Flomid FX Electromagnetic Flowmeter

The document describes an electromagnetic flowmeter called FLOMID FX. It has a modular design that can have either a compact or separate sensor and electronics unit. The sensor liner and electrodes are made of materials suitable for different applications like PP, PTFE, PVDF, and stainless steel. It provides accurate flow rate measurements independent of variables like flow profile, density, temperature, viscosity and pressure. Installation requires minimum straight pipe lengths upstream and downstream and it can be mounted in any position as long as the pipe remains full.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
32 views12 pages

Flomid FX Electromagnetic Flowmeter

The document describes an electromagnetic flowmeter called FLOMID FX. It has a modular design that can have either a compact or separate sensor and electronics unit. The sensor liner and electrodes are made of materials suitable for different applications like PP, PTFE, PVDF, and stainless steel. It provides accurate flow rate measurements independent of variables like flow profile, density, temperature, viscosity and pressure. Installation requires minimum straight pipe lengths upstream and downstream and it can be mounted in any position as long as the pipe remains full.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 12

FLOMID FX

Electromagnetic Flowmeter
Introduction
Modular design in two versions:
• Compact, sensor and electronics unit forming a single unit
• Separate, sensor and electronics separated, joined by a cable
Various possibilities of pipe fittings:
• Flange: DIN, ANSI, JIS norms etc.
• Sandwich: for DIN, ANSI, JIS norm flanges etc.
• Sanitary: DIN 11851 thread, ISO 2852 Clamp, SMS 1145 thread
• Sanitary welding: DIN 11850, ISO 2037 norms
Sensor liners PP, PTFE, PVDF & hard rubber
Electrodes in AISI-316L, Hastelloy C22 (UNS-06022),
Zirconium, Titanium
Exterior construction in AISI-316L

Benefits
● Flow rate measurament is independent of the flow profile
● Square wave ac magnetic field to avoid electrolysis
● Readings are independent of density, temperature,
viscosity and pressure
● Absence of obstructing elements gives zero pressure loss
and will allow the pass of solids
● Can be mounted in any position, provided that the pipe
is always full
● Low power consumption
● Good stability with temperature and age
● No moving parts provides zero maintenance
● Can be installed with short straight pipe sections
(5DN / 3DN)
● Good chemical resistance of construction materials
● All electronic control units for use with all sensors

Technical data FLOMID FX sensor


● Accuracy: + 0.5%
● Working pressure: PN16 standard,
on demand up to PN250
● Compling according to DIN, ANSI, JIS, Sanitary
● DN3...DN500
● Working temperature:
PP, -10...+80 oC
PTFE, PVDF, -20...+120 oC
Hard rubber, -20...+90 oC
(temperatures for sensor with separated electronics)
Measurement Principle
The measurement principle is based on FARADAY`S
induction law. A conductive liquid (>5µS/cm) creates a
voltage Tm when flowing in a pipe of diameter D, at an
average velocity of Vm, through a magnetic field B (which is
perpendicular to the flow direction). This voltage is
proportional to the velocity of the liquid. Two electrodes,
mounted perpendicular to the magnetic field inside the pipe,
transmit the voltage Tm to the control unit.

Tm = B x Vm x D

Tm = Voltage across the electrodes


Vm = Liquid velocity
B = Magnetic field strength
D = Pipe diameter

Basic Installation Recommendations

For the mechanical installation the most important factor to Mixtures of different liquids. The sensor should be installed
be taken into consideration is that the pipe must always be a minimum of 30DN from the point of mixture to avoid
full and the electrodes in contact with the liquid. To guarantee unstabilities in the readings.
this, the metering body should be mounted with the
electrodes in a horizontal plane and in a position so that the
pipe will always be full, without air pockets.

Metallic Earth Disc Plastic Disc with Electrode

Vibration A good earth connection between the measuring body and


Fasten the pipe before and after the sensor. the liquid can be considered as the most important factor for
Caution: Excessive vibration will need separate mounting of the correct working of the unit. In the event that the pipe is
sensor and electronics. made of electrically conductive material, just connect the two
With free runs of piping over 10 m long, mechanical supports earth wires to the pipe flanges, one to each side of the
are recommended to minimize external forces. instrument body. If the inside of the pipe is plastic (or of any
other non-conductive material) two earth discs and two more
rubber washers, one on each side of the body, must be
installed and the earth wires connected to the earth discs.
This earth should be only used for the measuring body, do not
connect other equipment to it since electrical interferences
may cause problems.

