Optima MS: Technical Manual
Optima MS: Technical Manual
MANUAL
TECHNICAL MANUAL
OPTIMA MS
Optima_ms_enTDM.fm 3
4.3.9 Checking the analogue sensor input........................................................................ 35
4.3.10 Checking the software version ................................................................................. 36
4.3.11 Checking the motor clamp ....................................................................................... 36
4.3.12 Checking the calibration values ............................................................................... 37
4.3.13 Checking the infusion motor..................................................................................... 37
4.3.14 Checking the last 10 events before a locking alarm or error .................................... 38
4.3.15 Checking the pressure alarm limit ............................................................................ 38
4.3.16 Flow rate check ........................................................................................................ 39
4.3.17 Air detector check .................................................................................................... 39
4.3.18 Battery autonomy check........................................................................................... 39
4.3.19 Continuity test .......................................................................................................... 40
4.3.20 Contrast check ......................................................................................................... 40
4.3.21 Timestamp check ..................................................................................................... 40
4.3.22 Quality Control Certificate ........................................................................................ 41
4.4 Cleaning and disinfection ................................................................... 43
4.5 Storage.................................................................................................. 44
5 Troubleshooting.............................................................................................. 45
5.1 Breakdown guide ........................................................................................... 45
5.2 Error messages ............................................................................................. 49
6 Intervention procedures................................................................................. 53
N°1, Procedure: CPU board ....................................................................... 55
N°2, Procedure: Display board .................................................................. 59
N°3, Procedure: Pumping system ............................................................. 63
N°4, Procedure: LCD display .................................................................... 67
N°5, Procedure: Air detector board........................................................... 71
N°6, Procedure: Tube clamp switch.......................................................... 75
N°7, Procedure: Pressure sensor.............................................................. 79
N°8, Procedure: OCS clamp motor ........................................................... 85
N°9, Procedure: Power supply board ....................................................... 89
N°10, Procedure: Battery ........................................................................... 91
N°11, Procedure: Front cover, base and angle bracket .......................... 93
7 Calibration ....................................................................................................... 97
7.1 Calibration procedure.......................................................................... 97
7.1.1 Access to the calibrations ........................................................................................ 97
7.1.2 EtAL1 Door calibration ........................................................................................ 99
7.1.3 EtAL2 Calibration of the flow rate corrective coefficient.................................... 100
7.1.4 EtAL9 Calibrating the upstream and downstream pressure sensors................ 101
44 Optima_ms_enTDM.fm
8 Spare parts catalogue .................................................................................. 105
8.1 Upper cover ........................................................................................ 105
8.2 Base..................................................................................................... 107
8.3 Optima pump angle bracket .............................................................. 109
Optima_ms_enTDM.fm 5
66 Optima_ms_enTDM.fm
1 Presentation
1.1 General presentation
The Optima pump is a new pump based on the same pumping mechanism as the MVP
pump. It is used for infusion of intravenous agents.
There are four types of the Optima pump:
MS: for specific Fresenius Vial PVC and PVC free tubes. The drop detector is optional. The
pump is equipped with an automatic occlusion checking system (OCS).
The Optima pump has its own control keyboard and distinct displays. It has an internal
battery which ensures a minimum autonomy of 4 hours for a flow rate of 125 ml/h.
A choice of easily accessible configurations enables optimal use of the functions according to
the needs of each department.
Presentation
optima01_001a_en.fm 7
1.2 Operation diagram
Battery
Keyboard RS 232
Power supply nurse call
Display PCB
Connector
Air Dripsensor
bubble Keyboard / Display (optional)
sensor UC
Motor
Rotation
controller
Peristatic mechanism
Transmission
Door sensor
Door
Lock
Occlusion sensor
Clamp detector
Luer Derivation Down-stream extension Pump Clamp Up-stream extension Drip chamber
Perfusion set
Presentation
8 optima01_001a_en.fm
1.3 Precautions for use
The symbol ! which appears on the device’s condensed user instructions recommends a
complete reading of the user instructions, in accordance with norm EN 60 601-1.
Fresenius Vial cannot, in any circumstances, be held responsible for any medical or other
problem due to incorrect use of the device.
Kindly consult the user manual for more details.
optima01_001a_en.fm 9
1.5.4 Conformity and norms
Conforms with the European Directive on Medical Materials 93/42 EEC: CE0459.
Conforms with the European Directive 89/336/CEE: Electromagnetic compatibility.
Conforms with norms EN 60601.1 and PrEN60601-2-24.
Protection against current leakage: type CF.
Protection against electric shocks: Class I.
Watertightness: IP31.
Presentation
10 optima01_001a_en.fm
2 Description and operation
2.1 Physical description
The Optima pump is made out of a front cover, a base and an angle bracket.
The front cover contains:
a display board linked to the control panel,
a CPU board,
an air detector board,
the OCS system.
optima02_001a_en.fm 11
2.1.1 Display board and front panel
The display board is situated under the control panel and contains all the essential parts for
the man-machine interaction:
keyboard,
synoptic and control indicators,
7-segment displays,
LCD screen.
Pin n°1
J1
Pin n°1
J2
J3
Pin n°1
11 GND Ground
12 CSAFCL Display selected
13 GND Ground
14 A0AFF LCD address
15 GND Ground
16 RESET/
17 GND Ground
12 optima02_001a_en.fm
J1 connector to CPU board
J2 connector to LCD
J3 connector to keyboard
optima02_001a_en.fm 13
2.1.2 The CPU board
The CPU board contains a 80C32 micro-processor. It is connected to the display board by
means of the J1 connector.
J1 J2 J4
J3
Pin n°1
J7
J6
J9 J10
J8
J11 J12
CPU board
Info: Pin 1 of connector J1 is opposite to that indicated by the CPU board screen printing.
