BS-120 - 200 - 200e Sop - V2.0
BS-120 - 200 - 200e Sop - V2.0
BS-120 - 200 - 200e Sop - V2.0
1 / 47
International IVD application team
For application training only V2.0
Contents
1. Prepare for Power On ·····································································································4
1.1 Preparation List ···················································································································4
1.2 Attention & cautions ···········································································································4
2. Power On ·······················································································································5
2.1 Main procedure ···················································································································5
2.2 Startup of the BS-120/200 software ····················································································5
2.2.1 Lamp background measurement ·············································································5
2.2.2 Cuvettes replacement ······························································································6
2.2.3 Washing solution check ····························································································7
2.3 Attention & Cautions ···········································································································7
3. Prepare for analysis ········································································································8
3.1 Reagent ·······························································································································8
3.1.1 Reagent parameter configuration ············································································8
3.1.1.1 Database import (for Mindray reagent) ························································8
3.1.1.2 Manually type-in (for mindray reagents) ······················································9
3.1.1.3 Manually type-in (for other reagent brand) ················································10
3.1.2 Reagent storage(Mindray only) ··············································································10
3.1.3 Reagent positioning································································································13
3.1.4 Attention & precautions ·························································································14
3.2 Calibrator···························································································································15
3.2.1 Calibrator brief introduction ··················································································15
3.2.2 Calibrator storage condition ···················································································16
3.2.3 Calibrator material preparation ·············································································17
3.2.4 Calibrator parameter setup ····················································································18
3.2.4.1 Positioning and assay value input ·······························································18
3.2.4.2 Calibration rule setup ··················································································20
3.3 Control·······························································································································20
3.3.1 Control brief introduction & storage ······································································20
3.3.2 Control material preparation ·················································································22
3.3.3 Control parameter setup ························································································22
3.3.3.1 Positioning and assay value input ·······························································22
3.3.3.2 QC rules setup ·····························································································23
3.3.3.3 Westgard rule introduction,, ········································································23
3.4 Samples ·····························································································································25
3.4.1 Sample types ··········································································································25
3.4.2 Sample preparation ································································································26
3.4.3 Sample storage conditions ·····················································································26
4. Perform Routine Analysis ······························································································26
4.1 Calibration ·························································································································26
4.1.1 Purpose of Calibration ····························································································26
4.1.2 Calibration frequency requirement ········································································27
4.1.3 Calibration request ·································································································27
2 / 47
International IVD application team
For application training only V2.0
4.1.4 Calibration results ··································································································28
4.2 QC program ·······················································································································30
4.2.1 Purposes of QC program ························································································30
4.2.2 QC request··············································································································30
4.2.3 QC results ···············································································································31
4.3 ISE module (Optional) ·······································································································32
4.3.1 ISE introduction ······································································································32
4.3.2 ISE reagent··············································································································33
4.3.3 ISE calibration ·········································································································34
4.3.4 ISE QC program ······································································································35
4.3.5 ISE Maintenance schedule and reagent consumption ···········································37
5. Analyze daily patient samples ·······················································································38
5.1 Normal routine sample request ························································································38
5.2 Emergency samples ···········································································································38
5.3 Sample request in batch ····································································································39
5.4 Other functions ·················································································································40
5.4.1 Sample dilution ······································································································40
5.4.1.1 Why sample dilution ···················································································40
5.4.1.2 Sample dilution on BS series ·······································································41
6. Shut down the machine ································································································42
6.1 Exit the operation software·······························································································42
6.2 Switch off the machine······································································································42
7. Easy maintenance ·········································································································44
3 / 47
International IVD application team
For application training only V2.0
1 Power supply and power cord of the analyzer, computer and printer
2 Connections between the analyzer and computer, as well as computer and printer
4 Deionized water tank (full) and the connection with the analyzer (sensor and tube)
5 Waste tank (empty) and the connection with the analyzer (sensor and tube)
11 Urine diluent (38) and Cleaning solution for ISE(37) in reagent disk (if ISE installed)
The washing solution on reagent position 39 is 10% CD80 solution. If necessary the
concentration of CD-80 solution can be elevated to increase the washing efficiency. However,
the original concentrated CD80 liquid is not recommended for the direct usage due to the
self-crystallization under low temperature.
