Manual Analizador de Oxigeno GPR 2900
Manual Analizador de Oxigeno GPR 2900
Manual Analizador de Oxigeno GPR 2900
Approvals: CE
Power: Universal; specify 100/120/220/240 VAC or 12-28 VDC 100-240 VAC or 12-28 VDC Power
Range ID: 4-20 mA non-isolated or 1-5V; option: relay contacts
(eliminates alarms)
2 Field Selectable Alarm Setpoints
Response Time: 90% of final FS reading < 10 seconds
Power and System Failure Alarms
Sample System: None
Sensitivity: < 0.5% of FS range
Sensitivity 0.5% Full Scale
Sensor Model: GPR-11-60-4 for non-acid (CO2) gas streams 4 Ranges Standard
XLT-11-24-4 for gases containing > 0.5% CO2
Optional Equipment
ISO 9001:2008 Certified
Sample conditioning accessories - contact factory
INTERTEK Certificate No. 485
* Specification subject to change without notice.
2855 Metropolitan Place, Pomona, CA 91767 USA ♦ Tel: 909-392-6900, Fax: 909-392-3665, www.aii1.com, e-mail: [email protected] Rev 10/15
Advanced Instruments Inc.
GPR-2900
Oxygen Analyzer
Owner’s Manual
2855 Metropolitan Place, Pomona, California 91767 USA ♦ Tel: 909-392-6900, Fax: 909-392-3665, e-mail: [email protected]
Advanced Instruments Inc.
Table of Contents
Introduction 1
Quality Control Certification 2
Safety 3
Features & Specifications 4
Operation 5
Maintenance 6
Spare Parts 7
Troubleshooting 8
Warranty 9
Material Safety Data Sheets 10
Correlating readings - LCD display to 4-20mA output Appendix B
Drawings A/R
1 Introduction
Your new oxygen analyzer is a precision piece of equipment designed to give you years of use in variety of industrial oxygen
applications.
This analyzer is designed to measure the oxygen concentration in inert gases, gaseous hydrocarbons, hydrogen, a variety of gas
mixtures and acid gases containing from 0-100% CO2 (for CO2 background, the XLT series sensor must be used) present.
In order to derive maximum performance from your new oxygen analyzer, please read and follow the guidelines provided in this
Owner’s Manual.
The serial number of this analyzer may be found on the inside the analyzer. You should note the serial number in the space
provided and retains this Owner’s Manual as a permanent record of your purchase, for future reference and for warranty
considerations.
Every effort has been made to select the most reliable state of the art materials and components designed for superior
performance and minimal cost of ownership. This analyzer was tested thoroughly by the manufacturer for best performance.
However, modern electronic devices do require service from time to time. The warranty included herein plus a staff of trained
professional technicians to quickly service your analyzer is your assurance that we stand behind every analyzer sold.
Advanced Instruments Inc. appreciates your business and pledge to make effort to maintain the highest possible quality
standards with respect to product design, manufacturing and service.
2
Advanced Instruments Inc.
Software version:
Power: ( ) 100/120/220/240 VAC
( ) 9-28 VDC non-loop
Enclosure: Panel mount 7W x 4”H x 4”D (panel cutout 6”W x 3”H)
Test System start-up diagnostics satisfactory
Alarm relays activate/deactivate with changes in O2 concentration
Analog signal output 4-20mA
Alarm relays activate/deactivate with changes in O2 concentration
Alarm bypass
Range ID 4-20mA reflects range changes
Baseline drift on zero gas < ± 2% FS over 24 hour period
Noise level < ± 1.0% FS
Span adjustment within 10-50% FS
Final Overall inspection for physical defects
Options:
Notes 1 of 1 due ASAP
3
Advanced Instruments Inc.
3 Safety Guidelines
General
This section summarizes the essential precautions applicable to the GPR-2900 Oxygen Analyzer. Additional precautions specific
to individual analyzer are contained in the following sections of this manual. To operate the analyzer safely and obtain
maximum performance follow the basic guidelines outlined in this Owner’s Manual.
Caution: This symbol is used throughout the Owner’s Manual to Caution and alert the user to recommended safety and/or
operating guidelines.
Danger: This symbol is used throughout the Owner’s Manual to identify sources of immediate Danger such as the presence of
hazardous voltages.
Retain Instructions: The safety precautions and operating instructions found in the Owner’s Manual should be retained for
future reference.
Heed Warnings: Follow all warnings on the analyzer, accessories (if any) and in this Owner’s Manual.
Follow Instructions: Observe all precautions and operating instructions. Failure to do so may result in personal injury or
damage to the analyzer.
Installation
Oxygen Sensor: DO NOT open the sensor. The sensor contains a corrosive liquid electrolyte that could be harmful if touched
or ingested, refer to the Material Safety Data Sheet contained in the Owner’s Manual appendix. Avoid contact with any liquid or
crystal type powder in or around the sensor or sensor housing, as either could be a form of electrolyte. Leaking sensors should
be disposed of in accordance with local regulations.
Mounting: The analyzer is approved for indoor use, outdoor use requires mounting in a secondary enclosure with the
appropriate NEMA rating. Mount as recommended by the manufacturer.
Power: Supply power to the analyzer only as rated by the specification or markings on the analyzer enclosure. The wiring that
connects the analyzer to the power source should be installed in accordance with recognized electrical standards and so they
are not pinched particularly near the power source and the point where they attach to the analyzer. Never yank wiring to
remove it from an outlet or from the analyzer.
Heat: Situate and store the analyzer away from sources of heat.
Liquid and Object Entry: The analyzer should not be immersed in any liquid. Care should be taken so that liquids are not
spilled into and objects do not fall into the inside of the analyzer.
Handling: Do not use force when using the switches and knobs. Before moving your analyzer be sure to disconnect the
wiring/power cord and any cables connected to the output terminals located on the analyzer.
4
Advanced Instruments Inc.
Maintenance
Serviceability: Except for replacing the oxygen sensor, there are no parts inside the analyzer for the operator to service.
Only trained personnel with the authorization of their supervisor should conduct maintenance.
Oxygen Sensor: DO NOT open the sensor. The sensor contains a corrosive liquid electrolyte that could be harmful if touched
or ingested, refer to the Material Safety Data Sheet contained in the Owner’s Manual appendix. Avoid contact with any liquid or
crystal type powder in or around the sensor or sensor housing, as either could be a form of electrolyte. Leaking sensors should
be disposed of in accordance with local regulations.
Troubleshooting: Consult the guidelines in Section 8 for advice on the common operating errors before concluding that your
analyzer is faulty.
Do not attempt to service the analyzer beyond those means described in this Owner’s Manual. Do not attempt to make repairs
by yourself as this will void the warranty as per Section 10 and may result in electrical shock, injury or damage. All other
servicing should be referred to qualified service personnel.
Cleaning: The analyzer should be cleaned only as recommended by the manufacturer. Wipe off dust and dirt from the outside
of the unit with a soft damp cloth then dry immediately. Do not use solvents or chemicals.
Nonuse Periods: If the analyzer is equipped with a range switch advance the switch to the OFF position and disconnect the
power when the analyzer is left unused for a long period of time.
5
Advanced Instruments Inc.
5 Operations
Principle of Operation
The GPR-2900 Oxygen Analyzer incorporates a variety of percentage range advanced galvanic fuel cell type sensors. This model
is configured for panel mounting and requires a 6”W x 3”H cutout with 4 holes for the studs located on the back side of the
analyzer’s front panel.
Optional mounting configurations include a 19” rack or wall mount enclosure with or without a sample system. Contact the
factory for additional information on options.
Oxygen, the fuel for this electrochemical transducer, diffusing into the sensor reacts chemically at the sensing electrode to
produce an electrical current output proportional to the oxygen concentration in the gas phase. The sensor’s signal output is
linear over all ranges and remains virtually constant over its useful life. The sensor requires no maintenance and is easily and
safely replaced at the end of its useful life.
Proprietary advancements in design and chemistry add significant advantages to an extremely versatile oxygen sensing
technology. The expected life of our new generation of percentage range sensors now range to five and ten years with faster
response times and greater stability. Another significant development involves expanding the operating temperature range for
percentage range sensors from -30°C to 50°C.