In installations where strong axial ground currents can be


found, the earth connections should have at least 50 mm2 in
contact with the liquid.

2
Straight sections of pipe are necessary for the correct For abrasive liquids or liquids with particles in suspension, it
working of the flowmeter. Curves, pipe reductions, control is recommended to mount the flowmeter in a rising pipe. For
elements and pumps should be at least 5 DN upstream and a horizontal pipe the installation can be made as in the
3 DN downstream from the flowmeter. following drawing with a valve system for cleaning.

Valves should be downstream from the flowmeter to keep the


pipe full and to avoid vacuums which can damage the
flowmeter liner.

Pumps should be upstream from the flowmeter to avoid


vacuums.

When installing near curves, control elements and pumps, a


minimum straight section of pipe must be installed before and
after these elements (Minimum 5 DN before and 3 DN after). drain

Measuring Precision (DIN 19200)


Ambient temperature 20 oC + 2 oC
Liquid temperature 20 oC + 2 oC
Power supply + 1%
Heat up time 30 min.
Minimum straight pipe 10 DN y 5 DN

V > 0.4 m/s E = + 0.5% of reading


0.18
V < 0.4 m/s E=+ = [%] of reading
V[m/s]

Pressure loss when reducing the pipe diameter


In installations where, due to the low flow rate, it is necessary
to reduce the pipe diameter to be able to install a flowmeter
with a nominal diameter less than the pipe diameter,
this should be done using a reduction cone with an angle
of less than 8o in order to avoid turbulences which can
give false readings.
∆p(m bar)

The graph shows the


pressure loss (∆P)
for water at different
d1 velocities and pipe
Diameter relationship
d2 reductions.

3
Flow rate table Metering body selection
The diagram shows the relation
between the liquid speed and the
flow rate for different body sizes.
The metering body diameter
should be chosen to obtain a
maximum normal working liquid
velocity of about 3-4 m/s. The
minimum working liquid velocity
should not be below 0.5 m/s.

A recommended working liquid


velocity is between 1 and 3 m/s.
When the liquid has solids in
suspension, it is better to work
between 3 and 5 m/s to avoid
sedimentation in the pipe and
metering body.

4
FLOMID 0 FX
Compact electronics

Separate electronics

ANSI B16.5 sandwich mounting


(Dimensions in inches)
PN Weight
DN (Ibs) A B C g (Ibs) F L h

*1/2” 16 262 212 23 46 6,5 45 64 178


Dimensions (150) (10,3) (8,3) (0,90) (1,81) (14,3) (1,7) (2,51) (7,0)
3/4” 16 265 215 25 50 6,8 47 64 182
DIN 2501/2502 sandwich mounting
(150) (10,4) (8,4) (0,98) (1,96) (14,9) (1,8) (2,51) (7,1)
(Dimensions in mm)
1” 16 275 222 31 61 7,2 47 64 188
Weight F (150) (10,8) (8,7) (1,22) (2,40) (15,8) (1,8) (2,51) (7,40)
DN PN A B C g (kg) L h 1 1/4” 16 243 193 35 70 8,2 - 79 160
(150) (9,6) (7,5) (1,37) (2,75) (18,0) - (3,11) (6,2)
3* 16 262 212 23 46 6,5 45 64 178 1 1/2” 16 250 200 41 82 9 - 79 167
6* 16 262 212 23 46 6,5 45 64 178 (150) (9,8) (7,8) (1,61) (3,22) (19,84) - (3,11) (6,5)
10* 16 262 212 23 46 6,5 45 64 178 2” 16 256 206 46 92 9,5 - 99 173
15* 16 265 215 25 50 6,8 47 64 182 (150) (10,0) (8,1) (1,81) (3,62) (20,9) - (3,89) (6,8)
20* 16 272 222 31 61 7,2 47 64 188 2 1/2” 16 263 213 53 106 10,3 - 99 180
25* 16 243 193 35 70 8,2 - 79 160 (150) (10,3) (8,3) (2,08) (4,17) (22,7) - (3,89) (7,1)
32 16 250 200 41 82 9 - 79 167 3” 16 275 225 63 126 11 - 119 192
40 16 256 206 46 92 9,5 - 99 173 (150) (10,8) (8,8) (2,48) (4,96) (24,2) - (4,68) (7,5)
50 16 263 213 53 106 10,3 - 99 180 4” 16 293 243 81 162 14 - 164 210
65 16 275 225 63 126 11 - 119 192 (150) (11,5) (9,5) (3,18) (6,37) (30,8) - (6,45) (8,2)
80 16 283 233 71 142 12,2 - 119 200 5” 16 310 260 96 192 17,5 - 164 226
100 10 293 243 81 162 14 - 164 210 (150) (12,7) (10,7) (3,77) (7,55) (38,5) - (6,45) (8,8)
125 10 310 260 96 192 17,5 - 164 226 6” 16 322 272 108 217 19,6 - 164 238
150 10 322 272 108 217 19,6 - 164 238 (150) (12,6) (10,7) (4,25) (8,54) (43,2) - (6,45) (9,3)
*h+24 mm *Equal dimensions for 1/8” and 3/8”

