Navy Diagnostic Imaging Equipment Performance Survey Manual
Navy Diagnostic Imaging Equipment Performance Survey Manual
Navy Diagnostic Imaging Equipment Performance Survey Manual
ii
NMCPHC TM 6470.1
JUNE 2013
iii
THIS PAGE INTENTIONALLY LEFT BLANK
iv
NMCPHC TM 6470.1
JUNE 2013
TABLE OF CONTENTS
Page
v
NMCPHC TM 6470.1
JUNE 2013
TABLE OF CONTENTS
Page
F. Survey Procedures for Portable and Mobile Radiographic Equipment .......................... 2-6
1. Visual Inspection ....................................................................................................... 2-6
Chapter 3 Dental Radiographic Units..................................................................................... 3-1
A. Minimum Required Personnel Qualifications ................................................................ 3-1
B. Introduction (Intraoral) ................................................................................................... 3-1
C. Introduction (Panoramic) ................................................................................................ 3-1
D. Introduction- Cone Beam Computed Tomography (CBCT) .......................................... 3-1
E. Testing Periodicity .......................................................................................................... 3-1
F. Equipment ....................................................................................................................... 3-2
G. References ...................................................................................................................... 3-2
H. Performance Test Requirements for Intraoral Dental Units ........................................... 3-2
1. Exposure and Timer Reproducibility ........................................................................ 3-2
2. Timer Accuracy ......................................................................................................... 3-2
3. Linearity of mR/mAs................................................................................................. 3-3
4. kVp Accuracy and Precision ..................................................................................... 3-3
5. Beam Quality (Half-Value Layer [HVL] Determination) ......................................... 3-3
6. Source to Skin Distance and X-Ray Field Size/Cone Alignment ............................. 3-4
7. Entrance Skin Exposure (ESE).................................................................................. 3-4
I. Performance Test Requirements for Dental Panoramic Units ......................................... 3-4
1. Exposure and Timer Reproducibility ........................................................................ 3-4
2. Duration of Exposure Cycle ...................................................................................... 3-4
3. Linearity of mR/mAs................................................................................................. 3-5
4. Beam Quality- Half Value Layer (HVL) Determination .......................................... 3-5
5. X-Ray Beam/Image Detector Slit Alignment ........................................................... 3-5
6. Entrance Skin Exposure (ESE).................................................................................. 3-6
7. Calibration file (if applicable) ................................................................................... 3-6
8. Laser Alignment/Phantom Ball and Pin Test (if applicable) .................................... 3-6
J. Performance Test Requirements for Dental Cone Beam Computed
Tomography (CBCT) ...................................................................................................... 3-7
Chapter 4 Fluoroscopy ............................................................................................................. 4-1
A. Minimum Required Personnel Qualifications ................................................................ 4-1
B. Testing Periodicity .......................................................................................................... 4-1
C. Equipment ....................................................................................................................... 4-1
D. References ...................................................................................................................... 4-1
E. Performance Tests for Fluoroscopy ................................................................................ 4-1
1. kVp Accuracy ............................................................................................................ 4-1
2. Air Kinetic Energy Released in Material (KERMA) Rate (AKR) Measurements ... 4-2
vi
NMCPHC TM 6470.1
JUNE 2013
TABLE OF CONTENTS
Page
3. Maximum Air Kinetic Energy Released in Material (KERMA) Rate (AKR) .......... 4-3
4. Transmission Through Primary Barrier .................................................................... 4-3
5. Beam Quality (Half Value Layer) ............................................................................. 4-3
6. Minimum Source to Skin Distance (SSD) ................................................................ 4-4
7. Minimum and Maximum Fluoroscopic Image Size (Beam Limitation Devices) ..... 4-5
8. Fluoroscopy Display Field Alignment ...................................................................... 4-5
9. Beam Central Alignment ........................................................................................... 4-6
10. Pin-cushion and “S-ing’ Distortion ......................................................................... 4-6
11. High Contrast Resolution ........................................................................................ 4-7
12. Low Contrast Sensitivity ......................................................................................... 4-7
13. Automatic Exposure Control (AEC) Systems ......................................................... 4-8
14. kVp Compensation .................................................................................................. 4-9
Chapter 5 Computed Tomography (CT) Units ...................................................................... 5-1
A. Introduction .................................................................................................................... 5-1
B. Minimum Required Personnel Qualifications ................................................................ 5-1
C. Testing Periodicity .......................................................................................................... 5-1
D. Equipment....................................................................................................................... 5-1
E. References ....................................................................................................................... 5-1
F. Performance Tests for ComputedTomography (CT) Units ............................................ 5-2
1. Table Loading............................................................................................................ 5-2
2. Table Positioning ....................................................................................................... 5-2
3. Table/Gantry Alignment............................................................................................ 5-2
4. Kilovoltage (kV) Accuracy ....................................................................................... 5-2
5. Tube Current Linearity .............................................................................................. 5-3
6. Beam Quality ............................................................................................................. 5-4
7. Image Quality: American College of Radiology (ACR) Phantom Tests the
Phantom ..................................................................................................................... 5-4
8. Position Accuracy...................................................................................................... 5-5
9. CT Number Accuracy ............................................................................................... 5-6
10. Low Contrast ........................................................................................................... 5-6
11. Uniformity, Artifact, & Scaling Accuracy .............................................................. 5-7
12. High Contrast .......................................................................................................... 5-8
13. Slice Thickness Accuracy ....................................................................................... 5-8
14. Radiation Dosimetry................................................................................................ 5-9
15. Scatter .................................................................................................................... 5-11
vii
NMCPHC TM 6470.1
JUNE 2013
TABLE OF CONTENTS
Page
viii
NMCPHC TM 6470.1
JUNE 2013
TABLE OF CONTENTS
Page
ix
NMCPHC TM 6470.1
JUNE 2013
TABLE OF CONTENTS
Page
D. Equipment..................................................................................................................... 11-1
E. References ..................................................................................................................... 11-1
Chapter 12 Entrance Skin Kinetic Energy Released in Material (KERMA) (ESK) ........ 12-1
A. Introduction .................................................................................................................. 12-1
B. Minimum Required Personnel Qualifications .............................................................. 12-1
C. Testing Periodicity ........................................................................................................ 12-1
D. Equipment..................................................................................................................... 12-1
E. References ..................................................................................................................... 12-1
F. Entrance Skin KERMA (ESK) for General Radiographic Equipment ......................... 12-1
1. Purpose .................................................................................................................... 12-1
2. Regulations .............................................................................................................. 12-1
3. Procedure ................................................................................................................. 12-1
G. Entrance Skin KERMA (ESK) for Dental Intraoral Units ........................................... 12-2
1. Procedure ................................................................................................................. 12-2
H. Entrance Skin KERMA Rate (ESKR) for Fluoroscopy Units...................................... 12-3
1. Procedure ................................................................................................................. 12-3
I. Digital Spot Entrance Skin KERMA Rate (ESKR) ....................................................... 12-3
1. Introduction ............................................................................................................. 12-3
2. Procedure ................................................................................................................. 12-3
3. Interpretation of Results .......................................................................................... 12-4
Chapter 13 Radiation Shielding Design and Evaluation for Medical and Dental
X-Ray Facilities ................................................................................................... 13-1
A. Introduction .................................................................................................................. 13-1
B. Minimum Required Personnel Qualifications .............................................................. 13-1
C. Design and Evaluation Periodicity ............................................................................... 13-1
D. Equipment..................................................................................................................... 13-1
E. References ..................................................................................................................... 13-1
Chapter 14 Performance Tests for Monitors in Picture Archiving and
Communication System (PAC) .......................................................................... 14-1
A. Introduction .................................................................................................................. 14-1
B. Testing Periodicity ........................................................................................................ 14-1
C. Reference ...................................................................................................................... 14-1
D. General Performance Tests for Picture Archiving and Communication
System (PACS) Monitor Equipment ............................................................................ 14-1
1. Geometric distortion ................................................................................................ 14-1
2. Display Reflection ................................................................................................... 14-2
x
NMCPHC TM 6470.1
JUNE 2013
TABLE OF CONTENTS
Page
xi
NMCPHC TM 6470.1
JUNE 2013
TABLE OF CONTENTS
Page
xii
NMCPHC TM 6470.1
JUNE 2013
FIGURES
Page
xiii
xiv
NMCPHC TM 6470.1
JUNE 2013
TABLES
Page
xv
xvi
NMCPHC TM 6470.1
JUNE 2013
1. Forms. The following Navy and Marine Corps Public Health Center (NMCPHC) forms are
available electronically at: https://navalforms.documentservices.dla.mil/web/public/forms.
2. Reports. The reporting requirements for this manual are exempt from reports control per
SECNAV M-5214.1 of December 2005, paragraph 7j.
xvii
xviii
NMCPHC TM 6470.1
JUNE 2013
ACRONYM LIST
xix
NMCPHC TM 6470.1
JUNE 2013
xx
NMCPHC TM 6470.1
JUNE 2013
xxi
NMCPHC TM 6470.1
JUNE 2013
xxii
NMCPHC TM 6470.1
JUNE 2013
Chapter 1
1-1
NMCPHC TM 6470.1
JUNE 2013
1-2
NMCPHC TM 6470.1
JUNE 2013
1-3
NMCPHC TM 6470.1
JUNE 2013
Qualified Modalities
1. Surveyors should be fully qualified as Basic Diagnostic Imaging Equipment Surveyor before progressing to
Advanced Diagnostic Imaging Equipment Surveyor.
2. Surveyors may be qualified for sublevels (individual systems) within Advanced Diagnostic Imaging
Equipment.
3. Surveyors may perform tests on equipment without being certified; however, the senior regional physicist must
review and approve their survey report.
1-4
Basic Diagnostic Imaging Equipment
NMCPHC TM 6470.1
Surveyor Personal Qualification Standard JUNE 2013
Knowledge Factors:
All applicable knowledge factors in accordance with AAPM Report 90 have been adequately demonstrated.
Practical Factors:
Perform a minimum of three (3) Under Instruction evaluations of each modality:
Each evaluation must include all parameters performed during acceptance testing.
Applicant is qualified to perform all modalities listed as Basic Diagnostic Imaging equipment
BUMED Date
1-5
THIS PAGE IS LEFT INTENTIONALLY BLANK
1-6
Advanced Diagnostic Imaging Equipment
NMCPHC TM 6470.1
Surveyor Personal Qualification Standard JUNE 2013
Knowledge Factors:
All applicable knowledge factors in accordance with AAPM Report 90 have been adequately demonstrated.
Practical Factors:
Perform a minimum of three (3) Under Instruction evaluations of each modality:
Each evaluation must include all parameters performed during acceptance testing.
1-7
Advanced Diagnostic Imaging Equipment
NMCPHC TM 6470.1
Surveyor Personal Qualification Standard JUNE 2013
Applicant is fully qualified as Basic Diagnostic Imaging Equipment surveyor and is recommended for
qualification the following modality: (circle all that apply)
BUMED Date
1-8
NMCPHC TM 6470.1
JUNE 2013
Chapter 2
2-1
NMCPHC TM 6470.1
JUNE 2013
e. Procedure. Set the x-ray tube at 40 inches a. Purpose. To ensure that the x-ray generator is
source-to-table distance, if possible. If using producing the exposure time as set on the
an ionization chamber, place the center of the control panel.
chamber 4 inches above the x-ray table top
and center the chamber in the light field (solid b. Regulations. The accuracy of the timer should
state detectors are usually have lead backing). be within ±5% of the selected timer setting or
Determine the distance from the focal spot to ±1 ms for exposure times less than 10 ms or 1
the center of the ion chamber/solid state pulse for exposure times less than 10 pulses.
detector. Collimate the light field to a narrow
beam geometry (e.g. 4x4 cm field) to include c. Equipment. Exposure meter with timer
the ion chamber/solid state detector. Make combination
radiation exposures at the selected technique.
For efficiency, the evaluator is reminded that d. Procedure. Utilize the procedure described
some meters will read out both exposure and for reproducibility measurements. Measure
time, therefore, record both for future and record the full range of clinically useful
measurements. exposure times.
2-2
NMCPHC TM 6470.1
JUNE 2013
adjustment by a qualified service engineer. d. Procedure. Place the kVp meter on the x-ray
Timer accuracy is critical as it directly table top. Set the distance from the focal spot
influences image quality and patient dose. to the table top as indicated in the kVp meter
owner’s manual. Collimate the beam to the
4. Linearity of mA/mAs active area of the kVp meter. Set the desired
starting kVp, mA, and time stations on the
a. Purpose. To ensure that similar exposures are generator using the manufacturer's suggested
obtained for the same mAs and kVp techniques. Evaluate kVp settings from 50 kV
regardless of the exposure time and mA used. up to the maximum kV incrementing by 5 kV.
During periodic evaluation it may be
b. Regulations. The average ratios of exposure necessary to evaluate only kVp settings from
to the indicated mAs product (mR/mAs) 60 kV to the maximum kV incrementing by
obtained at any two consecutive tube current 20 kV unless further measurements are
settings should not differ by more than 0.10 necessary. Make an exposure and record the
times their sum. display value of the kVp meter.
Where X1 and X2 are the average mR/mAs (1) The Half Value Layer (HVL) should
values obtained at each of two consecutive always be measured after assuring the kVp
tube current settings. (21CFR1020.31) is correct.
c. Equipment. Exposure meter with timer (2) The major cause of kVp variation is
combination calibration. Some generators maintain
their calibration well and others drift
d. Procedure. Utilize the setup described for constantly. It is important to note that a
reproducibility measurements. Measure and change in kVp may not always show as a
record the exposures at 5 different mA change in image density because changes
settings while keeping kVp and time constant. in the mA will often compensate for the
With some x-ray units, the mA cannot be change in kVp.
varied without varying time. In this instance
mA must be constant and time varied. Divide (3) Since the kVp affects the radiographic
the mR output by mAs setting, record contrast, it must be checked to assure that
mR/mAs as calculated. it is acceptable.
e. Interpretation of Results. If each of the (4) Other major causes of variations in kVp
average ratios between mA stations deviate are line voltage drops and electrical
from the criteria listed in 4.b, consult a component failure.
qualified service engineer. Linearity of
mA/mAs is critical as it directly influences f. Interpretation of Results. Refer units
image quality and patient dose. deviating from the criteria listed in 5b for
adjustment by a qualified service engineer.
5. Kilovoltage Accuracy Proper kVp calibration is critical as it directly
influences image quality and patient dose.
a. Purpose. To ensure that the x-ray generator is
producing the kVp as indicated on the control 6. Beam Quality
panel.
a. Purpose. To assure that the permanently
b. Regulations. The accuracy must be ± 5 % of installed filtration at the x-ray tube is
the nominal control panel setting or within maintained at an appropriate level to help
manufacture specifications. minimize patient exposure.
2-3
NMCPHC TM 6470.1
JUNE 2013
c. Equipment. Ionization chamber or solid state determine if the unit contains too much
detector, five 1 mm Type 1100 Aluminum filtration.
(Al) sheets, two 0.5 mm Type 1100 Al sheets
(if available) 7. Light Field Intensity
d. Procedure. Place the ion chamber 5 cm above a. Purpose. To ensure that the light field
the table top (solid state detectors are usually intensity is adequate to illuminate the field.
have lead backing and do not need to be above
the table). Collimate the light field to a narrow b. Regulations. The light should provide an
beam geometry to include the ion average illumination of not less than 160 lux
chamber/solid state detector. The Al sheets (15 foot candles) at 100 cm or at the
should be placed between the ion maximum Source Image Distance (SID),
chamber/solid state detector and the x-ray tube whichever is less. (21CFR1020.31)
at a distance X/2, where X = focal spot to
detector distance. Make sure the Al sheets c. Equipment. Light meter capable of providing
intercept the entire beam (light field). Make either lux or foot candles.
two exposures without any Al sheets in the
beam, one before and one after, ensuring that d. Procedure. Place the light meter on the x-ray
the geometry has not changed. (An exposure table top. Set the SID to 100 cm or the
made using 80 kVp, 0.10 sec and 320 mA to maximum available whichever is less.
achieve an output of approximately 300 mR Collimate the x-ray beam to a 25 x 30 cm
will ensure that you have a high enough field. Illuminate the field. Measure and record
exposure to make the measurements the illumination in the 4 quadrants. Calculate
accurately). Add Al sheets and make an average.
additional exposures until the exposure is less
than half of the original exposure. e. Interpretation of Results. Consult a qualified
Recommend using 2, 3, and 4 mm Al. service engineer if the alignment deviates
Remove all Al sheets and make one exposure. from the criteria listed in 7b.
If exposure is not within 2% of the initial
exposure, made with 0 mm of Al, repeat the 8. Light Field/X-Ray Beam Alignment
measurement series ensuring that the
technique and geometry selected remain the a. Purpose. To ensure that the x-ray field and
same throughout the procedure. the light field are congruent.
e. Possible Pitfalls.
b. Regulations. The light field/x-ray field
alignment should be within ± 2% of the SID
(1) The entire ion chamber must be in the x-
(21CFR1020.31).
ray beam. When placing the sheets of Al
in the beam, be sure that the entire beam is
c. Equipment. Digital rulers, fluorescent film
intercepted by the Al sheet. Once
selected, the technique factors must not be
d. Procedure. Place rulers at edge of light field
altered for subsequent exposures.
according to manufacturer’s directions. If
using film, mark the edge of the light field on
(2) The kVp should be checked before
the film and expose and measure distance
measuring the HVL to ensure that it is
from edge of exposed area to mark on film.
within acceptable limits.
e. Interpretation of Results. Consult a qualified
(3) The Al used for HVL measurements
service engineer if the alignment deviates
should be type 1100.
from more than 2% of the SID.
f. Interpretation of Results. If the HVL is not
greater than the minimum requirements listed 9. X-ray Field Size- Indicated vs. Actual
in 21CFR1020.30, consult a qualified service
engineer. If the HVL is greater than 3.5 mm of a. Purpose. To ensure that the actual and
Al, further evaluation should be conducted to indicated X-ray field are congruent.
2-4
NMCPHC TM 6470.1
JUNE 2013
a. Regulations. The indicated vs. actual x-ray 12. Automatic Exposure Control (AEC)
field should be within ± 2% of the SID. System (if applicable)
(21CFR1020.31)
a. Purpose. To ensure that the Automatic
b. Equipment. X-ray beam alignment tool and Exposure Control (AEC) system is responding
measurement plate adequately. The system compensates for
variations in technique factors and patient
c. Procedure. Measure the deviation between thickness such that resulting films appear with
the displayed and the actual x-ray field. constant, optimal densities. The following
AEC parameters should be evaluated during
d. Interpretation of Results. Consult a qualified testing: reproducibility, balance, maximum
service engineer if the alignment deviates exposure time, kVp compensation, thickness
from more than 2% of the SID. compensation and density control tracking (if
available).
