Volumetric Infusion Pumps Argus 717 V Argus 718 V: Service Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 68

Volumetric Infusion Pumps ARGUS 717 V en

Service Manual ARGUS 718 V

Firmware from 5.09

14.544_D_SM_en_A71XV_5.09 
Overview

1 5
4

3
7
8 8
9
12

10 11
13
6

Front 10. pressure sensor ‹Upstream›


1. Indicator for mains operation (Battery recharge) 11. pressure sensor ‹Downstream›
2. Status light 12. Automatic ‹free-flow› block
3. Drop light 13. Air bubble detector
4. LC display
5. Soft keys
6. Door handle
7. Drop detector
8. General alarm
9. Drop detector holder

Rear 19
14. Power connection
15. DC connection & interface for a medical PC
16. Nurse call
17. Optional interfaces: RS-232 galvanic isolated; Ethernet
18. Drop detector connection
19. Handle
18

15

14

17 16

2 / 68 14.544_D_SM_en_A71XV_5.09
Table of contents

READ CAREFULLY BEFORE USE 1. Introduction...........................................................................5


1.1. Important Terms and Definition................................................5
KEEP FOR FUTURE REFERENCE 1.2. Intended use............................................................................5
1.3. Safety Notification....................................................................5
1.4. Essential Performance Testing...................................................6
1.5. Accessories..............................................................................6
1.6. System Overview......................................................................7
2. Operating Elements..............................................................8
2.1. Symbols used on the pump......................................................8
2.2. Key functions...........................................................................9
2.3. Softkey functions ....................................................................9
3. CODAN ARGUSservice ARGUS 71x V.................................10
3.1. Compatibility......................................................................... 11
3.2. Installation............................................................................. 11
3.3. Starting CODAN ARGUSservice..............................................12
3.4. Login.....................................................................................12
3.5. Configuration Tools................................................................13
3.6. Tool Box Control....................................................................13
3.6.1. Service Interval.......................................................................14
3.6.2. CODAN ARGUSmedDB Medication Library ............................14
3.6.2.1. Download Medication Library.................................................14
3.6.2.2. Delete Medication Library.......................................................15
3.6.3. Configuration of infusion sets................................................15
3.6.4. Extras.....................................................................................15
3.6.4.1. Reset Configuration...............................................................15
3.6.4.2. Change Serial Number...........................................................16
3.6.4.3. Volume Calibration................................................................16
3.6.4.4. Calibration.............................................................................17
3.6.4.5. Change PIN............................................................................17
3.6.4.6. Battery Information................................................................18
3.6.4.7. Create Report File...................................................................18
3.7. History Control.......................................................................19
3.8. Date & Time Control..............................................................19
3.8.1. How to set Date and Time......................................................20
3.9. Parameter Control..................................................................20
3.9.1. How to change a device parameter configuration..................21
3.9.2. How to store a device configuration in a file ..........................21
3.9.3. How to replicate a pre-saved device configuration..................21
3.10. Flasher Control.......................................................................22
3.10.1. How to Flash..........................................................................23
3.11. Calibration Control.................................................................24
3.12. Get Connections....................................................................24
3.12.1. Status Panel...........................................................................25
3.12.2. Connecting / Disconnecting...................................................26
3.12.3. Clearing a Technical Error.......................................................26
4. Communication Interfaces..................................................27
5. Maintenance........................................................................28
5.1. Maintenance Mode................................................................28
5.1.1. Pump Settings........................................................................30
5.1.2. Volume Calibration................................................................30
5.1.3. Pressure Calibration................................................................32
5.1.4. Keypad & Display Test............................................................34
5.2. Set Calibration.......................................................................34
5.3. Technical Errors......................................................................35
5.4. Coin Cell Battery....................................................................36
5.5. Configuration Mode..............................................................36
5.5.1. Configuration Parameters......................................................37
5.5.2. Nurse Call..............................................................................43

14.544_D_SM_en_A71XV_5.09  3 / 68
6. Replacement of Parts......................................................... 44
6.1. Exploded View ARGUS 71x V complete................................. 44
6.2. Exploded View ARGUS 71x V Casing backside inside..............45
6.3. Exploded View ARGUS 71x V Casing backside outside............46
6.4. Exploded View ARGUS 71x V Casing front inside....................47
6.5. Exploded view of ARGUS 71x V Casing front outside............. 48
6.6. Part Numbers and Descriptions...............................................49
6.7. Technical Error List..................................................................50
6.8. Assembly Instructions.............................................................51
6.8.1. Disassembling of the Housing................................................51
6.8.2. Assembling of the Housing....................................................51
6.8.3. Replacement of Battery Pack..................................................51
6.8.4. Disassembling of the Main Board...........................................51
6.8.5. Assembling of the Main Board...............................................51
6.8.6. Replacement of the Coin Cell.................................................51
6.8.7. Replacement of the Keypad...................................................51
6.8.8. Replacement of Stop Flow Clamp ARGUS 717 V.....................52
6.8.9. Replacement of Anti-Free Flow Clamp ARGUS 718 V..............52
6.8.10. Replacement of Pressure Sensor.............................................52
6.8.11. Replacement of Air Detector..................................................52
6.8.12. Replacement of Window Global Alarm...................................52
6.8.13. Replacement of Interface Board Bluetooth..............................52
6.8.14. Disassembling of Door...........................................................52
6.8.15. Assembling of Door...............................................................52
6.8.16. Disassembling of Pump Unit...................................................52
6.8.17. Assembling of Pump Unit.......................................................52
6.8.18. Replacement of Peristaltic Membrane NBR.............................52
6.8.19. Disassembling of Sensor Board...............................................53
6.8.20. Assembling of Sensor Board...................................................53
6.8.21. Replacement of Stepper Motor..............................................53
6.8.22. Disassembling of Transformer and Mains Receptacle..............53
6.8.23. Assembling of Transformer and Mains Receptacle..................53
6.8.24. Disassembling of Power Board................................................53
6.8.25. Assembling of Power Board....................................................53
6.8.26. Replacement of Edge Board...................................................53
6.8.27. Replacement of Fuse..............................................................53
6.8.28. Replacement of the Combination Clamp................................53
6.8.29. Replacement of the Rubber Feet............................................53
7. Isolation of the Patient.......................................................54
8. Isolation of the User...........................................................55
9. Wiring Diagram...................................................................56
10. Block Diagram.....................................................................57
11. Safety Standard Check (SSC)..............................................58
11.1. Reminder “Safety Check is Due”............................................58
11.2. Volume Control Measurement................................................58
11.3. Pressure Control Measurement...............................................58
11.4. Functional Air Bubble Detection Test......................................59
11.5. Functional Drop Detection Test...............................................59
11.6. Battery Pack Test....................................................................60
11.6.1. Check with CODAN ARGUSservice Utility...............................60
11.6.2. Check without CODAN ARGUSservice Utility..........................60
11.7. Nurse Call Test.......................................................................61
12. Repair Order Form.............................................................. 64
13. Care of the pump................................................................65
13.1. Important notes.....................................................................65
13.2. Cleaning and disinfection.......................................................65
13.3. Storage and transport............................................................65
14. Product Specifications.........................................................66

4 / 68 14.544_D_SM_en_A71XV_5.09
1.  Introduction

This manual contains the latest data available and is subject to and serviced as specified in the instructions for use. The device
changes in accordance with technical improvements. shall be operated by trained medical professionals, under the
It is intended for the exclusive use by authorized persons who continuous supervision of qualified healthcare professionals that
have been trained in maintenance and repair of devices by have been properly instructed and trained in the use of the
CODAN ARGUS AG. The service manual is meant to be used volumetric pump.
together with the user manual. The information in this document The system is intended for use on human adults, juveniles, infants
is relevant to the service personnel (as defined in IEC60601- and neonates.
1:2005+Amd1:2012, see Section „1.1. Important Terms and
Definition“) servicing the device. Other operators (e.g. nurses, Contra-indication
doctors, etc.) handling the device are not obliged to know the The ARGUS 717 V and ARGUS 718 V are not designed and
content of this document. approved:
• for home use.
1.1 Important Terms and Definition • for ambulance or helicopter use.
Service Personnel • in a hyperbaric chamber.
Individual or entity accountable for installing, assembling, • in a MRI environment.
maintaining and repairing the device. • with non-calibrated IV-sets.
• for use by unqualified, untrained staff.
For numbers in decimal notation, a point is used for the separator.
For further information please contact the official distributor or
When ARGUS 71x V is used, both kinds of infusion pumps are the manufacturer’s Customer Service Department.
meant - ARGUS 717 V and ARGUS 718 V.


 1.3 Safety Notification
CODAN ARGUSservice
This is a PC-Windows based service software for the configuration  WARNING
and maintenance of ARGUS 71x V. • Configuration changes with the CODAN ARGUSservice utility,
as well as repairs and manipulations requiring the opening of
Meaning of signal words in this document: the ARGUS 71x V may only be performed by authorized and
trained personnel. Otherwise, patient or user safety is not
 DANGER guaranteed. CODAN ARGUS AG shall not assume any
The signal word DANGER indicates a hazard with a high level of responsibility for any manipulations which have been carried
risk which, if not avoided, will result in death or serious injury. out on the ARGUS 71x V by a non-authorized person.
• To completely isolate the ARGUS 71x V from mains power, it
 WARNING has to be unplugged.
The signal word WARNING indicates a hazard with a medium • The Safety Standard Check (SSC) has to be performed at least
level of risk which, if not avoided, could result in death or serious every 24 months or after 10’000 hours of operation. The SSC
injury. has to be performed in accordance with Section „11. Safety
Standard Check (SSC)“.
 CAUTION • Modifications outside of this service manual are not allowed.
The signal word CAUTION indicates a hazard with a low level of • The pumps should not be used adjacent to or stacked with
risk which, if not avoided, could result in minor or moderate other equipment.
injury.
In case of repair, send the device together with the filled out
Notice repair order form to the local distributor. See Section „12. Repair
The signal word NOTICE serves to advise the user of additional Order Form“. Further information is available at:
important information. There is no correlation with safety-
relevant risks. CODAN ARGUS AG
Oberneuhofstrasse 10
1.2 Intended use CH-6340 Baar
The volumetric pumps ARGUS 717 V and ARGUS 718 V are Phone +41 41 785 09 44
purposed to deliver fluids and medications through any clinically Fax +41 41 785 09 40
accepted route of administration connected to a patient in a [email protected]
predefined way. www.codanargus.com
The devices are intended for infusion therapies including but not
limited to: The latest ARGUS 71x V firmware version and CODAN
• Drugs such as cytostatic agents, anesthetics, etc. ARGUSservice software are available under www.codanargus.
• Blood and blood components com/Login section. First-time users are required to register in
• Total Parenteral Nutrition (TPN) order to enter the Login area of the website.
• Lipids
• Colloids and crystalloids Besides the regular SSC, no additional maintenance is necessary.
The volumetric pumps ARGUS 717 V and ARGUS 718 V are The device does not contain wearing parts.
designed to be safe for continuous operation (24 hours per day)
during expected life-time under the assumption that it is operated

14.544_D_SM_en_A71XV_5.09  5 / 68
1. Introduction

1.4 Essential Performance Testing


The essential performance of the ARGUS 71x V can be tested for
functionality as follows:
• Alarming - during start-up, check to ensure that the status
LED shortly lights up red and a short beep is emitted.
• Nurse Call - ensure that the nurse call configuration parameter
is enabled (see Sections „3.9. Parameter Control“ and „5.5.1.
Configuration Parameters“). Then connect the ARGUS 71x V
to a nurse call system, insert an infusion set and initiate an
alarm on the pump. The nurse call system should indicate an
alarm. If access to a nurse call system is not possible, then
follow the above steps and listen for the clicking sound of the
relay switch from within the ARGUS 71x V which indicates
the activation of the nurse call.

These tests are to be performed as part of the Safety Standard


Check (SSC) (see Sections „1.3. Safety Notification“ and „11.
Safety Standard Check (SSC)“).

1.5 Accessories
The accessories available are listed in the catalogue „Spare parts
&Accessories“. See www.codanargus.com/Downloads.

 WARNING
The ARGUS 71x V may only be used with spare parts and
accessories recommended by CODAN ARGUS AG. The functional
safety of the device is not guaranteed if non approved materials
or spare parts are used. The safety of the patient may be
endangered.

6 / 68 14.544_D_SM_en_A71XV_5.09
1. Introduction

1.6 System Overview


The ARGUS 71x V can be used in different combinations of
systems with CODAN ARGUS docking stations and pumps. For
charging, a docking station ARGUS 60 P, ARGUS 100 P, ARGUS
300 P, ARGUS 500 P and ARGUS 600 P can be used.

For systems with the ability to communicate (e.g. PDMS) the


ARGUS 71x V support the ARGUSprotocol. So they are compatible
to docking stations ARGUS 60 P, ARGUS 100 M, ARGUS 300 M,
ARGUS 500 M, ARGUS 600 M and infusion pumps ARGUS 606
S, ARGUS 600 S and ARGUS 707 V and ARGUS 708 V.

It is possible to connect the pump to a staff alert system (nurse


call system). This can be done directly over the nurse call
connection on the rear side of the pump (see Section “Overview”
enumeration 16.) or over a docking station which is able to
forward the alarm over the interface connection (see Section
“Overview” enumeration 15.)

