HTC InFire Speed
HTC InFire Speed
HTC InFire Speed
2018
Operating Instructions
English
Operating instructions
=
Contents Dentsply Sirona
Operating Instructions inFire HTC speed
Contents
1 Dear customer,......................................................................................................... 5
2 General data............................................................................................................. 6
2.1 Structure of the document............................................................................. 6
2.1.1 Identification of the danger levels..................................................... 6
2.1.2 Formats and symbols used .............................................................. 7
2.2 Warranty........................................................................................................ 7
4 Technical description................................................................................................ 11
4.1 Description of the furnace ............................................................................. 11
4.2 Certification ................................................................................................... 12
4.3 Intended use ................................................................................................. 12
4.4 Technical data............................................................................................... 13
5 Setup ........................................................................................................................ 14
5.1 Installation site .............................................................................................. 14
5.2 Electrical connection ..................................................................................... 15
6 Operation.................................................................................................................. 17
6.1 Controls and displays .................................................................................... 17
6.1.1 Key functions.................................................................................... 18
6.1.2 Description of the standby display ................................................... 18
6.2 Switching on the unit ..................................................................................... 19
6.3 Practical use.................................................................................................. 19
6.3.1 Initial startup..................................................................................... 20
6.3.2 Sintering zirconium oxide ................................................................. 20
6.3.2.1 Preparing the sintering tray ............................................... 20
6.3.2.2 Preparing the furnace door................................................ 21
6.3.3 Program selection ............................................................................ 26
6.3.4 Loading a program ........................................................................... 26
6.3.5 Starting the program ........................................................................ 26
6.3.6 Lifting the sintering tray out of the furnace ....................................... 26
6.4 Operating modes........................................................................................... 26
6.4.1 Super speed sintering (program starts at 1,580 °C / 2,876 °F, pre- 26
heating required)
6.4.1.1 Performing super speed sintering ..................................... 27
6.4.2 Speed sintering (program starts at room temperature, no pre- 29
heating required)
61 30 459 D 3497
2 D 3497.201.01.20.02 10.2018
Dentsply Sirona Contents
Operating Instructions inFire HTC speed
båÖäáëÜ
6.5.7 Replacing the furnace door (only for devices with serial numbers < 37
5000)>
6.6 Programs...................................................................................................... 38
6.6.1 Plan for assigning furnace programs............................................... 38
6.6.2 Fixed programs ............................................................................... 39
6.6.3 Select program number/load program............................................. 39
6.6.4 Starting the program........................................................................ 40
6.6.5 Starting a program with the Casting-Time function ......................... 40
6.6.6 Freely programmable programs ...................................................... 40
6.6.6.1 Entering program values .................................................. 40
6.6.6.2 Entering the program values for fast sintering (speed 42
sintering)
6.6.6.3 Programming example for fast sintering (speed sintering) 42
6.6.6.4 Storing program values .................................................... 43
7 Default settings........................................................................................................ 44
7.1 Display and acoustic signal .......................................................................... 44
7.2 Adjustment of the start time (casting time function) ..................................... 45
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 3
Contents Dentsply Sirona
Operating Instructions inFire HTC speed
10 Maintenance............................................................................................................. 48
10.1 Changing fuses F1/F2 ................................................................................... 48
10.2 Replacing the battery (only for devices with serial numbers < 5000)>.......... 49
10.3 Furnace chamber insulation .......................................................................... 49
11 Malfunctions ............................................................................................................. 50
11.1 Error messages from electronic components................................................ 50
11.2 Miscellaneous errors and their causes.......................................................... 51
12 Disposal.................................................................................................................... 52
Index......................................................................................................................... 53
61 30 459 D 3497
4 D 3497.201.01.20.02 10.2018
Dentsply Sirona 1 Dear customer,
Operating Instructions inFire HTC speed
1 Dear customer,
Thank you for purchasing your inFire HTC speed high-temperature
furnace from Sirona.
It will support you with your work for many years, as it was developed and
built with state-of-the-art technology.
Nevertheless, improper use and handling can cause damage and
hazards. Please read and follow these operating instructions carefully.
Always keep them within easy reach.
To prevent personal injury or material damage, it is important to observe
all safety information.
Your inFire HTC speed Team
båÖäáëÜ
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 5
2 General data Dentsply Sirona
2.1 Structure of the document Operating Instructions inFire HTC speed
2 General data
Please read this document completely and follow the instructions exactly.
You should always keep it within reach.
Original language of the present document: German.
DANGER
An imminent danger that could result in serious bodily injury or death.
WARNING
A possibly dangerous situation that could result in serious bodily injury
or death.
CAUTION
A possibly dangerous situation that could result in slight bodily injury.
NOTICE
A possibly harmful situation which could lead to damage of the product
or an object in its environment.
IMPORTANT
Application instructions and other important information.
61 30 459 D 3497
6 D 3497.201.01.20.02 10.2018
Dentsply Sirona 2 General data
Operating Instructions inFire HTC speed 2.2 Warranty
båÖäáëÜ
● List Designates a list.
“Command / menu item” Indicates commands, menu items or
quotations.
2.2 Warranty
The sinter support is excluded from the warranty as it is a consumable
item.
