Product Information PI
TOSHIBA X-RAY TUBE
D-063 / D-063R / D-063S / D-063SR
Stationary Anode X-ray Tube
Especially designed for mobile type surgical
X-ray TV apparatus.
Provided with an insulation cylinder.
(D-063S / D-063SR)
These tubes have focus 0.6, and are available
for maximum tube voltage 110 kV.
General Data
Application ........................................................ Diagnostic Radiography for Mobile C-arm
Electrical:
Circuit:
High Voltage Circuit .................................................. Constant Potential High-Voltage Generator
Grounding ......................................................................................................... Center-Grounded
Nominal X-ray Tube Voltage ................................................................................................... 110 kV
Nominal Focal Spot Value ............................................................................................................ 0.6
Nominal Anode Input Power (at 1.0s) .................................................................................... 2020 W
Exposure Duty Cycle ..................................................................................................... 1:60 or more
(Exposure Time : interval time)
Mechanical:
Dimensions:
Overall Length ......................................................................................... See dimensional outline
Maximum Diameter ................................................................................. See dimensional outline
Target:
Anode Angle ................................................................................................................. 8 degrees
Material .......................................................................................................................... Tungsten
Inherent Filtration .................................................................................... At least 0.8 mm Al at 50 kV
X-ray Coverage ......................................................................................... φ245 mm at SID 970 mm
Weight:
D-063 ...................................................................................................................... Approx. 320 g
D-063R .................................................................................................................... Approx. 440 g
D-063S .................................................................................................................... Approx. 450 g
D-063SR ................................................................................................................. Approx. 570 g
★The information contained herein is presented only as a guide for the applications of our products.
No Responsibility is assumed by TOSHIBA ELECTRON TUBES & DEVICES CO.,LTD.(TETD) for any infringements of patents or other rights of the
third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of TETD or others.
★The information contained herein may be changed without prior notice. It is therefore advisable to contact TETD before proceeding with the design of
equipment incorporating this product.
D-063 / D-063R / D-063S / D-063SR
Cooling Method .............................................. Oil immersed (60℃ Max.) and convection oil cooling.
Tube Holding:
D-063 / D-063R .......................... Holding the glass envelope of the anode end and cathode end,
or the screw of the anode shank.
D-063S / D-063SR ........................................................................ Holding the insulation cylinder.
Absolute Maximum and Minimum Ratings
(At any time, these values must not be exceeded.)
Maximum X-ray Tube Voltage ................................................................................................ 110 kV
Between Anode (or Cathode) and Ground ................................................................................ 55 kV
Minimum X-ray Tube Voltage ................................................................................................... 40 kV
Maximum X-ray Tube Current .................................................................................................. 20 mA
Filament Characteristics:
Maximum Filament Current ................................................................................................... 3.7 A
Filament Frequency Limits ...................................................... DC or AC (Sine Wave) 0 ~ 20 kHz
Thermal Characteristics:
Anode Heat Content:
D-063 / D-063S ............................................................................................................... 28 kJ
D-063R / D-063SR .......................................................................................................... 34 kJ
Maximum Anode Heat Dissipation:
D-063 / D-063S .............................................................................................................. 265 W
D-063R / D-063SR ......................................................................................................... 600 W
Maximum Radiographic Exposure Time ..................................................................................... 10 s
Maximum Fluoroscopic Duration:
D-063 / D-063S ................................................................................................. 265 W-20minutes
D-063R / D-063SR ............................................................................................ 600 W-20minutes
-2-
D-063 / D-063R / D-063S / D-063SR
Environmental Limits
Operating Limits (in dielectric oil):
Oil Temperature ........................................................................................................... 10 ~ 60 ℃
Oil Pressure ............................................................................................................. 70 ~ 140 kPa
Shipping and Storage Limits:
Temperature .............................................................................................................. -40 ~ 70 ℃
Humidity ........................................................................................................................ 10 ~ 90 %
(No condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa
-3-
D-063 / D-063R / D-063S / D-063SR
Maximum Rating Charts
(Absolute maximum rating charts)
Constant Potential High-Voltage Generator
Nominal Focal Spot Value: 0.6
100kV 90kV
20 80kV-50kV
110kV
18 40kV
TUBE CURRENT [mA]
16
14
12
10
0.1 0.2 0.3 0.5 0.7 1 2 3 5 7 10
EXPOSURE TIME [s]
Filament & Emission Characteristics
20 4.5
40kV
80kV
16 4.0
FILAMENT VOLTAGE [V]
110kV
TUBE CURRENT [mA]
12 3.5
Ef
8 3.0
4 2.5
0 2.0
2.9 3.1 3.3 3.5 3.7
FILAMENT CURRENT [A]
Note: This graph indicates typical characteristics.
