A-Dec - Manual Cuspidor

Download as pdf or txt
Download as pdf or txt
You are on page 1of 22

Owner's Guide

CASCADE 7284
®

AND
RADIUS 7285
®

CUSPIDOR

85.2609.00
ALPHABETICAL
EQUIVALENT TO THE NUMERAL
OF THE MONTH MANUFACTURED

SERIAL (SN)
A January
NUMBER
B February
C March
D April
E May
F June
G July
H August MODEL (REF)
I September
NUMBER
J October
K November
L December SERIAL NUMBER
IDENTIFICATION
YEAR MANUFACTURED

Cuspidor Serial Number Location:


• On the underside of the cuspidor

For service information contact your local authorized A-dec dealer.


Check with local codes and A.D.A. (Americans with Disabilities Act)
Requirements for Installation of this product.
Warranty

Warranty:
A-dec warrants all products in this owner's guide against
defects in material or
workmanship for one year from time of delivery. A-dec's
sole obligation under the warranty is to provide parts for
the repair, or at its option, to provide the replacement
product (excluding labor). The buyer shall have no other
remedy. All special, incidental, and coincidental damages
are excluded. Written notice of breach of warranty must
be given to A-dec within the warranty period. The
warranty does not cover damage resulting from improper
installation or maintenance, accident or misuse. The
warranty does not cover damage resulting from the use of
cleaning, disinfecting or sterilization chemicals and
processes. The warranty also does not cover light bulbs.
Failure to follow instructions provided in the A-dec
owner's guide (operation and maintenance instructions)
may void the warranty. A-dec warrants A-dec dental chair
cylinders, both lift and tilt, for ten years from the date of
purchase of the chair or the cylinder. This warranty is
retroactive to A-dec chair cylinders already in the field.
The warranty covers chair cylinders A-dec finds to have
manufacturing related irregularities. Stool cylinders are
covered under A-dec's one-year warranty.

NO OTHER WARRANTIES AS TO MERCHANTABILITY OR


OTHERWISE ARE MADE.
Cascade 7284 and Radius 7285 Cuspidor

Return Policy:
U.S. and Canadian dealers wishing to return overstock
(unopened) merchandise to A-dec for credit consideration
must include a copy of the original invoice number. A
return authorization form from an A-dec Territory
Manager must be included with serial numbered
equipment or A-dec/W&H handpieces. A 15% restocking
fee will be assessed. Merchandise that cannot be returned
for credit includes parts assembled to the dental unit,
chair, light, or dental furniture; obsolete parts; and
specials. Dental furniture cannot be returned for credit.
Standard color upholstery ordered for obsolete chairs or
stools cannot be returned for credit. In the case of a
defective warranty item, a copy of the replacement in-
voice, serial number of the unit under which it was
replaced, and a description of the symptoms of the
defect must be returned with the part to:
A-dec Inc. 2601 Crestview Drive,
Newberg, Oregon 97132, USA.
Cascade 7284 and Radius 7285 Cuspidor

Equipment Alterations Policy:


Certain modifications or alterations of A-dec equipment
which expand the use of A-dec equipment beyond its
design and intent, or which override any safety features
of A-dec equipment may jeopardize doctor, patient or
staff safety. Field modifications that alter the electrical
and/or mechanical safety of A-dec dental devices are in
conflict with Underwriters Laboratory (UL) construction
file requirements and are not sanctioned by A-dec.
Examples of field modifications that diminish safety
design include, but are not limited to, rendering access to
the line voltage without the use of tools, modification of
supporting elements that increase or shift loading
characteristics, and the addition of any powered device
that exceeds the design limits of the dental system. The
use of accessory equipment not complying with the
equivalent safety requirements of A-dec equipment may
lead to a reduced level of safety of the resulting system.
It is the responsibility of the equipment distributor and
the installer to assure that the installation complies with
all building code requirements. The responsibility to
determine whether a modification or alteration of A-dec
equipment falls within these constraints is with the
person(s) who initiates, approves and/or performs such
modification or alteration. A-dec will not respond to
inquiries on an individual basis. This person(s) will be
deemed to have assumed all associated risks with such
alteration or modification and will hold A-dec harmless
from resulting claims, including product liability claims.
Additionally, such modification or alteration voids A-dec's
warranty and may invalidate UL or other regulatory
agency approval.

All product names used in this document are trademarks or


registered trademarks of their respective holders.

Printed in U.S.A. • Copyright © 2006• All Rights Reserved

v
Cascade 7284 and Radius 7285 Cuspidor

Cascade 7284 Cuspidor

Radius 7285 Cuspidor


Cascade 7284 and Radius 7285 Cuspidor

Serial number location, service information, and


warranty information are located on the inside of
the front cover and front page.