To configure, the pump can be connected to a medical PC. Typical


connection is over a serial cable to the edge board interface on
the rear side of the pump (see 14. in Section “Overview”). It’s
also possible to establish a connection over LAN (see 17. in
Section “Overview”). For this an ethernet interface board is
necessary.

For use with a barcode scanner a bluetooth interface board is


necessary (depends of the main board version).

Figure 1:  Pump combinations

14.544_D_SM_en_A71XV_5.09  7 / 68
2.  Operating Elements

2.1 Symbols used on the pump

These symbols are attached to the rear of the device on the type
plate or directly on the pump housing. They give important
information on the general use of the device.

CODAN Device name


ARGUS
717V
Specification of flow direction and
recommended temperature range for
infusion solutions
Power voltage / power frequency
230V ~50–
60Hz Holder for drop detector

Power from the mains


12VA The designated terminal is reserved exclusively for
the drop detector!
Drop-protected in a horizontal position up to
IP22 ± 15° inclination Nurse call: additional, immediate alarm system. Can
Protection against solid foreign objects of be connected to the hospital‘s internal nurse call
≥ 12.5mm Ø system
Reference number Option: Communications interface for connection
capability to LAN or serial interface. Contact
CODAN ARGUS AG for further information.
Manufacturing date

Serial number

Fuse type

Battery type

Device contains recyclable materials.


Disposal according to WEEE 2002/96/EC and
country-specific regulations.
Protection level type CF
(level of protection against electric shock)

Protection class 2
Dual insulation

Note: please refer to the accompanying


documentation!

Meets the requirements of MDD 93/42/EEC

Manufacturer

8 / 68 14.544_D_SM_en_A71XV_5.09
2. Operating Elements

2.2 Key functions

ON/OFF KEY
Switch the device on or off

 WARNING
When turning off, all previous parameters of a therapy will be set
to zero.

The pump cannot be switched off while an infusion is running.


This can be deactivated in the configuration. A switch-off delay
can also be configured.

START/STOP and ENTER KEY


• Start or stop infusion
• Activate main or submenu (Select function)
• Confirm input or selection (Enter function)
An activation delay can be configured.

NUMERIC KEYS
These keys are used to enter rate, volume or time. The keys
correspond to the numbers of the display (starting from the
right). The value of the corresponding number is increased by one
unit by repeatedly pressing one of these keys.

CLEAR FUNCTION
Press both keys at the same time to set the value inputs to zero

MENU KEY
Call up basic menu
Mute alarm

2.3 Softkey functions

The function of the soft keys is controlled by the program and


changes depending on the operating status

SELECTION KEY
Select submenu

EXIT KEY
Exit menu or submenu

14.544_D_SM_en_A71XV_5.09  9 / 68
3.  CODAN ARGUSservice ARGUS 71x V

In order to configure the ARGUS 71x V, PC utility CODAN


ARGUSservice has to be used. Please verify that you are in
possession of the latest CODAN ARGUSservice version. For
ARGUS  71x V firmware 5.06, the minimum compatible version
of CODAN ARGUSservice is 5.06. This utility is a Microsoft
Windows based PC software offering various configuration and
maintenance functions. The communication to the PC is via Serial
RS232. For the connection between ARGUS 71x V and PC the
appropriate interface cable (USB: REF: 601 496; DSUB9: REF 601
555) is needed.

The following sections describe the functionality of CODAN


ARGUSservice and how it is to be used. These include:
• View, change and save the current device configuration
• Perform a firmware upgrade/downgrade
• View and save the device history
• Set the service interval
• Set the device time
• Set or change the device access pin
• Change the device serial number
• Create a device report file to send back to CODAN ARGUS AG
• Confirm and reset a technical error on the device

Switch the device off after usage with CODAN ARGUSservice is


completed. Changes to the configuration will not take effect until
the device is rebooted.

 WARNING
The ARGUS 71x V must not be connected to a patient while being
serviced with CODAN ARGUSservice.

In order to use the ARGUS 71x V with CODAN ARGUSservice


make sure that
1. The ARGUS 71x V is connected to mains power.
2. The interface cable is connected to the ARGUS 71x V serial
interface and your PC.
3. The pump is not inserted in a docking station.

10 / 68 14.544_D_SM_en_A71XV_5.09
3. CODAN ARGUSservice ARGUS 71x V

3.1 Compatibility
Use the table below to determine the proper CODAN
ARGUSservice version for your device, depending on the firmware
version of the device.

It is recommended that CODAN ARGUSservice is used with


Microsoft Windows 7 (32bit or 64 bit).

Device Generation Type Device Model Firmware Version of Minimum Compatible CODAN
Device ARGUSservice Version
First generation devices Syringe pumps ARGUS 600 S 4.xx V4
ARGUS 606 S
Volumetric pumps ARGUS 707 V
ARGUS 708 V
Docking stations ARGUS 60 M
ARGUS 100 M
New generation devices Volumetric pumps ARGUS 717 V, 5.04 5.04
ARGUS 718 V 5.05 5.05
5.06 5.06
Docking stations ARGUS 300 M, 5.04 5.06
ARGUS 500 M,
ARGUS 600 M

The ideal display size setting in Windows is smaller 100%


(default). This can be set under:
Start -> Control Panel -> Appearance and Personalization ->
Adjust screen resolution (http://windows.microsoft.com/en-us/
windows7/Make-the-text-on-your-screen-larger-or-smaller?v=t).
If another display setting is used, some of the buttons might not
appear correctly or not at all.

3.2 Installation
The newest version of CODAN ARGUSservice can be downloaded
from the CODAN ARGUS AG website (www.codanargus.com/
Login) On the website, go to the Login page and either register
or enter your user name and password to gain access to the
Login section. (Figure 2) Figure 2:  CODAN ARGUSservice setup icon

To install the CODAN ARGUSservice software, just double click


the installation binary (Figure 2).

Once the installation starts, continue to follow the on-screen


instructions.
When asked to select the components for installation, there is
the possibility to install only the components for the CODAN
ARGUSservice utility. The installation of the service tool for first
generation devices (see Section „3.1. Compatibility“) is optional
(Figure 3).
The option to install the driver for the USB interface cable is also
given. If the driver is already installed, this does not have to be
selected again.

Notice
Sometimes the driver for USB interface cable does not install,
even if selected. If this should be the case, then manually install
the driver after the installation of CODAN ARGUSservice is
complete. This can be done by going to the ARGUS folder in the
Windows start menu and selecting “Installation interface cable
driver” (Figure 3).
Figure 3:  Installation type selection

14.544_D_SM_en_A71XV_5.09  11 / 68
3. CODAN ARGUSservice ARGUS 71x V

3.3 Starting CODAN ARGUSservice


After the installation, CODAN ARGUSservice can be started using
either the desktop short-cut or quick launch icon, depending on
the installation options chosen (Figure 4).

When started, a window appears where you may select the Figure 4:  left: Desktop short-cut; right: Start menu short-cut
appropriate CODAN ARGUSservice version to start (Figure 4). The
selection depends on the generation of the device which is to be
serviced (see table in Section „3.1. Compatibility“).

Figure 5:  CODAN ARGUSservice selection at start-up

3.4 Login
After selecting new generation devices, a login window will
appear (Figure 6).

Login with the following credentials:


User: Administrator
Pin: 1220

The above given pin is defined as the default pin code. The pin
can be changed by the administrator. After logging in successfully,
the Configuration Tools page appears.

Notice Figure 6:  CODAN ARGUSservice login


It is possible to configure each device with a separate PIN. That
means, if you login, only devices with the same login data are
accessible. For connection to devices with different login data, it
is necessary to login on each device individually.

12 / 68 14.544_D_SM_en_A71XV_5.09
3. CODAN ARGUSservice ARGUS 71x V

3.5 Configuration Tools


The Configuration Tools page shows the various controls (utilities)
that are available (Figure 7). To enter a control, just click on the
control. Before being able to use the functions within a control,
a connection needs to be established to at least one device. (The
following enumeration refers to the numbered items in Figure 7.)

1. Set background colour of application.


2. Application configuration, application information, minimize,
maximize, close.
3. Login: Possibility to access the login window (see Section
„3.4. Login“and change the login settings.
4. Current version of CODAN ARGUSservice.
5. Title of current page.
6. Tool Box control: services such as medication library download,
change pin, set configuration and calibration (see Section
„3.6. Tool Box Control“).
7. History control: View and save the device history (see Section
Figure 7:  CODAN ARGUSservice Configuration Tools page
„3.7. History Control“).
8. Date & Times control: Set date and time on the device (see
Section „3.8. Date & Time Control“).
9. Parameter control: View, change and save device configuration
(see Section „3.9. Parameter Control“).
10. Flasher control: Upgrade and downgrade the device firmware
(see Section „3.10. Flasher Control“).
11. Calibration Control: calibrate the device (see Section „3.11.
Calibration Control“).

3.6 Tool Box Control


The Tool Box control (Figure 8) provides various functionalities to
change the settings on a device. (The following enumeration
refers to the numbered items in Figure 8.)

1. Get Connections: manages the connections to the attached


devices (see Section „3.12. Get Connections“)
2. The connection overview panel shows the device connections
that are selected.
3. Exit button: Click on this button to exit current screen.
4. Illustration of the attached device.
5. Device information: This includes device type, the COM port
number to which the device is attached and the serial number
and firmware version on the device.
6. Device operating information: For the ARGUS 71x V this is
total operating hours and infused volume.
7. Downloaded medication database information. Figure 8:  CODAN ARGUSservice Tool Box control page
8. Medication database utility: see Section „3.6.2. CODAN
ARGUSmedDB Medication Library“.
9. Configuration utility for infusion sets: see Section „3.6.3.
Configuration of infusion sets“.
10. Service interval utility: This utility sets the criteria for the next
safety standard check (see Section „3.6.1. Service Interval“).
11. Extra tools: see Section „3.6.4. Extras“.

14.544_D_SM_en_A71XV_5.09  13 / 68
3. CODAN ARGUSservice ARGUS 71x V

3.6.1 Service Interval


Use the Service Interval utility to set the date for the next safety
standard check (SSC). The reminder can be set to appear after a
certain number of calendar months (recommended 24) and/or
after a certain number of operational hours (recommended
10’000). If both criteria are set, the reminder will appear after
whichever expires first. If both criteria are set to 0, the reminder
is disabled. When clicking on Set Reminder, the “Last SSC” date
is set to the current date.

To open service interval utility, click on the Service Interval button


and the screen shown in Figure 8 will appear. (The following
enumeration refers to the numbered items in Figure 9).
1. Current: Shows the date of when the last SSC was performed
and when the next SSC (if months is enabled) will be due. On
a new device where no SSC has yet been performed, both
dates will be n/a.
2. Service interval in calendar months: This is the number of
calendar months after which the reminder to perform a safety
standard check will appear. If set to 0, this criteria is disabled.
3. Service interval in operational hours: This is the number
operational hours after which the reminder to perform a Figure 9:  CODAN ARGUSservice SSC reminder
safety standard check will appear. If set to 0, this criteria is
disabled.
4. Next SSC: This is the date for the next SSC. 

5. Set Reminder: This sets the criteria for the next SSC. The
above mentioned interval will be downloaded to the device.
After clicking on Set Reminder, wait until the download is
complete.
6. Exit button: Click on this button to exit the SSC screen.

Notice
The expiration of the specified service interval time will be
indicated by
• Docking station: a short beep once an hour and the alternate
green-orange blinking of the Status LED.
• ARGUS 71x V: a notification after the start-up animation.
This continues until the SSC is performed and a new reminder is
set.

3.6.2 CODAN ARGUSmedDB Medication Library


If a medication library is available, it can be download on device
or deleted using the MedDB utility. Please contact your local
distributor or the manufacturer.

3.6.2.1 Download Medication Library


A new medication library can be downloaded to the device. It is
Figure 10:  Downloading MedDB
not necessary to delete any previous CODAN ARGUSmedDB
version.

1. Ensure that a connection to the desired device exists and that


the device is connected to mains power and will remain so
during the entire flashing process.
2. Click on the MedDB button and then on Download MedDB.
3. In the window that appears, select the medication library file
and click on open.
4. During the flashing process of the medication library, please
do not disconnect the device.
5. The download process is completed when the pop-up
window “Writing MedDB to Device Finish“ appears.

14 / 68 14.544_D_SM_en_A71XV_5.09
3. CODAN ARGUSservice ARGUS 71x V

3.6.2.2 Delete Medication Library


This function deletes the medication library on the device.

1. Press on the MedDB button and select Delete MedDB.


2. Follow the on-screen instructions.
3. During the deleting process of the medication library, please
do not disconnect the device.
4. The deleting process is completed when the pop-up window
“Erasing MedDB Finish“ appears.