-4-
D-063 / D-063R / D-063S / D-063SR
Anode Heating / Cooling Curve
D-063 / D-063S
28
265W
25
HEAT STORAGE [kJ]
20
150W
15
10
COOLING
HEATING
0
0 100 200 300 400 500 600 700 800 900
TIME [s]
D-063R / D-063SR
35
600W
30
HEAT STORAGE [kJ]
25
300W
20
15
COOLING
10
100W
5
HEATING
0
0 60 120 180 240 300 360 420 480 540 600
TIME [s]
-5-
D-063 / D-063R / D-063S / D-063SR
Dimensional Outline of D-063
Unit: mm
CENTRAL X-RAY
φ16
M5 10DEEP M3
65±1
120±5
CENTRAL X-RAY
φ52MAX
FOCAL SPOT
TERMINAL SLEEVE
12MAX
C2 F 4 )
2 0.
(φ
N
CENTRAL X-RAY
SHORT PIN
C1 F
F :FILAMENT
C1,C2:CATHODE
N :NOT FOR USE
-6-
D-063 / D-063R / D-063S / D-063SR
Dimensional Outline of D-063R
Unit: mm
5
0
.
0
±
1
.
4
φ
-
4
CENTRAL X-RAY
1
.
0
±
8
2
φ
φ40
M5 10DEEP
RADIATOR
21
26
86±1
141±5
CENTRAL X-RAY
φ52MAX
FOCAL SPOT
TERMINAL SLEEVE
12MAX
C2 F )
.4
20
(φ
N
CENTRAL X-RAY
SHORT PIN
C1 F
F :FILAMENT
C1,C2:CATHODE
N :NOT FOR USE
-7-
D-063 / D-063R / D-063S / D-063SR
Dimensional Outline of D-063S
Unit: mm
CENTRAL X-RAY
φ16
M5 10DEEP M3
65±1
60
φ12
130±3
125
CENTRAL X-RAY
FOCAL SPOT
65
4MAX
TERMINAL SLEEVE φ60
C2 F )
.4
φ 20
(
N
CENTRAL X-RAY
SHORT PIN
C1 F
F :FILAMENT
C1,C2:CATHODE
N :NOT FOR USE
-8-
D-063 / D-063R / D-063S / D-063SR
Dimensional Outline of D-063SR
Unit: mm
5
0
.
0
±
1
.
4
φ
-
4
CENTRAL X-RAY
1
.
0
±
8
2
φ
φ40
M5 10DEEP
RADIATOR
21
26
86±1
60
φ12
151±3
125
CENTRAL X-RAY
FOCAL SPOT
65
4MAX
TERMINAL SLEEVE φ60
C2 F 4 )
2 0.
(φ
N
CENTRAL X-RAY
SHORT PIN
C1 F
F :FILAMENT
C1,C2:CATHODE
N :NOT FOR USE
-9-
D-063 / D-063R / D-063S / D-063SR
OVERSEAS SUBSIDIARIES AND AFFILIATES
EU REPRESENTATIVE
・TOSHIBA ELECTRONICS EUROPE GMBH
HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296-107 FAX +49 (211) 5296-402
For Sales & Technical Services, please contact the following representative:
・TOSHIBA ELECTRONICS EUROPE GMBH
HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296-107 FAX +49 (211) 5296-402
・TOSHIBA AMERICA ELECTRONIC COMPONENTS, INC.
2150 EAST LAKE COOK ROAD, SUITE 310
BUFFALO GROVE, ILLINOIS 60089 USA
PHONE +1 (847) 484-2400 FAX +1 (847) 541-7287
・TOSHIBA ELECTRON DEVICES & MATERIALS (SHANGHAI) CO., LTD. (TEMS)
RM1606, SH-PLAZA,
No.336, XIZANG ROAD (MIDDLE), SHANGHAI, 200001, CHINA
PHONE +86 (21) 6361-0077 FAX +86 (21) 6351-5760
(Product Development Engineering Department)
1385 SHIMOISHIGAMI, OTAWARA-SHI, TOCHIGI-KEN, 324-8550, JAPAN
PHONE: +81-287-26-6666 FAX: +81-287-26-6060
http://www.toshiba-tetd.co.jp/
・Toshiba Electron Tubes & Devices Co., Ltd. meets the Environmental Management System Standard, ISO14001.
・Toshiba Electron Tubes & Devices Co., Ltd. meets internationally recognized Standards for Quality Management System ISO9001, ISO13485.