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 7
3 Safety information Dentsply Sirona
Operating Instructions inFire HTC speed
3 Safety information
Perform connection by following the directions given in the present
operating instructions.
As manufacturers of dental and laboratory equipment and in the interest
of the operational safety of your system, we stress the importance of
having maintenance and repairs performed exclusively by our own
personnel or through our authorized representatives. Furthermore,
safety-critical system components must always be replaced with original
spare parts.
We suggest that you request a certificate specifying the nature and extent
of the work performed from the persons who performed it. Furthermore,
the certificate should also include any possible changes in rated
parameters or working ranges as well as the date, the name of the firm
and a signature.
Modifications to this system which may affect the safety of the operator or
third parties are prohibited by law.
For reasons of product safety, this product may be operated only with
original Sirona accessories or third-party accessories expressly approved
by Sirona. The user is responsible for any damage resulting from the use
of non-approved accessories.
If any equipment not approved by Sirona is connected, it must comply
with the applicable standards:
● EN 60 950 for information technology equipment, and
● EN 61 010-1 for laboratory equipment.
CAUTION
Liability exclusion for any other use
The inFire HTC speed high-temperature furnace is exclusively intended
for sintering oxide ceramics for use in dental technology. We are not
liable for damages due to any other use.
The inFire HTC speed high-temperature furnace with integrated gas
management system and associated accessories enable the sintering
of pre-sintered non-precious metals.
CAUTION
Risk due to improper operation
The inFire HTC speed high-temperature furnace may only be operated
by employees who are familiar with the content of these operating
instructions. Signs and labels on the laboratory furnace must be kept in
legible condition at all times. They may not be removed.
61 30 459 D 3497
8 D 3497.201.01.20.02 10.2018
Dentsply Sirona 3 Safety information
Operating Instructions inFire HTC speed
CAUTION
Risk due to incorrect setup
The inFire HTC speed high-temperature furnace may only be set up in
dry rooms and must never be in contact with any liquids. Furniture items
and other equipment in the vicinity of the furnace must not contain any
explosive, flammable or easily ignitable materials. It is prohibited to keep
or store ignitable or flammable gases or liquids in the room where the
furnace is set up.
CAUTION
Damage resulting from unauthorized alterations
The inFire HTC speed high-temperature furnace may only be altered
with our express prior approval. We are not liable for unauthorized
alterations of any kind.
➢ Switch the unit off and disconnect the power plug prior to performing
båÖäáëÜ
maintenance tasks.
WARNING
Separate circuit
The inFire HTC speed high-temperature furnace must always be
connected to a separate circuit with at least a 16A fuse. The fuses must
be of the C type.
WARNING
Risk of burns due to objects falling out
As there is a serious risk of burns in this case, it is essential to install a
fire-proof table and floor covering.
I. For safety reasons, the furnace must be operated when standing. Only
in this way is it possible for the user to quickly avoid parts that fall out.
II. When using super speed sintering, wearing heat-resistant protective
clothing is recommended. In accordance with professional association
regulations, the operator of the site is responsible for providing the user
with appropriate protective clothing.
III. The operator of the site is responsible for giving the user appropriate
safety instructions.
IV. If a fracture occurs during the sintering process as a result of a failure
to adhere to the recommended drying procedure, there is a risk that hot
objects will fall out as soon as the furnace door is opened.
V. The furnace should only be loaded and unloaded using the crucible
fork. For safety reasons, a trained user must be present when the
furnace is being loaded or unloaded.
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 9
3 Safety information Dentsply Sirona
Operating Instructions inFire HTC speed
WARNING
Risk of burns due to hot surfaces
The inFire HTC speed high-temperature furnace may develop hot
surfaces in some areas after extended use. Use the necessary caution
around the unit, especially when the furnace is operated for extended
periods.
➢ Never touch the hot surfaces.
WARNING
Risk of burns due to hot surfaces
Once the program has ended and the furnace has shut down, the
cooling fan will run until the interior temperature of the furnace has been
reduced to normal. The power plug may not be disconnected prior to this
point.
WARNING
Fire hazard due to covering ventilation slots on the upper side of the
casing
Covered ventilation slots can overheat the furnace and cause the object
placed on top to ignite.
➢ Do not place any objects on the casing upper side.
Note on the prevention, recognition and This device is a piece of electrical equipment with a supply voltage below
elimination of unintended 1000 VAC and is intended for commercial use. It is to be installed in
electromagnetic effects dental laboratories or other areas with a controlled electromagnetic
environment. The applicable EMC requirements comply with EN
61326:2006.
61 30 459 D 3497
10 D 3497.201.01.20.02 10.2018
Dentsply Sirona 4 Technical description
Operating Instructions inFire HTC speed 4.1 Description of the furnace
4 Technical description
båÖäáëÜ
Please refer to the rating plate at the rear of the device for information
about the model.
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 11
4 Technical description Dentsply Sirona
4.2 Certification Operating Instructions inFire HTC speed
4.2 Certification
CE mark
CAUTION
CE mark for connected products
Further products which are connected to this unit must also bear the CE
mark. These products must be tested according to the applicable
standards.
