Product Information: Toshiba X-Ray Tube D-054 / D-054S / D-054SB

Download as pdf or txt
Download as pdf or txt
You are on page 1of 9

Product Information PI

TOSHIBA X-RAY TUBE


D-054 / D-054S / D-054SB

Stationary Anode X-ray Tube


 Especially designed for dental radiography unit.

 Low target angle adaptive for dental


tomography.

 Provided with an insulation cylinder (D-054S)


and lead cylinder. (D-054SB)

 These tubes have 0.5 focus, and are available


for maximum tube voltage 100 kV.

 Installed in the same enclosure with the high


voltage transformer.

General Data

Electrical:
Circuit:
High Voltage Generator ......................................................... Two-Peak High-Voltage Generator
Constant Potential High-Voltage Generator
Grounding ......................................................................................................... Center-Grounded
Nominal X-ray Tube Voltage ................................................................................................... 100 kV
Nominal Focal Spot Value ............................................................................................................ 0.5
Nominal Anode Input Power (at 1.0s) (See Rating Chart)
Two-Peak High-Voltage Generator ..................................................................................... 840 W
Constant Potential High-Voltage Generator ...................................................................... 1750 W
Exposure Duty Cycle ..................................................................................................... 1:60 or more
(Exposure time : interval time)

Mechanical:
Dimensions:
Overall Length ......................................................................................... See dimensional outline
Maximum Diameter ................................................................................. See dimensional outline
Target:
Anode Angle ................................................................................................................. 5 degrees
Material .......................................................................................................................... Tungsten
Inherent Filtration .................................................................................... At least 0.8 mm Al at 50 kV
X-ray Coverage .................................................................................... 75 × 380 mm at SID 550 mm

★The information contained herein is presented only as a guide for the applications of our products.
No Responsibility is assumed by TOSHIBA ELECTRON TUBES & DEVICES CO.,LTD.(TETD) for any infringements of patents or other rights of the
third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of TETD or others.
★The information contained herein may be changed without prior notice. It is therefore advisable to contact TETD before proceeding with the design of
equipment incorporating this product.
D-054 / D-054S / D-054SB

Weight:
D-054 ...................................................................................................................... Approx. 320 g
D-054S .................................................................................................................... Approx. 460 g
D-054SB ................................................................................................................. Approx. 780 g
Cooling Method .............................................. Oil immersed (60℃ Max.) and convection oil cooling.
Tube Holding:
D-054 ......................................... Holding the glass envelope of the anode end and cathode end,
or the screw of the anode shank.
D-054S .......................................................................................... Holding the insulation cylinder.
D-054SB ............................................................................................... Holding the lead cylinder.

Absolute Maximum and Minimum Ratings


(At any time, these values must not be exceeded.)

Maximum X-ray Tube Voltage .................................................................................................. 100 kV


Between Anode (or Cathode) and Ground ................................................................................. 50 kV
Minimum X-ray Tube Voltage ..................................................................................................... 50 kV
Maximum X-ray Tube Current: ................................................................................. See rating charts
Two-Peak High-Voltage Generator ..................................................................................... 24 mA
Constant Potential High-Voltage Generator ........................................................................ 22 mA
Maximum Filament Current ........................................................................................................ 3.5 A
Maximum Filament Voltage ........................................................................................................ 4.2 V
Filament Frequency Limits ........................................................... DC or AC (Sine Wave) 0 ~ 20 kHz
Thermal Characteristics:
Anode Heat Content ............................................................................................................ 35 kJ
Maximum Anode Heat Dissipation ...................................................................................... 250 W
Maximum Radiographic Exposure Time ...................................................................................... 20 s

Environmental Limits

Operating Limits (in dielectric oil):


Oil Temperature ........................................................................................................... 10 ~ 60 ℃
Oil Pressure ............................................................................................................. 70 ~ 140 kPa
Shipping and Storage Limits:
Temperature .............................................................................................................. -40 ~ 70 ℃
Humidity ........................................................................................................................ 10 ~ 90 %
(No condensation)
Atmospheric Pressure .................................................................................................50 ~ 106 kPa

-2-
D-054 / D-054S / D-054SB

Maximum Rating Charts


(Absolute maximum rating charts)

Two-Peak High-Voltage Generator

Nominal Focal Spot Value: 0.5


50kV
24

60kV
20
TUBE CURRENT [mA]

70kV

80kV
15
90kV
100kV

10

5
1 2 3 5 7 10 20
EXPOSURE TIME [s]