CONTENTS
About Your Cuspidor .............................. 2
Cuspidor Controls ................................. 3
Touch Pad (optional) ............................... 4
Master On/Off Toggle .............................. 4
Cuspidor Limit Switch ............................. 5
Gravity Drain Cleaning ............................ 5
Cleaning the Cuspidor ............................. 6
Care Instructions .................................. 7
Bowl Rinse Adjustment ........................... 7
Bowl Rinse Flow Adjustment ................ 7
Bowl Rinse Time Adjustment ................ 8
Cup Fill Adjustment ............................... 9
Utility Controls ................................... 10
Maintenance ...................................... 11
Adjustments and Specifications ................ 11
Identification of Symbols ........................ 12
Classification of Equipment (EN 60601-1) ..... 12

1
Cascade 7284 and Radius 7285 Cuspidor

About Your Cuspidor


The cuspidor is equipped with a timer for the bowl
rinse. This timer can be adjusted to increase or
decrease the length of time water flows into
the bowl.
The cuspidor comes standard with an
automatically timed cup fill.
On most units, the cuspidor on/off function is
controlled by the master on/off toggle on the
Cascade control head. For cuspidor only units, the
cuspidor on/off function is controlled by the on/off
toggle on the post box (Cascade 7284 only).

CONTOURED
WATER SPOUT

BOWL
SPOUT

CUP LOCATION BOWL


SCREEN

CUP FILL
CUSPIDOR
BUTTON BOWL RINSE HOUSING
BUTTON

Figure 1. Cascade 7284 Cuspidor and


Radius 7285 Cuspidor

2
Cascade 7284 and Radius 7285 Cuspidor

Cuspidor Controls
Cup Fill Button. Allows water to flow into a cup.
When your cuspidor has an automatic timer, press
the button once and water will flow for
approximately four seconds. If you have a manual
cuspidor, water will flow as long as the button is
pressed. (Refer to page 9 for cup fill adjustments.)
Bowl Rinse Button. Starts the timed bowl rinse.
The bowl rinse provides rinse water for the cuspidor
bowl. (Refer to page 7 for bowl rinse flow adjust-
ments.)
Bowl Rinse Time Adjustment. Adjusts the length
of time that water flows to the bowl rinse. (Refer to
page 8 for bowl rinse time adjustments.)

CUP FILL BOWL RINSE


BUTTON BUTTON

Figure 2. Cuspidor Control Locations

3
Cascade 7284 and Radius 7285 Cuspidor

Touch Pad (Optional)


Your cuspidor may be equipped with a touch pad.
For chair touch pad instructions, refer to the Cascade
1040 Chair Owner’s Guide (A-dec Publication No.
85.2605.00) or Decade 1011/1021 Chair Owner’s
Guide (A-dec Publication No. 85.2635.00).
For Cascade Master Series® touch pad instructions,
refer to your Cascade Master Series Touch Pad
Owner's Guide, (A-dec Publication No. 85.2627.00).

Master On/Off Toggle


The master on/off toggle can be found on the post
box, control head or Master Series touch pad. Moving
the toggle to the ON position provides air and water
to the system. Move the toggle to the OFF position
while the unit is unattended to prevent water
damage in case of a leak.

TOGGLE ON
CONTROL HEAD

TOGGLE ON
POST BOX TOGGLE ON
MASTER SERIES
TOUCH PAD

Figure 3. Master On/Off Toggle Locations

4
Cascade 7284 and Radius 7285 Cuspidor

Cuspidor Limit Switch


Cuspidors mounted to A-dec Cascade or Decade®
chairs are equipped with a cuspidor limit switch. This
limit switch prevents potential equipment damage by
stopping the chair immediately whenever the cuspidor
is inadvertently lowered over an object, such as a
stool.
To test the cuspidor limit switch, slightly lift the
cuspidor and listen for a “click”. The “click”
indicates that a switch has been activated. To
check the operation of the limit switch, start
lowering the chair. While the chair is in motion, lift
the cuspidor slightly. The chair will stop and all
chair functions except Base Up will be disabled.
The Base Up function remains active so you can
raise the chair to remove any object caught under
the cuspidor. When the object has been removed,
all chair functions will operate normally.

Gravity Drain Cleaning


At the end of each day, flush the drain to remove
debris which can collect in the flexible drain tubing.
If the drain is not flushed regularly, debris may build
up and impair cuspidor draining. To flush the drain,
raise the chair to its highest position and run the
bowl rinse several times, long enough so that water
flows down the drain for about 60 seconds.
Caution
Do not empty the solids collector screen into your
cuspidor. Doing so could plug the cuspidor drain.

