LH10 - MarkII MANUAL Rev B 595431 For Emailing
LH10 - MarkII MANUAL Rev B 595431 For Emailing
LH10 - MarkII MANUAL Rev B 595431 For Emailing
As a result of the acquisition of Fusion UV by Heraeus Noblelight, the name of the company
changed from Fusion UV Systems Inc. to Heraeus Noblelight Fusion UV Inc.
© Copyright 2014 Heraeus Noblelight Fusion UV Inc.
All Rights Reserved.
This product is protected by the following US and foreign patents 5471109, 6509656, 6646384
B2, 6740892, 6908586 B2, 7037460 B2, 7055990 B2, 7407617, EP1354386, EP1468426A2,
as well as patents now pending, both US and foreign.
Heraeus Noblelight Fusion UV Inc. reserves the right to make changes without prior
announcement.
Table of Contents
PREFACE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiii
Document Conventions ........................................................................... xiii
Indications of Note, Caution, Warning, and Danger ................................. xiii
Icon Usage ........................................................................................ xiv
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
System Description .................................................................................. 1
Irradiator ............................................................................................ 2
Lamp Bulb ...................................................................................... 3
Bulb Types ...................................................................................... 3
Power Supply ...................................................................................... 3
Control Options ................................................................................... 4
System Components ................................................................................ 5
SAFETY .............................................. 7
Ultraviolet Radiation ................................................................................. 8
Microwave Radiation ................................................................................. 8
Personnel Effects ..................................................................................... 9
Radio Interference ................................................................................ 10
Ozone ................................................................................................ 10
Mercury Fill In UV Bulb ........................................................................... 10
Bulb Disposal .................................................................................... 10
Safety and Clean-up in Event of Bulb Breakage ...................................... 11
Temperature ......................................................................................... 11
High Voltage ......................................................................................... 12
UV Curable Materials .............................................................................. 12
MECHANICAL INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Customer Responsibilities ....................................................................... 13
Power Supply Mounting .......................................................................... 14
Irradiator Mounting ................................................................................ 14
UV Light Shielding .............................................................................. 16
Irradiator Cooling ............................................................................... 16
Remote Blowers ..................................................................................... 16
Differential Pressure Operation ................................................................ 17
Irradiator Exhaust Venting ...................................................................... 17
Quartz Plates ........................................................................................ 18
Lamps Mounted in Opposing Configuration ................................................ 19
RF Detector ........................................................................................... 20
ELECTRICAL INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Power Source ........................................................................................ 23
System Power Connection, J101 .............................................................. 25
Input Power Connection ...................................................................... 25
Rotating Phases ................................................................................. 26
Irradiator Cable Connections ................................................................... 27
RF Cable Connection, J105 ...................................................................... 28
Emergency Stop (E-STOP) Connection, J107 .............................................. 29
Master/Slave Connection, J102 ................................................................ 30
Customer Input/Output Connections, J106 ................................................ 32
Relay Outputs ................................................................................... 34
Remote Power Control ........................................................................ 35
Analog Power Control ..................................................................... 35
Digital Power Control ...................................................................... 35
Remote Inputs .................................................................................. 37
Customer Interlocks ........................................................................... 37
Remote Magnetron Current ................................................................. 39
Summary of LHP10 Mark II Connectors ..................................................... 39
Single Lamp Control in Remote Mode using Customer I/O ........................... 40
Multiple Lamp Control in Remote Mode using Dry-Contact I/O ...................... 40
Communication Bus Connection, J111 (Optional) ........................................ 42
Optional DeviceNet Bus Connection ...................................................... 42
Multiple Lamp Control in Remote Mode using DeviceNet .......................... 43
Optional Profibus Connection ............................................................... 46
Multiple Lamp Control in Remote Mode using Profibus ............................. 47
Optional EtherNet/IP Connection .......................................................... 50
Multiple Lamp Control in Remote Mode using EtherNet/IP ........................ 50
Blower Connection, J108 (Optional) .......................................................... 53
Electrical Installation Checklist ................................................................. 53
CONFIGURATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Navigation Guide ................................................................................... 55
Startup/Initializing ................................................................................. 56
Entering the Configuration Menu ............................................................. 57
Download/Upload through USB Port ......................................................... 57
Read Configuration from USB .............................................................. 58
Save Configuration to USB .................................................................. 59
Save Fault Data to USB ...................................................................... 61
Update Engine or CPU Firmware ........................................................... 63
OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Front Panel Indicators/Controls ................................................................ 83
Lamp Disable Switch .......................................................................... 84
Lamp Off (OFF) LED ........................................................................... 84
Standby (S/B) LED ............................................................................. 84
Lamp On (ON) LED ............................................................................ 84
Master LED ....................................................................................... 85
Power Level Arrow Keys ...................................................................... 85
Fault State ........................................................................................ 85
Operating the Light Hammer 10 Mark II with the Front Panel Switches .......... 85
Single Unit Operation with Front Panel Switches ..................................... 85
Multiple Unit Operation with Front Panel Switches ................................... 86
Operating the Light Hammer 10 Mark II Remotely ...................................... 86
Single Unit Remote Operation .............................................................. 86
Multiple Unit Remote Operation ............................................................ 87
Reduced Power Operation ....................................................................... 88
Quick Restart Mode ................................................................................ 88
ROUTINE MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Safety Interlock Tests ............................................................................. 91
Power Supply Routine Maintenance .......................................................... 91
Air Filter Replacement ........................................................................ 92
Irradiator Routine Maintenance ................................................................ 93
Bulb, Screen, and Reflector Maintenance ............................................... 93
Bulb Cleaning .................................................................................... 94
Reflector Cleaning .............................................................................. 94
Screen Cleaning ................................................................................. 94
DIAGNOSTICS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Navigating the View Data/Diagnostic Menu ................................................ 95
Data View Mode ..................................................................................... 96
Display Parameters ............................................................................ 97
Enter Diagnostic Mode ........................................................................ 99
Display PS and Modules SN ................................................................100
TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Basic Checks ........................................................................................108
Communication Module Status LEDs ....................................................108
DeviceNet Module Status LEDs ........................................................109
EtherNet/IP Module Status LEDs .....................................................110
Profibus Module Status LEDs ..........................................................111
Parameter Values .................................................................................111
Troubleshooting Guide for the Light Hammer 10 Mark II ............................113
Determining the Likely Cause of a Fault ...............................................113
Fault Tables .....................................................................................113
Troubleshooting Guide for the Legacy Operation of the Light Hammer 10
Mark II .............................................................................................141
UNPACKING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187
Returned Material Authorization Number ..............................................187
Packing Materials ..............................................................................187
WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
Warranty Policy ....................................................................................189
Return Shipping Instructions ..................................................................190
Returned Material Authorization Number ..............................................190
General Packing Instructions ..............................................................190
Power Supply ...................................................................................190
Irradiators .......................................................................................190
Magnetrons ......................................................................................191
UV Bulbs ..........................................................................................191
Control Cards ...................................................................................191
RF Detectors ....................................................................................191
Waste Electrical and Electronic Equipment (WEEE) Recommendations ..........191
UV Bulbs ..........................................................................................192
Reflectors ........................................................................................192
Magnetrons ......................................................................................192
Printed Circuit Boards ........................................................................192
Conveyor Belts .................................................................................192
CERTIFICATES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201
System Designation ..............................................................................201
System Operating/Environmental Requirements .......................................201
Agency Approvals .................................................................................201
Irradiator .............................................................................................202
Blower Specifications .............................................................................204
Power Supply .......................................................................................205
DeviceNet Communications Bus .............................................................206
RF Detector ..........................................................................................206
SCHEMATICS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 213
GLOSSARY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 215
INDEX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 217
List of Figures
Figure 1: Light Hammer 10 Mark II System ........................................................ 1
Figure 2: Vertical Mounting Design for LHI10-series Irradiators............................ 15
Figure 3: Irradiator Conversion -- Internal Air Pressure Switch Differential Mode .... 17
Figure 4: Quartz Plates - Example Mounting Design ........................................... 18
Figure 5: Quartz Plate Mounting Distances ........................................................ 19
Figure 6: Lamps in Opposing Configuration ....................................................... 20
Figure 7: RF Detector..................................................................................... 21
Figure 8: Front and Rear Panels of LHP10 Mark II .............................................. 24
Figure 9: Mark II Input Power Connections (shown with cover removed)............... 26
Figure 10: LHP10 Mark II Rear Panel Connectors ................................................. 27
Figure 11: J107 E-Stop Jumper Connection......................................................... 30
Figure 12: Master/Slave Connections ................................................................. 31
Figure 13: Input Connections (from internal power source) ................................... 32
Figure 14: Input Connections (from external power source) .................................. 33
Figure 15: Asserted Input (Closed Contact) ........................................................ 36
Figure 16: Digital Power Control (using internal power sinking input) ..................... 36
Figure 17: Single Lamp Connections .................................................................. 40
Figure 18: Single Row Multi-Lamp Connections via Master/Slave ........................... 41
Figure 19: Dual DeviceNet Connector................................................................. 42
Figure 20: Pinout for DeviceNet Connector Male M12 ........................................... 43
Figure 21: Single Row Multi-Lamp Connections via DeviceNet................................ 44
Figure 22: Single Row Multi-Lamp Daisy-chain Connections via DeviceNet .............. 45
Figure 23: Profibus Connector ........................................................................... 46
Figure 24: Single Row Multi-Lamp Connections via Profibus .................................. 48
Figure 25: Single Row Multi-Lamp Daisy-chain Connections via Profibus ................. 49
Figure 26: EtherNet/IP Connector...................................................................... 50
Figure 27: Single Row Multi-Lamp Connections via EtherNet/IP ............................. 51
Figure 28: Single Row Multi-Lamp Daisy-chain Connections via EtherNet/IP ............ 52
Figure 29: LHP10 Mark II Front Panel Indicators .................................................. 83
Figure 30: Front Panel of LHP10 Mark II Power Supply ........................................ 108
Figure 31: LHI10 and LHI10B Mark II Irradiators ................................................ 147
Figure 32: Screen Screw Tightening Sequence ................................................... 149
Figure 33: Reflector Assembly.......................................................................... 149
Figure 34: Removing or Replacing the Bulb........................................................ 150
Figure 35: “209” End Reflector......................................................................... 151
Figure 36: Removing or Replacing the Reflector ................................................. 152
Figure 37: Gasket Installation .......................................................................... 152
Figure 38: Reflector Screw Tightening Sequence ................................................ 153
Figure 39: Correct Reflector Alignment.............................................................. 154
List of Tables
Table 1: Icons ............................................................................................. xiv
Table 2: Irradiator Models ............................................................................... 2
Table 3: Bulb Types........................................................................................ 3
Table 4: Power Supply Models.......................................................................... 3
Table 5: Power Supply Communication Bus Options ............................................ 5
Table 6: Maximum Line Currents .................................................................... 25
Table 7: Leg Currents 50 and 60 Hz Nominal Value, non-blower unit ................... 26
Table 8: Leg Currents 50 and 60 Hz Nominal Value, blower control module
installed ......................................................................................... 27
Table 9: RF Detector Connector, J105 ............................................................. 28
Table 10: E-STOP Cable Connection Lengths...................................................... 29
Table 11: Internal Power Source Connections .................................................... 33
Table 12: External Power Source Connections .................................................... 33
Table 13: System Status Output Relays ............................................................ 34
Table 14: Analog Inputs for Power Level Control Connections, J106 ...................... 35
Table 15: Digital Inputs for Power Level Control Connections, J106 ....................... 35
Table 16: J106 Inputs (Remote Operation) ........................................................ 37
Table 17: Customer Interlocks ......................................................................... 38
Table 18: Magnetron Current Monitor ............................................................... 39
Table 19: Summary of Connectors ................................................................... 39
Table 20: DeviceNet Connections ..................................................................... 42
Table 21: Profibus Connections ........................................................................ 46
Table 22: Front Panel Indicators....................................................................... 84
Table 23: Operation with Remote Controls......................................................... 87
Table 24: DeviceNet Network Status LED ......................................................... 109
Table 25: DeviceNet Module Status LED ........................................................... 109
Table 26: EtherNet/IP Network Status LED ....................................................... 110
Table 27: EtherNet/IP Module Status LED......................................................... 110
Table 28: Profibus Operation Mode LED............................................................ 111
Table 29: Profibus Status LED......................................................................... 111
Table 30: Parameter Value Ranges .................................................................. 112
Table 31: Common Faults .............................................................................. 114
Table 32: Other Faults ................................................................................... 120
Table 33: Lamp System Maintenance Log ......................................................... 145
Table 34: Cooling Requirements...................................................................... 164
Table 35: Light Hammer 10 Mark II System Component Designations .................. 201
Table 36: Light Hammer 10 Mark II System Component Designations .................. 201
Table 37: Test Standards ............................................................................... 201
Table 38: Irradiator Model LHI10-series Specifications ....................................... 202
Document Conventions
The products and components discussed in this document are
designed and tested for safe operation; however, they use
technologies and resources that have inherent risks. This section
describes the conventions used throughout this manual when
discussing such technologies, so that users can plan and act
accordingly when installing, operating, or troubleshooting
equipment.