Electronics:
The signal generated by the sensor is processed by state of the art low power micro-processor based digital circuitry. The first
stage amplifies the signal. The second stage eliminates the low frequency noise. The third stage employs a high frequency filter
and compensates for signal output variations caused by ambient temperature changes. The result is a very stable signal.
Sample oxygen is analyzed very accurately. Response time of 90% of full scale is less than 10 seconds (actual experience may
vary due to the integrity of sample line connections, dead volume and flow rate selected) on all ranges under ambient
monitoring conditions. Sensitivity is typically 0.5% of full scale low range.
Additional features of the micro-processor based electronics include manual or auto ranging, isolated 4-20mA signal for signal
output and range ID. Whenever the analyzer is calibrated, a unique algorithm predicts and displays a message indicating a
‘weak sensor’ suggesting the sensor be replaced in the near future.
Users interested in adding their own sample handling or conditioning system are encouraged to consult the factory to ensure all
applicable conditions are addressed to ensure proper operation of the analyzer. Advanced Instruments Inc. offers a full line of
sample handling, conditioning and expertise to meet your application requirements. Contact us at 909-392-6900 or e-mail us at
[email protected]
6
Advanced Instruments Inc.
Temperature: The rate oxygen molecules diffuse into the sensor is controlled by a Teflon membrane otherwise known as an
'oxygen diffusion limiting barrier' and all diffusion processes are temperature sensitive, the fact the sensor's electrical output
will vary with temperature is normal. This variation is relatively constant 2.5% per ºC. A temperature compensation circuit
employing a thermistor offsets this effect with an accuracy of +5% or better and generates an output function that is
independent of temperature. There is no error if the calibration and sampling are performed at the same temperature or if the
measurement is made immediately after calibration. Lastly, small temperature variations of 10-15º produce < 1% error.
Accuracy: In light of the above parameters, the overall accuracy of an analyzer is affected by two types of errors: 1) those
producing 'percent of reading errors', illustrated by Graph A below, such as +5% temperature compensation circuit, tolerances
of range resistors and the 'play' in the potentiometer used to make span adjustments and 2) those producing 'percent of full
scale errors', illustrated by Graph B, such as +1-2% linearity errors in readout devices, which are really minimal due to today's
technology and the fact that other errors are 'spanned out' during calibration.
Graph C illustrates these 'worse case' specifications that are typically used to develop an analyzer's overall accuracy statement
of < 1% of full scale at constant temperature or < 5% over the operating temperature range. QC testing is typically < 0.5%
prior to shipment.
Example 1: As illustrated by Graph A any error, play in the multi-turn span pot or the temperature compensation circuit,
during a span adjustment at 20.9% (air) of full scale range would be multiplied by a factor of 4.78 (100/20.9) if used for
measurements of 95-100% oxygen concentrations. Conversely, an error during a span adjustment at 100% of full scale range is
reduced proportionately for measurements of lower oxygen concentrations.
Zero Calibration: In theory, the electrochemical galvanic fuel cell type oxygen has an absolute zero meaning it produces no
signal output when exposed to an oxygen free sample gas. In reality, expect the analyzer to generate an oxygen reading when
sampling oxygen free sample gas due to contamination or quality of the zero gas; minor leakage in the sample line connections;
residual oxygen dissolved in the sensor’s electrolyte; and, tolerances of the electronic components. The Zero Offset capability of
the analyzer is limited to 50% of lowest most sensitive range available with the analyzer.
7
Advanced Instruments Inc.
Recommendation 1: Zero calibration, see Determining True Zero Offset below, is recommended only for online analyzers
performing continuous analysis below 5% of the lowest most sensitive range available with a ppm analyzer, e.g. analysis below
0.5 ppm on the 10 ppm range, or below 0.1% (1000 ppm) with a percent analyzer. Note 1: Once the zero offset adjustment is
made, zero calibration is not required again until the sample system connections are modified, or, when installing a new oxygen
sensor. As a result, zero calibration is not practical and therefore not recommended for higher ranges or portable analyzers.
Determining True Zero Offset: Allow the analyzer approximately 24 hours to stabilize with flowing zero gas as evidenced by a
stable reading or horizontal trend on an external recording device. Note 2: 24 hours is required to assure the sensor has
consumed the oxygen that has dissolved into the electrolyte inside the sensor while exposed to air or percentage levels of
oxygen. For optimum accuracy, utilize as much of the actual sample system as possible.
Span Calibration: Involves adjusting the analyzer electronics to the sensor’s signal output at a given oxygen standard.
Regardless of the oxygen concentration of the oxygen standard used, a typical span calibration takes approximately 10 minutes.
Note 3: The amount time required to get the analyzer back on line for normal use is influenced by a.) the level of oxygen
analysis anticipated during normal operation (also determines the initial analyzer selection), and, b.) whether the sensor is new
or has been in service for at least two weeks.
General guidelines for analyzers to come online following span calibration and exposure to a zero/purge/sample gas with an
oxygen content below the stated thresholds:
¾ measurements above 1000 ppm or 0.1% require less than 3 minutes
¾ measurements above 100 ppm (parts-per-million analyzer) require less than 10 minutes
¾ measurements below 10 ppm (part-per-million analyzer) require 20 minutes if the sensor has been in service at ppm
levels for at least two weeks, and, 60 minutes if the sensor is new assuming the zero/purge/sample gas has an oxygen
concentration below 1 ppm
Recommendation 2: For 'optimum calibration accuracy' calibrate with a span gas approximating 80% of the full scale range
one or two ranges higher than the full scale range of interest (normal use) to achieve the effect illustrated on Graph A and
Example 1. Always calibrate at the same temperature and pressure of the sample gas stream.
Note 4: Calibrating with a span gas approximating 10% of the full scale range near the expected oxygen concentration of the
sample gas is acceptable but less accurate than ‘optimum calibration accuracy’ method recommended – the method usually
depends on the gas available. Calibrating at the same 10% of the full scale range for measurements at the higher end of the
range results in magnification of errors as discussed in Graph A and Example 1 and is not recommended. Of course the user can
always elect at his discretion to accept an accuracy error of +2-3% of full scale range if no other span gas is available.
Air Calibration: Based on the inherent linearity of the electrochemical galvanic fuel cell type oxygen sensor enables the user to
calibrate the analyzer with ambient air (20.9% oxygen) and operate the analyzer within the stated accuracy spec on the lowest
most sensitive range available with the analyzer – there is no need to recalibrate the analyzer with span gas containing a lower
oxygen concentration. Except for Oxygen Purity Analyzers intended to measure elevated oxygen levels ranging from 50-100%
oxygen, calibrating either a ppm or percent analyzer with ambient air on the CAL or 0-25% range meets the 80% criteria
discussed in Recommedation 2.
Recommendation 3: Air calibrate the analyzer (with the exception of Oxygen Purity Analyzers intended to measure elevated
oxygen levels ranging from 50-100% oxygen) when operating a percent analyzer, installing and replacing a ppm oxygen sensor,
to verify the oxygen content of a certified span gas or when a certified span gas is not available to calibrate a ppm analyzer
(immediately following air calibration reintroduce a gas with a low oxygen concentration to expedite the return to ppm level
measurements as per Note 3).
8
Advanced Instruments Inc.
Installation Considerations
Gas Sample Stream: Ensure the gas stream composition of the application is consistent with the specifications and review the
application conditions before initiating the installation. Consult the factory if necessary to ensure the sample is suitable for
analysis.
Note: In natural gas applications such as extraction and transmission, a low voltage current is applied to the pipeline itself to
inhibit corrosion. As a result, electronic devices can be affected unless adequately grounded.
Contaminant Gases: A gas scrubber and flow indicator with integral metering valve are required upstream of the analyzer to
remove interfering gases such as oxides of sulfur and nitrogen or hydrogen sulfide that can produce false readings, reduce the
expected life of the sensor and void the sensor’s warranty if not identified at time of order placement. Installation of a suitable
scrubber is required to remove the contaminant from the sample gas to prevent erroneous analysis readings and damage to the
sensor or optional components. Consult the factory for recommendations concerning the proper selection and installation of
components.