We declare conformity for the inFire HTC speed ceramic sintering furnace
on the basis of the following standards:
● Safety: IEC 61010-1:2001 2nd edition and IEC 61010-2-010:2003
2nd edition
● EMC: EN 61326:2006
● Risk assessment and risk reduction EN ISO 12100:2010
EAC certification
RoHS compliance
This symbol indicates that this product does not contain any toxic or
hazardous substances or components above the maximum
concentration value set out in the Chinese standard SJ / T 11364-2014,
and can be recycled following disposal and should not be carelessly
discarded.
61 30 459 D 3497
12 D 3497.201.01.20.02 10.2018
Dentsply Sirona 4 Technical description
Operating Instructions inFire HTC speed 4.4 Technical data
båÖäáëÜ
Dimensions of unit (W x H x D in mm): 500 x 820 x 565
Dimensions of packaged unit (W x H x D in mm): 630 x 1300 x 730
Furnace chamber: Diameter 130 mm x 80 mm height
Maximum sintering temperature: 1,650 °C (3,002 °F)
Approx. weight without packaging: 80 kg
Approx. weight including packaging: 96 kg
Maximum inlet pressure: 8 bar
(applies to all devices with built-in gas management
module)
Rated line voltage 200 V AC – 240 V AC
(all units except REF: 6401462 and 6416247):
Rated line frequency 50/60Hz
(all units except REF: 6401462 and 6416247):
Rated current consumption 15 A
(all units except REF: 6401462 and 6416247):
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 13
5 Setup Dentsply Sirona
5.1 Installation site Operating Instructions inFire HTC speed
5 Setup
NOTICE
Danger of overheating
If the unit overheats, the electronics switch off.
➢ Never obstruct the ventilation slots!
➢ Provide ventilation clearance on all sides.
➢ A minimum clearance of 500 mm must be adhered to, above and to
the sides.
61 30 459 D 3497
14 D 3497.201.01.20.02 10.2018
Dentsply Sirona 5 Setup
Operating Instructions inFire HTC speed 5.2 Electrical connection
båÖäáëÜ
Building installation The following electrical installation requirements are to be fulfilled for the
high-temperature inFire HTC speed furnace:
● The furnace requires a separate electric circuit.
● The electrical building installation must be protected by a type C
circuit breaker with a rating of at least 16 A.
● If an additional circuit breaker is used, it must be designed for a
tripping current of at least 30 mA.
● The furnace requires the connection of a protective ground wire to the
electric outlet for safe electrical operation.
● The distance between the electric outlet and the furnace must be
selected so that the 2.5 m long power cord supplied with the furnace
is sufficiently long (extension cords are not allowed!).
● The supply voltage must lie within the nominal voltage range of 200
VAC to 240 VAC. A separate building installation must be installed to
meet this requirement in the USA and Japan. In the USA, the 240 V
outlet must be designed for connecting a NEMA 6-15-type plug; in
Japan, the 200 V outlet must be designed for connecting a NEMA L6
(L6-20J)-type plug.
Unit Do not adjust the line voltage!
The unit automatically adjusts to the line voltage.
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 15
5 Setup Dentsply Sirona
5.2 Electrical connection Operating Instructions inFire HTC speed
RS232 interface
A type RS232 serial interface (A) is located on the rear side of the unit.
The use of this interface is restricted to the service engineer and is
described in the service manual.
61 30 459 D 3497
16 D 3497.201.01.20.02 10.2018
Dentsply Sirona 6 Operation
Operating Instructions inFire HTC speed 6.1 Controls and displays
6 Operation
båÖäáëÜ
Membrane keyboard
A LCD display
B Stage keys
C Numeric keys
D Memory key
E Load program key
F Casting time key
G Main switch (ON/OFF)
H START/STOP key
I Lift key
J Function key
K Stage LEDs
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 17
6 Operation Dentsply Sirona
6.1 Controls and displays Operating Instructions inFire HTC speed
Casting time key For setting the casting time; the switch-on time
for the unit is calculated automatically
61 30 459 D 3497
18 D 3497.201.01.20.02 10.2018
Dentsply Sirona 6 Operation
Operating Instructions inFire HTC speed 6.2 Switching on the unit
Description of the columns on the left side with the freely programmed
furnace programs
Holding time The holding time of the corresponding step is set / displayed in this
column.
båÖäáëÜ
The laboratory furnace heats up and cools down in up to four temperature
steps. It is possible to heat or cool down within the program run.
NOTICE
Risk of breakage
The sintering trays to be used are fragile. They break with strong
temperature fluctuations and under mechanical stress.
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 19
6 Operation Dentsply Sirona
6.3 Practical use Operating Instructions inFire HTC speed
NOTICE
Risk of damage to the furnace chamber
If incorrectly positioned, the sintering tray collides with the furnace
chamber insulation.
➢ Ensure that the sintering tray is positioned centrally on the covering,
before closing the furnace or starting the process.
➢ Place the sintering tray on the crucible tray and fill the tray with
sintering pearls as required. Only use original Sirona sintering pearls
for zirconium oxide. The correct fill level of the sintering trays varies
depending on requirements. To prevent a delay, when positioning the
restorations to be sintered, ensure that there is sufficient support and
that the sintering pearls can move freely.
