Introduction To Digital Radiography
Introduction To Digital Radiography
Direct Conversion
Indirect Conversion
Radiography. This technique utilizes DDAs (Digital Detector Arrays) instead of Film or
CR (Computed Radiography) in order to create an instant Image. The Radiation
reaches the DDA, which has passed through the object, converted by
(1)a Scintillator into visible light in indirect methods or
(2)a photon detector into photoelectric effects in direct methods and
then translated into a digital Image. The physics (Angles, Penetration, technique etc.)
remain similar and only mild changes are required to make the transition to Digital
Radiography.
Instead of X-ray film, digital radiography uses a digital image capture device. This
gives advantages of immediate image preview and availability; elimination of
costly film processing steps; a wider dynamic range, which makes it more
forgiving for over- and under-exposure; as well as the ability to apply special
image processing techniques that enhance overall display quality of the image.
Flat-panel structure
2.Direct FPDs. Amorphous selenium (a-Se) FPDs are known as “direct” detectors
because X-ray photons are converted directly into charge. The outer layer of the flat
panel in this design is typically a high-voltage bias electrode. X-ray photons create
electron-hole pairs in a-Se, and the transit of these electrons and holes depends on
the potential of the bias voltage charge. As the holes are replaced with electrons, the
resultant charge pattern in the selenium layer is read out by a TFT array, active matrix
array, electrometer probes or microplasma line addressing.
As of 2008, many color LCD TVs and monitors use this technology. TFT panels are frequently
used in digital radiography applications in general radiography. A TFT is used in both direct and
indirect capture[jargon] as a base for the image receptor in medical radiography.
AMOLED (active-matrix organic light-emitting diode) screens also contain a TFT layer.
The most beneficial aspect[neutrality is disputed] of TFT technology is its use of a separate
transistor for each pixel on the display. Because each transistor is small, the amount of charge
needed to control it is also small. This allows for very fast re-drawing of the display.
https://en.wikipedia.org/wiki/Thin-film_transistor
Studies dealing with CCD-based digital general radiography are rare. Phantom studies have been conducted to
investigate slot-scan CCD systems and compare them with screen-film combinations and various digital
detectors. In all of these studies, CCD-based systems were comparable to flat-panel detectors in terms of
image quality and allowed slightly superior low-contrast visualization. Clinical studies performed with slot-scan
detectors are mainly concentrating on applications in mammography and digital dental radiography. The
performance of lens-coupled CCD systems is somewhat inferior to that of slot-scan systems because of their
technical principle, substantially lower quantum efficiency, and lower signal-to-noise ratio.
Drawing illustrates an amorphous silicon–based indirect conversion DR system. X-ray energy is converted into
visible light in a scintillator layer. The emitted light is then converted into electrical charges by an array of
silicon-based photodiodes and read out by a TFT array.
a) Conversion Efficiency X-ray quantum efficiency Light conversion per X-ray photon
b) MTF / spatial resolution
c) Light output spectrum
d) Light decay / after glow
Note:
X ray → a-Se (Charge) → TFT Array
One advantage of these systems is greater clinical usefulness, since the detectors can be mounted on thorax
stands and bucky tables. To date, there have been only a few clinical studies conducted with selenium-based
flat-panel detectors. However, these studies indicate that the image quality provided by selenium-based flat-
panel detectors is equivalent to that provided by other flat-panel detectors and selenium drum detectors.
Another promising clinical application of selenium-based flat-panel detectors is in the field of mammography
(33).
Amorphous selenium–based direct conversion DR systems. (a) Drawing illustrates a selenium drum–based
system. A rotating selenium-dotted drum with a positive electrical surface charge is exposed to x-rays. Alteration of
the charge pattern of the drum surface is proportional to the incident x-rays. The charge pattern is then converted
into a digital image by an analog-to-digital (A/D) converter. (b) Drawing illustrates a selenium-based flat-panel
detector system. Incident x-ray energy is directly converted into electrical charges within the fixed photo-conductor
layer and read out by a linked TFT array beneath the detective layer.