J1 display board connector
16 RESET/
17 GND Ground
18 GNDTON/OFF Signal ON/OFF
19 VBAT Battery voltage
20 5V Power supply
14 optima02_001a_en.fm
J12 drip detector connector
Pin Signal Description
1 Pulse-drip Low signal yield
2 GND Ground
3 GND Ground
4 CDLED-drip Control LED 5V/ 40 mA
J4 motor connector
Pin Description
1 +Reel A
2 - Reel A
3 +Reel B
4 - Reel B
J6 power supply connector
J8 battery connector
Pin Description
1 + DC motor
2 - DC motor
optima02_001a_en.fm 15
J2 sensors connector
J9 air detector connector
16 optima02_001a_en.fm
2.1.3 The power supply board and battery
The power supply board is positioned on the angle bracket. It supplies power to the electronic
unit from the 230 VAC network. It also ensures a battery charge of 2.7/3 Ah.
J6
J4
J5
J3
F1 F2
J1 J2
J2 Mains connector
Pin Description
1 PH Phase (100 - 240 Vac)
2 N Neutral (100 - 240 Vac)
Pin Description
1 9V Sect 9V output convertor AC/DC
2 GND
3 9V Ext 9V external power supply not used
Description and operation
Pin Description
1 +Vbat
2 +Thermal sensor
3 GND
4 VBAT NP
optima02_001a_en.fm 17
J6 battery input connector
Pin Description
1 + VBAT
2 VBAT NP
3 + Thermal sensor
4 - Thermal sensor
5 GND
Description and operation
18 optima02_001a_en.fm
2.2 Operational description
From an operational point of view, the Optima pump is made up of 3 sub-assemblies:
a sub-assembly to maintain to check the correct positioning of the tube,
a motorisation sub-assembly,
an external connection sub-assembly.
optima02_001a_en.fm 19
Description and operation
20
optima02_001a_en.fm
3 Description of the menus
3.1 Parameter configuration menu
The configuration menu adapts the Optima to the particular needs of each department. It
accesses the menus that customise the infusion parameters.
Fresenius Vial recommends the presence of its qualified staff, or a member of your Technical
Department, who will assist you in implementing the configuration procedures you wish to
select.
The configuration mode is activated by holding down the "MODE" key when the device
is switched on.
The following menus are displayed:
User setting,
Wards setting,
Maintenance.
Each menu consists of sub-menus. The sub-menus can be accessed by using the
"START" key.
To access some of the sub-menus, you need to type in an access code. This code can be
defined and/or de-activated in the Service configuration mode. As long as the device is kept
switched on, the code does not need to be keyed in again.
optima03.1_001a_en.fm 21
Test Test1 operating period and maintenance date
Maintenance
Test2 LED, LCD screen, 7-segment display and buzzer test
Test3 keyboard test
Test4 display’s the battery voltage
Test5 display’s the last 10 alarms codes
Test6 display’s the total running time
Test9 display’s the analogue sensor values
TestA display’s the software version
TestC clamping motor test
TestE display’s the calibration values
TestF infusion motor test
TestJ display’s the code last 10 alarm codes before error
Configuration SAV1 preventive maintenance
SAV2 tube selection
Calibration Etal1 door calibration
Etal2 corrective infusion coefficient
Etal9 upstream and downstream pressure sensors calibration
22 optima03.1_001a_en.fm
3.1.1 Access to the menu
Key Function
ON switches on the device.
OFF switches off the device and exits the configuration mode when
pressed for more than three seconds.
SILENCE ALARM
The selection keys scroll through the numbers and letters, the tenths,
units and tens values.
Activate or de-activate choices.
ENTER moves the cursor.
By pressing "STOP" you can cancel the modification and return to the After-
sales Service configuration menu.
You can exit the configuration mode at any time by pressing the "OFF" key.
Description of the menus
optima03.1_001a_en.fm 23
3.1.3 SAV2, tube selection configuration
The tubes which have already been calibrated for pressure are marked with a "v".
By pressing "STOP", the modification is cancelled and you can return to the
After-sales Service configuration menu.
You can exit the configuration mode at any time by pressing the "OFF" key.
Description of the menus
24 optima03.1_001a_en.fm
Description of the menus
The calibration menu is reserved for qualified staff only. Its access is protected by a
secret code. To find out about the operating mode for the various calibrations, refer
to the "Calibrations" chapter.
optima03.3_001a_en.fm 25
Description of the menus
26
optima03.3_001a_en.fm
Description of the menus
The After-Sales Service tests can also be carried out more quickly and simply from a PC
using our maintenance software. (Contact our After-Sales Service department.)
A log of more than 760 time-stamped events can be consulted using our maintenance
software "VPControl" version 2.1 or later versions. (Contact our After-Sales Service
department.)
optima03.4_001a_en.fm 27
Description of the menus
28
optima03.4_001a_en.fm
4 Preventive maintenance
4.1 Recommendations
The Optima volumetric pump can only be checked, serviced or repaired by Fresenius Vial or
by a certified and approved maintenance service. Any abnormal functioning of the device
must be brought to the attention of your in-house qualified technical personnel or our After-
Sales Service.
Should a repair be necessary, kindly send the device, in its original packaging if possible, and
accompanied by a precise description of the fault, to the official Fresenius Vial
representative.
For all information concerning the repair and use of the device kindly contact our After-Sales
Service or our Sales department.
Fresenius Vial is not liable for loss or damage to the device during its transport to our After-
Sales Service.
Frequency Description
12 months Carry out a regular service inspection.
3 years Replace the battery (see procedure sheet “Door-battery and battery”).
Frequency Description
1st inspection in Carry out the first service inspection.
3rd year
Then every 2 Carry out a service inspection.
years
3 years Replace the battery (see procedure sheet “Door-battery and battery”).