Take note of the following items when filling the sintering tray:
● Free movement of the sintering pearls
● Sufficient support for sagging restorations
● Positioning the restorations with the chewing surface facing
downwards (in contact with the sintering pearls)
● Interdental areas should be free of sintering pearls
61 30 459 D 3497
20 D 3497.201.01.20.02 10.2018
Dentsply Sirona 6 Operation
Operating Instructions inFire HTC speed 6.3 Practical use
båÖäáëÜ
Door insulation Door insulation with sintering pearls
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 21
6 Operation Dentsply Sirona
6.3 Practical use Operating Instructions inFire HTC speed
Replacement door for metal sintering option Design with sintering bells
61 30 459 D 3497
22 D 3497.201.01.20.02 10.2018
Dentsply Sirona 6 Operation
Operating Instructions inFire HTC speed 6.3 Practical use
båÖäáëÜ
● Speed sintering tray
● Stacking disc
● Crucible tray
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 23
6 Operation Dentsply Sirona
6.3 Practical use Operating Instructions inFire HTC speed
1. Fit the lower door insulation brick A using the small ceramic tube B
and place it in the door panel.
2. Position the upper door insulation brick on the small ceramic tube on
the lower door insulation brick. The upper door insulation brick guides
make it possible to load and unload using the crucible fork.
4. Position the split support disc on the upper door insulation brick.
NOTICE
Risk of damage to the furnace chamber
If incorrectly positioned, the sintering tray collides with the furnace
chamber insulation.
➢ Ensure that the sintering tray is positioned centrally on the covering,
before closing the furnace or starting the process.
61 30 459 D 3497
24 D 3497.201.01.20.02 10.2018
Dentsply Sirona 6 Operation
Operating Instructions inFire HTC speed 6.3 Practical use
If you wish to stack, you require a stacking disc that is not split.
✔ The first sintering tray is constructed as described above.
NOTICE
Risk of damage to the furnace chamber
If incorrectly positioned, the sintering tray collides with the furnace
chamber insulation.
➢ Ensure that the stacking disc and sintering tray are positioned
centrally, before closing the furnace or starting the process.
båÖäáëÜ
2. Position the second sintering tray centrally on the stacking disc and
start the sintering process.
The furnace should only be loaded and unloaded using the supplied
crucible fork. Removal is via the guide grooves in the upper door
insulation brick.
1. After sintering, slide the crucible fork to the end of the guide slots.
2. Lift the sintering tray with the split support disk and place it on a
fireproof tray to cool down, e.g. on the crucible tray.
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 25
6 Operation Dentsply Sirona
6.4 Operating modes Operating Instructions inFire HTC speed
The fast heating process involved in super speed sintering means that the
restorations must be completely dry. Any residual moisture is very likely
to cause damage (cracks or breakages) to a restoration. In general, two
drying procedures are necessary when using inCoris TZI: one
immediately after milling and one after coloring. To safely dry crowns
made from colored inCoris TZI, the following drying procedure is
recommended by way of example:
Example of the drying procedure:
10 minutes at 150 °C (±5 °C); it is imperative the drying temperature is
maintained.
Recommendation for the unit: drying closets or other heating devices with
precise temperature controls and a maximum temperature deviation of ±
5 °C.
For more detailed information, take note of the manufacturer's
specifications for the material.
61 30 459 D 3497
26 D 3497.201.01.20.02 10.2018
Dentsply Sirona 6 Operation
Operating Instructions inFire HTC speed 6.4 Operating modes
båÖäáëÜ
The furnace closes automatically and initiates the pre-heating
phase.
WARNING
Risk of burns if used incorrectly
The furnace opens at 1,580 °C (2,876 °F). There is only a risk of burns
if the furnace is used incorrectly.
The internal company safety regulations for working on laboratory
furnaces apply.
➢ Always stand when loading the furnace.
➢ Always wear long trousers and closed shoes.
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 27
6 Operation Dentsply Sirona
6.4 Operating modes Operating Instructions inFire HTC speed
4. With the help of the crucible fork, remove the crucible with the
restoration and covering, so as to load the furnace.
61 30 459 D 3497
28 D 3497.201.01.20.02 10.2018
Dentsply Sirona 6 Operation
Operating Instructions inFire HTC speed 6.4 Operating modes
Sintering
After the holding time, the furnace opens at 1,580 °C (2,876 °F) and then
cools down to 1,000 °C (1,832 °F) by itself. 1 minute before the door
opens, the furnace displays a "countdown" of 60 seconds.
At 1,000 °C (1,832 °F), the display starts to flash. This is followed by an
acoustic signal which repeats every 15 seconds. The signals are to
indicate that the sinter may be removed.
1. Lift the crucible from the furnace door using the crucible fork and
place it on the fire-proof resting cube supplied with the device.
2. Press the start/stop button to close the furnace door. This helps retain
residual heat to shorten the heating-up phase if the program is to be
repeated.
båÖäáëÜ
If you do not press the start/stop button after removing the
crucible from the furnace door, the furnace door will close
automatically after 5 minutes.