Icon Usage
Several icons are used throughout the documentation to identify
the types of hazardous resources or technologies used when
working with a particular aspect of the system.
Table 1 lists these icons and their definitions. Exercise extreme
care by reading the material thoroughly and following all
procedures when working with the various components.
Table 1: Icons
Icon Definition
CAUTION
Table 1: Icons
Icon Definition
CAUTION: Ozone
Table 1: Icons
Icon Definition
System Description
The Light Hammer 10 Mark II UV Lamp System from Heraeus
Noblelight Fusion UV Inc. consists of a lamp module (irradiator),
a power supply unit, and the associated interconnecting cables.
Irradiator
The LHI10-series irradiator is the heart of the UV system. It
contains both the 254 mm (10 inch) long electrodeless lamp bulb
mounted in the elliptical reflector and the magnetrons which
energizes the bulb. The reflector is elliptical in cross-section and
the lamp bulb is mounted at the focus of the ellipse, producing
an intense strip of light at the other focus of the reflector 53.3
mm (2.1 inches) below the bottom rails of the lamp. The LHI10
irradiator operates at 600 W/inch.
Irradiator
Connectors Air Delivery
Model
Lamp Bulb
The lamp bulb is made of quartz and has a tapered shape which
optimizes its UV, thermal, and microwave properties. The short
quartz stubs at either end of the 254 mm (10 inch) bulb provide
mechanical support for quick mounting into spring-loaded
receptacles at the cavity ends. They are not electrodes and have
no electrical function.
Bulb Types
Many UV-curable formulations respond to different regions of the
UV Spectrum. Depending on the components of the formulation
and the application, bulbs whose UV output is strongest in the
specific wavelength range of interest may be an appropriate
choice.
Power Supply
Each lamp system has its own DC power supply. The power
supply unit houses the high voltage circuitry necessary to
energize the magnetrons, as well as all control and interlock
electronics, and the built-in cooling fans.
The high voltage circuit inside the power supply consists of two
solid state switching engines, which supply a regulated source of
power (approximately -5000V, 890 mA DC) to the associated
magnetron within the lamp.
Control Options
There are five different control options available with the Light
Hammer 10 Mark II system. These options include:
• DeviceNet™ Legacy
• DeviceNet™ Mark II
• Profibus®
• EtherNet/IP™
System Components
The basic Light Hammer 10 Mark II system consists of the
following items:
• An irradiator, LHI10-series.
• Filtered cooling air for the irradiator if the Blower module is not
installed. Refer to SPECIFICATIONS for the cooling
requirements and specifications. Refer to the IRRADIATOR
MAINTENANCE section for air flow and pressure requirements.
Ultraviolet Radiation
Ultraviolet radiation, which is emitted during normal operation
of the Heraeus Ultraviolet Lamp System, can be dangerous to
the eyes and skin of personnel. Adequate shielding around the
UV system and the region to be irradiated must be provided. If
unsafe levels of UV are allowed to escape during operation,
then all personnel in the vicinity of the lamps should be
required to wear goggles with UV-certified lenses. Protection is
needed for indirect as well as direct eye exposure. Discomfort
from excessive eye exposure to ultraviolet light typically
occurs about six hours after exposure. Personnel who
experience eye pain after possible exposure to direct rays from
the UV lamp should see a doctor. Furthermore, in the event
that adequate shielding is not possible, all personnel within
three meters of the lamp should wear gloves and long-sleeved
shirts to protect hands and arms.
Microwave Radiation
The Heraeus UV Lamp System is powered by high frequency
microwave energy. This form of energy is identical to that used
in home microwave ovens and, as in the case of ovens, can be
dangerous if misused or inadequately shielded. In the Heraeus
UV lamp, the shielding is adequate only as long as the lamp
and the screen on the bottom of the lamp are intact. Any rip or
large hole in the screen may lead to microwave radiation
leakage in dangerous amounts. The power to the lamp is
interlocked to shut off if there is excessive microwave radiation
leakage. The lamp system should never be turned on if the
lamp housing and screen are not intact, or if the microwave
leak detector and interlock are not functioning. Heraeus UV
equipment is interlocked to shut down if microwave leakage in
excess of 5 mW/cm2 (milliwatts per square centimeter) is
detected.
Personnel Effects
OSHA (U.S. Department of Labor, Occupational Safety, and
Health Administration - Standard 29CFR 1910.97) and ANSI
(American National Standards Institute - Standard C95.1-1999)
are the only major national organizations with voluntary
guidelines for safe limits of occupational microwave radiation
exposure.
Radio Interference
Heraeus UV lamps are classified as non-consumer industrial,
scientific and medical (ISM) equipment, as defined in Federal
Communication Commission (FCC) Rules and Regulations
Volume 47, Part 18. As required by these rules, Heraeus
Noblelight Fusion UV Inc. verifies that their systems are
capable of compliance with applicable technical standards
governing radiated emissions when the equipment is properly
maintained and is installed in an appropriate light shield. This
system complies with EN 55011 (CISPR-11, Group 2, Class A).
Ozone
Ozone is a gaseous form of oxygen, which is formed by
ultraviolet light. It has a characteristic pungent odor to which
most people are quite sensitive. People can typically detect
concentrations of several parts per hundred million. At high
concentrations it can cause discomfort and, at sufficiently high
levels, it can be dangerous.
Temperature
Surface temperatures of the lamp during normal operation
may exceed 49°C (120°F). Metal components near the lamp
may also be hot. Before attempting service procedures on the
lamp, allow the unit to operate in STANDBY so the blower
continues to run. The bulb will cool completely in less than one
minute. If the blower is not allowed to run, the bulb will retain
heat for several minutes. Observe appropriate precautions to
avoid burns.
High Voltage
Insulated electrical cables carry power at -5000 volts from the
power supply to the lamp. These voltages can be dangerous, so
some precautions should be taken in both the location and
protection of these lines. High voltages are exposed within the
power supply chassis, which should never be operated without its
cover.
UV Curable Materials
Please consult your UV materials supplier for recommended
precautions regarding the proper handling and use of these
products.
Customer Responsibilities
The basic system components of the Light Hammer 10 Mark II
system are shown in Figure 1. In addition to the components
shown, the user must supply the following items or purchase
them from Heraeus Noblelight Fusion UV Inc.:
Use the lift flanges on the front and rear of the chassis as the
lifting points for the power supply.
Irradiator Mounting
The lamp can be mounted over a moving substrate or web,
supported by the bottom rails. The mounting structure must,
however, provide UV shielding and adequate venting for the
cooling air. Each user will normally design the mounting structure
to suit the particular application, and the user is responsible for
providing the required shielding and ventilation.
The lamp should be mounted so that the bottom rails sit 53.3 mm
(2.1 inches) above the surface to be cured. This allows the
elliptical reflector to focus the most intense strip of light very
near the substrate. The lamp also has four threaded holes on
either side for mounting. Refer to OUTLINE DRAWINGS for the
mounting dimensions.
NOTE: Carefully grasp the sides of the irradiator to lift it into or
out of the lightshield/mounting structure. When lifting or moving
the irradiator, take care not to damage the fine mesh of the RF
screen.
PRESSURE
TEST POINT
LHI10 Series
Irradiator
MAKEUP
EXHAUST AIR
53.3mm
(2.1")
PRODUCT CL TRAVEL
NOTES: 1. Refer to Irradiator chapter for air pressure and flow verification procedure.
2. All pressures and flows measured at standard air density: 1.20kg/m3 (0.075 lbs/ft3).
3. Supply and exhaust blower specifications must be corrected for air density
variations at customer site.
UV Light Shielding
Adequate light shielding must be provided for each irradiator in
the system. If shielding is not used, or if UV light escapes around
the irradiator support, the operator’s eyes and skin must be
protected from the effects of UV. Refer to the SAFETY section.
The louvers that provide make-up air to cool the light shield must
be of a light shielding design. (For important information on UV
safety considerations, refer to the SAFETY section.)
Irradiator Cooling
Proper operation of the lamp requires adequate cooling of the
magnetron and the bulb. Refer to the IRRADIATOR
MAINTENANCE section for instructions on how to measure air
pressure within the irradiator.
Filtered air from a blower is fed into the lamp housing, over the
magnetrons, and into the cavity through the microwave slots and
an array of small cooling holes. It then passes out of the lamp
through the wire mesh screen at the bottom and is exhausted.
The downward air flow through the cavity serves to cool the lamp
bulb. In addition, the outward flow of filtered cooling air helps
keep the lamp and reflector clean which extends the time
between necessary cleanings of the lamp bulb and reflector.
NOTE: Refer to Table 39 in the SPECIFICATIONS section for
HVAC and operating requirements for the air specifications. If
a quartz plate assembly is used, refer to the cooling
requirements for this option.
Remote Blowers
A remote blower package is available from Heraeus Noblelight
Fusion UV Inc. The package includes an air distribution plenum
(for multi-lamp systems), an inlet filter box, and flexible ducts or
hoses for cooling the system. Diameter and length must be
specified. Excessive length of flexible hose will impede air flow to
the irradiator. If the blower is farther than a few feet (or a meter)
from the irradiator, smooth duct work with no sharp bends
should be used.
NOTE: When sizing a blower, duct losses, site altitude, and
filter losses must be considered to ensure that adequate
capacity is specified. Heraeus’ Air Handling Manual may be
used as a guide.
NOTE: The more often the lamp system cooling air filter is
replaced, the less frequent the bulb, screen, and reflector will
need to be cleaned.
Irradiator Pressure
Test Point
Air
Pressure
Silicone Switch
Air Tubing
Pressure 187 ID, Silicone
Switch 0.312 OD Tubing
187 ID,
0.312 OD
Lightshield
LHI10/LHI10_02 LHI10_01/LHI10_03
Figure 3: Irradiator Conversion -- Internal Air Pressure Switch
Differential Mode
Quartz Plates
To protect the material being cured from the warm air blowing
out of the irradiator, quartz plates may be mounted below the UV
lamp to block the air carrying heat away from the bulb. A typical
mounting design is shown in Figure 4; the typical mounting
distances are shown in Figure 5. Details of the mounting
structure depend on the position and location of the irradiator,
quartz plate, and the UV shielding.