Note:
CR = Computed Radiography = Cassette Radiography
photostimulable
phosphor
Computed radiography (CR) is often distinguished from Direct Radiography (DR). CR and DR
have many similarities. Both CR and DR use a medium to capture x-ray energy and both produce
a digital image that can be enhanced for soft copy diagnosis or further review. Both CR and DR
can also present an image within seconds of exposure. CR generally involves the use of a
cassette that houses the imaging plate similar to traditional film-screen systems, whereas DR
typically captures the image directly onto a flat panel detector without the use of a cassette.
Image processing or enhancement can be applied on DR images as well as CR images due to
the digital format of each. There are many different types of DR detectors in use in medicine and
industry. Each type has its own merits and distinctions and may be applied to certain imaging
requirements based on these attributes
Vignetting
Abstract Developments in digital detector technologies have been taking place and
new digital technologies are available for clinical practice. This chapter is intended
to give a technical state-of-the-art overview about computed radiography (CR)
and digital radiography (DR) detectors. CR systems use storage-phosphor image
plates with a separate image readout process and DR technology converts X-rays
into electrical charges by means of a readout process using TFT arrays. Digital
detectors offer several advantages when compared to analogue detectors. The
knowledge about digital detector technology for use in plain radiograph exami-
nations is thus a fundamental topic to be acquired by radiology professionals and
students. In this chapter an overview of digital radiography systems (both CR
and DR) currently available for clinical practice is provided.
Introduction
Several digital systems are currently available for the acquisition of projection
radiographs. Digital radiography systems have been replacing traditional analogue
or screen–film (SF) systems over the last three decades. The transition from an SF
environment to a new digital environment should be considered as a complex
process. Technical factors concerning image acquisition, management of patient
dose, and diagnostic image quality are some issues that could influence this process.
In a transition process from SF to digital, patient radiation doses could increase
40–103% [1]. When compared to SF, digital technology could increase patient
radiation doses due to the wide dynamic range they have. However, the dynamic
range is useful because it contributes for a better clinical image quality when
compared to traditional SF systems [2]. This is an important difference among
analogical and digital technologies. The risk of overexposure with no adverse effect
on image quality could be present. Digital imaging systems could facilitate over- or
underexposure that influences a patient’s dose. Overexposure could provide good-
quality images, but may cause unnecessary patient dose. Although several
advantages over SF systems are identified, considerable variations in image quality
and effective dose can be achieved among different digital detectors [3].
According to Busch [4] the choice of the radiographic technique, the radiation
dose delivered to the patient, and the diagnostic quality of radiographic image are
three core aspects of the imaging process aiming the management of patient dose
and image quality. This is a challenge for radiographers because clinical advantages
and limitations of digital technologies for projection radiography are also depen-
dent on the radiographer’s options for a particular patient examination.
The knowledge about digital detector technology for use in plain radiograph
examinations is thus a fundamental issue to be acquired by radiology professionals
and students. Several literature reviews concerning digital radiology detectors have
been provided by some authors [5–11]. In this chapter an overview of computed
radiography (CR) and digital radiography (DR) currently available for clinical
practice is provided.
Developments in digital detector technologies have been taking place and new
digital technologies are available for clinical practice. Table 2.1 shows a timetable
of developments in digital technologies since the early 1980s.
The first digital radiography system using the basic principle of the conversion of
the X-ray energy into digital signals utilizing scanning laser stimulated lumines-
cence (SLSL) was developed by Fuji (Tokyo, Japan) and introduced in the market in
the beginning of the 1980s [12]. In the mid-1980s, the storage phosphor systems
Digital x-ray
Technologies
CR DR
Computed Digital Digital x-ray
Radiography Radiography Technologies
Storage Storage
Scintillator Photoconductor Photoconductor Scintillator
Detector Phosphors phosphor
properties CsI a:Se a:Se CsI 2+
BaFBr:Eu2+ BaFBr:Eu
became a new clinical application as a new imaging method for exposures at the wall
stand, the Bucky table, and bedside imaging. The high technical requirements and
financial costs, associated with limited image quality and difficult handling—with-
out a reduction of examination time—delayed the transfer of storage phosphor
systems into routine clinical use, which started to increase at the beginning of the
1990s [4]. Today the storage-phosphor radiography systems or CR systems play a
fundamental role in the field of digital projection radiography. Other important
innovation was the development of flat-panel detectors in the middle of 1995.
