers of sensor channels. A simple command set, entered through industry
standard interface, allows the user to generate powerful application-specific programs to automate
system testing.

Manual Number: 46953


Manual Revision: B
ECO Number: 47035

PCB PIEZOTRONICS, INC. 3425 WALDEN AVENUE DEPEW, NY 14043-2495 PHONE 716-684-0001 FAX 716-684-0987
®
ICP is a registered trademark of PCB Group, Inc.
Table of Contents

Table of Contents i
Table of Figures iii
1-0. INTRODUCTION AND SPECIFICATIONS 1
1-1. Introduction: Safety Considerations 1
1-2. System Description 2
®
1-2.1 ICP Input/Output Mode 2
1-3. Block Diagram 2
1-4. Installation 2
1-4.1 Grounding Techniques 3
1-5. Operation: Standard AC Line 3
1-6. Maintenance and Repair 3
2-0. THEORY OF OPERATION 3
2-1. Sensor Excitation 3
2-2. Input Protection 4
2-3. ICP® Input Fault Detection 4
2-4. Input Interface 4
2-5. Gain 4
2-6. Normalized Output Sensitivity 5
2.7. Low Pass Input Filter 5
2.8. Overload Detection 5
2-9. Connector Configuration 6
2-10 Non-Volatile Memory 6
2-11 Transducer Electronic Data Sheet (TEDS) Interface: 6
3-0. FRONT PANEL MODULE CONTROL 6
3-1. Introduction 6
3-2. Unit Initialization 6
3-3. Main Menu Format 7
3-4. Submenu Format and Command Selections 7
1) Gain Functionality 7
2) Sensor Input Mode 9
3) Input Filter 9
4) Output Filter 9
5) ICP Current Adjustment 10
6) Voltage Excitation 10
7) Input Coupling 11
8) Auto zero/ Auto balance 11

i
MODEL 483C41 GENERAL OPERATION MANUAL ii

9) Auto Range 11
10) Switched Output 12
11) Clamp 12
12) Bias Measurement 12
13) Channel Output Measurement 13
14) Calibration 13
15) LED Test 14
16) Firmware Version 14
17) Unit ID 14
18) Reset to Factory Default Settings 15
19) Display Content Options 15
20) Save Current Settings 16
21) Reserved 16
4-0. COMPUTER INTERFACE PROGRAMMING GUIDE 16

4-1. Introduction 16
4-2. Ethernet Communication 16
4-3. Communication Guidelines 20
4-4. Unit Initialization Procedure 20
4-5. Command Summary 21
4-6. Command Format 21
4-7. Multiple Board Models 22
4-8. Commands 22
GAIN 22
SENS 22
FSCI 23
FSCO 23
INPT 24
IEXC 24
FLTR 24
OFLT 25
CALB 25
RTED 26
ALLC 26
RBIA 26
STUS 27
UNIT 27
UNID 28
LPCR 28
LEDS 28
RSET 29
SAVS 29
Communication Responses 29
MODEL 483C41 GENERAL OPERATION MANUAL iii

Table of Figures

Figure 1 Typical Block Diagram ..........................................................................2


®
Figure 2 ICP Sensor Excitation ..........................................................................3
Figure 3 Input Fault Window Comparator with LED Indicator .........................4
Figure 4 Input Amplifier Configuration ...............................................................4
Figure 5 Selectable Gain Amplifier Configuration .............................................4
Figure 6 Overload Window Comparator .............................................................6
MODEL 483C41 GENERAL OPERATION MANUAL 1

1-0. INTRODUCTION AND SPECIFICATIONS


1-1. Introduction: Safety Considerations
WARNING SYMBOLS AND TERMS

The following symbols and terms may be found on the equipment described in this manual.

This symbol on the unit indicates that the user should refer to the operating instructions located in the manual.

This symbol on the unit indicates that high voltage may be present. Use standard safety precautions to avoid personal
contact with this voltage.

This symbol indicates that the test fixture, Model 483C41, must be connected to earth ground via the power cord.

The WARNING heading used in this manual explains dangers that might result in personal injury or death. Always read
the associated information very carefully before performing the indicated procedure.

The Caution heading used in this manual explains hazards that could damage the instrument.

WARNING 1: The power supply/signal conditioner should not be opened by anyone other than qualified service
personnel. This product is intended for service by qualified personnel who recognize shock hazards and are familiar with
the safety precautions required to avoid injury.

WARNING 2: This equipment is designed with user safety in mind; however, the protection provided by the equipment
may be impaired if the equipment is used in a manner not specified by PCB Piezotronics, Inc.

Caution 1: Cables can kill your equipment. High voltage ElectroStatic Discharge (ESD) can damage electrical devices.
Similar to a capacitor, a cable can hold a charge caused by triboelectric transfer, such as that which occurs in the
following:

 Laying on and moving across a rug.


 Any movement through air.
 The action of rolling out a cable.
 Contact to a non-grounded person

The solution for product safety: 1) Connect the cables only with the AC power off. 2) Temporarily “short” the end of the
cable before attaching it to any signal input or output.

Caution 2: ESD considerations should be made prior to performing any internal adjustments on the equipment. Any
piece of electronic equipment is vulnerable to ESD when opened for adjustments. Internal adjustments should therefore
be done ONLY at an ESD-safe work area. Many products have ESD protection, but the level of protection may be
exceeded by extremely high voltage that is typically present in normal situations.

EQUIPMENT RATINGS
For complete specifications, please refer to the enclosed Specification Sheet. This equipment operates optimally at +32 to
+120°F (0 to +50°C), in an environment having <85% relative humidity.

This model requires 100 to 240 VAC / 47 to 63 Hz to operate. In turn, it supplies up to 20 mA of constant current
excitation at 24 VDC to any connected ICP sensors or ICP in-line charge converters.
MODEL 483C41 GENERAL OPERATION MANUAL 2

1-2. System Description


The 483C41 is an eight channel signal conditioner designed to interface measurement signals to readout or recording
devices. This model also serves to provide excitation power for ICP ® sensors and in-line charge converters. The
following are the features of the 483C41:

 Selectable input signal modes per channel including:


 ICP® with variable 2 to 20 mA source current
 Charge mode
 Voltage mode
 8th order low pass Butterworth input filter with cutoff frequencies of:
0.1, 0.3, 1.0, 3.0, 10, and 30 kHz (-10% point)
 Internal oscillator for 0.1 Vpk or 100 pCpk signal at 100 or 1000 Hz
 Incremental gain, x0.1 to x200 in 0.1 increments
 Input fault and overload LED
 Illuminator lights and a power button on the front panel
 Rear panel BNC jack input/output connectors
 Non-volatile memory
 Ethernet interface for computer control with PCB supplied software (part number EE75)
 Front panel display and keypad
 Normalized Output
 Capable of reading TEDS

1-2.1. ICP® Input/Output Mode


This model contains a regulated 24 VDC power supply that provides constant current for up to 8 individual channels.
Both the output and input connections utilize BNC jack connectors and are brought out through the rear panel.

1-3. Block Diagram

Figure 1 Typical Block Diagram

1-4. Installation
This signal conditioner model comes in a standard 19-inch rack mount enclosure with a 1U (1.75”) height. The enclosure
should be located in such a way as to allow convenient access to the power outlet for disconnect purposes. Since these
models have low power consumption, they can be located in confined environments.
MODEL 483C41 GENERAL OPERATION MANUAL 3

1-4.1. Grounding Techniques


Integrating this model into an application that links the outputs to other test equipment powered by line voltages may lead
to errors or loss of signal-to-noise ratio due to ground loops. The evidence of ground loops is easily seen whenever the
fundamental frequency (50 or 60 Hz) or a multiple of the fundamental frequency is present in the system when the
sensors are “at rest.” In order to maintain the operating specification of noise and reduce the effects of line interference,
proper grounding techniques should be used. The following procedure may be helpful:

1. Make sure the signal ground lines of all equipment are tied together. The signal grounds of the channels are typically
tied together via the case of the input and output BNC connectors. The individual channels of these models have their
signal ground line tied together internally at the power supply.

2. Insure that the sensor does not pick up line noise from the body under test. The case of the sensor should be isolated
from the body under (ground) using an isolation pad. The isolation pad breaks the loop formed by the signal path of the
sensor to the signal conditioner and the return (ground) from the signal conditioner back to the sensor.