3.6.3 Configuration of infusion sets


1. Available Sets: List of available infusion sets can be loaded
from firmware (see 7.). Up to 4 infusion sets can be dragged
and dropped from “Available Sets” to “Sets on device”
2. Infusion set recommended by CODAN ARGUS AG
3. Sets on Device: A list of max. 4 infusion sets which are
configured on the pump.
4. The check mark indicates which set is selected for modifying
the parameters (see 6.).
5. The red M symbol indicates that the set has been modified or
added.
6. Parameter: List of parameters of selected infusion set.
7. Load Set’s from File: The newest firmware file contains the
latest infusion set list
8. Restore Set Configuration: This button resets the modified
but not downloaded parameters and infusion sets (see 6.) in Figure 11:  Cofiguration of IV-Sets page
“Sets on Device” (see 3.).
9. Download Set Configuration to Device: The button downloads
the set configuration onto the device.
10. Go back to Tool Box screen.
11. Recycle bin: It is possible to remove infusion sets from Sets on
Device. Move the infusion set to the recycle bin with drag
and drop.

3.6.4 Extras
Under Extras (Figure 12), there are additional utilities.

3.6.4.1 Reset Configuration


To reset to the customer configuration, click on Reset
Configuration (Figure 12).

Figure 12:  Tool Box control extras for ARGUS 71x V

14.544_D_SM_en_A71XV_5.09  15 / 68
3. CODAN ARGUSservice ARGUS 71x V

3.6.4.2 Change Serial Number


Changing the serial number of a device is needed after a change
of the mainboard. For this procedure use the Change Serial
Number utility. (The following enumeration refers to the
numbered items in Figure 13).

1. The current serial number on the device is already listed (in


normal case 0).
2. Enter the new serial number e.g. 10000560.
3. A code is generated e.g. 0-3-10000560. Contact CODAN
ARGUS AG with the generated code to receive a permission
key.
4. Enter the received permission key.
5. Click on Program Serial Number.
6. Exit button: Click on this button to exit current screen without
changing the serial number.

3.6.4.3 Volume Calibration Figure 13:  Change device serial number


Use the Volume Calibration utility to set the pump’s specific
correction factors. The calibration can either be performed based
on a rate measurement or a volume measurement. The calibration
is specific to an infusion set type. This is an alternative method
to perform the calibration on the pump directly.

1. Calibration type: Click here to switch between volume and


rate calibration.
2. The IV Set panel provides calibration information about the
set.
3. Name of the infusion set.
4. Check mark: Only visible if selected.
5. Valid calibration range.
6. Current calibration value and new calibration value.
7. Target: the desired target value of the calibration
measurement. Volume calibration is in ml, rate calibration is
in ml/h.
8. Actual: the actual measured value. Volume calibration is in ml
or g, rate calibration is in ml/h. Figure 14:  Volume calibration page
9. Deviation: The percentage deviation between desired target
value and actual measured value is calculated by CODAN
ARGUSservice: Formula: % deviation = (actual-target)/
actual×100
10. Reset correction for selected set.
11. Download correction value to pump.
12. Apply correction value to select infusion set.

16 / 68 14.544_D_SM_en_A71XV_5.09
3. CODAN ARGUSservice ARGUS 71x V

3.6.4.4 Calibration
Use a balance with a minimum resolution of 0,1g.
1. Switch the pump on. The pump must be in the regular
infusion mode.
2. Insert the infusion set and close the front door.
3. Verify that the inserted infusion set corresponds with the
name being displayed on the pump.
4. Open roller clamp.
5. Purge the infusion set.
6. Set the balance to zero by pressing TARA.
7. Start the infusion at a rate of 250 ml/h and a VTBI of 30 ml.
The calibration time is therefore approximately 7 minutes.
8. Wait for the alarm “END INFUSION”.
9. The resulting weight is needed for setting the correction
factor in the toolbox of the CODAN ARGUSservice utility.
10. Change to the configuration mode of the pump.
11. Select the desired calibration type, either Volume or Rate.
12. Enter 30ml as target and the measured value as actual value.
13. The deviation is automatically calculated. It may not be
greater than +/- 15%.
14. Select the infusion set for which the calibration is being
performed.
15. Click on Apply Correction to Set.
16. Click on Download Correction to Device.

3.6.4.5 Change PIN


Use the Change PIN utility (Figure 15) to change the pin for a user.
Currently, only the pin for the user Administrator can be changed
(see Section „3.4. Login“). To open the utility, click on Change
PIN in Figure 12. (The following enumeration refers to the
numbered items in Figure 15, as well as being sequences of the
steps to follow in order to change the device PIN).

1. From the drop-down list under User, select the user for which
the pin is to be changed.
2. Under Old pin enter the current pin for the selected user.
3. Show PIN: Enabling this check-box, makes the pins visible.
4. Enter a new PIN. The PIN has to be exactly 4 digits long.
5. Enter the new PIN again to confirm it.
6. Confirm the change with OK. OK button will be enabled if
both poins match.
7. Or press cancel to exit the window without changing the PIN. Figure 15:  Change device login PIN

If the information was entered correctly, a confirmation pop-up


window will appear (Figure 16), otherwise, an error message is
shown on the top (Figure 17).

Notice
In the current releases of CODAN ARGUSservice (Version 5.06
and older) there is a bug which does not allow the proper
firmware upgrade of a device when the Administrator PIN is not
the default value (1220).
Figure 16:  Successful PIN change

To recover a lost PIN send the report file (see Section „3.6.4.7.
Create Report File“) to CODAN ARGUS AG. Then the PIN will be
sent to you.

Figure 17:  Invalid PIN error message

14.544_D_SM_en_A71XV_5.09  17 / 68
3. CODAN ARGUSservice ARGUS 71x V

3.6.4.6 Battery Information


The Battery Information shows the battery charge state.

Main Power States Device is connected to a power supply


(mains, external DC supply or docking)
charge cond Charge battery using conditional
charging current (deeply discharged
battery)
charge normal Charge battery using normal charging
current
charge trickle Charge battery using trickle charging
current (full battery)
charge overtemp Stop charging if battery is overheated
(and let it cool down) Figure 18:  Battery charge state

Battery Power Device is running on battery power


States
discharge normal Discharge battery (battery charge state
above pre-alarm level)
discharge low Discharge battery (battery charge state
below pre-alarm level, but not empty)
discharge empty Discharge battery (battery is empty)

3.6.4.7 Create Report File


Use the Create Report File utility (Figure 12) to generate a
complete report of the device which can be sent to CODAN
ARGUS AG. The report contains the current setting of all device
parameters and states, as well as the device history. In case of an
error or problem with the device, this report file is helpful in
analysing the issue.

After a complete download of all device parameters (this may last


several minutes) (Figure 20), a window will appear to enter a file
name for the report (Figure 19). Save the report with a desired
file name and then send it to CODAN ARGUS AG with a problem
description to be analysed. Figure 19:  Enter file name for saving report file

Figure 20:  Progress window when creating report file

18 / 68 14.544_D_SM_en_A71XV_5.09
3. CODAN ARGUSservice ARGUS 71x V

3.7 History Control


Use the History Control (Figure 21) to access the device history.
The history stores up to the last 1000 events. In case of a total
loss of power, the data in the history is saved. It is also possible
to:

• Read the history continuously, while the device is operating


(however without a connection to a patient).
• Save the device history to a file, as well as a PDF.
• View history files that have been saved previously.

(The following enumeration refers to the numbered items in


Figure 21).
1. Get Connections: It manages the connections to the attached
devices (see Section „3.12. Get Connections“).
2. Connection overview: The overview shows the selected
devices from the Get Connections control. The COM port
with the blue background illustrates the device with the
currently displayed history.
3. Exit button: Click on this button to exit the History Control.
4. The history list: The newest events are listed on top.
5. Read history continuously: When checked, the history will be
read continuously from the device and the history list will be
updated with the newest events.
6. Description: Optionally add a description here before saving
the history.
7. Reload History: Update the history list.
8. Save History: Save history to a file. Do not forget to add a
description.
9. Open History-File: Open a history file, that has previously been Figure 21:  CODAN ARGUSservice History Control
saved, to be viewed.
10. Create PDF-File: Save the history file as a PDF, which can then
be printed or be viewed without the use of CODAN
ARGUSservice.
11. Device information of currently active connection.

3.8 Date & Time Control


Use the Date & Time Control (Figure 22) to set the date and time
on one or more devices. The date and time on the selected
devices will be set to the system time of the PC. (The following
enumeration refers to the numbered items in Figure 22).

1. Get Connections: It manages the connections to the attached


devices (see Section „3.12. Get Connections“).
2. COM port number of connected and selected devices.

3. Exit button: Click on this button to exit the Date & Time
Control.
4. COM port of device.
5. Synchronize device: This button sets the date and time on the
device according to the date and time values shown in 6 of
Figure 22.
6. Current date and time on device.
7. Synchronize all devices: This button sets the date and time on
all selected devices according to the date and time values Figure 22:  CODAN ARGUSservice Date & Time Control
shown in 6 of Figure 22.
8. The date and time value that will be set on the device(s). The
date and time on the selected devices will be set to the system
time of the PC.

14.544_D_SM_en_A71XV_5.09  19 / 68
3. CODAN ARGUSservice ARGUS 71x V

3.6.8.1 How to set Date and Time


The following steps are to be carried out to set the date and time:

1. Use the Get Connections control to select the desired


device(s).
2. Ensure that the correct date and time is shown in 6 of Figure
22. It is essential that the time zone, time and date on the PC
is set correctly to the location where the device will be used.
3. To program only one individual device, click on “Synchronize
device”. To program all selected devices, click on “Synchronize
all devices”.

3.9 Parameter Control


Use the Parameter Control (Figure 23) to view, change, and save
the parameter settings on a selected device.

The parameters are listed in a parameter tree according to their


associated category. For each parameter the following are visible:
name, value, unit and a description. (The following enumeration
refers to the numbered items in Figure 23).

1. Get Connections: It manages the connections to the attached


devices (see Section „3.12. Get Connections“).
2. Connection overview: The connection overview consists of all
currently available devices, which were selected in the Get
Connections control. The COM port with the blue boundary
shows the connection to the device with the currently
displayed configuration.
3. Exit button: Click on this button to exit the Parameter Control.
4. Device Information: Information about the device associated Figure 23:  CODAN ARGUSservice Parameter Control
with the configuration currently being shown.
5. Configuration type: the configuration that is currently
displayed.
6. Parameter tree: To view the parameters of a category, click on
the “+” sign next to the category.
7. Parameter with default settings. The background colour is
light grey.
8. Parameter with non-default values. The background colour is
light orange.
9. Parameter that is currently selected for editing. The selected
parameter is highlighted.
10. Parameter Information: A complete meta-information about
the selected parameter.
11. Upload Configuration: Upload the current parameter
configuration from the device.
12. Download Configuration: Download the parameter
configuration to the device.
13. Save Configuration: Save the parameter configuration to a
file. The configuration can be saved to be loaded at a later
point in time.
14. Load Configuration: Load a parameter configuration from a
file. The configuration can then be changed and/or be
downloaded to a device.
15. Create PDF: Create a PDF version of the current parameter
configuration.
16. Replicate Configuration: Allows the configuration of one
device to be downloaded to other connected devices.

20 / 68 14.544_D_SM_en_A71XV_5.09
3. CODAN ARGUSservice ARGUS 71x V

3.9.1 How to change a device parameter configur-


ation
The following instruction steps show how a configuration is
uploaded from a device. It explains also how to edit and download
the configuration to the device again.

1. In the “Device Connection” click on the device, which is to


be configured.

2. Click on “Upload Configuration”.
3. Wait for the configuration to be uploaded. The complete
configuration can be seen in the parameter tree.
4. Select the desired parameter to change and edit the value.
The “Parameter Information” provides the range and
increment step size for the value.
5. After all desired parameters are modified, click on “Download
Configuration” to program the device with the new
parameter values. Only parameter values that have been
changed are downloaded.

Notice
After downloading the configuration successfully with CODAN
ARGUSservice, switch off the device to save the changed
parameters.

3.9.2 How to store a device configuration in a file


It is possible to save the configuration of a device to a file. This
file can be used to configure other device (of the same type) (see
Section „3.9.3. How to replicate a pre-saved device
configuration“).

1. Click on the target device in the “Connection Overview” for


which the configuration is to be saved.
2. Click on “Upload Configuration”.
3. Wait for configuration to upload completely.
4. Click “Save Configuration”.
5. Enter a file-name and select the location.
6. Click “Save”.

3.9.3 How to replicate a pre-saved device config-


uration
After saving a device configuration to a file, it is possible to
download this file to another device (of the same type) at a later
point in time.

1. Create a configuration file according to Section „3.9.2. How


to store a device configuration in a file“.
2. Click on the target device in the “Connection Overview” for
which the configuration is to be downloaded.
3. Click on “Load Configuration”.
4. Select the pre-saved configuration file.
5. Click on “Download Configuration”.
6. Wait for the download to complete.

14.544_D_SM_en_A71XV_5.09  21 / 68
3. CODAN ARGUSservice ARGUS 71x V

3.10 Flasher Control


The Flasher Control (Figure 24) permits flashing of a new firmware
file to a device. It is possible to upgrade or downgrade a firmware.
(The following enumeration refers to the numbered items in
Figure 24).