Expected Sensor Life: With reference to the publish specification located as the last page of this manual, the expected life of
all oxygen sensors is predicated on oxygen concentration (< 1000 ppm or air), temperature (77°F/25°C) and pressure (1
atmosphere) in “normal” applications. Deviations are outside the specifications and will affect the life of the sensor. As a rule of
thumb sensor life is inversely proportional to changes in the parameters.
Optimum Accuracy: Determine if Zero Calibration is recommended for your application. If it is Zero Calibration should
precede Span Calibration and both should be repeated after the analyzer has been allowed to stabilize, typically 24-36 hours
after installation. For Span Calibration use a certified span gas with an oxygen content (balance nitrogen) approximating 80% of
the next higher full scale range above the intended measuring range is recommended for optimum accuracy, see Calibration
and Accuracy.
Assuming the initial zero is performed according to the procedure described herein, the analyzer should not require Zero
Calibration again until the either the sensor is replaced or a change is made to the sample system or gas lines, and, it should
not require Span Calibration again for up to 3 months under “normal” application conditions as described in the published
specifications. One of the unique features of analyzers based on the electrochemical galvanic fuel cell type oxygen sensor is the
fact that it can be field calibrated at the user’s discretion to whatever standard of certified span gas the user elects to use.
Zero Calibration: In theory, the oxygen sensor produces no signal output when exposed to an oxygen free sample gas. In
reality, expect the analyzer to generate an oxygen reading when sampling oxygen free sample gas due to contamination or
quality of the zero gas; minor leakage in the sample line connections; residual oxygen dissolved in the sensor’s electrolyte; and,
tolerances of the electronic components.
Zero calibration, see Determining True Zero Offset below, is recommended only for online analyzers performing continuous
analysis below 5% of the lowest most sensitive range available with a ppm analyzer, e.g. analysis below 0.5 ppm on the 10 ppm
range, or below 0.1% (1000 ppm) with a percent analyzer. Note : Once the zero offset adjustment is made, zero calibration is
not required again until the sample system connections are modified, or, when installing a new oxygen sensor. As a result, zero
calibration is not practical and therefore not recommended for higher ranges or portable analyzers.
Determining True Zero Offset: Allow the analyzer approximately 24 hours to stabilize with flowing zero gas as evidenced by a
stable reading or horizontal trend on an external recording device. Note: 24 hours is required to assure the sensor has
consumed the oxygen that has dissolved into the electrolyte inside the sensor while exposed to air or percentage levels of
oxygen. For optimum accuracy, utilize as much of the actual sample system as possible.
Span Calibration: Involves adjusting the analyzer electronics to the sensor’s signal output at a given oxygen standard, e.g. a
certified span gas with an oxygen content (balance nitrogen) approximating 80% of the next higher full scale range above the
intended measuring range is recommended for optimum accuracy, see Calibration and Accuracy.
Recommendation: Based on the inherent linearity of the galvanic oxygen sensor enables the user to calibrate the analyzer with
ambient air (20.9% oxygen) and operate the analyzer within the stated accuracy spec on the lowest most sensitive range
available with the analyzer – there is no need to recalibrate the analyzer with span gas containing a lower oxygen
concentration. Calibrating either a ppm or percent analyzer with ambient air (with the exception of Oxygen Purity Analyzers
intended to measure elevated oxygen levels ranging from 50-100% oxygen) on the CAL or 0-25% range meets the 80% criteria
discussed above.
9
Advanced Instruments Inc.
Air calibrate the analyzer (with the exception of Oxygen Purity Analyzers intended to measure elevated oxygen levels ranging
from 50-100% oxygen) when operating a percent analyzer, installing and replacing a ppm oxygen sensor, to verify the oxygen
content of a certified span gas or when a certified span gas is not available to calibrate a ppm analyzer (immediately following
air calibration reintroduce a gas with a low oxygen concentration to expedite the return to ppm level measurements).
Materials: Assemble the necessary zero, purge and span gases and optional components such as valves, coalescing or
particulate filters, and, pumps as dictated by the application; stainless steel tubing is essential for maintaining the integrity of
the gas stream for ppm and percentage range (above or below ambient air) analysis; hardware for mounting.
Temperature: The sample must be sufficiently cooled before it enters the analyzer and any optional components. A coiled 10
foot length of ¼” stainless steel tubing is sufficient for cooling sample gases as high as 1,800ºF to ambient.
Pressure & Flow: All electrochemical oxygen sensors respond to partial pressure changes in oxygen. The sensors are equally
capable of analyzing the oxygen content of a flowing sample gas stream or monitoring the oxygen concentration in ambient air
(such as a confined space such in a control room or an open area such as a landfill or bio-pond).
Sample systems and/or flowing gas samples are generally required for applications involving oxygen measurements at levels
other than ambient air and when the pressure exceeds ambient. In these situations, the use of stainless steel tubing and fittings
is critical to maintaining the integrity of the gas stream to be sampled and the inlet pressure must always be higher than the
pressure at the outlet vent which is normally at atmospheric pressure. The sensor is exposed to sample gas that must flow or
be drawn through metal tubing via 1/8” compression inlet and vent fittings, a stainless steel sensor housing with an o-ring seal
to prevent the leakage of air and stainless steel tubing.
Flow rates of 1-5 SCFH cause no appreciable change in the oxygen reading. However, flow rates above 5 SCFH generate
backpressure and erroneous oxygen readings because the diameter of the integral tubing cannot evacuate
the sample gas at the higher flow rate. The direction the sample gas flows is not important, thus either
tube fitting can serve as the inlet or vent – just not simultaneously.
A flow indicator with an integral metering valve upstream of the sensor is recommended as a means of
controlling the flow rate of the sample gas. A flow rate of 2 SCFH or 1 liter per minute is recommended for
optimum performance.
Caution: Do not place your finger over the vent (it pressurizes the sensor) to test the flow indicator when
gas is flowing to the sensor. Removing your finger (the restriction) generates a vacuum on the sensor and
may damage the sensor (voiding the sensor warranty). To avoid generating a vacuum on the sensor (as
described above) during operation, always select and install the vent fitting first and remove the vent
fitting last.
Application Pressure - Positive: A flow indicator with integral metering valve positioned upstream of
the sensor is recommended for controlling the sample flow rate between 1-5 SCFH. To reduce the possibility of leakage for low
ppm measurements, position a metering needle valve upstream of the sensor to control the flow rate and position a flow
indicator downstream of the sensor. If necessary, a pressure regulator (with a metallic diaphragm is recommended for optimum
accuracy, the use of diaphragms of more permeable materials may result in erroneous readings) upstream of the flow control
valve should be used to regulate the inlet pressure between 5-30 psig.
Application Pressure - Atmospheric or Slightly Negative: For accurate ppm range oxygen measurements, an optional
external sampling pump should be positioned downstream of the sensor to draw the sample from the process, by the sensor
and out to atmosphere. A flow meter is generally not necessary to obtain the recommended flow rate with most sampling
pumps.
Caution: If the analyzer is equipped with an optional flow indicator with integral metering valve or a metering flow control
valve upstream of the sensor - open the metering valve completely to avoid drawing a vacuum on the sensor and placing an
undue burden on the pump. If pump loading is a consideration, a second throttle valve on the pump’s inlet side may be
necessary to provide a bypass path so the sample flow rate is within the above parameters.
10
Advanced Instruments Inc.
Moisture & Particulates: Installation of a suitable coalescing or particulate filter is required to remove condensation, moisture
and/or particulates from the sample gas to prevent erroneous analysis readings and damage to the sensor or optional
components. Moisture and/or particulates do not necessarily damage the sensor, however, collection on the sensing surface can
block or inhibit the diffusion of sample gas into the sensor resulting in a reduction of sensor signal output – and the appearance
of a sensor failure when in fact the problem is easily remedied by blowing on the front of the sensor. Consult the factory for
recommendations concerning the proper selection and installation of components.
Gas Connections: Inlet and outlet vent gas lines for ppm analysis require 1/8” or ¼” stainless steel compression fittings; hard
plastic tubing with a low permeability factor can be used percentage range measurements.
Power Connection: Locate the appropriate source of to meet the analyzer or analyzer requirements, ensure that is properly
grounded and meets the area classification.