3. Make sure that all equipment signal grounds of the test system are tied to the Earth ground at a single point and the
connections linking the equipment ground are made using a wire that can provide a very low impedance connection.

1-5. Operation: Standard AC Line


Plug the line cord into a 120V/60 Hz or 230V/50 Hz power source and toggle the “ON/OFF” switch.

CAUTION: Refer to the rear panel for proper input voltage and currents.

1-6. Maintenance and Repair


It is not recommended that the customer attempt repair of this model in the field. Should trouble occur, contact the
factory for assistance. If the unit becomes dusty and dirty, it may be wiped off with a soft cloth.

2-0. THEORY OF OPERATION


2-1. Sensor Excitation
ICP® refers to a low impedance voltage mode sensor with an integrated circuit and a piezoelectric sensing element in a
single housing that provides a voltage output. The sensor is powered by a +24 VDC power supply having a constant
current, variable from 2 to 20 mA.

Sensor excitation occurs as the constant current of all channels are set. 483C41 allows the constant current to be adjusted
up to 20 mA to provide the required excitation for most applications. Special situations, such as driving extra-long cables
(more than 1000 ft) with high frequency or fast rise time pulses, may require increasing the drive current to 12 mA or
higher.

Figure 2 ICP® Sensor Excitation


When driving fast rise time pulses over long lines, system performances can be optimized by “tuning” the drive current to
the line; i.e. by finding the best current setting for the particular test of physical parameters. To determine the optimal
current setting, experiment with your particular test set up. To insure optimal accuracy in constant-current adjustment,
make sure all channels have sensors or simulated loads, similar to the one provided by the model 401B04 sensor
simulator, consisting of a voltage amplifier similar to those found in ICP  sensors.
MODEL 483C41 GENERAL OPERATION MANUAL 4

2-2. Input Protection


The input section has protection to limit the amplitude of the incoming signal to within +24 volts to ground. Reference
spec sheet for maximum input signal.

2-3. ICP® Input Fault Detection


483C41 monitors two input fault conditions in ICP® mode, “short” and “open,” which indicate problems with sensor
input and is displayed through the front panel LEDs. Either case implies that the sensor is NOT functioning properly. An
input is shorted when it has a ground path for the sensor excitation and open when the sensor fails to draw the excitation.

Two voltage comparators consist of a window comparator that has two reference voltages (V ref) representing thresholds
for “short” and “open.” When the sensor’s bias voltage (Vbias) exceeds the comparator range, the front panel input fault
LED lights.

NOTE: Red LED implies input fault.

Figure 3 Input Fault Window Comparator with LED Indicator


2-4. Input Interface
The input signal conditioning for each channel in ICP® mode provides a unity gain buffer with high-input impedance
amplifiers. AC coupling eliminates the DC bias from the input signal unless DC coupling is enabled (483C41 is AC
coupled ONLY).

Figure 4 Input Amplifier Configuration

Voltage mode is similar to ICP® mode, except the excitation current is 0 mA. In charge mode, the signal from the charge
sensor is routed into a charge amplifier with a gain of 0.1, 1.0, or 10 mV/pC, followed by incremental voltage gain.

2-5. Gain
The Programmable Gain Amplification (PGA) block consists of a decimal gain amplifier. See Figure 5.

Figure 5 Selectable Gain Amplifier Configuration


MODEL 483C41 GENERAL OPERATION MANUAL 5

2-6. Normalized Output Sensitivity


The definition of normalized output sensitivity is the calculation of the individual channel’s gain, depending on the
sensitivity of the sensor attached and the desired output sensitivity. This is simply a function of the desired output
sensitivity (in Volts/ unit) divided by the sensor’s sensitivity.

1 Volt
Gain = / Sensitivity (mV/unit)
Additionally, a feature to improve flexibility allows changing the output level to a value of 5 Volts/ unit instead of
normalized 1 Volt/ unit. Some users request the ability calculate the gain using a known input signal along with the sensor
sensitivity, and desired output level. The result of the request impacts the formula:

1 Volt/
Gain = / [Sensitivity (mV/unit) * Full Scale Input (units)]

Adding some simple error checking to insure the limits of the sensor and gain limits of the signal conditioner are not
exceeded completes the normalization process.

Gainnormalized = FSOT (V) / SENS * FSIN


Additional Considerations:
The storage of individual channel gains is stored in non-volatile memory locations. The new variables for each channel’s
sensitivity, Full Scale Output level, and Full Scale Input are stored in non-volatile memory locations when SAVS
command is sent to the unit.

The error checking should provide a flag if the desired normalized output level is not feasible due to gain limitations. The
gain required may be too large given the sensor sensitivity defined, or too small which implies the sensor will not be
capable measure the expected value. The typical sensor will output a signal up to 5 Volts. The maximum swing may be
used in the error checking.

The 483C41 has the capability to provide signal conditioning for charge mode sensors in addition to the standard ICP
type. The sensitivity (mV/ pC) of the charge conversion will allow charge mode sensors (pC/ unit) to be normalized.

2.7. Low Pass Input Filter


The 483C41’s 8 pole low pass input filter is useful for attenuating signals above the user’s band of interest, and for anti-
aliasing of signals before digitization. High quality components are used to achieve 6 cutoff frequencies: 0.1, 0.3, 1.0,
3.0, 10.0, and 30.0 kHz (-10%). The Butterworth topology provides a maximally flat response in the passband and is
critically damped. Continuous-time stages are employed so that the clock noise associated with switched capacitor
elements is not an issue. This is especially significant because the filter is located prior to the gain stages.

2.8. Overload Detection


The overload feature uses the same window comparator principle previously discussed. The +V ref is equal to the default
overload value of ±10 volts. When the input voltage (Vsignal of Figure 6) to the window comparator exceeds the
reference voltage limits, overload has occurred, and the comparator’s output, which is normally “high,” becomes “low.”
This “low” state illuminates the overload LED and triggers the latch of overload detection circuitry. During regular
measuring time, the latch holds the occurrence of overloads until the user reads its status through the computer interface.
MODEL 483C41 GENERAL OPERATION MANUAL 6

Figure 6 /Overload Window Comparator

2-9. Connector Configuration


This model has eight BNC jack output connectors, located on the rear panel, which are used regardless of the input mode
being used. The rear panel also includes eight BNC jack input connectors that are used for all input modes.

2-10. Non-Volatile Memory


This feature provides a mechanism to save and restore a programmed configuration. The programmable settings are
saved when the unit receives a SAVS command and restored when the unit powers up. The unit’s non-volatile memory
may be reset to the factory default settings by using the RSET Command.

2-11. Transducer Electronic Data Sheet (TEDS) Interface:


TEDS sensors attach to the input connectors the same way a traditional ICP sensor is attached to the signal conditioner.
The TEDS interface requires the Ethernet computer interface, and the input mode option. The digital control and input
mode features automatically perform the TEDS read or write function. After the communication with the TEDS sensor is
complete the input mode returns to the input mode (charge or ICP) prior to the communication.

The current TEDS function will perform a read of the “raw” TEDS data and return the 32 bytes of TEDS sensor
EEPROM data and the 8 byte Application register contents, if it was programmed, and returns the data in an ASCII Hex
representation.

3-0. FRONT PANEL MODULE CONTROL


3-1. Introduction
483C41 is equipped with the capability of controlling functions through the front panel module. This module consists of
a display and five (5) push buttons. The display shows various menus in which all of the available options are given. The
user may scroll through the menu, and choose any option to change the settings.

3-2. Unit Initialization


Upon pressing the ON/OFF button, the unit will turn on and display the following:

PCB 483C41
FW Ver n.n
The unit is initialized with the factory default settings.
The factory defaults are:
 Gain = 1.0
 Sensitivity = 10.0
 Full Scale input = 1000.0
 Full Scale output = 10.0
 Input Mode = ICP®
 Input Filter = Disabled
MODEL 483C41 GENERAL OPERATION MANUAL 7

 Current excitation = 4.0 mA


 Output Filter = Disabled
 Coupling = AC
 Clamp = OFF
 Gain Type = Gain

3-3. Main Menu Format


Initially, pressing any button will cause the first two menu lines to be displayed and the cursor will blink on the ‘Gain’
option.

GAIN SENS INPT


FSOT FSIN FLTi
The cursor may be moved around the main menu by using the arrow keys. To show the other selections of the main
menu, press the down  or up ↑ arrow key. This action scrolls vertically to the subsequent lines on the menu. These lines
are shown below.