5
DIN, ANSI Flanges FLOMID 2 FX / FLOMID 4 FX
MX Compact electronics
208 162
94 110 130

A
º
lxn
D

g
k

DN L3 + 0,5%
min + 1mm
MCX Compact electronics
46 122 120
80
B

º
lxn

h
D

g
k

c
DN L3 + 0,5%
min + 1mm
FX Sensor (Separate electronics)

FLANGES mm (inch) DIMENSIONS mm (inch)

DIN 2632-2635 ANSI B16.5 A B L D K g lxnº h D K g lxnº Weight


DN PN DN PN mm mm mm DIN DIN DIN DIN DIN ANSI ANSI ANSI ANSI kg
(bar) (lbs) (inches) (inches) (lbs)

1-10 16 3/8” 150 285 (11,2) 235 (9,3) 150 (5,90) 90 60 40 14x4 205 (8,1) 88,9 4,5 (9,92)
15 16 1/2” 150 285 (11,2) 235 (9,3) 150 (5,90) 95 65 45 14x4 205 (8,1) 88,9 60,3 34,9 16x4 4,8 (10,60)
20 16 3/4” 150 285 (11,2) 235 (9,3) 150 (5,90) 105 75 58 14x4 205 (8,1) 98,4 69,8 42,9 16x4 6 (13,22)
25 16 1” 150 285 (11,2) 235 (9,3) 150 (5,90) 115 85 68 14x4 205 (8,1) 107,9 79,4 50,8 16x4 7,5 (16,53)
32 16 1 1/4” 150 330 (13,0) 280 (11,1) 150 (5,90) 140 100 78 18x4 250 (9,9) 117,5 88,9 63,5 16x4 8 (17,63)
40 16 1 1/2” 150 330 (13,0) 280 (11,1) 150 (5,90) 150 110 88 18x4 250 (9,9) 127,5 98,4 73 16x4 8,5 (18,74)
50 16 2” 150 347 (13,6) 297 (11,7) 200 (7,87) 165 125 102 18x4 267 (10,5) 152,4 120,6 92,1 19x4 10 (22,05)
65 16 2 1/2” 150 367 (14,5) 317 (12,5) 200 (7,87) 185 145 122 18x4 287 (11,3) 177,8 139,7 104,8 19x4 11 (24,25)
80 16 3” 150 367 (14,5) 317 (12,5) 200 (7,87) 200 160 138 18x4 287 (11,3) 190,5 152,4 127 19x4 12,5 (27,55)
100 16 4” 150 418 (16,4) 368 (14,5) 250 (9,84) 220 180 158 18x4 338 (13,3) 228,6 190,5 157,2 19x8 15 (33,06)
125 16 5” 150 443 (17,4) 393 (15,5) 250 (9,84) 250 210 188 18x4 363 (14,3) 254 215,9 185,7 23x8 18,5 (40,78)
150 16 6” 150 473 (18,6) 423 (16,7) 300 (11,81) 285 240 212 22x8 393 (15,5) 279,4 241,3 215,9 23x8 20,6 (45,41)
200 10 8” 150 523 (20,6) 473 (18,6) 350 (13,77) 340 295 268 22x8 443 (17,5) 342,9 298,4 269,9 23x8 32 (70,54)
250 10 10” 150 575 (22,6) 525 (20,7) 400 (15,74) 395 350 320 22x8 495 (19,5) 406,4 361,9 323,8 25x12 45 (99,20)
300 10 12” 150 621 (24,5) 571 (22,5) 500 (19,68) 445 400 370 22x8 541 (21,3) 482,6 431,8 381 25x12 52 (114,63)
350 10 14” 150 671 (26,4) 621 (24,5) 500 (19,69) 505 460 430 22x8 591 (23,3) 533,4 476,2 412,7 30x12 62 (136,68)
400 10 16” 150 721 (28,4) 671 (26,4) 600 (23,62) 565 515 482 26x16 641 (25,3) 596,9 539,7 469,9 30x16 76 (167,54)
500 10 18” 150 825 (32,5) 775 (30,5) 600 (23,62) 670 620 585 26x20 745 (29,4) 635 577,8 533,4 33x16 98 (216,04)