5
Cascade 7284 and Radius 7285 Cuspidor

Cleaning the Cuspidor


Caution
Clean the cuspidor with the contoured water spout in place.
This will help prevent cleaning solutions from causing
potential equipment damage and contamination of the
water flowing to the cup.

The cuspidor is designed for quick and easy


cleaning. The contoured spouts and the smooth cuspi-
dor housing help facilitate more complete cleaning.
When cleaning, do not remove the contoured water
spout (see Figure 4). Also remember to empty and
clean the bowl screen.

Caution
Do not empty the bowl screen into your cuspidor. Doing
so could plug the cuspidor drain.

CONTOURED
WATER SPOUT BOWL
SCREEN

Figure 4. Cleaning the Cuspidor

Caution
After cleaning, always install the bowl screen in the
cuspidor bowl drain to prevent debris from plugging the drain.

6
Cascade 7284 and Radius 7285 Cuspidor

Care Instructions
For recommended asepsis instructions, refer to
your Equipment Asepsis Owner’s Guide, A-dec
Publication No. 85.0696.00.

Bowl Rinse Adjustment


The flow of water to rinse the bowl and the length
of time that water flows to rinse the bowl are set at
the factory. Water conditions in your location may
require a change in these settings.

Bowl Rinse Flow Adjustment


The bowl rinse flow adjustment location is inside
the post box or under the cuspidor bowl (see Figure
5). To remove the side cover from the post box
(Cascade 7284 only), insert your fingers under the
lower edge of the cover and pull out and down on
the side cover. If you have a Radius 7285 Cuspidor,
loosen the thumbscrew and lower the cover.

POST BOX
INDICATOR
SIDE COVER
DOT

BOWL
RINSE FLOW
ADJUSTMENT

Figure 5. Bowl Rinse Flow Adjustment


7
Cascade 7284 and Radius 7285 Cuspidor

Locate the bowl rinse flow adjustment (see Figure


5). Notice the position of the indicator dot which
shows the present setting of the flow control.
Using an adjustment key or a 1¼8-inch hex key,
turn the adjustment key clockwise to increase the
flow rate and counterclockwise to decrease the flow
rate.
Check the flow rate by pressing the bowl
rinse button. When the adjustment has been
completed, reinstall the cover.

Bowl Rinse Time Adjustment

To change the length of time that water flows for


the bowl rinse, locate the time adjustment screw
(see Figure 6). Use an adjustment key or a 1¼8-inch
hex key and turn the adjustment key clockwise (as
viewed from the bottom) to increase the length of
time water flows to rinse the bowl or counterclock-
wise to decrease the length of time. Check the
length of time for water flow and adjust until it is
satisfactory.

ADJUSTMENT SCREW

Figure 6. Bowl Rinse Time Adjustment

8
Cascade 7284 and Radius 7285 Cuspidor

Cup Fill Adjustment


Timed or Manual Cup Fill
The timer for the automatic cup fill is set at the
factory. If you ordered an optional manual cup fill,
the water flow is set during installation to
prevent splashing. In either case, the amount of
water flowing into the cup may be changed.
If an adjustment is necessary to either the manual
or timed cup fill, remove the side cover on the post
box (Cascade 7284 Cuspidor) by inserting your
fingers under the lower edge of the cover and pulling
out and down on the cover (see inset in Figure 5 on
page 7). If you have a Radius 7285 Cuspidor, loosen
the thumbscrew and lower the cover.

CUP FILL FLOW


INDICATOR
ADJUSTMENT
DOT

Figure 7. Cup Fill Flow Adjustment

9
Cascade 7284 and Radius 7285 Cuspidor

Locate the cup fill flow adjustment (see Figure 7).


Notice the location of the indicator dot showing the
position of the current setting. Use an adjustment key
or a 1¼8-inch hex key and turn the key clockwise to
increase the flow rate or counterclockwise to
decrease the flow rate. Check the cup fill flow.
When the adjustment has been completed, reinstall
the cover.

Utility Controls
The air and water manual shutoff valves, filter and
regulator package, as well as any vacuum or drain
outlets, are located in the floor box mounted over
the office plumbing on the floor or wall. If you have
a Cascade 7284 Cuspidor, some controls are located
in the post box at the side of the chair.
Refer to the Floor Boxes Owner’s Guide (A-dec
Publication No. 85.2611.00) for complete
information about the floor box.