The DI water quality is very important. The water filter in the DI water tank should be
regularly checked before switching on the power, and replacement is required if the filter
turns to dark color. The normal replacement frequency is about 4-6 months; the DI water
supplier should be changed if the filter is replaced less than 3 months.
The ISE module is optional with the main chemistry analyzer. The area might be empty if no
ISE module mentioned on your purchase order. Since the installation of ISE module is strictly
prohibited out of Mindray factory, please make sure to double check with customers
whether the ISE module is required within few years.
4 / 47
International IVD application team
For application training only V2.0
2. Power On
6 Starting the Operating Software, Logon in(Default username: Admin; password: Admin)
Since BS-120 and BS-200 require the disposable cuvettes, there are some more steps before
entering into the main display.
Fig 2.2.1-1
Open the reaction disk window and unload the cuvette on position 1, then close the window and
click ‘OK’ to process the measurement.
N.B. there are two circus holder points on the cuvettes segments, please makes sure this holder
points fix with the reaction disk position tightly. The cuvette segments might fall off the reaction
disk under the high speed rotation if this step is not completed properly.
5 / 47
International IVD application team
For application training only V2.0
Fig 2.2.2-1
(1) Open the reaction disk window and load a clean cuvette segment, then close the window
(2) Click ‘OK’ to ensure the background measurement
(3) If the background checking is passed, the current status will be changed into ‘unused’
(4) Repeat step 1 and 2 until there are enough clean and unused cuvettes for the tests
(5) Click ‘Next’ to leave this page.
6 / 47
International IVD application team
For application training only V2.0
Fig 2.2.3-1
Please make sure again the washing solution is placed on reagent position 39 (For BS-120, the
detergent position is 35)
Routine wash means the washing process only require the DI water to wash the probe.
Enhanced wash means the washing process consume some CD80 detergent to extra wash probe.
There are two switches on the machine, the back one is the main power and the one the
right/left hand side of machine is to control the analysis part except refrigerator. The main
switch is turned on all the time and the analyzing switch can be turned off when the daily
works are finished every day.
The disposable cuvettes on BS-120 and BS-200 are different. There are 5 cuvettes on each
BS-120 segment and 10 cuvettes on each BS-200 segment. Those cuvettes are not allowed to
reuse. Even though the background checking is passed, some unexpected and transparent
contaminators remain on the walls.
7 / 47
International IVD application team
For application training only V2.0
3.1 Reagent
Test parameters should be input before entering any reagent information. There are two
pathways to enter the parameter application data:
(1) When you contact Mindray application specialist, you will receive the database files called
‘DB200.mdb’. The database name is exactly the same for all three modules but they cannot
be swapped. Please make sure you receive the correct file.
(2) Exit the operation software
(3) Copy the database file and paste it into the ‘Database’ folder under the software installation
folders. You need to replace the previous file with the given one. (Please make sure that the
previous database backup has been done properly since the new database will erase the
history documents and the customer might lose all the patient results if no backup.)
Fig 3.1.1.1-1
8 / 47
International IVD application team
For application training only V2.0
Fig 3.1.1.1-2
(1) Please contact mindray application specialist and ask for a latest copy of application
parameter sheet.
Fig 3.1.1.2-1
(2) Go to ‘ParametersTestParameters’ page
(3) Click ‘Add’ to create a new parameter account.
9 / 47
International IVD application team
For application training only V2.0
(4) Fill in the parameter information by using the sheets
(5) Click ‘OK’ to save any changes
If the customers use other brand reagent, please contact your regional application specialist to
check about the details. With your query, please enclose the following information:
(1) The reagent insert with the experimental procedures.
(2) The insert paper should include wavelength, reagent principle, reagent/sample volume,
manual operation procedure, incubation/reaction time and etc.
