SOP ASTM D1946-90 (Rev. 1.7)
SOP ASTM D1946-90 (Rev. 1.7)
SOP ASTM D1946-90 (Rev. 1.7)
Signature: Signature:
1.1 Analytes. Reformed gases and other gaseous mixtures containing the following
components: hydrogen, oxygen, nitrogen, carbon monoxide, methane, ethane, and
ethylene.
NOTE: a TCD detector is utilized for the inorganics and a FID for the organics. See
Entech’s most current version of the SOPs for GC/TCD analysis and GC/FID
analysis.
1.2 Applicability. This practice covers the determination of the chemical composition of
reformed gases and similar gaseous mixtures. The information about the chemical
composition may be used to calculate physical properties of the gas, such as heating
(calorific) value and relative density. Combustion characteristics, products of
combustion, toxicity, and interchangeability with other fuel gases may also be
inferred from the chemical composition.
Concentration Concentration
Range (Mol.%) Range (Mol.%)
FID: Methane 0.01 - 100 TCD: Hydrogen 0.5 - 100
Ethane 0.01 - 100 Nitrogen 0.1 - 100
Propane 0.01 - 100 Oxygen 0.1 - 50
Isobutane 0.01 - 10 Carbon monoxide 0.1 - 100
n-Butane 0.01 - 10 Carbon dioxide 0.1 - 100
neo-Pentane 0.01 - 2
iso-Pentane 0.01 - 2
n-Pentane 0.01 - 2
n-Hexame 0.01 - 2
n-Heptane 0.01 - 2
1.5 Interferences.
The sample matrix can interfere with peak resolution. A proper analytical
condition may be established by choosing an appropriate GC column, adjusting
column carry gas flow, and setting the GC temperature program accordingly.
Moisture in the sample or air may also affect detectors response. The instrument
should be heated to operation condition prior standard and sample analysis.
Since the TCD exhibits universal response and detects all gas compounds except the
carrier gas, interference may occur. Choosing the appropriate GC condition or
changing the column flow rate to shift the retention time may help to eliminate
resolution interferences.
3.0 Definitions.
4.0 Safety.
4.1 This method may involve hazardous materials, operations, and equipment. It is
the responsibility of the user of this method to establish appropriate safety and
health practices and procedures prior performing this method. Always consult the
instrument operator’s manual, chemical’s MSDS, and sampling source
information for specific precautions regarding the analytical procedure.
4.2 Follow the procedures in the most current version of Entech Engineering’s Accident
Prevention Plan and Employee Safety Manual.
FID Injection port: Valco UW type 10-port sampling and switching valve with a
250ul capacity sample loop installed.
Column A: Supleco Hayesep Q polymer (80/100 mesh) stainless steel packed column, 6’long
x 1/8” OD.
Column B: Supleco Hayesep DB polymer (100/120 mesh) stainless steel packed column,
30’long x 1/8” OD.
Syringe: SGE 1-liter jumbo syringe (model 1000MAR-LL-GT), SGE 100ml syringe
(model 100MAX-LL-GT), Hamilton 10ml syringe (model GASTIGHT #1010), Hamilton
NOT A CONTROLLED DOCUMENT ONCE PRINTED
This document is the confidential property of Entech Engineering Inc. It must be returned on
demand and must not be disclosed or reproduced without prior written consent from a duly
authorized representative of Entech Engineering Inc
SOP ID Number: SOP- ASTM D1946-90
Entech Engineering Inc. Revision: 1.7
P.O. Box 890746 • Houston, Texas 77289-0746 •281-332-3118 Date: June 24, 2016
Page: 4 of 16
Standard Operating Procedure
Cylinder gauge and regulator: CGA 350 and CGAS 580 regulator with gauges.
Gases: Hydrocarbon-free compress air system or Ultra zero grade compressed air; Ultra
High Purity (UHP) compressed nitrogen; UHP compressed hydrogen.
NOTE: In order to trace the supply gases were used during the analysis procedure, when
install or remove supply gas cylinders, record the cylinder’s serial number onto the
laboratory’s cylinder gases logbook with date and operator’s initial.
6.1 Moisture-free mixtures of known compositions are required for comparison with the
test sample. They must contain known percentages of the components that are to be
determined in the unknown sample. All components in the reference standard must be
homogenous in the vapor state at the time of use.
6.2 Primary reference standard gas mixture – Air Liquide certified master class
calibration standard and Alphagaz class gases or equivalent. Alternative calibration
standard – Scott Gases, PortaGas, and SigmaAldrich certified standards.
Follow the procedures presented in project proposal and EPA Reference Method,
whichever is applicable.