Preventive maintenance
optima04.1&2_001a_en.fm 29
Preventive maintenance
30 optima04.1&2_001a_en.fm
Preventive maintenance
4.3 Checks
In order to ensure the preventive maintenance follow-up, a regular servicing inspection is
recommended every 12 months (see "Quality Control Certificate").
Recharge the battery during 16h before starting the checking procedure !
Keyboard description.
Key Function
ON switches on the device.
OFF switches off the device when pressed for over three seconds.
SILENCE ALARM
The selection keys scroll through the numbers and letters on the tenths,
units and tens values, and activate or de-activate the infusion menus.
optima04.3_001a_en.fm 31
4.3.3 Operating period and maintenance date
This test displays teSt.1, press "START".
the Optima’s
running time since
The operating time and the last service inspection date
its last servicing
are displayed on the LCD screen.
inspection. It also
If "Preventive test" flashes, service inspection is required.
displays and
modifies the "last"
servicing
inspection date.
When the servicing
inspection date is
modified, the
operating time is
resetted.
You can interrupt the test at any time by pressing "STOP" and returning to
the test selection.
You may interrupt the test at any time by pressing "STOP" and returning
to the test selection.
If the display is faulty, replace the display board (see "Display Board procedure).
Preventive maintenance
32 optima04.3_001a_en.fm
4.3.5 Keyboard check
This test checks teSt.3, press "START".
that the keyboard
keys are
functioning The following screen appears:
correctly. Press each key one by one,
The keys that have been pressed appear as a
reverse video image on the LCD screen.
If two keys or more are held down simultaneously, "beep! beep! beep!" will be heard,
indicating an error.
optima04.3_001a_en.fm 33
Alarm or error codes description:
Alarm Description
A10 Battery alarm
A12 End of infusion alarm
A16 Downstream occlusion alarm
A17 Air detection alarm
A18 Air volume alarm
A19 Upstream occlusion alarm
A20 Tube absence alarm
A21 Mobile clamp absence alarm.
A22 Door alarm
A23 Drip detector absence alarm
A24 Drip detector alarm: under flow
A25 Drip detector alarm: over flow
A26 Free flow
A27 Occlusion problem during auto-set
OFF Normal stop
FAIL Abnormal stop
34 optima04.3_001a_en.fm
Error Type Description
85 Locking Restart µC secondary
16 Locking Date and time in µC secondary verification
56 (1) Association condition output/step number
17 Locking Keyboard matrix (key press and columns inactive)
27 Locking Keyboard check default (path not fine tuned)
37 (2) ON/OFF key locked
47 Fatal Switching off impossible
28 Locking Language file incompatible
(1) Type of error depends on the context.
(2) Specific treatment after these errors have been generated.
(3) Error detected on stand-by and displayed at next switched on.
If the value is out of limits, recalibrate the force sensor. (See "EtAL9 upstream and
downstream pressure sensors calibration".)
Preventive maintenance
optima04.3_001a_en.fm 35
4.3.10 Checking the software version
This test displays teSt.A, press "START".
the software
version and
The LCD screen displays:
revision numbers.
the name of the Optima pump.
the software version.
the date.
Press the tenths keys to continue the test.
You may interrupt the test at any moment by pressing "STOP" and
returning to the test selection.
Preventive maintenance
36 optima04.3_001a_en.fm
4.3.12 Checking the calibration values
This test displays teStE, press "START".
the calibration
values recorded in
the EEPROM. Move to another value using the keys.
Display Description
Por1 Hall effect sensor, door closed without tube.
The value between brackets after the title is the number of calibrations carried out. The
display of the upstream and downstream pressure values is carried out according to the
number of active tubes.
You may interrupt the test at any time by pressing "STOP" and return to
the maintenance menus.
Preventive maintenance
optima04.3_001a_en.fm 37
4.3.14 Checking the last 10 events before a locking alarm or error
This test enables teSt.J, press "START".
the visualisation of The number of times the machine has been switched on
the codes for the since the last locking error is displayed on the LCD
last 10 events that screen.
have occurred
"A" for an alarm. (See table teSt.5)
during operation of
the device before "E" for an error. (See table teSt.5)
the last locking
error. By validating once again, a new test may be seleted.
1 m 50cm
PU M P
M a n o m e te r
T a p 3 w a ys
C o n ta in e r
Press "ON".
Position a tube in the device and lock the door inserting
the clamp.
Position the manometer at the outlet of the tube.
Select a flow rate of 125 ml/h.
Adjust the pressure threshold to 750 mmHg.
Let the tap run in a container.
Start up the infusion and leave it running for 30
seconds.
Position the tap to the manometer.
Check that the alarm is triggered for a value of 1 bar ±
0.15 bar.
Stop the infusion by pressing "STOP".
Preventive maintenance
If the measured pressure value, depending to the limit selected is out of range, recalibrate
the pressure sensors. (See"ETAL9 Calibration of the upstream and downstream
pressure sensors".)
38 optima04.3_001a_en.fm
4.3.16 Flow rate check
Connect the Optima pump to the mains.
Before any flow rate test, run out an infusion of 20 ml to stabilise the flow rate.
To carry out this test, we recommend to use the IS Flow rate software available with
Fresenius Vial.
optima04.3_001a_en.fm 39
4.3.19 Continuity test
Use a multimeter.
To carry out this procedure, exit the test mode. Press "OFF.
Using a calibrated multimeter in ohms, check the presence of an electrical resistance
greater than 10 MΩ between:
phase and subD 15 pins connector (with any pin)
neutral and subD 15 pins connector (with any pin) .
40 optima04.3_001a_en.fm
Preventive maintenance
Conformity
N° Procedure Obtained value
Yes No
1 Check the general appearance of the case, labels, serial number and product
code.
2 Check the contrast.
If a fault is observed, refer to chapter ’Contrast Check’.