1. Get Connections: It manages the connections to the attached


devices (see Section „3.12. Get Connections“).
2. COM port number of connected and selected devices.
3. Exit button: Click on this button to exit the Flasher Control.
4. Hex-File(s): The firmware file selected for the flashing process.
5. Device Information: Information about a connected device,
including COM port, firmware version that is currently
running on the device, and new firmware version that has
been selected for the upgrade.
6. Progress bar of the flashing process.
7. Start flashing: This button starts the upgrade if a valid
firmware file is selected. Figure 24:  CODAN ARGUSservice Flasher Control
8. Open FW File: select the firmware file for flashing.
9. Clear Memory: clear the selected firmware file.
10. Flash all: Can flash the firmware file on all devices together
at the same time.

 WARNING
Do not make any software updates as long as the device is in use
and/or connected to a patient!

CAUTION
A standard safety check (SSC) has to be performed after every
firmware update!

Notice
• In the current release of CODAN ARGUSservice (Version 5.06
and earlier), there is a bug which does not allow the proper
firmware upgrade of a device when the Administrator pin is
not the default value. Therefore, before starting the flash
procedure, ensure the Administrator pin on the device which
is to be updated is the default value (default value is 1220).
• Before flashing, ensure that the device is connected to mains
power.
• During the flashing process, do not remove the serial port
connection or the connection to mains power.
• An interrupted or cancelled flash procedure might make a
restart of the device impossible. It is therefore forbidden to
manually switch off or reboot the ARGUS 71x V during a flash
procedure.
• The current firmware version of the device is shown by getting
connection from CODAN ARGUSservice to the device.

22 / 68 14.544_D_SM_en_A71XV_5.09
3. CODAN ARGUSservice ARGUS 71x V

3.10.1 How to Flash

Steps to flash:
1. Ensure that a connection to the desired device exists and the
device is listed in the Flasher tool. Also ensure that the device
is connected to mains power and will remain so during the
entire flashing process.
2. Click “Open FW File”.
3. Select the desired firmware file and confirm with open.
4. Confirm any pop-up windows that might appear.
5. Check the correct name and path for the selected firmware
file listed in “Hex-File(s)”.
6. In the device status window of each individual device, the
current firmware version and the new firmware version (after
flashing) are listed. Verify these.
7. Start the flashing process by clicking on “Start flashing” (7 in
Figure 24) or “Flash All”.
8. Select language (Figure 25)
9. The flashing process consists of the states shown in Figure 26.
10. During the flashing process, the device will reboot. The
flashing process is not complete yet.
11. After flashing is completed, the status bar for the device will Figure 25:  Select Language
indicate “Flashing finish”. The device will be in the
configuration mode again.
12. Check in configuration mode if proper firmware version was
flashed.

Figure 26:  CODAN ARGUSservice device flashing process

14.544_D_SM_en_A71XV_5.09  23 / 68
3. CODAN ARGUSservice ARGUS 71x V

3.11 Calibration Control


The calibration control displays the calibrated values with ranges
and information from each infusion set.

1. Get Connections: It manages the connections to the attached


devices (see Section „3.12. Get Connections“)
2. COM port number of connected and selected devices.
3. Set 1: Number of selected infusion set
4. Pressure Calibration: Contains calibrated values (see 3. Until
7.), date of last pressure calibration and if calibration was
successful.
5. CAL_SET_1_DOWNSTREAM_SENSITIVITY: Downstream
pressure sensitivity (0 – 3300 mV/bar)
6. CAL_SET_1_DOWNSTREAM_POTI_GAIN: Digital
potentiometer position (0 - 127) for gain calibration of
downstream pressure sensor.
7. CAL_SET_1_DOWNSTREAM_POTI_OFFSET: Digital
potentiometer position (0 - 127) for offset calibration of
downstream pressure sensor.
8. CAL_SET_1_UPSTREAM_POTI_GAIN: Digital potentiometer
position (0 - 127) for gain calibration of upstream pressure
sensor.
9. CAL_SET_1_UPSTREAM_POTI_OFFSET: Digital potentiometer
Figure 27:  Calibration Control
position (0 - 127) for offset calibration of upstream pressure
sensor.
10. Volume Calibration: Contains correction factor (see 9.) and if
calibration was successful.
11. CAL_SET_1_RATE_CALIBRATION_CORRECTION: Correction
factor (800 - 1200) for volume calibration.
12. Set selection: Selection of desired infusion sets.
13. Create PDF-File: Save the calibration as a PDF, which can then
be printed or be viewed without the use of CODAN
ARGUSservice.
14. Exit button: Click on this button to exit the Calibration
Control.

3.12 Get Connections


The Get Connections control (Figure 28) is used to manage the
connections to the attached devices. It is accessed by opening
one of the controls in the Configuration Tools and then by
clicking on Get Connections. Only COM (Serial) port connections
are supported currently. The connection can either be made
through a physical or a virtual COM port. (The following
enumeration refers to the numbered items in Figure 28).

The overview shows all available COM ports on the system.


Status panel: For each COM port, there is a status panel (see
Section „3.12.1. Status Panel“).
1. COM port list.
2. No device on COM port.
3. Connected device on COM port.
4. Connected and selected device on COM port.
5. Search & select all devices: This functionality goes through all
COM ports and determines if a known device is attached. If Figure 28:  CODAN ARGUSservice Get Connections page
so, a connection is established to the device and the device
is selected.
6. Disconnect selected: This button terminates the connection
to all selected devices.
7. Disconnect all: This button terminates the connection of all
attached devices.
8. Exit button: Click on this button to exit the current control
and return to the previous control.

24 / 68 14.544_D_SM_en_A71XV_5.09
3. CODAN ARGUSservice ARGUS 71x V

Notice
If a device with a valid connection is switched off manually and
switched on again, the connection to this device needs to be
re-established in the Get Connections control. A manual
reconnection needs to be performed for the specific device.

3.12.1 Status Panel


The status panel (Figure 29) provides the following information
(The following enumeration refers to the numbered items in
Figure 29):

1. Device login state:


If the device login state is blue, login was successful.
If the device login state is grey, login failed.
2. Identification of COM port.
3. The check-mark symbol indicates a selected connection.
4. Device information of connected device: type, serial number,
firmware version.
5. Status and baud rate of connection. Figure 29:  Device connection status panel

Additionally, the colour of the panel reflects the state of the


connection:

D
 ark red: Technical error on device or
Grey: Disconnected an error with COM port

Light green: Changing state, either connecting


or disconnecting Pink: Unknown state

Dark Green: Connected

Notice
In order to be able to use a device with CODAN ARGUSservice,
there needs to be a connection to the device AND the connection
needs to be selected (3 in Figure 29).
It is possible to be disconnected with an error. In this case the
colour of the panel will be red, but the status (5 in Figure 29) will
indicate disconnected.

14.544_D_SM_en_A71XV_5.09  25 / 68
3. CODAN ARGUSservice ARGUS 71x V

3.12.2 Connecting / Disconnecting


It is possible to connect or disconnect all attached devices at once
or individually.

Connect / disconnect to all devices:


1. Click “Search & Select All Devices” to connect to and select
all attached devices
2. Click “Disconnect All” to disconnect all devices at once

Connect to an individual device:


1. Left click on the status panel of respective device (Figure 30).

Reconnect / disconnect to an individual device: Figure 30:  Right click options of device status panel
1. Right click on the status panel of respective device (Figure 30).
2. From the drop down list, select the desired action, either
“Disconnect” or “Reconnect”.

3.12.3 Clearing a Technical Error


If a device has a technical error, the error can be cleared by using
CODAN ARGUSservice:

1. Make sure the device was turned on and off at least once
since the technical error occurred.
2. Connect to device.
3. Right click on the panel of the device (Figure 31).
4. Select “Reset Techerror”.
5. Wait for the technical error to be cleared and a report file to
be created (Figure 32). Enter a file name for the report file. It
can take several minutes to gather the information to create
the report file.
6. Send this report file together with a problem description to
your local CODAN ARGUS representative or directly to
CODAN ARGUS AG. Figure 31:  Resolving a technical error

Figure 32:  Creating report file after technical error

26 / 68 14.544_D_SM_en_A71XV_5.09
4.  Communication Interfaces

The device has one RS232 serial interface and one Ethernet
interface to connect to external systems. Over both interfaces,
an external system can communicate with the device and the
inserted pumps using the ARGUSprotocol.

Using the ARGUSprotocol, the status and current run-time


information from the devices can be requested. For more detailed
information on this feature, please contact your local distributor
or CODAN ARGUS AG.

 WARNING
Connect to ARGUS 71x V only to external (non CODAN ARGUS)
devices which comply with IEC60950-1:2005.

14.544_D_SM_en_A71XV_5.09  27 / 68
5.  Maintenance

5.1 Maintenance Mode

Entering the Maintenance Mode:


1. Keep the 2nd numeric key depressed and press the ON/OFF
key in order to get access to the configuration mode.
2. Enter the Administrator PIN to log in. The Administrator PIN
is set to 1220 by default but it can be changed with the
CODAN ARGUSservice utility, see Section „3.6.4.5. Change
PIN“. The PIN code shall prevent unauthorized use.
3. The pump settings menu is shown after login.
4. Use the EXIT Soft Key to leave a menu or sub-menu to get to
PUMP SETTINGS.

Menu Overview:
This overview describes all available menus and their sequence
of availability.
The Menu key guides the user through all available menus. Press
the Menu key until the desired menu is displayed.

Pump Settings
In the Pump Settings Menu some configuration parameters can
be set permanently and directly without using CODAN
ARGUSservice.

Volume Calibration
This menu is a part of the set calibration and guides the service
technician through the volume calibration process (see Section
„5.1.2. Volume Calibration“).

Pressure Calibration
This menu is a part of the set calibration and guides the service
technician through the pressure calibration process (see Section
„5.1.3. Pressure Calibration“).

Special Functions
Permanent selection of one special function. Just one special
function is possible at the same time. This menu is also accessible
in normal mode. If one of the functions is activated in the
maintenance mode, then the user is not able to change this
settings in normal mode.
• Delayed infusion
• Barcode mode

28 / 68 14.544_D_SM_en_A71XV_5.09
5. Maintenance

Keypad & Display Test


This menu is split into 3 user interface tests. Each of these tests
guides the service technician through the process (see Section
„5.1.4. Keypad & Display Test“).

Battery Info
Information about the remaining battery capacity in % and in
hours/minutes. If the specified times are no longer achieved, the
Technical Service Department can replace the batteries.

Info Calibration Status


This menu shows the calibration status of the configured sets. A
tick means: calibration done, x means: not calibrated. Infusion
sets must be pressure and volume calibrated in order to be used.
Column 1: Infusion set
Column 2: Pressure calibration status
Column 3: Volume calibration status

Info Medication Database


The medication library downloaded to the device can be
identified.

Version Info
Column 1: Firmware Release with Version and Revision
Column 2: Bootloader Release with Version and Revision
Column 3: Serial number of the device

14.544_D_SM_en_A71XV_5.09  29 / 68
5. Maintenance

5.1.1 Pump Settings


The following pump settings are adjustable:

Setting Default Range/Unit Description


Value
DISPLAY 7 1 – 15 Display contrast
CONTRAST
DISPLAY 240 0 – 255 Display backlight brightness
BACKLIGHT
DEF. BUZZER 10 1 – 10 Buzzer loudness. The volume cannot be turned off!
VOLUME
DEF. SPEAKER 10 1 – 10 Speaker loudness. The volume cannot be turned off!
VOLUME
LANGUAGE English English or 2nd Language selection of English or a secondary language for user interface in normal mode.
Language
UPSTREAM ON ON/OFF UPSTREAM OFF: The deactivation of the upstream sensor makes the use of a drop
SENSOR detector compulsory!
UPSTREAM ON: The pump may be used with (recommended) or without drop detector.

Notice
The buzzer and speaker volume may also be changed temporarily
during normal infusion. See user manual “14.2.1. Changing the
volume of the audible signal (buzzer and speakers”.

5.1.2 Volume Calibration


Enter the configuration mode and go to the menu “Volume
Calibration”:
1. Press Start.

2. Select the infusion set and press Start.

30 / 68 14.544_D_SM_en_A71XV_5.09
5. Maintenance

3. Insert the infusion set and press Start. The infusion starts
and the volume increases to 5.0 ml.

4. Set the balance to zero by pressing “TARA” and press Start.


The infusion starts and the volume increases to 30.0 ml.

5. Check the result of the balance.

6. Enter now the value of the balance (e.g. 29.80g) and press
Start.

7. Press Start to store the correction factor.

8. YES: Press Start and continue with step 9.


NO: Press Start and repeat from step 1.

9. Use the EXIT key to go back to “Pump Settings”.

14.544_D_SM_en_A71XV_5.09  31 / 68
5. Maintenance

5.1.3 Pressure Calibration


Start pump in configuration mode and go to the menu “Pressure
Calibration”:
1. Press Start

2. Select the infusion set and press Start.

3. Open the door, remove the infusion set and press Start.

4. The signals at 0 bar for up- and downstream sensor are


displayed. Press Start.

5. Insert the infusion set, close door, open the 3-way stop
cock and press Start.

6. Close the 3-way stop cock. Press Start as soon as the


manometer reaches 1.0 bar.

32 / 68 14.544_D_SM_en_A71XV_5.09
5. Maintenance

7. Open the 3-way stop cock and release the pressure. Close
the 3-way stop cock again and press Start as soon as the
manometer reaches 0.2 bar.

8. Press Start as soon as the manometer reaches 1.2 bar


The plausibility screen is displayed.