Flat-panel detectors were initially developed to be integrated detectors in the
radiology equipment, but more recently they are available as nonintegrated detectors
and working as a wireless or a non-wireless technology.
Digital systems are traditionally split into two broadly defined categories [10, 11]:
computed radiography and digital radiography. Although this taxonomy is commonly
accepted other classifications are described [13]: direct digital radiography and indirect
digital radiography technologies (including CR). In this case the detector classification
is related with the conversion process of X-ray energy to electric charge. Figure 2.1
shows a schematic figure that includes a comparative diagram of the taxonomy of
digital radiography technologies, the conversion process, and the detector properties.
Other taxonomic option is to give a classification according to the integration of
the digital detector within the radiology equipment: in this case integrated and
nonintegrated detectors terminology could be used.
Despite the taxonomy that is used the major difference among digital technology
systems related with X-ray detection and readout process. Concerning CR systems
they use storage-phosphor image plates with a separate image readout process, which
means an indirect conversion process; DR technology converts X-rays into electrical
charges by means of a direct readout process using thin-film transistor (TFT) arrays.
12 2 Digital Radiography Detectors: A Technical Overview
Table 2.2 shows the differences among detector technology concerning three
components of digital detectors [14]: the capture element, the coupling element,
and the charge readout element.
CR technology uses an indirect conversion process using a two-stage technique.
X-rays are captured at a storage-phosphor screen (SPS) (e.g.: BaFBr:Eu2+) and then
a photodetector captures the light emitted from the SPS and converts the captured
luminescence into a corresponding digital image.
DR detectors can use either a direct or an indirect process for converting X-rays
into electric charges. These detectors use direct-readout by means of a TFT array
despite the conversion process of the X-ray beam. Direct-conversion detectors have
an X-ray photoconductor—such as amorphous selenium (a-Se)—that converts
directly at only one stage X-ray photons into electric charges.
Indirect-conversion systems use a two-stage technique for conversion. They have
a scintillator, such as cesium iodide (CsI) that converts X-rays into visible light at a
first stage. That light is then converted—at a second stage—into an electric charge
by means of an amorphous silicon photodiode array [15].
Despite the process of X-ray detection and readout digital detectors offer several
advantages when compared to SF systems. These include wide dynamic range,
adjustable image processing, better image quality, rapid image acquisition, and
image access at remote locations [16].
Computed Radiography
Computed radiography was the first available digital technology for projection
radiography. CR technology is based in SPS and its first clinical application by
Fuji took place at the early 1980s.
This technology uses a photostimulable detector replacing the traditional SF
cassettes. The storage-phosphor plates are exposed inside the cassettes with stan-
dard dimensions for typical plain radiography and no change of generator, X-ray
tube, and Bucky wall or table mounted system is necessary. CR technology allows
the radiographer to obtain plain radiography images like in a traditional SF system.
Overview of Computed Radiography and Digital Radiography Detectors 13
a X-ray photon b
c d
Photostimulated
luminescence (PSL)
Laser beam
Fig. 2.2 SPS exposure and PSL; SPS storage-phosphor screens, PSL photostimulated luminescence
The difference is how the latent image is created and how this image processing
is done. The basic CR imaging cycle has three steps [13]: (1) expose, (2) readout,
and (3) erase.
Inside the radiography cassette an image plate (IP)—or SPS—having a detective
layer of photostimulable crystals is available. The detective layer consists of a
family of phosphors BaFX:Eu2+ where X can be any of the halogens Cl, Br, or I (or
an arbitrary mixture of them) [17]. A typical SPS can store a latent image for a
considerable period of time. However, according to the American Association of
Physicists in Medicine [18], it will lose about 25% of the stored signal between
10 min and 8 h after an exposure resulting in the loss of energy through spontaneous
phosphorescence.
The phosphor crystals are usually cast into plates into resin material in an
unstructured way (unstructured scintillators) [10]. When the SPS is exposed to
the X-ray the energy of the incident radiation is absorbed and excites electrons
to high-energy levels (Fig. 2.2a, b). These excited electrons remain trapped at
unstable energy levels of the atom. The absorbed X-ray energy is stored in crystal
structure of the phosphor and a latent image is then created at these high-energy
states giving a spatial distribution of these electrons at the SP detector. This trapped
energy can be released if stimulated by additional light energy of the proper
wavelength by the process of photostimulated luminescence (PSL) (Fig. 2.2) [18].