10. Central Beam Alignment
b. Regulations. For each AEC cell (phototimer)
a. Purpose. To ensure that the central x-ray measurement the reproducibility should be
beam is perpendicular to the table. within ± 5%. Back up timer should terminate
the exposure at 600 mAs or 2000 mAs for
b. Regulations. The perpendicularity of the tube potentials less than 50 kVp.
central beam should be within 5mm.
c. Equipment. 4 cm Al, 1.6 mm Lead (Pb) plate,
c. Equipment. X-ray beam alignment tool and ionization chamber/solid state detector
measurement plate
d. Procedure. For a radiographic system, set the
d. Procedure. If the x-ray beam alignment tool x-ray tube at 40 in (72 in for chest systems)
was used, measure the deviation between the SID distance and center to the bucky. Set
upper (magnified) bead and the lower bead. selector such that only one phototimer is
activated. To determine the location of the
e. Interpretation of Results. Consult a qualified phototimer(s) look at the chest unit pattern of
service engineer if the perpendicularity rectangles on the chest board surface. Use the
measured deviates by more than 5 mm. same layout for the table, noting that the
center chamber is usually located at the lateral
11. Indicated Source to Image Distance center of the table when the tube and bucky
(SID) are aligned. If using a Computed
Radiography (CR) machine, place a cassette
a. Purpose. To ensure the actual SID distance in the bucky tray (not necessary for digital).
and the indicated SID are congruent. Place 4 cm of Al in the beam. Ensure that the
Al covers all the AEC detector cells. Set the
a. Regulations. The actual SID should be within control in the AEC mode, 80 kVp and select
± 2% of the indicated SID. the detector to be checked (e.g. table, center,
left or right chamber). If using a CR system, a
b. Equipment. Tape measure single cassette should be used for all testing,
to reduce variability. This will require
c. Procedure. Position the tube assembly at 40 processing the cassette after each exposure.
inches from the image receptor. Measure the
SID. If automatic detent is available, position (1) Output Reproducibility. Use the setup
the assembly utilizing the detent. Measure the previously described in 12.d. Place the ion
SID. chamber/solid state detector along the
beam central axis at the phantom beam
d. Interpretation of Results. Consult a qualified entrance surface. Set the technique at
service engineer if the measurement deviates 80 kVp, 200 mA, AEC setting to neutral
from the criteria in Table 2.1. (0). If using a CR machine, ensure you
have a clean cassette for this test. Perform
three exposures, record the readings and
2-5
NMCPHC TM 6470.1
JUNE 2013
calculate their mean. All three readings 13. Entrance Skin Exposure Measurements
should line within ± 5% of their mean. (ESE)
Repeat for each detector (note: if using CR
and your process your cassette after each See Chapter 12
exposure you can obtain data (mAs) needed
for phototimer balance test).
F. Survey Procedures for Portable and
(2) Back-up Timer. Use the setup previously Mobile Radiographic Equipment
described. Place the lead sheet (lead apron
can also be used) in the area of the beam The following modifications of quality control
as to cover all of the AEC detector fields procedures and acceptance parameters from fixed
so that no radiation reached them. Set the x-ray units apply for portable and mobile units:
technique at 80 kVp, 200 mA, AEC setting
to neutral (0). Record the elapsed mAs. 1. Visual Inspection
The beam should terminate prior to the
accumulation of 600 mAs or 2000 mAs for a. The minimum source to skin distance must be
tube potentials less than 50 kVp. no less than 12 inches (30 cm). This can be
measured directly with a tape measure
(3) Phototimer Balance. Use the setup provided the location of the focal spot is
previously described. Set the technique at known.
80 kVp, 200 mA, AEC setting to neutral
(0). Record the elapsed mAs and use the b. The operator must be able to stand at least six
mR from the output reproducibility test to feet away from the x-ray tube during the
chart mR/mAs. actual exposure. This is normally
accomplished by attaching the exposure
(4) Patient Thickness Compensation. Use the switch to the unit with at least a six foot long
setup described previously. Set the cord.
technique at 80 kVp, 200 mA, AEC setting
to neutral (0). Vary the Al thickness over c. Each portable unit should be supplied with at
the range: 2, 4 cm Al. Record the elapsed least two lead aprons and gonadal shields.
mAs for each image and measure Optical
Density (OD) at the center of each image.
The densities should lie within the range
of ± 0.3 of the baseline density.
2-6
NMCPHC TM 6470.1
JUNE 2013
2-7
NMCPHC TM 6470.1
JUNE 2013
2-8
General Radiographic Unit Survey NMCPHC-TM 6470.1
JUNE 2013
Facility: Date:
Manufacturer:
Test
Performed
Pass Fail N/A Comments (failure comments must annotate minor or significant finding)
Exposure
Reproducibility
Timer
Reproducibility
Timer Accuracy
Linearity of
mA/mAs
kVp Accuracy
Beam Quality
Light Field/X-ray
beam alignment
Central Beam
Alignment
Indicated SID
PBL
Focal Spot
Size/Constancy
AEC
Safety Equipment/
Mechanical Checks
Additional Comments:
Purpose: Results:
Surveyor Name:
Surveyor Signature:
2-10
NMCPHC TM 6470.1
JUNE 2013
Chapter 3
a. Timer accuracy;
b. Beam quality; and,
c. Beam/film slit alignment.
4. Table 3-2 list required survey requirements
3-1
NMCPHC TM 6470.1
JUNE 2013
3-2
NMCPHC TM 6470.1
JUNE 2013
NOTE: This test cannot be performed on (2) Center the end of the cone on the kVp
fixed kVp or mA units. meter so that the x-ray field will COVer
the required area of the kVp meter.
a. Purpose. To ensure that similar exposures are
obtained for the mAs and kVp regardless of (3) Check 90 kVp at one-half second, 80 kVp
the exposure time and mA. at one second and 70 kVp at two seconds.
Four measurements should be obtained at
b. Regulations. The average ratios of exposure the most clinically used setting.
to indicated mAs (mR / mAs) obtained at two
tube current settings should not differ by more (4) For fixed kVp units, determine the actual
than 0.10 times their sum. kVp.
c. Equipment. Same as above (5) Allow for tube cooling between longer
shots, e.g., 1minute for 1 second, 2
d. Procedure. minutes for 2 seconds, etc.
(1) With the equipment in the same set-up as e. Interpretation of Results. The meter reading
above, record one of the reproducibility should be within 5 kVp of each setting. The
results as the first reading. coefficient of variation should be less than
0.02.
(2) Switch to another mA station if one exists
while holding kVp and timer settings 5. Beam Quality (Half-Value Layer
constant. [HVL] Determination)
(3) Make an exposure and record pulse a. Purpose. To assure that the permanently
exposure, then divide mR output by mAs installed filtration at the x-ray tube is
setting. maintained at an appropriate level to help
minimize patient exposure.
(4) Record this mR / mAs as calculated.
b. Regulations. The minimum value of the HVL
e. Interpretation of Results. These two mR/mAs should be as stated in Table 3-4 for the actual
results should be similar, specifically the kVp determined above.
difference between the two divided by sum of
the two should not exceed 10%. Repeat this c. Equipment. Electrometer with small ion
test at several kVp settings. chamber, sheets of type 1100 alloy Al
a. Purpose. To ensure that the x-ray generator is (1) Set the tube voltage potential to 80 kVp, if
producing the kVp as indicated on the control the unit has variable kVp.
panel.
(2) Take an exposure using the reproducibility
b. Regulations. The accuracy must be within 5 test set-up.
kVp of the control panel setting (Some units
have fixed kVp and must be within 5 kVp of (3) Measure the radiation and record the value
that value). as the exposure with 0 mm Al added.
c. Equipment. kVp meter (4) Tape 1 mm Al (use tape which does not
leave marks, such as paper surgical tape,
or whatever is conveniently available) on
3-3
NMCPHC TM 6470.1
JUNE 2013
the end of the cone and take a reading at 6. Source to Skin Distance and X-Ray
the same settings, recording this for 1 mm Field Size/Cone Alignment
Al added. Maintain good beam geometry.
a. Purpose. To determine the minimum source
(5) Repeat for 2, 3 and 4 mm Al. to patient distance and field size.
(6) Finally, remove all Al and take one last b. Regulation. The source to skin distance and
reading with zero mm Al. The four mm field size should be as stated in 21 CFR
trial is not needed if 3 mm added Al 1020.31(f) & (h).
reduces the initial reading by half. If the
final exposure is not within 2% of the c. Equipment. Measuring tape and fluorescent
initial exposure made with 0 mm Al, screen
repeat the measurement series ensuring
that the technique and geometry selected d. Procedure.
remain the same throughout the procedure.
(1) Measure and record the length of the
e. Interpretation of Results. removable cone, the distance between the
focal spot and end of the cone and the
(1) Use the average of the two 0 readings as inner diameter of the cone.
the unattenuated value.
(2) Use the fluorescent screen to ensure the
(2) The HVL may be determined x-ray beam at the end of the cone is the
mathematically using logarithmic same size as the cone.
interpolation or graphically using semi-log
paper. Refer to the general radiographic
beam quality section for further 7. Entrance Skin Exposure (ESE)
instructions.
See Chapter 12.
(3) The HVL must meet Food and Drug
Administration (FDA) standards for the I. Performance Test Requirements for
actual kVp used which was determined Dental Panoramic Units
above. FDA standards for HVLs, for
dental units, are included in Table 3-4. 1. Exposure and Timer Reproducibility
a. Procedure. (Same as for the dental intraoral
unit.) The ion chamber must be secured to the
chin rest with adhesive tape for measurement
to be taken.
3-4
NMCPHC TM 6470.1
JUNE 2013
3-5
NMCPHC TM 6470.1
JUNE 2013
b. Regulations. The beam dimensions should comparing the calibrated and non-calibrated
not exceed the slit opening. images, notice that the calibrated image fills
the frame window completely, without any
c. Equipment. Fluorescent screen or intraoral white areas on any sides of the image. Also,
film and tape the calibrated image has an even, consistent
quality while the non-calibrated image
d. Procedure. appears “noisier”, with sharper differences
between the rows.
(1) This test may be performed in real time by
using a piece of fluorescent screen taped to c. Equipment. Computer image processing and
the image receptor holder COVering the calibration software as provided by
slit. Mark the outline of the slit on the manufacturer upon installation.
screen. Dim the room lighting and
position yourself so as to be able to see the d. Procedure. This procedure should be
screen. Make an exposure and watch for performed by the manufacture technical
the slit area to glow. representative upon installation and
demonstrated during acceptance. The specific
(2) The slit alignment may also be recorded procedure should be obtained from the
on film for documentation as follows: manufacturer’s technical manual.
(a) Tape two pieces of intraoral film Note: For bi-directional systems, if the
diagonally across the slit, one at the top panoramic machine scans in both
and one at the bottom of the slit, or left-to-right and right-to-left
place a piece of ready pack film across directions, you will need to create
the film holder. two calibration files, one for each
direction
(b) Mark the slit location using a pin to
prick the film at the edge of the slit 8. Laser Alignment/Phantom Ball and Pin
opening and make an exposure only a Test (if applicable)
few seconds in duration.
a. Purpose. To ensure the positioning lasers are
(c) Develop the film. properly aligned to provide sharp and
symmetric images.
e. Interpretation of Results.
b. Regulation. The balls in the center of the
(1) Fluorescent screen: The entire slit should picture should be sharp and symmetric. If
be seen. they are too wide and blurred, the lateral laser
is positioned to far forward. The opposite is
(2) For film, a diagonal line should be seen true if the balls appear elongated and narrow.
across each film from corner to corner or
between pin marks. c. Equipment. Ball and pin phantom as supplied
by the manufacturer
6. Entrance Skin Exposure (ESE)
d. Procedure.
(See Chapter 12)
(1) Turn on the positioning lasers, and adjust
7. Calibration file (if applicable) the lateral laser position so its projected
line falls on the center ball of the ball and
a. Purpose. To ensure the digital image is pin phantom. If necessary, manually
properly calibrated to only view the area in adjust the laser to ensure proper alignment.
which radiation is being detected on the image If manual adjustment of the laser is not
receptors. available, contact the system field engineer
to have the laser alignment corrected.
b. Regulations. Average Pixel Value should be
within manufacture’s specifications. When
3-6
NMCPHC TM 6470.1
JUNE 2013
3-7
NMCPHC TM 6470.1
JUNE 2013
Abbreviations: A: acceptance, P: periodic, HVL: Half Value Layer, kVp: kilovolt peak, mA: milliamp, mAs:
milliamp seconds, SSD: Source To Skin Distance
3-8
NMCPHC TM 6470.1
JUNE 2013
10 repeat
A cov< 0.05
Exposure measurements
1.
Reproducibility if cov>0.05, do 4
P 4 repeat measurements
more
Duration of Exposure Measure during
2. A/P 1 Second
Cycle exposure reproducibility
< 0.1 of the sum of
Measure at 2 mA the measurements
3. mAs Linearity A/P
stations if available of adjacent mA
stations
HVL of 2.3 mm Al
4. X-ray Beam Quality A/P @ 80 kVp, 1st HVL
or greater
-View beam slit
-View of entire
A fluorescence using
image receptor slit
fluorescent screen or
X-ray Beam/Image
5. Receptor Slit -Diagonal line
-Expose 2 pieces of across each film
Alignment intraoral film taped
P from corner to
diagonally across beam corner
slit or use ready pack
-Mark slit on film
To be performed by
manufacturer IAW manufacturer’s
6. Calibration file A
representative upon specifications
installation.
Ensures lasers are
Laser Alignment /
properly aligned to IAW manufacturer’s
7. Phantom Ball and Pin A/P
produce sharp and specifications
Test
symmetric images
Abbreviations: A: acceptance, P: periodic, cov: coefficient of variation, HVL: Half Value Layer, kVp:
kilovolt peak, mA: milliamp, mAs: milliamp seconds
3-9
NMCPHC TM 6470.1
JUNE 2013
3-10
NMCPHC TM 6470.1
JUNE 2013
1
Dental x-ray systems designed for use with intraoral image receptors and manufactured after December 1, 1980.
2
Dental x-ray systems designed for use with intraoral image receptors and manufactured before or on December 1,
1980, and all other x-ray systems subject to this section and manufactured before June 10, 2006.
3
All x-ray systems, except dental x-ray systems designed for use with intraoral image receptors, subject to this
section and manufactured on or after June 10, 2006.
3-11
THIS PAGE IS LEFT INTENTIONALLY BLANK
3-12
Dental Radiographic Unit Survey NMCPHC-TM 6470.1
JUNE 2013
Facility: Date:
Manufacturer: Type:
Test
Performed
Pass Fail N/A Comments (failure comments must annotate minor or significant finding)
Exposure
Reproducibility
Timer
Reproducibility
Timer Accuracy
Linearity of
mA/mAs
kVp Accuracy
Beam Quality
kVp Precision
X-ray field
size/cone alignment
Minimum SID
Duration of
Exposure Cycle
X-ray Beam/Slit
Alignment
Safety Equipment/
Mechanical Checks
Additional Comments:
Purpose: Results:
Surveyor Name:
Surveyor Signature:
3-14
NMCPHC TM 6470.1
JUNE 2013
Chapter 4
Fluoroscopy
4-1
NMCPHC TM 6470.1
JUNE 2013
(3) Record average or effective kVp, as (6) Try to use a phantom to chamber distance
available, when using a meter offering of approximately 8 cm. This distance
multiple reading formats. allows for adequate shielding of medium
and larger image intensifiers. Very large
d. Interpretation of Results. Refer units IIs may require that the phantom be placed
deviating from the criteria in Table 4-1 for closer to the ion chamber.
adjustment by a qualified service engineer.
Proper kVp calibration is critical as it directly (7) Place the scatter grid in the beam path.
influences image quality and patient dose.
(8) Collimate the field to the phantom
2. Air Kinetic Energy Released in ensuring chamber is in the Field Of Vision
Material (KERMA) Rate (AKR) (FOV).
Measurements (9) Maintain consistent phantom/ solid state
detector/ion chamber/image intensifier
a. Purpose. positions to assure reproducibility (record
distances).
(1) To establish AKR for varying thicknesses.
d. Measurement Considerations.
(2) To verify that measured and displayed AKR
are within ±35 percent. (1) Make AKR measurements using all
available output modes and II size
(3) To verify long term AKR consistency. combinations. Include manual and pulse
modes, if available. ABC systems will
(4) To verify proper automatic brightness control demonstrate different output rates at each
of exposure rate with varying Image II size to compensate for the loss of
Intensifier (II) field size. magnification gain.
b. Equipment. 4 cm Al or 15 cm acrylic (2) For manual mode readings, adjust kVp and
phantom, solid state detector or ion chamber mA to provide a monitor image brightness
with exposure meter. equal to that of ABC normal mode.
c. Procedure. (3) Make AKR measurements with and
without the grid in place, as warranted.
(1) Typical and maximum AKR The grid generally remains in the beam but
measurements can be made with the same may be removed if no-grid studies are
equipment arrangement. performed.
(2) Place the solid state detector or ion (4) Use minimal “beam-on” time to prevent
chamber at the location specified by 21 unnecessary x-ray tube wear.
CFR 1020.32.
(5) If the unit is equipped with High Level
(3) Invert C-arms for testing. This allows for Control (HLC), a distinct tone must be
easier phantom placement. heard when HLC is active.
(4) Treat lithotripsy systems as standard/ C-
arm hybrids (i.e. meet both conditions). e. Cine Output Measurements.
(5) Place the phantom close to the solid (1) Most cine systems work at 7.5, 15 and
state/ion chamber. If using an ion 30 frs-1.
chamber, make sure that there is enough
distance as to minimize backscatter. The II (2) Measure AKR using the most commonly
is very sensitive; ensure that it is always used II size and ABC or manual
shielded by the phantom. 1100 Al alloy or techniques suitable for an average patient
acrylic are acceptable. (i.e. 4 cm aluminum/15 cm acrylic).
Evaluate all available II sizes during
acceptance.
4-2
NMCPHC TM 6470.1
JUNE 2013
f. Interpretation of Results. Typical AKR values influenced (10 to 15 % above equivalent free
should be significantly lower than their in air measurements) by backscatter from the
maximum output rate counterparts. Use lead sheet if it is too close to the phantom/lead
acceptance inspection values to set baselines assembly. If maximum exposure rates exceed
for future reference. Subsequent annual limits set in 21 CFR 1020.32, recommend that
evaluation results should agree reasonably the unit be temporarily removed from patient
well with original levels (e.g. 10 %). use and recalibrated by a qualified service
engineer as soon as possible. If practicable,
3. Maximum Air Kinetic Energy Released verify that the new maximum exposure rates
are acceptable before the service engineer
in Material (KERMA) Rate (AKR)
leaves the facility.
a. Purpose. To prevent excessive exposure to
patients subjected to fluoroscopic 4. Transmission Through Primary Barrier
examinations by verifying that the maximum
AKR conforms to the limits of 21 CFR a. Purpose. To verify that the radiation
1020.32. attenuation provided by the II housing is
adequate.
b. Regulations. 21 CFR 1020.32 specify
maximum exposure rates allowed for b. Equipment. 4 cm Al or 15 cm acrylic
fluoroscopic equipment. phantom, 1.6 mm Pb plate, solid state detector
or ion chamber with exposure meter.
c. Equipment. 4 cm Al or 15 cm acrylic
phantom, 1.6 mm Pb plate, solid state detector c. Procedure.
or ion chamber with exposure meter.
(1) Arrange the fluoroscopy unit, phantom,
d. Procedure. and Pb sheet in the same manner as for
evaluating maximum AKR.
(1) Set up the fluoroscopy unit, phantom, and
solid state detector/ion chamber as for (2) Place the large ion chamber 10 cm beyond
typical AKR measurements. the rear surface of the primary barrier (i.e.