9. Compare the value of plausibility check with the value of


the manometer (±100 mbar) and press Start.

10. Remove the infusion set and press Start.

11. The 0.0 bar calibration done screen is displayed and press
Start.

12. Confirm calibration with Start.

13. Press the EXIT Key.

14.544_D_SM_en_A71XV_5.09  33 / 68
5. Maintenance

5.1.4 Keypad & Display Test


This menu is designed for the intuitive performance of a safety
standard check (SSC). The aim of this test is to verify the correct
operation of the user interface including keypad and display.

Select all tests one by one and follow the instructions shown on
the display.

DISPLAY TEST This test detects possible defective pixels. The check contains of two steps. The correct function of
the display is confirmed by a black display at the end of this test.
KEYPAD TEST This test verifies the correct function of all keys except the ON/OFF Key. This one is automatically
checked at every switching on and off of the pump. The test is successfully completed when every
key was pressed once and when the check-mark is displayed.
LED AND BACKLIGHT This test verifies all light media. The test is successfully completed if all light media have shone as
indicated on the display.

5.2 Set Calibration


The set calibration consists of volume and pressure calibration
and is necessary after configuration of a new infusion set.

The volumetric infusion pumps ARGUS 71x V contain two


pressure sensors. The upstream sensor on the side of the
medication container (left input) and the downstream sensor on
the patient’s side (right output). The pressure calibration can be
performed directly on the pump (see Section „5.1.3. Pressure
Calibration“). After each pressure calibration a control
measurement is recommended (see Section „11.3. Pressure
Control Measurement“).

For pressure calibration and control measurement the following


equipment is needed:
• a manometer with a scaling of 0.1 bar
• a 3-way stop cock and a clamping shears
• the chosen infusion set

For the volume calibration, there are two ways to calibrate. It is


possible to either enter the correction factor directly on the device
(see Section „5.1.2. Volume Calibration“) or to use the CODAN
ARGUSservice PC utility tool (see Section „3.6.4.3. Volume
Calibration“). After each volume calibration a control
measurement is recommended (see Section „11.2. Volume
Control Measurement“).

For volume calibration and control measurement the following


equipment is needed:
• a scale with a minimum resolution of 0.1g
• the chosen infusion set

This notification in normal mode reminds the user to calibrate the


infusion set:

34 / 68 14.544_D_SM_en_A71XV_5.09
5. Maintenance

Notice
• The ARGUS 71x V can be used with a maximum of 4 different
infusion sets.
• Before an infusion set can be calibrated on the pump, it must
be configured by CODAN ARGUSservice utility. The parameter
“Enable infusion set” must be checked in order to perform a
calibration of the infusion set. See Section „3.6.3.
Configuration of infusion sets“.

WARNING
• Unless otherwise specified by the customer, the ARGUS 71x V
will be calibrated by the manufacturer with the CODAN
standard set. Other infusion sets have to be calibrated
separately.
• Only infusion sets being validated and approved by
CODAN  ARGUS  AG shall be used. For an up-to-date list
please contact CODAN ARGUS AG.
• After configuration of a new infusion set, pressure and
volume calibration are mandatory.
• A pressure calibration is necessary when a pressure sensor or
the main board was replaced before.
• If the pressure control measurement is not accurate, a
pressure calibration is recommended.

5.3 Technical Errors


A technical error is specified on the display via a numerical code
(example: «18082»). The second line contains information such
as firmware version, module number, module revision and line
number. This message is accompanied by a periodic audible
signal, a permanent red status light and a flashing alarm light. In
case of a technical error, the device needs to be analysed by the
technical service. The failure status will be stored in the history
log. Information about this technical error can be found with the
history tool of CODAN ARGUSservice. For further assistance
generate the report file and contact CODAN ARGUS AG.

Procedure in the event of a technical error message


• Press On/Off key, to mute the audible signal.
• Press On/Off key, to switch off the pump.
• The pump must be disconnected from the patient immediately
• and replaced with a substitute device.
Pass the defective pump to the Technical Service Department.

Procedure in emergencies, when interruption of the
therapy poses a risk to the patient
• Press On/Off key, to mute the audible signal.
• Press On/Off key, to switch off the pump.
• Press On/Off key again, the pump is now in normal mode
again, until a new technical error occurs.

14.544_D_SM_en_A71XV_5.09  35 / 68
5. Maintenance

5.4 Coin Cell Battery


The lithium ion coin cell battery (Type: CR2032, REF: 601 603) is
used to store the states and history of the device. When the
battery is replaced, this information is lost. It is recommended
that the information is saved using CODAN ARGUSservice before
the battery is replaced. For step-by-step instructions on how to
replace coin cell, see Section „6.8.6. Replacement of the Coin
Cell“.

During normal usage, the button cell battery will last over
10 years. However, if the ARGUS 71x V is not used for an extended
periods of time while being disconnected from mains power, the
battery may be empty within a shorter time span.

 WARNING
Repairs shall only be performed by trained specialists who are
authorized in maintenance and repair of the ARGUS 71x V by
CODAN ARGUS AG. Repairs shall only be performed with original
spare parts which must not be modified without prior
authorization from CODAN ARGUS AG. Not following these
instructions can compromise patient safety.

5.5 Configuration Mode

Entering the Configuration Mode:


Keep the 2nd numeric key depressed and press the ON/OFF key
in order to get access to the configuration mode.

The configuration contains various device-specific settings. These


settings make it possible to customize the functionality of the
ARGUS 71x V. There are two sets of this configuration:

Current Configuration: The active device configuration.


Customer Configuration: The default configuration that can be customized by order. The Order and Configuration Form is
available at CODAN ARGUS AG. The customer configuration can be restored with CODAN
ARGUSservice.

Changes made in configuration mode are stored permanently


into current configuration.

 WARNING
• Changes in the configuration mode (with the CODAN
ARGUSservice PC utility tool), preventive maintenance
measurements or repair constitute a modification of the
device and may only be carried out by authorized persons.
Otherwise, patient safety cannot be guaranteed.
• After each change of a configuration, a function check has
to be performed.
• Reducing the volume also affects the volume of the audible
alarm.

An active connection is displayed on the screen.

36 / 68 14.544_D_SM_en_A71XV_5.09
5. Maintenance

5.5.1 Configuration Parameters


This section contains an overview of all relevant settings accessible
with the parameter tool of CODAN ARGUSservice (see Section
„3.9. Parameter Control“).

Alarm Setting Default Value Range/Unit Description

Air Accumulation Detection True False/True Detection of accumulated air bubbles


Duration 32 8 - 64 min Time window for detected air accumulation. Volume
of air accumulation (see infusion set configurations).
Alarm Signal Pause no. 5 350 350 -1300 ms Duration of pause no. 5 (See Figure 34).
Interburst 2500 2500 - 15000 ms Interburst interval (See Figure 34).
Pause 50 50 - 125 ms Duration of pause (See Figure 34).
Pulse 75 75 - 200 ms Pulse length (See Figure 34).
Pre-alarm Signal Interburst 15000 15000 - Alarm low priority interburst interval (See Figure 35).
30000 ms
Pause 125 125 - 250 ms Duration of pause (See Figure 35).

Pulse 125 125 - 250 ms Pulse length (See Figure 35).

Note Signal Pause 750 500 - 2000 ms Duration of pause (See Figure 36).
Pulse 750 500 - 2000 ms Pulse length (See Figure 36).
Nurse Call Function False False/True Allows alarm transmissions
Pause width 2 0 - 3600 s Pause width
Pulse width 1 0 - 3600 s Pulse width
Standby Duration 2 1 - 60 min Time-out for unused device until alarm release
One-time False False/True Alarm release only before the first start of infusion
Standby alarm
Figure 33:  Auditory Signal for an Alarm

14.544_D_SM_en_A71XV_5.09  37 / 68
5. Maintenance

Auditory Signal for an Alarm:


A: Pulse length
B: Duration of pause
C: Duration of 3rd and 8th pause (calculated: A + 2B)
D: Duration of pause no.5
E: Interburst interval

Figure 34:  Auditory Signal for an Alarm

Auditory Signal for a Pre-Alarm:


A: Pulse length
B: Duration of pause
C: Alarm low priority interburst interval

Figure 35:  Auditory Signal for Pre-Alarm

Auditory Signal for a Note:


A: Pulse length
B: Duration of pause

Figure 36:  Auditory Signal for a Note

38 / 68 14.544_D_SM_en_A71XV_5.09
5. Maintenance

Operation Conditions Setting Default Range/Unit Description


Value
Buzzer Volume setting 10 1 - 10 Buzzer loudness

Sound at start of True False/True Buzzer signal release at start of infusion


infusion
General Automatic bolus False False/True Option in addition to the manual bolus
function
Fluid balancing False False/True Reset of the infused volume while VTBI remains
unaffected
infusion set menu True False/True Selection and confirmation of an infusion set brand

Drug selection False False/True User is forced to select a drug before infusion start
required (drug library must be available)
Rate modification False False/True User is forced to stop infusion before a rate change
in Stop Mode
Set duration 0 0 - 1000 h Notification time if infusion set is inserted too long
exceeded time (0 = disabled).
OFF key delayed 3 0-3s Adjustments of the response time

OFF key in Stop True False/True Device must be in Stop Mode to be switched off
Mode only
STOP key delayed 0 0 - 3000 ms Adjustments of the response time

Fallback time to 15 5 - 120 sec Time to fall back within main screen
infusion rate
Transport Mode 15 1 - 60 min Interval time for recurrence of reminders (Transport
reminder Mode must be enabled)
Speaker Volume setting 10 1 - 10 volume Speaker alarm loudness

Delayed Infusion Blue Leds False False/True Blue leds are enabled for indicating delayed
infusion mode
Barcode Mode Patient barcode False False/True Patient barcode required
required
Staff barcode False False/True Staff barcode required
required
Staff barcode False False/True Staff barcode required for rate change (Parameter
required for rate 'Staff barcode required' must also be enabled,
change otherwise no action)

 WARNING
When setting the volume of the buzzer or speaker, ensure that
the loudness is higher than the ambient noise level, otherwise
the operator might not recognize the occurrence of an alarm
condition. This could lead to a delayed reaction to an alarm
condition.

14.544_D_SM_en_A71XV_5.09  39 / 68
5. Maintenance

KVO Setting Default Range/Unit Description


Value
General KVO after VTBI False False/True KVO is active when VTBI (Volume to be infused)
was reached was reached
KVO at stop True False/True KVO is active whenever an infusion is stopped
(alarm conditions excluded)
Setting for high 3000 100 - 10000 KVO rate for infusion rates above 40 ml/h
infusion rates µl/h
Setting for low 100 100 - 5000 µl/h KVO rate for infusion rates up to 40 ml/h
infusion rates

Pre-Alarms Setting Default Range/Unit Description


Value
General Clear-time for low 2 1 - 240 min Time until pre-alarm is repeated after its last
battery clearance
Clear-time Near 2 1 - 240 min Time until pre-alarm is repeated after its last
End of Infusion clearance
Near End of False False/True Announcement that the VTBI will be reached soon
Infusion
Time for Near 10 1 - 240 min Time of the pre-alarm release
End of Infusion

Automatic pressure Setting Default Range/Unit Description


reduction Value
General Automatic True False/True Automatic activation of pressure reduction through
pressure reduction occlusion
at downstream

Service Intervals Setting Default Range/Unit Description


Value
General Interval setting 0 0 - 24 months Time period up to the reminder SAFETY CHECK IS
in months DUE
Interval setting in 0 0 - 10000 hrs Time period up to the reminder SAFETY CHECK IS
hours of operation DUE

40 / 68 14.544_D_SM_en_A71XV_5.09
5. Maintenance

Infusion set 1-4 Setting Default Range/Unit Description


Value
General Size of individual 250 50 - 1000 µl Air bubble size at which an air alarm is released
air bubble
Air accumulation 250 100 - 2000 µl Volume of accumulated air bubbles at which
volume an air alarm is released (see Alarm Settings: Air
Accumulation
Enable infusion set True False/True Enable infusion set

Bolus volume 10000 1000 - 999000 µl Limit of the bolus volume being administered
per bolus
Name of infusion CODAN text Name to be displayed on the pump
set STANDARD
Level setting for 700 100 - 1000 mbar Pressure level at which a downstream alarm is
occlusion alarm released

User Permissions Setting Default Range/Unit Description


Value
Definition 1 Name Administrator The user name

Pin code 1220 0 - 9999 Pin code of the user

Display Setting Default Range/Unit Description


Value
General Brightness 240 0 - 255 steps Display backlight brightness

Contrast 7 1 - 15 steps Display contrast

14.544_D_SM_en_A71XV_5.09  41 / 68
5. Maintenance

Menu Setting Default Range/Unit Description


Value
General Alarm Presets True False/True This menu shows the current alarm presets

Battery Info True False/True Access for user to check remaining battery capacity

Bolus function True False/True Access for user to bolus function

Key Lock False False/True Access for user to lock the keyboard by means of a pin

Key lock pin code 0 0 - 9999 4 digit pin code

Pressure Setting True False/True Access for user to modify temporarily the occlusion level

Automatic Purge False False/True Access for user to purge IV tubing automatically

Settings True False/True Access for user to modify temporarily buzzer volume,
standby alarm
Fallback time from 60 5 - 60 sec Time to fall back from menu to main screen
menu operation
Transport Mode Access for user to the transport mode
False/True
Pressure unit mbar enum mbar, mmhg, kpa, cmh20, psi, Pressure unit to be
settings displayed
Info Menu Accumulated False False/True Shows accumulated volume info menu
Volume
Special Functions Barcode Mode False False/True Shows Barcode Mode Menu

Delayed Infusion False False/True Shows Delay Infusion Menu

Start-up Setting Default Range/Unit Description


Value
Splash Screen Text Department text Department name to be displayed at the bottom of the
splash screen

42 / 68 14.544_D_SM_en_A71XV_5.09
5. Maintenance

5.5.2 Nurse Call


The ARGUS 71x V can be adapted to the nurse call system. The
connection can be found on the rear side of the device (see
Section “Overview”). A connection plan of the RJ9 plug is shown
in Figure 45.