The application specialist will send you the instruction or final setup data if the reagent is valid on
our machine. However Mindray cannot guarantee the accuracy or traceability of final results if
using other brand reagents.
Fig 3.1.1.3-1
The reagents are valid up to the expiration date on the label when stored unopened at
correct temperature and protected from light. The detailed shelf life for all mindray reagent
are shown on next few pages(Table 3.1.2-2/3/4). All the reagent should be stored at 2-8
degrees except bilirubin DSA reagents which require the room temperature conditions due
to the self sedimentation under low temperature.
Once opened, the reagents are stable for centain days when refrigerated on analyzer or
refrigerator and the reagents are strictly prohibited to freeze under any circumstances. The
detailed onboard stability dates are also shown in next few pages(Table 3.1.2-1). Please
make sure the reagent should keep fresh all the time to ensure the accuracy of the final
results.
The contamination of reagents must be avoided. The contaminating source might be from
the air, condensation water, human saliva, carryover of other reagent, abnormal samples,
machine spare parts and etc. Please make sure the reagent disk cover should be closed all
the time.
10 / 47
International IVD application team
For application training only V2.0
Some parameters require the instant working solutions which has lower stability, please
make sure the working solution should be strictly stored in the refrigenator all the time.
On the low throughput machine, some new parameter is not valided. Please contact your
mindray regional application specialist for the details.
11 / 47
International IVD application team
For application training only V2.0
Table 3.1.2-2 the following parameters with the theoretical shelf life: 18 months
12 / 47
International IVD application team
For application training only V2.0
Table 3.1.2-3 The following parameters with the theoretical shelf life: 15 months
Table 3.1.2-4 The following parameters with the theoretical shelf life: 12 months
(1) The reagent could randomly insert into the reagent holder, there are three different
configured reagent package for BS series analyzer, and on BS-120/200/200E, the reagent
package is in green colour.
Fig 3.1.3-1
13 / 47
International IVD application team
For application training only V2.0
(2) The standard configuration for BS-120/200/200E series reagent bottles are totally 40ml and
20ml for R1 and R2 respectively. And R1 and R2 are sharing with the same reagent disk and
probe. The dead volume for R1 and R2 bottles are 2.5ml and 1.5ml respectively.
(3) Please make sure the refrigenator of analyzer is swtiched on and the temperature inside is
dropped down. The reagent disk cover should be closed all the time to avoid any
condensation water production.
(4) After insert the reagent into the reagent disk, please note down the position number.
(5) Go to software page ‘ReagentReagent setup’, and enter into the reagent position and
expiration date information properly. The orange background means the logically incorrect
input of reagent information.
For the double reagent parameter, you need to fill in both R1 and R2 information correctly to
carry on the following tests.
Fig 3.1.3-2
During the transportation, the reagent will be placed in a sealed and heat isulated box with
some ice bag inside. Those ice bags will perform a cool environment between 2-8 degrees
for about 3 days. However it takes about a week for the whole boxes reach to the room
temperature. Please make sure you finish the custom clearance documentary works as soon
as possible.
When you open the boxes, please pay more attention on the reagent status. Please
immediately contact the express company if any of those situations happened:
14 / 47
International IVD application team
For application training only V2.0
The ice bags are thoroughly thawed and the inside temperature is almost room
temperture or above.
Lots of condensation water is remaining on the bottom of the box.
Any leakage of reagent or ice bags materials.
Any irritating smells
Any abnormal or foreign objects in the box
3.2 Calibrator
Mindray calibrator includes two group materials, the multi parameter calibrators and
individual single parameter calibrators.
Multi parameter calibrator includes multi sera calibrator, specific protein calibrator and lipids
calibrator, which could support more than one parameter. The single parameter calibrators
are usually unique for a certain parameter and also can be sold individually or included in
the total reagent package.
15 / 47
International IVD application team
For application training only V2.0
2 Specific Proteins Calibrator 105-001129-00 5×1 mL C3, C4, CRP, IgA, IgG, IgM
N.B. Please make sure the correct calibrator materials are purchased after you order reagents.