II housing) with the large flat surface
(2) Place the lead sheet on top of the phantom perpendicular to the beam axis.
between the solid state/ion chamber and
image intensifier. (3) Irradiate the phantom using the maximum
AKR technique. Record the radiation
(3) Make AKR measurements for all available level and compare it with the maximum
output modes. For manual modes, set kVp AKR recorded previously.
to its maximum level.
d. Interpretation of Results. Radiation levels at
(4) Maximum AKR measurements need only 10 cm beyond the II housing should not
be made at the largest II size. exceed 3.34 x 10-3 percent of the entrance
AKR. Refer units showing excessive radiation
(5) Radiation streaming around the lead plate transmission for repair by a qualified service
should not be visible during testing. engineer.
e. Interpretation of Results. Ensure that proper b. Regulations. 21 CFR Part 1020.30 specifies
solid state/ion chamber/lead/II distances are the minimum beam quality (HVL)
maintained. When using an ion chamber, requirements for a range of tube potentials.
maximum AKR measurements may be unduly
4-3
NMCPHC TM 6470.1
JUNE 2013
c. Equipment. 1100 aluminum alloy HVL (d) Repeat the measurement with 1
sheets, solid state detector or ion chamber through 8 mm Al between the tube and
with exposure meter, 4 cm aluminum or ion chamber; moving each Al sheet
15 cm acrylic phantom. from behind the ion chamber to in
front of it near the tube output. A
d. Procedure. constant Al thickness must remain in
the beam throughout the procedure to
(1) Set up the fluoroscopy unit, phantom, and prevent ABC from changing technique
solid state detector/ion chamber as for factors. Varying factors will lead to
typical AKR measurements. erroneous readings.
(2) If the unit allows manual kVp and mA (4) Determine HVL for the appropriate
control, use the following procedure: voltage potential (set manually or obtained
through ABC) mathematically.
(a) Manually set kVp = 90.
e. Interpretation of Results. 21 CFR 1020.30
(b) Place the 4 cm Al or 15 cm acrylic lists minimum HVLs for various voltage
phantom between the solid state potentials. If the beam does not meet the
detector/ion chamber and II. Allow minimum standard, refer the unit for
some separation between the two to adjustment by a qualified service engineer.
minimize the effect of backscatter (ion Insufficient filtration may lead to unnecessary
chamber). patient dose. A unit with a hard beam need
not be removed from service. However, a
(c) Under fluoroscopy, collimate the beam high HVL often indicates the presence of an
to an area just larger than the ion older tube that may fail shortly thereafter.
chamber. Ensure that the phantom
always intercepts the beam. Failure to 6. Minimum Source to Skin Distance
do so may damage the II. (SSD)
(d) Set mA to produce an output rate a. Purpose. To prevent unnecessary patient
between 300 and 500 mR/min. exposure resulting from an unduly short SSD.
(e) Measure the exposure rate without any b. Regulations. 21 CFR 1020.32 specifies the
Al sheets between the tube and ion minimum source to skin distance requirements
chamber. Repeat the measurement based on fluoroscopy unit mobility and
with 1, 2, 3, 4, and 5 mm Al between application.
the tube and ion chamber.
c. Equipment. Tape measure, etched brass plate,
(3) If the unit does not permit manual 14” x 17” (35 cm x 43 cm) CR cassette.
technique control (i.e. ABC only), use the
following procedure: d. Procedure.
(a) Place the Al or acrylic phantom and (1) For C-arm systems, determine minimum
collimate the beam as per steps 5.d.(2) SSD using a tape measure. Measure from
(b) and (c). the external target position mark to the end
of the collimator assembly or spacing cone
(b) Place all 8 mm Al sheets between the if permanently installed.
ion chamber and II (e.g. above the Al
or acrylic phantom). (2) For fixed SID, overhead tube systems,
measure minimum SSD in the same
(c) Measure the exposure rate without any manner as step (1).
Al sheets between the tube and ion
chamber, allowing ABC to set kVp for (3) For fixed under-table tube systems that
all the aluminum in the beam (i.e. allow tube access, measure minimum SSD
phantom + sheets). using a tape measure as the distance from
4-4
NMCPHC TM 6470.1
JUNE 2013
the target mark to the tabletop. For (4) Expose the cassette for 1 - 2 sec using a
systems with variable SSD, set the target low technique (50-60 kVp @ 1 mA).
to table distance to minimum before
measuring. (5) Close all collimators to minimum setting.
(4) For fixed SSD, under-table systems (6) Move the CR cassette over to align the
without tube access, measure minimum center of the image intensifier with a
SSD using triangulation. Calculate SSD corner of the cassette.
as:
(7) Re-expose the cassette per step (4).
OID
SSD =
w2 w1 - 1 (8) Measure the dimensions of the darkened
areas on the processed film. Correct the
measurements if a significant cassette to II
Where: distance existed during exposure.
OID = Brass plate to film image distance
e. Interpretation of Results. If the maximum or
w2 = Division length at SID
minimum field size dimensions exceed
w1 = Division length on the plate
tolerance limits, recommend that a qualified
service engineer recalibrate the collimators.
e. Interpretation of Results. If the source to skin
One method to eliminate the film based beam
distance is less than required, refer the unit for
limitation test procedure is to calibrate the
adjustment by a qualified service engineer.
collimator shutters so that they are just visible
along the edges of the live image at maximum
7. Minimum and Maximum Fluoroscopic field size. Once the collimators are properly
Image Size (Beam Limitation Devices) calibrated, maximum field size conformance
can be verified visually on the monitor image.
a. Purpose. To verify that the fluoroscopic
imaging system displays the geometrically 8. Fluoroscopy Display Field Alignment
appropriate anatomical area of interest.
a. Purpose. To verify that the fluoroscopy beam
b. Regulations. 21 CFR 1020.32 specifies that is properly collimated so that only the tissue
the minimum radiation field size at maximum volume corresponding to the active entrance
SID should be contained within a square of 5 area of the II is irradiated, & that the same
cm by 5 cm. volume is presented on the monitor.
c. Equipment. Etched brass plate, 14” x 17” b. Equipment. Etched brass plate, plastic
(35 cm x 43 cm) CR cassette (do not need for cylinder with stacked steel balls, 14” x 17”
DR). (35 cm x 43 cm) CR cassette or GAF film.
d. Procedure. c. Procedure.
(1) Arrange the unit for maximum SID, (1) Arrange the system for minimum SID,
largest available II size, grid in the beam, largest available II size, grid in the beam,
and all collimators fully open. and collimators fully open.
(2) Position the brass plate between the tube (2) Position the brass plate to obtain an Object
and image intensifier to fully intercept the to Image Distance (OID) of approx.
beam. 30 cm. and collimate the image as
necessary so that the plate fully intercepts
(3) Using appropriate protection, place the CR the beam.
cassette as close to the II face as possible
with the screen facing the tube. Center the (3) Place the cassette as close to the II face as
CR cassette over the II housing assembly. possible with the screen facing the tube.
Center the CR cassette over the II housing
assembly.
4-5
NMCPHC TM 6470.1
JUNE 2013
(4) Expose the CR cassette using normal circles indicating central axis deviations of
fluoroscopy to acquire a background film 1.5 and 3 degrees from the perpendicular.
density of approximately 1.2 ( 1 sec at 80
kVp and 200 mA). d. Interpretation of Results. If the beam axis/II
misalignment exceeds 1.5 degrees, refer the
(5) On both the monitor and CR images, system for imaging chain repositioning by a
determine the indicated distance between qualified service engineer.
opposing edges of the viewing field (LCD)
or radiation field (cassette) along the two 10. Pin-cushion and “S-ing’ Distortion
axes on the plate.
a. Purpose. To verify that the fluoroscopic
(6) Compare the axis lengths in the monitor image contains minimal spatial distortion and
and film images and calculate the artifacts.
difference between the two as a fraction of
SID. (1) It is difficult to quantify an amount of
acceptable distortion. However, any
(7) If the unit allows, increase SID to distortion should be horizontally and
maximum and repeat steps (5) and (6) vertically symmetrical. It should also be
during acceptance testing. In a properly visibly similar for fluoroscopic, cine, and
functioning unit, collimation should track digital spot images produced using the
automatically with changing SID. same II.
d. Interpretation of Results. If the difference (2) Two major forms of spatial distortion are
between the lengths of either monitor/film pincushion distortion and S-ing.
axis pair exceeds 3 % of SID or if the sum of Pincushion is characterized by bowing of
the differences for both axis pairs exceeds 4 % peripheral chords into the center of the
of SID, refer the system for recalibration by a image. S-ing is characterized by warping
qualified service engineer. of straight lines passing through the center
of the image into “S” shapes in the central
9. Beam Central Alignment quarter to third of the image.
b. Equipment. Etched brass plate, plastic (1) Verify that the unit meets the standards for
cylinder with stacked steel balls, 2-Dimension fluoroscopy display field and beam central
level. alignment.
(3) Reposition the plate to provide four equal d. Interpretation of Results. If the amounts of
axis arm lengths. On the monitor image, pincushion distortion or S-ing exceed the
locate the position of the upper steel ball levels prescribed in Table 4-1, refer the
relative to the pair of etched concentric system for adjustment by a qualified service
engineer. Due to the subjectivity of this test,
4-6
NMCPHC TM 6470.1
JUNE 2013
last hold hard copy reference images showing 12. Low Contrast Sensitivity
the level of distortion during acceptance may
be invaluable during subsequent periodic a. Purpose. To verify the system’s ability to
testing. display low contrast information.
4-7
NMCPHC TM 6470.1
JUNE 2013
formed with sub-second photon bursts, (3) Place the CR cassette in the bucky.
making them difficult to assess visually. Low
pulse rate images should not be held to the (4) Place a 4 cm aluminum or 15 cm acrylic
same standards as their continuous beam phantom in the beam in the same manner
counterparts. If the observed sensitivity does as for measuring AKR. Ensure that the
not meet the baseline set at acceptance, refer phantom covers all the AEC detector cells.
the unit for adjustment by a qualified service
engineer. (5) Set the II field to its largest setting,
collimating to the phantom periphery if
13. Automatic Exposure Control (AEC) necessary. Fluoroscopy the phantom
briefly, allowing the ABC to select an
Systems appropriate kVp. Several systems apply
the ABC selected kVp directly to the
a. Purpose. mechanical spot film technique. For those
that do not, the fluoroscopy kVp serves as
(1) AEC systems attached to fluoroscopy spot a useful guideline for manual spot film
film devices provide the same function as technique programming. For units without
their radiographic system counterparts; i.e. ABC, use 80 kVp.
compensation for variations in technique
factors and patient thickness such that (6) If more than one detector cell is available
resulting spot films appear with constant, and cells can be programmed to work
optimal densities. independently, select the center cell,
otherwise use all cells simultaneously.
(2) This evaluation assumes proper operation
of the processor used to develop spot (7) Use a single CR cassette for testing. This
films. It also assumes that the AEC will require processing the image after
system is calibrated for the CR each exposure.
combination used with the unit.
Therefore, the processor, CR cassette, and (8) Measure and record the Optical Density
film used for testing should be those (OD) at the center of the field. The OD
actually used during patient imaging. should be at least 1.2. The radiologist may
set a higher baseline density. The range of
(3) The following AEC parameters should be
densities should be within 0.15 of the
evaluated during testing: reproducibility,
baseline density.
maximum exposure time, kVp
compensation, patient thickness
d. Output Reproducibility.
compensation, density control function,
and multi - image format (field size)
(1) Use the basic imaging chain arrangement
compensation.
and phantom thickness. Place the solid
state detector or ion chamber along the
b. Equipment. 4 cm Al or 15 cm acrylic
beam central axis at the phantom beam
phantom, 1.6 mm Pb plate, 14 in x 17 in
entrance surface. Set technique factors as
(35 cm x 43 cm) CR cassette, solid state
follows: kVp from the fluoroscopy image
detector or ion chamber with exposure meter.
or, in the absence of ABC, 80 kVp; 200
mA, AEC setting to neutral (0).
c. Procedure.
(2) Irradiate the phantom, ion chamber and
(1) Arrange the unit in the same configuration
cassette holding exposed film three times.
used for measuring fluoroscopic AKR.
Record the exposure readings and
Ensure that if a grid is used clinically, it is
calculate their mean.
in the beam path during testing.
(3) All three readings should lie within 5%
(2) Record the SID, film/screen combination,
of their mean.
and film size used for future testing
reproducibility.
4-8
NMCPHC TM 6470.1
JUNE 2013
(1) Use the basic imaging chain arrangement (1) Use the basic imaging chain arrangement
and phantom thickness. Place the lead and phantom thickness. Place the solid
sheet over the AEC detector fields so that state detector or ion chamber just off the
no radiation reaches them. Set technique beam central axis at the phantom beam
factors as follows: kVp from the entrance surface. Set technique factors as
fluoroscopy image or, in the absence of follows: kVp from the fluoroscopy image
ABC, 80 kVp; 200 mA, AEC setting to or, in the absence of ABC, 80 kVp; 200
neutral (0). Retain the previously exposed mA.
film from the reproducibility test.
(2) Vary AEC density over the range of
(2) Irradiate the phantom until AEC shuts off available positive and negative settings,
the beam. Record the elapsed mAs. exposing a new CR cassette (not needed
for DR) for each setting. Record the
(3) The beam should terminate prior to the elapsed mAs, density at the center of each
accumulation of 600 mAs. film, and exposure for each image.
4-9
NMCPHC TM 6470.1
JUNE 2013
4-10
NMCPHC TM 6470.1
JUNE 2013
7a. Beam Limitation A/P Max SID, minimum collimation, Dark area 5 x 5
(Minimum Field Size) film at II face cm2
7b. Maximum Field Size A/P Max SID, widest collimation, Dark area
max II size, film at II face maximum nominal II
size
Collimator tracks with
changing SID
8. Fluoro Display Field A Largest II size, CR cassette at II Diff between CR
Alignment face, etched plate at table top image and screen X
For variable SID units: determine or Y axis lengths 3
at minimum and maximum SID % of SID
Sum of X & Y axis
diff 4 % of SID
P Test at largest II size, min SID only Same as acceptance
9. Beam Central A/P Minimum SID, max II size, widest 1.5 degrees from
Alignment collimation vertical
* Alternatively use 15 cm thick acrylic phantom, if available.
10a Pin-Cushion Distortion A/P Manual or ABC (NL) mode at Spatial linearity
largest II size visually uniform over
center 75 % of FOV
10b “S-ing” Distortion A/P Manual or ABC (NL) mode at Etched plate lines
largest II size visually linear along
X & Y axes in center
75 % of FOV
11. High Contrast A All available output modes Use Manufacturer’s
Resolution [manual, ABC (NL & HLC), pulse, suggested values.
cine, spot (digital & mech.)] at
each II size
Manual - 60 kVp
ABC only - minimum kVp provided
for 1 mm Al in beam
4-11
NMCPHC TM 6470.1
JUNE 2013
12. Low Contrast A All available output modes See at least 3.1 mm
Sensitivity [manual, ABC (NL & HLC), pulse, test tool holes at 2 %
cine, spot (digital & mech.)] at contrast (4 cm Al)
each II size (Pulse modes
Manual - 85 to 90 kVp, as exempt)
available
ABC only - kVp provided by unit
4-12
General Fluoroscopy Unit Survey NMCPHC TM 6470.1
JUNE 2013
Facility: Date:
Manufacturer: Type:
Test Performed Pass Fail N/A Comments (failure comments must annotate minor or significant finding)
kVp Accuracy
Maximum AKR
Transmission
Through Primary
Barrier
Minimum SSD
Minimum / Maximum
Field Size
Fluoro Display Field
Alignment
Beam Central /
Mechanical Spot Film
Alignment
Pin-Cushion / “S-ing”
Distortion
High Contrast
Resolution / Low
Contrast Sensitivity
Mechanical Spot Film
AEC
Mechanical / Digital
Spot Film ESE
Safety Equipment/
Mechanical Checks
Additional Comments:
Purpose: Results:
Surveyor Name:
Surveyor Signature:
4-14
NMCPHC TM 6470.1
JUNE 2013
Chapter 5
A. Introduction
5. Tape measure, 10 - 20 cm ruler (marked in mm)
X-ray CT scanners represent a departure from 6. Bubble level
conventional film/screen x-ray performance. CT 7. CT compatible kVp meter
scanners generate a thin, well collimated beam of x- 8. Computerized Tomography Unit Survey
rays to a cross section of the patient's body from
multiple rotational angles. The transmitted beam is E. References
collected by radiation detectors, and the information
is fed into a computer which analyzes the data and 1. American Association of Physicists in Medicine,
constructs an image which reflects variations of the Phantoms for Performance Evaluation and
physical attenuation characteristics of the material. Quality of CT Scanners, Report no.1, 1977.
2. American Association of Physicists in Medicine,
The CT is unique in its ability to detect exceptionally Standardized Methods for Measuring Diagnostic
fine variations in linear attenuation of adjacent X-rays Exposures, Report no.31, 1990.
structures and incorporate these differences into a 3. American Association of Physicists in Medicine,
diagnostic quality image suitable for further use in Specification and Acceptance Testing of
radiation therapy treatment planning and stereotactic Computed Tomography Scanners, Report no. 39,
surgical navigation. 1993.
4. Code of Federal Regulations, Title 21, Chapter 1,
Due to the potential complexity of CT scanner Part 1020.30, 1020.31, 1020.32; 1 April 1996
operation, a radiological technologist, trained and edition.
experienced on the CT scanner being evaluated 5. Curry, T.S. III, Dowdey J.E., Murry, R.C. Jr. 1990.
should be present during testing. The technologist Christensen's Physics of Diagnostic Radiology.
should operate the scanner while the physicist Lea & Febiger, Philadelphia.
performs the tests. 6. ImPACT, CT Scanner Acceptance Testing,
Version 1.02, 2001.
B. Minimum Required Personnel 7. C. H. McCollough, et al., “The phantom portion of
Qualifications the American College of Radiology (ACR)
Computed Tomography (CT) accreditation
Advanced Diagnostic Imaging Equipment (CT) program: Practical tips, artifact examples and
pitfalls to avoid,” Medical Physics, 31, 2423-2442
(2004).
C. Testing Periodicity 8. Gray, J.E., Winkler, N.T., Stears, J., Frank, E.D.
Quality Control in Diagnostic Imaging;
All units: Annually, upon acceptance and after major University Park Press, Baltimore, 1983
repairs, i.e. x-ray tube replacement. 9. American Association of Physicists in Medicine,
Quality Control in Diagnostic Radiology, Report
D. Equipment no.74, 2002.
10. American Association of Physicists in Medicine,
1. Electrometer with 10 cm CT and large (180 cc) Comprehensive Methodology for the Evaluation
volume ion chambers of Radiation Dose in X-ray Computed
2. Manufacturer’s QC Phantoms Tomography, Report no.111, 2010.
3. ACR, AAPM or Catphan performance phantom 11. ACR CT Accreditation website, ACR.org
4. Computed Tomography Dose Index (CTDI)
phantoms (16 cm diameter-head and 32 cm
diameter-body)
5-1
NMCPHC TM 6470.1
JUNE 2013
c. Equipment. Ruler
1. Table Loading
d. Procedure.
a. Purpose. To verify manufacturer’s weight
loading specifications for the patient support (1) Using the laser light, raise the scanning
device. table until the lateral lasers intersect the
horizontal plane.
b. Criteria. Refer to manufacturer’s
specifications. (2) Insert the table into the gantry opening.
c. Equipment. Weights and/or persons totaling (3) Scan the table using the standard head
the manufacturer’s loading specifications. technique. Using the electronic ruler and
grid, project the distance from grid center
d. Procedure. Distribute specified weight over to right and left table edges onto the grid.
table top in proportion to normal weight
distribution. Check full range of vertical and (4) Compare the two distances and determine
horizontal motion. Record maximum weight the difference between them.
and range of motion. Do not load table
beyond manufacturer’s specification. (5) Calculate misalignment as half the
difference between the two measurements.
e. Interpretation of Results. If table loading
requirements do not meet manufacturer’s e. Interpretation of Results. Misalignment of the
specifications consult a qualified service two table edges and isocenter should be
engineer. < 5 mm. If it is not consult a qualified service
engineer.