6
DIN 11851 FLOMID 1 FX DIN 11851 FLOMID 1 FX
MX Compact electronics
DN 10 15 20 25 32 40 50 65 80 100
NW Ø d1 10 16 20 26 32 38 50 66 81 100
Rd 28 Rd 34 Rd 34 Rd 52 Rd 58 Rd 65 Rd 78 Rd 95 Rd 110 Rd 130
Ø d5 x 1/8” x 1/8” x 1/6” x 1/6” x 1/6” x 1/6” x 1/6” x 1/6” x 1/4” x 1/4”
H 28 28 28 29 29 29 29 31 27 28
Ø d3 60 64 70 80 90 100 114 144 160 180
A 268 272 275 280 286 290 300 310 318 328
L1 200 220 240 280 330
L2 192 188 206 226 264 310
L3 120 134 154 186 234
F 45 48 54 - - - - - - -

FLOMID 1...3 FX
MXC Compact electronics

ISO 2853 FLOMID 3 FX


MX Compact electronics

ISO 2853 FLOMID 3 FX


DN 25 32 38 51 63,5 76
NW Ø d1 22,5 29,5 35,5 48,5 60,5 72,9
Rd 40 Rd 48 Rd 60 Rd 70 Rd 85 Rd 98
Ø d5 x 1/6” x 1/6” x 1/6” x 1/6” x 1/6” x 1/6”
H 25 25 25 25 25 25
Ø d3 80 90 100 114 144 160
A 280 286 290 300 310 318
L1 220 240 280
L2 134 154 186
Equival. DIN-DN 25 32 40 50 65 80

7
CLAMP ISO 2852:1993 FLOMID 6 FX CLAMP ISO 2852 : 1993 FLOMID 6 FX
MX Compact electronics Coupling
dimensions - - 1/2” - - 1” 1 1/4” 1 1/2” 2” 2 1/2” 3” 4”
Equival. DIN-DN 6 6 10 15 20 25 32 40 50 65 80 100
pipe
outside ø 8 12 12,7 17,2 21,3 25 33,7 38 51 63,5 76,1 101,6
Ø C6 6 10 10,7 15,2 19,3 22,6 31,3 35,6 48,6 60,3 72,9 97,6
Ø C7 34 34 34 34 34 50,5 50,5 50,5 64 77,5 91 119
H 30 30 30 30 30 31 31 31 31 34 34 34
Ø d3 60 60 60 64 70 80 90 100 114 144 160 180
A 268 268 268 272 275 280 286 290 300 310 318 328
L3 120 134 154 186 234
L1 200 220 240 280 330
F 45 45 45 48 54 - - - - - - -

MCX Compact electronics

DIN 11850 FLOMID D10 FX ISO 2853 FLOMID I38 FX


DN 10 15 25 40 50 65 80 DN 15 20 25 40 50 65 80
de 12(0,47) 18(0,71) 40(1,57) 52(2,05) 68(2,03) 83(3,27) 83(3,27) R 1/2” 3/4” 1” 1 1/2” 2” 2 1/2” 3”
di 10(0,39) 16(0,63) 26(1,02) 38(1,50) 50(1,97) 66(2,60) 81(3,19) L 50(2) 63(2,48) 60(2,36)
L 50(2) 63(2,48) 60(2,36)

DIN 11850 ISO 2037 ISO 2853 TRI-CLAMP

ISO 2037 FLOMID I70 FX


DN 10 15 25 40 50 65 80 TRI-CLAMP FLOMID D30 FX
de 12(0,47) 18(0,71) 25,6(1,01) 38,6(1,52) 51,6(2,03) 64,1(2,52) 76,6(3,02)
DN 25 40 50 65 80 100
di 10(0,39) 16(0,63) 22,6(0,89) 35,6(1,40) 48,6(1,97) 60,3(2,37) 72,9(2,87)
C7 50,4 50,4 64 77,8 91 119
L 50(2) 63(2,48) 60(2,36) di 22,1 34,8 47,5 60,2 72,9 97,4
*The rest of the dimensions as with FLOMID 1 (page 7) L 63(2,48) 60(2,36)
( ) dimensions in inches