3. Wait 2 minutes.
5. After 2 minutes retention time, place the object being sintered on the
fireproof tray using a pair of tweezers.
6. Wait a further 2 minutes. Afterwards, the sinter is cooled down to the
extent that you can touch it without risk of burns.
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 29
6 Operation Dentsply Sirona
6.5 Non-precious metal (NPM) sintering (optional) Operating Instructions inFire HTC speed
WARNING
61 30 459 D 3497
30 D 3497.201.01.20.02 10.2018
Dentsply Sirona 6 Operation
Operating Instructions inFire HTC speed 6.5 Non-precious metal (NPM) sintering (optional)
6.5.1.2 Use of argon in the inFire HTC speed with gas management
module
With a factory-installed NPM function, the inFire HTC speed has an
internal gas management module which requires an inlet pressure of 7-8
bar in order to provide the required volume flow of 1.1 l / min for NPM
sintering.
The technical components are designed for a maximum inlet pressure of
8 bar.
The supply of argon is controlled via an electrically switchable valve.
When the program is started the furnace control activates the supply of
argon and switches it off again when the program is terminated. The
switching valve closes automatically in the event of a power outage.
DANGER
båÖäáëÜ
Observe the following when connecting argon at the furnace gas
management module:
– The valve on the gas cylinder should be limited to a maximum outlet
pressure of 8 bar in order to prevent excessive stress being placed on
the installed components in the furnace.
- The outlet pressure of the cylinders must be set to a minimum of 7 bar
and a maximum of 8 bar.
– Following installation, a check is required to ensure that the gas lines
and connected sockets are secure and that there are no leaks so that
no argon escapes uncontrollably.
– In the event of malfunction the volume of one argon cylinder is enough
to flood an entire room. Therefore, adequate ventilation and air
extraction is to be provided at the place of installation, otherwise the
user is at risk of suffocation.
As the gas is heavier than air, appropriate ventilation in the floor area is
recommended. Shafts and ducts must also be protected from gas
infiltration.
– Do not store gas cylinders in the working area.
– Always check valves for tightness on full and empty cylinders when
changing a cylinder.
– Secure cylinders from falling over.
– Maintenance: a regular tightness test is required.
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 31
6 Operation Dentsply Sirona
6.5 Non-precious metal (NPM) sintering (optional) Operating Instructions inFire HTC speed
DANGER
Danger of suffocation
Observe the national safety regulations when using argon in
laboratories (for Germany: TRGS526, in particular 5.2.11 compressed
gas cylinders and valves)
Applies for devices with serial numbers < 5000
When changing the furnace door from inCoris CC to ceramic, the argon
supply must be compulsorily separated from the furnace's furnace door
with gas socket, as otherwise uncontrolled argon gas will escape
through the furnace door with gas socket during a ceramic sintering
process.
61 30 459 D 3497
32 D 3497.201.01.20.02 10.2018
Dentsply Sirona 6 Operation
Operating Instructions inFire HTC speed 6.5 Non-precious metal (NPM) sintering (optional)
The filling level of the compressed gas cylinder can be ascertained from
the manometer. A new argon compressed gas cylinder is filled with 200
bar. If the pressure displayed is only half of the pressure, i.e. 100 bar, then
the cylinder is half empty.
Example:
a 50-liter compressed gas cylinder of argon with a filling pressure of 200
bar contains approx. 10,000 liters of argon.
The consumption per inCoris CC sintering cycle is approx. 270 liters at a
flow rate of 1.1 l / min.
As a result approx. 35 inCoris CC sintering cycles are possible with a 50-
liter compressed gas cylinder of argon.
båÖäáëÜ
(not supplied depending on the standard equipment)
The set for the sintering of inCoris CC/CCB or “Crypton for inLab” consists
of the following components:
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 33
6 Operation Dentsply Sirona
6.5 Non-precious metal (NPM) sintering (optional) Operating Instructions inFire HTC speed
1. Fill the sintering tray with zirconium oxide beads. The filling level
should be between 1.5 and 2 cm. In this process, only use the
zirconium oxide beads for NPM sintering (REF 64 13 640).
2. Place the parts to be sintered in the sintering tray and position the
restorations so that the crown margins protrude from the zirconium
oxide beads by around 1 - 1.5 mm.
IMPORTANT
The zirconium oxide beads are white when new and change to a dark
color during the NPM sintering cycle.
The base plate, sintering tray, cover, sintering bell, and zirconium oxide
beads must undergo an NPM sintering cycle without sinter before the
first NPM sintering cycle with sinter.
3. Place the sintering tray onto the center of the base plate.
4. Cover the sintering tray with the lid and sintering bell.
5. Load the heating program and start it with the start/stop button.
When the sintering programs are changed from ceramic to metal
sintering, a confirmation query appears on the display. Make sure
that the right furnace door featuring a gas socket for metal sintering
is installed.
After checking to make sure that the right furnace door is installed,
confirm the confirmation query and then start the program.
The lifting instrument moves upwards and seals the heating
chamber.
The heating program runs fully automated.
WARNING
Risk of burning
The sintering tray is still hot after the sintering process. Its temperature
is approx. 300 °C/572°F.