SPECIAL NOTE FOR LHI10B Mark II IRRADIATORS: The
LHI10B Mark II should be used in lightshields with
unrestricted air flow. When a quartz plate assembly is used
below the irradiator, a negative pressure exhaust system is
required that eliminates back pressure at the base of the
irradiator.
Irradiator
Irradiator
mounting
structure
support
53 mm
(2.1”) D
Irradiator
Removal access cooling
for cleaning air flow Substrate
to be cured
Quartz plate
mounting structure
Irradiator
Quartz Plate
Assembly
53 mm D
Substrate to be
(2.1”)
C Cured
RF Detector
The Light Hammer 10 Mark II UV Lamp System includes an RF
detection device, which shuts the system down after 60 seconds
if dangerous levels of RF energy are present. During the sixty
seconds, a warning is provided over the communication bus. A
customer-supplied controller could be programmed to shut down
the entire system or shut down the unit under fault if microwave
energy in excess of permissible levels are detected. This
shutdown might be caused by a faulty RF screen or some other
type of damage to the lamp or by inadequate or damaged
shielding.
Figure 7: RF Detector
The presence of metal surfaces near the detector may cause false
readings; therefore, allow a maximum open area around the
device. Any hole in the light shield should be covered with
nonconducting silicone rubber.
NOTE: Ensure that the metal case of the RF detector does not
come in contact with any grounded surface. The mounting
bracket is isolated from the case of the RF detector.
ITEM SECTION
Power Source
Heraeus Noblelight Fusion UV Inc. UV equipment must be
connected in accordance with local wiring codes. Use applicable
standards for wire sizes, fuse sizes, and disconnect box locations.
Each power supply requires a separate fused three-branch circuit
with a line disconnect.
Current
Current (amps)
Voltage (amps)
System Frequency blower control
(VAC) non-blower
module installed
unit
For each power supply, remove the two J101 cover screws to
access the input power terminal block connector. Connect the
three-phase wiring and ground to the terminal block as follows:
phase 1 to L1, phase 2 to L2, and phase 3 to L3. Connect the
green or green/yellow ground wire to the dedicated ground
terminal for the chassis ground connection.
Secure the cable using the provided strain relief and re-install the
cover. All conduit and wire sizes should be in compliance with
local electrical code.
Input Power wired directly into Terminal Block Optional Pigtail Adapter connected
Rotating Phases
The Light Hammer 10 Mark II system with the LHI10 irradiator
provides a balanced HV load to the 3-phase power source. There
is no need to rotate phases. Table 7 shows the nominal leg
currents for the LHP10 Mark II power supply units without the
blower control module installed. Table 8 shows the nominal leg
currents for the LHP10 Mark II power supply units with the
blower control module installed.
J108 BLOWER
(optional)
J107
E-STOP
J105 FAULT
INTERLOCK COMMENTS
PINS CONDITION
NOTE: Heraeus recommends that the dry contacts used for the
E-STOP circuit be gold-plated. The total external wiring and
contact resistance for this connection must be less than 1Ω.
18 10 30
16 15 45
14 25 75
Input To LHI10
J101
Power Lamp
J103
Terminating Resistor
J104 Female
J105
RF Detector
J107
E-Stop Jumpers
J108
F M F M
J102
F standard
J106 Customer I/O Master/Slave DeviceNet
Cable Tee
MASTER
Input To LHI10
J101
Power Lamp
J103
J104
J105
RF Detector
J107
J108 standard
F M F M DeviceNet
J102 cable
F
J106 Customer I/O
Master/Slave
Cable
SLAVE #1
All connections
same for other
Slave systems
Input To LHI10
J101
Power Lamp
J103
J104
J105
RF Detector
J107
J108
F M F M
J102
F
J106 Customer I/O
Master/Slave
Cable
SLAVE #15
Terminating Resistor
Male
PLC
CONTACT
29 INTERNAL PS ( + ) 29 INTERNAL PS ( + )
EXTERNAL
28 USER POWER
JUMPER
INPUT
+
_
INPUT 28
EXTERNAL USER COMMON
PLC INTERNAL PS ( - ) JUMPER INTERNAL PS ( - )
CONTACT
27, 35, 37 27, 35, 37
J106 J106
Internal System Power 29 Output 5mA @ 14 Vdc When using this configuration, this
power is supplied internally and a
jumper from pin #28 to pin #29 is
required.
29 N/C 29 N/C
USER INPUT
28 POWER
PLC
EXTERNAL CONTACT
EXTERNAL
24V 24V
INPUT 28
USER COMMON
User Power (user 28 Input 10mA @ 24 Vdc User provides 12-24 Vdc to this pin.
provided)
Relay Outputs
The 37-pin Customer I/O connector (J106) is used to remotely
control the Light Hammer 10 Mark II unit. The status relay
outputs may be used to synchronize the customer’s equipment
with the Light Hammer 10 Mark II system. The unit can be
controlled through the use of dry contacts, analog, or digital
control signals.
K1 (Filament ON) 1, 2 Closed 1A@ 16-24 AWG K1 contacts close when filament
24 Vdc or 24 VAC voltage is applied.
K5 (System/ 10, 11 Closed 1A@ 16-24 AWG K5 contacts close when the lamp
Ready) 24 Vdc or 24 VAC reaches full UV output and the
user process can start.
K6 (Warning) 12, 13 Open 1A@ 16-24 AWG K6 contacts open when unit
24 Vdc or 24 VAC warning occurs.
0-10 VDC Mode: This input allows a user to control lamp power
in 1 percent steps using the 0-10 VDC control loop.
ANALOG
J106 MAX. CIRCUIT
POWER I/O TYPE COMMENTS
PIN RATINGS
CONTROL
DIGITAL
J106
POWER COMMENTS
PIN
CONTROL
J106
29 INTERNAL PS ( + )
JUMPER 28 USER POWER
23 PWR 0
24 PWR 1
25 PWR 2
26 PWR 3
27 INTERNAL PS ( - )
Remote Inputs
Three switches are required for remote Light Hammer 10 Mark II
operation; they control the RESET, STANDBY, and LAMP ON
functions. Connections are made through the J106 connector.
Single or Master/Slave systems may be operated remotely. For
Master/Slave systems, the remote switches must be connected
to the Master power supply. Refer to Table 16.
The remote Lamp Off (Reset), Standby, and Lamp On inputs are
only enabled when the user selects the remote control function.
The Lamp Off (Reset) always takes precedence over the Lamp On
input. Unused remote control inputs should not be jumpered.
Customer Interlocks
The LHP10 Mark II power supply has two customer interlock
inputs (External interlock and System Blower interlock). A fault
condition on either of these inputs will cause the system to shut
down; the fault must be corrected before the system can be
restarted. These interlock inputs are optional; however, the
system blower and external interlock inputs in the J106
connector must be jumpered if these inputs are not used. Refer
to Table 17 for more information on the Customer Interlocks.
MAX.
J106
SIGNAL I/O TYPE CIRCUIT COMMENTS
PINS
RATINGS
Cooling Air
Irradiator
Cable J10
J11
J101 3-phase
J103
Power Input
J104
J105
RF Detector
J107 E-Stop
Jumper Optional J12
J108
Blower Control
J102 MASTER/SLAVE
CONNECTION
J106
Master/Slave
Cable
Cooling Air
Irradiator
Cable J10
J11
J101
3-phase
Power Input
J103
J104
Terminating Resistor
J105
RF Detector
J107 E-Stop
Jumper Optional J12 Master/Slave
J108
Cable
Blower Control
J102
J106
J101 3-phase
Power Input
J103
J104 To Lamp
J105
RF Detector
J107 E-Stop Jumper
J101 3-phase
Power Input
J103
J104 To Lamp
J105
RF Detector
J107 E-Stop Jumper
Terminating Resistor
NS MS
1 Drain Bare
2 V+ Red
3 V- Black
4 CAN_H White
5 CAN_L Blue
2 1
3 4
Male (pins)
Terminating Resistor
Female
DeviceNet
Drop Cable
M F M F M
Input To LHI10
J101 F
Power Lamp
J103
J104
J105
RF Detector
J107
E-Stop Jumpers
J108
Communication Bus
J102
F M F M
J111
F standard
J106 Customer I/O DeviceNet
DeviceNet
Cable Tee
SLAVE #1
Input To LHI10
J101
Power Lamp
J103
J104
J105
RF Detector
J107
J108
J102 standard
F M F M DeviceNet
J111 cable
F
J106 Customer I/O
DeviceNet
Cable
SLAVE #2
All connections
same for other
Slave systems
Input To LHI10
J101 Power Lamp
J103
J104
J105
RF Detector
J107
J108
J102
F M F M
J111
F
J106 Customer I/O
DeviceNet
Cable
SLAVE #15
Terminating Resistor
Male
Input To LHI10
J101
Power Lamp
J103
J104
J105
RF Detector
J107
E-Stop Jumpers
J108
Communication Bus
J102
J111 Terminating Resistor
J106
Customer I/O standard
SLAVE #1 DeviceNet
cable
Input To LHI10
J101
Power Lamp
J103 Terminating Resistor
J104
J105
RF Detector
J107
J108 DeviceNet
Tee
J102
J111
standard
DeviceNet
J106 cable
Customer I/O
SLAVE #2
All connections
same for other
Slave systems DeviceNet
Drop Cable
Input To LHI10
J101 Power Lamp
J103
J104
J105
RF Detector
J107
J108
J102
J111
SLAVE #15
Figure 22: Single Row Multi-Lamp Daisy-chain Connections via
DeviceNet
OP ST
5 Shield Shield
Terminating Resistor
Female
DeviceNet
Drop Cable
M F M F M
Input To LHI10
J101 F
Power Lamp
J103
J104
J105
RF Detector
J107
E-Stop Jumpers
J108
Communication Bus
J102
F M F M
J111
F Profibus
J106 Customer I/O Tee
Profibus
Cable
SLAVE #1
Input To LHI10
J101
Power Lamp
J103
J104
J105
RF Detector
J107
J108
J102
M Profibus
F M F Cable
J111
F
J106 Customer I/O
Profibus
Cable
SLAVE #2
All connections
same for other
Slave systems
Input To LHI10
J101 Power Lamp
J103
J104
J105
RF Detector
J107
J108
J102
F M F M
J111 Profibus
Tee
F
J106 Customer I/O
Profibus
Cable
SLAVE #15
Terminating Resistor
Male
Input To LHI10
J101
Power Lamp
J103
J104
J105
RF Detector
J107
E-Stop Jumpers
J108
Communication Bus
J102
J111 Profibus Terminating Resistor
J106
Customer I/O
Profibus
SLAVE #1 Cable
Input To LHI10
J101
Power Lamp Profibus
J103 Terminating Resistor
J104
J105
RF Detector
J107
Profibus
J108
Tee
J102
J111
J106
Profibus
Customer I/O Cable
SLAVE #2
All connections
same for other
Slave systems Profibus
Cable
Input To LHI10
J101 Power Lamp
J103
J104
J105
RF Detector
J107
J108
J102
J111
SLAVE #15
Figure 25: Single Row Multi-Lamp Daisy-chain Connections via
Profibus
NS MS
Input To LHI10
J101
Power Lamp
J103
J104
J105
RF Detector
J107
E-Stop Jumpers
J108
Communication Bus
J102
J111
Input To LHI10
J101 shielded
Power Lamp
J103 CAT6
Cable
J104
J105
RF Detector
J107
J108
J102
J111
All connections
same for other
Slave systems
Input To LHI10
J101 Power Lamp
J103
J104
J105
RF Detector
J107
J108
J102
J111
Input To LHI10
J101
Power Lamp
J103
J104
J105
RF Detector
J107
E-Stop Jumpers
J108
J102 Communication Bus
J111 shielded
CAT6
J106 Cable
Customer I/O
SLAVE #1
Input To LHI10
J101
Power Lamp
J103
100 meter maximum
J104
without Switch Box
J105 and with use of CAT6
RF Detector double-shielded cable
J107
J108
J102
J111
shielded
J106 CAT6
Cable
Customer I/O
SLAVE #2
All connections
same for other
Slave systems
Input To LHI10
J101 Power Lamp
J103
J104
J105
RF Detector
J107
J108 shielded
CAT6
J102
Cable
J111
SLAVE #15
Figure 28: Single Row Multi-Lamp Daisy-chain Connections via
EtherNet/IP
ITEM SECTION
• EtherNet/IP
• Profibus
NOTE: The Master/Slave control method is the default method
and is available on all units.