2. Table Positioning
4. Kilovoltage (kV) Accuracy
a. Purpose. To ensure that table movement and
localization is accurate. a. Purpose. To ensure that the x-ray generator is
producing the kVp as indicated on the control
b. Criteria. Manufacturer’s specifications or panel.
+ 1 mm.
b. Criteria. The accuracy must be ± 5% of the
c. Equipment. Ruler. nominal control panel setting or within
manufacturer’s specifications.
d. Procedure. Tape a ruler to the fixed portion of
the patient support assembly. Make a mark on c. Equipment. Noninvasive tube potential
the table adjacent to the tape measure. Move measuring device calibrated for CT use.
the table both in and out of the gantry to
predetermined distances. Record the actual d. Procedure. Noninvasive devices may not
and selected distances traveled (typically 1, 10 work well in CT if detector sensitive region is
and 40 cm.). too large for narrow CT beams. Be certain that
the device is calibrated for the kV range and
e. Interpretation of Results. The table should beam quality to be tested (consult device
move smoothly and accurately to within 1 mm manufacturer) and is set for direct current (dc)
of target in either movement direction. or three-phase wave forms. Hardening
Consult a qualified engineer if the requirement corrections may be necessary with some
is not met. instruments to compensate for heavily filtered
CT beams.
3. Table/Gantry Alignment
(1) From operator’s console, rotate tube to
a. Purpose. To ensure proper alignment of the overhead Anterior/Posterior (AP) position.
table and gantry isocenter.
5-2
NMCPHC TM 6470.1
JUNE 2013
(2) Put table at lowest scan position, move 5. Tube Current Linearity
tabletop into gantry opening and place kV
sensor on table. Align detector(s) to scan a. Purpose. To ensure radiation exposures vary
alignment light. If instrument detector is linearly with tube current (mA).
large, place it at bottom of gantry opening
where the field size is greatest, with b. Criteria. (10 CFR 1020) The average ratios of
tabletop out of field. exposure to the indicated mAs product
(mR/mAs) obtained at any two consecutive
(3) Operate CT system in service mode with tube current settings should not differ by more
tube and tabletop stationary (consult than 0.10 times their sum: (X1-X2) <
service engineer). 0.10(X1+X2).
(4) Set widest collimator setting and expose Where X1 and X2 are the average mR/mAs
detector. Proceed only if instrument values obtained at each of two consecutive
obtains a reading without error indication. tube current settings.
(5) Measure tube potential at each generator c. Equipment. Calibrated Ionization chamber or
power level for each potential setting. solid state detector and electrometer.
Record which focus is selected with dual
focus xray tubes. d. Procedure.
e. Interpretation of Results.
5-3
NMCPHC TM 6470.1
JUNE 2013
directly influences image quality and (5) Add Al sheets and make additional
patient dose. exposures until the exposure is less than
half of the original exposure. Note that
(2) Significant deviation from linearity may CT beams are heavily filtered, and the
indicate miscalibration of potential, HVL is normally greater than 5 mm Al at
current, or exposure time. If the coefficient 90 keV.
of linearity of mGy/mAs between the
mean of all values and any single value (6) Remove all Al sheets and make one
(absolute difference divided by sum) is exposure. If exposure is not within 2% of
greater than 0.05 then significant the initial exposure, made with 0 mm of
miscalibration may be present. If time Al, repeat the measurement series ensuring
values are obtained, determine if measured that the technique and geometry selected
times correspond to scan time settings. remain the same throughout the procedure.
Errors with constant potential generators
can indicate shutter problems. e. Possible Pitfalls.
b. Criteria. Federal and many state regulations (2) The kVp should be checked before
specify minimum required HVLs at various measuring the HVL to ensure that it is
kVp values (21 CFR 1020.30). Refer to within acceptable limits.
manufacturer’s specifications for HVL.
(3) The Al used for HVL measurements
c. Equipment. Ionization chamber or solid state should be type 1100.
detector, a total of 10 mm Al Type 1100
plates; combination of 1 mm and 0.5 mm Al f. Interpretation of Results. The HVL should
plates. easily meet minimum HVL thickness per
21 CFR 20120. However, if the HVL
d. Procedure. significantly exceeds manufacturer’s
specifications a qualified service engineer
(1) Position the radiation detector as described should be contacted immediately. Excessive
in section 2 above. HVLs will reduce CT tube life.
(2) Set the CT system to operate in service 7. Image Quality: American College of
mode with tube and tabletop stationary as Radiology (ACR) Phantom Tests The
described in section 2 above. Phantom
(3) The Al sheets should be placed between The American College of Radiology (ACR) CT
the ion chamber/solid state detector and accreditation phantom has been designed to
the x-ray tube at a distance X/2, where X = examine a broad range of scanner parameters.
focal spot to detector distance. Make sure These include:
the Al sheets intercept the entire beam.
Make two exposures without any Al sheets a. Positioning accuracy
in the beam, one before and one after, to b. CT number accuracy
ensure that the geometry has not changed. c. Image thickness
d. Low contract resolution
(4) Acquire an exposure using 90 kVp, and e. High contrast resolution
mAs to provide an exposure of f. CT number uniformity
approximately 300 mR. g. Image noise
5-4
NMCPHC TM 6470.1
JUNE 2013
The ACR CT accreditation phantom (Gammex 464) choose a patient orientation of “supine
is a solid phantom containing four modules, and is head first” on the scanner.)
constructed primarily from a water-equivalent
material. Each module is 4 cm in depth and 20 cm in (2) Carefully position the phantom so that the
diameter. There are external alignment markings CT scanners alignment lights are
scribed and painted white (to reflect alignment lights) accurately positioned over the scribe line
on EACH module to allow centering of the phantom corresponding to the center of Module 1
in the axial (z-axis, cranial/caudal), coronal (y-axis, (FOOT END of the phantom). Use the set
anterior/posterior), and sagittal (x-axis, left/right) of alignment lights, internal or external,
directions. There are also “HEAD”, “FOOT” and that are used clinically. Align the phantom
“TOP” markings on the phantom to assist with in the sagittal, coronal, and axial planes.
positioning as shown in Figure 5-1. Zero (or landmark) the table at this point
(or be sure to note the table location, as all
scans will be acquired in reference to this
location). While maintaining careful
alignment, secure the phantom so it will
not move.
8. Position Accuracy
a. Purpose. To ensure the CT phantom is
aligned properly with the scanner and
alignment lights (lasers).
5-5
NMCPHC TM 6470.1
JUNE 2013
9. CT Number Accuracy
a. Purpose. To ensure that the CT numbers Polyethelne Bone
associated with air, water, acrylic,
polyethylene and bone are accurate and that
the system is properly calibrated.
5-6
NMCPHC TM 6470.1
JUNE 2013
(3) Figure 5-4 shows low contrast resolution within + 5 HU of the center ROI mean
image at WW = 100 and WL = 100 with value.
ROI placements.
NOTE: Manufacturer’s criteria for
uniformity may differ from the
ACR test protocol and
standards. In this case if the
facility is not applying for or
under ACR Accreditation
status, the results should meet
the manufacturer’s
specification.
(2) Use the absolute value of the difference – (3) Figure 5-5 shows Module 3 showing
that is, do not take into consideration placement of uniformity center and edge
whether the CNR is a positive or negative ROIs.
number.
b. Criteria.
5-7
NMCPHC TM 6470.1
JUNE 2013
5-8
NMCPHC TM 6470.1
JUNE 2013
d. Procedure. c. Equipment.
(1) View the images in Module 1 using the (1) Phantom Data Form, with average facility
adult abdomen protocol obtained from protocols for adult head and body, and
three different slice width scans. pediatric head (1 year old) and body
(5 year old, 40-50 lbs).
(2) Count the number of wires visualized in
each of the two slice thickness ramps (2) Calibrated CTDI (pencil) ionization
(Figure 5-3). chamber (typically 10 cm in length)
(3) Count the number of wires in the top (3) Calibrated electrometer.
section separately from the number of
wires in the bottom section. (4) Acrylic (PMMA) cylindrical phantoms,
having cylindrical holes at 1 cm from the
(4) The images should be viewed with a WW edge, and one at the center (Figure 5-7):
~ 400 and a WL ~ 0.
(a) Head CTDI phantom: 16 cm diameter
(5) Count any wire that appears to be 50% or
more as bright as the central wires. (b) Body CTDI phantom: 32 cm diameter
b. Criteria. ACR Accreditation Standards. (1) Position the phantom appropriately at the
isocenter of the scanner. Ensure that the
phantom is correctly aligned in all three
Pass/Fail Criteria Reference Levels
planes (sagittal, axial, and coronal).
Examination CTDLvol (mGy) CTDLvol (mGy)
Adult Head 80 75 (a) For the adult position the 16 cm
Adult Abdomen 30 25 phantom in the head holder.
Pediatric Head (1 yr) To be determined 45
(b) For the adult abdomen scans, position
Pediatric Abdomen (5 yr, 25 20
40-50 lb) the 32 cm phantom on the table top.
5-9
NMCPHC TM 6470.1
JUNE 2013
(d) For the pediatric head scans, position (5) Repeat the scan two more times and record
the 16 cm phantom on the table top. the measurements from each scan in the
Excel Dose Spreadsheet. If the data differ
(2) Connect the pencil chamber to the by more than 5%, check your equipment
electrometer and insert the pencil chamber and rescan the data until the three
into the central hole in the phantom. measurements agree within 5%.
Ensure that all other holes (those at 3, 6, 9,
and 12 o’clock positions) are filled with (6) The spreadsheet will calculate the average
acrylic rods. measurement from scans in mR, and the
CTDI at isocenter in phantom in mGy.
(3) Using the appropriate protocol as entered
in the phantom data form, acquire a single (7) Move the pencil chamber from the center
axial slice at the center of the phantom, position to the 12 o’clock peripheral
with no table increment. If the protocol is position. Ensure that an acrylic rod is then
normally scanned helically, change to an inserted into the vacated isocenter
axial scan, keeping the remaining technical position.
parameters unchanged. All CTDI dose
information must be acquired using axial (8) Repeat steps 3 through 7 and record the
scans. value in mGy as the CTDI at 12 o’clock
position.
(a) In multislice CT, CTDI is a function of
detector configuration. It is imperative (9) The spreadsheet will calculate the
that the detector configuration and total following:
beam width used matches the site’s
Clinical Protocol (NxT) as closely as (a) Average of the three measurements
possible. from the 12 o’clock position in mR
(b) If NxT used for dosimetry does not (b) Head CTDI at 12 o’clock position in
exactly match the clinical value, be mGy
sure to modify the table increment used
in the calculation to yield the same (c) CTDIw in mGy
pitch value as used clinically.
(d) CTDIvol in mGy
(4) Record the following in the Excel Dose
Spreadsheet for the appropriate (e) DLP in mGy-cm
examination:
(f) Effective Dose in mSv
(a) kVp
(10) Repeat steps 1 through 9 for adult head
(b) mA and body, and pediatric head and body
clinical CT protocols.
(c) Exposure time (sec)
e. Interpretation of Results.
(d) Z-axis collimation (T, in mm)
(1) Compare the measured CTDIVOL values
(e) Number of data channels used with the ACR recommended reference
levels.
(f) Table Increment (mm) used to yield
the clinical pitch (2) If the measured CTDIVOL values exceed
ACR Pass threshold criteria, consult the
(g) Active chamber length of pencil lead interpreting radiologist and senior CT
chamber technologist to adjust scanning protocols
to acceptable dose levels.
(h) Chamber correction factor
(i) Exposure in mR
5-10
NMCPHC TM 6470.1
JUNE 2013
15. Scatter
a. Purpose. To establish and maintain control of
the scatter pattern created by the CT scanner.
e. Interpretation of Results.
5-11
NMCPHC TM 6470.1
JUNE 2013
5-12
NMCPHC TM 6470.1
JUNE 2013
11. High Contrast A/P ACR Phantom Module 4 5 lp/cm for adult
Evaluating limiting abdomen protocol;
resolution (lp/cm) pattern 6 lp/cm for high
resolution adult
chest protocol
12. Slice Thickness A/P ACR Phantom Module 1 Measured slice
Accuracy Evaluate 3 different slice thickness should be
thickness and measure within + 1.5 mm of
the steps in the top and nominal setting.
bottom ramps.
13. Radiation Dosimetry A/P CTDI volume of adult Adult Head < 80
head & abdomen mGy
protocol, and pediatric Adult Abd < 30 mGy
head & abdomen Ped Head < TBD
protocol Ped Abd < 25
mGy
14. Scatter A Largest slice thickness 100 mR/yr to the
Body FOV, standard public in occupied
body algorithm spaces
CTDI Body Phantom
simulating patient,
measure @ occupied
locations
Abbreviations: A: acceptance, P: periodic, HU: Hounsfield units, NL: Normal, FOV: Field Of
Vision, ROI: Region Of Interest, DPSC: Defense Personnel Support Center, MTF: Modulation
Transfer Function, RAD: Radiation Absorbed Dose (1 rad = 100 erg/g), MSAD: Multiple Scan
Average Dose, CTDI: Computed Tomography Dose Index, mR: millirad = 1/1000 rad, kV: kilovolt,
SMPTE: Society of Motion Picture and Television Engineers
5-13
THIS PAGE IS LEFT INTENTIONALLY BLANK
5-14
Computerized Tomography Unit Survey NMCPHC TM 6470.1
JUNE 2013
Facility: Date:
Manufacturer:
Test Performed Pass Fail N/A Comments (failure comments must annotate minor or significant finding)
Table Loading
Table positioning
Table, Gantry
Alignment
kVp Accuracy
Tube Current
Linearity
Position Accuracy
(Laser Alignment)
CT number
Accuracy
Low Contrast
Uniformity
Artifact
Scaling Accuracy
High Contrast
Slice Thickness
Accuracy
Radiation
Dosimetry
Scatter
Additional Comments:
Purpose: Results:
Surveyor Name:
Surveyor Signature:
5-16
NMCPHC TM 6470.1
JUNE 2013
Chapter 6
Mammographic Units
A. Introduction E. References
The MQSA; Correction; Final Rule, 21 CFR 16 1. Code of Federal Regulations, Title 21, Chapter 1,
and 900, November 10, 1997, requires any facility Parts 1020.30, 1020.31, 1020.32; 3 May 1993
that produces, processes or interprets mammograms edition.
to be certified by the FDA. To be certified, facilities 2. Mammography Quality Standards Act (MQSA) of
must meet the federal regulations and must be 1992.
accredited by a FDA approved private or state 3. Quality Assurance in Mammography. American
accrediting body. College of Radiology (ACR), 1994.
MQSA
C. Testing Periodicity
Annually
D. Equipment
6-1
THIS PAGE INTENTIONALLY LEFT BLANK
6-2
NMCPHC TM 6470.1
JUNE 2013
Chapter 7
D. Equipment d Procedure.
1. Tissue equivalent phantom with average velocity (1) Rest ultrasound transducer on scanning
of sound of 1540 m/s 10 m/s and an surface of the phantom.
attenuation coefficient of 0.5 to 0.7 dB/MHzcm
2. Ultrasound Scanner System Survey (2) Adjust the focal zone and depth controls
until the vertical pin targets are clearly
visualized and freeze the image.
7-1
NMCPHC TM 6470.1
JUNE 2013
measured distance and compared with (3) Scan phantom and freeze image. Measure
known distance. and record the maximum depth of
visualization of background echoes from
2. Horizontal Distance Accuracy the phantom. This is the distance from the
top of the scan window to the deepest
a Purpose. To determine the accuracy of cylindrical or spherical object that is
distances measured perpendicular to the beam barely visible. Record the depth of
axis. penetration.
7-2
NMCPHC TM 6470.1
JUNE 2013
b Regulations. Visual determination of opposite the phantom cysts) and freeze the
satisfactory or unsatisfactory per Table 7-1. image. Record the settings for future use.
7-3
NMCPHC TM 6470.1
JUNE 2013
variations are normal, the pin width should b Regulations. Refer to manufacturer’s
remain relatively constant over time. A specifications.
service engineer should be contacted if beam
width changes by more than 1 mm from c Equipment. TM phantom
baseline values.
d Procedure.
8. Axial Resolution
(1) Adjust the focal zone and depth controls
a Purpose. To determine the instrument’s until the target cyst is clearly visualized,
ability to detect and clearly display closely the edges of the cyst are sharply defined,
spaced objects that lie on the beam’s axis. and the cyst interior is echo free. Freeze
the image.
b Regulations. Refer to manufacturer’s
specifications. NOTE: Bright spots at the top and
bottom of cysts are specular
c Equipment. TM phantom reflections and are considered
normal for some systems.
d Procedure.
(2) Measure and record the height and width
(1) For each axial resolution target group of each cyst.
(near, middle, and far), adjust the focal
zone and depth controls to scan the target (3) Rate the cyst for no distortion, minor
group and freeze the image. distortion, and/or major distortion.
(2) Determine the axial resolution by finding (4) If using a phantom that has cysts of
the two pin targets with the smallest different sizes, determine the smallest cyst
vertical spacing that are visible as distinct at near, middle, and far field zones.
objects and have no vertical overlap and
record the resolution level. (5) Repeat steps 1-3 for each phantom cyst.
Record settings for future use.
(3) Record machine baseline settings and use
the same settings for all follow up e Interpretation of Results. Cyst imaging
procedures. combines aspects of contrast resolution and
image uniformity into a single test. Cyst
NOTE: Pin targets larger than 0.15mm image quality can be affected by electrical
in diameter may produce a noise, side lobes in the transducer beam and
doubling artifact for transducer problems in the image processing hardware.
frequencies greater or equal to Flattened cyst indicates geometric distortion.
5 MHz. Echoes inside the cyst may be the result of
system noise or side lobe contamination. A
e. Interpretation of Results. Axial resolution service engineer should be contacted if you
depends on the transducer’s spatial pulse observe any major distortions e.g. height
length or pulse duration, which depend on the differs from width by 20% or more; are
center frequency and damping factor. Axial detected or any measureable change from
resolution should remain stable over time. A baseline.
service engineer should be contacted if any
consistent measurable change from baseline 10. Display Monitor and Hardcopy Fidelity
values.
a Purpose. To independently evaluate the
9. Cyst Imaging (Anechoic Imaging) monitor and printer capabilities.
c Equipment. TM phantom
7-4
NMCPHC TM 6470.1
JUNE 2013
d Procedure.
7-5
NMCPHC TM 6470.1
JUNE 2013
7-6
NMCPHC TM 6470.1
JUNE 2013
7-7
THIS PAGE IS LEFT INTENTIONALLY BLANK
7-8
Ultrasound Scanner System Survey NMCPHC TM 6470.1
JUNE 2013
Facility: Date:
Manufacturer:
Test Performed Pass Fail N/A Comments (failure comments must annotate minor or significant finding)
Display Monitor
Fidelity
Hardcopy Fidelity
Image Uniformity
Vertical Distance
Accuracy
Horizontal Distance
Accuracy
Anechoic Object
Imaging
Axial Resolution
Lateral Resolution
Dead Zone
Physical and
Mechanical
Inspection
Additional Comments:
Transducer Model Number: Serial Number:
Purpose: Results:
Surveyor Name:
Surveyor Signature:
7-10
NMCPHC TM 6470.1
JUNE 2013
Chapter 8
(1) Place the uniform, spherical phantom at Some manufacturers do not allow analysis of pre-
the center of the magnet. The phantom scan or manual shimming. In these cases, you may
should have a large spherical volume if have to use manufacturer software to provide a line
possible. Prepare the phantom for a width. Contact your field engineer to find out details.
simple T1 weighed spin echo scan.
e. Procedure (Method 2 – Phase Diff Map).