3 Check the clock. If an adjustment is needed, refer to chapter on datestamp
checking.
4 Display the operating period, teSt1and the maintenance date.
5 Check the indicator lights, teSt2.
6 Check the keyboard, teSt3.
7 Display the total running time, teSt6.
8 Occlusion check, teStF
Connect the water manometer to the tube outlet.
Start up test F.
Check that there is no motor error and that the maximum pressure is above1.6
bars (1216 mmHg).
9 Check clamping and unclamping, teStC. .
Confirm the two functions.
Check for any absence of motor alarm.
10 Check the drop detector (if the detector is present).
Start up an infusion.
Check the (green) drop indicator lights up each time a drop appears in the drop
chamber.
Disconnect the detector.
Check if the drop alarm is activated.
11 Check the air detector.
Create an air bubble in the tube of at least 5 cm long.
Check the presence of the air bubble alarm (sound and visual) while the air
bubble is in the air bubble detector.
12 Check the downstream pressure.
Install a calibrated water manometer at the tube outlet.
Adjust the pressure threshold to 750 mmHg.
Preventive maintenance
optfiche_001a_en.fm 41
Conformity
N° Procedure Obtained value
Yes No
13 Check the upstream pressure.
Start an infusion at 100 ml/h without drop detector.
Clamp the tube between the drop chamber and the pump.
Check that the upstream occlusion alarm is activated within 10 seconds.
Unclamp.
Check that the confirmation indicator lights up.
14 Check the clamp.
Close the door without the tube
Check if the message "Insert tube" is displayed.
Insert the tube without the clamp and close the door.
Check if the message "Insert clamp" is displayed.
Insert the clamp and close the door.
Check if the previous message has disappeared.
15 Check the battery operation.
Disconnect the mains lead.
Check the presence of the battery indicator.
Reconnect the mains lead.
Check if the battery indicator is switched off and that the mains indicator lights
up.
16 Check the flow rate.
Before any flow rate test, run an infusion of 20 ml to stabilise the flow rate.
Carry out a flow rate check in accordance with the norm PREN-60-601-2-24.
Check that the flow rate error is between - 5% and + 5%.
If there is too much discrepancy, refer to chapter 7 "Calibration / Calibration
procedure / EtAL2 calibrating the flow rate corrector coefficient".
17 Check the battery autonomy.
Start an infusion at 125 ml/h for 4 hours.
18 Carry out an electrical safety test according to norm EN 60601-1
Observations:
Preventive maintenance
42 optfiche_001a_en.fm
Diagnosis
¨TRICHLOROETHYLENE-DICHLOROETHYLENE:
AMMONIA.
AMMONIUM CHLORIDE
CHLORINE AND AROMATIC HYDROCARBON.
ETHYLENE DICHLORIDE-METHYLENE CHLORIDE
CETONE based cleaning products.
These aggressive agents could damage the plastic parts and lead to device malfunctions.
Also beware of ALCOHOL SPRAYS (20% -40% alcohol) that tarnish and crack the
plastic and fail to provide the cleaning action required prior to disinfection.
For further information, please contact the competent department in your hospital for the
supply of the appropriate cleaning and disinfecting products.
Preventive maintenance
optima04.5&6_001a_en.fm 43
4.5 Storage
In case of prolonged storage, it is advisable to disconnect the battery using the battery access
door on the bottom of the device. This operation should be done by an experienced
technician.
The equipment must be stored in a dry and cool place.
The recommended environmental temperature conditions for storage of the equipment are
between 0°C and 40°C.
Relative humidity tolerated: max. 85%, no condensation.
Fully recharge the battery before using the equipment to avoid any risks caused by micro
power cuts in the mains supply and to ensure maximum autonomy.
Preventive maintenance
44 optima04.5&6_001a_en.fm
5 Troubleshooting
5.1 Breakdown guide
Message on
Meaning Causes Action
LCD display
Door open Hall affect sensor Check set installation
alarm detects that the door Check the presence of the
is open door magnet
Calibration of the door
sensor
Replace the door sensor
Clamp The Optima does Check set installation
position not detect the Check the securing clamp
alarm presence of the set micro switch
securing clamp Replace it if necessary
Set insertion Wrong position of Check set installation
alarm the set Check the presence of the
door magnet
Calibration of the door
sensor
Replace the door sensor
OCS test Occlusivity test Check set installation
Alarm not successful. The Check the door and carry
pumping mechanism on a calibration if necessary
does not succeed to Verify the good condition of
increase the the membrane. Replace it if
pressure in the IV necessary; refer to the
set during the OCS maintenance protocol : PM
test 0094
Air bubble Air volume Prime the line to remove the
alarm detected is higher air
Than the limit Check set installation on the
programmed air detector Verify the air
Set is not adjustment of the air size
properly inserted in programmed
the air detector Replace the air detector or
air bubble board
Troubleshooting
optima05.1_001a_en.fm 45
Message on
Meaning Causes Action
LCD display
Drop alarm The container is Verify the volume remaining
(If drop empty on the container
detector is The rolling clamp Verify if the roller clamp is
present) is closed open
Over-flow , under Verify if the air inlet cap if
flow, free flow open when the pump is used
The position of with a bottle
the drip sensor Verify the positioning of the
Drop chamber is drop detector
full Verify the volume of the
Drop detector is liquid in the drop chamber
disconnected while Verify the IV set position
the pump is Verify the drip sensor is well
configured to be connected
used with Verify the connector if it's
The drip sensor damaged
is connected and Replace the drip sensor
disconnected
Free flow Over-flow Check set installation
alarm Set not properly Check drip sensor
(in case of positioned positioning
use with drip
sensor)
46 optima05.1_001a_en.fm
Message on
Meaning Causes Action
LCD display
Air alarm The IV set is Remove the IV set and re-
present on the air install it
bubble detector but Carry on a calibration of the
not on the pumping door
mechanism
The pump runs
over 24 hours, no air
bubble have
detected by the
sensor
Battery Low battery Connect the pump to the
alarm voltage mains power
Check the charging voltage,
pump connected to the mains.