This information is clearly shown on the calibrator insert paper, please make sure you have
already read and fully understand all the storage and stability content before use any materials.
16 / 47
International IVD application team
For application training only V2.0
Fig 3.2.3-1
(4) Reconstitute by adding 3.0 mL deionized water to the side of the bottle slowly.
The pipette or burette is strongly recommended to measure and transfer the deionized
water. And please try not to use measuring cylinder, syringe, or other rough scaled
measuring equipment since the inaccurate measurement would cause the incorrect
concentration of calibrator.
17 / 47
International IVD application team
For application training only V2.0
Fig 3.2.3-2
Secondly the quality of water is also very important. Please make sure the fresh deionized
water is required for the calibrator preparation process each time.
(5) Carefully replace the rubber stopper and mix contents by inverting several times and
swirling gently to ensure all the components of powder are dissolved.
Please mix the content really carefully and avoid forming the foams and bubbles as possible.
After mixing process, the calibrator solution should be kept still for at least 15 mins at room
temperature.
(6) Divide the calibrator solution into small portions to micro tubes as shown below. The dead
volume of this micro tube is about 200ul, so we recommend that 500ul calibrator material in
each vials.
The vials should be stored in the freezer at -20 degrees all the time. Once the material is
taken out of the freezer and thawed to room temperature, the calibrator should not be
re-frozen.
Fig 3.2.3-3
(1) Go to ‘CalibrationCalibrator’ page and click ‘Add’ to create a new calibrator account.
(2) Type in the calibrator name, exp. date, position and lot number
18 / 47
International IVD application team
For application training only V2.0
(3) Take the assay value insert paper out of calibrator package and read the target value for each
parameter.
Fig 3.2.4.1-1
Fig 3.2.4.1-2
19 / 47
International IVD application team
For application training only V2.0
3.3 Control
Mindray controls also include two group materials, the multi parameter controls and
individual single parameter controls. And each control includes two levels, the normal level
and pathological level.
Please check the table 3.3.1-1 very carefully and make sure you purchased the correct
control material for your test items.
20 / 47
International IVD application team
For application training only V2.0
The storage conditions of control are also closed to calibrator and please read the insert
paper carefully before any actions.
Fig 3.3.1-1
1 0 5 -0 0 1 1 4 5 -0 0 1 0 ×5 mL A L B, A L P , A L T , A M Y , A ST , D B- D SA , D B- V O X, T B- D SA , T B-
V O X, C a, T C , C K, C rea- Jaff, C rea- S, G L U - H K, G L U - O , G G T ,
1 M ulti C ontrol Sera N
H BD H , I gA , I gG , I gM , L D H - L , M g, P , T P , T G , U rea, U A , Fe, C H E ,
1 0 5 -0 0 1 1 3 3 -0 0 2 0 ×5 mL L I P , N a +, K +, C l -
1 0 5 -0 0 1 1 4 6 -0 0 1 0 ×5 mL A L B, A L P , A L T , A M Y , A ST , D B- D SA , D B- V O X, T B- D SA , T B-