(2) During pre-scan, the system should check
the central frequency and transmitter gain (1) Requires the availability to display phase
attenuation. Record this information. images, which may not be available on all
MRIs.
e. Interpretation of Results:. Have the facility
contact the system field engineer if center (2) Position the phantom in the center of the
frequency drift exceeds prescribed large volume (body) Radiofrequency (RF)
manufacturer tolerances or action limits coil. Scan the phantom using a simple
established by the facility’s quality control field echo (spoiled gradient echo)
program (typical expected range = nom freq sequence. Do not use a RF spin echo to
(Hz) + 100 Hz). avoid rephrasing field inhomogeneity
caused of phase differences.
8-1
NMCPHC TM 6470.1
JUNE 2013
(2) Scan the phantom twice using a Gradient c. Equipment. ACR MRI Phantom
Echo (GRE) sequence appropriate for the
scanner (e.g. flip angle = 25, TR = 52 ms, d. Procedure.
TE = 8.5 ms, thickness = 6 mm, scan time
1:06, 256 x 256 matrix, 400 mm FOV) and (1) Place the large ACR MRI phantom in the
alternately with small and large head coil and align to phantom crosshairs.
Bandwidths (BW). The low BW should Advance the table to the magnet isocenter.
be near minimum (e.g. < 20 Hz pixel-1);
the high BW should be near the system’s (2) Acquire a sagittal localizer (scout) image
available maximum (e.g. ~ 500 Hz pixel-1). using the following parameters: 1 slice,
sagittal spin echo, TR = 200 ms, TE = 20
(3) Measure and record the displayed phantom ms, slice thickness = 20 mm, FOV =
diameters for each band width (For axial 25 cm, matrix = 256 x 256, NEX = 1, scan
image, measure in the left to right time = 0:56 sec.
direction; in coronal and saggital images,
measure in the anterior to posterior (3) Display the resulting sagittal slice. Inspect
direction). the image to ensure the entire phantom is
visible
(4) Calculate the difference between the B0
field at a given voxel and the reference (4) Acquire axial slices of the relevant
value at the center of the field of view: phantom sections using 11 slices, starting
at the vertex of the crossed 45 degree
wedges at the inferior end of the ACR
8-2
NMCPHC TM 6470.1
JUNE 2013
phantom and ending at the vertex of the length difference is twice the actual slice
crossed 45 degree wedges at the superior displacement error. The magnitude of the bar
end. The ACR T1 series: (11 slices, spin length difference should be less than or equal
echo, TR = 500 ms, TE = 20 ms, FOV = to 5 mm.
25 cm, slice thickness = 5 mm, slice gap =
5 mm, matrix= 256 x 256, NEX = 1). 5. Slice Thickness Accuracy
(5) As necessary, the ACR T2 series: (11 a. Purpose. To determine accuracy of specified
slices, double echo spin echo, TR = 2000 slice thickness.
ms, TE / TE2 = 20 / 80 ms, FOV = 25 cm,
slice thickness = 5 mm, slice gap = 5 mm, b. Regulations. None
matrix= 256 x 256, NEX = 1)
c. Equipment. Same as above
(6) As necessary, determine and record the
facility’s Axial Head T1 and T2 weighted d. Procedure.
sequences.
(1) Display slice #1 with a magnification of
4. Slice Position Accuracy two to four while keeping the slice
thickness insert fully visible (The slice
a. Purpose. To determine whether locations of thickness inserts are the two opposed
acquired axial slices differ excessively from horizontal thick bright lines).
their prescribed locations for a properly
operating scanner. (2) Adjust the display level so that the signal
ramps are well visualized. Place a
b. Regulations. None rectangular ROI at the middle of each
signal ramp. Note the mean signal values
c. Equipment. ACR MRI Phantom for each of the two ROI’s and determine a
joint ramp man signal value. If the two
d. Procedure. ROI mean values differ more than 20%,
the ROI’s may include some of the area
(1) Display the T1-weighted series. surrounding the ramps. Adjust the ROI
Differences between the prescribed and sizes to only include the ramps and note
actual positions of slices #1 and #11 are the new mean values and joint mean value.
measured.
(3) Lower the display level to one half of the
(2) Display the slice magnified on the monitor average ramp signal. Leave the display
by a factor of 2 to 4. Keep the vertical window set to minimum.
bars of the crossed wedges with the
displayed portion of the image. (4) Use the display station length
measurement tool to measure the lengths
(3) Adjust the display window so that the ends of the top and bottom ramps. Record these
of the vertical bars are clear using a measurements.
narrow display window. The display level
should be set to a level one half that of the e. Interpretation of Results. The slice thickness
signal of the bright portions of the is calculated using the following formula:
phantom.
Slice Thickness = 0.2 x (top x bottom) / (top x bottom)
(4) Use the viewer’s measurement tool to
measure the difference in length between where “top” and “bottom” refer to measured
the left and right bars. If the left bar is lengths of the top and bottom ramps
longer, then assign a negative value to the respectively. This measurement is best for
number; if the right bar is longer, assign a slice thicknesses between 3 mm and 7 mm.
positive value. For the ACR T1-weighted axial imaging
series, the measured slice thickness should be
e. Interpretation of Results. The crossed wedges 5.0 mm + 0.7 mm.
have a 45 degree slope. Therefore the bar
8-3
NMCPHC TM 6470.1
JUNE 2013
a. Purpose. To ensure that the radio frequency (b) Perform these tests for at least one
coils are functioning properly. volume coil using the ACR T1 and T2
sequences during annual inspections
b. Regulations. None and all volume coils during acceptance.
(c) Select a second ROI that covers (b) Determine mean and standard
approximately 0.15% of the FOV (for deviation values as for the volume coil
a 256 x 256 matrix, the ROI should be above, using appropriate
about 100 pixels). Move this ROI to proportionalities, or as directed by the
the area of maximum intensity within manufacturer.
the “mean signal ROI”. This area can
be found by narrowing the window and (c) Calculate SNR:
adjusting the level value to find the
brightest area. Determine the ROI SNR = Mean Signal / Noise Standard Deviation
mean signal in this area. Record as
“high”. Repeat this measurement for (d) Observe signal intensity distribution
the location of lowest signal intensity. and note whether it generally appears
This value is “low”. the same as when previous
measurements were performed on the
(d) Select a third oval ROI (Approx 120 coil. During acceptance, save a hard or
cm2). Using the ROI, measure the softcopy image with window settings
mean signal values above, below, and for future reference.
to the left and right of the phantom.
Calculate “ghost signal” as (top (e) Perform this test for at least one
+bottom) and “background signal” as volume coil using coil T1 and T2
(left + right). Record all results. sequences during annual inspections
and all volume coils during acceptance.
(e) Calculate SNR:
e. Interpretation of Results. Volume and surface
SNR = Mean Signal / Noise Standard Deviation RF coil SNRs should match manufacturer’s
specifications and be consistent from one
(2) Calculate PIU annual inspection to the next. Volume coil
PIU should be equal to or greater than 87.5%
PIU = 100 x [1 - (High – Low) / (High + Low)] for 1.5 T systems and equal to or greater than
8-4
NMCPHC TM 6470.1
JUNE 2013
8-5
NMCPHC TM 6470.1
JUNE 2013
9. Low Contrast Detectability (LCD) each slice of the T1 weighted ACR axial
series.
a. Purpose. To assess the extent to which low
contrast objects are discernible in the images. (2) Check to make sure that the phantom
appears circular, there are no ghost images
b. Regulations. None in the background or overlying the
phantom, there are no streaks or artifact
c. Equipment. ACR MRI Phantom bright or dark spots visible and there are
no unusual or new features in the image.
d. Procedure. (3) Repeat for T2 weighted series, if desired.
(2) Count the number of complete spokes in c. Equipment. Precision luminance meter
each slice. The number of complete
spokes counted is the score for this slice. d. Procedure.
The ACR recommends choosing a single
slice in which all of the spokes are not Without full implementation of the DICOM
normally observable as the slice to score. Part 14 standard, the following limited set of
However, for an annual check it is easy to tests is recommended:
score all the slices and add them together.
(1) Luminance Maximum and Minimum.
(3) Repeat for T2 weighted series, if desired. Display a uniform image. Adjust
window/level to the brightest available
e. Interpretation of Results. Low contrast level. Record this value as “maximum”.
detectability should be equal to or greater than Adjust window/level to display all black
or equal to 9 spokes (total) for 1.5 T systems (lowest available). Record this value as
and equal to or greater than 37 spokes (total) “minimum”.
for 3 T units.
(2) Luminance Uniformity. Display a uniform
10. Image Artifact Assessment image. Measure the luminance at the
center and each of the four corners of the
a. Purpose. To ensure that overall MRI system image display. Record these values.
performance is remaining stable. Calculate the average value of the four
corners. Record this value. Calculate the
b. Regulations. None percentage difference of the luminance
values measured in the image display area
c. Equipment. ACR MRI phantom as follows:
(1) Adjust the display window and level to (3) Resolution. Display a SMPTE test pattern
show the full range of pixel values for with a magnification factor that is
appropriate for the monitor being tested.
Evaluate the high resolution patterns in the
8-6
NMCPHC TM 6470.1
JUNE 2013
image center and in each of the corners. 90% gray level patches. Record these
(SMPTE patterns vary depending on values
design, so be careful to use one with e. Interpretation of Results. Action levels should
appropriate resolution and patterns). be established by the quality control program.
If the action levels are exceeded, double check
(4) Spatial Accuracy. A SMPTE test pattern phantom positioning before seeking other
that displays a rectangular grid is typically action.
displayed with a magnification factor that
allows it to fill the entire screen. Display a 13. B0 Field Room Survey
similar grid pattern over the screen.
Compare the displayed SMPTE pattern to a. Purpose. To map out the magnetic field inside
the overlaid grid. the magnet room and in adjacent areas to
ensure patient and staff safety.
e. Interpretation of Results. Luminance
uniformity percentage difference should be b. Regulation. None
less than or equal to 5%. Verify that the 5%
patch can be distinguished in the 0 / 5% patch, c. Equipment. Gauss meter
the 95% patch can be distinguished in the
95 / 100% patch and that all gray level steps d. Procedure.
around the ring of gray levels are distinct from
adjacent steps. Maximum luminance, (1) Using a calibrated Gauss meter perform a
minimum luminance and luminance survey inside the magnet room. Record
uniformity should not vary by more than 15% these measurements using a room diagram.
from levels measured at acceptance.
(2) Take additional readings in areas directly
12. Hard Copy Image Quality Control adjacent to the magnet room and record
these values as well.
a. Purpose. To assess the stability of the Hard
copy imager e. Interpretation of Results. The 5 Gauss line should
be marked by signs to warn people (e.g. patients,
b. Regulations. None staff and visitors) with pacemakers and other
restrictive devices to avoid being in the area. In most
c. Equipment. Densitometer modern MRI suites, the 5 Gauss line lies within the
scanner room.
d. Procedure.
14. Physical and Mechanical Inspection
Necessary if the facility uses film for primary
interpretation. a. Purpose. To assess the physical and
mechanical integrity of the MRI unit and
(1) Display the SMPTE pattern and visually associated equipment.
confirm that the gray levels on the display
are subjectively correct. Verify that the b. Regulations. None
0% patch can be distinguished in the
0 / 5% patch, the 95% patch can be c. Equipment. None
distinguished in the 95 / 100% patch and
that all gray level steps around the ring of d. Procedure.
gray levels are distinct from adjacent
steps. (1) Visually inspect the MRI magnet for
obvious defects, damage or other safety
(2) Print the SMPTE pattern using a 6-on-1 hazards. Visually inspect the cryogen
format and capture the same image on all venting system to ensure there are no
six frames. obvious impediments to its proper
function.
(3) Using a film densitometer, measure the
optical density of the 0%, 10%, 40%, and
8-7
NMCPHC TM 6470.1
JUNE 2013
(2) Visually inspect the magnet equipment d. Procedure. Test fixed and portable (e.g. fixed
room for general and missile effect safety frame, stand alone and wand) detector systems
hazards. Inspect ancillary equipment and for proper operation. Record any deficient
supplies in the magnet room and adjacent devices.
spaces for MR compatibility (items should
be appropriately labeled as MRI safe or e. Interpretation of Results. The detector should
MRI conditional at the relevant magnetic operate according to the manufacturer’s
field strength). specifications.
(3) Visually inspect the magnet equipment 17. ACR MRI Safety Recommendations
room for electrical, trip and falling object Compliance
hazards.
a. Purpose. To evaluate facility compliance with
e. Interpretation of Results. MRI equipment ACR MRI safety zone and procedure
should be physically and mechanically stable, recommendations.
showing no obvious mechanical or electrical
conditions that suggest patient or operator b. Regulation. None
safety concerns. All ancillary equipment (e.g.
surface coils, phantoms, power injectors, c. Equipment. Gauss meter
anesthesia machines, etc should be in good
working order and compatible for the unit d. Procedure.
strength being used. Report violations to staff
for corrective action (some items may require (1) Walk about the MRI spaces to determine
immediate removal from the magnet room to whether the layout meets the ACR’s most
prevent missile hazards). recent safety zone scheme. Does the
arrangement make sense and are its
15. Patient Monitoring Evaluation instructions and intent obvious to an
untrained member of the general public?
a. Purpose. To ensure that the facility maintains Are the various zones posted correctly?
appropriate patient / unit operator
communication capability. (2) Review the facility’s MRI policy and
procedures manual to determine if it is
b. Regulation. None operating in conformance with the ACR’s
safety guidelines with respect to: safety
c. Equipment. None zones, patient screening, public access,
contrast administration, patient consent,
d. Procedure. Test patient communication fire and code blue safety.
devices (e.g. intercom and hand held operator
notification devices) for proper operation. (3) Review MRI staff and non-staff MRI
Record any deficient devices. safety training records.
c. Equipment. None
8-8
NMCPHC TM 6470.1
JUNE 2013
8-9
THIS PAGE IS LEFT INTENTIONALLY BLANK
8-10
Magnetic Resonance Imaging Unit Survey NMCPHC TM 6470.1
JUNE 2013
Facility: Date:
Manufacturer:
Test Performed Pass Fail N/A Comments (failure comments must annotate minor or significant finding)
Central Frequency
Evaluation
Magnetic Field
Homogeneity
ACR Phantom Image
Acquisition
Slice Thickness
Accuracy
RF Coil Checks
Geometric Accuracy
Spatial Resolution
Low Contrast
Image Artifact
Purpose: Results:
Surveyor Name:
Surveyor Signature:
8-12
NMCPHC TM 6470.1
JUNE 2013
Chapter 9
A. Introduction
The modern gamma camera is a highly sophisticated implementation of computer chips, and a better
imaging system but its underlying principle of understanding of the underlying problems and the
operation is still the same as the original Anger procedures in the reports still test the end result of
camera. The evolution of the gamma camera has lead these improvements.
to improved performance by optimizing the various
sub systems as new technology was implemented. B. Minimum Required Personnel
Today the most of the imaging degrading problems Qualifications
associated with the original camera have been either
Advanced Diagnostic Imaging Equipment (NMI)
engineered out or sophisticated methods have been
devised to correct data prior to generating the image. C. Testing performances
As the gamma camera evolved so did the methods of The testing required of a system will be dependent on
collecting and displaying images. Today the gamma how the system will be used -planar imaging, Single
camera can provide cross sectional images of patient Photon Emission Computed Tomography (SPECT)
anatomy as well as general planar images. The cross imaging or both -and whether it is a single head
sectional imaging places far greater demands on the system or a multi-head system. Multi-headed system
camera performance then does planar imaging and used for both planar and SPECT imaging will require
subsequently increase demands on quality control of substantially more testing and more time than a
the systems. single headed planar system. Needed documentation:
Nuclear Medicine Camera Survey Form.
To ensure the best possible imaging, all gamma
camera systems should be tested when first purchased D. References
to establish that the system meets its claimed
performance and to collect data to be used to verify 1. AAPM Report No. 6, “Scintillation Camera
continued system performance over the life of the Acceptance Testing and Performance
system. The tests provided in this document are based Evaluation,” 1980
on several reports prepared by the AAPM. Although 2. AAPM Report No.9, “Computer-Aided
the documents are somewhat dated, the fact that the Scintillation Camera Acceptance Testing,” 1982
gamma camera's underlying principle of operation 3. AAPM Report No. 22, “Rotating Scintillation
and its general design have not changed much over Camera SPECT Acceptance Testing and Quality
time, makes these reports still relevant. The Control,” 1987
performance gains have been achieved by improved 4. AAPM ReportNo.52, “Quantitation of SPECT
engineering of components, Performance,” 1995
5. National Electrical Manufacturers Association
Standard NU 1,”Performance Measurement of
Scintillation Cameras,” 1994
9-1
NMCPHC TM 6470.1
JUNE 2013
d. Verify the alignment of the system as (2) Select a 20% energy window width (or use
appropriate. If the system has indicator noting the manufacturers recommended window
when it is horizontal or perpendicular check width for measuring 99mTc.
these with a level. If the system indicates
angles of the head check this using a level and (3) Ensure that the background radiation level
protractor. is less than 500 counts per second.
e. The gamma camera head should be (4) Make two 99mTc sources (#1 and #2)
appropriately shielded to prevent interference each with between 2 and 7 mCi activity.
from extraneous radiation. This should be Place the sources in container with at least
checked the 111g;hest Plnp>'(nl radioisotope 6 mm lead shielding on sides and bottom.
used in the area of the gamma camera. The Place at least 6 mm of Cu on top of the
source (5 to 10 millicuries for 99mTc) should container to absorb scatter from the
be slowly moved around the back and sides of container.
the head at about one to two feet and the count
rate noted. Excessive background will reduce
the contrast of the images acquired with the
camera. If the manufacturer quotes a shielding
performance ensure the system meets the
standard. Repeat using a high energy source
(67Ga or 131 1) is such sources are used in the
9-2
NMCPHC TM 6470.1
JUNE 2013
(5) Place the source at a distance that gives a scatter and the NEMA method does not.
count rate of approximately 30,000 cps or Therefore the NEMA method gives a
if the dead time is quoted by the larger τ (~1.3x).
manufacturer use the following equation to
calculate the count rate: b. Maximum Intrinsic Count Rate. This
measurement is dependent on the dead time of
0.1 the system and is the maximum count rate the
1.2 system will record. At this level the count rate
will not increase with increasing activity and
Ensure that the entire crystal is being may in fact decline.
irradiated by the source.
(1) With the collimater removed and using the
(6) Using a preset time of 100 sec and a fixed energy window as the dead time test; place
time between each measurement measure approximately 4 mCi of 99mTc in
the sources in the following order. container shielded on the bottom and sides
by 6mm of lead and with 6mm Cu on top
(a) Measure source #1 and record counts. at a distance of 1.5 meters. Move the
source and or camera head until a
(b) Place source #2 next to source #1 and maximum count rate is observed. See
record counts. Figure 9-1.
Figure 9-1.
In the counts recorded at the camera head (2) Record the maximum count rate and
and the computer console are different; compare to any published manufacturers
calculate the dead time for each. A larger specification.