10
Cascade 7284 and Radius 7285 Cuspidor

Maintenance
Gravity Drain Cleaning ................................page 5
Cleaning the Cuspidor.................................page 6
Care Instructions
Equipment Asepsis
Owner’s Guide ..............................85.0696.00
Parts
Maintenance Parts
Owner’s Guide ..............................85.2634.00

Adjustments and Specifications


Touch Pad Functions
Cascade 1040 Chair
Owner’s Guide ..............................85.2605.00
Decade 1011/1021 Chair
Owner’s Guide ..............................85.2635.00
Cascade Master Series Touch Pad
Owner's Guide...............................85.2627.00
Bowl Rinse Adjustment ...............................page 7
Cup Fill Adjustment ...................................page 9
Utility Controls
Floor Boxes
Owner’s Guide ..............................85.2611.00
Minimum air, water and vacuum service requirements
for proper unit operation:
Air: 2.50 cfm (70.80 l/min) at 80 psi (551 kPa).
Water: 1.50 gpm (5.68 l/min) at 40 psi (276 kPa).
Vacuum: 12 cfm (339.84 l/min) at 8 inches of
mercury (27 kPa).
Maximum weight for the 7284: 160 lbs (72.58 kg).
Maximum moment at the 7284 unit mount arm created
by the freestanding unit is 250 ft-lbs
(339 N·m) along the axis of the post box.

11
Cascade 7284 and Radius 7285 Cuspidor

Maximum weight for the 7285: 50 lbs (22.68 kg).


Maximum moment at the 7285 unit mount arm created
by the freestanding unit is 40 ft-lbs
(54 N·m).

Specifications are subject to change without notice.

12
Cascade 7284 and Radius 7285 Cuspidor

Identification of Symbols
Symbol Description
Recognized by Underwriters Laboratories Inc. with respect to electric shock, fire and mechanical hazards
only in accordance with UL 60601-1 (2601-1) and under mutual recognition agreement with CAN/CSA C22.2,
No. 601.1.

Classified by Underwriters Laboratories Inc. with respect to electric shock, fire and mechanical hazards
only in accordance with UL 60601-1 (2601-1) and under mutual recognition agreement with CAN/CSA C22.2,
No. 601.1.

UL listed to UL 61010A-1, BS EN 61010-2-010 and Canadian (CAN/CSA C22.2, No. 1010.1-92)


safety standards.

Conforms to applicable European Directives (refer to Declaration of Conformity).

Protective earth (ground).

Functional earth (ground).

Attention, consult accompanying documents. No user servicable parts . Attention, line voltage. Only
licensed electrician should remove cover.

Type B applied part.

Class II equipment.

Caution: Metal surfaces can be hot during and following the dry cycle.

Classification of Equipment (60601-1)


Type/Mode Classification
Types of shock CLASS I EQUIPMENT: Dental chairs, dental lights, and power supplies
protection CLASS II EQUIPMENT: Chair, wall, and cart-mounted delivery systems

Degree of shock TYPE B APPLIED PART: Delivery systems only


protection

Degree of pro- ORDINARY EQUIPMENT: All products


tection against
water ingress

Mode of CONTINUOUS OPERATION: All models except dental chairs


operation CONTINUOUS OPERATION WITH INTERMITTENT LOADING: Dental chairs - 5% duty cycle

Flammable Not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous
Gasses: oxide, where such gasses may accumulate in concentration (closed space).

13
Cascade 7284 and Radius 7285 Cuspidor

Electrical Rating
Type Specification
Volts 100/110-120/220-240 VAC

Frequency 50-60 Hz

Current As configured and specified in equipment manual (products labeled 15A or greater require dedicated
circuit, identified in distribution panel).

Environmental Specifications
Temperature/Humidity Specification
Storage/Transportation -40°C to 70°C (-40°F to 158°F) - Relative humidity: 80% for up to 31°C, decreasing
Temperature: linearly to 50% at 40°C.

Operating Temperature: 10°C to 40°C (40°F to 104°F) - Relative humidity: 80% for up to 31°C, decreasing
linearly to 50% at 40°C.

Indoor Use: Altitude up to 2,000M (6,563 ft.), installation category II, pollution degree 2.
(UL 61010A-1 and CAN/CSA C22.2, No. 1010.1-92 only)

14
Notes
USA and Canada
2601 Crestview Drive
Newberg, Oregon 97132 USA
Phone: 1-800-547-1883
1-503-538-7478
Fax: 1-503-538-0276
www.a-dec.com

International
Phone: 1-503-538-9471
Fax: 1-503-538-5911

Distribution Centers
A-dec Australia
41-43 Bowden Street
Alexandria, NSW 2015, Australia
Phone: 61-(0)2-9699 4600
Fax: 61-(0)2-9699 4700
www.adec.com.au

A-dec United Kingdom


Austin House
11 Liberty Way
Nuneaton, Warwickshire
England CV11 6RZ
Phone: 0800-ADEC-UK (2332-85) Within UK
44 24 7635 0901 Outside UK
Fax: 44 24 7634 5106

85.2609.00 Printed in USA.


2006-02 Rev K Copyright © 2006,
(03597) Made with recycled paper All Rights Reserved.

You might also like