Navigation Guide
This section of the Manual is organized in the following manner:
Startup/Initializing
After the unit is powered, the following items are displayed.
Use the arrow keys (>or?) on the front SLAVE ADDRESS RANGE: 1 to 15
panel until the desired number default = 1
displays. Press and release the ON
button to select the displayed Slave
Address.
19. Using the arrow keys (>or?), select
Front Panel Lock when it displays.
Configure DeviceNet
The DeviceNet communication method requires a plug-in module
and proper configuration to operate. Contact Heraeus Noblelight
Fusion UV Inc. Technical Services if DeviceNet is required for
your application and it was not installed at the factory. However,
Master/Slave operation is always available on the system for
stand-alone operation.
Configure Profibus
The Profibus communication method requires a plug-in module
and software to operate. Contact Heraeus Noblelight Fusion UV
Inc. Technical Services if Profibus is required for your application
and it was not installed at the factory.
Configure EtherNet/IP
The EtherNet/IP communication method requires a plug-in
module and software to operate. Contact Heraeus Noblelight
Fusion UV Inc. Technical Services if EtherNet/IP is required for
your application and it was not installed at the factory.
There are four LED indicators, three lamp state buttons (ON, S/B,
OFF), a lamp disable switch, and a 2-line, 16-character display
for address, power level, data view, configuration, and error
reporting. The two arrow keys are used to set the power level and
to navigate through the menus. Refer to Figure 29.
Reset to Standby Flashing Yellow Unit is entering standby. During this time, the unit
runs checks on interlock status (Air Pressure, Cable
Interlock, RF Detector).
Standby to Lamp Flashing Green Unit is entering Lamp On state. During this time,
ON the power supply is performing its safety check.
Master LED
The blue Master LED is illuminated when the unit is configured as
a Master unit. For Slave units, this LED remains off (dark).
Fault State
In the case of a fault, the Lamp On, Standby, and Lamp Off LEDs
will flash together and an error code will display on the 2-line,
16-character display. Refer to the TROUBLESHOOTING section
for the cause of the error.
When a fault occurs, correct the fault and reset the unit by
pressing the OFF button before operation can be resumed.
• If a SYSTEM fault occurs, correct the fault and reset the Master
unit.
1. The OFF, STBY, and ON LEDs will blink and the display
indicates Waiting for User Reset. Press the OFF button to reset
the system.
These rules also apply to the power level. Any Slave unit cannot
have a power level higher than the Master unit unless the
optional power offset setting (+%power) is selected for the Slave
unit.
These rules also apply to the power level as well. Any connected
unit cannot have a power level higher than what is set by the
Remote Controller.
Operation States
Inputs on J106
Initial
Reset Standby Lamp On
Power On State
Closed indicates that the connected relay (switch) shorts this input to GND.
• Clean excessive dust and debris from the exterior of the power
supply by removing the air filter, fan assembly, and rear
access plate and then blowing compressed air through the air
flow path. Excessive dust and debris can limit cooling
efficiency and encourage arcing or tracking failures across HV
components. Periodically replace the air filters, located on the
front panel.
1. Unsnap the cover for each air filter. There are four clips; one
on each side.
Access to the cavity for changing the bulb or for routine cleaning
of the bulb and reflector is easily accomplished by removing the
screws which secure the screen assembly to the bottom of the
lamp. This procedure is explained in detail below in the “Bulb
Cleaning” section. The fine mesh screen must be intact and firmly
secured to the lamp if the system is to work properly and safely.
The RF leakage interlock automatically shuts the system off if
excess microwave energy passes through a damaged screen.
There are three hour meter setpoints that can be set to provide
a soft warning or reminder that a maintenance activity is
required. Refer to the “View/Reset Hour Meter” paragraph in the
CONFIGURATION section for further information.
Bulb Cleaning
Reflector Cleaning
If the reflector surface is dull or discolored, it should be cleaned.
Abrasives should NEVER be used on reflectors.
Screen Cleaning
Screens may be cleaned with the same cleaners cited in the “Bulb
Cleaning” paragraph. Screens may be soaked, if necessary.
Brush lightly and rinse well. Care should be taken not to damage
the gaskets or the fine screen wires.
• Display Parameters
• Display Hours
• Software Version
Display Parameters
During operation, the system displays the status of various
parameters. This display is active in RESET, STANDBY, and LAMP
ON states.
Display Hours
1. If the View Data Menu is not displayed,
press and hold the arrow keys (>and
?) simultaneously for three seconds to
enter this mode. After entering Data
View, DATA VIEW MODE displays on
the top line of the display.
2. When Display Hours displays and/or
scrolls across the LHP10 Mark II front
panel display, press and release the
button, located on the front panel of
the LHP10 Mark II, corresponding to NOTE: Operational state can be determined
by the LED associated with the button. When
the current operational state of the
the LED is illuminated, this indicates the
Lamp to enter the Display Hours
current Lamp mode.
sub-menu.
3. Use the arrow keys (>or?) on the front
panel to cycle through the various
component usage hours. The available
runtime hours include: Magnetron
Hours, Bulb hours, Reflector hours,
and Power Supply Standby hours,
Power Supply Lamp On hours.
Software Version
1. If the View Data Menu is not displayed,
press and hold the arrow keys (>and
?) simultaneously for three seconds to
enter this mode. After entering Data
View, DATA VIEW MODE displays on
the top line of the display.
2. When Software Version displays
and/or scrolls across the LHP10 Mark
II front panel display, press and
NOTE: Operational state can be determined
release the button, located on the
by the LED associated with the button. When
front panel of the LHP10 Mark II,
the LED is illuminated, this indicates the
corresponding to the current current Lamp mode.
operational state of the Lamp to enter
the Software Version sub-menu.
3. Use the arrow keys (>or?) on the front
panel to cycle through the installed
software version for the Engine A,
Engine B, the CPU, and Blower (if
installed). Some components have
more than one software version
installed (e.g., Engine A has four:
primary, primary-boot, secondary, and
secondary-boot).
Basic Checks
Check that input power is connected properly at J101. Actual line
voltage must be within ±10% of the voltage rating shown on the
equipment label. Labels are located on the front of the power
supply. If input voltage does not match, contact Heraeus
Noblelight Fusion UV Inc.
Parameter Values
The parameters shown in Table 30 can be viewed from the
Display Parameters menu. This menu is for viewing purposes
only and no changes can be made to the values displayed.
However, the displayed parameter values may change based on
the operational state of the Lamp. For instructions on accessing
this menu, refer to “Display Parameters” in the DIAGNOSTICS
section.
Fault Tables
The tables below list the error messages by classification and
error number.
NOTE: When operating in DeviceNet, Profibus, or EtherNet/IP
mode, Fault 11 acts as a timed warning. The warning time (60
seconds) allows customers to provide a controlled shutdown
to their process. After the predefined time, the fault is
immediate and the system shuts down.
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
0 0 no fault
1 1 MAG-A CURRENT HI Magnetron A current above threshold. 1. Magnetron end of life.
2. Defective Engine A.
2 2 MAG-B CURRENT HI Magnetron B current above threshold. 1. Magnetron end of life.
2. Defective Engine B.
3 3 MAG-A CURRENT LO Magnetron A current below threshold. 1. Magnetron end of life.
2. Defective Engine A.
4 4 MAG-B CURRENT LO Magnetron B current below threshold. 1. Magnetron end of life.
2. Defective Engine B.
5 NOT USED This error is only available on the LHP10 units
and not on the LHP10 Mark II units. There fore,
this error message will ONLY display on the
original LHP10 units. When LHP10 Mark II units
June 2014
TROUBLESHOOTING
10 10 AIR PRESSURE Air Pressure within irradiator out of specification 1. Air pressure interlock open.
range. 2. Remote blower(s) are not delivering
This error is caused by an electrical Open in the specified air pressure to irradiator.
power supply System Air Pressure circuit. This 3. Air pressure switch dirty.
fault can indicate low irradiator air pressure.
4. Defective CPU board.
114
Table 31: Common Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
11 11 RF DETECTOR RF Detector sensor not functioning properly. If error occurs from RESET to STANDBY:
This error is caused by a defective RF-1 model 1. RF detector failed self test.
RF detector. Possible internal short-circuit. If error occurs from STANDBY to LAMP ON:
1. Torn RF screen.
2. Failed lamp bulb.
12 NOT USED This error is only available on the LHP10 units
and not on the LHP10 Mark II units. There fore,
this error message will ONLY display on the
original LHP10 units. When LHP10 Mark II units
are operated in Legacy mode, the error message
is coded to match the Mark II error messages.
13 NOT USED This error is only available on the LHP10 units
and not on the LHP10 Mark II units. There fore,
this error message will ONLY display on the
June 2014
TROUBLESHOOTING
19 19 FIL-B CURRENT HI Filament B current above threshold. 1. Magnetron end of life.
2. Defective Engine B.
20 20 MAG-A VOLT HIGH Magnetron A voltage is above maximum 1. Magnetron end of life.
threshold. 2. Defective Engine A.
21 21 MAG-A VOLT LOW Magnetron A voltage is below minimum 1. Magnetron end of life.
115
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
22 22 NO RF DETECTOR No RF detector detected. 1. No RF detector installed.
This error indicates a disconnected or damaged 2. Defective RF harness.
RF detector. The LHP10 Mark II control card 3. Defective RF cable.
must recognize a working RF detector during
4. Defective CPU board.
power supply or irradiator operation.
5. Damaged internal harness.
6. Defective RF detector.
23 NOT USED This error is only available on the LHP10 units
and not on the LHP10 Mark II units. There fore,
this error message will ONLY display on the
original LHP10 units. When LHP10 Mark II units
are operated in Legacy mode, the error message
is coded to match the Mark II error messages.
24 24 NO BULB IGNITION Power Supply cannot maintain the start up If the UV bulb did not light, replace bulb.
power level. 1. Magnetron end of life.
June 2014
This error is caused when the control board does 2. Defective photodetector.
not detect a proper signal from the irradiator NOTE: A defective photodetector does
photodetector at UV bulb startup. not prevent the UV bulb from lighting.
HINT: Substitute another irradiator to verify 3. Defective Engine A or Engine B.
source of fault.
25 25 PHOTODETECTOR This error is caused by a fault photodetector 1. Defective photodetector.