V O X, C a, T C , C K, C rea- Jaff, C rea- S, G L U - H K, G L U - O , G G T ,
2 M ulti C ontrol Sera P
H BD H , I gA , I gG , I gM , L D H - L , M g, P , T P , T G , U rea, U A , Fe, C H E ,
1 0 5 -0 0 1 1 3 4 -0 0 2 0 ×5 mL L I P , N a +, K +, C l -
1 0 5 -0 0 1 1 3 8 -0 0 5 ×1 mL
Spec ific P roteins
3
C ontrol N
1 0 5 -0 0 1 1 4 7 -0 0 1 0 ×1 mL
C 3 , C 4 , C RP , I gA , I gG , I gM , T P , A L B
1 0 5 -0 0 1 1 3 9 -0 0 5 ×1 mL
Spec ific P roteins
4
C ontrol P
1 0 5 -0 0 1 1 4 8 -0 0 1 0 ×1 mL
P realbumin C ontrol
5 1 0 5 -0 0 1 1 3 7 -0 0 (N )3 ×1 mL +(P )3 ×1 mL PA
N &P
6 L ipids C ontrol N 1 0 5 -0 0 1 1 3 5 -0 0 6 ×3 mL A P O A 1 , A P O B, T C , T G , H D L - C , L D L - C
7 L ipids C ontrol P 1 0 5 -0 0 1 1 3 6 -0 0 6 ×3 mL A P O A 1 , A P O B, T C , T G
L ipoprotein(a) C ontrol
8 1 0 5 -0 0 1 1 4 3 -0 0 (N )2 ×1 mL +(P )2 ×1 mL L p(a)
N &P
10 C K- M B C ontrol N 1 0 5 -0 0 1 1 4 1 -0 0 4 ×3 mL
C K- M B
11 C K- M B C ontrol P 1 0 5 -0 0 1 1 4 2 -0 0 4 ×3 mL
12 H bA 1 c C ontrol P 1 0 5 -0 0 2 1 3 8 -0 0 4 ×1 mL
H bA 1 c
13 H bA 1 c C ontrol N 1 0 5 -0 0 2 1 4 0 -0 0 4 ×1 mL
14 Rhematois m C ontrol N 1 0 5 -0 0 2 1 3 6 -0 0 4 ×3 mL
H S- C RP , A SO , RF
15 Rhematois m C ontrol P 1 0 5 -0 0 2 1 3 7 -0 0 4 ×3 mL
16 H C Y C ontrol N 1 0 5 -0 0 2 1 4 1 -0 0 3 ×1 mL
HC Y
17 H C Y C ontrol P 1 0 5 -0 0 2 1 3 9 -0 0 3 ×1 mL
18 C O 2 C ontrol N 1 0 5 -0 0 2 1 4 2 -0 0 3 ×5 mL CO2
19 FU N C ontrol P 1 0 5 -0 0 2 1 4 3 -0 0 3 ×1 mL FU N
23 U I BC c ontrol 1 0 5 -0 0 2 3 0 7 -0 0 1 ×1 level×5 mL U I BC
26 M A L B c ontrol 1 0 5 -0 0 2 3 1 6 -0 0 1 ×1 level×1 mL M A LB 21 / 47
International
27 IVD application1 0team
T RF c ontrol 5 -0 0 2 3 1 8 -0 0 1 ×2 levels ×1 mL T RF
28 β- H B c ontrol 1 0 5 -0 0 2 3 2 0 -0 0 1 ×2 levels ×5 mL β- H B
1 0 5 -0 0 3 6 9 7 -0 0 1 ×1 mL
29 A D A C ontrol L A DA
1 0 5 -0 0 3 6 9 8 -0 0 3 ×1 mL
1 0 5 -0 0 3 6 9 3 -0 0 1 ×1 mL
15 Rhematois m C ontrol P 1 0 5 -0 0 2 1 3 7 -0 0 4 ×3 mL
16 H C Y C ontrol N 1 0 5 -0 0 2 1 4 1 -0 0 3 ×1 mL
HC Y
17 H C Y C ontrol P 1 0 5 -0 0 2 1 3 9 -0 0 3 ×1 mL
18 C O 2 C ontrol N 1 0 5 -0 0 2 1 4 2 -0 0 3 ×5 mL CO2
19 FU N C ontrol P 1 0 5 -0 0 2 1 4 3 -0 0 3 ×1 mL FU N
23 U I BC c ontrol 1 0 5 -0 0 2 3 0 7 -0 0 1 ×1 level×5 mL U I BC
26 M A L B c ontrol 1 0 5 -0 0 2 3 1 6 -0 0 1 ×1 level×1 mL M A LB
27 T RF c ontrol 1 0 5 -0 0 2 3 1 8 -0 0 1 ×2 levels ×1 mL T RF
28 β- H B c ontrol 1 0 5 -0 0 2 3 2 0 -0 0 1 ×2 levels ×5 mL β- H B
1 0 5 -0 0 3 6 9 7 -0 0 1 ×1 mL
29 A D A C ontrol L A DA
1 0 5 -0 0 3 6 9 8 -0 0 3 ×1 mL
1 0 5 -0 0 3 6 9 3 -0 0 1 ×1 mL
30 5 '- N T C ontrol L 5 '- N T
1 0 5 -0 0 3 6 9 4 -0 0 3 ×1 mL
1 0 5 -0 0 3 6 8 9 -0 0 1 ×1 mL , 1 ×1 mL
31 C ys C C ontrol L &H C ys C
1 0 5 -0 0 3 6 9 0 -0 0 3 ×1 mL , 3 ×1 mL
1 0 5 -0 0 3 6 9 1 -0 0 1 ×1 mL
32 β2 - M G C ontrol L β2 - M G
1 0 5 -0 0 3 6 9 2 -0 0 3 ×1 mL
1 0 5 -0 0 3 6 9 7 -0 0 1 ×1 mL , 1 ×1 mL
33 A FU C ontrol L &H A FU
1 0 5 -0 0 3 6 9 8 -0 0 3 ×1 mL , 3 ×1 mL
1 0 5 -0 0 3 6 9 5 -0 0 1 ×5 mL
34 T BA C ontrol L T BA
1 0 5 -0 0 3 6 9 6 -0 0 3 ×5 mL