8
Liner materials. Temperature, Pressure and Vacuum (20oC reference temperature)
PTFE PTFE / PVDF PP EBONITA

Flomid 2 FX, Flomid 0 FX Flomid 0 FX Flomid 2 FX,


DN Flomid 4 FX Flomid 1, 3, 5, 7 FX Flomid 4 FX

PN Vacuum(1) PN Vacuum(1) PN Vacuum(1) PN Vacuum(1)

DIN2501(3) ANSI mbar psi DIN2501 ANSI mbar psi DIN2501 ANSI mbar psi DIN2501 ANSI mbar psi
B 16.5(4) B 16.5 B 16.5 B 16.5

3 0 0
6 60 0,9 100 1,5
∆ ∆ ∆ ∆ ∆ ∆
10 16 40 150 300 16 150 300 16 150 300



15 500 7,5 600 9


20 0 0 100 1,5
25 80 1,2 ∆ ∆
32 ∆ ∆ 100 1,5 180 2,7
16 40 150 300 16 150 300 16 150 300 16 40 150 300



∆ ∆ ∆ ∆
40
50 500 7,5 600 9 700 10,5 200 3
65 150 2,2 150 2,2 200 3 120 1,8
∆ ∆ ∆ ∆ ∆ ∆ ∆ ∆
16 40 150 300 16 150 300 16 150 300 16 40 150 300



80 650 9,7 700 10,5 800 12 250 3,7


100 250 3,7 300 4,5 380 5,7 280 4,2
∆ ∆ ∆ ∆ ∆ ∆

125 16 150 750 11,2 10 150 300 800 12 10 150 300 900 13,5 10 150 ∆ ∆

∆ 450 6,7 480 7,2 ∆ 650 9,7 ∆


∆ ∆ ∆ ∆ ∆ ∆

150 40 800 12 900 13,5 1000 15 40 400 6


200 450 6,7 250 3,75
∆ ∆

900 13,5 ∆ ∆

250 500 7,5


∆ ∆
10 150 10 150
300 ∆ 1000 15 ∆ 450 6,7
350 750 11,2 500 7,5
∆ ∆ ∆ ∆
400 40 40
500 1000 15 600 9
Max. -20...+120oC -20...+120oC -10...+80oC -20...+90oC
Temp. ( -4...+248oF) ( -4...+248oF) ( 14...+176oF) ( 14...+194oF)
Limit(2) 130oC (266oF) 130oC (266oF) ___ ___

(1) In mbar absolute 40oC and 80oC reference temperature (3) bar
(2) Maximum 30’. On order up to 180oC and PN200 (4) lbs

Wiring

Upper coil Reset. Anode


Lower coil Reset. Cathode
Ground Pulse output. Emitter
Output 1. Collector Pulse output. Collector
Output 1. Emitter Output mA (+)
Output 2. Collector Output mA (-)
Output 2. Emitter Rear electrode (+)
Earth Rear electrode (-)
Neutral Front electrode (-)
Live Front electrode (+)

250 mA (T)

• Wire for separate electronics:


PAR-POS 2 x 2 x 0.34 (recommended)

9
Electronic Control Units Cycle
Different models of electronic control units are available mv

to comply with the options of flow indication, Signal voltage

maximum/minimum flow rate controls, analog outputs, pulse


outputs or flow switches. All are compatible with the different
Flomid FX sensor models.
0 Time
For compact mounting directly on the sensor
● Series MX, with local indication, signal, mA and pulse outputs

● Series MCX/T ... FT ... FI, without local indication, with

analog output and/or pulse output


mv
For wall and pipe bracket mounting
A = 20 ms
● Series MX (Local Indication) - M* ... T*

● Series MCX (Without Local Indication) - M* ... T*


Measuring time

*M = wall mounting
*T = pipe mounting Note: There is no difference between the compact and the
separated electronics except for the M or T accesory
(supplied for its mounting on a wall or on a pipe), and for the
cable for its connection to the sensor.