SHORT assembly which is located in the LHI10-series 2. Defective CPU board.
irradiator. The cause can usually be traced to a
photodetector that has lowered substantially in 3. Defective wiring and/or cables.
resistance when the irradiator is in the STANDBY
mode. An incorrect resistance reading will be
seen by the power supply control card as a
photodetector that has shorted. (DC voltage
drop across the photodetector is too low,
simulating the "UV Lamp ON" condition.)
TROUBLESHOOTING
116
Table 31: Common Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
26 26 LAMP OUT Photodetector sensor not sensing lamp on. 1. Defective photodetector.
This error usually means that the UV bulb has 2. Bulb failure.
come on; however, the photodetector signal 3. Defective magnetron.
from the irradiator did not meet the required
4. Overcooled bulb.
voltage setpoint within 10 seconds of bulb
ignition. This is usually a sign of photodetector 5. Torn screen on Irradiator.
interlock failure due to degraded irradiator
components.
HINT: Substitute another irradiator to verify
source of fault.
27 27 ESTOP DETECTED E-stop circuit is open. 1. E-stop interlock is open.
2. Incorrect E-stop wiring and/or contacts
used.
3. Defective E-stop harness.
June 2014
TROUBLESHOOTING
(3.3V, 5V, or 13V) is low. 2. Defective Engine A.
117
Table 31: Common Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
31 31 COMM TIMEOUT Slave unit did not receive message from Master Communication timeout fault.
unit.
This error means that this particular power
supply control board has lost the BUS
communication signal, which is carried between
all power supplies operating in a group
(MASTER/SLAVE(s)). The specific signals from
the PLC (controller) are not being received by
this power supply.
32 NOT USED This error is only available on the LHP10 units
and not on the LHP10 Mark II units. Therefore,
this error message will ONLY display on the
original LHP10 units. When LHP10 Mark II units
are operated in Legacy mode, the error message
is coded to match the Mark II error messages.
June 2014
33 33 CABLE INTERLOCK One of the Cable Interlocks is open. 1. Cable interlock input open.
2. Open HV cable in power supply or
irradiator.
3. Open control cable.
4. Defective CPU board.
34 34 RAM FAULT RAM Fault. CPU board software data mismatch.
35 NOT USED This error is only available on the LHP10 units
and not on the LHP10 Mark II units. There fore,
this error message will ONLY display on the
original LHP10 units. When LHP10 Mark II units
are operated in Legacy mode, the error message
is coded to match the Mark II error messages.
36 36 NETWORK POWER Network has lost power. DeviceNet network power problem.
37 20-37 MAG-B VOLT HIGH Magnetron B voltage is above maximum 1. Magnetron end of life.
TROUBLESHOOTING
threshold. 2. Defective Engine B.
38 21-38 MAG-B VOLT LOW Magnetron B voltage is below minimum 1. Magnetron end of life.
threshold. 2. Defective Engine B.
118
Table 31: Common Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
39 16-39 FIL-B RMS V LOW Filament B voltage supplied to lamp is below 1. Magnetron end of life.
threshold 2. Defective Engine B.
40 17-40 FIL-B RMS V HIGH Filament B voltage supplied to lamp is above 1. Magnetron end of life.
threshold 2. Defective Engine B.
41-50 reserved for future use
June 2014
TROUBLESHOOTING
119
Table 32: Other Faults
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
ENGINE A FAULTS
51 30-51 5V Low Engine A internal 5V is low. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
52 30-52 5V High Engine A internal 5V is high. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
53 30-53 12V Low Engine A internal 12V is low. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
54 30-54 12V High Engine A internal 12V is high. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
55 30-55 24V Low Engine A internal 24V is low. Defective Engine A hardware. Refer to
June 2014
TROUBLESHOOTING
phase 1 & 2 is high; input AC voltage is high. “Determining the Likely Cause of a Fault” on
page 113.
61 32-61 Phase 2-3 Low Engine A input voltage measurement between Defective Engine A hardware. Refer to
phase 2 & 3 is low; input AC voltage is low. “Determining the Likely Cause of a Fault” on
page 113.
120
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
62 32-62 Phase 2-3 High Engine A input voltage measurement between Defective Engine A hardware. Refer to
phase 2 & 3 is high; input AC voltage is high. “Determining the Likely Cause of a Fault” on
page 113.
63 32-63 Phase 3-1 Low Engine A input voltage measurement between Defective Engine A hardware. Refer to
phase 3 &1 is low; input AC voltage is low. “Determining the Likely Cause of a Fault” on
page 113.
64 32-64 Phase 3-1 High Engine A input voltage measurement between Defective Engine A hardware. Refer to
phase 3 &1 is high; input AC voltage is high. “Determining the Likely Cause of a Fault” on
page 113.
65 28-65 Conv. Freq. Low Switch mode power converter module switching Defective Engine A hardware. Refer to
frequency is low. “Determining the Likely Cause of a Fault” on
page 113.
66 6628- Conv. Freq. High Switch mode power converter module switching Defective Engine A hardware. Refer to
frequency is high. “Determining the Likely Cause of a Fault” on
page 113.
June 2014
67 7-67 Temp Δ PFC1 Lo Temperature delta between PFC1 and air inlet is
low.
68 5-68 Temp Δ PFC1 Hi Temperature delta between PFC1 and air inlet is 1. Inlet air is blocked.
high. 2. Fan filters are dirty.
3. Defective fan or fans.
4. Defective Engine A hardware.
69 7-69 Temp Δ PFC2 Lo Temperature delta between PFC2 and air inlet is
low.
70 5-70 Temp Δ PFC2 Hi Temperature delta between PFC1 and air inlet is 1. Inlet air is blocked.
high. 2. Fan filters are dirty.
3. Defective fan or fans.
4. Defective Engine A hardware.
TROUBLESHOOTING
71 7-71 Temp Δ Conv Lo Temperature delta between HV converter section
and air inlet is low.
72 5-72 Temp Δ Conv Hi Temperature delta between HV converter section 1. Inlet air is blocked.
and air inlet is high. 2. Fan filters are dirty.
3. Defective Engine A hardware.
121
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
73 32-73 Mains Freq. Low Input AC power frequency (50/60 Hz) is low. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
74 32-74 Mains Freq. High Input AC power frequency (50/60 Hz) is high. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
75 32-75 AC Unbalanced Load between the 3-phase input is unbalanced. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
76 28-76 HV Phase 1 Low Internal Engine A hardware fault. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
77 28-77 HV Phase 1 High Internal Engine A hardware fault. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
June 2014
78 28-78 HV Phase 2 Low Internal Engine A hardware fault. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
79 28-79 HV Phase 2 High Internal Engine A hardware fault. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
80 21-80 Mag V Low Magnetron low voltage fault. Output disabled. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Magnetron end of life.
2. Defective Engine A.
81 20-81 Mag V High Magnetron overvoltage fault. Output disabled. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Magnetron end of life.
TROUBLESHOOTING
2. Defective Engine A.
82 31-82 Comm. Fault Mod Bus communication fault. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Check cable connection between CPU and
Engine A.
2. Defective CPU board.
122
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
83 12-83 PFC Bus V-Pos Hi PFC positive bus voltage is high. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
84 12-84 PFC Bus V-Neg Hi PFC negative bus voltage is high. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
85 12-85 PFC Bus I-Pos Hi PFC positive bus current is high. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
86 12-86 PFC Bus I-Neg Hi PFC negative bus current is high. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
87 13-87 PFC Bus V-Pos Lo PFC positive bus voltage is low. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
June 2014
88 12-88 Converter-I High HV converter module overcurrent condition. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Magnetron end of life.
2. Defective Engine A.
89 13-89 PFC Bus V-Neg Lo PFC negative bus voltage is low. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
90 18-90 Filament-I High Filament A overcurrent condition. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Magnetron end of life.
2. Filament transformer defective.
3. Defective Engine A hardware.
91 12-91 PFC P-Pos High PFC positive bus power is high. Defective Engine A hardware. Refer to
TROUBLESHOOTING
“Determining the Likely Cause of a Fault” on
page 113.
92 12-92 PFC P-Neg High PFC negative bus power is high. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
123
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
93 13-93 Primary I-Pos Lo Positive primary side low current condition. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Magnetron end of life.
2. Defective Engine A hardware.
94 12-94 Primary I-Pos Hi Positive primary side overcurrent condition. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Magnetron end of life.
2. Defective Engine A hardware.
95 13-95 Primary I-Neg Lo Negative primary side low current condition. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Magnetron end of life.
2. Defective Engine A hardware.
96 12-96 Primary I-Neg Hi Negative primary side overcurrent condition. Refer to “Determining the Likely Cause of a
June 2014
TROUBLESHOOTING
3. Defective Engine A hardware.
103 7-103 Temp Air Low Ambient air temperature is lower than threshold.
104 7-104 Temp PFC1 Low Ambient air temperature is lower than threshold.
105 7-105 Temp PFC2 Low Ambient air temperature is lower than threshold.
124
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
106 6-106 Temp Air High Ambient air temperature exceeds threshold. 1. Inlet air is blocked.
2. Fan filters are dirty.
3. Defective fan or fans.
4. Defective Engine A hardware.
107 5-107 Temp PFC1 High Ambient air temperature exceeds threshold. 1. Inlet air is blocked.
2. Fan filters are dirty.
3. Defective fan or fans.
4. Defective Engine A hardware.
108 5-108 Temp PFC2 High Ambient air temperature exceeds threshold. 1. Inlet air is blocked.
2. Fan filters are dirty.
3. Defective fan or fans.
4. Defective Engine A hardware.
June 2014
109 7-109 Temp Conv Low Ambient air temperature is lower than threshold.
110 5-110 Temp Conv High Ambient air temperature exceeds threshold. 1. Inlet air is blocked.
2. Fan filters are dirty.
3. Defective fan or fans.
4. Defective Engine A hardware.
111 1-111 Mag-I High Magnetron current high. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Possible magnetron end of life.
2. Defective Engine A hardware.
114 17-114 Fil Over Power Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Possible magnetron end of life.
2. Defective Engine A hardware.
TROUBLESHOOTING
116 31-116 SPI Comm. Engine A internal communication fault.
117 27-117 ESTOP Input Open E-Stop signal is open to Engine A. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Check cable between CPU and Engine A.
2. Defective CPU board.
3. Defective Engine A hardware.
125
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
118 28-118 Interlock Circuit
119 6-119 Temp Air Out Hi Exhaust temperature from Engine A is higher Defective Engine A hardware. Refer to
than threshold. “Determining the Likely Cause of a Fault” on
page 113.
120 7-120 Temp Air Out Low Exhaust temperature from Engine A is lower than Defective Engine A hardware. Refer to
threshold. “Determining the Likely Cause of a Fault” on
page 113.
121 9-121 Fans rpm Low Two or more fan’s rpm is lower than threshold 1. Defective fans.
2. Defective Engine A hardware.
124 31-124 Comm. fault ModBus communication fault Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Check cable between CPU and Engine A.
2. Defective CPU board.
June 2014
TROUBLESHOOTING
134 30-134 12V High Engine A internal 12V is high. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
135 30-135 ADC Vref Low Engine A internal ACD reference voltage low. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
126
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
136 30-136 ADC Vref High Engine A internal ACD reference voltage high. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
137 30-137 50mV Vref Low Engine A internal 50mV reference voltage low. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
138 30-138 50mV Vref High Engine A internal 50mV reference voltage high. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
139 9-139 V Fans Low Fans voltage is low. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
140 9-140 V Fans High Fans voltage is high. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
June 2014
141 9-141 I-Fans Low Fans total current is low. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
142 9-142 I-Fans High Fans total current is high. Defective Engine A hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
151 20-151 Mag V High Vout Mag High Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Possible magnetron end of life.