After the X-ray exposure and the creation of the latent image, the SPS is scanned in
a separate CR reader device. The readout is a process that follows exposure of the
image plate and constitutes the second step of the CR imaging cycle. A red laser beam
14 2 Digital Radiography Detectors: A Technical Overview
scans the photostimulable screen stimulating the emission of blue light photons under
the excitation of the laser beam. When the detective layer of the IP is scanned pixel by
pixel with a high-energy laser beam of a specific wavelength, stored energy is set free
as emitted light having a wavelength different from that of the laser beam [10]. This
triggers the process of PSL resulting in the emission of blue light in an amount
proportional to the original X-ray [17] and setting free the excited electrons to their
lower energy level (Fig. 2.2c, d). This light is collected by photodiodes and converted
into electric charge while an analog-to-digital device converts it into a corresponding
digital image. Figure 2.3 shows the SPS scanning process.
Finally the third step of the basic CR imaging cycle is the residual signal erasure.
Residual latent image electrons are still trapped on higher energy levels after
readout. This energy is erased after the readout process using a high-intensity
white light source that flushes the traps without reintroducing electrons from the
ground energy level [18].
Digital Radiography
TFT array
Photodiode or charge collector Collects charges from the upper
Converts x-rays to light layer
or to electric charges
Electronic control
Triggers the switching diodes
in
Switching diodes
Connects each pixel to readout device
out
Analog-to digital
Multiplexer conversion
Readout the electronic signal
TFT arrays (Fig. 2.5) are typically deposited onto a glass substrate in multiple
layers, with readout electronics at the lowest level, and charge collector arrays at
higher levels.
Depending on the type of detector being manufactured, charge collection
electrodes or light-sensing elements are deposited at the top layer of this “electronic
sandwich” [20].
The advantages of this design include compact size and immediate access to digital
images. The performance of DR systems greatly exceeds the performance of CR
systems, which have conversion efficiencies of 20–35%, and of screen–film systems
for chest radiography, which have nominal conversion efficiencies of 25% [20].
Wireless DR flat-panel systems have become commercially available by 2009.
Wireless DR systems are nonintegrated detectors that could be used to obtain
radiographs in a similar way to CR. With wireless DR detector it is mandatory to
use a wireless LAN for communications between the DR detector unit and the
workstation console. This way each performed radiograph is transferred at almost
real time from the cassette DR to the workstation. The DR cassette includes a built-
in battery to power supply and this allows the detector’s necessary autonomy to
obtain several radiographs and to transfer the obtained radiographs to the system for
further viewing.
Analog-to digital
Out conversion
In
Before the flat panel is exposed to X-rays an electric field is applied across the
selenium layer. Then the X-ray exposure generates electrons and holes within the
a-Se layer: the absorbed X-ray photons are transformed into electric charges and
drawn directly to the charge-collecting electrodes due to the electric field. Those
charges—proportional to the incident X-ray beam—are generated and migrate
vertically to both surfaces of the selenium layer, without much lateral diffusion.
At the bottom of the a-Se layer, charges are drawn to the TFT charge collector,
where they are stored until readout. The charge collected at each storage capacitor
is amplified and quantified to a digital code value for the corresponding pixel.
During the readout, the charge of the capacitors of every row is conducted by the
transistors to the amplifiers.
Summary 17
Summary
Different digital technologies are currently available for clinical practice in plain
radiography. CR and DR technologies constitute a remarkable improvement based
on detector technology developments. The specific properties and capabilities of a
digital detector influence the choice of the radiographic technique, the radiation dose
delivered to the patient, and the diagnostic quality of radiographic image. Although
SF and digital technology (CR and DR) coexist at the present time in many countries
the trends in the near future seem to point towards the digital technology.
18 2 Digital Radiography Detectors: A Technical Overview
References
1. Vaño E, Fernández JM, Ten JI, Prieto C, González L, Rodrı́guez R, de Las Heras H. Transition
from screen–film to digital radiography: Evolution of patient radiation doses at projection
radiography. Radiology. 2007;243:461–6.
2. Persliden J. Digital radiology and the radiological protection of the patient. Eur Radiol
Syllabus. 2004;14:50–8.