Replace the battery
Preventive maintenance Preventive Preventive Proceed to the complete
24/10/2003 passed
maintenance maintenance date or check up of the pump and
requested time is passed update the maintenance date
or time ( test 1)
does not switch Power supply out of order Replace the power supply board
ON
If the problem remains, replace the
CPU board
optima05.1_001a_en.fm 47
Problems Causes Action
Battery alarm Battery is totally Check the charging voltage
while it has discharged Replace the battery
been
completely
charged
Battery alarm Power supply or CPU Check the charging voltage
whereas the boards are faulty Check the power supply mains fuses
Optima is
Replace the power supply if
connected to
necessary.
the mains
If the problem remains, replace the
CPU board and carry on a complete
configuration
The module Keyboard is defective Check the keyboard ( test 3)
MVP turns ON Replace the keyboard if necessary
or turns OFF CPU board is defective Check that there was no liquid
alone
infiltration on the CPU board
Replace the CPU board and carry
on a complete configuration of the
device
Some keys Keyboard is defective Check the keyboard (test 3)
from keyboard Replace the keyboard if necessary
out of order
48 optima05.1_001a_en.fm
Troubleshooting
Er11 Stepper motor rotation error Check the motor coupling - See PM
0094
Er21 Stepper motor rotation error Replace the mechanical block - See
PM 0094
Er31 OCS DC motor failure Check the motor clamp for OCS and
rubber O ring
Put some grease on the clamp finger
threaded screw. Replace the rubber O
ring if necessary (p/n: 174256)
Troubleshooting
optima05.2_001a_en.fm 49
Error codes Meaning Action
Er14 Motor period calculation Reconfigure the Optima pump
anomaly Replace the CPU board
Er16 Date / time failure Check the date and the time clock
Replace the CPU board and Carry
on a complete configuration of the
device via VP Control software
Er47 Switching off impossible Replace the CPU board and carry
on a complete configuration of the
device via VP Control software
50 optima05.2_001a_en.fm
Error codes Meaning Action
Troubleshooting
optima05.2_001a_en.fm 51
Troubleshooting
52 optima05.2_001a_en.fm
6 Intervention procedures
This chapter lists all the dismantling and re-assembling procedures.
Intervention procedures
optima06.0_001a_en.fm 53
Intervention procedures
54 optima06.0_001a_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
1 Posidriv Z1 screwdriver
1 anti-static bracelet
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure:
Access
Pivot the Optima pump on the top of the front cover.
Unscrew the 2 cruciform screws (marker 1), connecting the Optima angle bracket to the
base.
Remove the angle bracket being careful not to pull out the connectors.
Disconnect the J4 and J3 connectors, as well as the functional Earth.
Return the Optima pump to its original position.
1
2 2
optima06.0_0010a_en.fm 55
Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig. 2
Dismantling
J1
3
J9 J10 J6
J8
Intervention procedures
Remove the board from its housing, being careful with the display board J3 flat cable.
Remove the CPU board.
56 optima06.0_0010a_en.fm
Re-assembling
Carry out the dismantling procedure in reverse order to re-assemble the unit.
When closing the pump, ensure that the 2 cables connecting the power supply board and
the CPU board are positioned above the motor. In addition, check that the earth wire is
next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when the
pump is in operation.
Positioning of the earth wire.
Incorrect
Correct
Positioning of the wires connecting the power supply board and the CPU board.
Right Wrong Intervention procedures
optima06.0_0010a_en.fm 57
Wrap an adhesive silicon cord around the CPU board J12 and J11 connectors.
Each Optima pump has its own type of "display and CPU board"; it is important that you
do not mix up the different Optima pump references and that you order the one that
corresponds to your device.
Carry out regular servicing checks. (See Quality Control Certificate sheet.)
Intervention procedures
58 optima06.0_0010a_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
1 Posidriv Z1 screwdriver
1 anti-static bracelet
Maintenance level:
Level 2, specialist technician (see biomedical technical training documentation).
Procedure:
Access
Pivot the Optima pump on the top of the front cover.
Unscrew the 2 cruciform screws (marker 1) connecting the Optima pump angle bracket to
the base.
Remove the angle bracket, being careful not to pull out the connectors.
Disconnect the J4 and J3 connectors, as well as the functional Earth.
Return the Optima pump back to its original position.
1
1
2
2
Using an Allen key, gently slide out the two screws located under the door.
Remove the door.
Unscrew the 4 cruciform screws (marker 2 on Fig. 1 and 2) connecting the front cover to
the base.
Remove the front cover, being careful not to pull out the CPU board flat cable.
optima06.0_0020a_en.fm 59
Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig. 2
Dismantling
Disconnect the J1 connector connecting the display board to the CPU board.
Disconnect the display board J3 connector.
Unscrew the cruciform screw (marker 3) situated on the display board, connecting it to the
front cover.
Remove the display board.
J3
J1
Intervention procedures
J2
60 optima06.0_0020a_en.fm
Re-assembling
Each Optima pump has its own type of "display and CPU board"; it is important that you
do not mix up the different Optima pump references and that you order the one that
corresponds to your device.
Carry out the dismantling procedure in reverse order to re-assemble the new unit.
When fixing the display board, ensure that the plastic inserts are not tightened too much
or they will be damaged.
When positioning the front case, ensure that the sealing is perfectly positioned in its
housing after re-assembling.
When closing the pump, ensure that the 2 cables connecting the power supply board and
the CPU board are positioned above the motor. Also check that the earth wire is next to
the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when the
pump is in operation. (See photo on procedure sheet n°1.)