✔ The temperature display shows the temperature of the furnace
chamber and not that of the sintering bell.
➢ Wait until the furnace has cooled down completely.
➢ As an alternative, we recommend using commercial crucible and
investment tongs.
61 30 459 D 3497
34 D 3497.201.01.20.02 10.2018
Dentsply Sirona 6 Operation
Operating Instructions inFire HTC speed 6.5 Non-precious metal (NPM) sintering (optional)
6. Remove the sintering tray after the furnace has cooled down.
or
➢ Remove the sintering tray using commercial crucible and
investment tongs.
Tip: From serial number 5000, the sintering bell can be removed using the
supplied crucible fork.
6.5.6 Loading the furnace with the inCoris CC set, from serial
number >5000
1. Place the base plate on the upper door insulation brick.
båÖäáëÜ
2. Position the sintering tray with the cover centrally on the base plate.
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 35
6 Operation Dentsply Sirona
6.5 Non-precious metal (NPM) sintering (optional) Operating Instructions inFire HTC speed
61 30 459 D 3497
36 D 3497.201.01.20.02 10.2018
Dentsply Sirona 6 Operation
Operating Instructions inFire HTC speed 6.5 Non-precious metal (NPM) sintering (optional)
6.5.7 Replacing the furnace door (only for devices with serial
numbers < 5000)
WARNING
Risk of burning
The furnace door may be hot.
➢ Check the temperature of the furnace on the display.
➢ Change furnace doors only in the cold state.
The furnace door must be changed when sintering oxide ceramics. The
argon supply pipe for the furnace door with gas socket must also be
removed.
Please proceed as follows to change the furnace door:
1. To disconnect the pipe socket for the argon supply, push the retainer
ring for the argon socket towards the furnace enclosure.
båÖäáëÜ
2. Pull the argon supply pipe of the furnace door with gas socket out of
the socket.
DANGER
Danger of suffocation
If the pipe remains in the socket when sintering oxide ceramic materials,
then argon escapes from the base plate of the furnace door with gas
socket during the sintering process.
3. On the lower side of the furnace door there is a guide pin for locking
the door with the furnace fastening clamp. To unlock the guide pin,
pull this downward and then turn the pin in a counter-clockwise
direction.
The pin is now unlocked and stopped.
4. Pull the furnace doors forward and out of the fastening clamp.
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 37
6 Operation Dentsply Sirona
6.6 Programs Operating Instructions inFire HTC speed
5. Ensure that the door guide pin is locked in place in the locating hole
(A) provided for this purpose when inserting the furnace door for
oxide ceramics.
6. Push the furnace door for oxide ceramics until the end stop in the
furnace fastening clamp. Unlock the guide pin once more by pulling it
and turning it counter-clockwise. Now pull the door forward carefully
until the guide pin locks in place in the corresponding locating hole in
the furnace fastening clamp.
NOTICE
Sinter program does not start
If the guide pin does not lock in place correctly in the slot of the furnace
fastening clamp, then the door will not close properly once a sintering
program is started. The sintering program will not start.
6.6 Programs
Explanation With the inFire HTC speed sintering furnace, sintering is controlled via
callable programs [ → 39].
● inFire HTC speed has fixed programs for sintering certain dental
oxide ceramics.
● There are freely programmable programs for other dental oxide
ceramics.
● The inFire HTC speed with integrated or retrofitted gas management
system contains additional programs for the sintering of non-precious
metals. This sintering process is conducted in a protective gas
atmosphere with argon-flooding of the sintering bells that have been
developed especially for this purpose (see section "Non-precious
metal (NPM) sintering (optional) [ → 30]“).
61 30 459 D 3497
38 D 3497.201.01.20.02 10.2018
Dentsply Sirona 6 Operation
Operating Instructions inFire HTC speed 6.6 Programs
båÖäáëÜ
09 VITA In-Ceram YZ speed
10 VITA In-Ceram YZ classic
11 VITA In-Ceram AL classic
12 Sirona service
Loading a program ➢ Press the "S2" key (=YES) to confirm the loading process.
The standby display is shown.
To quit the menu immediately, press the "S1" key (=NO).
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 39
6 Operation Dentsply Sirona
6.6 Programs Operating Instructions inFire HTC speed
NOTICE
Entries with less than four digits
If the entry only consists of three digits (or less), the cursor must be
moved to the "Holding time" column by pressing the corresponding S
key.
61 30 459 D 3497
40 D 3497.201.01.20.02 10.2018
Dentsply Sirona 6 Operation
Operating Instructions inFire HTC speed 6.6 Programs
Programming steps S2 to S4 ➢ Follow the same procedure as described for step S1 to program steps
S2 to S4 (S2 key for step S2, etc.). In order to create a freely
programmable fast sintering program, you must set a heating velocity
of at least > 30 °C (86 °F/min) for S3 (see "Entering the program
values for fast sintering (speed sintering) [ → 42]").
NOTICE
Unrequired steps
If a step is not required, set all values of this step except the heating
velocity to "0". It is not possible to set the heating and cooling-down
speeds to "0", but this is of no importance for the program run.