Electronic Technical Data


Series MX (Compact) and Series MX-M, MX-T (Separated)

● Plastic coated (Polyamide 11) injected aluminium housing. ● Adaptive flowrate filter: Integration time programmable
IP 67 degree of protection 0.1 ... 25.5 seconds
● Programming via front tactile pushbuttons Filter reset window programmable
● Totalising and partial volume counters ● Linearity: 0.1%
● Programmable batching volume with remote input ● Zero drift: 0.05%
for starting ● Temperature drift: 0.015% / oC
● Magnetic field: 6.25 Hz square wave for ● Working ambient temperature range: -10 ...+60 oC
50 Hz mains
7.5 Hz square wave
for 60 Hz mains
● Mains supply: 12 V, 24 V, 110 V, 120 V, 220 V,
240 V ac, + 10%, 50 o 60 Hz
24 Vdc
(others on demand)
● Power consumption: <10 VA
● Min. Flowrate cut off: Programmable
● Measuring range: 0...10 m/s
● Display: 16 character x 2 lines LCD
6 languages to be selected.
Program for auto-edition of
other languages
● Flowrate indication: User programmable
measuring units
● Analog output: 4...20 mA programmable
● Programmable pulse output mode:
Flowrate x Pulses/unit volume < 2 pulses/s
● Programmable frequency output mode: 10..1000 Hz
● Logic outputs: Two programmable outputs for
flowrate alarms, empty pipe
detection, flow direction or batching

10
The MCX Series are a reliable and economical solution for a FLOMID MCX/T, MCX/FT & MCX/FI
majority of industrial processes that require specific controls
such as:

• Analog transmitters & pulse output (MCX/T ... ...


Output to panel indicators, control units, automats, etc.
• Analog transmitter with fast response (MCX/FT ... ...
For applications that require very fast response times.
• Pulse transmitter with fast response (MCX/FI ... ...
For batching processes that require very fast response
times.

Compact electronics MCX/T ... FT ...FI


Separate electronics MCX/T ... M/T*
MCX/FT ... M/T*
MCX/FI ... M/T*
*M = wall mounting
*T = pipe mounting Note: There is no difference between the compact and the
separated electronics except for the M or T accesory
(supplied for its mounting on a wall or on a pipe), and for the
Transmitters Technical Characteristics cable for its connection to the sensor.

Electronic control model

FLOMID MCX/T FLOMID MCX/FT FLOMID MCX/FI

IP-65 housing (polyamide 11 plastic covered) yes yes yes


Galvanic separation
between sensor and outputs yes no yes/no
Flow direction selected by jumpers yes yes yes
Magnetic Field
Square wave 6.25 Hz at 50 Hz mains yes no no
Square wave 7.5 Hz at 60 Hz yes no no
Square wave 12.5 Hz at 50 Hz no yes yes
Square wave 15 Hz at 60 Hz no yes yes
Response time (ms) ~ 2500 ms 200 ms 200 ms
Power supply +10% yes yes yes
12 V, 24 V, 110 V, 120 V, 220 V, 240 V confirm confirm confirm
at 50 & 60 Hz with the order with the order with the order
Analog output yes yes -
0-5 V, 0-10 V, 1-5 V, 2-10 V, confirm confirm -
0...4-20 mA with the order with the order -
Standard pulse output 0...100 Hz / 0...1000 Hz no -
1, 10, 100 p/l (others on demand) - - yes
Measuring range 0,15...10 m/s 0,15...10 m/s 0,15...10 m/s
3 full scale ranges selectable 0-100 % f.s.v*
by jumpers no 0-50 % f.s.v* no
0-25 % f.s.v*
Minimum flowrate 0,15 m/s 0,15 m/s 0,15 m/s
(3% of nominal full scale value at 5 m/s)
Linearity 0,3 % 0,3 % 0,3 %
Zero drift 0,15 % 0,15 % 0,15 %
Working ambient temp. range -10 +60 oC -10 +60 oC -10 +60 oC
Temperature drift 0,015 % / oC 0,015 % / oC 0,015 % / oC
* Full scale

11
MCX/T-M...T, MCX/FT-M...T, MCX/FI-M...T Flomid MCX/T, MCX/FT, MCX/FI
(Separate electronics) (Compact electronics)

Wall Mounting Electronics Pipe Mounting Electronics

Series MX-M Series MX-T

Series MCX-M Series MCX-T

We are at your service, please consult us.


TECFLUID develops and manufactures instruments for gases
and liquids, using the most advanced technology.
Request Information by telephone nº (34) 93 372 45 11

C/. Narcis Monturiol, 33 - E-08960 SANT JUST DESVERN (BARCELONA)


International: Telephone. (34) 93 372 45 11 - Fax (34) 93 473 44 49
www.tecfluid.com - e-mail: [email protected]

Data in this pamphlet is subject to modification without notification, if the


technical innovations in the product or manufacturing processes so require.

You might also like