τ at the computer console indicates
additional count losses at the computer c. Extrinsic Temporal Resolution. The above
measured is for the non clinical situation
interface. This is rare but if it does occur where no collimator is installed on the camera
the difference should small. head. This test measures the dead time in a
more clinical situation and requires the use of
(8) Calculate the count rate at which the count a scatter phantom. If such a phantom is not
loss is 20% due to the dead time (τ). available then skip this test.
1 10 0.23311 (1) Select the most common collimator used
%
8 clinically (usually a general purpose or
high sensitivity collimator).
Compare this value with the
manufacturers specification. Note that this
method of measure dead time eliminates
9-3
NMCPHC TM 6470.1
JUNE 2013
(7) Calculate the observed counting rate that (2) Obtain a 99mTc point source (l cc or less)
will generate a 25% count loss. with an activity that will give a count rate
(~30,000 cps) that generates more than a
1 4 0.216 10% count loss for the energy window
% used clinically (15% or 20%) at a distance
3
of five times the UFOV diameter. Use the
During imaging, the recommended intrinsic dead time value previously
upper limit for count loss due to dead calculated to verify that the count loss
does not exceed 10% (see chart below).
time is 25%. Use of activities in Record the activity and time it was
patients in excess of that which gives measured.
this count rate merely increases the
dose to the patient with no gain in (3) Place the source centered over the crystal
image quality of imaging time. See at a distance of 5 UFOV diameters. Be
Figure 9-2. extremely careful not to drop the source on
the detector as this could damage the
crystal (a thin sheet of styrofoam placed
on the crystal face is good protection).
Record the count rate indicated on the
9-4
NMCPHC TM 6470.1
JUNE 2013
console and/or camera gantry. See (10) Calculate the “integral uniformity”
Figure 9-3. (maximum deviation of the counts in the
pixels as a percent) using
9-5
NMCPHC TM 6470.1
JUNE 2013
9-6
NMCPHC TM 6470.1
JUNE 2013
e. If the system has an automated linearity e. Place a vial containing 99mTc in the vial
calculation tool then this tool can be used to shield. Adjust the source activity to give no
evaluate the system but a visual check should more than 10,000 cps when place on the face
also be done. If the image shows significant of the uncollimated gamma camera head.
deviations on observation and the program
does not then further analysis should be done f. Adjust the system to collect a 128 x 128 by 16
to determine the reason for the discrepancy. bit image and place the vial at the center of the
detector. Collect an image for about 5 sec.
6. Point Source Sensitivity Note the time since all subsequent images will
have to be collected for the same period of
time for the results to be valid. Observe the
This test is performed to determine the
image to ensure there is no pixel overflow. If
uniformity of the sensitivity of the crystal at
so then reduce the time for image collection
various locations. Ideally, the crystal should
until no overflow is seen. Record the counts in
have the same sensitivity at all points.
the image.
a. Disable the uniformity correction circuitry of g. Using the time determined above repeat the
the system if allowed. measurement with the vial located at 25%,
50% and 75% of the radius in both the plus
b. A lead container at least 6 mm in thickness and minus direction of the X axis. Repeat the
with a 3 mm hole in the bottom will be measurement in the Y axis and then along the
required. This can be made from a lead vial 45° angle between the X and Y axes for a total
9-7
NMCPHC TM 6470.1
JUNE 2013
7. Spatial Resolution
20
FWHM
10%
To accurately measure intrinsic resolution
requires a slit mask made of 3 mm lead with a 1 0
0 20 40 60 80 100 120
mm wide slit extending for a distance equal to
the UFOV of the camera being tested. If such a Figure 9-7.
slit mask is not available, a more straight
forward but less accurate method is to use a line (5) Repeat this measurement along the X and
bar phantom that has a set of bars just below and Y axes at points 25%, 50%, and 75% of
just above the intrinsic resolution. Extrinsic the distance from the edge of the UFOV to
resolution can be measured by imaging a simple the center point for a total of 13
line source or a bar phantom. measurements.
a. Intrinsic Resolution (slit mask method). (6) Optionally repeat this measurement using
radioisotopes of interest elOTI etc.).
(1) Remove the collimator for the camera
head and orient the crystal facing up. b. Intrinsic Resolution (bar pattern method).
Select a 20% energy window center on the
140 keY peak of 99mTc and use the NOTE: For this test to be useful at least one
maximum matrix size available (at least of the bar patterns must be smaller
512 by 512 by 16 bits. than the intrinsic resolution of the
system (~3-4 mm).
(2) Place the slit mask along the X axis flush
against the detector and suspend a point (1) Remove the collimator for the camera
source approximately five crystal head and orient the crystal facing up.
diameters away from the mask. Select a 20% energy window center on the
140 keY peak of 99mTc and use the
(3) Collect an image with at least 10,000 maximum matrix size available (at least
counts in the central pixel of the image of 512 by 512 by 16 bits)
the slit.
(2) Mask the outer edge of the crystal
(4) Determine the FWHM from the image of physically or electronically to limit the
the slit by direct measurement using linear image to the UFOV of the system.
interpolation. This is most accurate if the
pixel is approximately 1/10th the FWHM. (3) Place a 99mTc point source at a distance
If this is the case then also estimate of 5 UFOV distances from the crystal
FWTM. See Figure 9-7. face.
9-8
NMCPHC TM 6470.1
JUNE 2013
(6) Repeat the above rotating the bar phantom a. The collimator will be removed from the
90° each time until each quadrant has been camera and the camera oriented facing up.
tested with the highest frequency (smallest
bar size). b. The source to be used is 67Ga (93 keV, 184
keV and 296 keV energy peaks). The source
(7) Observe the images to assess the will be placed in a vial shield (at least 6 mm
uniformity of resolution across all lead) with a 3 mm hole in the bottom. The
quadrants. activity will be adjusted to give no more than
10,000 cps for all energy windows being
c. Extrinsic Resolution. tested.
(1) Install a general purpose or the most c. The vial containing the source will be placed
frequently used collimator to the camera on the-X axis at 75% of the UFOV radius
head and orient the crystal facing up.
Select a 20% energy window center on the d. Use a zoom that will ensure that at least 10
140 keV peak of 99mTc and use the pixels are included in the FWHM of the
maximum matrix size available (at least image. An image for each of the energy
512 by 512 by 16 bits). windows will be acquired (minimum of 1000
counts in the central pixel).
(2) Fill two capillary tubes (inner diameter
less than 1mm and at least 5 cm long) with e. Move the source a known distance along the
high concentration 99mTc to give no more X axis and repeat the measurement
than 10,000 cps.
f. Determine the pixel size by measuring the
(3) Place the capillary tube along the X axis at number of pixels between the two peaks and
a distance of 5 cm from the face of the dividing this in to the known distance.
collimator.
g. For each of the images (one at each image)
(4) Acquire an image with at least 10,000 determine the location of the center of the
counts in the central pixel of the Line peak.
Spread Function (LSF) image.
h. Calculate the difference between the position
(5) Place the second capillary tube 5 cm away of the 93keV peak and each of the other
from the first and parallel to it and collect peaks.
a second image. Calculate the pixel size
by dividing 5 cm by the number of pixels
between the peaks of the two LSF images.
9-9
NMCPHC TM 6470.1
JUNE 2013
1. Rotational Uniformity and Sensitivity b. Suspend the line source over the end of the
imaging table so that it is over the detector
This will test the system’s sensitivity to stray without any interfering objects.
magnetic fields and to the thermal changes in the
detector heads. c. Adjust the table so that the line source is at the
center of rotation and then set the radius of
a. Prepare a flood source with 99mTc ensuring the rotation to 20 cm or the smallest that is
activity is uniformly disbursed throughout the achievable. Make sure the line source and the
phantom. Alternatively and preferably use a collimator face are parallel to the axis of
solid 57Co flood source with a uniformity of at rotation and level.
least +/-5% or better.
d. Peak the system on the photopeak energy
b. Secure the flood source to the face of the using a 20% window, select a 128x128 matrix
collimator (a low energy general purpose and 128 views over 360 degrees.
collimator) so that it does not move during
motion of the imaging head. e. A pixel size in the range of 3-3.5 mm is
desirable (a 16 inch crystal with a 128 matrix
c. Center a 20% energy window on the will give a pixel size of 3.2 mm). If necessary,
photopeak energy of the radioisotope being select a magnification mode that will deliver
used and collect a 5 million count flood in a this pixel size or smaller.
64x64 matrix with the detector at 0 degree
position. Record the time necessary to acquire f. Acquire the tomographic study using a time
the 5 million counts. that will deliver at least 100k counts in the
first projection image.
d. Repeat the flood image at 90, 180, 270 and
360 degrees, using the time recorded for the g. Reconstruct the data using a ramp filter into
first flood image. 10 mm transverse sections, starting at the top
of the line. Uniformity and attenuation
e. Calculate the maximum sensitivity variation: correction do not need to be applied. Consult
the vendor to determine the appropriate filter
to use if in doubt.
% 100
h. Using the first section, draw a one pixel wide
In the absence of vendor performance profile in the X direction through the hottest
criteria, the sensitivity variation should not pixel in the reconstructed image of the line
exceed 1%. source. From the profile, calculate the FWHM
by linear interpolation.
f. Using the system’s computer, subtract the
90 degree flood image from the 270 degree i. Repeat the above for the Y direction and then
image. Review each of the subtracted images for a slice at the bottom of the stack and in the
for indications of structure noise and other middle.
non random noise. Save the images for
reference. j. Acquire a planar image of the line source
containing 500k counts with the collimator 20
2. Spatial Resolution cm from the line source.
This test will use a line source which is at least 30 cm k. At approximately the same positions as the
in length and no more than 2 mm inside diameter. three tomographic resolution results were
These can be made from catheters. determined, use a ~10 mm(3 pixel) wide count
profile through the planar image of the line
a. Fill the line source with 99mTc with a source to calculate FWHM.
concentration of 1 mCi/cm3. If the line source
is not rigid, then attach it to a rigid support l. Record the ratio of the SPECT/Planar FWHM
which will not generate catter (tape the to the three locations. These should not exceed
catheter to a piece of cardboard or Styrofoam) 1.1. If so, then the Center of Rotation (COR)
may be incorrect. Collect a new COR and
9-10
NMCPHC TM 6470.1
JUNE 2013
repeat the test. Large errors can be due to the (5) Use a magnification factor that will give a
X or Y axis’ being out of alignment with pixel size in the range of 6 to 7 mm if
reference to the line source. needed.
m. In the absence of vendor performance criteria, (6) Acquire the tomographic study using a
the ratio should not exceed 10% and must not time sufficient to give at least 500k counts
exceed 15%. in the first projection image.
9-11
NMCPHC TM 6470.1
JUNE 2013
(4) Repeat the above using the data set (4) If there are not vendor provided
corrected for uniformity. performance criteria then use the
following values as a guide. The data in
(5) In the absence of vendor provided Table 9-1 is from a Jaszczak phantom.
performance criteria, the following values
are considered acceptable: Table 9-1.—Data from Jaszczak Phantom.
9-12
Nuclear Medicine Camera Survey NMCPHC TM 6470.1
JUNE 2013
Facility: Date:
Manufacturer:
Test Performed Pass Fail N/A Comments (failure comments must annotate minor or significant finding)
Uniformity
System Resolution
Multi-Energy
Registration
Pixel Size
Count Rate
System Sensitivity
System Linearity
Center of Rotation
SPECT only
Slice Uniformity
SPECT only
Slice Thickness
SPECT only
In Place Resolution
SPECT only
Image Contract
SPECT only
Additional Comments:
Purpose: Results:
Surveyor Name:
Surveyor Signature:
9-14
NMCPHC TM 6470.1
JUNE 2013
Chapter 10
10-1
NMCPHC TM 6470.1
JUNE 2013
CT-simulator: Annually, monthly, and daily QA tests 2. AAPM TG-40, “Comprehensive QA for
per TG-40 and TG-66, upon acceptance and after Radiation Oncology: Report of AAPM Radiation
major repairs per TG-66. Therapy Committee Task Group 40,” Med. Phys.
21(4)(1994).
D. Equipment 3. AAPM TG-45, “AAPM Code of Practice for
Radiotherapy Accelerators: Report of AAPM
LINACs: Phantom (solid for monthly or distilled Radiation Therapy Task Group No. 45,” Med.
water for annual/monthly QA tests), MapCheck (if Phys. 21(7)(1994).
applicable), ionization chamber, electrometer, QA 4. AAPM TG-51, “Protocol for Clinical Reference
mechanical device (if applicable), other equipment as Dosimetry of High-Energy Photon and Electron
specified by the physicist per TG-142. Beams: Report of AAPM Radiation Therapy
Committee Task Group 51,” Med. Phys.
CT simulator: Laser alignment tool, head/body water 26(9)(1999).
phantom, radiographic/radiochromic film (when 5. AAPM TG-66, “Quality assurance for
available), laser QA device, other equipment as computed-tomography simulators and the
specified by the physicist per TG-66. computed-tomography-simulation process:
Report of the AAPM Radiation Therapy
Tomotherapy: Phantoms (solid/liquid water) for Committee Task Group No. 66,” Med. Phys.
weekly or monthly quality assurance (QA), 30(10)(2003).
radiochromic film (if available) or MapCheck for 6. AAPM TG-142, “Quality Assurance of Medical
dose Delivery Quality Assurance (DQA) – i.e. IMRT Accelerators,” Med. Phys. 36(9)(2009).
QA prior to treatment, specified by the physicist per 7. AAPM TG-148, “QA for helical Tomotherapy:
TG-148. Report of the AAPM Task Group 148,” Med.
Phys. 37(9)(2010).
8. NCRP Report No. 151, Structural Shielding
E. References Design and Evaluation for Megavoltage X-and
Gamma-Ray Radiotherapy Facilities, National
1. AAPM TG-21, “A Protocol for absorbed dose Commission on Radiation Protection and
from high-energy beams,” Med. Phys. Measurements, 2006
10(6)(1983).
10-2
NMCPHC TM 6470.1
JUNE 2013
Chapter 11
1. It is highly recommendation that performance Remote afterloader: Upon acceptance and after major
and safety evaluation procedures provided by repairs, annual QA, quarterly QA, and daily QA per
the equipment manufacturer and/or by AAPM TG-59.
nationally-recognized protocols (e.g. Nuclear
Regulatory Commission) are followed when D. Equipment
applicable, such as TG) reports published by
the AAPM) as well as reports published by Radiochromic (Gafchromic) film, applicators, source
the NCRP, ICRP, and other nationally- guide tubes, and radioactive source
recognized organizations.
E. References
2. Local policies (e.g. the applicable Navy
Radioactive Materials Permit) and procedures 1. AAPM TG-41, “Remote Afterloading
should specify facility-specific requirements Technology: Report of AAPM Task Group No.
and periodicities, or at a minimum, indicate 41.”
the use of specific nationally-recognized 2. AAPM TG-59, “High dose-rate brachytherapy
protocols. treatment delivery: Report of AAPM Radiation
Therapy Committee Task Group No. 59,” Med.
3. A Medical Physicist specialized in Radiation Phys. 25(4)(1998).
Therapy should determine which acceptance 3. NCRP Report No. 151, Structural Shielding
Design and Evaluation for Megavoltage X-and
Gamma-Ray Radiotherapy Facilities, National
Commission on Radiation Protection and
Measurements, 2006.
11-1
THIS PAGE INTENTIONALLY LEFT BLANK
11-2
NMCPHC TM 6470.1
JUNE 2013
Chapter 12
A. Introduction E. References
One of the largest contributors to total population 1. AAPM Report 31. “Standardized Methods
radiation exposure from man-made radiation sources for Measuring Diagnostic X-Ray Exposures,”
is from diagnostic (dental and medical) radiography. 1990.
One of the goals of the Conference of Radiation 2. “Average Patient Exposure/Dose Guides. A
Control Program Directors (CRCPD) is to reduce the report by Committee n Quality Assurance in
unnecessary component of dental and medical x-ray Diagnostic Radiology (H-7).” Conference of
exposure to a level As Low As Reasonably Radiation Control Program Directors, Inc.
Achievable (ALARA). CRCPD Publication 92-4, 1992.
3. “Nationwide Evaluation of X-Ray Trends
Reference 3, lists recommended patient exposure (NEXT),” Conference of Radiation Control
guides which reflect “State of current practice” in a Program Directors, Inc. Frankfort: CRCPD.
cross section of radiography facilities across the 1974 - 1994.
United States. They are to be used as a tool for 4. AAPM Report 70. Cardiac Catheterization
reducing unnecessary radiation exposure to patients, Equipment Performance, 2001.
while maintaining or improving image quality.
Calculated Entrance Skin KERMA (ESK) should be F. Entrance Skin Kinetic Energy Released
compared to these average values. in Material (KERMA) (ESK) for
Exposures that significantly exceed the levels General Radiographic Equipment
indicated in the guides for routine examinations are
likely to represent unnecessary patient doses and 1. Purpose
causes for such excessive exposure should be
investigated. A reasonable but arbitrary range of To ensure that ESK also formally known as entrance
acceptability is 20% of a guide value. skin exposure for standard radiographic techniques is
within standards.
B. Minimum Required Personnel
Qualifications 2. Regulations
Joint Commission for Accreditation of Healthcare
Basic Diagnostic Imaging Equipment
Organizations (JCAHO, 1994) requires ESK's, for
techniques used most commonly, be evaluated on an
C. Testing Periodicity annual basis.
Acceptance, annual, or after major repair
3. Procedure
D. Equipment a. Manual Mode.
1. Ion chamber (1) In addition to the procedures outlined
2. Patient-Equivalent Phantoms below, the following parameters must be
3. Tape measure known for each tube head to obtain
specific organ doses:
12-1
NMCPHC TM 6470.1
JUNE 2013
(3) Collimate the light field on the ion (9) To determine tissue/organ doses for
chamber using narrow beam geometry. projections common in diagnostic
radiology, use the ESK for each
(4) Set the desired technique factors at the projection, the additional information in
control panel. item 3.1 above and refer to HHS
Publication (FDA) 89-8031.
(5) Place the ion chamber in the center of the
useful beam, measure, and record SSD, (10) To estimate the dose to the embryo-fetus
SID, and the ESK. from radiographic examinations, refer to
reference HHS Publication (FDA) 79-8079
(6) Repeat as necessary for other commonly and NCRP Report 54.
used projections and technique factors.
b. Automatic Exposure Control Mode.
(7) Calculate the entrance skin KERMA for
each projection using the patient thickness (1) Measure and record the HVL of the x-ray
guidelines in Table 2 of HHS Publication beam.
(FDA) 89-8031 and the inverse square
law. The inverse square calculation is as (2) Set up x-ray unit for commonly used
follows: radiographic projections: chest, abdomen,
and extremity.
2
x
ESK (mGy) 1 AirKERMA (3) Place the patient-equivalent phantom over
the selected AEC detectors. (see figures
x2 12-1a and 12-1b, shown for a chest unit).
Where:
(4) Place the ion chamber in the useful beam,
ESK = corrected x-ray beam intensity at the skin measure, and record the SSD, SID, and
entrance without backscatter ESK values. Ensure that the detector does
x1 = focal spot to center of ion chamber distance not cover the AEC sensor.
x2 = distance from the focal spot to the surface of the
skin (x2 = x1 - patient thickness) (5) Follow the steps in 3.(a)(7) through
Air KERMA = Air KERMA at the center of the ion 3.(b)(4) above to calculate ESK, interpret
chamber results and to estimate organ dose/embryo-
fetal doses.