Please kindly follow the instruction on section 3.2.3 calibrator material preparation.
N.B. Please read the instruction paper carefully and dissolve the material with the correct
amount of ID water.
22 / 47
International IVD application team
For application training only V2.0
Go to ‘QCControl’ page:
(1) Click ‘Add’ to create a new control account.
(2) Edit the control name, lot numbers, position and expired date.
(3) Select the test items you required in your lab.
(4) Fill in the target and SD value by reading from the assay value insert paper of control
material. (Please refer to section 3.3.2 calibrator parameter setup.)
(5) Click ‘OK’ to confirm all your setup.
Go to ‘ParametersTestQC’ page:
(1) Select the parameter item that requires the QC test.
(2) Select the Westgard rule. (Refer to Section 3.3.3.3 Westgard rule introduction)
(3) Choose the control materials needed for the QC program for this test.
(4) Click ‘OK’ to confirm your settings.
13s refers to a control rule that is commonly used with a Levey-Jennings chart when the
control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a
single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.
1 Westgard JO, Barry PL, Hunt MR, Groth T. A multi-rule Shewhart chart for quality control in clinical chemistry. Clin Chem 1981;27:493-501.
2 Westgard JO, Klee GG. Quality Management. Chapter 17 in Textbook of Clinical Chemistry, 2nd edition. Burtis C, ed., WB Saunders Company, Philadelphia, 1994,
pp.548-592.
3 www.westgard.com
23 / 47
International IVD application team
For application training only V2.0
12srefers to the control rule that is commonly used with a Levey-Jennings chart when the
control limits are set as the mean plus/minus 2s. In the original Westgard multi rules QC
procedure, this rule is used as a warning rule to trigger careful inspection of the control data
by the following rejection rules.
22s - Reject when 2 consecutive control measurements exceed the same mean plus 2s or
R4s - reject when 1 control measurement in a group exceeds the mean plus 2s and another
exceeds the mean minus 2s.
24 / 47
International IVD application team
For application training only V2.0
41s - Reject when 4 consecutive control measurements exceed the same mean plus 1s or
the same mean minus 1s control limit.
10x - reject when 10 consecutive control measurements fall on one side of the mean.
3.4 Samples
25 / 47
International IVD application team
For application training only V2.0
(1) The serum sample should be centrifuged at 2500-3000rpm for 5 mins until all the serum is
separated from red cells. If the centrifugal machine cannot reach this speech, please extend
the working time.
(2) The plasma sample should be centrifuged at 3000 rpm for 10 mins until all the plasma is
separated from red cells. If the centrifugal machine cannot reach this speech, please extend
the working time. Then use a tip to pick up the visible fibro protein material present in
plasma.
(3) During the transportation, the violent vibration of samples is strictly prohibited because it
would cause the hemolysis problem and largely affect the final results.