Carry out regular servicing checks. (See Quality Control Certificate sheet.)
Intervention procedures
optima06.0_0020a_en.fm 61
Intervention procedures
62 optima06.0_0020a_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
1 Posidriv Z1 screwdriver
1 anti-static bracelet
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure:
Access
Pivot the Optima pump on the top of the front cover.
Unscrew the 2 cruciform screws (marker 1), connecting the Optima pump angle bracket to
the base.
Remove the angle bracket being careful not to pull out the connectors.
Disconnect the J4 and J3 connectors, as well as the functional Earth.
Return the Optima pump back to its original position.
1
1
2 2
Using an Allen key gently slide out the two screws located under the door from their
Intervention procedures
groove.
Remove the door.
Unscrew the 4 cruciform screws (marker 2 on Fig. 1 and 2) connecting the front cover to
the base.
Remove the front cover, being careful not to pull out the CPU board flat cable.
optima06.0_0030a_en.fm 63
Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig.2
Dismantling
Unscrew the 3 cruciform screws (marker 3) connecting the pumping system to the base.
3
Intervention procedures
64 optima06.0_0030a_en.fm
Using a cutter, remove the label.
Unscrew the 2 middle cruciform screws (marker 4) located under the label.
4 3
4
3
5 5
Re-assembling
Carry out the dismantling procedure in reverse order to re-assemble the new unit.
When closing the pump ensure that the 2 cables connecting the power supply board and
the CPU board are positioned above the motor. In addition, check that the earth wire is
Intervention procedures
Carry out regular servicing checks. (See Quality Control Certificate sheet.)
optima06.0_0030a_en.fm 65
Intervention procedures
66 optima06.0_0030a_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
1 Posidriv Z1 screwdriver
1 anti-static bracelet
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure:
Access
Pivot the Optima pump on the top of the front cover.
Unscrew the 2 cruciform screws (marker 1), connecting the Optima pump angle bracket to
the base.
Remove the angle bracket being careful not to pull out the connectors.
Disconnect the J4 and J3 connectors, as well as the functional Earth.
Return the Optima pump to its original position.
1
2 2
optima06.0_0040a_en.fm 67
Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig. 2
Dismantling
68 optima06.0_0040a_en.fm
Re-assembling
Recover the self-adhesive block from the old LCD.
Position the self-adhesive block on the LCD display, above the LCD display label.
Position the LCD display on the display board.
Press down the display on the display board so that the block adheres to the display
board.
Carry out the dismantling procedure in reverse order to re-assemble the new unit.
When closing the pump, ensure that the 2 cables connecting the power supply board and
the CPU board are positioned above the motor. In addition, check that the earth wire is
next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when the
pump is in operation. (See photo on Procedure n°1.)
.
Carry out regular servicing checks. (See Quality Control Certificate sheet.).
Intervention procedures
optima06.0_0040a_en.fm 69
Intervention procedures
70 optima06.0_0040a_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
1 Posidriv Z1 screwdriver
1 anti-static bracelet
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure:
Access
Pivot the Optima pump on the top of the front cover.
Unscrew the 2 cruciform screws (marker 1), connecting the Optima pump angle bracket to
the base.
Remove the angle bracket being careful not to pull out the connectors.
Disconnect the J4 and J3 connectors, as well as the functional Earth.
Return the Optima pump to its original position.
1
2 2
optima06.0_0050a_en.fm 71
Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig. 2
Dismantling
Unclip the non-oxidizing aluminium plate (marker 3) holding the air detector board in
place.
Unclip the board.
3
Intervention procedures
72 optima06.0_0050a_en.fm
Disconnect the J9 connector (marker 4) connecting the CPU board to the air detection
board.
Disconnect the connector (marker 5) connecting the board to the air detector.
5
4
Re-assembling
Carry out the dismantling procedure in reverse order to re-assemble the new unit.
When closing the pump, ensure that the 2 cables connecting the power supply board and
the CPU board are positioned above the motor. In addition, check that the earth wire is
next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when the
pump is in operation. (See photo in Procedure n°1.)Carry out regular servicing checks.
Carry out regular servicing checks. (See Quality Control Certificate sheet.)
Intervention procedures
optima06.0_0050a_en.fm 73
Intervention procedures
74 optima06.0_0050a_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
1 Posidriv Z1 screwdriver
1 anti-static bracelet
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure:
Access
Pivot the Optima pump on the top of the front cover.
Unscrew the 2 cruciform screws (marker 1), connecting the Optima pump angle bracket to
the base.
Remove the angle bracket being careful not to pull out the connectors.
Disconnect the J4 and J3 connectors, as well as the functional Earth.
Return the Optima pump to its original position.
1
2
2
optima06.0_0060a_en.fm 75
Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig. 2
76 optima06.0_0060a_en.fm
Dismantling
4
4
Re-assembling
Carry out the dismantling procedure in reverse order to re-assemble the unit.
Intervention procedures
When closing the pump, ensure that the 2 cables connecting the power supply board and
the CPU board are positioned above the motor. In addition, check that the earth wire is
next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when the
pump is in operation. (See photo in Procedure n°1.)
Carry out regular servicing checks. (See Quality Control Certificate sheet.)
optima06.0_0060a_en.fm 77
Intervention procedures
78 optima06.0_0060a_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
1 Posidriv Z1 screwdriver
1 anti-static bracelet
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure:
Access
Pivot the Optima pump on the top of the front cover.
Unscrew the 2 cruciform screws (marker 1), connecting the Optima pump angle bracket to
the base.
Remove the angle bracket being careful not to pull out the connectors.
Disconnect the J4 and J3 connectors, as well as the functional Earth.
Return the Optima pump to its original position.