2. Defective Engine A hardware.
ENGINE B FAULTS
201 30-201 5V Low Engine B internal 5V is low. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
TROUBLESHOOTING
202 30-202 5V High Engine B internal 5V is high. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
203 30-203 12V Low Engine B internal 12V is low. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
127
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
204 30-204 12V High Engine B internal 12V is high. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
205 30-205 24V Low Engine B internal 24V is low. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
206 30-206 24V High Engine B internal 24V is high. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
207 30-207 ADC Vref Low Engine B internal ACD reference voltage low. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
208 30-208 ADC Vref High Engine B internal ACD reference voltage high. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
June 2014
209 32-209 Phase 1-2 Low Engine B input voltage measurement between Defective Engine B hardware. Refer to
phase 1 & 2 is low; input AC voltage is low. “Determining the Likely Cause of a Fault” on
page 113.
210 32-210 Phase 1-2 High Engine B input voltage measurement between Defective Engine B hardware. Refer to
phase 1 & 2 is high; input AC voltage is high. “Determining the Likely Cause of a Fault” on
page 113.
211 32-211 Phase 2-3 Low Engine B input voltage measurement between Defective Engine B hardware. Refer to
phase 2 & 3 is low; input AC voltage is low. “Determining the Likely Cause of a Fault” on
page 113.
212 32-212 Phase 2-3 High Engine B input voltage measurement between Defective Engine B hardware. Refer to
phase 2 & 3 is high; input AC voltage is high. “Determining the Likely Cause of a Fault” on
page 113.
213 32-213 Phase 3-1 Low Engine B input voltage measurement between Defective Engine B hardware. Refer to
phase 3 &1 is low; input AC voltage is low. “Determining the Likely Cause of a Fault” on
TROUBLESHOOTING
page 113.
214 32-214 Phase 3-1 High Engine B input voltage measurement between Defective Engine B hardware. Refer to
phase 3 &1 is high; input AC voltage is high. “Determining the Likely Cause of a Fault” on
page 113.
215 28-215 Conv. Freq. Low Switch mode power converter module switching Defective Engine B hardware. Refer to
frequency is low. “Determining the Likely Cause of a Fault” on
128
page 113.
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
216 28-216 Conv. Freq. High Switch mode power converter module switching Defective Engine B hardware. Refer to
frequency is high. “Determining the Likely Cause of a Fault” on
page 113.
217 7-217 Temp Δ PFC1 Lo Temperature delta between PFC1 and air inlet is
low.
218 5-218 Temp Δ PFC1 Hi Temperature delta between PFC1 and air inlet is 1. Inlet air is blocked.
high. 2. Fan filters are dirty.
3. Defective fan or fans.
4. Defective Engine B hardware.
219 7-219 Temp Δ PFC2 Lo Temperature delta between PFC2 and air inlet is
low.
220 5-220 Temp Δ PFC2 Hi Temperature delta between PFC1 and air inlet is 1. Inlet air is blocked.
high. 2. Fan filters are dirty.
June 2014
TROUBLESHOOTING
“Determining the Likely Cause of a Fault” on
page 113.
226 28-226 HV Phase 1 Low Internal Engine B hardware fault. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
129
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
227 28-227 HV Phase 1 High Internal Engine B hardware fault. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
228 28-228 HV Phase 2 Low Internal Engine B hardware fault. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
229 28-229 HV Phase 2 High Internal Engine B hardware fault. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
230 21-230 Mag V Low Magnetron low voltage fault. Output disabled. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Magnetron end of life.
2. Defective Engine B.
231 20-231 Mag V High Magnetron overvoltage fault. Output disabled. Refer to “Determining the Likely Cause of a
June 2014
TROUBLESHOOTING
page 113.
235 12-235 PFC Bus I-Pos Hi PFC positive bus current is high. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
236 12-236 PFC Bus I-Neg Hi PFC negative bus current is high. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
130
page 113.
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
237 13-237 PFC Bus V-Pos Lo PFC positive bus voltage is low. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
238 12-238 Converter-I High HV converter module overcurrent condition. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Magnetron end of life.
2. Defective Engine B.
239 13-239 PFC Bus V-Neg Lo PFC negative bus voltage is low. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
240 19-240 Filament-I High Filament B overcurrent condition. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Magnetron end of life.
2. Filament transformer defective.
June 2014
TROUBLESHOOTING
2. Defective Engine B hardware.
245 13-245 Primary I-Neg Lo Negative primary side low current condition. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Magnetron end of life.
2. Defective Engine B hardware.
131
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
246 12-246 Primary I-Neg Hi Negative primary side overcurrent condition. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Magnetron end of life.
2. Defective Engine B hardware.
247 13-247 Bus Balance Low Internal Bus balance is low. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
248 12-248 Bus Balance High Internal Bus balance is high. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
250 31-250 SPI Comm. Engine B internal communication fault.
251 27-251 ESTOP Input Open E-Stop signal is open to Engine B. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Check cable between CPU and Engine B.
June 2014
TROUBLESHOOTING
4. Defective Engine B hardware.
258 5-258 Temp PFC2 High Ambient air temperature exceeds threshold. 1. Inlet air is blocked.
2. Fan filters are dirty.
3. Defective fan or fans.
4. Defective Engine B hardware.
132
259 7-259 Temp Conv Low Ambient air temperature is lower than threshold.
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
260 5-260 Temp Conv High Ambient air temperature exceeds threshold. 1. Inlet air is blocked.
2. Fan filters are dirty.
3. Defective fan or fans.
4. Defective Engine B hardware.
261 2-261 Mag-I High Magnetron current high. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Possible magnetron end of life.
2. Defective Engine B hardware.
264 17-264 Fil Over Power Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Possible magnetron end of life.
2. Defective Engine B hardware.
266 31-266 SPI Comm. Engine B internal communication fault.
June 2014
267 27-267 ESTOP Input Open E-Stop signal is open to Engine A. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Check cable between CPU and Engine B.
2. Defective CPU board.
3. Defective Engine B hardware.
269 6-269 Temp Air Out Hi Exhaust temperature from Engine B is higher Defective Engine B hardware. Refer to
than threshold. “Determining the Likely Cause of a Fault” on
page 113.
270 7-270 Temp Air Out Low Exhaust temperature from Engine B is lower than Defective Engine B hardware. Refer to
threshold. “Determining the Likely Cause of a Fault” on
page 113.
274 31-274 Comm. fault ModBus communication fault Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Check cable between CPU and Engine B.
TROUBLESHOOTING
2. Defective CPU board.
3. Defective Engine B hardware.
275 28-275 HV Over Power Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Possible magnetron end of life.
133
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
281 30-281 5V Low Engine B internal 5V is low. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
282 30-282 5V High Engine B internal 5V is high. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
283 30-283 12V Low Engine B internal 12V is low. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
284 30-284 12V High Engine B internal 12V is high. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
285 30-285 ADC Vref Low Engine B internal ACD reference voltage low. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
June 2014
286 30-286 ADC Vref High Engine B internal ACD reference voltage high. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
287 30-287 50mV Vref Low Engine B internal 50mV reference voltage low. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
288 30-288 50mV Vref High Engine B internal 50mV reference voltage high. Defective Engine B hardware. Refer to
“Determining the Likely Cause of a Fault” on
page 113.ware.
301 20-301 Mag V High Vout Mag High Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Possible magnetron end of life.
2. Defective Engine B hardware.
BLOWER MODULE FAULTS
TROUBLESHOOTING
351 10-351 5V Low Blower control module internal 5V is low. Defective Blower control module. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
352 10-352 5V High Blower control module internal 5V is high. Defective Blower control module. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
134
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
353 10-353 12V Low Blower control module internal 12V is low. Defective Blower control module. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
354 10-354 12V High Blower control module internal 12V is high. Defective Blower control module. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
357 10-357 ADC Vref Low Blower control module internal ACD reference Defective Blower control module. Refer to
voltage low. “Determining the Likely Cause of a Fault” on
page 113.
358 10-358 ADC Vref High Blower control module internal ACD reference Defective Blower control module. Refer to
voltage high. “Determining the Likely Cause of a Fault” on
page 113.
359 10-359 Phase 1-2 Low Blower control module input voltage Defective Blower control module. Refer to
measurement between phase 1 & 2 is low; input “Determining the Likely Cause of a Fault” on
AC voltage is low. page 113.
June 2014
360 10-360 Phase 1-2 High Blower control module input voltage Defective Blower control module. Refer to
measurement between phase 1 & 2 is high; input “Determining the Likely Cause of a Fault” on
AC voltage is high. page 113.
361 10-361 Phase 2-3 Low Blower control module input voltage Defective Blower control module. Refer to
measurement between phase 2 & 3 is low; input “Determining the Likely Cause of a Fault” on
AC voltage is low. page 113.
362 10-362 Phase 2-3 High Blower control module input voltage Defective Blower control module. Refer to
measurement between phase 2 & 3 is high; input “Determining the Likely Cause of a Fault” on
AC voltage is high. page 113.
363 10-363 Phase 3-1 Low Blower control module input voltage Defective Blower control module. Refer to
measurement between phase 3 &1 is low; input “Determining the Likely Cause of a Fault” on
AC voltage is low. page 113.
364 10-364 Phase 3-1 High Blower control module input voltage Defective Blower control module. Refer to
measurement between phase 3 &1 is high; input “Determining the Likely Cause of a Fault” on
TROUBLESHOOTING
AC voltage is high. page 113.
365 10-365 Conv. Freq. Low Switch mode power converter module switching Defective Blower control module. Refer to
frequency is low. “Determining the Likely Cause of a Fault” on
page 113.
366 10-366 Conv. Freq. High Switch mode power converter module switching Defective Blower control module. Refer to
frequency is high. “Determining the Likely Cause of a Fault” on
135
page 113.
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
367 10-367 Temp Δ PFC1 Lo Temperature delta between PFC1 and air inlet is
low.
368 10-368 Temp Δ PFC1 Hi Temperature delta between PFC1 and air inlet is 1. Inlet air is blocked.
high. 2. Fan filters are dirty.
3. Defective fan or fans.
4. Defective Blower control module.
369 10-369 Temp Δ PFC2 Lo Temperature delta between PFC2 and air inlet is
low.
370 10-370 Temp Δ PFC2 Hi Temperature delta between PFC1 and air inlet is 1. Inlet air is blocked.
high. 2. Fan filters are dirty.
3. Defective fan or fans.
4. Defective Blower control module.
371 10-371 Temp Δ Conv Lo Temperature delta between HV converter section
June 2014
TROUBLESHOOTING
Fault” on page 113.
1. Check cable connection between CPU and
Blower control module.
2. Defective CPU board.
3. Defective Blower control module.
136
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
383 10-383 PFC Bus V-Pos Hi PFC positive bus voltage is high. Defective Blower control module. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
384 10-384 PFC Bus V-Neg Hi PFC negative bus voltage is high. Defective Blower control module. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
385 10-385 PFC Bus I-Pos Hi PFC positive bus current is high. Defective Blower control module. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
386 10-386 PFC Bus I-Neg Hi PFC negative bus current is high. Defective Blower control module. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
387 10-387 PFC Bus V-Pos Lo PFC positive bus voltage is low. Defective Blower control module. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
June 2014
389 10-389 PFC Bus V-Neg Lo PFC negative bus voltage is low. Defective Blower control module. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
391 10-391 PFC P-Pos High PFC positive bus power is high. Defective Blower control module. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
392 10-392 PFC P-Neg High PFC negative bus power is high. Defective Blower control module. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
393 10-393 Primary I-Pos Lo Positive primary side low current condition. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Magnetron end of life.