3. Pascoal A, Lawinsky CP, Mackenzie A, Tabakov S, Lewis CA. Chest radiography: a compari-
son of image quality and effective dose using four digital systems. Radiat Prot Dosimetry.
2005;114:273–7.
4. Busch HP. Image quality and dose management for digital radiography—final report. In:
DIMOND. 3rd ed. European Commission. Available at http://www.dimond3.org/European
(2004).
5. Schaefer-Prokop CM, De Boo DW, Uffmann M, Prokop M. DR and CR: recent advances in
technology. Eur J Radiol. 2009;72:194–201.
6. Lança L, Silva A. Digital radiography detectors—a technical overview: Part 1. Radiography.
2009;15:58–62.
7. Lança L, Silva A. Digital radiography detectors—a technical overview: Part 2. Radiography.
2009;15:134–8.
8. Uffmann M, Schaefer-Prokop C. Digital radiography: the balance between image quality and
required radiation dose. Eur J Radiol. 2009;72:202–8.
9. Williams MB, Krupinski EA, Strauss KJ, Breeden 3rd WK, Rzeszotarski MS, Applegate K,
Wyatt M, Bjork S, Seibert JA. Digital radiography image quality: image acquisition. J Am Coll
Radiol. 2007;4:371–88.
10. Körner M, Weber CH, Wirth S, Pfeifer KJ, Reiser MF, Treitl M. Advances in digital radiogra-
phy: physical principles and system overview. Radiographics. 2007;27:675–86.
11. Samei E, Seibert JA, Andriole K, Badano A, Crawford J, Reiner B, Flynn MJ, Chang P.
AAPM/RSNA tutorial on equipment selection: PACS equipment overview. Radiographics.
2004;24:313–34.
12. Sonoda M, Takano M, Miyahara J, Kato H. Computed radiography utilizing scanning laser
stimulated luminescence. Radiology. 1983;148:833–8.
13. Schaetzing R. Computed radiography technology. In: Samei E, Flynn MJ, editors. Advances in
digital radiography: RSNA categorical course in diagnostic radiology physics. Oak Brook, IL:
Radiological Society of North America; 2003. p. 7–21.
14. Samei E. Performance of digital radiographic detectors: factors affecting sharpness and noise.
In: Samei E, Flynn MJ, editors. Syllabus: advances in digital radiography-categorical course in
diagnostic radiology physics. Oak Brook, IL: Radiological Society of North America; 2003. p.
49–61.
15. Chotas HG, Dobbins III JT, Ravin CE. Principles of digital radiography with large-area,
electronically readable detectors: a review of the basics. Radiology. 1999;210:595–9.
16. Chotas H, Ravin C. Digital chest radiography with a solid-state flat-panel X-ray detector:
contrast-detail evaluation with processed images printed on film hard copy. Radiology.
2001;218:679–82.
17. Rowlands J. The physics of computed radiography. Phys Med Biol. 2002;47:R123–66.
18. American Association of Physicists in Medicine. Acceptance testing and quality control of
photostimulable storage phosphor imaging systems. In: Report of AAPM Task Group 10.
Available at http://www.aapm.org/pubs/reports/RPT_93.pdf (2006).
19. Kotter E, Langer M. Digital radiography with large-area flat-panel detectors. Eur Radiol.
2002;12:2562–70.
20. Culley JD, Powell GF, Gingold EL, Reith K. Digital radiography systems: an overview.
Available at http://www.hologic.com/oem/pdf/DROverviewR-007_Nov2000.pdf (2000).
References 19
21. Cha KB, Lee SJ, Muralidharan P, Kim DK, Kim JY, Cho G, Jeon S, Huh Y. Novel
nanocrystalline Gd2O3(Eu) scintillator screens with a micro-pixel structure for high spatial
resolution X-ray imaging. Nucl Instrum Meth Phys Res A. 2011;652:717–20.
22. Cha KB, Kim JY, Cho G, Seo CW, Jeon S, Huh Y. Quasi-pixel structured nanocrystalline
Gd2O3(Eu) scintillation screens and imaging performance for indirect X-ray imaging sensors.
Nucl Instrum Meth Phys Res A. 2011;648:S12–5.
http://www.springer.com/978-1-4614-5066-5
DR Images