During an alarm or pre-alarm the nurse call becomes activated


(see in table below). After muting the alarm or pre-alarm the
nurse call relay goes back to the normal contact.

The kind of the function can be configurated in 3 ways (see


Section „5.5.1. Configuration Parameters“):

Signal form Setting Value


Single pulse alarm Nurse Call Function True
Nurse Call Pause Width 3600 s
Nurse Call Pulse Width Pulse duration
Pulse sequence alarm (with Nurse Call Function True
Reminder) Nurse Call Pause Width Reminder time
Nurse Call Pulse Width Pulse duration
Static alarm Nurse Call Function True
Nurse Call Pause Width 0s
Nurse Call Pulse Width 3600 s
Nurse Call disabled Nurse Call Function False
Nurse Call Pause Width Any
Nurse Call Pulse Width Any

14.544_D_SM_en_A71XV_5.09  43 / 68
6.  Replacement of Parts

6.1 Exploded View ARGUS 71x V - complete

Figure 37:  Exploded view ARGUS 71x V - complete

44 / 68 14.544_D_SM_en_A71XV_5.09
6. Replacement of Parts

6.2 Exploded View ARGUS 71x V -Casing backside inside

Figure 38:  Exploded View ARGUS 71x V - Casing backside inside

14.544_D_SM_en_A71XV_5.09  45 / 68
6. Replacement of Parts

6.3 Exploded View ARGUS 71x V - Casing backside outside

Figure 39:  Exploded View ARGUS 71x V - Casing backside outside

46 / 68 14.544_D_SM_en_A71XV_5.09
6. Replacement of Parts

6.4 Exploded View ARGUS 71x V - Casing front inside

Figure 40:  Exploded View ARGUS 71x V - Casing front inside

14.544_D_SM_en_A71XV_5.09  47 / 68
6. Replacement of Parts

6.5 Exploded view of ARGUS 71x V - Casing front outside

Figure 41:  Exploded View ARGUS 71x V - Casing front outside

48 / 68 14.544_D_SM_en_A71XV_5.09
6. Replacement of Parts

6.6 Part Numbers and Descriptions

No. Order No. Spare Part


1 601281 Casing front complete ARGUS 717 V
1 601485 Casing front complete ARGUS 718 V
2 601279 Keypad (ARGUS 717 V)
2 601340 Keypad (ARGUS 718 V)
3 601238 Window global alarm left (upstream side)
4 601239 Window global alarm right (downstream side)
5 600960 Door (ARGUS 717 V)
5 601049 Door (ARGUS 718 V)
6 601231 Door handle (with dowel pin, securing screw)
7 601438 Main board with LCD (ARGUS 717 V)
7 601439 Main board with LCD (ARGUS 718 V)
8 600658 Lithium coin CR1632.IB 3V /125mAh (coin cell for mainboard)
9 600976 Air bubble detector
10 601188 Pressure sensor Z1.0
11 601057 Anti-free flow clamp (ARGUS 718 V)
11 600966 Stop flow clamp (ARGUS 717 V)
12 601491 Buzzer (>90dB RM15-D22,6mm)
13 601278 Sensor board
14 601343 Ribbon cable 6-pol (60mm ready-made)
15 600965 Stepper motor
16 600963 Stepper motor cable
17 601275 Pump unit (peristaltic block)
18 600604 Membrane for peristaltic NBR (000-70A-0099 (57618) )
19 601389 Casing backside 115V (with transformer, mains receptacle and edge board)
19 601274 Casing backside 230V (with transformer, mains receptacle and edge board)
20 601282 Edge board
21 600478 Edge board protective insert against drops
22 601277 Power board
23 601103 Transformer 115 V (with cable, plug)
23 601062 Transformer 230 V (with cable, plug)
24 600486 Mains receptacle
25 600699 Non-slip rubber foot (ø10x3 black)
26 601342 Ribbon cable 24-pol (120 mm ready-made)
27 601074 Battery Ni-Mh / 12V / 1800 mAh (with temperature control)
28 601283 Socket head cap screw M4x12 (ZS-M4x12-IK-V-DIN 912)
29 601284 Socket head cap screw M4x70 (ZS-M4x70-IK-V-DIN 912)
30 600491 Edge board hood
31 600233 Socket head cap screw M5x20 (ZS-M5x20-IK-A2-DIN 7984)
32 601167 Screw cover
33 600484 Fuse T125mAL250V (230VAC mains power supply)
33 600485 Fuse T250mAL250V (115VAC mains power supply)
34 600978 Combination clamp

14.544_D_SM_en_A71XV_5.09  49 / 68
6. Replacement of Parts

6.7 Technical Error List


Code Number Probable defective hardware or Code Number Probable defective hardware or
firmware firmware
8011 - 8014 1. MB 8210 - 8213 1. MB
8015 - 8026 1. BP, 2. MB, 3. PB 18001 - 18007 1. MB, 2. PB
8027 - 8038 1. MB 18008 1. MB
8039 - 8040 1. PB 18009 - 18011 1. MB, 2. PB
8041 - 8046 1. MB 18012 1. PB
8047 - 8053 1. PB 18019 1. ML
8054 - 8073 1. MB 18020 1. UPS
8074 - 8082 1. MB, 2. PB 18021 1. DPS
8088 - 8090 1. MB 18031 1. PB
8094 - 8096 1. MB, 2. PB, 3. BIB 18034 - 18035 1. PB, 2. T
8105 - 8109 1. BP, 2. MB, 3. PB 18036 - 18037 1. MB, 2. PB, 3. T
8111 1. FW, 2. MB 18038 - 18039 1. PB
8112 - 8114 1. H 18040 - 18043 1. MB
8115 1. PB 18044 - 18045 1. BP, 2. PB
8116 - 8117 1. MB 18046 - 18047 1. PB
8119 1. MB 18048 - 18049 1. MB
8122 - 8125 1. MB 18050 1. MV, 2. PB, 3. T
8126, 8130 1. SM, 2. MB 18051 - 18052 1. MB
8132 - 8134 1. MB 18053 - 18055 1. BP, 2. PB
8135 1. MB, 2. PB 18056 - 18057 1. MB
8137 - 8141 1. MB, 2. PB, 3. SB 18058 - 18059 1. MB
8143 - 8144 1. MB, 2. PB, 3. SB 18060 - 18063 1. UPS, 2. MB
8174 1. KP 18064 1. IB
8177 1. MB 18065 1. MB
8182 1. UPS, 2. MB 18066 1. MB
8183 1. DPS, 2. MB 18067 1. PB
8184 1. ML 18068 1. SB
8186 1. MB 18069 - 18070 1. AT, 2. MB
8199 1. BIB 18071 - 18072 1. AT, 2. BP
8201 - 8204 1. BIB 18074 - 18076 1. PB
8207 1. PB all others 1. FW

Abbreviations:
AT ambient temperature MB main board
BIB bluetooth interface board ML medication library
BP battery pack MV mains voltage
DPS downstream pressure sensor PB power board
FW firmware SB sensor board
H description in the history file SM stepper motor
IB interface board T transformer
KP keypad UPS upstream pressure sensor

If the error is located in the firmware, check if new firmware is


available. Some technical errors give a hint in the history which
hardware component is defective. In the table above, the possible
components are in the order of probability as error source. If the
defective component can be found, first check the connections of
the component. If the error occurs again, replace the components
after order of the probability and check if the error occurs again. If
no action has success, inform the CODAN ARGUS technical services.

50 / 68 14.544_D_SM_en_A71XV_5.09
6. Replacement of Parts

6.8 Assembly Instructions Notice


• The battery pack of the ARGUS 71x V is equipped with a
 WARNING temperature and current sensors for an accelerated charging
• The ARGUS 71x V may only be used with original accessories time with high current-flow. Therefore, only battery packs
and spare parts being approved by CODAN ARGUS AG. being provided by CODAN ARGUS AG shall be used.
• Repairs shall only be performed by authorized specialists who • The charging time depends on the present intensity of use
have been trained in maintenance and repair of the and the status of the pump (e.g. run or stop mode, high or
ARGUS 71x V infusion pumps. low infusion rate, etc.).
• Disconnect the device from mains and remove all interface
connections prior to repairs. 6.8.4 Disassembling of the Main Board
• The internal mains connectors below the transformer are not 1. Perform steps in Section „6.8.1. Disassembling of the
insulated. Housing“
• A complete recalibration is mandatory after: 2. Take the casing front side.
1. Calibration of a new set 3. Unplug all cables from pressure sensors, air detector, sensor
2. Replacement of a pressure sensor board, power board, keypad and motor cable.
3. Replacement of the peristaltic block 4. Remove all 4 screws on the main board.
4. Replacement of the housing 5. Withdraw the main board.
5. Replacement of the main board
6. Replacement of battery pack 6.8.5 Assembling of the Main Board
• Remove the battery connector (X15) prior to replacing any 1. Insert the new main board (#7).
parts. 2. Fix the mainboard with 4 screws and tighten them with a
• Always completely re-assemble the ARGUS 71x V before torque of 0.5 Nm.
attaching it to mains power again. 3. Plug in all connectors.
4. Perform steps in Section „6.8.2. Assembling of the Housing“
CAUTION
Ensure that proper ESD precautions are met before opening the 6.8.6 Replacement of the Coin Cell
device to avoid damaging any electronic components. Notice
Beware of the information written in Section „5.4. Coin Cell
References to the exploded view drawings are in the following Battery“.
format: (#x).
1. Perform steps in Section „6.8.4. Disassembling of the Main
6.8.1 Disassembling of the Housing Board“
1. Remove the screw covers. 2. Remove the coin cell by lifting it with an insulated object. Be
2. Remove the screws on the back side of the housing. aware of the risk of local short circuits.
3. Separate the casing front side from the casing backside. 3. Any standard CR1632 coin cell can be used (#8).
4. Unplug the ribbon cable 24-pol from main and power board 4. Perform steps in Section „6.8.5. Assembling of the Main
Board“
6.8.2 Assembling of the Housing 5. Turn the pump on (in regular user mode) in order to re-initialize
1. Merge the casing front (#1) with the casing backside (#19) the internal real time clock.
together. 6. Switch the pump off again.
2. Plug in the ribbon cable 24-pol (#26) from main and power 7. Use the CODAN ARGUSservice utility (Section „3.8.1. How to
board set Date and Time“) to set the clock.
3. Fix the screws (2pcs #28 and 2pcs #29) in the casing backside
and tighten them with a torque of 1.1 Nm. 6.8.7 Replacement of the Keypad
4. Place the screw covers (4pcs #32). 1. Perform steps in Section „6.8.4. Disassembling of the Main
5. A final recalibration of pressure and volume must be Board“
performed. 2. Push the window in the keypad from behind the front casing.
3. Clean the surface of the housing (e.g. isopropyl alcohol),
6.8.3 Replacement of Battery Pack where to keypad was located.
1. Perform steps in Section „6.8.1. Disassembling of the 4. Remove the protective film of the new keypad (#2)
Housing“ 5. Insert the flex cable from the keypad through the provided
2. Take the casing back side. opening and bring the keypad in the right position in the
3. Unplug the battery pack. front casing.
4. Insert a new battery pack (#27). 6. Press the whole keypad surface on the front casing, so that
5. Perform steps in Section „6.8.2. Assembling of the Housing“ the keypad is glued well.
6. Update the time on the device with CODAN ARGUSservice 7. Perform steps in Section „6.8.5. Assembling of the Main
(Section „3.8.1. How to set Date and Time“). Board“