NOTE: Exposure in air (mR) may
be converted to ESK
(mGy) by multiplying by
0.00876 mGy/mR.
12-2
NMCPHC TM 6470.1
JUNE 2013
G. Entrance Skin Kinetic Energy Released I. Digital Spot Entrance Skin Kinetic
in Material (KERMA) (ESK) for Dental Energy Released in Material (KERMA)
Intraoral Units Rate (ESKR)
1. Procedure 1. Introduction
a. Place the probe about ½-inch from end of a. Fluoroscopy is routinely used as a localization
cone. mechanism for radiographic images that are
analyzed at a later time. In many fluoroscopic
b. Use a technique commonly used on the dental examinations, the digital spot exposure
unit to make an exposure. This is the ESK component can be substantial, especially if the
which should be recorded along with all use of contrast is involved. Therefore,
settings used. accurate digital spot ESKR measurements are
essential to maintaining a database for
c. Record focal spot to chamber distance which determining patient exposures.
represents the source to skin distance and
approximates the source to image distance. b. Digital spot exposure measurement assumes
Use the actual kVp as determined. proper generator calibration and satisfactory
operation of the imaging chain components.
H. Entrance Skin Kinetic Energy Released Image intensifier entrance exposure rate
in Material (KERMA) Rate (ESKR) for (Rfr-1) must be properly set to the
manufacturer’s recommendation. Digital spot
Fluoroscopy Units exposure testing is typically performed last
during an acceptance inspection or annual
1. Procedure performance evaluation.
Refer to Chapter 4. for Entrance Skin KERMA
2. Procedure
Rate (ESKR) for fluoroscopic measurement
procedures.
a. Arrange the fluoroscopy unit and ion chamber
in the configuration appropriate to the
NOTE: 1100 alloy Al sheets are less suitable
machine type; i.e. undertable tube, overtable
as a fluoroscopy ESKR phantom than
tube, or c-arm. Ensure that if a grid is used in
acrylic slabs since the Al represents
clinical studies, it is in the beam path during
significantly different equivalent
testing.
patient thicknesses at different kVp
values. Al also does not provide the
b. Place a 4 cm Al or 15 cm acrylic phantom in
same amount of scatter as the thicker
the beam in the same manner as for measuring
acrylic block. Experimental data has
ESKR. Ensure that the phantom sits between
demonstrated up to 2X higher ESKR
the ion chamber and the image intensifier
using acrylic. However, not all
tube.
evaluators will have acrylic phantoms
in their equipment inventories. To
c. If the unit provides specific digital spot
maintain consistency, clearly identify
settings for different anatomical applications,
the type of phantom used, record the
program the system for non-contrast
testing conditions, and perform
abdominal studies. If different dose levels are
subsequent evaluations using the
also provided, select an average setting. If
original phantom and conditions.
anatomical or dose programming are not
provided, use the system’s ABC) to determine
the kVp to be used during testing. For manual
only systems, program the unit for 80 kVp.
Select an appropriate medium level current
(e.g. 200 mA). Program a digital spot film
system to operate at its minimum frame rate (1
frs-1 is most desirable). If appropriate, set the
12-3
NMCPHC TM 6470.1
JUNE 2013
3. Interpretation of Results
a. Calculate ESKR as a function of mAs.
Determine maximum, minimum, and average
exposure/mAs. If current values differ from
their acceptance or historical counterparts by
more than 10 %, refer the system for
adjustment by a qualified service engineer.
12-4
NMCPHC TM 6470.1
JUNE 2013
Chapter 13
A. Introduction
On a case by case basis, NMCPHC may evaluate
Shielding design for medical facilities must
individual shielding design experience as an
protect workers and members of the public from
exemption to board certification. Shielding
excessive exposure to ionizing radiation from
experience in these cases should include a minimum
radiographic and nuclear imaging, and radiation
of 2 designs of the applicable type(s): linear
therapy.. In Title 10, Code of Federal Regulations
accelerators, HDR therapy units, nuclear medicine
Part 20, the United States Nuclear Regulatory
suites, computed tomography, fluoroscopy, and
Commission requires that the effective dose does not
radiographic units.
exceed 1mSv (100 mrem) per year for a member of
the general public and 50 mSv (5 rem) per year for
Designs used to demonstrate competency may be
radiation workers from radioactive materials.
example shielding problems approved and reviewed
by any individual who has been granted a shielding
The National Council on Radiation Protection
design qualification from the MPAB.
and Measurements has established similar
recommendations for ionizing radiation exposure
from other sources of ionizing radiation including x- C. Design and Evaluation Periodicity
rays. The concept of ALARA mandates that the
exposure be kept below these limits. Thus when Shielding design is required for all new
designing shielding, the target limit should be lower radiographic, nuclear medicine, therapy installations,
to ensure that the regulatory limits are not likely to be and after major renovations to existing facilities. As
exceeded. recommended in NCRP Report No. 147, sites
designed using the previous NCRP Report No. 49
The current recommended design criteria is 0.02 standard do not need to be upgraded as long as the
mSv (2 mrem) per week for members of the general room has not changed in functionality.
public and 0.1 mSv (10 mrem) per week for radiation
workers. It is the responsibility of the shielding D. Equipment
designer to keep abreast of changes in these limits. 3
Magnetic Resonance Imaging facilities will also 1. Large volume ion chamber (≥ 180 cm
require shielding designs that reduce magnetic and nominal volume) to check radiation levels
radiofrequency interference. The design of these
shields are typically beyond the capabilities of site E. References
personnel and will generally be included as part of
the purchase contract for such units. 1. United States Nuclear Regulatory Commission.
Standards for Protection Against Radiation.
B. Minimum Required Personnel Washington D.C.: Chapter 10 Part 20;2007.
2. National Council on Radiation Protection and
Qualifications
Measurements. Structural Shielding Design for
Medical X-Ray Imaging Facilities. Bethesda,
Board certification from ABR in the appropriate
MD: NCRP; Report No. 147; 2004.
subspecialty or American Board of Health Physicists
3. American Association of Physicists in Medicine.
(ABHP) is the normal pathway for obtaining the
Task Group Report 108: Shielding for PET/CT
Advanced Radiological Systems Shielding Design
Facilities. 2006
qualification. This is in line with the NCRP 147 and
NCRP 151 recommendation that only “Qualified 4. National Council on Radiation Protection and
Experts” should perform shielding designs. Measurements. Structural Shielding Design for
Megavoltage Radiotherapy Facilities. Bethesda,
MD: NCRP; Report No. 151; 2005.
13-1
NMCPHC TM 6470.1
JUNE 2013
13-2
NMCPHC TM 6470.1
JUNE 2013
Chapter 14
A. Introduction
Display testing in the form of acceptance testing software used to perform quality control tests. Prior
or frequent QC provides a means by which the user to the annual evaluation of the monitor the ambient
can be assured that the display quality is adequate lighting level should be measured and documented
and is maintained throughout the useful life of the and a verification of the DICOM luminance
device. It will also determine when a display device calibration should be done (to include verifying the
should be decommissioned before diagnosis is Lmax and Lmin). Most monitors have software installed
adversely affected. Furthermore, a comprehensives to perform the verification of the DICOM luminance
display. calibration.
The purpose of this section is to provide b. Equipment. AAPM TG18-QC test pattern
guidance for visually testing display systems. For
more comprehensive quantitative or advanced c. Procedure.
testing, refer to the primary reference, AAPM TG 18.
(1) The pattern should be maximized to fill
This section does not cover performance testing the entire display area. For displays with
for mammography monitors. Refer to the rectangular display areas, the patterns
manufacturer’s manual for specific testing should cover at least one orientation of the
requirements for these monitors. display area and be placed at the center of
the area used for image viewing.
B. Testing Periodicity
(2) The linearity of the pattern should be
Acceptance, annually, and after major repairs checked visually at a viewing distance of
30 cm across the display area and at the
C. Reference edges.
14-1
NMCPHC TM 6470.1
JUNE 2013
2. Display Reflection
a. Purpose. Assess and minimize display contrast patterns are seen such as an
reflection. identification badge on a white shirt or a
picture frame on a light wall, then the ambient
b. Description. Reflections can have two general illumination in the room should be reduced.
forms, specular and diffuse. Specular The threshold of visibility for low-contrast
reflection is said to occur when the angle of patterns in the TG18-AD test pattern should
the incident light rays equals that of the not be different when viewed in total darkness
emerging rays as dictated by geometrical and when viewed in ambient lighting
optics. Such a reflection produces a virtual condition. If the ambient lighting renders the
image of the source as would a mirror. In “dark-threshold” not observable, the ambient
diffuse reflection, the light is randomly illuminance on the display surface is causing
scattered out of the specular direction and no excess contrast reduction and the room
virtual image of the source is produced. There ambient lighting needs to be reduced.
are two types of diffuse reflection. One is
where the scattering angles of the emergent 3. Luminance Response
light are broadly distributed and poorly
correlated with the angle of the incident light a. Purpose. To ensure the luminance of the
such as with a Lambertian reflector where the display device is consistent and uniform.
direction of the incident light has little affect
on the observed reflected luminance. The b. Equipment. AAPM TG18-CT and TG18-MP
other type of diffuse reflection is where light test patterns
is randomly scattered into a narrow
distribution of angles in the vicinity of the c. Procedure.
specular direction.
(1) The TG18-CT pattern should be evaluated
c. Equipment. AAPM TG18-AD test pattern for visibility of the central half-moon
targets and the four low-contrast objects at
d. Procedure. the corners of each of the 16 difference
luminance regions at a viewing distance of
(1) Turn the display device off or put in power 30 cm.
save mode. Ensure the ambient lighting in
the room is the same as when the (2) The TG18-MP test pattern should be
radiologist is interpreting images. viewed to evaluate the bit-depth resolution
of the display. The evaluation includes
(2) Examine the display’s faceplate at a ascertaining the horizontal contouring
distance of 30-60 cm with an angular bands, their relative locations, and
view of +/-15 degrees for specularly grayscale reversals at a viewing distance
reflected light sources or illuminated of 30 cm.
objects.
d. Interpretation of Results.
(3) Load the TG18-AD test pattern and view
the low contrast patterns with normal (1) The appearance of the TG18-CT test
ambient lighting. pattern should clearly demonstrate the low
contrast target in each of the 16 regions.
(4) Turn off all lights and view the test pattern Since this pattern is viewed in one state of
in total darkness. visual adaptation, it is expected that the
contrast transfer will be better at the
e. Interpretation of Results. In examining the brightness for which the visual system is
display’s faceplate under normal ambient light adapted as opposed to the darkest or the
condition, no specularly reflected patterns of brightest regions. A common failure is not
high contrast objects should be seen. If light to be able to see the targets in one or two
sources such as that from a film illuminator or of the dark regions.
window are seen, the position of the display
device in the room is not appropriate. If high
14-2
NMCPHC TM 6470.1
JUNE 2013
14-3
NMCPHC TM 6470.1
JUNE 2013
c. Procedure.
14-4
NMCPHC TM 6470.1
JUNE 2013
Chapter 15
The Environment of Care standards set forth by the Properly cared for lead aprons have a life expectancy
JCAHO require performance inspections on medical of approximately 10 years. They should be properly
equipment. Lead aprons are considered medical hung on hooks/racks when not in use. They should
equipment. In addition, all state radiation control never be folded, creased or draped across another
programs require evaluations of lead aprons. In piece of equipment. They should be kept clean of
recent years, references (a) and (b) have addressed dirt, grease and other contaminants.
the topic of the proper care, evaluation and disposal
of lead aprons.
3. Evaluation of Lead Aprons
B. Guidance Pursuant to references (a) and (b), lead aprons should
be evaluated at least annually. This evaluation
The Bureau of Medicine and Surgery (BUMED) should consist of a visual inspection and physical
guidance for the use, care, evaluation and disposal of evaluation. If a suspicious area is found, i.e. tears,
lead aprons (including thyroid and gonadal shields) perforations, or thinning creases at the time of the
utilized in medical facilities is as follows: annual check or on any other occasion, an x-ray
evaluation should be performed. Depending on the
availability of the facility, the x-ray evaluation may
C. References be done with either fluoroscopy or x-rays. A record
of the date of the check, the type of check (visual,
1. Implementation of an X-ray Radiation Protective tactile, and/or x-ray), the results, and who performed
Equipment Inspection. the check, should be kept for three years. It is
2. Program, Operational Radiation Safety; 82(2): expected that an inspector will be able to verify that
S51-S53, 2002. the evaluations have been done by either looking at
3. Inspection of Lead Aprons: A Practical Rejection the aprons and/or the documentation system.
Model,
4. Operational Radiation Safety; 95(Supplement 2):
S133-S136, 2008. 4. Rejection Criteria
15-1
NMCPHC TM 6470.1
JUNE 2013
Length of Defect
0.25 0.50
Double Layered Whole Body 13.5 cm 27 cm
Gonads 4.4 cm 8.7 cm
Single Layered Whole Body 5.9 cm 5.4 cm
Gonads 1.9 cm 1.7 cm
Thyroid 1.9 cm 1.8 cm
15-2
NMCPHC TM 6470.1
JUNE 2013
Chapter 16
A. Introduction E. Reference
Digital Radiography (DR) consists of specialized 1. KCARE DDR Commissioning and Annual QA
equipment using Photoconductors (Direct DR or Protocol Manual, January 2005. 2. The Essential
DDR) to replace the cassette and cassette tray Physics of Medical Imaging. Bushberg, Jerrold,
altogether with a specialized digital detector system. Anthony Seibert, Edwin Leidholdt, John Boone.
The photoconductor material – Selenium – is placed Philadelphia, 2002.
on top of a Thin Film Transistor (TFT) array. The
detector captures ionizations and measures the charge F. General Requirements for
deposited directly. The data from the detector are Radiographic Equipment
directly digitized and sent out as a picture to the
PACS.
1. Dosimetry
This document describes a series of tests to assess
a. Purpose. To measure entrance detector
digital detector performance. The tests are intended
exposures.
to detect artifacts and test image quality and
sensitivity. All images should be acquired with a
b. Regulations. Measures dose required for later
minimal amount of pre-processing and a linear Look
tests 3, 4, 5, 6, 10.
Up Table (LUT) applied, unless otherwise stated.
Some of the images should be assessed on either a
a. Equipment. Ionization chamber, 1.0 mm Cu
reporting quality monitor or printer. These devices
filter.
should be monitored as part of a QA program.
Ambient lighting conditions should be consistent and
b. Procedure. Set a Focus to Detector Distance
appropriate.
(FDD) of at least 150 cm (see Figure 16-1).
Record this distance. Position a chamber at
B. Minimum Required Personnel least 30 cm in front of the detector (away from
Qualifications backscatter). If possible either point the tube
away from the detector or shield the detector
Basic Diagnostic Imaging Equipment with a lead apron. Collimate to the chamber.
Expose the chamber such that the inverse
C. Testing Periodicity square law corrected receptor entrance air
KERMA is approximately 10 uGy, using 70
Annually, upon acceptance, and applicable portions kVp and 1.0 mm Cu filtration. Record the
after major repairs measured dose and repeat twice. Under the
same beam conditions determine the mAs
D. Equipment required to deliver 1 uGy, 4 uGy, 12 uGy, and
50 uGy.
1. Tape measure
2. Adhesive tape
3. 1.0 mm copper filtration (> 10 x 10 cm)
4. TO20 threshold contrast test object
5. Resolution test object (e.g. Huttner 18)
6. M1 geometry test object
7. Contact mesh
8. Ionization chamber
9. Small lead or copper block (~ 5 x 5 cm)
10. Direct Digital Radiography System Survey
16-1
NMCPHC TM 6470.1
JUNE 2013
16-2
NMCPHC TM 6470.1
JUNE 2013
d. Procedure. Ensure the grid is removed from b. Regulations. The measurements should be
the system if possible and there is no used to set a baseline for future QA tests. The
attenuation in the beam. Set the FDD to be indicated sensitivity indices should not differ
approximately 180 cm. Close the collimators by greater than 20% of equivalent exposure,
and cover the detector with a lead apron. Set between exposures.
a low exposure (e.g. 50 kVp and 0.5 mAs).
Open the collimators and place the attenuating c. Equipment. 1.0 mm Cu filter
material (Cu or Pb 5 x 5 cm) on the detector
such that it covers part of the field. Make an d. Procedure. If possible, remove the grid from
exposure at 70 kVp and mAs to deliver a the system. Set a field size to cover the entire
receptor dose of order 4 uGy. Obtain another detector and a FDD as for test 1 – Dosimetry.
blank image as described above. This Expose the detector to a known dose of 10
exposure should be made 1 minute after the uGy at 70 kVp with 1.0 mm Cu at the tube
previous one. Set a very narrow window and head. Set a mAs as established from test 1 –
adjust the level. Visually inspect the image Dosimetry. Record the organ program, LUT
for any remnant of the previous image. If a name and detector dose indicator, without
remnant is visible, use ROI analysis to changing the window and leveling. Repeat at
quantify the difference in pixel value between least 3 times. Also repeat for 1 uGy and 12
the ghosted and unghosted areas. See uGy (1 image for each).
Figure 16-2. If ROI analysis is not available
the image should be transferred to a reporting e. Interpretation of Results. If indices differ by
workstation. The ROI values should be used greater than 20% of baseline, consult with
to calculate indicated receptor doses using the qualified service engineer.
STP equation established in test 3 – Linearity,
STP, and Dark Noise. 6. Uniformity
a. Purpose. To assess the uniformity of the
recorded signal from a uniformly exposed
Attenuated detector. A non-uniform response could
Region affect clinical image quality.
16-3
NMCPHC TM 6470.1
JUNE 2013
16-4
NMCPHC TM 6470.1
JUNE 2013
blurring and stitching artifacts. Repeat with a could be compared to those form other similar
finer mesh if available. systems if available. Print or save the images
for future reference, if possible. Future annual
e. Interpretation of Results. If blurring or QA testing should be performed only at 4 uGy
stitching artifacts with loss of information are and be +/- 30% of baseline.
present consult with qualified service
engineer. c. Equipment. 1.0 mm Cu filtration
16-5
NMCPHC TM 6470.1
JUNE 2013
1. Dosimetry A/P Set FDD at 150 cm and chamber 30 cm Measured dose required
from detector. Collimate to chamber, for later tests 3, 4, 5, 6,
expose at 70 kVp using 1.0 mm Cu to 10 10.
uGy. Repeat twice. Determine mAs
required to deliver 1 uGy, 4 uGy, 12 uGy,
and 50 uGy.
2. Dark Noise A/P Close collimator, cover detector with lead Used as baseline for
apron. Expose at 50 kVp and 0.5 mAs, future QA tests. On future
record detector dose indicator value and annual tests, the tolerance
pixel value. should be +/- 50%
exposure equivalent.
3. Linearity, A Open collimators and expose entire area The trend-line plotted in
System of the detector at 10 kVp with 1.0 mm Cu Excel should have an R2
Transfer filter at the tube head. Set a mAs and fit value > 0.95. There is
Properties, FDD to deliver a dose of 1 uGy (as no tolerance for the STP
and Dark determined in test 1 – Dosimetry). Record equation; however the
Noise detector dose indicator value. Repeat for pixel value to dose
doses of 4 uGy, 12 uGy, and 50 uGy. relationship should be a
Record pixel value from the center of each simple relationship (e.g.
image from the acquisition workstation or log, linear, or square root).
reporting workstation. Plot a graph of There is also no tolerance
pixel value versus receptor dose (e.g. with for the dark noise image;
Microsoft Excel). A zero dose point can however the data may be
be obtained using the result of test 2 – useful as a baseline when
Dark Noise. Obtain equation of the trend rechecking the dark noise
line for this graph. in the event of other image
quality problems.