Constant Potential High-Voltage Generator

Nominal Focal Spot Value: 0.5


80kV 70kV
22
60kV

20 90kV
50kV
100kV
TUBE CURRENT [mA]

15

10

5
1 2 3 5 7 10 20
EXPOSURE TIME [s]

-3-
D-054 / D-054S / D-054SB

Emission & Filament Characteristics

Two-Peak High-Voltage Generator

Nominal Focal Spot Value: 0.5

25 7

20 6

FILAMENT VOLTAGE [V]


50kV
TUBE CURRENT [mA]

15 5
100kV

10 4
Ef

5 3

0 2
2.9 3.0 3.1 3.2 3.3 3.4 3.5
FILAMENT CURRENT [A]

Note: This graph indicates typical characteristics.

Constant Potential High-Voltage Generator

Nominal Focal Spot Value: 0.5


25 6

80kV

20 5
FILAMENT VOLTAGE [V]

100kV
TUBE CURRENT [mA]

15 Ef 4

10 3

50kV
5 2

0 1
2.9 3.0 3.1 3.2 3.3 3.4 3.5

FILAMENT CURRENT [A]

Note: This graph indicates typical characteristics.

-4-
D-054 / D-054S / D-054SB

Anode Heating / Cooling Curve


35

30 225W

175W
HEAT STORAGE [kJ]

25

HEATING
20

15

10
COOLING

0
0 2 4 6 8 10
TIME [min]

-5-
D-054 / D-054S / D-054SB

Dimensional Outline of D-054


Unit: mm

M6 8DEEP φ15.7

(8)

73±1
5°

138±3
CENTRAL
φ45MAX X-RAY

FOCAL SPOT
14MAX


20.
C2 F φ


CENTRAL X-RAY

SHORT PIN
C1 F

F    :FILAMENT
C1,C2:CATHODE
N    :NOT FOR USE

-6-
D-054 / D-054S / D-054SB

Dimensional Outline of D-054S


Unit: mm

φ54
INSULATION CYLINDER
M6 8DEEP
φ15.7

73±1
φ25

5°

146±3
143

CENTRAL
X-RAY

FOCAL SPOT
6MAX


0.
φ2
C2 F

CENTRAL X-RAY

C1 F
SHORT PIN

F    :FILAMENT
C1,C2:CATHODE
N    :NOT FOR USE

-7-
D-054 / D-054S / D-054SB

Dimensional Outline of D-054SB


Unit: mm

φ58

φ54
INSULATION CYLINDER
M6 8DEEP
φ15.7
30

73±1
φ30
φ25
40

5°

146±3
143

CENTRAL
X-RAY
65

FOCAL SPOT

LEAD SYLINDER
6MAX


0.
φ2
C2 F

CENTRAL X-RAY

C1 F
SHORT PIN

F    :FILAMENT
C1,C2:CATHODE
N    :NOT FOR USE

-8-
D-054 / D-054S / D-054SB

OVERSEAS SUBSIDIARIES AND AFFILIATES

EU REPRESENTATIVE
・TOSHIBA ELECTRONICS EUROPE GMBH
HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296-107 FAX +49 (211) 5296-402

For Sales & Technical Services, please contact the following representative:

・TOSHIBA ELECTRONICS EUROPE GMBH


HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296-107 FAX +49 (211) 5296-402

・TOSHIBA AMERICA ELECTRONIC COMPONENTS, INC.


2150 EAST LAKE COOK ROAD, SUITE 310
BUFFALO GROVE, ILLINOIS 60089 USA
PHONE +1 (847) 484-2400 FAX +1 (847) 541-7287

・TOSHIBA ELECTRON DEVICES & MATERIALS (SHANGHAI) CO., LTD. (TEMS)


RM1606, SH-PLAZA,
No.336, XIZANG ROAD (MIDDLE), SHANGHAI, 200001, CHINA
PHONE +86 (21) 6361-0077 FAX +86 (21) 6351-5760

(Product Development Engineering Department)


1385 SHIMOISHIGAMI, OTAWARA-SHI, TOCHIGI-KEN, 324-8550, JAPAN
PHONE: +81-287-26-6666 FAX: +81-287-26-6060
http://www.toshiba-tetd.co.jp/

・Toshiba Electron Tubes & Devices Co., Ltd. meets the Environmental Management System Standard, ISO14001.
・Toshiba Electron Tubes & Devices Co., Ltd. meets internationally recognized Standards for Quality Management System ISO9001, ISO13485.

You might also like