EXCi EXCv FLTo


ACDC ZERO ARNG
SWOT CLMP Bias
CHRD Cal LEDS
Ver? Unit Reset
Disp Savs Res

3-4. Submenu Format and Command Selections


1) Gain Functionality – Includes five menu selections
The following menu options all pertain to how the gain functionality operates in the 483C41.

Gain Type (GType on front panel): Allows the user to choose how the gain of each channel can be set. The selections
in the submenu are Gain and Normalize (Gain is the factory default setting). Pressing ENTER brings up the submenu
shown below. The up  and down  arrow key can be used toggle between the two options.

Gain Set Option


Gain
Both options return control to the main menu after a selection is made. If Gain is selected then only the GAIN menu item
can be used to set the gain of a channel. If Normalize is selected then the GAIN menu item is disabled and SENS, FSOT
and FSIN will be used to set the gain using this equation: Gain = (FSOT * 1000) / (FSIN * SENS).
MODEL 483C41 GENERAL OPERATION MANUAL 8

If the user attempts to use a menu selection (SENS, FSOT, FSIN, GAIN) that conflicts with the GType setting the
message ‘Invalid Item for Gain Entry Type’ will be displayed.

Gain (GAIN on front panel): Directly sets the gain value of each channel. For ICP®/voltage input modes: .1 to 200.0.
For charge mode: .01 to 2000.0. GType must be set to the Gain selection. See section 1-4.2c for instructions on setting
the gain directly. The gain step is .1.

NOTE: A microprocessor inside the conditioner transparently sets the charge and voltage gains when in charge mode.
This is done so sub-optimum values cannot be selected (ie – setting a low charge gain with a high voltage gain). Highest
resolution is achieved if the charge gain is set as high as possible.

Sensitivity (SENS on front panel): The sensor sensitivity is entered here using the  and  arrow keys and then
hitting Enter. GType must be set to the Normalize selection.

Full Scale Input (FSIN on front panel): The estimated full scale mechanical input (ex – engineering units, g’s, psi, bar,
etc.) is entered here using the  and  arrow keys and then hitting Enter. GType must be set to the Normalize
selection.

Full Scale Output (FSOT on front panel): The desired full scale output voltage is entered here using the  and 
arrow keys and then hitting Enter. GType must be set to the Normalize selection.

The submenus for SENS, FSIN and FSOT appear like this:

Chan Sensitivity Chan FS In Chan FSOT


2 00100.000 2 01000.000 2 0010.0
The 483C41 has the ability to automatically normalize the output sensitivity of each channel based on the sensitivity of
the sensor (SENS), the full scale value of the input (FSIN) in engineering units (g’s for example) and the full scale output
(FSOT) of the amplifier expressed in volts (this is the output that a signal at an FSIN level will produce). As stated
previously, the gain equation is:

Gain = (FSOT * 1000) / (FSIN * SENS)

Normalized Output Example: A full scale output of 5 volts is desired and a full scale input of 380 engineering units is
expected. The sensor sensitivity is 9.96 mV/g. These values should be entered into FSOT, FSIN and SENS respectively.
The gain will be automatically set to 1.3 to give a 5 volt full scale output based on the expected input and sensor
sensitivity.

Gain = (5 * 1000) / (380 * 9.96) = 1.3

Normalization is useful when standardizing the output of channels with sensors of different sensitivities. Consider the
following example where 1V/unit standardization on each channel is desired:

Channel Sensor Sensitivity Gain Setting Actual Gain Needed


mV
1 10.10 /unit 99.00 99.01
mV
2 101.32 /unit 9.9 9.869
mV
3 22.30 /unit 44.8 44.84

NOTE: The GAIN menu selection will not show what the gain is set at when in the Normalize GType setting.
MODEL 483C41 GENERAL OPERATION MANUAL 9

2) Sensor Input Mode (INPT on front panel):


The 483C41 allows the user to switch between ICP, Voltage, Charge modes. The factory default for all channels is
ICP. To change the mode select the INPT option from the main menu using the left / right  arrow keys until the
cursor is over the INPT option, and then press the Enter button.

Gain SENS INPT


FSOT FSOT FLTi
The following submenu will appear. Choose the channel you wish to change by using the up  and down  arrow keys.
Next, use the left / right  arrow key to move the flashing cursor over to the mode column. To change the INPT, use
either the up  or down  arrow key to scroll through the options, once you have the desired setting selected hit ENTER
to apply it.

Chan INPUT TYPE


2 ICP
NOTE: Setting a channel to Voltage or Charge Input will cause the ICP© current source for that channel to be turned off.
Conversely, If ICP© is set from Voltage or Charge then the ICP© current source will be set to 4mA.

3) Input Filter (FLTi on front panel):


Each channel’s Input Filter setting can be set independently of the other channels. To change the Input Filter setting from
the main menu, use the left / right  arrow key until the cursor is over the FLTi option, and then press the Enter
button.

GAIN SENS INPT


FSOT FSIN FLTi
The following submenu will appear. Choose the channel you wish to change by using the up  and down  arrow keys.
Next, use the left / right  arrow key to move the flashing cursor over to the INP Filter column. To change the Filter
setting, use either the up  or down  arrow key to select from the following options; 0.100 kHz (100 Hz), 0.300 kHz
(300 Hz), 1.000 kHz, 3.000 kHz, 10.000 kHz, 30.000 kHz, OFF, once you have the desired setting selected hit ENTER
to apply it.

Chan INP Filter


2 ON-0.100 k

4) Output Filter (FLTo on front panel):


A screen stating ‘This Option is Not Installed’ will appear. The 483C41 does not come standard with an output filter, but
the option is available. Contact the factory for more information.

Each channel’s Output filter setting can be set independently of the other channels. To change the Output Filter setting
from the main menu, use the left / right  arrow key until the cursor is over the FLTo option, and then press the Enter
button.
MODEL 483C41 GENERAL OPERATION MANUAL 10

EXCi EXCv FLTo


ACDC ZERO ARNG
The following submenu will appear. Choose the channel you wish to change by using the up  and down  arrow keys.
Next, use the left / right  arrow key to move the flashing cursor over to the Out Filter column. To change the Filter
setting, use either the up  or down  arrow key to select either ‘On-nn.nk’ or ‘Off’, once you have the desired setting
selected hit ENTER to apply it.

NOTE: The ON setting also indicates the filter corner, in kHz, implemented by the filter for the channel.

Chan Out Filter


2 ON-10.0k
5) ICP Current Adjustment (EXCi on front panel):
The Current excitation value can be set from 0 – 20 mA. Where 0 = Off. To set the current excitation value, move the
cursor to the ‘EXCi’ location, as shown below.

EXCi EXCv FLTo


ACDC ZERO ARNG
Pressing the ENTER button, will cause the following submenu to appear. The ICP current may be altered by using the
up  or down  arrow keys.

Chan ICP Current


4 02 mA
The 483C41 will turn off the current excitation signal if the unit is not in ICP© input mode and will not allow it to be set
unless it is in ICP© mode. If a channel is selected from the Iexc menu that is inappropriate then NA will appear as shown
below in place of the ICP© current settings. Also, because of this limitation, the ‘ALL’ channel selection is not allowed
for this setting.

Chan ICP Current


2 NA
NOTE: The factory default setting for the ICP current is 4 mA.

6) Voltage Excitation (EXCv on front panel):


The Voltage excitation option is not applicable to the 483C41 units.
MODEL 483C41 GENERAL OPERATION MANUAL 11

EXCi EXCv FLTo


ACDC ZERO ARNG
The following message will show if it is activated.

This Option is
Not Installed
7) Input Coupling (ACDC on front panel):
The input coupling option is not applicable to the 483C41 units. 483C41 is AC coupled only.

EXCi EXCv FLTo


ACDC ZERO ARNG
The following message will show if it is activated.

This Option is
Not Installed
8) Auto zero/ Auto balance (ZERO on front panel):
The Auto Zero/Auto Balance option is not applicable to the 483C41 units.

EXCi EXCv FLTo


ACDC ZERO ARNG
The following message will show if it is activated.

This Option is
Not Installed
9) Auto Range (ARNG on front panel):
The auto-range option is not applicable to the 483C41 units.
MODEL 483C41 GENERAL OPERATION MANUAL 12

EXCi EXCv FLTo


ACDC ZERO ARNG
The following message will show if it is activated.