Procedure:
1. The GPR-2900 front panel measures 7”W x 4”H x 4.5”D. This
compact configuration is designed for panel mounting directly
to any flat vertical surface, wall or bulkhead plate with the
appropriate 6”W x 3”H cut out and four ¼” diameter holes for
insertion of the mounting studs located on the back side of
the front panel.
2. When mounting the analyzer position it approximately 5 feet
off the floor for viewing purposes and allow sufficient room
for access to the terminal connections at the rear of the
enclosure. Note: The proximity of the analyzer to the sample
point and use of optional sample conditioning components have an impact on sample lag time.
3. Position the sensor flow housing along any flat surface. The base of sensor flow housing is fabricated with two 6/32
mounting holes. The oxygen sensor is not position sensitive but it is recommended to orient the sensor flow housing with
the female threaded opening facing the ceiling.
4. Next screw the sensor into sensor flow housing, snug the sensor’s sealing o-ring but do not over tighten or crush it.
5. Register the female plug molded to the cable with the mating male connector attached to the rear of the sensor.
6. Remove the shorting device, normally a formed spring type wire, from the sensor, push the registered mating connectors
together and turn the knurled lock ring on the molded plug until finger tight on the sensor’s connector.
7. Connect the four wires of the cable, following the color coding above the terminal block, at the rear of the analyzer.
8. Attach the flow housing to the mounting position determine above.
11
Advanced Instruments Inc.
Gas Connections:
The GPR-2900 with its standard flow through configuration is designed for positive pressure samples and requires connections
for incoming sample and outgoing vent lines. The user is responsible for calibration gases and the required components, see
below. Flow rates of 1-5 SCFH cause no appreciable change in the oxygen reading. However, flow rates above 5 SCFH generate
backpressure and erroneous oxygen readings because the diameter of the integral tubing cannot evacuate the sample gas at
the higher flow rate. A flow indicator with an integral metering valve upstream of the sensor is recommended as a means of
controlling the flow rate of the sample gas. A flow rate of 2 SCFH or 1 liter per minute is recommended for optimum
performance.
Caution: Do not place your finger over the fitting designated as the vent (it pressurizes the sensor) or to test the flow indicator
when gas is flowing to the sensor. Removing your finger (the restriction) generates a vacuum on the sensor and may damage
the sensor (voiding the sensor warranty).
Procedure:
1. Caution: Do not change the factory setting until instructed.
2. Designate the female quick disconnect fittings, right side of the analyzer, as inlet and vent respectively.
3. Regulate the pressure and flow as described in Pressure & Flow above.
4. Install one mating male vent fitting into the female quick disconnect fitting designated as the VENT – connection of 1/8”
dia. metal vent line (requires an 1/8” male NPT to tube adapter) is optional.
5. Connect the second mating male fitting to 1/8” dia. metal sample line using a 1/8” male NPT to tube adapter.
6. Connect the third mating male fitting to 1/8” dia. metal span gas line using a 1/8” male NPT to tube adapter.
7. Install either the sample or span mating male fitting into the female quick disconnect fitting designated as SAMPLE.
8. Set the flow rate to 2 SCFH (open the flow control valve completely if using an external sampling pump positioned
downstream of the sensor).
9. Allow gas to flow through the analyzer for 3-5 minutes and proceed to Calibration or Sampling.
Electrical Connections:
The appropriate AC or DC power requirement must specified be
specified at order placement.
Filtered terminal blocks are provided for power failure, set point
alarm relays and signal output connections. The PCB also
includes a transformer to power the relays, a sensor bypass
switch for troubleshooting the electronics
Procedure:
1. As illustrated above the sensor, power and alarm relays and signal output connections are hard wired to screw type
terminal blocks located at the rear of the analyzer.
2. Use a small bladed screwdriver to loosen the appropriate terminal screws as illustrated above.
3. Strip the wires of the cable no more than 3/16 inch.
12
Advanced Instruments Inc.
4. To connect to an active relay or “fail safe”, connect the live cable to the common terminal C and the secondary cable to the
normally open NO terminal.
5. To break the connection upon relay activation, connect the secondary cable to the normally closed NC terminal.
6. Insert the stripped end of the cables into the appropriate terminal slots assuring no bare wire remains exposed that could
come in contact with the back panel of the analyzer enclosure.
7. Tighten the terminal screws to secure the wires of the cable.
Danger: While connecting the cables to the relay terminals, ensure there is no voltage on the cables to prevent electric shock
and possible damage to the analyzer. Caution: Assure the stripped wire ends of the cable are fully inserted into the terminal
slots and do not touch each other or the back panel of the analyzer enclosure.
Alarm Relays
Alarm 1 and Alarm 2 represent two threshold type alarms that can be configured in the field from the analyzer’s menu driven
LCD display as follows:
The alarm set point represents a value. When the oxygen reading exceeds (high alarm) or falls below (low alarm) the alarm set
point, the relay is activated and the LCD displays the alarm condition. When activated the alarms trigger SPDT Form C non-
latching relays @ 5A, 30VDC or 240VAC resistive. To prevent chattering of the relays, a 2% hysteresis is added to the alarm set
point. This means that the alarm will remain active until the oxygen reading has fallen 2% below the alarm set point (high
alarm) or risen 2% above the alarm set point (low alarm) after the alarm was activated. Aside from being totally defeated in the
Alarm Bypass mode, the timeout feature is useful while replacing the oxygen sensor or during calibration when the oxygen
reading might well rise above the alarm set point and trigger a false alarm.
Note: When making connections the user must decide whether to configure/connect Alarm 1 and Alarm 2 in failsafe mode
(Normally Open – NO – where the alarm relay de-energizes and closes in an alarm condition) or non-failsafe mode (Normally
Closed – NC – where alarm relay energizes and opens in an alarm condition).
For range ID the output of 4mA, 8mA, 12mA, 16mA, 20mA correspond to the most sensitive to least sensitive analysis range.
Caution: The integral 4-20mA converters are internally powered and do not require external power. DO NOT supply any
voltage to any of the two terminals of the 4-20mA converter.
13
Advanced Instruments Inc.
Caution: DO NOT open the oxygen sensor. The sensor contains a corrosive liquid electrolyte that could be harmful if touched
or ingested, refer to the Material Safety Data Sheet contained in the Owner’s Manual appendix. Avoid contact with any liquid or
crystal type powder in or around the sensor or sensor housing, as either could be a form of electrolyte. Leaking sensors should
be disposed of in manner similar to that of a common battery in accordance with local regulations.
Procedure:
1. As described above the following steps should already be completed:
a) Locate and secure the sensor flow housing with two 6/32 mounting screws, in the preferred position the threaded
female should be facing the ceiling;
b) connect the gas lines;
c) electrical connections.
2. Open the barrier bag containing the new sensor.
3. The sensor is equipped with a wire spring type shorting device, remove the shorting wire from the pins of the male
receptacle attached to the new sensor.
4. Line up the registration point of the female plug molded to the cable and male receptacle on the oxygen sensor.
5. Push the female plug (including the knurled lock ring) into the male receptacle attached to the oxygen sensor.
6. Screw and finger tighten the knurled lock ring of the cable plug into to the male connector attached to the sensor.
7. Place the analyzer in the auto ranging mode, see below.
8. With the sensor exposed to ambient air wait for 5-10 minutes until the reading stabilizes.
9. Perform and air calibration as described below.
10. Screw the new sensor, finger tight plus ¼ turn, into the threaded hole in the flow housing and ensure the o-ring seal is
engaged but not ‘crushed’.
11. Begin sampling.
Caution: Do not contaminate the span gas cylinder when connecting the regulator. Bleed the air filled regulator (faster and
more reliable than simply flowing the span gas) before attempting the initial calibration of the instrument.
Required components:
¾ Certified span gas cylinder with an oxygen concentration, balance nitrogen, approximating 80% of the full scale range
above the intended measuring range.
¾ Regulator to reduce pressure to between 5 and 30 psig.
¾ Flow meter to set the flow between 1-5 SCFH,
¾ Suitable fittings and 1/8” dia. 4-6 ft. in length of metal tubing to connect the regulator to the flow meter inlet
¾ Suitable fitting and 1/8” dia. 4-6 ft. in length of metal tubing to connect from the flow meter vent to tube fitting designated
SAMPLE IN on the GPR-2900.
