CALIBRATION CERTIFICATE FORMAT Pipettor SAMPLE

This document provides a [1] certificate of calibration and calibration report for a 2-20uL pipettor. [2] The pipettor was calibrated on March 3, 2021 at San Pedro Hospital in Davao City using comparison method and standards traceable to NIST. [3] The pipettor was found to have errors of -0.00008 uL at 2 uL and +0.0028 uL at 20 uL settings.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
1K views2 pages

CALIBRATION CERTIFICATE FORMAT Pipettor SAMPLE

This document provides a [1] certificate of calibration and calibration report for a 2-20uL pipettor. [2] The pipettor was calibrated on March 3, 2021 at San Pedro Hospital in Davao City using comparison method and standards traceable to NIST. [3] The pipettor was found to have errors of -0.00008 uL at 2 uL and +0.0028 uL at 20 uL settings.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
You are on page 1/ 2

CERTIFICATE OF CALIBRATION

Manufacturer : DLAB Certifcation Number : CAL20210303001


Model : 2-20uL Description : Pipettor
Serial Number : YM204AH0103631
Customer Name : Calibration Location :
San Pedro Hospital On-Site
Davao City

Method Used : Comparison Method Temperature : 25°C


Calibration Date : March 3, 2021 Humidity : 60 % RH
Recommended Due Date : March 3, 2022

This calibration was conducted using standards traceable to National Institute of Standards and Technology (NIST). The
user may be obliged to have the product recalibrated at appropriate intervals. This report applies only to the item
identified and shall not be reproduced, except in full, without the written approval of the calibration laboratory.
This calibration results pertains to the date of calibration. Should the unit be modified or damaged in any way, or
develop inconsistent readings, the result may not be valid and the recalibration of the unit will be required. The user
should determine the suitability of the instrument for its intended use.

Approved by:

_____________________
Iris Clint Q. Caro
Service Supervisor
CALIBRATION REPORT

Manufacturer : DLAB Certifcation Number : CAL20210303001


Model : 2-20uL Description : Pipettor
Serial Number : YM204AH0103631 Calibration Date : March 3, 2021

CALIBRATION DATA

REFERENCE
REFERENCE STANDARD READING (g)
UNIT UNDER TEST STANDARD
ERROR
SETTING(uL) MEAN READING
1 2 3 4 5
(g)
2 0.0019 0.0020 0.0018 0.0022 0.0017 0.00192 -0.00008

20 0.023 0.024 0.020 0.023 0.024 0.0228 +0.0028

Calibration Equipment Used:

MANUFACTURER MODEL SERIAL NUMBER CAL DUE DATE TRACEABILITY

BIOBASE BA2204C 039805 March 13, 2022 NIST

Performed by:

___________________