This Option is
Not Installed
10) Switched Output (SWOT on front panel):
The Switched Output option is not applicable to the 483C41 units.

SWOT CLMP Bias


CHRD Cal LEDS
The following message will show if it is activated.

This Option is
Not Installed
11) Clamp (CLMP on front panel):
The Clamp option is not applicable to the 483C41 units.

SWOT CLMP Bias


CHRD Cal LEDS
The following message will show if it is activated.

This Option is
Not Installed

12) Bias Measurement (Bias on front panel):


The 483C41 is capable of measuring the bias voltage present on each of its channels. By moving the cursor upon the
“Bias” location, the display appears as follows:
MODEL 483C41 GENERAL OPERATION MANUAL 13

SWOT CLMP Bias


CHRD Cal LEDS
Using the ENTER button, the submenu appears as shown in the following diagram. Choose the channel you wish to
change by using the up  and down  arrow keys. The bias reading for the channel will be displayed automatically when
the channel is selected.

Chan Bias
2 0023.7
If unit had a short circuit, the bias voltage would be <2.0V. Proper bias voltage is between 2.0V and 22V. Anything over
22V would indicate an open circuit.

13) Channel Output Measurement (CHRD on front panel):


The Channel Output Measurement option is not applicable to the 483C41 units.

SWOT CLMP Bias


CHRD CAL LEDS
The following message will show if it is activated.

This Option is
Not Installed
14) Calibration (CAL on front panel):
The user may select the Internal Calibration setting by pressing the ENTER key on the front panel when the cursor is
located on the CAL option. When either the 100 or 1k Hz setting is selected the channel is forced into a charge mode, if it
isn’t set that way already, and an internal 100pC signal is generated. This signal is then converted to a voltage based on
the selected charge sensitivity (1mV/pC if the ICP setting was selected) after which the channels gain is applied to
provide an internal cal signal.

CHRD CAL LEDS


Ver? Unit Reset
The following submenu will appear. Choose the channel you wish to change by using the up  and down  arrow keys.
Next, use the left / right  arrow key to move the flashing cursor over to the Cal Mode column. To change the Cal
Mode setting, use either the up  or down  arrow key to select either ‘Off’, ‘100Hz’, ‘or ‘1kHz’. Once you have the
desired setting selected select ENTER to apply it.
MODEL 483C41 GENERAL OPERATION MANUAL 14

Chan Cal Mode


1 100Hz
15) LED Test (LEDS on front panel):
The testing of the front panel LEDs allows checking whether or not they are functioning properly. To test them, place the
cursor over the word “LEDS” on the display and press enter.

CHRD Cal LEDS


Ver? Unit Reset
The screen shown below will be displayed.

LED Test
CANCEL
Use the up or down arrow keys to select ‘Execute’ or ‘Cancel’. Both options return control to the main menu. If ‘Execute’
is selected then the LED’s will blink 3 times if ‘Cancel’ is selected then no action is taken.

16) Firmware Version (Ver? on front panel):


To determine the firmware version of the unit select the ‘Ver?’ menu option and hit enter.

CHRD Cal LEDS


Ver? Unit Reset
A screen similar to the one shown below will be displayed

PCB 483C41
FW Ver 1.6
17) Unit ID (Unit on front panel):
The user may change the Unit ID setting by pressing the ENTER key on the front panel when the cursor is located on the
Unit option.

CHRD Cal LEDS


Ver? Unit Reset
MODEL 483C41 GENERAL OPERATION MANUAL 15

The following submenu will appear. Use the up  or down  arrow key to change the Unit Id setting to a new value. Once
you have the desired setting selected hit ENTER to apply it.

Current Unit #
1
NOTE: Changing the Unit ID is only necessary if you have the Ethernet or RS-232 communication option installed and
you are communicating with more than one unit. If this is the case then all units in the network MUST have a unique ID.

18) Reset to Factory Default Settings (Reset on front panel):


The reset option provides a mechanism to reset the unit to its factory default settings. To do this select the reset option
from the menu by placing the cursor over the Reset option and hitting enter.

CHRD Cal LEDS


Ver? Unit Reset
Pressing the ENTER brings up a submenu that allows the operator to continue by selecting ‘Execute’ or cancel the
function by selecting ‘Cancel’. Both options return control to the main menu. If ‘Execute’ is selected then the defaults are
restored.

Factory Defaults
Execute
19) Display Content Options (Disp on front panel):
The Display option allows the user to Lock or Unlock the display. To enter this mode, select the ‘DISP’ option from the
menu by placing the cursor over the display option and hitting enter.

Ver? Unit Reset


Disp Res Res
Pressing the ENTER brings up a submenu with two options: Lock or Unlock.

Display Opts

Lock

If Lock is selected the any attempt to change another setting will cause the message “Unlock Display to Proceed” to be
displayed. The user must then reset the display option to UNLOCK to change any other setting.
MODEL 483C41 GENERAL OPERATION MANUAL 16

20) Save Current Settings (Savs on front panel):


When activated the save settings command forces the units current settings to become the default power up settings.
These new settings will be implemented the next time the unit is powered on.

Ver? Unit Reset


Disp Savs Res
Pressing the ENTER brings up a submenu that allows the operator to continue by selecting ‘Execute’ or cancel the
function by selecting ‘Cancel’. Both options return control to the main menu. If ‘Execute’ is selected then the defaults are
set.

Save Settings
Execute
21) Reserved (Res on front panel):
No operation. These menu items are reserved for future use.

Ver? Unit Reset


Disp Savs Res

4-0. COMPUTER INTERFACE PROGRAMMING GUIDE


4-1. Introduction
The Ethernet Interface enables the 483C41 to be fully controlled by a computer or laptop controller. With this interface,
the unit is able to become part of a fully automated system.

4-2. Ethernet Communication


483C41 is equipped with an Ethernet Interface so the unit’s IP address must be set up before any remote communication
can commence.

The steps to set the IP address are as follow:


 Connect, either directly with a crossover CAT5 cable or through a hub, to a single unit. This unit is presumed to
have a Unit ID of 1,
 Set the IP address using an independent utility from Lantronix called DeviceInstallerTM. This utility is provided
on the PCB MCSC application software installer CD and also available for download at the Lantronix website:
http://www.lantronix.com/device-networking/utilities-tools/device-installer.html
Once installed it can be accessed from the MCSC software from the ‘Help | Configure Ethernet Devices’ menu
item.
 After the DeviceInstallerTM application is installed, run it and the following screen will appear. Click on the
Search icon and the program will search for the Ethernet device internal to the signal conditioner unit. When
found, as shown below, details about the device show in the list. You can verify it is the correct unit by
comparing the Hardware (MAC) Address displayed to the one listed on the side of the signal conditioner unit.
MODEL 483C41 GENERAL OPERATION MANUAL 17

 Selecting an item from the list, shown below, highlights the item in the list and enables more icons.

 Select Assign IP and the following screen will appear. Select whether you want to assign a static IP address that
is appropriate for your network or have it assigned from a network server. Subsequent steps are self-explanatory.

 Now you can use the assigned IP address to address the unit.
MODEL 483C41 GENERAL OPERATION MANUAL 18

Important Note: The communication protocol requires a unit id as part of the command header. The unit id is
not the IP address. To send commands to the unit Via Ethernet you must address the TCP-IP packets with the
proper IP address and ensure the packet payload contains the correct Unit Id in the command header.

The XPort device should be configured properly for communication with the signal conditioner. If problems
occur please check the communication parameters of the device as shown below.

Navigate to the actual device as shown in the left panel above and then click on the ‘Web Configuration’ tab and clock on
the ‘Go’ button. This will cause the following screen to appear, just select ‘OK’. There are no login parameters defined.

Following this screen the right panel will appear as shown below. Click on the ‘Connection’ item and check to see that
the parameters are shown as below and pay particular attention to the ‘Active Connection’ and ‘Endpoint’ Port items.
They should be set as shown below.

Next, check the Serial Settings. They should be 19200 baud, 8 data bits, no parity, 1 stop bit. And no flow control.

If any setting is changed click ‘Ok’ on the page and then activate the ‘Apply Settings’ item on the left side of the right
panel.
MODEL 483C41 GENERAL OPERATION MANUAL 19
MODEL 483C41 GENERAL OPERATION MANUAL 20

4-3. Communication Guidelines


1) Data transfer from the host terminal to the unit must contain an ending delimiter of <CR><LF>.