Procedure:
1. With the span gas cylinder valve closed, install the regulator on the cylinder.
2. Open the regulator’s exit valve and partially open the pressure regulator’s control knob.
3. Open slightly the cylinder valve.
4. Loosen the nut connecting the regulator to the cylinder and bleed the pressure regulator.
5. Retighten the nut connecting the regulator to the cylinder
6. Adjust the regulator exit valve and slowly bleed the pressure regulator.
7. Open the cylinder valve completely.
8. Set the pressure between 5-30 psig using the pressure regulator’s control knob.
9. Caution: Do not exceed the recommended flow rate. Excessive flow rate could cause the backpressure on the sensor
and may result in erroneous readings and permanent damage to the sensor.
14
Advanced Instruments Inc.
START-UP TEST
ELECTRONICS – PASS
TEMP SENSOR – PASS
BAROMETRIC SENSOR – PASS
REV. 1.61
Note: The analyzer display defaults to the sampling mode when 30 seconds elapses without user interface.
3.3%
AUTO SAMPLING
10 % RANGE
24.5 C
LO1 1% 10% HI2
Menu Navigation
The four (4) pushbuttons located on the front of the analyzer
operate the micro-processor:
1. blue ENTER (select)
2. yellow UP ARROW
3. yellow DOWN ARROW
4. green MENU (escape)
Main Menu
Access the MAIN MENU by pressing the MENU key:
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
CALIBRATION
CONFIG ALARMS
BYPASS ALARMS
Range Selection
The GPR-2900 analyzer is equipped with four (4) standard measuring ranges (see specification) and provides users with a
choice of sampling modes. By accessing the MAIN MENU, users may select either the AUTO SAMPLING (ranging) or MANUAL
SAMPLING (to lock on a single range) mode.
15
Advanced Instruments Inc.
Note: For calibration purposes, use of the AUTO SAMPLE mode is recommended. However, the user can select the full scale
MANUAL SAMPLE RANGE for calibration as dictated by the accuracy of the analysis required – for example, a span gas with an
8.0% oxygen concentration with the balance nitrogen would dictate the use of the 0-10% full scale range for calibration and a
0-5% or 0-1% measuring ranges.
Auto Sampling:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight AUTO SAMPLE.
3. Press the ENTER key to select the highlighted menu option.
4. The display returns to the sampling mode:
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
CALIBRATION
3.3 %
AUTO SAMPLING
CONFIG ALARMS 5 % RANGE
BYPASS ALARMS 24.5 C
LO1 1% 10% HI2
The display will shift to the next higher range when the oxygen reading (actually the sensor’s signal output) exceeds 99.9% of
the upper limit of the current range. The display will shift to the next lower range when the oxygen reading drops to 85% of the
upper limit of the next lower range.
For example, if the analyzer is reading 1% on the 0-10% range and an upset occurs, the display will shift to the 0-25% range
when the oxygen reading exceeds 9.9%. Conversely, once the upset condition is corrected, the display will shift back to the 0-
10% range when the oxygen reading drops to 8.5%.
Manual Sampling:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight MANUAL SAMPLE.
3. Press the ENTER key to select the highlighted menu option.
4. The following display appears:
5. Advance the reverse shade cursor using the ARROW keys to highlight the desired MANUAL RANGE.
6. Press the ENTER key to select the highlighted menu option.
7. The following displays appears with the range selected and oxygen concentration of the sample gas:
MANUAL RANGE
25%
10% >>>
3.3%
MANUAL SAMPLING
5% 5 % RANGE
1% 24.5 C
LO1 1% 10% HI2
8. The display will not shift automatically. Instead, when the oxygen reading (actually the sensor’s signal output) exceeds
110% of the upper limit of the current range an OVER RANGE warning will be displayed.
16
Advanced Instruments Inc.
9. Once the OVER RANGE warning appears the user must advance the analyzer to the next higher range via the menu and
keypad Press MENU, select MANUAL SAMPLING, press ENTER, select the appropriate MANUAL RANGE and press ENTER
again.
Alarms
The CONFIG ALARMS features a system that can be configured in the field. Two field adjustable alarm relays with dry contacts
operate independently of one another which means the alarms can be set-up as:
¾ HI and LO
¾ LO and LO, LO,
¾ HI and HI,HI
Additional feature includes delaying the activation of the local audible alarm and relay contacts for up 99 minutes to enable
users to distinguish between transient occurrences and true upset conditions which is particularly useful on remote applications
without affecting the 4-20mA signal output. The local audible alarm can be silenced or disabled as well without affecting the 4-
20mA signal output.
Note: A separate feature, BYPASS ALARMS described below, enables the user to disable the local audible alarm and relay
contacts during calibration or servicing. The alarms are enabled when the alarm condition is corrected.
5. Advance the reverse shade cursor using the ARROW keys to highlight the SET ALARM 1 option.
6. Press the ENTER key to select the highlighted menu option.
7. The following displays appears with PERCENT as the default alarm value :
8. Advance the reverse shade cursor using the ARROW keys to highlight the desired option.
9. Press the ENTER key to select the highlighted menu option.
17
Advanced Instruments Inc.
10. Note: The PERCENT alarm value is entered with one decimal, the PPM alarm value is entered as an integer.
01.0
GAS CONCENTRATION
SET ALARM 1 VALUE
PERCENT
PPM >>> PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
002
GAS CONCENTRATION
SET ALARM 1 VALUE
PERCENT
>>> PRESS UP OR DOWN
PPM
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
11. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the alarm value.
12. Press the ARROW keys to enter the alarm value.
13. Repeat steps 11 and 12 until the complete span value has been entered.
14. Note: If an alarm is set as a PERCENT value and subsequently changed to a PPM value, the PERCENT value is not retained
and is reset to 00.0. This holds if the alarm was first set as PPM value and then changed to a PERCENT value.
15. Save the alarm value by pressing the ENTER key or abort by pressing the MENU key.
16. The system returns to the SAMPLING mode and displays:
3.3%
AUTO SAMPLING
10 % RANGE
24.5 C
LO1 2% 10% HI2
18
Advanced Instruments Inc.
5. Advance the reverse shade cursor using the ARROW keys to highlight the SET ALARM DELAY.
6. Press the ENTER key to select the highlighted menu option.
7. The following displays appear with last alarm delay value :
99
MAIN MENU
DELAY IN MINUTES
SET ALARM 1
SET ALARM 2 >>> PRESS UP OR DOWN
SET ALARM DELAY TO CHANGE VALUE
ALARM 1 HI/LO ENTER TO SAVE
ALARM 2 HI/LO MENU TO RETURN
ALARMS AUDIBLE/SILENT
8. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the alarm value.
9. Press the ARROW keys to enter the alarm value.
10. Repeat steps 17 and 18 until the complete span value has been entered.
11. Save the alarm value by pressing the ENTER key or abort by pressing the MENU key.
12. The system returns the SAMPLING mode and displays:
3.3%
AUTO SAMPLING
10 % RANGE
24.5 C
LO1 2% 10% HI2
19
Advanced Instruments Inc.
5. Advance the reverse shade cursor using the ARROW keys to highlight the ALARM 1 option, which appears as either ALARM
1 HI or ALARM 1 LO.
6. Press the ENTER key to toggle and change the displayed setting. After 3 seconds, the system returns to SAMPLING mode.
3.3%AUTO SAMPLING
10 % RANGE
24.5 C
LO1 2% 10% HI2
5. Advance the reverse shade cursor using the ARROW keys to highlight the ALARMS AUDIBLE/SILENT option, which appear
as either ALARMS AUDIBLE or ALARMS SILENT.
6. Press the ENTER key to toggle and change the displayed setting. After 3 seconds, the system returns to SAMPLING mode.
3.3%
AUTO SAMPLING
10 % RANGE
24.5 C
LO1 2% 10% HI2
20
Advanced Instruments Inc.
Bypass Alarms:
This feature, separate from CONFIG ALARMS above, enables the user to disable the local audible alarm and relay contacts
during calibration or servicing. The alarms are enabled when the alarm condition is corrected.
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight BYPASS ALARMS.