4. Image A Set the FDD to 180 cm. Close collimators If no evidence of ghosting
Retention and cover the detector with a lead apron. is found from visual
Set a low exposure (e.g. 50 kVp and 0.5 inspection of the images
mAs). Open collimators and place the then the test is passed
attenuating material (Cu or Pb 5 x 5 cm) and there is no need to
on the detector such that it covers part of perform ROI analysis.
the field. Make an exposure at 70 kVp There should be <5%
and mAs to deliver a receptor dose of 4 (remedial) difference
uGy. Obtain another blank image as between the STP
described above. This exposure should corrected pixel values in
be made 1 minute after the previous one. the ghosted region and
Set very narrow window and adjust the the surrounding areas.
level. Visually inspect the image for any
remnant of the previous image. If a
remnant is visible, use ROI analysis to
quantify the difference in pixel value
between the ghosted and unghosted
areas. See Fig 2. The ROI values should
be used to calculate indicated receptor
doses using the STP equation established
in test 3 – Linearity, STP, and Dark Noise.
16-6
NMCPHC TM 6470.1
JUNE 2013
5. Detector A/P Set a field size to cover the entire detector The measurements should
Dose and a FDD as for test 1 – Dosimetry. be used to set a baseline
Indicator Expose the detector to a known dose of for future QA tests. The
Consistency 10 uGy at 70 kVp with 1.0 mm Cu at the indicated sensitivity
tube head. Set a mAs as established indices should not differ by
from test 1 – Dosimetry. Record the greater than 20% of
organ program, LUT name and detector equivalent exposure,
dose indicator, without changing the between exposures.
window and leveling. Repeat at least 3
times. Also repeat for 1 uGy and 12 uGy
(1 image for each).
6. Uniformity A/P Visually inspect all images obtained in test The images should not
5 – Detector Dose Indicator Consistency have obvious artifacts.
for uniformity and artifacts. The uniformity The ratio of the standard
of one of images should be assessed deviation of the 5 STP
using ROI analysis if available; to corrected ROI values to
measure the mean and standard deviation their mean (the coefficient
of the pixel values in position a-e, as of variation) should be
indicated in figure 3 (i.e. the center of the less than 10%.
image and the center of the four
quadrants). For detectors that are tiled
detectors an ROI should be drawn at the
centre of all tiles. The ROIs should be of
order 10000 pixels. If ROI analysis is not
available at the acquisition workstation
then images should be transferred to a
reporting workstation. If uniformity is poor
in the direction of the anode cathode axis
this is likely to be a result of the anode
heel affect. To confirm this, the test should
be repeated with the tube rotated through
90o. The five values obtained from ROI
analysis should be used to calculate five
indicated receptor dose values using the
STP equation obtained in test 4 – Image
Retention.
16-7
NMCPHC TM 6470.1
JUNE 2013
7. Scaling A Position the M1 test object directly onto (See Figure 4: Scaling
Errors detector with FDD of 150 cm. Expose the Errors Test) The
detector at 50-60 kV with no attenuation in measured distances x and
the beam and 10 mAs. Using distance y should agree within 3%
measuring software tools measure the of the actual distances at
dimensions (x and y) of five central the center or 5% at the
squares in both the horizontal and vertical corners. All calculated
directions. Calculate the aspect ratio x/y. aspect ratios should be
Select any corner of the image and within 1.00 +/- 0.03 at the
measure the horizontal (a) and vertical (b) center or 5% at the
sizes of two squares as indicated in corners.
Figure 4. Calculate the aspect ratio a/b.
Repeat this for one other corner. If
possible download the image as a DICOM
file. Open image using a DICOM viewer
such as Santeviewer. Hold the cursor
7. Scaling A over a corner of a square in the grid.
Errors Record the position within the image (i.e.
(continued) x and y coordinates). Move cursor to the
corner of the square of the grid 10cm from
the first corner in the x direction. Record
the coordinates again. Calculate the pixel
pitch, p(mm)=100/n
where n=number of pixels covering 10cm
of the grid. Repeat for the y direction.
This test is only necessary for
commissioning. Compare the pixel pitch
to that stated by the manufacturer. The
difference should be no greater than the
estimated measurement error.
8. Blurring A/P The test should be made with the grid No blurring should be
and Stitching both in and out of the detector. Ensure present. If stitching
Artifacts there is no attenuation in the beam and artifacts are present there
that FDD is set as large as possible. With should be no loss of
a contact mesh on the detector, make an information.
exposure at 50-60 kVp and 10 mAs using
fine focus. An MS4 or MS5 test object is
appropriate. Visually inspect the image
for blurring and stitching artifacts. Repeat
with a finer mesh if available.
16-8
NMCPHC TM 6470.1
JUNE 2013
9. Limiting A/P Ensure the grid is removed from the These measurements
Spatial system, there is no attenuation in the should be used to set a
Resolution beam and the FDD is set as large as baseline for future QA
possible. Place the resolution test object tests. Print or save the
onto the detector aligned at 45o to its images for future
edges. Set 50-60 kV and expose the reference, if possible.
cassette using 10 mAs on fine focus. Future annual QA testing
Adjust the window level and magnification should be +/- 20% of
to optimize the resolution. Score the baseline.
number of resolvable groups of lines from
the screen. The image should be scored
at a magnification of order x 5. If this
facility is not available on the review
workstation then images should be
transferred to the reporting workstation for
scoring. Look up the corresponding
resolution. Repeat the measurement
twice with the resolution test object placed
at a slight angle to the lateral or
longitudinal axis. If the system has more
than one detector element, measurements
at 45 should be made for each element.
10. Threshold A/P Ensure the grid is removed from the The results of this test are
Contrast system, if possible. With the tube, used to set a baseline for
Detail detector, and 1.0 mmCu filtration in the future QA tests. Results
Detectability same positions as for test 5 – Detector could be compared to
Dose Indicator Consistency. Place the those form other similar
TO20 (or equivalent) test object on the systems if available. Print
detector. Collimate down to the size of the or save the images for
test object. Set 70 kVp and the future reference, if
appropriate mAs to deliver ~4 Gy. possible. Future annual
Ascertain whether clinical images are QA testing should be
most commonly viewed soft or hard copy. performed only at 4 uGy
If images are viewed softcopy, score them and be +/- 30% of
on a reporting workstation optimizing baseline.
window and level settings for each detail
size. If they view hardcopy, adjust the
window to optimize the visibility of the
details, ensuring that background noise is
perceptible, and print the image out on the
largest film size. View the image on a
masked light box. Calculate an image
quality factor, IQF,
1 n H T Ai Dref
0.5
IQF ref
n i 1 H T Ai D
where:
HT(A) = threshold contrast detail
16-9
NMCPHC TM 6470.1
JUNE 2013
Abbreviations: A: acceptance; P: periodic/annual; ROI: Region Of Interest; FDD: Focus to Detector Distance; STP: System
Transfer Properties; uGy: micro-gray
16-10
Direct Digital Radiography System Survey NMCPHC TM 6470.1
JUNE 2013
Facility: Date:
Manufacturer:
Test Performed Pass Fail N/A Comments (failure comments must annotate minor or significant finding)
Uniformity
System Resolution
Energy Resolution
Multi-Energy
Resolution
Pixel Size
Count Rate
System Sensitivity
System Linearity
Center of Rotation
SPECT Only
Slice Uniformity
SPECT Only
Slice Thickness
SPECT Only
In Plane Resolution
SPECT Only
Image Contrast
SPECT Only
Additional Comments:
Purpose: Results:
Surveyor Name:
Surveyor Signature:
16-12
NMCPHC TM 6470.1
JUNE 2013
Chapter 17
17-1
NMCPHC TM 6470.1
JUNE 2013
F. General Requirements for Computed d. Procedure. Pre-expose the ion chamber free
Radiographic Equipment in air to approximately 5 - 10 mR at 72” SID
using 80 kVp with 0.5 mm Cu and 1.0 mm Al
filtration. Expose the full plate in the same
1. Phosphor Plate Throughput fashion. If unable to obtain 72” SID, expose
the plate to half the value necessary for the
a. Purpose. To verify the manufacturer’s plate
exposure and rotate the plate in between
throughput.
exposures. All plates / cassettes should be
tested. Process the plate with the vendor
b. Regulations. The throughput should be within
recommended algorithm (system diagnosis /
10% of the manufacturer’s stated rate.
flat field / 200 speed-Agfa, Test / sensitivity,
Semi EDR- Fuji, Pattern-Kodak). Take global
c. Equipment. Watch with a second sweep hand
ROIs of each image and record the average
(stopwatch). pixel values and standard deviation. Record
the proprietary exposure indicator values for
d. Procedure. If needed, pre-identify the plates.
each image.
Choose a processing type appropriate for the
plate size / type. Start the time when the first
e. Interpretation of Results. If artifacts appear
plate is placed either in the buffer (multi-plate
on the image, clean the plate with the
system) or into the single reader slot. Run at
appropriate cleaner. If the artifact is still
least ten plates (max available if this is not
visible after re-exposure, consider taking the
possible) or four plates each exposed to ~ 2
plate out of service (the technologists should
mR. Stop the time when the image is
also be making this decision between annual
displayed on the vendor processing station and
evaluations). If the pixel values are not
extrapolate to an hourly rate. Repeat this for
appropriate, consider taking the “guilty plate /
all plate sizes and types.
plates” out of service. Plates should be
replaceable to the activity at no cost.
e. Interpretation of Results. If the measured rate
does not meet the vendor’s specifications,
consult with a qualified service engineer and 3. Exposure Indicator
the local networking staff. Some vendors’
specifications are network dependent. a. Purpose. To verify that system is calibrated in
the manner intended by the manufacturer.
2. Phosphor Plate Uniformity (Reproducibility) b. Regulations. The values should be within 2%
of the manufacturer’s programmed value. The
a. Purpose. To verify that the plates respond in a
values must be within 10% of manufacturer’s
uniform manner and that all respond in a
programmed value (Agfa lgM = 2.2 < + 0.045;
similar manner as a group. To verify that
Fuji S = 200 < + 20; Kodak EI = 2000 < + 45).
artifacts will not obscure the clinical image.
c. Equipment. Same as for previous test.
b. Regulations. The pixel values should be
within 10% of the average pixel value for all
d. Procedure. Expose three plates of each size to
of the plates. The standard deviation of each approximately 1 mR at 72” SID with the
average pixel value amongst all plates should
appropriate vendor energy, filter as per
be less than 25 for Agfa, less than 20 for Fuji
Table 17-1. Process the plate with the
and Kodak. The standard deviation of the appropriate vendor algorithm (Fuji test / sens
exposure indicator values (lgM-Agfa,
(L = 1) semi-EDR; Agfa-previous; Kodak-
Sensitivity-Fuji, EI-Kodak) should be less
pattern).
than 0.02 for Agfa, less than 5% for Fuji
(standard deviation / mean sensitivity) and
e. Interpretation of Results. If values are outside
less than 20 for Kodak. The images should
of the 2% range, pay closer attention to this
also be artifact free.
throughout the year. If they fall outside of
10%, consult the vendor for proper
c. Equipment. 180 cc ion chamber, 1.5 mm
calibration. This value is crucial in daily use
Copper filter (Agfa)
by the technologist in monitoring dose and
determining the need for repeat images.
17-2
NMCPHC TM 6470.1
JUNE 2013
d. Procedure. Pre-expose the ion chamber free c. Equipment. Line pair resolution pattern (0.01
in air to 0.1, 1 and 5 or 10 mR at 72” SID to a - 0.1 mm Pb) and 180 cc ion chamber
beam with 80 kVp in the same fashion as in
the previous test. Expose the same plate nine d. Procedure. Pre-expose an ion chamber free in
times (three using each exposure level) in the air at 72” SID to a beam using 60 kVp, no
same fashion. Process per previous test - added filtration and enough mAs for 5 mR.
Agfa (Fuji- Test / ave 4.0 Semi-EDR). Place Expose a cassette of each size in the same
ROIs on the images to get pixel values; fashion. If multiple test patterns are available,
technologists should also be making this you may expose each size once. If not, the
decision in between annual evaluations. Plot pattern must be exposed two times each
the average pixel value of the three, and the (center x (y-acceptance), periphery x).
proprietary indicators versus Log of the Collimation is necessary to avoid wash out,
exposure in Roentgens. but too much collimation may cause the
software to misjudge auto collimation. It may
e. Interpretation of Results. Consult the field also misjudge if multiple patterns are on a
service engineer if values vary greater than small cassette. Process the plates with an
expected. appropriate algorithm for the plate size (as in
the throughput test). View the image with the
5. Laser Beam Evaluation magnification tool (10 X on narrow window)
on a 2 K workstation (if not available at the
a. Purpose. To ensure that the CR reader unit site, consider printing on film, or viewing on a
laser is sampling all plate data points and is 1 K workstation).
not skipping data lines.
e. Interpretation of Results. If the resolution is
b. Regulations. There should be no signal drop below the expected value, first ensure that the
out and no more than occasional jitters. resolution for the radiographic tube is
adequate and then consult with a qualified
c. Equipment. Metal coated or steel straight service engineer. The laser may need to be
edged ruler and 180 cc ion chamber replaced.
17-3
NMCPHC TM 6470.1
JUNE 2013
c. Equipment. Low contrast test pattern and 180 the test tool in the image center (i.e. apply
cc ion chamber. collimation). Process the plate with an
appropriate algorithm for the plate size. Re-
d. Procedure. Pre-expose an ion chamber in a expose at a very low exposure without the test
tabletop 40” SID free in air mode using 75 tool and using slightly smaller collimation.
kVp, 1.5 mm added copper filtration, and View on the image processing station.
enough mAs for 0.1, 1 and 5(Agfa) 10(Fuji)
mR respectively. Expose three cassettes in the e. Interpretation of Results. Consult a qualified
same manner with the test tool on top of each service engineer if a residual image is seen,.
cassette. Process the image using a general
low contrast algorithm (Agfa-flat field, 200 9. Phosphor Plate Dark Noise
speed Fuji-test / sensitivity, semi-EDR,
Kodak-pattern). View using window and a. Purpose. To ensure that electronic artifacts do
level as appropriate on a 2K/printer/1K not exist.
workstation (depending on availability).
b. Regulations. There should be no visible
e. Interpretation of Results. Consult with the artifacts. (Agfa- lgM < 0.28 / SAL < 130,
qualified service engineer if low contrast average pixel value < 350 and stand dev < 5,
resolution is noticeably degraded. Consider Fuji- Average pixel value < 280, stand dev <
using a DSA step wedge phantom to 4, Kodak EI < 80, stand dev < 4).
compensate for anode-heel effect.
c. Equipment. None
8. Erasure Thoroughness
d. Procedure. Process three newly erased plates
a. Purpose. To ensure that vestiges of previous per the manufacturer’s specified algorithms
images do not appear in the current image. (Agfa-system diagnosis / flat field / speed 200,
Fuji-Test / sensitivity (L = 1), fixed EDR (S =
b. Regulations. Ghost images of prior exposures 10,000). Perform ROIs on the resulting
should not appear in a current image. images. Record average pixel values,
standard deviation and proprietary values.
c. Equipment. High contrast tool (Pb or line pair
phantom) e. Interpretation of Results. Consult with the
qualified service engineer if values are not
d. Procedure. Expose a plate at 72” SID to an 80 below the specified limit or artifacts are seen.
kVp beam with a high incident exposure and
17-4
NMCPHC TM 6470.1
JUNE 2013
1. Phosphor A/P If possible, at least ten of each cassette Should be within ± 10%
Cassette size, or four of each size exposed to ~ 2 of manufacturer’s spec.
Throughput mR then extrapolate to hourly rate
2. Phosphor A/P Expose each plate to 10 mR at 80 kVp Agfa PVSD < 25, Fuji and Kodak
Plate Uniformity with 0.5 mm Cu and 1 mm Al filtration PVSD < 20 or all PVs within 10%,
(Reproducibility) and 72” SID. Gather pixel data with other values vendor specific, all
global ROIs and vendor specific values plates should be artifact free
3. Exposure A/P Expose three plates of each size to 1 Exposure Indicator should be within ±
Indicator mR at 72” SID, gather pixel data with 2% of manufacturer’s programmed,
global ROIs and vendor specific values absolutely must be within 10% (Agfa
(Agfa: 75 kVp, 1.5 mm Cu, no delay; lgM = 2.2 < +/- 0.045, Fuji S = 200 <
Fuji: 80 kVp no filter, 10 minute delay; +/- 20, Kodak EI = 2000 < +/- 45)
Kodak: 80 kVp, 0.5 mm Cu + 1 mm Al,
15 minute delay)
4. Linearity A/P Expose the same plate three times Noise should decrease with
each to 0.1 mR, 1 mR and 5-10 mR(if exposure, the linearity coefficients
can get to 10 with an exposure and slope values are vendor specific;
indicator) with 80 kVp beam; gather all plots should result in a straight line
pixel data with ROIs and vendor specific (Agfa-SlopelgM - 1 < +/- 0.1, Fuji-
values; plot the vendor specific values SlopeS +1 < +/- 0.1 or SlopePV/256 – 1
and pixel values versus lg (exposure) < +/- 0.1, Kodak-SlopeEI/1000 – 1 <
+/- 0.1 or SlopePV/1000 – 1 < +/- 0.1)
5. Laser Beam A/P Expose a plate to 5 mR with ruler There should be no signal drop out or
Evaluation centered on cassette and perpendicular no more than an occasional ± jitter
to the scan lines (slight angle)
6. Spatial A/P Expose a line pair phantom on each The resolution must be within 10% of
Resolution plate size to 5 mR using an unfiltered the manufacturer’s stated (Rhor / fNyquist
60 kVp beam at 72” SID in the x and y > 0.9, Rver / fNyquist > 0.9, R45° / (1.41
directions and at 45°, for both center fNyquist) > 0.9
and periphery
7. Low Contrast A/P Expose a low contrast phantom on a The low contrast should improve with
single plate to 0.1 mR, 1 mR and 5-10 increased exposure, it should be
mR comparable to film/screen, if annual,
should be similar to previous year’s
8. Erasure A/P Expose a large cassette to 50 mR with There should not be a ghost image in
Thoroughness high contrast object and process (72’’ second image; Agfa PV < 630, PVSD
SID, no filter). Re- expose processed < 5; Fuji PV < 280, PVSD < 4; Kodak
plate to 1 mR without object and ~ 5 cm PV < 80, PVSD < 4
reduced collimation; reprocess one
more time using “dark noise” settings
for quantitative data
9. Phosphor A/P Take three recently processed (erased) The plates should not show any
Plate Dark Noise plates and process per manufacturer’s visible artifacts, ROI values vary by
recommendations vendor
17-5
NMCPHC TM 6470.1
JUNE 2013
17-6
Computed Radiography System Survey NMCPHC TM 6470.1
JUNE 2013
Facility: Date:
Manufacturer:
Test Performed Pass Fail N/A Comments (failure comments must annotate minor or significant finding)
Phosphor Cassette
Phosphor Plate
Uniformity
Exposure Indicator
Linearity
Laser Beam
Evaluation
Spatial Resolution
Low Contrast
Beam Central /
Mechanical Spot Film
Alignment
Distance Accuracy
and Aspect Ratio
Erasure
Thoroughness
Phosphor Plate Dark
Noise
Additional Comments:
Purpose: Results:
Surveyor Name:
Surveyor Signature:
17-8