1
1
2
2
optima06.0_0070a_en.fm 79
Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig. 2
80 optima06.0_0070a_en.fm
Dismantling
2
3
2
4 4
5
Intervention procedures
optima06.0_0070a_en.fm 81
Separate the pressure sensors from their support (see arrows)
Unsolder the pressure sensor wires (marker 6).
Remove the pressure sensors.
6
Intervention procedures
82 optima06.0_0070a_en.fm
Re-assembling
Carry out the dismantling procedure in reverse order to re-assemble the unit.
After re-assembling the pressure sensors, calibrate them. (See ETAL9, Calibrating
the upstream and downstream pressure sensors.)
When closing the pump, ensure that the 2 cables connecting the power supply board and
the CPU board are positioned above the motor. In addition, check that the earth wire is
next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when the
pump is in operation. (See photo in Procedure n°1.)
Carry out regular servicing checks. (See Quality Control Certificate sheet.)
Intervention procedures
optima06.0_0070a_en.fm 83
Intervention procedures
84 optima06.0_0070a_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
1 Posidriv Z1 screwdriver
1 anti-static bracelet
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure
Access
Pivot the Optima pump over the top of the front cover.
Unscrew the 2 cruciform screws (marker 1), connecting the Optima angle bracket to the
base.
Remove the angle bracket being careful not to pull out the connectors.
Disconnect the J4 and J3 connectors, as well as the functional Earth.
Return the Optima pump to its original position.
2 1
2
optima06.0_0080a_en.fm 85
Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig. 2
86 optima06.0_0080a_en.fm
Dismantling
4
3
4
4
4
Intervention procedures
optima06.0_0080a_en.fm 87
Disconnect the J10 connector (CPU board).
Unscrew the clamp (marker 6), while holding the motor reducer.
Unscrew the 3 cruciform screws (marker 5) holding the
motor reducer to the front cover.
Remove the motor reducer.
6
5
5
5
Re-assembling
Carry out the dismantling procedure in reverse order to reassemble the unit.
When re-assembling the motor clamp, apply a drop of silicon grease around the screw
path (marker 6).
When closing the pump, ensure that the 2 cables connecting the power supply board and
the CPU board are positioned above the motor. In addition, check that the earth wire is
next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when the
pump is in operation. (See photo in Procedure n°1.)
Carry out regular servicing checks. (See Quality Control Certificate sheet.)
Intervention procedures
88 optima06.0_0080a_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
1 Posidriv Z1 screwdriver
1 anti-static bracelet
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure
Access
Pivot the Optima pump on the top of the front cover.
Unscrew the 2 cruciform screws (marker 1), connecting the Optima pump angle bracket to
the base.
Remove the angle bracket being careful not to pull out the connectors.
Disconnect the J4 and J3 connectors, as well as the functional Earth.
1
Intervention procedures
optima06.0_0090a_en.fm 89
Unscrew the 2 cruciform screws (marker 4) holding the power supply shelf in place.
Disconnect the earth wire (marker 5) and the mains connector (marker 6).
Remove the power supply board.
Re-assembling
Carry out the dismantling procedure in reverse order to re-assemble the unit.
When closing the pump, ensure that the 2 cables connecting the power supply board and
the CPU board are positioned above the motor. In addition, check that the earth wire is
next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when the
pump is in operation. (See photo in Procedure n°1.)
Carry out regular servicing checks. (See Quality Control Certificate sheet.)
.
Intervention procedures
90 optima06.0_0090a_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
1 Posidriv Z1 screwdriver
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure:
Access
Pivot the Optima pump on the top of the front cover.
Unscrew the 2 cruciform screws (marker 1) connecting the Optima pump angle bracket to
the base.
Remove the angle bracket, being careful not to pull out the connectors.
1
Intervention procedures
optima06.0_0100a_en.fm 91
Dismantling
Unscrew the 2 cruciform screws (marker 2) holding the battery in place.
Disconnect the battery.
Remove the battery.
Re-assembling
Carry out the dismantling procedure in reverse order to re-assemble the unit.
We recommend carrying out a complete charging and discharging cycle to guarantee that
the battery will function correctly.
When closing the pump, ensure that the 2 cables connecting the power supply board and
the CPU board are positioned above the motor. In addition, check that the earth wire is
next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when the
pump is in operation. (See photo in Procedure n°1.)
Carry out regular servicing checks. (See Quality Control Certificate sheet.)
Intervention procedures
92 optima06.0_0100a_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed
1 Posidriv Z1 screwdriver
1 anti-static bracelet
1 soldering iron
Soldering wire "RADIEL Sn60Pb RI 1" (no cleaning necessary) or equivalent
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.).
Procedure:
Access
Pivot the Optima pump on the top of the front cover.
Unscrew the 2 cruciform screws (marker 1), connecting the Optima angle bracket to the
base.
Remove the angle bracket being careful not to pull out the connectors.
Disconnect the J4 and J3 connectors, as well as the functional Earth.
Return the Optima pump to its original position.
1
1
2
2
Intervention procedures
optima06.0_0110a_en.fm 93
Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig.2
94 optima06.0_0110a_en.fm
Dismantling the angle bracket
Dismantle the battery. (See corresponding Procedure.)
Dismantle the power supply board. (See corresponding Procedure.)
When closing the pump, ensure that the 2 cables connecting the power supply board and
the CPU board are positioned above the motor. In addition, check that the earth wire is
next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when the
pump is in operation. (See photo in Procedure n°1.)
Carry out regular servicing checks. (See Quality Control Certificate sheet.)
.
Intervention procedures
optima06.0_0110a_en.fm 95
Intervention procedures
96 optima06.0_0110a_en.fm
7 Calibration
7.1Calibration procedure
The calibration menu is reserved for qualified staff only. Its access is protected with
a secret code.