3. The following displays appears:
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
CALIBRATION
CONFIG ALARMS
BYPASS ALARMS
4. Press the ENTER key to bypass and disable both the local audible alarm and relay contacts. After 3 seconds, the system
returns to SAMPLING mode.
Note: The appropriate alarm setting will alternately reverse shades indicating the alarm condition exists but the BYPASS
ALARMS feature has disabled the local audible alarm and relay contact. The alarms are enabled when the alarm condition is
corrected.
3.3%
AUTO SAMPLING
10 % RANGE
24.5 C
LO1 2% 10% HI2
21
Advanced Instruments Inc.
Zero Calibration
In theory, the oxygen sensor produces no signal output when exposed to an oxygen free sample gas. In reality, expect the
analyzer to generate an oxygen reading when sampling oxygen free sample gas due to:
¾ Contamination or quality of the zero gas
¾ Minor leakage in the sample line connections
¾ Residual oxygen dissolved in the sensor’s electrolyte
¾ Tolerances of the electronic components
Recommendation: Zero calibration is recommended only for online analyzers intended for continuous precise measurements
below 0.1% on the 0-1% range and then only when the sample system connections are modified and when installing a new
oxygen sensor. It is not practical on higher ranges or portable analyzers for the following reasons:
1. Determining the true zero offset reguires the user allow the analyzer approximately 24 hours to stabilize with flowing zero
gas as evidenced by a stable reading with no downward trend on an external recording device. Note: Approximately 24-36
hours is required to assure the sensor has consumed the oxygen that has dissolved into the electrolyte inside the sensor
while exposed to air or percentage levels of oxygen. For optimum accuracy, utilize as much of the actual sample system as
possible.
2. Thus it is not practical to find the true zero offset particularly in the case of applications requiring higher level oxygen
measurements because of the low offset value, normally 50% of the most sensitive analysis range, is not material to the
accuracy of higher level measurements. Nor is it practical to zero a portable analyzer every time it is moved from one
sample point to another.
3. Caution: Prematurely zeroing the analyzer can cause a negative reading in both the ZERO and SAMPLE modes.
4. Satisfying users that the zero offset is reasonably acceptable for their application can be accomplished much quicker.
Unless the zero gas is contaminated or there is a significant leak in the sample connections, the analyzer should read less
than 1000 ppm oxygen within 2-3 minutes after being placed on zero gas thereby indicating it is operating normally.
Procedure:
Zero calibration should precede the span calibration and once performed should not have to be repeated with subsequent span
calibrations. Normally, zero calibrations are performed when a new sensor is installed or changes are made in the sample
system connections.
Refer to Span Calibration below for the detailed procedure. Differences include the displays illustrated below, substituting a
suitable zero gas for the span gas and the time required to determine the true zero offset of specific oxygen sensor, analyzer
and sample system combination.
The maximum zero calibration adjustment permitted is 50% of the lowest full scale range available. Accordingly, the analyzer’s
ZERO has not been adjusted prior to shipment because the factory conditions are different from the application condition at the
user’s installation.
5. Advance the reverse shade cursor using the ARROW keys to highlight ZERO CALIBRATE.
6. Press the ENTER key to select the highlighted menu option.
7. The following displays appear:
22
Advanced Instruments Inc.
0.000 PPM
ZERO
CALIBRTION
ENTER TO CALIBRATE
MENU TO ABORT
8. Press the ENTER key to calibrate or MENU key to abort and return to SAMPLING mode.
9. Allow approximately 60 seconds for the calibration process while the processor determines whether the signal output or
reading has stabilized within 50% of the full scale low range.
10. Both the Zero Calibrate and Span Calibrate functions result in the following displays:
PASSED FAILED
CALIBRATION OR CALIBRATION
Default Zero:
The software will eliminate any previous zero calibration adjustment and display the actual the signal output of the sensor at
any specific oxygen concentration. For example, assuming a zero gas is introduced, the display will reflect an oxygen reading
representing basically the zero calibration adjustment as described above. This feature allows the user to test the sensor’s signal
output without removing it from the sensor housing.
Procedure:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appear:
5. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT ZERO.
6. Press the ENTER key to select the highlighted menu option.
7. The following display appears and after 3 seconds the system returns to the SAMPLING mode:
FACTORY
DEFAULTS
SET
3.3%AUTO SAMPLING
10 % RANGE
24.5 C
LO1 2% 10% HI2
23
Advanced Instruments Inc.
Output Zero:
In rare instances the 4-20mA signal output may not agree to the reading displayed by the LCD. This feature enables the user to
adjust the 4mA signal output when the LCD displays 00.00. Compute the adjustment value as described in Appendix B or
consult the factory.
Note: Adjust the 20mA signal output with the OUTPUT SPAN option described below. The true adjustment value must be
determined empirically by trial and error. Adjust the initial adjustment value for additional percent errors.
Procedure:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appear:
5. Advance the reverse shade cursor using the ARROW keys to highlight OUTPUT ZERO.
6. Press the ENTER key to select the highlighted menu option.
7. The following display appears
100.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
8. Compute the adjustment value as described in Appendix B. Note: The true adjustment value must be determined
empirically by trial and error. Adjust the initial adjustment value for additional percent errors.
090.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
9. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the OUTPUT ZERO OFFSET value.
10. Press the ARROW keys to enter the OUTPUT ZERO OFFSET value.
11. Repeat steps 9 and 10 until the complete OUTPUT ZERO OFFSET value has been entered.
12. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
13. The system returns to the SAMPLING mode.
24
Advanced Instruments Inc.
Span Calibration
Maximum drift from calibration temperature is approximately 0.11% of reading per °C. The analyzer has been calibrated at the
factory. However, in order to obtain reliable data, the analyzer must be calibrated at the initial start-up and periodically
thereafter. The maximum calibration interval recommended is approximately 3 months, or as determined by the user’s
application.
Calibration involves adjusting the analyzer electronics to the sensor’s signal output at a given oxygen standard, e.g. a certified
span gas with an oxygen content (balance nitrogen) approximating 80% of the next higher full scale range above the intended
measuring range is recommended for optimum accuracy, see Calibration and Accuracy.
Recommendation – based on the inherent linearity of the galvanic oxygen sensor air calibrate the analyzer as described
below when installing and replacing the oxygen sensor (exception UHP and MS versions of the Pico Ion Sensor); or, to verify the
oxygen content of a certified span gas; or, when a certified span gas is not available (immediately following air calibration
reintroduce a gas with a low oxygen concentration to expedite the return to ppm level measurements).
Caution: Prematurely initiating the SPAN CALIBRATION key before the galvanic fuel cell sensor based analyzer reading has
stabilized can result in erroneous readings.
For example, to assure an accurate air calibration when installing a new ppm oxygen sensor from its packaged oxygen free
atmosphere allow the oxygen sensor 2-3 minutes to reach equilibrium with the oxygen content of the ambient air surrounding it
before pressing the SPAN CALIBRATE key. A ppm oxygen sensor that has not been allowed to reach equilibrium will generate a
lower current output than a ppm oxygen sensor that has reached equilibrium. Pressing the SPAN CALIBRATE key before the
ppm oxygen sensor has reached equilibrium forces the micro-processor to prematurely read the (erroneous low) current output
of the ppm oxygen sensor and introduce larger (erroneous) than required electronic gain adjustment and display (also
erroneous) CALIBRATION SUCCESSFUL message to the user. The error will become evident when a zero gas with a low oxygen
concentration is introduced into the ppm analyzer to purge it down below 10 ppm. The analyzer reading may stop and appear
to stabilize as high as 1800 ppm – giving the user the (erroneous) impression there is a problem with the ppm oxygen sensor
when in fact the problem lies with the user’s failure to follow the recommended calibration procedure.
Procedure:
The user must ascertain that the oxygen reading (actually the sensor’s signal output) has reached a stable value within the
limits entered below before entering the span adjustment. Failure to do so will result in an error.
This procedure assumes a span gas under positive pressure and is recommended for a analyzer without an optional sampling
pump, which if installed downstream of the sensor should be placed in the OFF position and disconnected so the vent is not
restricted during calibration. To assure an accurate calibration, the temperature and pressure of the span gas must closely
approximate the sample conditions.
