OPERATION RUN MODE DISPLAY FUNCTIONS

When the Quick Start procedure is complete...

1. to enter the RUN mode

"----" is displayed briefly...

The AIRANGER SPL rapidly takes measurements, verifies


repeatability, and calculates Readings.

The Reading is displayed continuously.

The Bar Graph indicates material level (0 to 100%),


regardless of the Operation selected. 1 Transducer 8 Data Out
2 Reading 9 Transducer Active
If "▲" is displayed the vessel level has exceeded 80% and
not yet fallen below 75% (High Alarm). 3 Units 10 Auxiliary Reading
4 High Alarm 11 Relay Number
If "▼" is displayed the vessel level has fallen below 20% and 5 Bar Graph Relay Status
not yet exceeded 25% (Low Alarm).
6 Low Alarm 12 Normal Operation
When in alarm, the corresponding relay is de-energized. 7 Filling / Emptying 13 Failsafe Operation

Alarm Relay #
High 1 RUN MODE KEYPAD FUNCTIONS
Low 2
Display mA output in 12 .

Display temperature in 12 .
2. % Display Empty / Fill Rate in 12 .

Readings are displayed in % (based upon Operation). Display Failsafe Time Left in 12 .

level space or distance Key in Parameter # to display value in 12 .

Empty ➞ Span = 0 ➞ 100% Empty ➞ Span = 100 ➞ 0% Display Material Level in 12 .

3. Display Distance in 12 .

Initiate program mode access ( see % ).


View mA output value in the Auxiliary Reading display.
% Display Units / % in 3 (complete program mode access).
level space or distance

Empty ➞ Span = 4 ➞ 20mA Empty ➞ Span = 20 ➞ 4mA PERFORMANCE EVALUATION


4. Monitor system performance, under all operating conditions.

View the Failsafe Time Left in % (before "LOE" is displayed). If alternate display, relay, or mA output operation is required,
If the value reaches 0, the "old" Reading, Bar Graph, relays, or "LOE" is displayed, see instruction manual.
and mA output are "held" and "LOE" flashes.
Connect (or enable) alarm/control equipment only after
When a valid measurement is made, the value resets to 100. satisfactory performance is verified.
The AIRANGER SPL advances to the "new" level; operating
displays and outputs accordingly.

5 6
GENERAL STEP 1 Which type of Operation is desired? STEP 4 Which Transducer is connected (check nameplate)?
(P001) (P004)

Quick Start details the minimum recommended start up 0 = off (point out of service) ULTRASONIC ULTRASONIC/TEMPERATURE
procedure for most AIRANGER SPL installations. 1 = level (how full the vessel is)
0 = not entered 100 = ST-H 106 = XPS-40
2 = space (how empty the vessel is)
Refer to the individual product instruction manuals for 1 = ST-25 101 = XCT-8 107 = XLT-30
3 = distance (from the transducer face) 2 = ST-50 102 = XPS-10 108 = XLT-60
complete installation and interconnection instructions for the:
3 = ST-100 103 = XCT-12 109 = XLS-30
e.g. for level 4 = LR-21 104 = XPS-15 110 = XLS-60
AIRANGER SPL 5 = LR-13 105 = XPS-30 111 = XKS-6
Ultrasonic Transducers 112 = XRS-5
TS-3 Temperature Sensors (if used)
BIC-II Buffered Interface Converter (if used) to advance e.g. for LR-21

Ensure all process control equipment is disabled until STEP 2 Which Material is being monitored?
satisfactory AIRANGER SPL performance is verified. (P002)
to advance
1 = liquid (flat level surface)
2 = solid (uneven sloped surface)
INITIAL START UP STEP 5 Which measurement Units are desired?
(P005)
e.g. for liquid
1. Place the programmer in the AIRANGER SPL 1 = metres (m) 4 = feet (ft)
enclosure lid recess. 2 = centimetres (cm) 5 = inches (in)
3 = millimetres (mm)
2. Apply power to the AIRANGER SPL,
to advance
e.g. for feet
!!!! is displayed.
STEP 3 Which Measurement Response is desired?
(P003)

1 = 0.1 m/min slow to advance


3. 2 = 1 m/min •
% to enter the program mode
and display . . . 3 = 10 m/min •
STEP 6 Where is Empty (transducer face to vessel bottom)?
4 = 1.7 m/sec • (P006)
5 = 17 m/sec fast (Presets Span to the maximum recommended value).

(Slower response improves measurement stability!) 0.000 to 9999

e.g. for 10 m/min e.g. for 60 Units

to advance
to advance
During the following procedure:
STEP 7 What is the Span (Empty to highest material level)?
a) If the selection desired is already displayed... (P007)
0.000 to 9999
no entry is required. (Advance to the next step.)
e.g.
b) If an entry error is made...

and try again.

Proceed to OPERATION.