Upon the arriving the laboratory, check the integrity of the sample bag for leaking or
deflecting. Logging samples into the chain-of-custody (COC) according to the procedures
presented in sample acceptance SOP with appropriate remarks.
Refer to section 9.3 to setup a proper data acquisition. Remember to include the gas
standard’s cylinder lot number or lab designated unique number in raw data file.
Connect a low level standard bag to the sampling valve. When the GC is ready, open
the bag valve by turning it counter clockwise and press the bag to feed standard into
the sample loop for 20 seconds. Immediately, press the GC’s “START” button to
start the GC run, and close the bag valve by turning it clockwise.
Note: when pressing the bag, air bubbles should appear at the sample loop exit; if no
bubbles appear, no sample is going into the loop. It is possible that the sample bag
valve may be closed or there is not enough pressure on the bag. Always keep a
consistent pressure on the bag.
Repeat steps in 10.1 to run the standard in triplet. Repeat steps in 10.1 for mid and
high level standards.
Collect all the results from standards analysis, calculate the individual response for
each standard run, and the average response for each level. Use the known
NOT A CONTROLLED DOCUMENT ONCE PRINTED
This document is the confidential property of Entech Engineering Inc. It must be returned on
demand and must not be disclosed or reproduced without prior written consent from a duly
authorized representative of Entech Engineering Inc
SOP ID Number: SOP- ASTM D1946-90
Entech Engineering Inc. Revision: 1.7
P.O. Box 890746 • Houston, Texas 77289-0746 •281-332-3118 Date: June 24, 2016
Page: 6 of 16
Standard Operating Procedure
Integrated bag sample should be analyzed within the same period as the laboratory’s
quality control sample for each project. The bag sample will be retained in the laboratory
for 90 days after the report has been submitted to the client.
Follow the procedures presented in Entech Engineering Inc. Quality Assurance & Quality
Control (QA/QC) Manual, Sections 5.2.3 through 5.2.6.
NOT A CONTROLLED DOCUMENT ONCE PRINTED
This document is the confidential property of Entech Engineering Inc. It must be returned on
demand and must not be disclosed or reproduced without prior written consent from a duly
authorized representative of Entech Engineering Inc
SOP ID Number: SOP- ASTM D1946-90
Entech Engineering Inc. Revision: 1.7
P.O. Box 890746 • Houston, Texas 77289-0746 •281-332-3118 Date: June 24, 2016
Page: 7 of 16
Standard Operating Procedure
Follow the procedures listed in the most current version of Entech Engineering’s QA and QC
Manual.
Follow the procedures presented in Entech Engineering Inc. QA/QC Manual, Section
6.2.3.
14.0 Calculations.
14.1 The number of significant digits retained for the quantitative value of each
component shall be such that accuracy is neither sacrificed nor exaggerated. The
expressed numerical value of any component in the sample should not be presumed
to be more accurate than the corresponding certified value of that component in the
calibration standard.
14.2
(Known concentration of the compound in standard)
Response Factor =
(Compound peak area)
Where,
Follow the procedures presented in EPA 40 CFR 60, Appendix A, Reference Method 18,
Section 8.2.1.6 to determine the moisture content in the sample
15.3.1 FID - If applicable, perform a mid level standard analysis as CCV at the
beginning of a set sample analysis to verify the instrument response drift is
within 5 percent of initial calibration. If a CCV failed the criteria due to the
matrix interference from pervious sample analysis, a new calibration, as
described in 10.1, may be required.
16.1 The sample introduced into the GC column should have a volume between 0.2 and
0.5 ml. Sufficient accuracy can be obtained for the determination of all but the very
minor components with this sample size. When increased sensitivity is required for
the determination of components present in low concentrations, a sample size of up to
NOT A CONTROLLED DOCUMENT ONCE PRINTED
This document is the confidential property of Entech Engineering Inc. It must be returned on
demand and must not be disclosed or reproduced without prior written consent from a duly
authorized representative of Entech Engineering Inc
SOP ID Number: SOP- ASTM D1946-90
Entech Engineering Inc. Revision: 1.7
P.O. Box 890746 • Houston, Texas 77289-0746 •281-332-3118 Date: June 24, 2016
Page: 9 of 16
Standard Operating Procedure
16.2 Precision: The following data should be used to judge the acceptability of the results:
0 to 1 0.05
1 to 5 0.1
5 to 25 0.3
Over 25 0.5
0 to 1 0.1
1 to 5 0.2
5 to 25 0.3
Over 25 1.0
During and upon completion of a project, Entech will be responsible for adhering to
applicable federal and state pollution prevention and waste management procedures.
Note: Inspect and calibrate instrument before and after each field use.