For calibration purposes, use of the AUTO SAMPLE mode is recommended. However, the user can select the full scale MANUAL
SAMPLE RANGE for calibration as dictated by the accuracy of the analysis required – for example, a span gas with an 80 ppm
oxygen concentration with the balance nitrogen would dictate the use of the 0-100 ppm full scale range for calibration and a 0-
10 ppm measuring range. Select as described above.
25
Advanced Instruments Inc.
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
3.3%
AUTO SAMPLING
CALIBRATION 10 % RANGE
CONFIG ALARMS
BYPASS ALARMS 24.5 C
LO1 2% 10% HI2
GAS CONCENTRATION
PERCENT
PPM
20. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the alarm value.
21. Press the ARROW keys to enter the alarm value.
22. Repeat steps 20 and 21 until the complete SPAN value has been entered.
26
Advanced Instruments Inc.
23. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
24. Allow approximately 60 seconds for the calibration process while the processor determines whether the signal output or
reading has stabilized within 60% of the full scale low range. Both the Zero Calibrate and Span Calibrate functions result in
the following displays:
PASSED FAILED
CALIBRATION OR CALIBRATION
25. If the calibration is successful, the analyzer returns to the SAMPLING mode after 30 seconds.
26. If the calibration is unsuccessful, return to the SAMPLING mode with span gas flowing through the analyzer, make sure the
reading stabilizes and repeat the calibration before concluding the equipment is defective.
27. Before disconnecting the span gas line and connecting the sample gas line, restart if necessary the flow of sample gas and
allow it to flow for 1-2 minutes to purge the air inside the line.
28. Disconnect the span gas line and replace it with the purged sample gas line.
29. Wait 10-15 minutes to ensure the reading is stable and proceed to sampling.
Default Span
The software will set the SPAN adjustment based on the average oxygen reading (actually the sensor’s signal output) at any
specific oxygen concentration. For example, when a span gas is introduced, the micro-processor will display an oxygen reading
within +50% of the span gas value. This feature allows the user to test the sensor’s signal output without removing it from the
sensor housing.
Procedure:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appears:
5. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT SPAN.
6. Press the ENTER key to select the highlighted menu option.
7. The following displays appear and after 3 seconds the system returns to the SAMPLING mode:
FACTORY
DEFAULTS
SET
3.1%AUTO SAMPLING
10 % RANGE
24.5 C
LO1 2% 10% HI2
27
Advanced Instruments Inc.
Output Span:
In rare instances the 4-20mA signal output may not agree to the reading displayed by the LCD. This feature enables the user to
adjust the 20mA signal output should the LCD display not agree.
Note: Adjust the 4mA signal output with the OUTPUT ZERO option described above.
Procedure:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appear:
5. Advance the reverse shade cursor using the ARROW keys to highlight OUTPUT SPAN.
6. Press the ENTER key to select the highlighted menu option.
7. The following display appears
100.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
8. Compute the adjustment value as described in Appendix B or consult the factory. The true adjustment value must be
determined empirically by trial and error. Adjust the initial adjustment value for additional percent errors.
099.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
9. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the OUTPUT SPAN OFFSET value.
10. Press the ARROW keys to enter the OUTPUT SPAN OFFSET value.
11. Repeat steps 9 and 10 until the complete OUTPUT SPAN OFFSET value has been entered.
12. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
13. The system returns to the SAMPLING mode.
28
Advanced Instruments Inc.
Sampling
GPR-2900 Oxygen Analyzer requires positive pressure to flow the sample gas by the sensor to measure the oxygen
concentration in a sample gas. If not available see Pressure & Flow section.
Note: Prematurely initiating the ZERO CALIBRATION procedure can cause the analyzer to display a negative reading in both
the ZERO and SAMPLE modes. Prematurely initiating the SPAN CALIBRATION procedure can cause erroneously high offsets and
inaccurate readings.
Procedure:
Following calibration the analyzer returns to the SAMPLE mode after 30 seconds.
1. Select the desired sampling mode - auto or if manual, the range that provides maximum resolution – as described above.
2. Use metal tubing to transport the sample gas to the analyzer.
3. The main consideration is to eliminate air leaks which can affect oxygen measurements above or below the 20.9% oxygen
concentration in ambient air - ensure the sample gas tubing connections fit tightly into the 1/8” male NPT to tube adapter,
and, the NPT end is taped and securely tightened into the mating male quick disconnect fittings which mate with the
female fittings on the analyzer
4. Assure there are no restrictions in the sample line.
5. For sample gases under positive pressure the user must provide a means of controlling the inlet pressure between 5-30
psig and the flow of the sample gas between 1-5 SCFH, a flow rate of 2 SCHF is recommended
6. For sample gases under atmospheric or slightly negative pressure an optional sampling pump is recommended to draw the
sample into the analyzer. Generally, no pressure regulation or flow control device is involved.
7. Caution: If the analyzer is equipped with an optional sampling pump and is intended for use in both positive and
atmospheric/slightly negative pressure applications where a flow meter valve is involved – ensure the valve is completely
open when operating the sampling pump. Refer to the Pressure & Flow section above.
8. Assure the sample is adequately vented for optimum response and recovery – and safety.
9. Allow the oxygen reading to stabilize for approximately 10 minutes at each sample point.
Standby
¾ The analyzer has no special storage requirements.
¾ The sensor should remain connected during storage periods.
¾ Store the analyzer with the power OFF.
¾ If storing for an extended period of time protect the analyzer, cable and sensor from dust, heat and moisture.
29
Advanced Instruments Inc.
6 Maintenance
There are no moving parts in the analyzer given the modular nature of the electronics and sensor. Cleaning the electrical
contacts when replacing the sensor is the extent of the maintenance requirements of this analyzer.
Sensor Replacement: Periodically, the oxygen sensor will require replacement. The operating life is determined by a number of
factors that are influenced by the user and therefore difficult to predict. The sections dealing with Specification and Installation
Considerations define the normal operating conditions and expected life of the standard sensor utilized by the GPR-2900
analyzer. Expected sensor life is inversely proportional to changes in oxygen concentration, pressure and temperature.
Serviceability: Except for replacing the oxygen sensor, there are no parts inside the analyzer for the operator to service. Only
trained personnel with the authorization of their supervisor should conduct maintenance.
Caution: DO NOT open the oxygen sensor. The sensor contains a corrosive liquid electrolyte that could be harmful if touched
or ingested, refer to the Material Safety Data Sheet contained in the Owner’s Manual appendix. Avoid contact with any liquid or
crystal type powder in or around the sensor or sensor housing, as either could be a form of electrolyte. Leaking sensors should
be disposed of in accordance with local regulations.
Procedure:
1. Unscrew the knurled lock ring connecting the cable to the sensor.
2. Disconnect and remove the female plug (including the knurled lock ring) molded to the cable from the male receptacle
attached to the old oxygen sensor.
3. Unscrew the old oxygen sensor from the threaded hole in the sensor flow housing.
4. Open the barrier bag containing the new sensor.
5. The new oxygen sensor is equipped with a wire spring type shorting device, remove the shorting wire from the pins of the
male receptacle attached to the new oxygen sensor.
6. Line up the registration point of the female plug molded to the cable and male receptacle on the oxygen sensor.
7. Push the female plug (including the knurled lock ring) into the male receptacle attached to the oxygen sensor.
8. Screw and finger tighten the knurled lock ring of the cable plug into to the male connector attached to the sensor.
9. Place the analyzer in the auto ranging mode, as described above.
10. With the sensor exposed to ambient air wait for 5-10 minutes until the reading stabilizes.
11. Perform and air calibration as described above.
12. Screw the new sensor, finger tight plus ¼ turn, into the threaded hole in the flow housing and ensure the o-ring seal is
engaged but not ‘crushed’.
13. Begin sampling.
7 Spare Parts
Recommended spare parts for the GPR-2900 Oxygen Analyzer include:
30
Advanced Instruments Inc.
8 Troubleshooting
Symptom Possible Cause Recommended Action
Reading does not reflect Sensor was not calibrated at the pressure, Recalibrate the analyzer using the known oxygen
expected values flow rate and temperature anticipated in content of ambient air, 20.9%, and if the sensor
the sample gas stream responds we should presume it is correct.