Example: <CR><LF> -Carriage Return and Line feed. (In ASCII,


<CR> is 13; <LF> is 10.)

2) The number of characters for any command string, from the first character to the <CR>, may not exceed 255.
3) Status request commands, i.e., commands ending with a “?”, may only be sent within one transmission.
Example: 7:0:GAIN?<CR> <LF>
-Reads the gain setting of all channels.

4-4. Unit Initialization Procedure


1) To begin, connect the signal conditioner to the network using the RJ-45 connector on the back of the unit.
2) Turn on power to the signal conditioner. The power indicator should light up to indicate the power is on.
3) With the terminal or computer that is set up to transmit data, send the command:

“1:0:LEDS=0” <CR><LF>

This command flashes the front panel LEDs three times.


MODEL 483C41 GENERAL OPERATION MANUAL 21

4-5. Command Summary


The table below is a summary of the command set. Some commands may not be valid in all units. The commands are
sent and received from/to the host computer in ASCII text format.

Command Type Scope Meaning


GAIN R/W Channel Set or read gain
SENS R/W Channel Sensor Sensitivity
FSCI R/W Channel Expected Full Scale Input Value
FSCO R/W Channel Full Scale Output
INPT R/W Channel Input Signal Mode Selection
IEXC R/W Channel Read/Adjust ICP current setting
FLTR R/W Channel Disable or Select Input Filter Corner
OFLT R/W Channel Enables/Disables Output Filter (optional)
CALB R/W Channel Selects the Calibration Mode
RTED R Channel Read TEDS data (DS2430A)
ALLC R Channel Read majority of channel settings
RBIA R Unit Measure Bias (returns all channels)
STUS R Unit input fault / overload status, for all channels
UNIT R Unit Read Unit Configuration information
UNID R/W Unit Set Unit ID
LPCR R Channel Returns the programmable input filter corners
LEDS FCN Unit Front panel LED test function
RSET FCN Unit Restores factory default channel settings
SAVS FCN Unit Saves the current settings to NVRAM
Command type definitions:

 R/W – the setting can be read from or written to the unit or channel.
 R – The information can only be read from the unit or channel.
 FCN – The command invokes a function in the unit.

4-6. Command Format


The communication protocol incorporates the concept of 'Directed' and 'Global' commands at both the Unit and Channel
level with the following characteristics;
 Unit or Channel numbers =0 are global commands that affect either all units or all channels of a particular unit
or both.
 Directed commands that set a unit parameter are always acknowledged (ACK) with an ASCII message that
indicates '<Unit#>:<CMD>:ok’ if implemented with no errors or NAK with; <Unit#>:<CMD>:=<error#>’ if an
error was encountered.
 Directed commands that request a particular parameters setting (query) result in a query response being returned
 No response is ever given to a Global Unit command.

Command Format:
'Unit#:Ch#:Cmd[=|?]{<value1 >{,< value2 >}}{;Ch#:Cmd[=|?]{<value1>{,< value2>}}}\n\r

 Each message must be preceded by a Unit# & Channel# (both of which could be 0)
 Messages may contain multiple commands separated by a semicolon ‘;’.
 The second and subsequent commands in a message shall not contain a unit number but shall contain a channel
number.
 Each command in a message will evoke a response message if one is warranted (not global)
 Query’s ('?') can only be directed to one unit but if the channel=0 then each channel's setting will be returned in
the order 0-MAXCHANNELS separated by a ':'
MODEL 483C41 GENERAL OPERATION MANUAL 22

Command examples:
1:0:GAIN=100.2\r\n unit 1,all channels gain set to 100.2
1:1:GAIN=100.2;2:GAIN=120.3\r\n unit 1, channel 1 gain = 100.2; channel 2 gain = 120.3
1:3:GAIN=100.2;0:FLTR=1\r\n unit 1, channel 3 gain = 100.2; all channel’s filter = ON

General Query Response Format


Unit#:Cmd:Ch#=<value>{; Ch#=<value>}…>{; Ch#=<value>}…\r\n

Responses to a query with a channel number=0 will return the setting of each channel in a list separated by
semicolons ‘;’.
If the target of the query is a unit setting (ex. current excitation) then the channel number returned is the 1 st
channel of the board that processed the command.

4-7. Multiple Board Models


The 483C41 has multiple main boards which appear as a contiguous set of channels; they are physically separate and
operate somewhat independently. A global channel command to set 8 channels on a unit will be handled by each internal
board simultaneously. However only the board with the channels designated as 1-4 will ACK the command, the other
board will remain silent.

Likewise, a global Query command will be responded to by the channel 1-4 board but will be ignored by the channel 5-8
board. However, in order to facilitate efficient communications with the second board, a second unit address was is used
to query the second board. A user’s control application can direct global channel setting queries to the second board of a
unit by addressing the query to the secondary unit address of the unit. The secondary unit address is defined to be the
normal unit address plus 128 (ex. If a unit’s address is 1 then the 2nd board’s address would be 129)

4-8. Commands
GAIN
SET GAIN: This command sets the programmable gain of a channel.

Setting:
The amplifier gain can be set directly by sending a Gain command:

1:0:GAIN=100.2\r\n (unit 1,all channels gain set to 100.2)

When a channels gain is set directly the unit will adjust the FSI parameter of the gain equation using the
following equation; FSI = (((FSO*1000)/Gain)/Sens) to ensure it remains valid.

Setting Response: 1:GAIN:ok

Query:
The Gain query returns all of the parameters used to determine it in a single response
Query Format: Unit#:Ch#:CMD?
Response format: Unit#:Cmd:Ch#:=Gain:SENS:FSO:FSI;
Query: 1:5:GAIN?
Response: 1:GAIN:5= 5.0: 10.0: 10.0: 200.0;

Global Gain Query: 1:0:GAIN?


Global Gain Response: 1:GAIN:1= 5.0: 10.0: 10.0: 200.0;2= 5.0: 10.0: 10.0: 200.0;3= 5.0:
10.0: 10.0: 200.0;4= 5.0: 10.0: 10.0: 200.0;

SENS
The SENS command provides a mechanism to have the transducer Sensitivity influence the Gain setting of the
channel. Channel Gain is calculated using the equation;
MODEL 483C41 GENERAL OPERATION MANUAL 23

Gain = FSO*1000/(FSI*SENS). If a Sensitivity is entered that caused the gain to exceed the amplifiers
capability the FSI component will be adjusted to keep the equation valid.

Setting:
1:0:SENS=20.2\r\n (unit 1,all channels transducer sensitivity set to 20.2)
Setting Response: 1:SENS:ok

Query:
The SENS query returns the channels transducer sensitivity
Query Format: Unit#:Ch#:CMD?
Response format: Unit#:Cmd:Ch#:= SENS;
Query: 1:1:SENS?
Response: 1:SENS:1= 6.0;

Global Gain Query: 1:0:SENS?


Global Gain Response: 1:SENS:1= 6.0;2= 10.0;3= 10.0;4= 10.0;

FSCI
The FSCI command provides a scaling mechanism to automatically set the gain based on a known input level (in
EU) and what output level (in Volts) you would like that Full Scale input level to be represented by. For instance
1000g’s = 10Volts. These 2 values along with the transducer sensitivity set the gain. Channel Gain is calculated
using the equation: Gain = FSO*1000/(FSI*SENS).
Setting:
1:1:FSCI=1000.000\r\n (unit 1, channel 1 FSI set to 1000.0)
Setting Response: 1:FSCI:ok

Query:
The FSCI query returns the channels Full Scale Input value in engineering units
Query Format: Unit#:Ch#:CMD?
Response format: Unit#:Cmd:Ch#:= <fsci>;
Query: 1:1: FSCI?
Response: 1:FSCI:1=1000.0;

Global FSCI Query: 1:0:FSCI?


Global FSCI Response: 1:FSCI:1=1000.0;2=1000.0;3=1000.0;4=1000.0;

FSCO
The FSCO command provides a scaling mechanism to automatically set the gain based on a known input level
(in EU) and what output level (in Volts) you would like that Full Scale input level to be represented by. For
instance 1000g’s = 10Volts. These 2 values along with the transducer sensitivity set the gain. Channel Gain is
calculated using the equation:

Gain = FSO*1000/(FSI*SENS).