14.544_D_SM_en_A71XV_5.09  51 / 68
6. Replacement of Parts

6.8.8 Replacement of Stop Flow Clamp 6.8.13 Replacement of Interface Board Bluetooth
ARGUS 717 V 1. Perform steps in Section „6.8.1. Disassembling of the
1. Perform steps in Section „6.8.1. Disassembling of the Housing“.
Housing“ 2. Take the casing back side.
2. Take the casing front side. 3. Unplug the cable of the Interface Board from the Mainboard
3. Remove the 2 screws. 4. Remove the Interface Board.
4. Remove the clamp. 5. Insert new Interface Board (601551).
5. Insert new clamp (#11). 6. Plug in cable from the Interface Board.
6. Fix the clamp with the 2 screws and tighten them with a 7. Perform steps in Section „6.8.2. Assembling of the Housing“.
torque of 0.6 Nm
7. Perform steps in Section „6.8.2. Assembling of the Housing“. Notice
• it is only possible to use a Bluetooth Interface Board with
6.8.9 Replacement of Anti-Free Flow Clamp ARGUS Mainboard same or newer than index M in article number
718 V 601276.M (located on label, near plug for keypad).
1. Perform steps in Section „6.8.1. Disassembling of the
Housing“ 6.8.14 Disassembling of Door
2. Take the casing front side. 1. Open the door.
3. Unplug the cable from the clamp. 2. Gently press the hinge stopper backwards.
4. Remove the 2 screws. 3. Pull the shaft towards the centre with help of needle nose
5. Remove the clamp. pliers until the shaft is out of the holder.
6. Insert new clamp (#11). 4. Repeat step (2) and (3) for the other side of the door.
7. Fix the clamp with the 2 screws and tighten them with a 5. Remove the Door.
torque of 0.5 Nm
8. Plug in the cable from the clamp. 6.8.15 Assembling of Door
9. Perform steps in Section „6.8.2. Assembling of the Housing“. 1. Insert Door (#5).
2. Push the two shafts with needle nose pliers into the holders.
6.8.10 Replacement of Pressure Sensor 3. Close the door.
1. Perform steps in Section „6.8.1. Disassembling of the
Housing“. 6.8.16 Disassembling of Pump Unit
2. Take the casing front side. 1. Perform steps in Section „6.8.1. Disassembling of the
3. Unplug the cable from the downstream or/and the upstream Housing“.
Pressure Sensor. 2. Perform steps in Section „6.8.14. Disassembling of Door“.
4. Remove the screw of the selected Pressure Sensors. 3. Unplug all cable from mainboard, sensor board and motor.
5. Insert new Pressure Sensor (#10). 4. Remove the 4 screws on the front casing located above and
6. Fix the Pressure Sensor with the screw and tighten them with below.
a torque of 0.3 Nm. 5. Be careful with the flat cable of the display when withdrawing
7. Perform steps in Section „6.8.2. Assembling of the Housing“. the pump unit.

6.8.11 Replacement of Air Detector 6.8.17 Assembling of Pump Unit


1. Perform steps in Section „6.8.1. Disassembling of the 1. Insert a pump unit (#17) and ensure that the flat cable is not
Housing“. Take the casing front side. damaged.
2. Unplug the cable from the Air Detector. 2. Fix the pump unit with 4 screws and tighten them with a
3. Remove the 2 screws and Air Detector. torque of 1.0 Nm.
4. Insert new Air Detector (#9). 3. Plug in the ribbon cable 6-pol (#14) from sensor board .and
5. Fix the Air Detector with the 2 screws and tighten them with the stepper motor cable (#16) into mainboard.
a torque of 0.6 Nm. 4. Perform steps in Section „6.8.15. Assembling of Door“.
6. Perform steps in Section „6.8.2. Assembling of the Housing“. 5. Perform steps in Section „6.8.2. Assembling of the Housing“.

6.8.12 Replacement of Window Global Alarm 6.8..18 Replacement of Peristaltic Membrane NBR
1. Perform steps in Section „6.8.1. Disassembling of the 1. Perform steps in Section „6.8.16. Disassembling of Pump
Housing“. Unit“.
2. Take the casing front side. 2. Change Membrane NBR (#18).
3. Push with a thin pointed tool from behind into the holder 3. Perform steps in Section „6.8.17. Assembling of Pump Unit“.
hole of the Window Global Alarm.
4. Remove the Window Global Alarm.
5. Insert new Window Global Alarm (#3 and #4).
6. Perform steps in Section „6.8.2. Assembling of the Housing“.

52 / 68 14.544_D_SM_en_A71XV_5.09
6. Replacement of Parts

6.8.19 Disassembling of Sensor Board 6.8.25 Assembling of Power Board


1. Perform steps in Section „6.8.16. Disassembling of Pump 1. Insert a new power board (#22).
Unit“. 2. Fix the power board with 7 screws and tighten them with a
2. Remove the 2 screws from the Sensor Board. torque of 0.6 Nm.
3. Separate the Sensor Board. 3. Plug in all connectors.
4. Perform steps in Section „6.8.23. Assembling of Transformer
6.8.20 Assembling of Sensor Board and Mains Receptacle“.
1. Insert new Sensor Board (#13).
2. Fix the 2 screws on sensor board and tighten them with a 6.8.26 Replacement of Edge Board
torque of 0.4 Nm. 1. Perform steps in Section „6.8.1. Disassembling of the
3. Perform steps in Section „6.8.17. Assembling of Pump Unit“. Housing“.
2. Take casing back side.
6.8.21 Replacement of Stepper Motor 3. Remove the screw witch fix the edge board with the clamp
1. Perform steps in Section„6.8.19. Disassembling of Sensor holder.
Board“. 4. Unplug the edge board cable from the power board.
2. Remove the screw nut, which stabilizes the step motor. 5. Remove the edge board.
3. Remove the 2 spacer bolts on the motor metal plate. 6. Insert new edge board (20).
4. Separate the motor metal plate. 7. Fix the edge board with the screw on the clamp holder and
5. Remove the 4 screws which fixes the axle on the pump unit. tighten them with a torque of 0.5 Nm.
6. Separate the Stepper Motor from pump unit. 8. Perform steps in Section „6.8.2. Assembling of the Housing“.
7. Insert new Stepper Motor (#15).
8. Fix the 4 screws which fixes the axle on the pump unit and 6.8.27 Replacement of Fuse
tighten them with a torque of 0.6 Nm. 1. Use a screw driver to open the fuse holder of the mains
9. Insert the motor metal plate. receptacle.
10. Fix the 2 spacer bolts on the motor metal plate. 2. Change the fuse (#33).
11. Fix the screw nut, which stabilizes the step motor. 3. Close the fuse holder again.
12. Perform steps in Section „6.8.20. Assembling of Sensor
Board“. 6.8.28 Replacement of the Combination Clamp
1. Remove the M5x20 screw of the combination clamp.
6.8.22 Disassembling of Transformer and Mains 2. Fix the new combination clamp (#34) with the M5x20 screw
Receptacle (#31) on the clamp holder and tighten them with a torque of
1. Perform steps in Section „6.8.1. Disassembling of the 1.5 Nm.
Housing“.
2. Take casing back side. 6.8.29 Replacement of the Rubber Feet
3. Plug out the transformer cable from the power board. 1. Remove the rubber feet.
4. Remove both screws of the transformer. 2. Clean surface (e.g. isopropyl alcohol).
5. Remove both screws of the mains receptacle. 3. Glue the new rubber feet (#25) on the bottom of casing.
6. Remove the transformer and mains receptacle.

6.8.23 Assembling of Transformer and Mains


Receptacle
1. Insert the transformer (#23) and mains receptacle (#24).
2. Fix the transformer with two screws and tighten them with a
torque of 1.0 Nm.
3. Fix the mains receptacle with two screws and tighten them
with a torque of 0.5 Nm.
4. Plug in the transformer cable with Power Board.
5. Perform steps in Section „6.8.2. Assembling of the Housing“.

6.8.24 Disassembling of Power Board


1. Perform steps in Section„6.8.22. Disassembling of Transformer
and Mains Receptacle“.
2. Separate all connectors of edge board, battery pack and main
board.
3. Remove all 7 screws of the power board.
4. Withdraw the power board.

14.544_D_SM_en_A71XV_5.09  53 / 68
7.  Isolation of the Patient

Isolation of the Patient


No. Values Description
1 2 MOPP Transformer with isolation voltage of 4kV, medical approved, EN 61558-2-6
2 2 MOPP No electrical connection to housing (see material datasheets)
3 2 MOPP Air sensor is the most critical component to infusion set (conductive material on plates). infusion set wall thickness
of 0.5mm (0.1mm => 1kV)
4 0 MOPP Patients can be connected to conductive medication in infusion set

Figure 42:  Isolation of the Patient

54 / 68 14.544_D_SM_en_A71XV_5.09
8.  Isolation of the User

Isolation of the User


No. Values Description
5 2 MOPP Transformer with isolation voltage of 4kV, medical approved, EN 61558-2-6
6 2 MOPP No electrical connection to housing (see material datasheets)
7 2 MOPP Power cable standard isolation
8 1 MOPP Isolation with relay (1.5kV)
9 2 MOPP Double isolation (each 1.5kV)

Figure 43:  Isolation of the User

14.544_D_SM_en_A71XV_5.09  55 / 68
9.  Wiring Diagram

Figure 44:  Wiring Diagram ARGUS 71x V

Figure 45:  Connection Plan Nurse Call

56 / 68 14.544_D_SM_en_A71XV_5.09
10.  Block Diagram

Figure 46:  Block Diagram ARGUS 71x V

14.544_D_SM_en_A71XV_5.09  57 / 68
11.  Safety Standard Check (SSC)

11.1 Reminder “Safety Check is Due”


When a safety standard check is due, a message appears after
each startup procedure of the pump. Follow the steps in the SSC
check list.

The reminder is triggered under the following conditions:
• Serial number of the device is missing
• Service interval (months or operating hours) has expired, if
not configured to zero
• Time and date are not set
• Lithium coin cell battery has low voltage, replacement is
necessary

Notice
The exact reason for the reminder “Safety Check is Due” may be
found in the history. Look for Safety Standard Check (SSC) in the
history.

11.2 Volume Control Measurement


Use the same volume calibration equipment as described in
Section „5.2. Set Calibration“.
1. Switch the pump on. The pump must be in the regular
infusion mode.
2. Prepare the balance for measurement and insert the infusion
set in the pump.
3. Close the front door and open the roller clamp.
4. Verify that the inserted infusion set corresponds with the
infusion set displayed in the pump screen.
5. Purge the extension line.
6. Set the balance to zero by pressing TARA.
7. Define a therapy with a rate of 250ml/h and a total volume
of 30ml.
8. Start the infusion.
9. Wait until the END INFUSION alarm.
10. The resulting net weight shall read 30 g ±5% (corresponds
to 28.5ml – 31.5ml).
11. If the weight is out of limit, the set calibration has to be
repeated.

11.3 Pressure Control Measurement


Use the same pressure calibration equipment as described in
Section „5.2. Set Calibration“.
1. Switch the pump on; the pump is in regular infusion mode.
2. Insert the infusion set and start the infusion at a rate of
200ml/h.
3. Close the 3-way stop cock on patient’s side to simulate an
occlusion.
4. As soon as the manometer reaches 700 mbar, the pump must
stop and release an occlusion alarm (tolerance is +/-150
mbar).
5. Make sure the upstream sensor is activated (see Section
„5.1.1. Pump Settings“).
6. Pump Settings are configurable on the device.
7. Start an infusion without drop detector, at a rate of 200 ml/h
and a VTBI of 10 ml.
8. Simulate an occlusion by closing the roller clamp on the side
of the medication container.
9. The pump must stop after several seconds releasing an
upstream occlusion alarm.

58 / 68 14.544_D_SM_en_A71XV_5.09
11. Safety Standard Check (SSC)

WARNING
The pressure calibration procedure always includes all of the
following steps and is complete when the control measurement
was successful. The previous values remain stored in a premature
termination of the calibration.

11.4 Functional Air Bubble Detection Test


Configuration Prerequisites:
Setting Value
Infusion set/ General/Size of individual air bubble 250 µl
Infusion set/ General/Air accumulation volume 250 µl
Alarm/ Air Accumulation/Detection True
Alarm/ Air Accumulation/Duration 32 min

1. Switch the pump in configuration mode and activate the


upstream sensor in the pump settings.
2. Switch off and on again, so that the pump is in regular
infusion mode.
3. Insert the infusion set in the pump.
4. Close the front door and open the roller clamp.
5. Verify that the inserted infusion set corresponds with the
infusion set displayed in the pump screen.
6. Purge the extension line.
7. Start a therapy with a rate of 250ml/h and a total volume of
30ml.
8. Incline the drop chamber of the infusion set until the pump
intakes air inside a segment of minimum 3,5 cm from the
infusion line.
9. Change the rate to 1000ml/h.
10. Verify that after a few seconds the air alarm becomes
displayed.
11. Open the door and move the infusion set until the air is
outside of the pump.
12. Start the therapy again.
13. Verify that no alarm occurs.

11.5 Functional Drop Detection Test


1. Switch the pump on; the pump is in regular infusion mode
with drop detector.
2. Insert the infusion set in the pump.
3. Close the front door and open the roller clamp.
4. Clamp the drop detector to the drop chamber (see section
“4.3 Preparing the pump” in user manual ARGUS 71x V)
5. Purge the extension line.
6. Start a therapy with a rate of 250ml/h.
7. Verify that the drop light flashes green each time a drop falls.
8. Unclamp the drop detector from the drop chamber.
9. Verify that after a few seconds the drop occlusion alarm
becomes displayed.

14.544_D_SM_en_A71XV_5.09  59 / 68
11. Safety Standard Check (SSC)

11.6 Battery Pack Test


The battery pack of the ARGUS 71x V is equipped with a
temperature and current sensors for an accelerated charging time
with high current-flow. Therefore, only battery packs being
provided by CODAN ARGUS AG shall be used.
It is not permitted to reuse or to modify the built-in safety
electronics. The charging time depends on the present intensity
of use and the status of the pump (e.g. run or stop mode, high
or low infusion rate, etc.). The replacement of a battery pack may
only be performed by authorized specialist who were trained by
CODAN ARGUS AG.