Oxygen reading drifts Indication sensor is nearing the end of its Replace sensor
toward zero or significant useful life
number of turns of the
span control adjustment is
required to calibrate the
analyzer.
Slow response time Liquid covering sensing membrane Gently remove with alcohol and lint free towel.
Erratic oxygen reading Presence of interference gases. Consult factory, replace sensor
High oxygen reading Inadequate control of pressure and See Installation Considerations, section 5
flowrate
31
Advanced Instruments Inc.
9 Warranty
The design and manufacture of GPR Series oxygen analyzers, monitors and oxygen sensors are performed under a certified
Quality Assurance System that conforms to established standards and incorporates state of the art materials and components
for superior performance and minimal cost of ownership. Prior to shipment every analyzer is thoroughly tested by the
manufacturer and documented in the form of a Quality Control Certification that is included in the Owner’s Manual
accompanying every analyzer. When operated and maintained in accordance with the Owner’s Manual, the units will provide
many years of reliable service.
Coverage
Under normal operating conditions, the monitor, analyzers and sensor are warranted to be free of defects in materials and
workmanship for the period specified in accordance with the most recent published specifications, said period begins with the
date of shipment by the manufacturer. The manufacturer information and serial number of this analyzer are located on the rear
of the analyzer. Advanced Instruments Inc. reserves the right in its sole discretion to invalidate this warranty if the serial
number does not appear on the analyzer.
If your Advanced Instruments Inc. monitor, analyzer and/or oxygen sensor is determined to be defective with respect to
material and/or workmanship, we will repair it or, at our option, replace it at no charge to you. If we choose to repair your
purchase, we may use new or reconditioned replacement parts. If we choose to replace your Advanced Instruments Inc.
analyzer, we may replace it with a new or reconditioned one of the same or upgraded design. This warranty applies to all
monitors, analyzers and sensors purchased worldwide. It is the only one we will give and it sets forth all our responsibilities.
There are no other express warranties. This warranty is limited to the first customer who submits a claim for a given serial
number and/or the above warranty period. Under no circumstances will the warranty extend to more than one customer or
beyond the warranty period.
Limitations
Advanced Instruments Inc. will not pay for: loss of time; inconvenience; loss of use of your Advanced Instruments Inc. analyzer
or property damage caused by your Advanced Instruments Inc. analyzer or its failure to work; any special, incidental or
consequential damages; or any damage resulting from alterations, misuse or abuse; lack of proper maintenance; unauthorized
repair or modification of the analyzer; affixing of any attachment not provided with the analyzer or other failure to follow the
Owner’s Manual. Some states and provinces do not allow limitations on how an implied warranty lasts or the exclusion of
incidental or consequential damages, these exclusions may not apply.
Exclusions
This warranty does not cover installation; defects resulting from accidents; damage while in transit to our service location;
damage resulting from alterations, misuse or abuse; lack of proper maintenance; unauthorized repair or modification of the
analyzer; affixing of any label or attachment not provided with the analyzer; fire, flood, or acts of God; or other failure to follow
the Owner’s Manual.
Service
Call Advanced Instruments Inc. at 909-392-6900 (or e-mail [email protected]) between 7:30 AM and 5:00 PM Pacific Time Monday
thru Thursday or before 12:00 pm on Friday. Trained technicians will assist you in diagnosing the problem and arrange to
supply you with the required parts. You may obtain warranty service by returning you analyzer, postage prepaid to:
Be sure to pack the analyzer securely. Include your name, address, telephone number, and a description of the operating
problem. After repairing or, at our option, replacing your Advanced Instruments Inc. analyzer, we will ship it to you at no cost
for parts and labor.
32
Advanced Instruments Inc.
General Requirements
Use Potassium Hydroxide or Acetic Acid - electrolyte, Lead - anode
Handling Rubber or latex gloves, safety glasses
Storage Indefinitely
Physical Properties
Boiling Point Range KOH = 100 to 115° C or Acetic Acid = 100 to 117° C
Melting Point Range KOH -10 to 0° C or Acetic Acid – NA, Lead 327° C
Freezing Point KOH = -40 to -10° C or Acetic Acid = -40 to -10° C
Molecular Weight KOH = 56 or Acetic Acid – NA, Lead = 207
Specific Gravity KOH = 1.09 @ 20° C, Acetic Acid = 1.05 @ 20° C
Vapor Pressure KOH = NA or Acetic Acid = 11.4 @ 20° C
Vapor Density KOH – NA or Acetic Acid = 2.07
pH KOH > 14 or Acetic Acid = 2-3
Solubility in H2O Complete
% Volatiles by Volume None
Evaporation Rate Similar to water
Appearance and Odor Aqueous solutions: KOH = Colorless, odorless or Acetic Acid = Colorless, vinegar-like odor
Reactivity Data
Stability Stable
Conditions Contributing to Instability None
Incompatibility KOH = Avoid contact with strong acids or Acetic Acid = Avoid contact with strong bases
Hazardous Decomposition Products KOH = None or Acetic Acid = Emits toxic fumes when heated
Conditions to Avoid KOH = None or Acetic Acid = Heat
33
Advanced Instruments Inc.
Spill or Leak
Steps if material is released Sensor is packaged in a sealed plastic bag, check the sensor inside for electrolyte leakage. If the
sensor leaks inside the plastic bag or inside an analyzer sensor housing do not remove it without
rubber or latex gloves and safety glasses and a source of water. Flush or wipe all surfaces
repeatedly with water or wet paper towel (fresh each time).
Disposal In accordance with federal, state and local regulations.
Special Precautions
Precautions Do not remove the sensor’s protective Teflon and PCB coverings. Do not probe the sensor with
sharp objects. Wash hands thoroughly after handling. Avoid contact with eyes, skin and clothing.
Empty sensor body may contain hazardous residue.
Transportation Not applicable
34
Advanced Instruments Inc.
Appendix B
Correlating Readings - LCD Display and 4-20mA Output
In rare instances the 4-20mA signal output may not agree to the reading displayed by the LCD. The Output Zero and Output
Span features enable the user to adjust the 4mA signal output to correlate with the reading displayed by the LCD.
In the field or in the absence of the preferred gases, use the OUTPUT SPAN feature and adjust the 20mA signal output using
the span gas available.
Example: Analyzer reading is 60 ppm oxygen (100 ppm range) on 84 ppm span gas, 4-20mA signal output at PLC is 24mA
Output Zero
7. Access the MAIN MENU by pressing the MENU key.
8. Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
9. Press the ENTER key to select the highlighted menu option.
10. The following displays appear:
11. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT ZERO.
12. Press the ENTER key to select the highlighted menu option.
13. The following display appears:
35
Advanced Instruments Inc.
100.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
14. Enter the calculated adjustment value. Note: Once the initial adjustment is made and checked at the PLC it may be
necessary to fine tune the initial adjustment by repeating. Any additional percent error must be added or subtracted from
the initial adjustment value
000.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
15. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the adjustment OUTPUT ZERO OFFSET value.
16. Press the ARROW keys to enter each the numerical value of each digit of the adjustment OUTPUT ZERO OFFSET value.
17. Repeat until the complete OUTPUT ZERO OFFSET value has been entered.
18. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
19. The system returns to the SAMPLING mode.
Output Span
20. Access the MAIN MENU by pressing the MENU key.
21. Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
22. Press the ENTER key to select the highlighted menu option.
23. The following displays appear:
24. Advance the reverse shade cursor using the ARROW keys to highlight OUTPUT SPAN.
25. Press the ENTER key to select the highlighted menu option.
26. The following display appears:
100.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
36
Advanced Instruments Inc.
27. Enter the calculated adjustment value, refer to example described above. Note: Once the initial adjustment is made and
checked at the PLC it may be necessary to fine tune the initial adjustment by repeating. Any additional percent error must
be added or subtracted from the initial adjustment value
064.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
.
28. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the adjustment OUTPUT SPAN OFFSET value.
29. Press the ARROW keys to enter the numerical value of each digit of the OUTPUT SPAN OFFSET value.
30. Repeat until the complete OUTPUT SPAN OFFSET value has been entered.
31. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
32. The system returns to the SAMPLING mode.
37