Setting:
1:1:FSCO=10.000\r\n (unit 1, channel 1 FSI set to 10.0)
Setting Response: 1:FSCO:ok

Query:
The FSCO query returns the channels Full Scale Output Value in volts
Query Format: Unit#:Ch#:CMD?
Response format: Unit#:Cmd:Ch#:= <fsci>;
Query: 1:1: FSCO?
Response: 1:FSCO:1=10.0;
MODEL 483C41 GENERAL OPERATION MANUAL 24

Global FSCO Query: 1:0:FSCO?


Global FSCO Response: 1:FSCO:1=10.0;2=10.0;3=10.0;4=10.0;

INPT
The INPT command sets the input mode for a given channel. The mode selection is sent as an integer value. The
possible input modes and their associated values are:

 CHARGE 0
 VOLTAGE 1
 ICP® 2

Setting:
1:1:INPT= 2\r\n (unit 1, channel 1 input mode set to ICP)
Setting Response: 1:INPT:ok

Query:
The INPT query returns the channels input mode selection
Query Format: Unit#:Ch#:CMD?
Response format: Unit#:Cmd:Ch#:= <INPT>;
Query: 1:1:INPT?
Response: 1:INPT:1= 2;

Global INPT Query: 1:0:INPT?


Global INPT Response: 1:INPT:1= 2.0;2= 2.0;3= 2.0;4= 2.0;

IEXC
The IEXC command sets the current excitation level for a channel in ICP mode. The current excitation value is
sent as an integer value from 0 (off) to 20mA. If a channel’s input mode changes to Voltage mode then this
value will be forced to 0 automatically.

Setting:
1:1:IEXC= 2\r\n (unit 1, channel 1 is set to 2mA)
Setting Response: 1:IEXC:ok

Query:
The IEXC query returns the units excitation value.
Query Format: Unit#:Ch#:CMD?
Response format: Unit#:Cmd:Ch#:= <IEXC>;
Query: 1:1:IEXC?
Response: 1:IEXC:1=2;\r\n

Global IEXC Query: 1:0: IEXC?


Global IEXC Response: 1:IEXC:1=2;2=4;3=4;4=4;

FLTR
The FLTR command selects the Input Filter corner (1-6) or Disables the Filter (0).
The standard factory installed corner selections are:

 0= OFF
 1=30 kHz
 2=10 kHz
 3=3 kHz
 4=1 kHz
 5=300 Hz
MODEL 483C41 GENERAL OPERATION MANUAL 25

 6=100 Hz

Setting:
1:1:FLTR= 1\r\n ( unit 1, channel 1, Input Filter Enabled; Corner=30 kHz)
Setting Response: 2: FLTR:ok

Query:
The FLTR query returns the channels Output Filter.
Query Format: Unit#:Ch#:CMD?
Response format: Unit#:Cmd:Ch#:= <0-6>;
Query 1:1:FLTR?;
Response:: 1:FLTR:1=1;\r\n

Global FLTR Query: 1:0:FLTR?


Global FLTR Response: 1:FLTR:1=1;2=0;3=0;4=0;\r\n

OFLT
The OFLT command enables or disables the Output Filter.
The Output Filter value is sent as an integer value of either 0 -Disable or 1-Enable.

Setting:
2:1:OFLT= 1\r\n ( unit 2, channel 1, Output Filter Enabled )
Setting Response: 2: OFLT:ok

Query:
The FLTR query returns the channels Output Filter.
Query Format: Unit#:Ch#:CMD?
Response format: Unit#:Cmd:Ch#:= <0|1>;
Query: 2:1: OFLT?
Response: 2: OFLT:1=1;

Global OFLT Query: 1:0: OFLT?


Global OFLT Response: 1: OFLT:1=1;2=0;3=0;4=0;

CALB
The CALB command selects the Internal Calibration Reference Oscillator setting. When either the 100 or 1k Hz
setting is selected the channel is forced into a charge mode, if necessary, and an internal 100pC signal is
generated. This signal is then converted to a voltage after which the channels gain is applied to provide an
internal cal signal.

The CALB value is sent as an integer value of either 0 –Disable, 1-1000 Hz Enable, 2-100 Hz Enable

Setting:
1:1: CALB= 1\r\n (unit 1, channel 1, Oscillator=1 kHz )
Setting Response: 1:CALB:ok

Query:
The CALB query returns the channels Internal Reference Oscillator setting.
Query Format: Unit#:Ch#:CMD?
Response format:
Unit#:Cmd:Ch#:= <0|1>;
Query: 1:1: CALB?
Response: 1:CALB:1=1;\r\n

Global CALB Query: 1:0:CALB?


MODEL 483C41 GENERAL OPERATION MANUAL 26

Global CALB Response: 1:CALB:1=1;2=0;3=0;4=0;\r\n

RTED
The RTED query returns the TEDS information that is stored in the sensor or other, TEDS capable, in-line
module attached to a specific channel. The signal conditioners are IEEE 1451.4 compliant in that they will read
the DS2430A Application Register Status to see if it indicates the 64 bit Application Register has data in it. If so
it will read the Application register contents and return it followed by the contents of the TEDS EEPROM.
NOTE: The 1st byte of the DS2430A EEPROM data should contain the checksum of both the Application
Register contents and the EEPROM contents if the TEDS is IEEE 1451.4 compliant. No attempt is made to
validate or interpret the TEDS data.

Setting:
N/A – Command is Read only

Query:
The RTED query returns the TEDS data associated with the specified channel. This command must be directed
to a specific channel. It will return an indicator that specifies if the DS2430A Application register has been used
to store the basic TEDS data and up to 40 bytes in ASCII Hex format (8 bytes of Application register content if
it was burned and 32 bytes of the EEPROM content).

Query Format: Unit#:Ch#:CMD?


Response format: Unit#:Cmd:Ch#= <APP Reg Status>:<APP Reg Contents (8 bytes if APP Reg
Status=1)><DS2430A EEPROM Contents (32 bytes)>
Where: APP Reg Status=1 if the APP Register had data or 0 if it did not

Query: 1:1:RTED?
Response: (for TEDS chip on channel 1 with valid app register data)
1:RTED:1=1:168010a00975000012648016a88ae8e112801f2000f60ec4046dd18737f3206a380555e76
5390800

ALLC
The ALLC command is used to read several channel settings at once.

Setting:
N/A – Command is Read only

Query:
This command must be a directed command use of the global channel indicator is not allowed.

Query Format: Unit#:Ch#:CMD?


Response format: Unit#:Cmd:Ch#:= GAIN:<Gain value>; SENS:< SENS value>; FSCI:< FSCI
value>; FSCO:< FSCO value>; INPT:< INPT value>; FLTR:< FLTR value>; IEXC:< IEXC value>;
OFLT:< OFLT value>; CPLG:< CPLG value>; CLMP:< CLMP value>; CALB:< CALB value>;
VEXC:< VEXC value>; SWOT:< SWOT value>
Query: 1:1:ALLC??
Response:
1:ALLC:1=GAIN:\s\s\s1.0;SENS:\s\s10.0;FSCI:1000.0;FSCO:\s\s10.0;INPT:\s\s\s2.0;FLTR:1;IEXC:2;
OFLT:0;CPLG:0;CLMP:0;CALB:1;VEXC:\s\s\s0.0;SWOT:0;\r\n

NOTE: Some items returned may not be applicable for every model

RBIA
The RBIA command is used to read all channels Bias Levels.
MODEL 483C41 GENERAL OPERATION MANUAL 27

Setting:
N/A – Command is Read only

Query:
This command is a global command and will return all channel bias readings regardless of the channel id in the
command.

Query Format: Unit#:Ch#:RBIA?


Response format: Unit#:Cmd:Ch#:=CH1 bias;… CHn#:=CHn bias;
Query: 1:1:RBIA?
Query Response: 1:RBIA:1= 25.5;2= 25.5;3= 25.5;4= 25.5;

STUS
The STUS command is used to read the unit and all channel status indicators.

Setting:
N/A – Command is Read only

Query:
This command is a global command and will return all channel bias readings regardless of the channel id in the
command.

Query Format: Unit#:Ch#:STUS?