The sample should be centrifuged within 30 mins after blood draw and remove any visible
impurities or fibrins in plasma.
If the tests are carried out within 8 hours, the sample could be stored at room temperature.
If the tests are running after 8 hours to 48 hours, the serum/plasma should be placed in
refrigerator at 2-8 degrees.
If the specimens are not completed within 48 hours, the serum/plasma should be frozen at
-15 to -20 degrees. Frozen samples could be thawed only once( analytes degeneration may
occur in samples that are repeated frozen and thawed)
The sample should be stored in the dark and dry place all the time.
4.1 Calibration
The instrument measures the light intensity through photoelectric conversion, linear
amplification and AD conversion. And according to different reaction method, the response will
be calculated from absorbance readings.
However our end users expect the concentration as the final results that printed on the report
and send to patients, response means nothing in the clinical record.
In this case, calibration is introduced to plot relationship curves between response and
concentration. After calibration, once instrument reads certain absorbance, the machine will
automatically calculate and translate into the concentration as the common results on the report.
Calibration curves include linear and non-linear types. It is strong recommended to use two-point
26 / 47
International IVD application team
For application training only V2.0
linear method to plot a straight linear line for the relationship. Please do not use one-point linear
method otherwise mentioned specially.
Fig 4.1.2-1
Go to ‘Calibration Calibration Request’ page
(1) Select the items for calibration.
N.B. if the item is in gray background and unselected, it means that during the setup process,
there are some invalid errors occurred. Please put the mouse on that item and a simple hint
will display to provide the direct solution for errors. However only one hint displayed on the
screen even though there are more than one error.
Fig 4.1.2-2
(2) Select the request type.
27 / 47
International IVD application team
For application training only V2.0
Calibration: the instrument will straightly run the calibration tests by using calibrators
selected.
Reagent blank: the instrument will only use DI water (position 40 on BS-200/200E and
position 36 on BS-120) as sample to run the tests and record the blanks.
Calibration + reagent blank: the instrument will run both processes together.
(3) Make sure the calibrator material and DI water are placed properly in the correct position,
and click ‘ok’ to confirm. Words ‘Requested tests exist’ will display on the top of screen if
successfully request.
Fig 4.1.2-3
(4) If you want to cancel any requested calibration test, select the item again and when the item
background turns back to white, click ‘ok’ to confirm the change.
28 / 47
International IVD application team
For application training only V2.0
N.B. If the regional application specialist requires your calibration information, please feedback
the following message to them:
(1) Reproducibility of the responses results
The purpose of this step is to ensure the correct results of all the repeated absorbance
reading for the same material. (The replicate of calibration process is recommended as 3)
29 / 47
International IVD application team
For application training only V2.0
4.2 QC program
“Quality Control” is a measure of precision or how well the measurement system reproduces the
same result over time and under varying operating conditions. The purpose of QC is to detect,
reduce and correct deficiencies in Lab’s internal analytical process.
QC program is strongly recommended as daily routine analysis procedure, because it could
monitor the basic status of the whole system and track the tiny performance changes of reagents,
or spare parts of machine.
Both normal and pathological controls should be worked out before running the samples.
4.2.2 QC request
Fig 4.2.2-1
Since the QC program is the daily routine works, a shortcut is created on the left column.
(1) Select the test items for QC program.
N.B. if the item is in gray background and unselected, it means that during the setup process,
there are some invalid errors occurred. Please put the mouse on that item and a simple hint
will display to provide the direct solution for errors. However only one hint displayed on the
screen even though there are more than one errors.
Fig 4.2.2-2
30 / 47
International IVD application team
For application training only V2.0
(2) Input the tests replicate number.
(3) Make sure the control materials are placed in the correct position in the sample disk and click
‘OK’ to confirm the request.
(4) Click ‘Start’ to run the QC program.
4.2.3 QC results
31 / 47
International IVD application team
For application training only V2.0
An Ion Selective Electrode (ISE) measures the individual potential voltage for a specific ion against
the constant potential voltage from a stable reference electrode. The potential difference
between two electrodes depends on the activity of the specific ion in sample.