NOT A CONTROLLED DOCUMENT ONCE PRINTED
This document is the confidential property of Entech Engineering Inc. It must be returned on
demand and must not be disclosed or reproduced without prior written consent from a duly
authorized representative of Entech Engineering Inc
SOP ID Number: SOP- ASTM D1946-90
Entech Engineering Inc. Revision: 1.7
P.O. Box 890746 • Houston, Texas 77289-0746 •281-332-3118 Date: June 24, 2016
Page: 10 of 16
Standard Operating Procedure
Check laboratory routine supplies and standards periodically to ensure an effective and
efficient operation.
19.4 Column
Evaluate performance proudly. Replace when it falls below performance
requirement, such as contamination, column bleeding, and fail to resolve adjacent
peaks.
Follow the procedures listed in the most current version of Entech Engineering’s QA and QC
Manual.
Follow the procedures presented in Entech Engineering Inc. QA/QC Manual, Section 7.0.
Follow the procedures presented in Entech Engineering Inc. QA/QC Manual, Section 9.0.
23.0 Complaints.
Follow the procedures presented in Entech Engineering Inc. QA/QC Manual, Section 9.0.
NOT A CONTROLLED DOCUMENT ONCE PRINTED
This document is the confidential property of Entech Engineering Inc. It must be returned on
demand and must not be disclosed or reproduced without prior written consent from a duly
authorized representative of Entech Engineering Inc
SOP ID Number: SOP- ASTM D1946-90
Entech Engineering Inc. Revision: 1.7
P.O. Box 890746 • Houston, Texas 77289-0746 •281-332-3118 Date: June 24, 2016
Page: 11 of 16
Standard Operating Procedure
24.0 References.
1. July 2001. ASTM Standard Test Method D1946-90 “Standard Practice for Analysis of
Reformed Gas by Gas Chromatography (GC)” ASTM 1990.
2. Entech Engineering Inc., Quality Assurance & Quality Control Manual, most recent
version.
3. Thomas A. Ratliff, Jr., The Laboratory Quality Assurance System. (New York: Wiley,
1991).
1.6 – 1.7 10.0 and 15.0 Updated section content sly 06/24/16
Table – 1
NOTE:
(1) FID low and mid level standard: dilute 25 ml and 50 ml of high level standard with nitrogen
to the final volume of 1 liter, respectively.
(2) TCD low and mid level standard: (a) dilute 100 ml and 500 ml of high level standard with
helium to the final volume of 1 liter for nitrogen, carbon monoxide, and carbon dioxide,
respectively; (b) dilute 100 ml and 200 ml of high level standard with helium to the final
volume of 1 liter for oxygen, respectively; (c) dilute 200 ml and 500 ml of high level standard
with helium to the final volume of 1 liter for hydrogen, respectively;
(3) High level standard concentration may vary due to manufacturing process.
NOT A CONTROLLED DOCUMENT ONCE PRINTED
This document is the confidential property of Entech Engineering Inc. It must be returned on
demand and must not be disclosed or reproduced without prior written consent from a duly
authorized representative of Entech Engineering Inc
SOP ID Number: SOP- ASTM D1946-90
Entech Engineering Inc. Revision: 1.7
P.O. Box 890746 • Houston, Texas 77289-0746 •281-332-3118 Date: June 24, 2016
Page: 13 of 16
Standard Operating Procedure
Table – 2
GC configuration
GC Operating Parameters
Table - 3
FID Initial Calibration
(Example)
This document is the confidential property of Entech Engineering Inc. It must be returned on demand and must not be disclosed
or reproduced without the prior written consent from a duly authorized representative of Entech Engineering Inc.
SOP ID Number: SOP- ASTM D1946-
Entech Engineering Inc. 90
P.O. Box 890746 • Houston, Texas 77289-0746 •281-332-3118 Revision: 1.7
Date: April 26, 2016
Page: 15 of 16
Standard Operating Procedure
Table - 4
TCD Initial Calibration
(Example)
This document is the confidential property of Entech Engineering Inc. It must be returned on demand and must not be disclosed
or reproduced without the prior written consent from a duly authorized representative of Entech Engineering Inc.
SOP ID Number: SOP- ASTM D1946-
Entech Engineering Inc. 90
P.O. Box 890746 • Houston, Texas 77289-0746 •281-332-3118 Revision: 1.7
Date: April 26, 2016
Page: 16 of 16
Standard Operating Procedure
Table - 5
Reference Chromatogram
FID
TCD
This document is the confidential property of Entech Engineering Inc. It must be returned on demand and must not be disclosed
or reproduced without the prior written consent from a duly authorized representative of Entech Engineering Inc.