Response format: Unit#:Cmd:Ch#:<unit status bit map>;<CH1 status bit map>;…;<CHn status bit
map>;
Query: 1:1:STUS?
Query Response: 1:STUS:1:0;1;5;5;5;

Where :
Unit Bit 0 = 1= BAD EEPROM read for channel settings on power up
Unit Bit 1 = 1= BAD EEPROM read for Unit options on power up
Unit Bit 2 = 1= BAD EEPROM read for cal factors on power up
For Unit bit map 0=no errors
Channel Bit 0= Open input fault (0=Fault; 1 = no fault)
Channel Bit 1= Short input fault (0=Fault; 1 = no fault)
Channel Bit 2= Overload condition (0=Overload; 1 = no Overload)
For channel bit map 7=no errors

UNIT
Query:
The UNIT query returns the unit configuration information which includes the installed options, unit number,
Model id and starting channel number
Query Format: Unit#:Ch#:CMD?
Response format: Unit#:Cmd:Model string:Firmware Ver string: Serial Number:Cal Date:
Unit#:#channels:Starting Ch#: Gain Option byte, Input Option Byte, Filter option byte,Misc1 option byte,Misc2
option byte: fixed or 1st option of input Filter corners: output filter corners<cr><lf>
Query: 1:1:UNIT?
Response :1:UNIT:483C41\s\s\s\s\s\s\s\s\s\s:FW\sVer\s4.00\s\s\s\s\s:12345:06-28-
2011:1:4:1:16,10,16,140,132:30.00000:30.00000:30.00000:30.00000:0.00000:0.00000:0.00000:0.00000:\r\n

Model & Firmware version strings are self-explanatory. Following are:


Serial Number (U16): Cal Date (10 character string): Filter corner (kHz)
Unit Id: Number of Channels: Starting Channel Number
Followed by the option bytes:
Gain Options
MODEL 483C41 GENERAL OPERATION MANUAL 28

OPT_GAIN_INC 0x10 Incremental Gain


Input Options
OPT_INP_CIM 0x02 ICP|VOLT|CHG
OPT_INP_INTCAL 0x08 Internal Cal
Filter
OPT_FILTER_OUT 0x02 Output filter (optional)
OPT_FILTER_PGM_BTR 0x10 Programmable Butterworth LP filter
Miscellaneous Options
OPT_MISC_TEDS 0x04 TEDS
OPT_MISC_IEXC 0x08 current excitation
More Miscellaneous Options
OPT_MISC2_NOPWR 0x80 No Soft (logical) Power Button

UNID
The UNID command is used to set the units ID number. The Unit Id number is critical to remote
communications since it indicates to the unit which commands it should accept and respond to. Units are
typically shipped with a unit id of 1. If more than one unit is in the system and they will be communicated with
remotely the user must set a unique id in each unit. This can be done through the front panel interface, if one
exists, or by connecting to each unit individually and sending this command.

Setting:
1:1:UNID= 2\r\n ( unit 1, channel 1, New ID=2 )

Setting Response: 2:UNID:ok


NOTE: The new Unit Id becomes effective immediately

Query:
This command can be sent as a query but its usefulness is marginal being as it is a directed command and as
such it is necessary to include the Unit Id in the command and the response will simply validate the commands
unit id parameter.
Query Format: Unit#:Ch#:UNID?
Response format: Unit#:Cmd:Ch#:=unit id
Query: 2:1:UNID?
Query Response: 2:UNID:1=2;

LPCR
Query:
The LPCR query returns the unit low pass filter corners
Query Format: Unit#:Ch#:CMD?
Response format: Unit#:Cmd:,<num corners>< n filter corners> <cr><lf>
Query: 1:1:LPCR?
Response: 1:LPCR:6.000:30.000:10.000:3.000:1.000:0.300:0.100:\r\n
Global Query: 1:0:LPCR?
Global Response 1:LPCR:6.000:30.000:10.000:3.000:1.000:0.300:0.100:
6.000:30.000:10.000:3.000:1.000:0.300:0.100:6.000:30.000:10.000:3.000:1.000:0.300:0.100:
6.000:30.000:10.000:3.000:1.000:0.300:0.100:\r\n

CAUTION! Query only - The user should NEVER send this as a setting command as these values depend on the
HW installed and will be set at the factory.

LEDS
The LEDS command is used to test the LED functionality of the front panel. When sent as a command the
LED’s on the front panel will flash 3 times. This command invokes a function and therefore has no query
capability.
MODEL 483C41 GENERAL OPERATION MANUAL 29

Setting:
2:1:LEDS= 0\r\n ( unit 2, channel 1, Any value )

Setting Response: 2:LEDS:ok

Query:
N/A

RSET
The RSET command is used to restore the factory default channel settings. This command invokes a function
and therefore has no query capability.

Setting:
2:1:RSET = 0\r\n ( unit 2, channel 1, Any value )

Setting Response: 2:RSET:ok

Query:
N/A

The factory Defaults are:


 Gain = 1.0
 Gain Type = Gain
 Sensitivity =10.0
 Full Scale input = 1000.0
 Full Scale output = 10.0
 Input Mode = ICP
 Current excitation = 4.0mA
 Output Filter = Disabled
 Internal Reference Oscillator = OFF

SAVS
The SAVS command is used to store the current channel setting as the default settings that will be restored on
power up. This command invokes a function and therefore has no query capability.

Setting:
2:1:SAVS = 0\r\n ( unit 2, channel 1, Any value )

Setting Response: 2:SAVS:ok

Query:
N/A

Communication Responses
Typically the unit will return <Unit>:<Cmd String>:OK when the command is successful. Errors are indicated with
negative numbers. The unit may return one of the following:

<Unit>:<Cmd String>:OK<CR> <LF>


Represents that the last command was entered in the correct format and was performed properly.

<Unit>:<Cmd String>:-1<CR> <LF>


Option Error. The unit is not equipped with the option necessary to implement the command sent
MODEL 483C41 GENERAL OPERATION MANUAL 30

<Unit>:<Cmd String>:-2<CR> <LF>


Channel Error. The channel number in the command is invalid.

<Unit>:<Cmd String>:-3<CR> <LF>


Command Error. The command is not recognized.

<Unit>:<Cmd String>:-4<CR> <LF>


Unit Error. The unit number in the command is invalid.

<Unit>:<Cmd String>:-5<CR> <LF>


Unit Error. The function invoked by the command encountered an error or a query only command (ex.
RBIA) was sent as a setting.

<Unit>:<Cmd String>:-6<CR> <LF>


Command Parameter Error. A channel setting parameter was found to be out of range.
Model Number Revision: E
483C41 EIGHT-CHANNEL ICP®/VOLTAGE/CHARGE SENSOR SIGNAL CONDITIONER ECN #: 49172