The BS-120/200/200E ISE module is from Medica (USA) and the overview appearance of the
measuring channel is shown below:
32 / 47
International IVD application team
For application training only V2.0
In order to activate the ISE module on the software, operator should log in the software with
engineers password and follow the process as shown below:
33 / 47
International IVD application team
For application training only V2.0
(4) Tri Level Quality Control Kit
Please read all the reagent insert carefully for the further details.
The calibration of ISE module is two-point linear method by using reagent A and B respectively.
After the reagents, electrodes and spare parts are well fixed on the machine, operator can carry
out the calibration process as below:
Link: ‘Maintenance’ ‘ISE’ ‘Daily Maint’ ‘Calibration’
34 / 47
International IVD application team
For application training only V2.0
Table 4.3.3-1 ISE slope range
Electrode Na+ K+ Cl-
(1) Please refer to ‘chapter 3.3.3.1 Control parameter setup’ to enter the assay value and range
from the control insert paper.
35 / 47
International IVD application team
For application training only V2.0
Please contact the local agent or mindray application specialist for any further information.
36 / 47
International IVD application team
For application training only V2.0
37 / 47
International IVD application team
For application training only V2.0
The process is very similar to 5.1 normal sample request part, but the STAT option must be
selected. The software will give those emergency samples the highest priority in all experiments.
If the STAT samples are required during the tests, please follow the steps below to insert
emergency samples.
Step 1: When the instrument is in the experiment process, click ‘probe stop’ button on the left
column to pause. (this button can only be selectable during the experiment)
38 / 47
International IVD application team
For application training only V2.0
Step 2: Read the information on the pumped window carefully and click ‘OK’ to continue.
39 / 47
International IVD application team
For application training only V2.0
middle. For example, in the picture above, there are 6 samples with the initial position 5 and
finished at position 11. The samples should be placed in the correct order from 6 to 11.
(b) All the samples should share the same test panel. And there test items should be exactly the
same.
(c) The patient information cannot be input directly but should be edited individually after
request.
For example, Mindray ALT reagent insert paper clearly declare that the linearity range of this
reagent is 4-1000 U/L, which means when the real substrate concentration is within 4-1000 U/L,
mindray can offer the end user a guaranteed and accurate results.
However, if the sample substrate is exceeded this declared range, the results are not making any
sense and cannot be printed in the final patient reports.
The most common method to avoid the linearity problem for those high concentration samples is
to dilute with certain ratio and finally multiply back with the ratio number in the final results.
40 / 47
International IVD application team
For application training only V2.0
41 / 47
International IVD application team
For application training only V2.0
When the daily operation and tests are all finished, please follow the steps below to switch off
the instrument. (If there is any unfinished test, the request will be invalided after exit software)
Fig 6.1-1
(3) Replace the used cuvettes with the clean one. This step is optional. Please click ‘OK’ to
replace the cuvettes or click ‘Next’ to skip this step and exit software directly.
Fig 6.1-2
Please kindly turn off the switch on the left or right hand side of machine and turn the main
power on to keep the refrigerator working all the time. If the electricity supplement cannot be
guaranteed in the continuous and stable manner, please make sure the reagents are placed in the
commercial refrigerator in advance.
Please make sure the front cover on the top should be always closed to avoid any dust or other
contaminators.
Please avoid any contact with the reaction liquid when change the cuvettes or empty the waste
tank, and those liquid might contain the harmful virus or bacteria inside and should be labelled as
biohazard material if necessary.
42 / 47
International IVD application team
For application training only V2.0
Please immediately rinse the contacted skin if get any spillage of liquids from cuvettes or waste
tank, and then go to doctors’ for the further treatment if necessary.
43 / 47
International IVD application team
For application training only V2.0
7. Easy maintenance
44 / 47
International IVD application team
For application training only V2.0
45 / 47
International IVD application team
For application training only V2.0
46 / 47
International IVD application team
For application training only V2.0
For the further information, please kindly contact the service engineers team.
47 / 47
International IVD application team