2. Defective Blower control module.
394 10-394 Primary I-Pos Hi Positive primary side overcurrent condition. Refer to “Determining the Likely Cause of a
TROUBLESHOOTING
Fault” on page 113.
1. Magnetron end of life.
2. Defective Blower control module.
137
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
395 10-395 Primary I-Neg Lo Negative primary side low current condition. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Magnetron end of life.
2. Defective Blower control module.
396 10-396 Primary I-Neg Hi Negative primary side overcurrent condition. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Magnetron end of life.
2. Defective Blower control module.
397 10-397 Bus Balance Low Internal Bus balance is low. Defective Blower control module. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
398 10-398 Bus Balance High Internal Bus balance is high. Defective Blower control module. Refer to
“Determining the Likely Cause of a Fault” on
page 113.
June 2014
401 10-401 ESTOP Input Open E-Stop signal is open to Blower control module. Refer to “Determining the Likely Cause of a
Fault” on page 113.
1. Check cable between CPU and Blower
control module.
2. Defective CPU board.
3. Defective Blower control module.
403 10-403 Temp Air Low Ambient air temperature is lower than threshold.
404 10-404 Temp PFC1 Low Ambient air temperature is lower than threshold.
405 10-405 Temp PFC2 Low Ambient air temperature is lower than threshold.
406 10-406 Temp Air in High Ambient air temperature exceeds threshold. 1. Inlet air is blocked.
2. Fan filters are dirty.
3. Defective fan or fans.
4. Defective Blower control module.
TROUBLESHOOTING
407 10-407 Temp PFC1 High Ambient air temperature exceeds threshold. 1. Inlet air is blocked.
2. Fan filters are dirty.
3. Defective fan or fans.
4. Defective Blower control module.
138
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
408 10-408 Temp PFC2 High Ambient air temperature exceeds threshold. 1. Inlet air is blocked.
2. Fan filters are dirty.
3. Defective fan or fans.
4. Defective Blower control module.
409 10-409 Temp Conv Low Ambient air temperature is lower than threshold.
410 10-410 Temp Conv High Ambient air temperature exceeds threshold. 1. Inlet air is blocked.
2. Fan filters are dirty.
3. Defective fan or fans.
4. Defective Blower control module.
SYSTEM FAULTS
501 DUPLICATE ADDR Duplicate Address in system Two or more units have the same address.
502 MISSING SLAVE(S) No response from a Slave configured in the The number of Slave units connected does not
June 2014
TROUBLESHOOTING
602 WRONG COMM Incorrect Module installed 1. Incorrect Configuration.
MODULE 2. Incorrect module installed.
3. Defective Communication Module or CPU.
139
Table 32: Other Faults (continued)
©Heraeus-Light Hammer 10 Mark II
Mark II Legacy
Error Error Error Message Error Description Possible Causes
Number Number
USB FAULTS
651 CHECKSUM
652 MSD not found Mass storage device (MSD) not connected at the
USB port
653 File not found Required file not found for Read operation.
654 File Read File may have incorrect value or be corrupted
655 Format error. Please FAT of MSD not supported or corrupted. Format the MSD and try again.
format the MSD using
PC and try again
656 MORE FACT FILE More than one Factory configuration file found in
MSD
657 MORE USER FILE More than one User configuration file found in
MSD
June 2014
MODBUS FAULTS
701 31-701 Engine A Comm. Engine A ModBus communication is not working. Check the cable connection.
702 31-702 Engine B Comm. Engine B ModBus communication is not working. Check the cable connection.
703 31-703 Blower Comm. Blower ModBus communication is not working. Check the cable connection.
MISC. FAULTS
751 30-751 POWER FAIL CPU board power failure
ACCESS ACCESS KEYPAD LOCKED Front panel locked. User has tried to access the front panel when
ERROR ERROR the front panel lock is enabled.
TROUBLESHOOTING
140
TROUBLESHOOTING
Location Position Date Hours of By Repairs made (Record all Replacement (serial By
Operation items replaced) number)
June 2014
SYSTEM LOGS
145
SYSTEM LOGS
Irradiator Description
Microwave energy is directed into the cavity formed by an
aluminum reflector and a screen over the bottom of the reflector.
The lamp bulb is located inside this chamber. The screen keeps
microwaves inside the chamber but allows UV light to pass
through onto the material to be cured.
2. Grasp the screen frame by both ends, taking care not to twist
or bend the screen itself, and remove the screen assembly.
4. Apply pressure to the frame only and press firmly into place.
Refer to Figure 32 for screen screw tightening sequence.
Tighten each to a snug position; finally tighten to 0.9 N-m
(8 inch-lbs) using a torque-controlled screwdriver.
Perform the following steps to remove and replace the lamp bulb.
3. Clean any deposits from the surface of the bulb. See “Bulb
Cleaning” in the ROUTINE MAINTENANCE section.
Retaining hole
Retaining hole
“209” End Reflector
Standard bulb position
Retaining hole
“209” bulb position
3. Remove the four (4) flat head screws and lift the reflector out.
Remove and discard the four (4) gasket strips. If evidence of
arcing is present (dark or burned spots), clean the gasket
grooves with emery cloth to remove all oxide.
5. Insert both end reflectors, being sure that they both seat fully
and evenly. The replacement end reflectors must be installed
before installing the new RF gasket.
6. Install the four pre-cut RF gaskets. Insert one end flush with
the top of the mounting surfaces. Working toward the other
end of the groove, press the gasket strip down firmly with a
thumbnail so that it rests at the bottom of the groove. Refer
to Figure 37.
Magnetron “B”
Filament A Filament B
Terminal Terminal
I/O Interface
Connector J11 Magnetron
(LHI10 Model) Anchor Nut
(4 per Mag)
Waveguide
Pressure
Assembly
Switch
Photodetector
Assembly
Filament
Transformer Reflector
2. Tilt the connector plate forward and unhook it from the slot in
the irradiator frame.
3. Cut the cable ties and slide the PVC tubing off of the
magnetron connections.
6. Reposition the PVC tubing over the terminal screws and install
cable ties.
10.Reposition the PVC tubing over the terminal screws and install
cable ties.
2. Tilt the connector plate forward and unhook it from the slot in
the irradiator frame.
3. Cut the cable ties and slide the PVC tubing off of the
magnetron connections.
6. Reposition the PVC tubing over the terminal screws and install
cable ties.
10.Reposition the PVC tubing over the terminal screws and install
cable ties.
Quartz window 2000 Pa 2500 Pa 7.7 m3/min @ 2875 Pa 8.9 m3/min @ 3125 Pa
(25-30 mm) (8.0" H2O) (10.0" H2O) 270 scfm @ 11.5" H2O) 315 scfm @ 12.5" H2O)
(Note 3)
F10T2/C10T2 For air requirements when using the F10T2 or C10T2 Lightshields, please refer to the F10T
Lightshield and C10T Manuals.
(no quartz window)
NOTES:
High
Pressure High
Pressure Hose
Measuring Pressure
Irradiator Pressure Hose Pressure
Gauge Test Point Measuring
Gauge
Low Low
Pressure Pressure
Hose Hose
Lightshield Lightshield
5. Remove the two Phillips head screws from the front of the
connector plate that hold the pressure switch in place. Refer
to Figure 40.
Safety Procedures
Before servicing the Light Hammer 10 Mark II UV Lamp System,
first turn off and disconnect all power to the unit.
The high voltage circuit inside the LHP10 Mark II power supply
consists of two solid state switching engines, which supply a
regulated source of power (approximately -5000V, 890 mA DC)
to each magnetron.
The CPU board within the power supply monitors cooling air
pressure, detection of RF, power levels, bulb ignition, and
interlocks. If a problem arises, the control system turns off all
high voltage power to the lamp, and flashes a fault indicator on
the power supply control panel.
1. Unsnap the cover for each air filter. There are four clips; one
on each side.
The power supply provides the electrical power and control to the
irradiator/blower unit.
Repairs
Removing Blower Cover
2. Lift the blower cover and filter straight up and off of the
irradiator.
Blower Replacement
5. Replace the blower and reverse the above steps. Tighten the
blower mounting nuts (M8) to 10-10.5 N-m (7.3-7.7 lb-ft) and
the blower cover screws (M5) to 3.4 N-m (30 inch-lbs).
J102 J111
J105 MASTER/SLAVE COMM BUS
J104 RF DETECTOR J106 J107
IRRAD CNTL CUSTOMER I/O E-STOP
J103
HV CNTL J108 BLOW ER
(option al)
J108
J107
BLOWER (optional)
E-STOP
J111 COMM BU S
J101 AC INPUT
J103 J104 J105 J106 J102
HV CNTL IR RAD C NTL RF DETECTOR CU STOMER I/O MA STER/SLAVE J101 AC INPU T
EARTH GROUND
AIR EXHAUST OUTLET
USB PORT
LED DISPLAY
LAMP DISABLE
kVA
217 mm
+
+
(8.54”)
+
+
+
+
+
+
MA STE R
ON
POWER LEVEL
S/B
FILTERS (3)
+
+
+
+
+
764.7 mm
(30.11”)
700.5 mm
(27.58”)
616 mm
(24.25”)
582 mm
42.5 mm 308 mm (22.91”)
(1.67”) (12.13”) REAR
LIFTING
FRONT 34 mm (1.34”) TAB
LIFTING
TAB
157.5mm
(6.2”)
131.72 mm
(5.19”)
Recommended 23.5 mm
Recommended
Front 656 mm (0.93”)
Rear
clearance (25.83”)
clearance
203 mm (8.0”) 305 mm (12.0”)
(minimum) (minimum)
133.5
(5.26)
84.6
(3.33)
FOR REFERENCE ONLY
NOT TO BE USED FOR
SPECIFICATION
Contact Heraeus Noblelight Fusion UV Inc.
for current revision of DWG 573540
before designing an installation.
44.5
(1.75) NOTE 3
434
M6 (8) (17.07)
J10
HV
384
150 (15.13) J11
(5.91) Control
Center
185 of
(7.3) Gravity
115 154 Pressure
(4.52) (6.1) Switch
25
(0.99)
206
NOTES: (8.12)
1. Dimensions in mm (inches). 173
2. Geometric center of Focus. 200 (6.81)
Optimum substrate location 212 (7.88)
may vary based on the (8.35)
specific process.
3. See Specifications for the
cooling air requirements.
29 70
(1.13) (2.76) M5 (8)
267
(10.53)
133.5
(5.26)
FOR REFERENCE ONLY
NOT TO BE USED FOR 84.6
SPECIFICATION (3.33)
Contact Heraeus Noblelight Fusion UV Inc.
for current revision of DWG 574030
before designing an installation.
OPTIONAL AIR
INLET DUCT
J10 J11
161
(6.34)
44.5
(1.75) NOTE 3
M6 (8)
384 Pressure
150 (15.13) Switch
(5.91)
254
(10.0)
Center
of
Gravity
185
115 (7.3)
(4.52)
25
(0.99)
206
(8.12)
NOTES: 173
1. Dimensions in mm (inches). 200 (6.81)
2. Geometric center of Focus. 212 (7.88)
Optimum substrate location (8.35)
may vary based on the
specific process.
3. See Specifications for the
cooling air requirements.
29 70
(1.13) (2.76) M5 (8)
267
(10.53)
44.5
(1.75)
Optional NOTE 3
air inlet M6 (8)
plenum
J10
HV
381
150 (14.98) J11
(5.91) Control
25
(0.99)
206
(8.12)
NOTES: 173
1. Dimensions in mm (inches). 200 (6.81)
2. Geometric center of Focus. 212 (7.88)
Optimum substrate location (8.35)
may vary based on the
specific process.