Performance ENGLISH SI OPTIONAL VERSIONS


Channels 8 8 Optional versions have identical specifications and accessories as listed for the standard model
Sensor Input Type(s) ICP®, Voltage, ICP®, Voltage, Charge except where noted below. More than one option may be used.
Charge
Gain(ICP®/ Voltage Input) 0.1 to 200 0.1 to 200
Gain(mV/pC)(Charge Input) 0.01 to 2000 0.01 to 2000
Gain Increment(minimum)(ICP®/ Voltage Input) 0.1 0.1
Gain Increment(minimum)(Charge Input) 0.01 0.01
Accuracy(ICP®/ Voltage Input)(Gain, 0.1 to 0.4) ±5% ±5%
Accuracy(ICP®/ Voltage Input)(Gain, 0.5 to 200) ±1% ±1%
Accuracy(Charge Input)(Gain, 0.01 to 0.04) ±6% ±6%
Accuracy(Charge Input)(Gain, 0.05 to 2000) ±1% ±1%
Input Range(maximum)(Charge Input) 100,000 pC pk 100,000 pC pk
Input Range(maximum)(ICP® Input) 10 Vpk 10 Vpk [4]
Input Range(maximum)(Voltage Input) 5 Vpk 5 Vpk
Low Frequency Response(-5 %)(ICP®/ Voltage Input) ≤ 0.05 Hz ≤ 0.05 Hz
Low Frequency Response(-5 %)(Charge Input) 0.5 Hz 0.5 Hz [5]
Electrical Filter Roll-off 160 dB/decade 160 dB/decade
Filter Type(8-pole Butterworth) Low Pass Low Pass
High Frequency Response(-3 dB)(Gain from 0.01 to 99.9) >100 kHz >100 kHz
High Frequency Response(-3 dB)(Gain from 100 to 2000) >80 kHz >80 kHz
Electrical Filter Corner Frequency(-10 %) 0.1-0.3-1-3-10-30 kHz 0.1-0.3-1-3-10-30 kHz [6]
Electrical Filter Pass Band Amplitude Accuracy 1% 1%
Phase Response(at 1 kHz) ±2° ±2°
Non-Linearity 1% 1%
Cross Talk <-72 dB <-72 dB
TEDS Sensor Support Yes Yes
Fault/Bias Monitor LEDS Open/Short/Overload Open/Short/Overload
Control Interface NOTES:
Digital Control Interface Ethernet Ethernet [1] Use T1.6A fuse.
Human Interface Keypad Keypad [2] User adjustable, factory set at 4 mA (± 1.0 mA). Each channel individually adjustable.
Display 2 rows, 16 columns 2 rows, 16 columns [3] Typical.
Environmental [4] Max input signal is dependant on sensor bias.
Temperature Range(Operating) +32 to +120 °F 0 to +50 °C [5] The low frequency tolerance is accurate within ±25% of the specified frequency.
Electrical [6] The high frequency tolerance is accurate within ±5% of the specified frequency.
[7] See PCB Declaration of Conformance PS024 for details.
Power Required(direct input to unit) AC Power AC Power
AC Power(47 to 63 Hz) 100 to 240 VAC 100 to 240 VAC
AC Power ≤ 0.7 Amps ≤ 0.7 Amps [1]
Excitation Voltage(To Sensor) >24 VDC >24 VDC
DC Offset <50 mV <50 mV
Constant Current Excitation(To Sensor) 2 to 20 mA 2 to 20 mA [2]
Output Voltage(minimum) 10 V 10 V
Output Current(minimum) 10 mA 10 mA
Output Impedance <50 Ohm <50 Ohm
Broadband Electrical Noise(1 to 10,000 Hz)(Gain x1) 50 µV/rms 50 µV/rms [3]
Spectral Noise(1 Hz) 8 µV/√Hz 8 µV/√Hz [3]
Spectral Noise(10 Hz) 2 µV/√Hz 2 µV/√Hz [3]
Spectral Noise(100 Hz) 0.7 µV/√Hz 0.7 µV/√Hz [3]
Spectral Noise(1 kHz) 0.7 µV/√Hz 0.7 µV/√Hz [3]
Spectral Noise(10 kHz) 0.6 µV/√Hz 0.6 µV/√Hz [3]
Broadband Electrical Noise(1 to 10,000 kHz)(Gain x10) 75 µV rms 75 µV rms [3]
Spectral Noise(1 Hz) 18 µV/√Hz 18 µV/√Hz [3]
Spectral Noise(10 Hz) 1.5 µV/√Hz 1.5 µV/√Hz [3]
Spectral Noise(100 Hz) 1.0 µV/√Hz 1.0 µV/√Hz [3]
Spectral Noise(1 kHz) 1.0 µV/√Hz 1.0 µV/√Hz [3]
Spectral Noise(10 kHz) 1.0 µV/√Hz 1.0 µV/√Hz [3]
Broadband Electrical Noise(1 to 10,000 Hz)(Gain x100) 350 µV rms 350 µV rms [3]
Spectral Noise(1 Hz) 100 µV/√Hz 100 µV/√Hz [3]
Spectral Noise(10 Hz) 10 µV/√Hz 10 µV/√Hz [3]
Spectral Noise(100 Hz) 6 µV/√Hz 6 µV/√Hz [3]
Spectral Noise(1 kHz) 5 µV/√Hz 5 µV/√Hz [3]
Broadband Electrical Noise(1 to 10,000 Hz)(0.1 mV/pC) 52.0 µV/rms 52.0 µV/rms [3]
Spectral Noise(10 kHz) 5 µV/√Hz 5 µV/√Hz [3]
Spectral Noise(1 Hz) 10.0 µV/√Hz 10.0 µV/√Hz [3]
Spectral Noise(10 Hz) 1.5 µV/√Hz 1.5 µV/√Hz [3]
Spectral Noise(100 Hz) 0.6 µV/√Hz 0.6 µV/√Hz [3]
Spectral Noise(1 kHz) 0.6 µV/√Hz 0.6 µV/√Hz [3]
Spectral Noise(10 kHz) 0.6 µV/√Hz 0.6 µV/√Hz [3]
Broadband Electrical Noise(1 to 10,000 Hz)(1.0 mV/pC) 52.0 µV/rms 52.0 µV/rms [3]
Spectral Noise(1 Hz) 14.0 µV/√Hz 14.0 µV/√Hz [3]
Spectral Noise(10 Hz) 2.0 µV/√Hz 2.0 µV/√Hz [3]
Spectral Noise(100 Hz) 0.7 µV/√Hz 0.7 µV/√Hz [3]
Spectral Noise(1 kHz) 0.7 µV/√Hz 0.7 µV/√Hz [3]
Spectral Noise(10 kHz) 0.7 µV/√Hz 0.7 µV/√Hz [3]
Broadband Electrical Noise(1 to 10,000 Hz)(10.0 mV/pC) 56.0 µV/rms 56.0 µV/rms [3]
Spectral Noise(1 Hz) 15.0 µV/√Hz 15.0 µV/√Hz [3]
Spectral Noise(10 Hz) 2.0 µV/√Hz 2.0 µV/√Hz [3]
Spectral Noise(100 Hz) 0.6 µV/√Hz 0.6 µV/√Hz [3]
Spectral Noise(1 kHz) 0.6 µV/√Hz 0.6 µV/√Hz [3]
Spectral Noise(10 kHz) 0.6 µV/√Hz 0.6 µV/√Hz [3]
Oscillator(+/- 2%)(Internal Generator - ICP®/ Voltage Mode) 0.1 V pk 0.1 V pk
Oscillator(+/- 2%)(Internal Generator - Charge Mode) 100 pC pk 100 pC pk
Oscillator(+/- 2%) 100/1000 Hz 100/1000 Hz
Overload Threshold(± 0.5 Vpk) ± 10 Vpk ± 10 Vpk
Physical
Electrical Connector(Input, sensor) BNC Jack BNC Jack
Electrical Connector(Output) BNC Jack BNC Jack SUPPLIED ACCESSORIES:
Electrical Connector(Ethernet) RJ-45 RJ-45 Model 017AXX Power Cord (1)
Size (Height x Width x Depth)(nominal) 1.75 in x 19 in x 13.7 44.5 mm x 482.6 mm x 348 Model EE75 PCB MCSC Control Software. (1)
in mm
Weight 8 lb 3.6 kg
Entered: LK Engineer: CPH Sales: KK Approved: DY Spec Number:

Date: 2/26/2019 Date: 2/26/2019 Date: 2/26/2019 Date: 2/26/2019 46951


[7]
All specifications are at room temperature unless otherwise specified. Phone: 716-684-0001
In the interest of constant product improvement, we reserve the right to change specifications without notice.
Fax: 716-684-0987
ICP® is a registered trademark of PCB Group, Inc. 3425 Walden Avenue, Depew, NY 14043 E-Mail: [email protected]
4 3 2 1
PCB Piezotronics Inc. claims proprietary rights in
the information disclosed hereon. Neither it nor any REVISIONS
reproduction thereof will be disclosed to others
without the written consent of PCB Piezotronics Inc.
REV DESCRIPTION DIN

46903 B REMOVED EXT CAL INPUT CONNECTOR 42318

D D

EXTERNAL AC POWER
INPUT AND FUSE
ETHERNET INPUT
16X BNC CONNECTORS

C 17.0 C
[432]

12.00
[304.7] 13.7
[347]

1.00
[25.3]

B B

18.3
[465]

1.73
[43.9] 1.25
[31.8]

DISPLAY/KEYPAD AC POWER SWITCH

A A

UNLESS OTHERWISE SPECIFIED TOLERANCES ARE: DRAWN CHECKED ENGINEER


DIMENSIONS IN INCHES DIMENSIONS IN MILLIMETERS
[ IN BRACKETS ] JDM 3/19/14 JDM 3/19/14 HK 3/19/14

DECIMALS XX ±.03 DECIMALS X ± 0.8 TITLE


3425 WALDEN AVE. DEPEW, NY 14043
XXX ±.010 XX ± 0.25 (716) 684-0001 E-MAIL: [email protected]
OUTLINE DRAWING
ANGLES  2 DEGREES ANGLES  2 DEGREES CODE DWG. NO.
MODEL 483C41
SENSOR SIGNAL CONDITIONER
IDENT. NO.
52681 46903
FILLETS AND RADII FILLETS AND RADII
.003 - .005 0.07 - 0.13 SCALE: .75 X SHEET 1 OF 1

4 3 2 1

You might also like