Keyboard description.
Key Function
ON switches on the device.
OFF switches off the device when pressed for more than three
seconds.
SILENCE ALARM
The selection keys scroll through the numbers and letters on the LCD
screen using the tenths, units and tens values.
Activate or de-activate the infusion menu.
ENTER confirms the selected calibration, or moves the cursor.
Calibration
optima07.1_001a_en.fm 97
How to activate the calibration mode
Keep the "MODE" button pressed, during the device switch on.
The various menus are displayed.
Press "START".
EtAL1: calibration of the door.
98 optima07.1_001a_en.fm
7.1.2 EtAL1 Door calibration
This menu records EtAL1, press "START".
the 3 door Por1: close the door without tube.
positions in the
Press "START".
EEPROM.
The value is read and recorded. The value should be between
1196 < X < 3750 mV.
Por2: open the door without tube.
Press "START".
The value is read and recorded. The value should be between
2197 < X < 2695 mV.
Por3: position a tube filled with water and close the door.
Press "START".
The value is read and recorded. The value should be between
1699 < X <3398 mV.
By validating once again, another calibration can be
selected.
Calibration
optima07.1_001a_en.fm 99
7.1.3 EtAL2 Calibration of the flow rate corrective coefficient
Before any flow rate test, carry out an infusion of 20 ml to stabilise the flow rate. The
adjustment range allows a correction up to -20% to +10%.
This menu adjusts Place a tube filled with water on the device, taking into
the device flow rate account the positioning of the clamp, and close the door.
according to the Select a flow rate of 100 ml/h.
used tube. Start the infusion.
Perform a flow rate measurement according to the norm
601-1-2-24.
Note the programmed flow rate and the obtained flow rate.
EtAL2, press "START".
Pressing "STOP" exits the menu without calculating the new coefficient.
Calibration
100 optima07.1_001a_en.fm
7.1.4 EtAL9 Calibrating the upstream and downstream pressure
sensors
This menu records The Optima pump can operate with various tube types. The choice of tube is made through
the pressure value the "MODE" key options. The pressure calibrations must be done for each type of tube (active
in the EEPROM or chosen with the "MODE"key).
according to the
selected tube.
If a pressure sensor, the door sensor or the CPU board are changed, the calibration
values for each tube must be reset to zero. If the CPU board is changed, the calibration is
carried out using maintenance software with which the Optima pump is being configured.
Installation:
According to the diagram below:
P ocket
1 m 50 cm
Pum p
M a n o m e te r
T a p 3 w a ys
C o n ta in e r
The manometer must be placed at the same level as the Optima pump.
Calibration
optima07.1_001a_en.fm 101
EtAL9, press "START".
Selecting "NO" is done when the tube change is done through the "MODE" key options.
The calibration described below is done for an active tube or one chosen through the
"MODE" key options.
Calibrating the downstream and upstream pressure sensor with the door open.
PaV1,PaM1: open the door of the Optima
pump, no tube installed.
Press "ENTER".
102 optima07.1_001a_en.fm
Calibrating the downstream pressure sensor at 0 mm Hg with the door closed, filled
tube installed.
Position the tube according to the preceding diagram and
open the 3-channel tap in the direction of the Pump-
Manometer.
Press "START", to warm up the tube.
Wait until the pressure reaches around 750 mm Hg
(1 bar +/- 0.05 bar), and open the 3-channel tap in the
pump-container direction to lower the pressure (0 mmhg).
Press "STOP" to stop infusion.
Calibrating the downstream pressure sensor at 750 mm Hg with the door closed filled
tube installed.
Position the 3 channel tap in the direction pump-manometer.
Press "START" to start up the pump.
Wait until the pressure reaches 750 mm Hg ±30 mm Hg
(1 bar +/- 0.05 bar).
Press "STOP" to stop infusion.
Calibrating the upstream pressure sensor at 0 mm Hg with the door closed filled
tube installed.
Keep the door closed.
Open the 3-channel tap in the direction pump-container.
Press "ENTER" to record the pressure, when the pressure level
is back to 0 mm Hg.
optima07.1_001a_en.fm 103
Calibration
104 optima07.1_001a_en.fm
8 Spare parts catalogue
8.1 Upper cover
107
100
104
102
109
106
101
111 105
110 103
108
optima08.1_001a_en.fm 105
Spare parts catalogue
106 optima08.1_001a_en.fm
Spare parts catalogue
8.2 Base
207
205
206
203
200
201 208
202
204
...
....
.....: see following page
optima08.2_001a_en.fm 107
....
....
....
....
....
.... ....
....
....
208
209
108 optima08.2_001a_en.fm
Spare parts catalogue
304 305
306
302 307
300
301
303
308
optima08.3_001a_en.fm 109
Spare parts catalogue
110 optima08.3_001a_en.fm
Useful addresses
SALES DEPARTEMENT
Fresenius Vial
Le Grand Chemin, 38590 Brézins
CUSTOMER SERVICE Tel.: 00 33 (0) 4 76 67 10 81
or 00 33 (0) 4 76 67 10 54
or 00 33 (0) 4 76 67 11 08
Fax: 00 33 (0) 4 76 67 11 34
TRAINING DEPARTMENT
Fresenius Vial Tél. : 00 33 (0) 4 76 67 10 76
Le Grand Chemin, 38590 Brézins Tél. : 00 33 (0) 4 76 67 11 40
Fax : 00 33 (0) 4 76 67 11 22
This document may not be reproduced in part or in whole without the written permission of
Fresenius Vial S.A.S.
Fresenius Vial S.A.S. - siège social : Le Grand Chemin - 38590 Brézins (FRANCE)
Société par Actions Simplifiées au capital de 13 744 520 Euros
SIREN Grenoble B 408 720 282.
Design and construction : SEDOC