11.6.1 Check with CODAN ARGUSservice Utility


The following steps are to be carried out to check the battery
pack with CODAN ARGUSservice:
1. Charge the battery until a capacity of 100% (see Section
„3.6.4.6. Battery Information“).
2. Discharge the full battery pack by running the pump at a rate
of 25ml/h for at least 6 hours.
3. Disconnect the pump from the mains supply.
4. Wait until the battery alarm is released.
5. Open the history tool of the CODAN ARGUSservice utility (see
Section „3.7. History Control“).
6. Read out the last time stamps from the alarm event “Indication
of alarm: (A12) END OF BATTERY CONNECT PUMP TO
MAINS!” and notification event “Power source = Battery“.
7. The elapsed time is the difference between this two time
stamps.
8. If the battery pack does not reach the required capacity (see
Item Battery operation in Section „14. Product Specifications“),
the battery pack should be replaced (see Section „6.8.3.
Replacement of Battery Pack“).
9. Connect the pump to mains for 7 hours to charge the battery
pack.

11.6.2 Check without CODAN ARGUSservice Utility


The following steps are to be carried out to check the battery
pack without CODAN ARGUSservice:
1. Charge the battery until a capacity of 100% (see Battery Info
menu).
2. Discharge the full battery pack by running the pump at a rate
of 25ml/h for at least 6 hours.
3. Disconnect the pump from the mains supply and start a
stopwatch. Figure 47:  Battery Info Menu
4. Wait until the battery alarm is released and stop the
stopwatch.
5. Read out the elapsed time.
6. If the battery pack does not reach the required capacity (see
Item Battery operation in Section „14. Product Specifications“),
the battery pack should be replaced (see Section „6.8.3.
Replacement of Battery Pack“).
7. Connect the pump to mains for 7 hours to charge the battery
pack

60 / 68 14.544_D_SM_en_A71XV_5.09
11. Safety Standard Check (SSC)

11.7 Nurse Call Test


1. S witch the pump on; the pump is in regular infusion mode.
2. Insert the infusion set in the pump.
3. Close the front door and open the roller clamp.
4. Verify that the inserted infusion set corresponds with the
infusion set displayed in the pump screen.
5. Purge the extension line.
6. Start a therapy with a rate of 250ml/h and a total volume of
30ml.
7. Open the door.
8. Verify that after the door open alarm occurs, the relay
switches the contact on the RJ9 Plug. In this case the contact
switches from Pin2/Pin3 to Pin2/Pin4 and back (Signal is
depending of the configuration).

14.544_D_SM_en_A71XV_5.09  61 / 68
11. Safety Standard Check (SSC)

Device identification :
ARGUS 717 V ARGUS 718 V
Serial-No.: Hospital:
Pump-No.: Department:
Used infusion set: Customer:
The SSC has to be performed at least every 24 months or after 10’000 hours of operation.
The check has to be done in accordance to the user and service manuals.

Test steps for ARGUS 717 V and ARGUS 718 V


1 Visual check for damage, cleanness and completeness. Housing, labels, accessories, connectors, power cable, etc.
2 Switch pump on in configuration mode and note the Firmware release:
SW-Rel. on the start-up screen and compare the right device
type. Is the device type identical to the pump?
3 Check if a firmware update is required. Latest firmware is available on CODAN ARGUS AG website
(www.codanargus.com)
4 Perform display, keypad, LED and backlight tests (see Section Was all 3 tests successful?
„5.1.4. Keypad & Display Test“).
5 Switch pump off and start the pump in normal mode. Check Acoustic signal is audible?
of acoustic signal and indication of red status LED during The red status LED lights up?
start-up procedure.
6 Connect/disconnect the pump to/from the mains. The mains LED lights up only when connected to mains?
7 Air bubble test (see Section „11.4. Functional Air Bubble Detection Test“):
Start therapy with a rate of 250.0 ml/h and a total of 30ml. Air bubble alarm is displayed?
Simulate air in the infusion line.
Confirm the alarm and remove the air in the infusion line. No more air bubble alarm occurs.
Start therapy again.
8 Drop detector test (see Section „11.5. Functional Drop Detection Test“):
Start a therapy with a rate of 250ml/h. Remove drop Drop occlusion alarm is displayed.
detector from infusion set during therapy.
9 Interfaces tests:
Check the external connector “Nurse Call“ (see Section Relay contact switches
„11.7. Nurse Call Test“).
Check the interface to docking station (connect pump to The indicator for mains operation must light green
docking station). The green status LED on the pump lights up.
10 Pressure sensor tests (see Section „11.3. Pressure Control Measurement“):
Downstream sensor: Measured
Control measurement with pressure limit of 700mbar. pressure limit:
Upstream Sensor: An acoustic alarm will go off after a few seconds.
Kink infusion set on bottle side.
11 Volume accuracy test (see Section „11.2. Volume Control Measurement“):
Start a therapy with KVO disabled, rate 250ml/h and total Measured volume:
30ml.
12 Battery pack test:
Charge battery of running pump for at least 9 hrs. Mains power displayed
Battery Check at a rate of 35ml/h. Battery symbol is displayed during test
Run the battery test until battery alarm goes off
13 Inspection certificate: Visual check of mains connector
Standard: IEC 62353 measurements attached

62 / 68 14.544_D_SM_en_A71XV_5.09
11. Safety Standard Check (SSC)

Additional test steps for ARGUS 717 V


14 Test the function of the stop flow clamp Proper function
15 Recharge the battery after this test

Additional test steps for ARGUS 718 V


14 Test the function of the anti free flow clamp Proper function
15 Recharge the battery after this test

The pump has passed the safety standard check and is safe for use.
Date: Name: Signature:

14.544_D_SM_en_A71XV_5.09  63 / 68
12.  Repair Order Form

Purchase Order
Proforma Invoice Number:
Customer Name:

Address:

Contact Person:

Tel. Number:

Device Identification :
ARGUS 717 V ARGUS 707 V ARGUS 600 S ARGUS 404
ARGUS 718 V ARGUS 708 V ARGUS 606 S ARGUS 414
ARGUS 300 P ARGUS 500 P ARGUS 600 P ARGUS 60 P ARGUS 100 P
ARGUS 300 M ARGUS 500 M ARGUS 600 M ARGUS 60 M ARGUS 100 M
Accessory:

Serial-No.: Production Code:

Detailed failure or problem description:

Excepted work / repair to be done:


Repair Description:
Warranty repair
Replacement
Other

Date: Name: Signature:

64 / 68 14.544_D_SM_en_A71XV_5.09
13.  Care of the pump

13.1 Important notes


Disconnect the pump from the mains before cleaning!

Remove all devices and connecting cables.

CAUTION
• Avoid any liquid penetrating into the device or the device
plug. If liquid is spilt over the device, the latter must be
immediately disconnected from the mains plug or removed
from the docking station. Then the pump must be dried
immediately and thoroughly cleaned.
• Warnings relating to the liquids (medications) must be
observed for cleaning.
• If there is any likelihood of liquid having penetrated inside the
device, the device must be inspected by the Technical Service
Department before continuing in use.

Notice
• The pump and the accessories must be kept clean and dry. In
order to maintain the full functionality, regular cleaning
within the product specifications is recommended (see
Section „14. Product Specifications“).
• Do not use any abrasive cleaning agents.
• The pump is unsuitable for sterilization in autoclaves and
must not be immersed in liquids.

13.2 Cleaning and disinfection


Clean the device only by wiping with a damp cloth. The use of
lukewarm water is normally adequate. Care must be taken that
the pump‘s connections are clean and dry so that no electrical
damage occurs when connecting to a docking station or the
mains.
For disinfection, only agents containing diluted alcohol (isopropyl)
may be used. The in-house specialist department for hygiene can
give information on suitable disinfectants.

13.3 Storage and transport


Devices may only be stored in clean conditions at a cool and dry
place in accordance with Section „14. Product Specifications“

Notice
• The battery must be fully charged after 3 months at the latest
so that full capacity is preserved. Otherwise batteries can
over-discharge and leak, at worst.
• For transport, make sure that you choose appropriate
packaging with sufficient protection against impact effects!
We recommend using the original packaging.

14.544_D_SM_en_A71XV_5.09  65 / 68
14.  Product Specifications

Technical Data Software 5.0X Remarks


Performance Data
Applications intensive care, anesthesia, standard care, oncology, blood dose calculation, TIVA
transfusion, parenteral nutrition, neonatal care
Dosing units g, mg, μg, ng, IU, mmol per hour, minute, kilogram, day
Flow rate 0.1 – 1‘200 ml/h steps of 0.1 ml/h, modification without infusion interruption
Infusion volume 0.1 – 999.9 ml / 1‘000 – 9‘999 ml steps of 0.1 ml / steps of 1.0 ml
Infusion time 1 min – 99 h 59 min steps of 1 min
Programming rate in ml/h, volume in ml, duration in h and min
Automatic rate calculation on the basis of volume and time rate in ml/h, volume in ml, infusion duration in h and min
Bolus administration manually or automatically with / without dose limit, limits via ARGUSmedDB
Bolus rate 0.1 – 1‘200 ml/h without infusion interruption
Bolus volume 0.1 – 999 ml without infusion interruption
Purging of the infusion line 0.1 – 1‘200 ml/h
Accuracy ±5%, depending on the used, supported infusion set under normal conditions according to IEC/EN 60601-2-24,
technical deviation < ±1%
KVO (Keep Vein Open) rate 0.1 – 10.0 ml/h depending on flow rate, deactivation possible
Standby mode 1 – 60 min

History and Interface


History and event log up to 1‘000 events with time stamp
Nurse call connection to the central system 24 V / 0.2 A, configurable pulse, potential-free
RS-232 interface electrically isolated, connectable to PDMS (Patient Data
Management System)

Operation
Display type monochrome LCD, 98.4 x 26.2 mm, 240 x 64 pixel illuminated, adjustable brightness
Display contents flow rate in units and ml/h, infused volume, target volume
(VTBI), infusion time, remaining infusion time, battery capacity,
pressure level, medication name and setup, bolus delivery and
information menu etc.
Menu overview intuitive and configurable for specific applications display text in national language, overview of infusion data
Menu configuration by means of ARGUSservice software / temporary settings by client-specific configuration possible
operator
Infusion lines CODAN standard- and cyto-sets and other validated sets tube materials: PVC, PVC-free, NoDEHP / display of the set
name
ARGUS 718 V: CODAN IVP dedicated set
©
with anti-freeflow-clamp

Safety
Pressure release automatic pressure release after occlusion
Air bubble detection individual air bubbles configurable from 50 – 1‘000 µl
accumulated air bubbles configurable from 100 – 2‘000 µl within 8 – 64 min
Drop detector 10 – 65 drops/ml in transport mode alarm is less sensitive
Alarms, pre-alarms visual and audible signals for rapid detection with text explanation and symbols, large alarm window
Alarm types occlusion (up- and downstream), air bubbles, no drops and
too many drops, end of infusion, standby, low battery, end of
battery, maintenance due etc.
Key lock activation / deactivation via numerical code
Pressure limit for occlusion 100 – 1‘000 mbar (also in mmHg, kPa, cmH2O, Psi) 10 configurable steps
DERS: Medication database on request on request
(Dose Error Reduction System)

Power Supply
Battery 1’800 mAh, NiMH maintenance-free
Battery operation 6 hours at 25 ml/h
Battery charging time 9 hours in standby mode
Mains power supply 230 VAC ±10%, 50 – 60 Hz option: 115 VAC ±10%, 50 – 60 Hz
Power consumption 12 VA maximum

Product Specification
Dimensions 190 x 160 x 130 mm (W x L x D) without combination clamp
Housing ASA high performance plastic
Operating-/ medication temperature 5 – 40°C / 18 – 30°C
Temperature for storage and transport 0 – 40°C
Relative humidity 20 – 90% without condensation
Weight with battery 2.0 kg including battery and power supply unit
Maintenance safety standard check (SSC): every 24 months or after 10‘000 operating hours

Compliance
IP protection class IP22 drop-protected in horizontal position up to ±15° inclination
Protection class II electric shock protection according to IEC 61140
Applied part type CF according to EN 60601-1
MDD classification IIb
Electrical safety EN 60601-1, EN 60601-1-4, EN 60601-2-24
Electromagnetic compatibility EN 60601-1-2
Certification CODAN ARGUS AG ISO 13485, ISO 9001
Device CE mark CE 0123 according to directive 93/42/EEC

Accessories
Docking stations ARGUS docking stations in 3 sizes for different combinations version M: monitoring (PDMS link) + power supply
version P : only power supply
Attachment system combination clamp suitable for pole- and rail systems
Software ARGUSservice program for simple maintenance and configuration

66 / 68 14.544_D_SM_en_A71XV_5.09
14.544_D_SM_en_A71XV_5.09  67 / 68
14.544_D_SM_en_A71XV_5.09

CODAN ARGUS AG
Oberneuhofstrasse 10
CH-6340 Baar
Phone +41 41 785 09 44
Fax +41 41 785 09 40
[email protected], www.codanargus.com

You might also like