3. See Specifications for the
cooling air requirements.
29 70
(1.13) (2.76) M5 (8)
267
(10.53)
235
(9.25)
203 25
(8.00) 102 (1.00)
(4.00)
M6 blind insert
for mounting
transition inlet (6)
108 221 40
(4.25) (8.70) (1.56)
190
169 (7.49)
(6.64)
Optional NOTE 3
air inlet
plenum
336 Pressure
150 (13.23) Switch
(5.91)
254
(10.0)
115
(4.52)
25
(0.99)
206
NOTES: (8.12)
1. Dimensions in mm (inches). 173
2. Geometric center of Focus. 200 (6.81)
Optimum substrate location 212 (7.88)
may vary based on the (8.35)
specific process.
3. See Specifications for the
cooling air requirements.
29 70
(1.13) (2.76) M5 (8)
267
(10.53)
235
(9.25)
203 25
(8.00) 102 (1.00)
(4.00)
M6 blind insert
for mounting
transition inlet (6)
108 221 40
(4.25) (8.70) (1.56)
190
169 (7.49)
(6.64)
268
(10.54)
FOR REFERENCE ONLY
NOT TO BE USED FOR
SPECIFICATION
Contact Heraeus Noblelight Fusion UV Inc.
204 for current revision of DWG 584220
(8.02) before designing an installation.
101
(3.99)
LIFT HANDLES
2 PLACES NOTES:
1. All dimensions in mm (inches).
2. Cooling air inlet 224 mm (8.81) x 67 mm
J10 (2.65) with integral O-ring seal.
HV 3. Handles may be removed (user option).
J11 4. Mating mount plate shall incorporate two
Control
0.375 in. Diameter alignment pins,
Pressure
Switch 250 mm apart.
5. Captive, spring-loaded, threaded fasteners
(2) secure irradiator to mating M8 holes in
mounting plate. Use an M5 hex wrench.
6. Refer to the Specifications listed in
Equipment the Air Flow Requirements Table in
ID Plate this Document for the HVAC and
337
(13.25) differential pressure operating
requirements.
235
(9.24)
Alignment Slot
M8 Mounting
Bolt (2)
250
(9.84)
Alignment 115
(4.54)
Bushing
101
(3.98)
662
(26.07)
J12
Blower
M6 (8) NOTE 3
J10
HV
384
150 (15.13) J11
(5.91) Control
115
(4.52)
25
(0.99)
NOTES:
1. Dimensions in mm (inches).
2. Geometric center of Focus.
229 173
Optimum substrate location (9.00) 200 (6.81)
may vary based on the 212 (7.88)
specific process. (8.35)
3. See Specifications in Manual
for the cooling air requirements.
Packing Materials
Save the packing materials. They will be valuable if any
component must be returned to Heraeus Noblelight Fusion UV
Inc. for service.
Warranty Policy
Items manufactured by Heraeus Noblelight Fusion UV Inc. are
warranted (under normal and proper use) to be free from defects
in material and workmanship. Heraeus Noblelight Fusion UV Inc.
(HNFN) will repair or replace, at its option, any defective parts
when returned to HNFN by the purchaser, transportation paid,
within the warranty period stated on your Sales Contract. See the
Terms & Conditions included with your sales order for the length
of the warranty period for irradiators, power supplies, conveyors,
lamp bulbs, magnetrons, reflectors, and conveyor belts.
Power Supply
Power supplies should be returned in their original packaging. If
this is not available, the power supplies should be packaged as
follows:
Irradiators
Irradiators should be returned in their original packaging. If this
is not available, lamps must be double boxed, as follows:
Magnetrons
If available, use the original shipping box. Otherwise,
magnetrons must be double boxed as described below.
UV Bulbs
Place the intact bulb in a rigid tube (use Heraeus Noblelight
Fusion UV Inc. original plastic shipping tube, if available). Secure
ends to prevent impact against the tube and wrap the tube in
three layers of plastic “bubble wrap” or equivalent. Ship with
loose styrofoam fill in the box.
Control Cards
The integrated circuits in control board assemblies are
susceptible to damage from static electrical charges which may
be transmitted by handling or packaging materials. To guard
against static, place control cards in anti-static bags, if available.
If not, wrap control cards in aluminum foil. Isolate the cards from
impact with additional wrapping of “bubble wrap” or equivalent.
RF Detectors
RF Detectors should be wrapped in two layers of “bubble wrap”
or equivalent and placed in a box with loose styrofoam fill.
UV Bulbs
UV bulbs contain a small amount of mercury. In some
communities, disposal of items containing mercury are
restricted. Intact bulbs may be returned to Heraeus Noblelight
Fusion UV Inc. in their original containers for disposal. If this
option of disposal is not feasible, dispose of the bulb according to
your local regulatory requirements.
Reflectors
Reflectors are aluminum with a reflective coating. In some
communities, aluminum items are collected separately from
other waste items. Follow your local regulatory requirements for
the environmentally sound disposal of reflectors.
Magnetrons
Magnetrons are primarily constructed of different types of
metals. In some communities, metal items are collected
separately from other waste items. Follow your local regulatory
requirements for the environmentally sound disposal of
magnetrons.
Conveyor Belts
Disposal of conveyor belts will be dependent upon the
composition of the conveyor belt. Each type of belt may have
specific disposal requirements in your local area. Follow your
local regulatory requirements when disposing of these items.
• Declaration of Conformity/Incorporation
• TÜV Certificates
System Designation
Table 35: Light Hammer 10 Mark II System Component
Designations
Component Parameter
Irradiator LHI10-series
Component Parameter
Altitude 0-2000m
Agency Approvals
Table 37: Test Standards
EN 61000-6-2
EN 61000-4-x
Irradiator
Table 38: Irradiator Model LHI10-series Specifications
Component Parameter
ELECTRICAL:
MECHANICAL:
Dimensions see OUTLINE DRAWINGS
Substrate Location 53 mm (2.1 inch) from face of lamp, for maximum irradiance
COOLING:
Minimum Cooling at 100% Power Refer to Table 39 for the cooling requirements.
Air Delivery Top Connection: 152 mm (6 inch) diameter for LHI10 and LHI10_01
Rear Connection: 40mm x 221 mm (1.56 inch x 8.7 inch) for LHI10_02
and LHI10_03
Quartz window 2000 Pa 2500 Pa 7.7 m3/min @ 2875 Pa 8.9 m3/min @ 3125 Pa
(25-30 mm) (8.0" H2O) (10.0" H2O) 270 scfm @ 11.5" H2O) 315 scfm @ 12.5" H2O) (Note 3)
F10T2/C10T2 For air requirements when using the F10T2 or C10T2 Lightshields, please refer to the F10T
Lightshield and C10T Manuals.
(no quartz window)
NOTES:
Blower Specifications
Table 40: LHI10B Mark II Blower Specifications
LHI10B Mark II
ELECTRICAL:
MECHANICAL:
Maximum Dimensions Irradiator/Blower: 267 mm x 313 mm x 662 mm
(W x D x H) (10.53 in. x 12.32 in. x 26.07 in.)
OPERATING CONDITIONS:
Performance LHI10B Mark II blower delivers a minimum pressure of 1750 (7.0 inches to the
LHI10B Mark II irradiator test port. (Use non-swept reflectors only.)
Air Filtration In order to prevent fouling of the bulb or reflectors, cooling air should be
filtered using a filter with a minimum ASHRAE average efficiency of 34%. If
the National Institute of Standards and Technology (NIST) ratings are used, a
filter with a minimum of 60% efficiency is recommended. Efficiency is based
on NIST discoloration type test, using atmospheric dust as the test aerosol.
Filters must be non-woven, bonded polyester fiber with a maximum
continuous operating temperature of 250°F. The bonding agent must be flame
and fungus retardant as well as moisture proof. The air velocity is 200-450
fpm.
Replacement filters are available from Heraeus Noblelight Fusion UV Inc.
Noise 76 dbA (at 1 meter)
Power Supply
Table 41: Power Supply Model LHP10 Mark II
ELECTRICAL:
3-phase 50/60 Hz
Max. Line Current 200 - 480 V: 30-12 A
MECHANICAL:
Clearance Allow 305 mm (12 inches) clearance front and rear of the power supply
for cooling air flow and cable connections.
CONTROLS:
COOLING:
RF Detector
Table 43: RF-1 RF Detector Specifications
Component Parameter
-or -
For the most current Replacement Parts Listing for the Light
Hammer 10 Mark II system, please visit:
http://www.fusionuv.com/LH10_MarkII_Parts_listing.aspx
On-Line Store
Customers can now place parts orders online*. Our Web Store
provides a new level of efficiency and convenience and features:
• 24/7 access
If you are not yet registered for the Web Store but would like to
do so, please contact Sales Support on 1-301-990-8600 or
[email protected].
INDEX Master/Slave 65
navigation guide 55
A Profibus 74
read config data 58
AC power source save config data 59
power supply 23 save fault data 61
Air pressure startup/initializing 56
switch replacement, irradiator 165 View/Reset Hour Meter 79
test point 165 Configuration Menu 57
verification, irradiator 163 Control cards, packing instructions 191
Ambient operating temperature 201 Cooling requirements
Analog power control irradiator 163
0-10 Vdc 35 Credit for returns 190
4-20 mA 35 Cross reference for faults 113
ANSI Standard C95.1-1999 9 Cross reference for Legacy faults 141
Authorization number for returns 187, 190 Curable materials 12
Current
B
leg 26
Basic troubleshooting checks 108 line 25
Blower Customer responsibilities 13
cover removal 171
remote 16 D
replacement 171 Data View menu items
sizing 16 Diagnostic menu items 96
specifications 204 Data View Mode 97
Blower control module, leg currents 27 display hours 101
Bureau of Radiological Health 8 display S/N 100
EEPROM data 105
C
enter diagnostic mode 99
Cable factory settings 103
irradiator connections 27 how enter 96
irradiator/blower connection 27 software version 102
power supply/blower connection 53 user settings 104
RF connection 28 DeviceNet
Checklist configuration 72
electrical installation 53 connections 42
mechanical installation 22 status LEDs 109
Communication bus terminating resistor 43
DeviceNet 42 Diagnostic Mode
DeviceNet configuration 72 how enter 99
EtherNet configuration 76 Differential pressure configuration 17
EtherNet/IP 50 Digital power control 35
Master/Slave configuration 65 Dimensions
Profibus 46 irradiator 175
Profibus configuration 74 power supply 174
specification 206 quartz plate 18
Communication Port configuration 65 RF detector 184
Components Direct view photodetector replacement 162
designations 201 Display Parameters 97
irradiator 156 display parameters 97
Configuration Disposal
DeviceNet 72 conveyor belts 192
EtherNet/IP 76 magnetrons 192
firmware update 63 printed circuit boards 192
V
Vertical mounting
design example 15
irradiator 15
View
bulb hours 101
EEPROM data 105
factory settings 103
mag hours 101
parameters 97
reflector hours 101
serial numbers 100
software version 102
user settings 104
View/Reset Hour Meter 79
W
Warranty
period 189
voiding 189
WEEE recommendations 191
Heraeus K.K.
Noblelight Fusion Division Phone: +81 3 6902 6602
Sumitomo Fudosan Otowa Fax: +81 3 6902 6613
Bldg. 1F, 2F, 5F email: [email protected]
2-9-3 Otsuka, Bunkyo-ku,
112-0012, Tokyo Japan