While steps S1 through S3 may be set to 0, a temperature setting must
be entered for step S4.
Step S4 always specifies the cooling parameters.
båÖäáëÜ
Heating velocity Temperature Holding time
°C/min °C min
Step 4 12 300 0
Step 3 12 1540 120
Step 2 10 0 0
Step 1 10 0 0
°C
1540
120 min
300
t
Graphical representation
This example describes the programming of program number 11 with the
values of the table shown above.
1. Press the Load Program key.
2. Enter the program number 11.
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 41
6 Operation Dentsply Sirona
6.6 Programs Operating Instructions inFire HTC speed
Programming step S3
1. Press the S3 key. The cursor will blink in the "Heating velocity"
column of line S3.
2. Enter the value 12. The cursor will blink in the "Temperature" column.
3. Enter the value 1540. The cursor will blink in the "Holding time"
column.
4. Enter the value 120.
Programming step S4
NOTICE
Entries with less than four digits
If you did not enter the temperature as 0300 and used 300 instead, the
cursor must be manually moved to the "Holding time" column by
pressing the S1 key.
Steps S1 and S2
Since steps S1 and S2 are not needed in this example, you can leave
their values at 0 with the exception of the heating velocity.
6.6.6.2 Entering the program values for fast sintering (speed sintering)
For a fast or speed sintering program with S3, the cooling speed must be
set to > 30 °C (86 °F/min); cooling is then performed via the door opening.
61 30 459 D 3497
42 D 3497.201.01.20.02 10.2018
Dentsply Sirona 6 Operation
Operating Instructions inFire HTC speed 6.6 Programs
Entering a program title (optional) Program titles can be entered on the left side of the display in four lines
(the cursor flashes at the entry position).
Use the "S4" key to move the cursor one space to the right.
Numeric entry
båÖäáëÜ
Use the numeric keys (0-9) for entering numbers.
Text entry
Repeatedly pressing function key "F" displays the letters of the alphabet
in succession.
Once the desired letter is reached, press the "S4" key to insert this letter
in the display.
Deleting a character
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 43
7 Default settings Dentsply Sirona
7.1 Display and acoustic signal Operating Instructions inFire HTC speed
7 Default settings
Key assignment
● S4 key:
This key changes the language of the LCD screen.
● S3 key:
This key activates and deactivates the acoustic signal that sounds
after the end of a program.
● S2 key:
This key sets the current time with the numeric keys (Monday = 1,
Tuesday = 2... Sunday = 7).
The current time always has to be adjusted manually when clocks are
changed for daylight saving time.
● S1 key: This key switches to Dialog 2.
Dialog box 2
Key assignment
● S4 key:
This key sets the temperature at which the furnace door should open
(maximum 300 ℃ /572 °F).
● S3 key:
This key selects a 12-hour or 24-hour time display.
● S2 key:
This key is used to set the temperature scale to be used.
– „°C" = Celsius
– „F" = Fahrenheit
● S1 key:
This key quits the dialog.
The changes are saved in the memory.
61 30 459 D 3497
44 D 3497.201.01.20.02 10.2018
Dentsply Sirona 7 Default settings
Operating Instructions inFire HTC speed 7.2 Adjustment of the start time (casting time function)
båÖäáëÜ
NOTICE
Furnace empty?
No sintering trays may be located in the furnace during the correction
program run.
NOTICE
Long program run
This program takes about 3-4 hours to run.
➢ You can quit the program by pressing the START/STOP key if
necessary.
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 45
8 Description of special functions Dentsply Sirona
8.1 Sintering with ventilation (sintering with a small air gap) Operating Instructions inFire HTC speed
61 30 459 D 3497
46 D 3497.201.01.20.02 10.2018
Dentsply Sirona 9 Regular care
Operating Instructions inFire HTC speed 9.1 General care instructions
9 Regular care
NOTICE
Damage to the heating unit
Avoid all contamination in the heating room to prevent damage to the
heating unit.
NOTICE
Acidic coloring liquids
båÖäáëÜ
Coloring liquids may contain acids which can escape during the
sintering process, significantly reducing the service life of the heating
elements.
Recommendation: Use Sirona water-based coloring liquids:
- inCoris TZI coloring liquids.
IMPORTANT
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 47
10 Maintenance Dentsply Sirona
10.1 Changing fuses F1/F2 Operating Instructions inFire HTC speed
10 Maintenance
1. Unlock the cover of the fuse holder (S) with a screwdriver (press the
cover slightly and turn it counterclockwise).
2. Take the cover plus fuse out of the fuse holder.
3. Pull the fuse out of the cover.
4. Insert a replacement fuse in the cover.
5. Place the cover with the replacement fuse back in the fuse holder.
6. Lock the cover (S) again using a screwdriver (press the cover slightly
and turn it clockwise).
The fuse replacement process is now finished.
61 30 459 D 3497
48 D 3497.201.01.20.02 10.2018
Dentsply Sirona 10 Maintenance
Operating Instructions inFire HTC speed 10.2 Replacing the battery (only for devices with serial numbers < 5000)
NOTICE
Furnace overheating
In a worst-case scenario the battery may experience deep discharging,
after which cooling cannot be guaranteed if the furnace overheats in the
event of a power failure. If the furnace overheats it will be totally
damaged.
NOTICE
båÖäáëÜ
Do not disconnect the device from the power supply
Even if the device is switched off on the main switch of the inFire HTC
speed, the recharging of the battery is ensured by a direct connection to
the charging electronics.
In order to guarantee the functioning of the emergency cooling, the
device may not be disconnected from the power supply, as otherwise
the recharging of the integrated battery will not be possible.
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 49
11 Malfunctions Dentsply Sirona
11.1 Error messages from electronic components Operating Instructions inFire HTC speed
11 Malfunctions
61 30 459 D 3497
50 D 3497.201.01.20.02 10.2018
Dentsply Sirona 11 Malfunctions
Operating Instructions inFire HTC speed 11.2 Miscellaneous errors and their causes
båÖäáëÜ
after the unit is switched on.
No display. Yellow/white (from serial Protection fuse of the furnace is Switch the furnace off, wait 30
numbers > 5000) pilot light of ON/OFF defective. seconds and then switch it back on.
switch is on. The LEDs do not flash If this produces no results: Replace
briefly after the unit is switched on. fuses F1 and F2 [ → 48]. Call
customer service.
No display. Yellow/white (from serial No line voltage Check the fuses in the fuse box,
numbers > 5000) pilot light of ON/OFF check electrical connection. Call
switch is not on. customer service.
Furnace triggers circuit breaker. Unsuitable circuit breaker Check the circuit breakers as
specified.
Defective electronics
Call customer service.
Furnace triggers the circuit breaker in FI is too sensitive or heating is Check whether FI is ≥ 30 mA. Call
the fuse box. defective customer service if necessary.
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 51
12 Disposal Dentsply Sirona
Operating Instructions inFire HTC speed
12 Disposal
In accordance with Directive 2012/19/EU and national disposal
regulations regarding old electrical and electronic devices, please be
advised that such items must be disposed of in a special way within the
European Union (EU). These regulations require the environmentally
friendly recycling/disposal of old electrical and electronic devices. Such
items must not be disposed of as domestic refuse. This has been
expressed using the icon of the “crossed out trash can”.
Disposal procedure
We feel responsible for our products from the first idea to their disposal.
For this reason, we give you an option to return our old electronic and
electrical devices.
If you wish to dispose of your devices, please proceed as follows:
In Germany
To initiate return of the electrical device, please send a disposal request
to enretec GmbH. You have the following options here:
● Use the “Returning an electrical device” button under the “eom” menu
item on the enretec GmbH homepage (www.enretec.de).
● Alternatively, you can also contact enretec GmbH directly.
enretec GmbH
Kanalstraße 17
16727 Velten
Tel.: +49 3304 3919-500
E-mail: [email protected]
In accordance with the national disposal regulations regarding old
electrical and electronic devices (ElektroG), as the manufacturer, we
assume the costs for disposing of the electrical and electronic devices in
question. Disassembly, transport and packaging costs shall be borne by
the owner/operator.
Prior to disassembly/disposal of the product, it must be fully prepared
(cleaned/disinfected/sterilized).
If your unit is not permanently installed, it will be collected from the
practice. If it is permanently installed, it will be picked up curbside at your
address by appointment.
Other countries
For country-specific information on disposal, contact your local dental
dealers.
61 30 459 D 3497
52 D 3497.201.01.20.02 10.2018
Dentsply Sirona Index
Operating Instructions inFire HTC speed
Index
A L
Acoustic signal, 44, 44 Language, 44
Lift key, 18
C
Load program key, 18
Casting time key, 18
Casting-time function, 40, 45 M
CE mark, 12 Memory key, 18
Celsius, 44
N
Controller, 17
NPM sintering, 31
D Numeric keys, 18, 26, 39
Danger of overheating, 14
O
Daylight saving time, 44
Operating devices, 17
DEFECTIVE SENSOR, 50
Operation, 17
DEFECTIVE THYRISTOR, 50
Display, 44 R
Disposal of old electrical and electronic devices, 52 Replacement fuse, 48
REPLACING THE BATTERY, 50
E
Risk of breakage, 19
enretec GmbH, 52
S
F
Safety instructions, 6
Fahrenheit, 44
SECURITY TURN OFF, 50
Fast sintering program, 42
SENSOR + <-> -, 50
Function key, 18
SENSOR SHORT, 50
Fuse, 48
Sintering temperature, 13
G Software version, 19
Gas management module, 31 Speed sintering program, 42
Stage keys, 18
H
Start/stop button, 18
Hardware version, 19
Status display, 19
Heating and cooling down speed, 19
Super speed process, 28
Holding temperature, 19
Super speed program, 27
Holding time, 19
61 30 459 D 3497
D 3497.201.01.20.02 10.2018 53
Index Dentsply Sirona
Operating Instructions inFire HTC speed
T
Temperature, 19
Temperature scale, 44
Temperature steps, 19
The correction program, 45
Time display, 44
V
Ventilation slots, 14
61 30 459 D 3497
54 D 3497.201.01.20.02 10.2018
We reserve the right to make any alterations which may be required due to technical improvements.